<Topics Covered in Chapter 8>

• Cardiac Output Fick Technique Method Indicator Dilution Thermodilution• Electromagnetic Flowmeter (Not good)• Ultrasound Flowmeters• Plethysmography

Medical Instrumentation II

Chap 8. Part I. Flow and Volume of Blood

Page 3: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

1. Indicator Dilution Method

• Fick Method• Indicator Dilution using

rapid injection• Thermodilution

Medical Instrumentation II

Page 4: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

iF oF FFF oi )(densityCi )(densityCo

dt

dm(Infusion rate) -> We Known

FCdt

dmFC oi (Mass balance

equation)

min/min// gllg

io CCdt

dmflowF )(Conclusio

n :

Page 5: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Fick Method

Medical Instrumentation II

dt

dm

Substitution of Equation!

ao CC

vi CC

(O₂ density of Artery Blood)

(O₂ density of Vein Blood)

(O₂ Consumption, L/min)measure By Spirometer

Page 6: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Indicator Dilution using rapid injection

Medical Instrumentation II

Dye Dilution MethodIndocyanine green (cardiogreen)

Extrapolation

Indictor dilution curve

)(tC

tBolus injection

Recirculation

Page 7: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

Find Area!

)(tCdv

dm dvtCdm )(

FtCdt

dvtCdvtC

dt

d

dt

dm )()()(

TT

dttCFdmm00

)(

T

dttC

mF

0)(

Conclusion :

Page 8: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

Weak Point : Once dye get an injection, it remains in our body.

The dye should be excreted in the urine. In order

to remeasure the blood flow, you should wait for a long time.

Page 9: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Thermodiltion (standard)

Medical Instrumentation II

T

tBolus injection

Weak Point : We unknown temperature of Right artium.

Using 4 lumen catheter(Swan-Ganz catheter)

Page 10: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

]/[)(

3

0

smdttTC

QF T

bbb

Conclusion :

][JCTVQ iiii (Heat content of injectate)

]/[ 3mkgb

]/[ KkgJCb

(Blood density)

(Specific heat of blood)

Page 11: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

-> pressure element, approvaled by 1MHz frequency.Ex)

t

No reflection

Primary Echo

skin

2. Ultra Sound

Page 12: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

2

f

v

MHzf

3

Near field

Far field

D

2.1sin

Page 13: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Transit Time Flowmeter

Medical Instrumentation II

Du

Velocity of RBC, WRC, and plate.

c Velocity of ultrasound.

Page 14: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

i ) Down stream ( = Blood flow direction)

ii ) Up stream ( = Blood flow reverse direction)

cosuc

DtD

coscos uc

D

uc

Dttt DU

cosuc

DtU

2222

cos2

cos

cos2

C

Du

uc

Du

Because… C >> ucosφ

Conclusion :

Page 15: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Continuous Doppler Flowmeter

Medical Instrumentation II

u

c

u

f

f

o

d

cff d

cfo

Doppler frequency shift

Ex) aviation, ship, submarine…

Page 16: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

Pulsed Doppler Flowmeter

D

1u2u

3u2u

1u

12321 uuuuu

Flow profile measuremen

t

Page 17: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Thermal Convection Flowmeter

Medical Instrumentation II

Thermister

f

2T

1T

i ) Bias current -> Thermister heatingii ) T₂Thermister is cooled by thermal convection

Weak Point : Invasive and Probe position is difficult.

Page 18: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

-> Volume, blood flow measurement

V

`i Howland Current

Source

Cuff

Artery openVein close

mmHg60

3. Plethysmography

Page 19: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

t

Z)(v

Venous Thrombosis

NormalVolume

Cuff On Cuff Out

Slope = flow

Arterial

Page 20: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Patient monitoring (Respiration monitoring)

Part II. Pulmonary Function Test

Medical Instrumentation II

1. ArchitectureTrachea

Bronchus

Alveoli

Area TT Au ,

AA Au ,

AATT AuAuQ

A

TTA A

Auu

TA uu Because… O₂ & CO₂ diffusion at alveoli

Page 21: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

2. Physiological Functions

• O₂ transport into blood stream• CO₂ removal from blood

stream• Homeostasis of pH, Po₂, Pco₂

Page 22: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

3. Respiration

• Ventilatory mechamis - RC equivalent circuit model

• Gas transport - Mass balance equation(Diffusion

process)

Page 23: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

4. Pressure measurementDifferential pressure transducer

1P 2P

Diaphragm

i )

ii )1P 2P

Diaphragm

2C1C

1x 2x

Page 24: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

iii )

Mesh screen

1P 2P

Differential pressure sensor

Q

Flow

Mouth

RQPP 21

R

PPQ 21 Conclusio

n :

Page 25: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Airpump

Medical Instrumentation II

Ballon sensorHole Ballon

Gas

Pressure sensor

Pv5

MPiv

Control

Pressure

Metal switch

Switch On ->Switch Off ->

v0v5

iPP iPP

Page 26: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

5. Flow measurement

• Rotating Vane• Ultrasound Flowmeter• Thermal Convection Flowmeter• Pneumotachometer (standard)

Page 27: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

6. Volume measurement-> Plethysmography

t

TLC

FRC

RV

TV (Tidal Volume)

ERC

IC VC

Weak Point : It is Relative value, not absolute value.

Page 28: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

Water Spirometer

Water

Air

Pen

Rotating Paper

Mouth

CO₂ absorber

Page 29: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

Dry Spirometer

Air (gas)

Mouth

Inspiratory gas

Expiratory gas

Valve

Piston

Gas is Separated by two section

Page 30: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

7. FRC measurementHe dilution technique - patient at FRC

- prepare a spirometer of volume VSP and

initial He, concentration of fIHE

- patient breaths tooth spirometer - find He concentration of expiration F

SPFHeFHeIHe VFFFFRC

Page 31: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

N₂ washout technique - patient at FRC - prepare a spirometer of O₂, VSP

- patient breaths to the spirometer - Measure N₂ concentration in (FFN₂)

the expiratory.

2)(8.0 FNSP FVFRCFRC

Page 32: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

8. Gas concentration measurement

• Mass spetroscophy• Gas chromatography• Infrared spectrography• Raman spectrography• Emission spectrography• Paremagnetic O₂ analyzer

Page 33: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

9. Gas transport measurement

Gas distribution test(Single breath N₂ wachout) - RC -> TLC O₂ inspiration - Pause (hold breath) - Expiration and measure fN2

Page 34: 생체계측 II Report #9 2005200427 송성진 Medical Instrumentation II

Medical Instrumentation II

Diffusion test(CO diffusion capacity) - Inspire ω, He in air - fICO, fIHe

- Hold breath for 10sec - Expire to measure FFCO, FFHe

He

He

CO

CO

I

F

F

I

B

ACO f

F

F

f

Pt

vD ln

)47(

60