a. bouty; s. broussard; a. lotode; y. mariette; ch. veyssiere: cea/dsm/irfu/sis
DESCRIPTION
DSS tests in LEGNARO 22 nd to 25 th April 2008. A. BOUTY; S. BROUSSARD; A. LOTODE; Y. MARIETTE; Ch. VEYSSIERE: CEA/DSM/IRFU/SIS D. BAZZACCO: Univ. Padova E. FARNEA; A. GADEA; R. MENEGAZZO: LNL P. JONES: Univ. Jyväskylä T. ENGERT; T. HABERMANN; I. KOJOUHAROV: GSI - PowerPoint PPT PresentationTRANSCRIPT
09/07/2008 1AGATA – Ch. VEYSSIERE
A. BOUTY; S. BROUSSARD; A. LOTODE; Y. MARIETTE; Ch. VEYSSIERE: CEA/DSM/IRFU/SIS
D. BAZZACCO: Univ. Padova
E. FARNEA; A. GADEA; R. MENEGAZZO: LNL
P. JONES: Univ. Jyväskylä
T. ENGERT; T. HABERMANN; I. KOJOUHAROV: GSI
J. CACITTI; M. TRIPON: GANIL
DSS tests in LEGNARO
22nd to 25th April 2008
09/07/2008 2AGATA – Ch. VEYSSIERE
Goal of these tests
Realized with the detector S001
•To test the AXIS low voltage crate:
– 48V
– ±6V; ±12V for the preamps
– 6.5V for the HV Box (Proto #1)
•To test the HV box in an other environment (GSI (2006) -> LNL)
•To Test the last version of the GUI developed by GSI
09/07/2008 3AGATA – Ch. VEYSSIERE
The “AXIS” Box
09/07/2008 4AGATA – Ch. VEYSSIERE
The AGATA Box
09/07/2008 5AGATA – Ch. VEYSSIERE
S001 – HV Box (proto #1)
09/07/2008 6AGATA – Ch. VEYSSIERE
Poor Man’s Digitiser
09/07/2008 7AGATA – Ch. VEYSSIERE
Means for the tests
•Comparatives tests:
– Patrice Medina 48V / AXIS
– ±6V; ±12V NIM/ AXIS
– 6.5V AXIS + HV Saclay/ Silena module
Following analysis from Dino Bazzacco: PStestsWeek17-1.doc
09/07/2008 8AGATA – Ch. VEYSSIERE
Mechanical mounting of the LV crate
Reference spectrum at high and low gain
AXIS crate mounted (just grounded)
EHG= 2.52 keV at 1332 keV ELG = 2.75 keV at 1332 keV
09/07/2008 9AGATA – Ch. VEYSSIERE
Mechanical mounting of the PLC
PLC mounted (outside of the cabinet) and switched on
Replacement of the 24V PS of the PLC by a linear floating PS; 6V and 12V AXIS on
EHG= 2.57 keV at 1332 keV ELG = 2.63 keV at 1332 keV
09/07/2008 10AGATA – Ch. VEYSSIERE
Test of LV from AXIS (6V, 12V) versus NIM
Final frequency spectra (100 k sample) before shutting down the HVRed: HG LV from AxisGreen: HG LV from NIM binBlue: LG LV from AxisCyan: LG LV from NIM bin
The spectra are very similar, with the ones using Axis slightly better at the lowest frequencies and those using NIM having lower amplitude in the peaks at 23 and 46 kHz.
09/07/2008 11AGATA – Ch. VEYSSIERE
Test of the 48V (1/2)
With the Patrice 48V
With the AXIS 48V
09/07/2008 12AGATA – Ch. VEYSSIERE
Test of the 48V (2/2)
48V AXIS with improvement on the grounding (copper braid shortened)
With the Patrice 48V
Green and blue when adding a 400 and 600W resp.
09/07/2008 13AGATA – Ch. VEYSSIERE
Using a GPBox (dual core preamp. inside)
Red: 6, 12 and 48 V Axis
Green: LV NIM, 48V Patrice
09/07/2008 14AGATA – Ch. VEYSSIERE
Comparison of AGATA and Silena HV
•Using AXIS LV•Energy resolution (in keV) derived from INP
Unfiltered Filteredwith 3 s
Filteredwith 6 s
HV Silenahigh gain
15.2 1.75 1.43
HV AGATA
high gain
15.4 1.66 1.44
HV Silena low gain
18.6 1.82 1.59
HV AGATA
low Gain
18.8 1.74 1.62
09/07/2008 15AGATA – Ch. VEYSSIERE
Noise spectra after 6 s shaping trapezoidal shaping
Red HV Silena (high gain signal)Green HV AGATA “” Blue HV Silena (low gain signal)Cyan HV AGATA “”
09/07/2008 16AGATA – Ch. VEYSSIERE
Integrated power spectra after ~6 s trapezoidal shaper
Red HV Silena (high gain signal)Green HV AGATA “” Blue HV Silena (low gain signal)Cyan HV AGATA “”
keV
09/07/2008 17AGATA – Ch. VEYSSIERE
Problems to be solved:
•Voltage regulators associated to long cables generate oscillations: regulators to be changed,
•Temperature too high in the 48V module:
– Geometry to be modified to improve the air path on the heat sink,
– Vicor 600W PS to be tested. To check its influence on the signal,
– Water cooling?
•When profibus is reconnected, the crate stops briefly,
•Safety of the 6 and 12V should not be linked with the 6.5V (when 6 and 12V stop, HV stops also),
•48V offset to be reduced,
•Connectors (front panel Sub-D, 24V connectors, Profibus-DP)
09/07/2008 18AGATA – Ch. VEYSSIERE
Some problems can be avoided…
•Reconnecting the PA when voltage is set induces an inrush current that trigs the safety of the 6 and 12V: not to be done… Removing the safety on the overcurrent might be dangerous in some cases.
July tests of revised AXIS box at Legnaro – on platform
-Production Digitiser-Detector (Single detector ?)-Integration with rest of DSS ongoing
09/07/2008 19AGATA – Ch. VEYSSIERE