abt-ba-e-0911

Upload: dexterpoli

Post on 03-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 ABT-BA-e-0911

    1/71

    KERN & Sohn GmbHZiegelei 1

    D-72336 Baling en

    E-Mai l : [email protected]

    Tel: +49-[0]7433- 9933-0

    Fax: +49-[0] 7433-9933-149

    Internet: w ww.kern-sohn.com

    Operating instructionAnalytical balance

    KERN ABTVersion 1.112/2009GB

    #

    ABT-BA-e-0911

  • 7/28/2019 ABT-BA-e-0911

    2/71

    ABT-BA-e-09112

    GBKERN ABTVersion 1.1 12/2009

    Operating instruction

    Analytical balance

    Table of Contents

    1 Technical data................................................................................................ 52 Declaration of conformity ............................................................................. 83 Keyboard and display overview ................................................................... 93.1 Keyboard overview.....................................................................................................93.2 Overview of display..................................................................................................114 Basic Information (General)........................................................................ 124.1 Proper use.................................................................................................................124.2 Improper Use ............................................................................................................124.3 Warranty....................................................................................................................124.4 Monitoring of Test Resources.................................................................................135 Basic Safety Precautions............................................................................ 135.1 Pay attention to the instructions in the Operation Manual...................................135.2 Personnel training....................................................................................................136 Transport and storage................................................................................. 136.1

    Testing upon acceptance ........................................................................................13

    6.2 Packaging.................................................................................................................. 137 Unpacking, Setup and Commissioning ..................................................... 147.1 Installation Site, Location of Use ............................................................................147.2 Unpacking .................................................................................................................147.2.1 Scope of delivery ...................................................................................................................... 157.2.2 Installation ................................................................................................................................ 167.3 Mains connection .....................................................................................................177.4 Connection of peripheral devices...........................................................................177.5 Initial Commissioning ..............................................................................................187.5.1 Turning On the Power .............................................................................................................. 188 Adjustment................................................................................................... 198.1 Automatic adjustment via PSC ...............................................................................198.1.1 PCS function ON/OFF.............................................................................................................. 208.2 Automatic adjustment via Clock-CAL ....................................................................218.2.1 Setting the time for Clock-CAL ................................................................................................. 228.3 Adjustment via preset process ...............................................................................238.3.1 Selecting default adjustment method....................................................................................... 238.3.2 Default setting: Adjustment with internal weight....................................................................... 248.3.3 Default setting: Adjustment test with internal weight................................................................ 248.3.4 Default setting: Adjustment test with external weight............................................................... 258.4 Carrying out alternative adjustment methods.......................................................268.4.1 Adjustment with internal weight................................................................................................ 268.4.2

    Adjustment test with internal weight......................................................................................... 27

    8.4.3 Adjustment test with external weight........................................................................................ 28

  • 7/28/2019 ABT-BA-e-0911

    3/71

    ABT-BA-e-0911 3

    9 Verification ................................................................................................... 2910 ISO/GLP log.................................................................................................. 3010.1 Setting the log printout of your adjustment data ..................................................3010.2 Balance ID Number Setting .....................................................................................3110.3 Setting the date printout ..........................................................................................3210.3.1 Printout Date and Time, without weight value ..................................................................... 3211 Basic Operation ........................................................................................... 3311.1 Weighing ...................................................................................................................3311.2 Taring......................................................................................................................... 3311.3 Changing the display ...............................................................................................3411.4 Changing readability ................................................................................................3511.5 Underfloor weighing.................................................................................................3612 The menu...................................................................................................... 3712.1

    Changing settings ....................................................................................................37

    12.2 How to change settings ...........................................................................................3712.3 Call up menu.............................................................................................................3812.4 Selecting the menu...................................................................................................3912.5 Menu overview..........................................................................................................4012.6 Menu Lock.................................................................................................................4312.7 Resetting the menu ..................................................................................................4312.8 Settings control display...........................................................................................4413 Description of individual functions............................................................ 4513.1 Stability filter.............................................................................................................4513.1.1 Standard mode..................................................................................................................... 4513.1.2 Anti-convection mode........................................................................................................... 4613.1.3 High-stability Mode............................................................................................................... 4613.1.4 Apportioning mode............................................................................................................... 4713.2 Standstill display ......................................................................................................4813.3 Auto Zero...................................................................................................................4913.4 Setting date ...............................................................................................................5013.5 Setting time...............................................................................................................5113.6 Capacity display .......................................................................................................5214 Application Functions ................................................................................. 5314.1 Parts counting ..........................................................................................................5314.2 Percent determination..............................................................................................5414.3 Density determination..............................................................................................5514.3.1 Density determination of solids ............................................................................................ 5514.3.2 Determining density of liquids.............................................................................................. 5614.4 Auto Print ..................................................................................................................5814.5 Add up mode.............................................................................................................5914.6 Recipe mode .............................................................................................................61

  • 7/28/2019 ABT-BA-e-0911

    4/71

    ABT-BA-e-09114

    15 Data output................................................................................................... 6315.1 RS 232C interface.....................................................................................................6315.2 Data Formats.............................................................................................................6315.3

    Remote control instructions....................................................................................64

    15.4 Standard Settings.....................................................................................................6515.5 User Settings ............................................................................................................6615.5.1 Communication speed settings............................................................................................ 6715.5.2 Delimiter settings.................................................................................................................. 6715.5.3 Parity settings....................................................................................................................... 6715.5.4 Stop bit settings.................................................................................................................... 6815.5.5 Input-output data format settings ......................................................................................... 6815.5.6 Handshake settings ............................................................................................................. 6816 Service, maintenance, disposal.................................................................. 6916.1 Cleaning ....................................................................................................................6916.2 Service, maintenance...............................................................................................6916.3 Disposal..................................................................................................................... 6917 Instant help................................................................................................... 70

  • 7/28/2019 ABT-BA-e-0911

    5/71

    ABT-BA-e-0911 5

    1 Technical data

    KERN ABT 100-5M ABT 120-4M

    Readability (d) 0.01 mg 0.1 mg

    Weighing range (max) 101 g 120 g

    Minimum load (Min) 1 mg 10 mg

    Verification value (e) 1 mg 1 mg

    Verification class I I

    Reproducibility 0.05 mg 0.1 mg

    Linearity 0.15 mg 0.2 mg

    Stabilization time 10 sec 3 sec

    Adjustment weight internal

    Weighing units(calibrated appliances)

    g, ct

    Smallest part weight forpiece counting

    1 mg 1 mg

    Reference quantities atpiece counting

    10, 20, 50,100

    Weighing plate, stainlesssteel 80mm

    Dimensions of the hous-ing (B x D x H)

    217 x 356 x 338

    DimensionsGlass wind screen [mm]

    Weighing room 168 x172 x223

    Net weight (kg) 7

    Permitted environmentalcondition

    +10 C to +30 C

    Humidity of air max. 80 % relative (not condensing)

    Voltage 230 V/50Hz

  • 7/28/2019 ABT-BA-e-0911

    6/71

    ABT-BA-e-09116

    KERN ABT 220-4M ABT 320-4M

    Readability (d) 0.1 mg 0.1 mg

    Weighing range (max) 220 g 320 g

    Minimum load (Min) 10 mg 10 mg

    Verification value (e) 1 mg 1 mg

    Verification class I I

    Reproducibility 0.1 mg 0.1 mg

    Linearity 0.2 mg 0.2 mg

    Stabilization time 3 sec

    Adjustment weight internal

    Weighing units(calibrated appliances)

    g, ct

    Smallest part weight forpiece counting

    1 mg 0,1 mg

    Reference quantities atpiece counting

    10, 20, 50,100

    Weighing plate, stainlesssteel 80mm

    Dimensions of the hous-

    ing (B x D x H)

    217 x 356 x 338

    DimensionsGlass wind screen [mm]

    Weighing room 168 x172 x223

    Net weight (kg) 7

    Permitted environmentalcondition

    +10 C to +30 C

    Humidity of air max. 80 % relative (not condensing)

    Voltage 230 V/50Hz

  • 7/28/2019 ABT-BA-e-0911

    7/71

    ABT-BA-e-0911 7

    KERN ABT 120-5DM ABT 220-5DM

    Readability (d) 0.01/0.1 mg 0.01/0.1 mg

    Weighing range (max) 42 g/120 g 82 g/220 g

    Minimum load (Min) 1 mg 1 mg

    Verification value (e) 1 mg 1 mg

    Verification class I I

    Reproducibility 0.02/ 0.1 mg 0.05/ 0.1 mg

    Linearity 0.05/0.2 mg 0.1/0.2 mg

    Stabilization time 3sec./10 sec.

    Adjustment weight internal

    Weighing units(calibrated appliances)

    g, ct

    Smallest part weight forpiece counting

    1 mg

    Reference quantities atpiece counting

    10, 20, 50,100

    Weighing plate, stainlesssteel 80mm

    Dimensions of the hous-

    ing (B x D x H)

    217 x 356 x 338

    DimensionsGlass wind screen [mm]

    Weighing room 168 x172 x223

    Net weight (kg) 7

    Permitted environmentalcondition

    +10 C to +30 C

    Humidity of air max. 80 % relative (not condensing)

    Voltage 230 V/50Hz

  • 7/28/2019 ABT-BA-e-0911

    8/71

    ABT-BA-e-09118

    2 Declaration of conformity

    KERN & Sohn GmbHD-72322 Balingen-Frommern

    Postfach 4052

    E-Mail: [email protected]

    Tel: 0049-[0]7433- 9933-0

    Fax: 0049-[0]7433-9933-149Internet: www.kern-sohn.de

    Declaration of conformity

    Declaration of conformity for apparatus with CE markKonformittserklrung fr Gerte mit CE-Zeichen

    Dclaration de conformit pour appareils portant la marque CEDeclaracin de conformidad para aparatos con marca CE

    Dichiarazione di conformit per apparecchi contrassegnati con la marcatura CE

    English We hereby declare that the product to which this declaration refers conforms with the fol-lowing standards.

    Deutsch Wir erklren hiermit, dass das Produkt, auf das sich diese Erklrung bezieht, mit den nach-stehenden Normen bereinstimmt.

    Franais Nous dclarons avec cela responsabilit que le produit, auquel se rapporte la prsentedclaration, est conforme aux normes cites ci-aprs.

    Espaol Manifestamos en la presente que el producto al que se refiere esta declaracin est deacuerdo con las normas siguientes

    Italiano Dichiariamo con ci che il prodotto al quale la presente dichiarazione si riferisce confor-

    me alle norme di seguito citate.

    Electronic Balance: KERN ABT

    Mark applied EU Directive Standards

    2004/108/EC EMC EN45501EN55022

    2006/95/EC Low Voltage EN60950

    Date: 30.11.2009 Signature:

    Gottl. KERN & Sohn GmbH

    Management

    Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149

  • 7/28/2019 ABT-BA-e-0911

    9/71

    ABT-BA-e-0911 9

    3 Keyboard and display overview

    3.1 Keyboard overview

    In weighing mode:

    KeyDesigna-

    tion

    Pressed once and releasedKeep pressed for about 3 seconds

    [ON/OFF]Switches between the opera-

    tion and standby modes.Switches the key notification buzzer on/off.

    [CAL] Invokes adjustment or menuselection.

    Invokes adjustment or menu selection.

    [TARE] Tare or set weight display to zero

    [UNIT] Changing the display

    [PRINT]

    Issue of weighing value to

    external appliances (printer,

    PC)

    Issue of date and time to external appliances.

    ABT 120-4M

    ABT 220-4M

    ABT 320-4M[1d/10d]

    ABT 100-5M

    ABT 120-5DM

    ABT 220-5DM

    For verified applications cutting off the last

    digit is only possible for models with a readout

    of 0,01 mg.

    For all other models the button 1/10d is inop-

    erable.

  • 7/28/2019 ABT-BA-e-0911

    10/71

    ABT-BA-e-091110

    In menu:

    KeyDesigna-

    tionPressed once and released

    Keep pressed for about 3 sec-

    onds

    [ON/OFF] Back to menu Return to weighing mode

    [CAL] Menu selection

    [TARE]Parameter selection

    Storing settings

    [UNIT] Entering numeric values Increases the numeric value of flashing digit by 1.

    [PRINT] Entering numeric values. Shifts flashing digit.

    [1d/10d] No effect.

  • 7/28/2019 ABT-BA-e-0911

    11/71

    ABT-BA-e-0911 11

    3.2 Overview of display

    Display Designation Description

    Standstill display

    Indicates that the weighed value is stable. Highlights current set-

    ting during menu item selection

    Weight symbol

    Appears during adjustment. Flashes prior to start of automatic

    adjustment.

    Appears during parameter selection for adjustment. Blinks to ad-

    vise necessity of adjustment.

    Number symbol Indicates numeric value entry.

    Menu symbolAppears during menu selection. Always shown when the menu is

    locked.

    Asterisk Indicates that the displayed numeric value is not a mass value.

    Add-on symbol Indicates set-up of Add-on mode.

    Communication

    symbol

    Is lit up during communication with external appliances via RS-

    232C cable. Shown when communication functions are ON.

    Battery symbolIndicates a low battery voltage when using the balance with the

    optional battery pack.

    Auto Print symbol Indicates set-up of the Auto Print function.

    Standby mark Appears during standby status of balance.

    Inverse triangle sym-

    bol

    Illuminates as part of the solid specific gravity measurement dis-

    play.

    Capacity displayUnits display

  • 7/28/2019 ABT-BA-e-0911

    12/71

    ABT-BA-e-091112

    4 Basic Information (General)

    4.1 Proper use

    The balance you purchased is intended to determine the weighing value of material

    to be weighed. It is intended to be used as a non-automatic balance, i.e. the mate-rial to be weighed is manually and carefully placed in the centre of the weighing plate.As soon as a stable weighing value is reached the weighing value can be read.

    4.2 Improper Use

    Do not use balance for dynamic add-on weighing procedures, if small amounts ofgoods to be weighed are removed or added. The stability compensation installed inthe balance may result in displaying an incorrect measuring value! (Example: Slowlydraining fluids from a container on the balance.)Do not leave permanent load on the weighing plate. This may damage the measuring

    system.Impacts and overloading exceeding the stated maximum load (max) of the balance,minus a possibly existing tare load, must be strictly avoided. Balance may be dam-age by this.Never operate balance in explosive environment. The serial version is not explosionprotected.The structure of the balance may not be modified. This may lead to incorrect weigh-ing results, safety-related faults and destruction of the balance.The balance may only be used according to the described conditions. Other areas ofuse must be released by KERN in writing.

    4.3 Warranty

    Warranty claims shall be voided in case

    Our conditions in the operation manual are ignored

    The appliance is used outside the described uses

    The appliance is modified or opened

    Mechanical damage or damage by media, liquids, natural wear and tear

    The appliance is improperly set up or incorrectly electrically connected

    The measuring system is overloaded

  • 7/28/2019 ABT-BA-e-0911

    13/71

    ABT-BA-e-0911 13

    4.4 Monitoring of Test Resources

    In the framework of quality assurance the measuring-related properties of the bal-ance and, if applicable, the testing weight, must be checked regularly. The responsi-ble user must define a suitable interval as well as type and scope of this test. Infor-mation is available on KERNs home page (www.kern-sohn.com with regard to themonitoring of balance test substances and the test weights required for this. InKERNs accredited DKD calibration laboratory test weights and balances may becalibrated (return to the national standard) fast and at moderate cost.

    5 Basic Safety Precautions

    5.1 Pay attention to the instructions in the Operation Manual

    Carefully read this operation manual before setup and commissioning, even if youare already familiar with KERN balances.

    5.2 Personnel training

    The appliance may only be operated and maintained by trained personnel.

    6 Transport and storage

    6.1 Testing upon acceptance

    When receiving the appliance, please check packaging immediately, and the appli-ance itself when unpacking for possible visible damage.

    6.2 Packaging

    Keep all parts of the original packaging in case you need to return the appliance.Only use original packaging for returning.Before sending, disconnect all connected cables and loose/movable parts.Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate,power unit etc., to prevent slipping and damage.

    http://www.kern-sohn.com/http://www.kern-sohn.com/
  • 7/28/2019 ABT-BA-e-0911

    14/71

    ABT-BA-e-091114

    7 Unpacking, Setup and Commissioning

    7.1 Installation Site, Location of Use

    The balances are designed in a way that reliable weighing results are achieved in

    common conditions of use.You will work accurately and fast, if you select the right location for your balance.

    Therefore, observe the following for the installation site:

    Place the balance on a firm, level surface;

    Avoid extreme heat as well as temperature fluctuation caused by installingnext to a radiator or in the direct sunlight;

    Protect the balance against direct draughts due to open windows and doors;

    Avoid jarring during weighing;

    Protect the balance against high humidity, vapours and dust;

    Do not expose the device to extreme dampness for longer periods of time.Non-permitted condensation (condensation of air humidity on the appliance)may occur if a cold appliance is taken to a considerably warmer environment.In this case, acclimatize the disconnected appliance for ca. 2 hours at roomtemperature.

    Avoid static charge of goods to be weighed or weighing container.

    If electro-magnetic fields or static charge occur, or if the power supply is unstablemajor deviations on the display (incorrect weighing results) are possible. In that case,the location must be changed.

    7.2 Unpacking

    Carefully remove the balance from the packaging, remove plastic cover and setup

    balance at the intended workstation.

  • 7/28/2019 ABT-BA-e-0911

    15/71

    ABT-BA-e-0911 15

    7.2.1 Scope of delivery

    Serial accessories:

    Balance

    Weighing plate

    Carrier of weighing plate

    Screening ring

    Mains adapter

    Adapter cable holder

    Protective cover

    Operating Manual

  • 7/28/2019 ABT-BA-e-0911

    16/71

    ABT-BA-e-091116

    7.2.2 Installation

    Attach the adapter cable holder. Peelthe protective sheet of adhesive of

    the adapter cable holder, and stick iton the back of the balance as shownin the figure.

    Carrier of weighing plate, attachweighing plate and circular screen inorder.

    Level balance with foot screws untilthe air bubble of the water balance is

    in the prescribed circle.

    When the key panel and the displaymust be protected from dirt and wear,place the cover over the key panel.

  • 7/28/2019 ABT-BA-e-0911

    17/71

    ABT-BA-e-0911 17

    1

    7.3 Mains connection

    Power is supplied via the external mains power supply. The stated voltage valuemust be the same as the local voltage. Only use original KERN mains power sup-plies. Using other makes requires consent by KERN.

    7.4 Connection of peripheral devices

    Before connecting or disconnecting of additional devices (printer, PC) to the data in-terface, always disconnect the balance from the power supply.With your balance, only use accessories and peripheral devices by KERN, as theyare ideally tuned to your balance.Terminal for external devices:

    1 RS -232 interface2 Mains connector3 Anti-theft safety device (for chains or other fastenings)4 Connection earthing terminal

    2

    34

  • 7/28/2019 ABT-BA-e-0911

    18/71

    ABT-BA-e-091118

    7.5 Initial Commissioning

    A warming up time of 4 hours after switching on stabilizes the measuring values.The accuracy of the balance depends on the local acceleration of gravity.Strictly observe hints in chapter Adjustment.

    7.5.1 Turning On the Power

    1. Supply power to balance via mainsadapter. After the balance performs aself check, calibration will be automati-cally executed. During this process, thedisplay will change as follows. CHE 5,CHE 4...... CHE 0, CHE 2, CHE 1, CAL 2 CAL 0 , CAL-End, oFF.This adjustment can be interrupted im-

    mediately after switching-on the powersupply by actuating the [ON/OFF]-key.However, at least one adjustment mustbe carried out before the balance canbe used.

    2. Actuate the [ON/OFF] key. After all dis-plays appear, the gram (g) symbol ap-pears.

    3. Pressing the [ON/OFF] key again turnson the standby symbol and puts the

    balance in standby (warm up) status.Current time (see ch. 13.5) will be dis-played.

  • 7/28/2019 ABT-BA-e-0911

    19/71

    ABT-BA-e-0911 19

    8 Adjustment

    As the acceleration value due to gravity is not the same at every location on earth,each balance must be coordinated - in compliance with the underlying physicalweighing principle - to the existing acceleration due to gravity at its place of location

    (only if the balance has not already been adjusted to the location in the factory). Thisadjustment process must be carried out for the first commissioning, after eachchange of location as well as in case of fluctuating environment temperature. To re-ceive accurate measuring values it is also recommended to adjust the balance peri-odically in weighing operation.Observe stable environmental conditions. Stabilisation requires a warm-up period of1 or 4 hours (for models d = 0.01 mg). Ensure that there are no objects on theweighing plate.

    8.1 Automatic adjustment via PSC

    The default setting for balances of the series ABT ("PSC ON") prompts automaticadjustment via the PSC function.The moment a change in temperature is detected this function uses the internal tareweight to carry out fully-automatic adjustment with the help of a temperature sensor.If PSC is left ON, when there is a temperature change that would influence sensitiv-ity, span calibration executes automatically to maintain the sensitivity of the balance.Adjustment in weighing mode is carried out automatically under the following condi-tions:

    (1) If a change in environmental temperature (0.5 C) occurs;

    (2) When about four hours has passed since the previous calibration.

    (3) When the balance is switched from standby status to weighing modeand condition (1) or (2) has been met.

    (4) When balance was disconnected from the mains.

    If one of the above conditions was met in weighing mode, the weight symbol flashesfor about two minutes in order to notify the pending adjustment; afterwards the dis-play will show "PSC run". During operation, the display will automatically change andthe motor sound of the weight loading system is heard. In order to ensure properPSC operation, prevent vibrations and air flow. As soon as the display in grams re-appears after completing adjustment via PSC, the balance returns to weighing mode.The sensitivity before and after adjustment is slightly different. Also, no measure-ments can be made during adjustment. To prevent a start-up of adjustment during ameasuring process you have to actuate the [ON/OFF] key as soon as the weightsymbol starts flashing. This will interrupt automatic adjustment.

  • 7/28/2019 ABT-BA-e-0911

    20/71

    ABT-BA-e-091120

    8.1.1 PCS function ON/OFF

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Actuate the [CAL] key repeatedly until the current "PSC:" set-ting appears:

    To change the setting, actuate the [TARE] key as soon as

    "PSC:" appears.

    You can use the [CAL] key to select between the following set-tings:

    PSC-on Function activated

    PSC-oF Function deactivated

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

    PSC and Clock-CAL may be turned on and off separately. The setting control display

    (see ch. 12.6) shows the weight symbol ( ), when PSC or Clock-CAL or both func-tions are activated.

  • 7/28/2019 ABT-BA-e-0911

    21/71

    ABT-BA-e-0911 21

    8.2 Automatic adjustment via Clock-CAL

    (only models ABT 100-5M, ABT 120-5DM, ABT 220-5DM)

    With the help of its internal adjusting weight and integrated clock the balance can beset to carry out automatic adjustment at set times (up to three times daily). Clock-CAL is a very convenient function, when calibration reports are desired to be madefor regular calibrations, or when wishing span calibrations during break times to avoidinterruption of measurement work.The weight symbol blinks for about two minutes as notification of span calibrationbefore it begins. Automatic adjustment can be stopped by actuating the [ON/OFF]key during this message.

    Clock-CAL function ON/OFF:

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Actuate the [CAL] key repeatedly until the current "tCal:" settingappears:

    To change the setting, actuate the [TARE] key as soon as

    "tCal:" appears.

    You can use the [CAL] key to select between the following set-tings:

    tCAL-on Function activated

    tCAL-oF Function deactivated

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    22/71

    ABT-BA-e-091122

    8.2.1 Setting the time for Clock-CAL

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    (for setting 1)

    Press the [CAL] key repeatedly until tCAL t* appears:

    Actuate the [TARE] key until "t*HH:MM" is displayed. The *-position symbolises a number between 1 and 3 (3 set times for

    automatic adjustment). The currently set time is displayed in"HH:MM" (HH for hours, MM for minutes) format with the first digiton the left flashing. The time appears as _ _:_ _ when no time is

    set. The symbol, and the # symbol appear to indicate it is inthe numerical input status.

    Pressing the [PRINT] key moves the blinking digit to the next digitto the right. Press the [UNIT] key to change the value of the blink-

    ing digit. When the [UNIT] key is pressed, the numerical of theblinking digit increases by 1 at a time. The numerals progress in

    this order: 0

    1

    2

    ...

    9

    _

    0... Set the hours within the rangeof 00 to 23 and the minutes within the range of 00 to 59. Completeyour setting by pressing the [TARE] key. The display returns to tCAL t*.

    In order to set another time, press the [CAL] key to move to thenext t CAL t* setting and set the time in the same way.

    Actuate the [ON/OFF] key repeatedly. Return to menu/weighingmode.

    Clearing the Settings

    The clock settings tCAL t1 to tCAL t3 may each be reset byusing procedure 3 to set the time setting to _ _:_ _.

    PSC and Clock-CAL may be turned on and off separately. The setting control display (see ch. 12.6)

    shows the weight symbol ( ), when PSC or Clock-CAL or both functions are activated.

  • 7/28/2019 ABT-BA-e-0911

    23/71

    ABT-BA-e-0911 23

    8.3 Adjustment via preset process

    It is possible to start the preset adjustment method without having to access themenu. When in weighing mode, the preset adjustment method can be carried out bysimply actuating the [CAL] key, followed by [TARE].

    8.3.1 Selecting default adjustment method

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Press the [TARE] key. The display shows E CAL

    Press the [CAL] key repeatedly until the desired setting appears.The standstill display ( ) appears when current adjustment isdisplayed.

    Select amongst the following four types:

    Not documented

    Adjustment test with external weight(see ch. 8.3.4)

    Adjustment test with internal weight(see ch. 8.3.2)

    Adjustment test with internal weight(see ch. 8.3.3)

    Confirm your setting by pressing the [TARE] key. The display

    shows CAL dEF.

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

    The selected adjustment can be carried out by simply actuating

    the [CAL] key followed by [TARE].

  • 7/28/2019 ABT-BA-e-0911

    24/71

    ABT-BA-e-091124

    8.3.2 Default setting: Adjustment with internal weight

    Condition: Function i tCAL activated (see ch. 8.3.1)

    Press the [CAL] key The display shows i-CAL

    Press the [TARE] key. The display changes automatically in thisorder: CAL 2, CAL 1, CAL 0, and CAL End. After success-ful adjustment the balance automatically returns to weighingmode.In case of an adjustment error (e.g. objects on the weighingplate) the display will show an error message, repeat adjustment.

    8.3.3 Default setting: Adjustment test with internal weight

    During adjustment tests the balance automatically compares the saved value of theadjustment weight with the actual value. This is only a check, i.e. no values arechanged.

    Condition: Function i tESt activated (see ch. 8.3.1)

    Press the [CAL] key The display shows i-tESt

    Press the [TARE] key. The display changes automatically in thisorder: tESt 2, tESt 1, tESt 0

    After this, the difference to the previous adjustment will be dis-played for several seconds.

    After "tESt End" was displayed, the balance will automatically

    return to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    25/71

    ABT-BA-e-0911 25

    8.3.4 Default setting: Adjustment test with external weight

    Condition: Function E tESt activated (see ch. 8.3.1)

    Press the [CAL] key The display shows E-tESt

    Press the [TARE] key. Testing will be started and zero display isflashing (Ensure that no items are present on the weighing plate).

    (Example)

    Wait until the weighing value for adjustment is shown andflashes.

    Place the required adjusting weight on the weighing plate.

    Wait until the blinking zero display reappears. (this may takeabout 30 seconds.)

    Remove the weight from the weighing plate.

    After this, the difference to the previous adjustment will be dis-played for several seconds.

    After "tESt End" was displayed, the balance will automatically

    return to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    26/71

    ABT-BA-e-091126

    8.4 Carrying out alternative adjustment methods

    Here you start adjustment by selecting a setting in the menu.

    8.4.1 Adjustment with internal weight

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press the [TARE] key. "CAL" is displayed.

    Press the [TARE] key. "E CAL" is displayed.

    Repeatedly press the [CAL] key until "i CAL" appears.

    Press the [TARE] key. The display changes automatically in thisorder: CAL 2, CAL 1, CAL 0, and CAL End.

    After successful adjustment the balance automatically returns toweighing mode.In case of an adjustment error (e.g. objects on the weighing plate)the display will show an error message, repeat adjustment.

  • 7/28/2019 ABT-BA-e-0911

    27/71

    ABT-BA-e-0911 27

    8.4.2 Adjustment test with internal weight

    During adjustment tests the balance automatically compares the saved value of theadjustment weight with the actual value. This is only a check, i.e. no values arechanged.

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press the [TARE] key. "CAL" is displayed.

    Press the [TARE] key. "E CAL" is displayed.

    Repeatedly press the [CAL] key until "i tESt" appears.

    Press the [TARE] key. The display changes automatically in thisorder: tESt 2, tESt 1, tESt 0

    After this, the difference to the previous adjustment will be dis-played for several seconds.

    After "tESt End" was displayed, the balance will automaticallyreturn to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    28/71

    ABT-BA-e-091128

    8.4.3 Adjustment test with external weight

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press the [TARE] key. "CAL" is displayed.

    Press the [TARE] key. "E CAL" is displayed.

    Repeatedly press the [CAL] key until "i tESt" appears.

    Press the [TARE] key. Testing is started and zero display isflashing.(Ensure that there are no objects on the weighing plate).

    (Example)

    Wait until the weighing value for adjustment is shown andflashes.

    Place the required adjusting weight on the weighing plate.

    Wait until the blinking zero display reappears. (this may takeabout 30 seconds.)

    Remove the weight from the weighing plate.

    After this, the difference to the previous adjustment will be dis-played for several seconds.

    After "tESt End" was displayed, the balance will automaticallyreturn to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    29/71

    ABT-BA-e-0911 29

    9 Verification

    General introduction:According to EU directive 90/384/EEC balances must be verified if they are used asfollows (legally controlled area):

    a) For commercial transactions if the price of goods is determined by weighing.b) For the production of medicines in pharmacies as well as for analyses in themedical and pharmaceutical laboratory.

    c) For official purposes.d) For manufacturing final packages.

    In cases of doubt, please contact your local trade in standard.

    Verification instructions

    An EU type approval exists for balances described in their technical data as verify-able. If a balance is used where obligation to verify exists as described above, itmust verified and re-verified in regular intervals.Re-verification of a balance is carried out according to the respective national regula-tions. The validity for verification of balances in Germany is e.g. 2 years.The legal regulation of the country where the balance is used must be observed!After verification the balance is sealed at the indicated positions.Verification of the balance is invalid without the "seal".

    Position of the official seals:

    Balances with obligation to verify must be taken out of operation if:

    - The weighing result of the balance is outside the error limit. Therefore, inregular intervals load balance with known test weight (ca. 1/3 of the max.load) and compare with displayed value.

    - The reverification deadline has been exceeded.

    Seals

  • 7/28/2019 ABT-BA-e-0911

    30/71

    ABT-BA-e-091130

    10 ISO/GLP log

    Quality assurance systems require printouts of weighing results as well as of correctadjustment of the balance stating date and time and balance identification. The easi-est way is to have a printer connected.

    10.1 Setting the log printout of your adjustment data

    This function enables automatic log issue after each adjustment. These logs may beissued by using an optional printer. Fully automatic and periodical adjustments arecarried out and logs created in combination with the Clock-CAL function (see ch. 8.2).First make adjustments as specified in chap. 15.4 under KERN-YKB-01N.Then set the log printout as follows:

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Repeatedly actuate the [CAL] key until "GLP:**" appears (The symbolise the current setting).

    To change settings, actuate the [TARE] key when "GLP:**" ap-pears.

    You can use the [CAL] key to select between the following set-tings:

    GLP-on Function activated

    GLP-oF Function deactivated

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    Actuate the [ON/OFF] key repeatedly. Return to menu/weighingmode.

  • 7/28/2019 ABT-BA-e-0911

    31/71

    ABT-BA-e-0911 31

    10.2 Balance ID Number Setting

    This setting is for the balance ID number that is output along with the adjustment re-port.

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Repeatedly actuate the [CAL] key until "id:****" appears (The symbolise the current setting).

    Press the [TARE] key. In the upper part of the display panel, thesymbol and the # symbol appear in order to indicate nu-

    merical input status. The leftmost digit of blinks. The numeralof the blinking digit can be changed.

    When the [UNIT] key is pressed, the numerical of the blinking digitincreases by 1 at a time. You can determine the value of the flash-ing digit, or shift the flashing digit by one position to the right, by

    pressing the [PRINT] key. Confirm your setting by pressing the[TARE] key.

    Actuate the [ON/OFF] key repeatedly. Return to menu/weighingmode.

  • 7/28/2019 ABT-BA-e-0911

    32/71

    ABT-BA-e-091132

    10.3 Setting the date printout

    This setting determines whether or not the date and time on the balances built-inclock is printed out along with the log.

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Repeatedly actuate the [CAL] key until "Prtdt:" appears (The symbolise the current setting).

    To change settings, actuate the [TARE] key when "Prtdt:" ap-pears.

    You can use the [CAL] key to select between the following set-tings:

    Prtdt-on Date and time are printed.

    Prtdt-oF Date and time are not printed.

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    Actuate the [ON/OFF] key repeatedly. Return to menu/weighingmode.

    10.3.1 Printout Date and Time, without weight value

    To print out only date and time, without weight value,press [PRINT] button for approx. 3 sec.

  • 7/28/2019 ABT-BA-e-0911

    33/71

    ABT-BA-e-0911 33

    11 Basic Operation

    11.1 Weighing

    Note: Stabilisation requires a warm-up period of 1 or 4 hours(for models d = 0.01 mg).

    Turn on balance by pressing the [ON/OFF] key. The balance will carry out aself-test Your balance is ready to weigh as soon as the 0.0000 g" display ap-pears.Note: The [TARE] key can be used to set the balance to zero at any time.

    Place goods to be weighed on balance. Wait until the standstill display ( )appears, then read the weighing result.

    11.2 Taring

    The dead weight of any weighing container may be tared away by pressing a button,so that the following weighings show the net weight of the goods to be weighed.

    Place empty tare container on the weighing plate. The total weight of thecontainer is displayed.

    To start the taring process press the [TARE] key. The weight of the con-tainer is now internally saved.

    Place the goods to be weighed into the tare container.

    Read the weight of the goods on the display.

    Note:The balance is able to only store one taring value at a time.When the balance is unloaded the saved taring value is displayed with negative sign.Remove all items from the weighing plate in order to delete the stored tare value andsubsequently press the [TARE] key.The taring process can be repeated any number of times. The limit is reached whenthe whole weighing range is exhausted.

  • 7/28/2019 ABT-BA-e-0911

    34/71

    ABT-BA-e-091134

    11.3 Changing the display

    You can change the display to activated units, piece counting, percentage and den-sity determination mode by pressing the [UNIT] key several times.

    The default setting provides the following options:

    [g] [%] [Pcs] [ct] [g]

    Different settings have to be activated in the menu:

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press [TARE] key

    (Example)Repeatedly press the [CAL] key until "Unit.SEL" appears.

    Press [TARE] key

    To select settings, use the [CAL] key:

    Current settings are indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key.

    The [TARE] key is also used to deactivate a unit or function ifthe corresponding setting with standstill display is shown in thedisplay.

    Repeatedly press the [ON/OFF] key. This takes you back to themenu/weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    35/71

    ABT-BA-e-0911 35

    11.4 Changing readability

    (only models ABT 100-5M, ABT 120-5DM, ABT 220-5DM)

    Models ABT 120-5DM and ABT 220-5DM are set to lower range with a readabilityof 0.01 mg as soon as they are connected to power supply and turned on. To switchto the upper range with a minimum display of 0.1mg, press the [1d/10d] key.If the lower max. weighing range (82g for ABT 220-5DM, 42 g for ABT 120-5DM) isexceeded during the weighing process, the balance changes automatically to the up-per range.If you tare the balance in the upper range, this setting will be maintained throughout.To return to lower range, you will have to actuate the [TARE] key after the stabilitysymbol ( ) has appeared.

    Lower range:

    Upper range:

  • 7/28/2019 ABT-BA-e-0911

    36/71

    ABT-BA-e-091136

    11.5 Underfloor weighing

    Objects unsuitable for placing on the weighing pan due to size or shape may beweighed with the help of the flush-mounted platform.Proceed as follows:

    Switch off balance.

    Open the closing lid (1) on the bottom of your balance.

    Place your balance over an opening.

    Suspend the goods to be weighed from the hook and carry out the weighing.

    Fig. 1: Underfloor weighing device

    CAUTION

    Always ensure that all suspended objects are stable enough to hold the desiredgoods to be weighed safely (danger of breaking).

    Never suspend loads that exceed the stated maximum load (max) (danger ofbreaking)

    Always ensure that there are no persons, animals or objects that might be damagedunderneath the load.

    NOTE

    After completing the underfloor weighing the opening on the bottom of the balancemust always be closed (dust protection).

    (1)

    !

    !

  • 7/28/2019 ABT-BA-e-0911

    37/71

    ABT-BA-e-0911 37

    12 The menu

    To adapt the behaviour of the balance to your requirements, go to the balance modemenu. Usually the default setting of the balance mode menu is such that you do notneed to make any changes. If you encounter special conditions of use, go to the bal-

    ance mode menu in order to set your balance according to your individual require-ments.

    12.1 Changing settings

    To change the settings of specific functions, select the related functions.

    Functions are changed by taking the following three steps:

    Call up menu

    Set function Confirm and store

    When setting a function the ON/OFF, CAL and TARE keys have special functions.

    12.2 How to change settings

    CAL key = Select menu and pass through menu items from top to bot-

    tom (

    ).

    TARE key = Select function.After selecting the function on the display by pressing the CAL key, thechange is invoked by pressing the TARE key.

    CAL key = Select one of the possible settings within the function. Pass-ing through menu items from top to bottom.

    TARE key = Store the setting momentarily appearing on the display byactuating the TARE key.

    The standstill display shows the present setting for the function.

    ON/OFF key = Exit function

    Brief actuation ofON/OFFkey:Back to previous menu.

    Long actuation ofON/OFFkey:Back to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    38/71

    ABT-BA-e-091138

    12.3 Call up menu

    Please have a go and try to change a function yourself. Change the "Auto-Zero" func-tion to OFF and then again to ON.

    Turn on balance by pressing the ON/OFF key.

    Actuate the CAL key until FUnC.SEL appears

    Press the TARE key once until CAL appears

    Press the CAL key once until trC :on appears

    To confirm, press the TARE key once."Auto-Zero function is now selected.The standstill display shows the current setting.

    Press the CAL key

    thereby means:trC-oF Auto-Zero is switched offtrC-on Auto-Zero is switched onSelect trC-oF

    Press the TARE key onceThe colons appear when storing the changed trC: oF

    Press ON/OFF keyTo quit the menu keep the ON/OFFkey pressed longer than 2 sec.

    Note:If you wish to make several changes in the weighing mode menu you do not have toexit the menu after each change. You can make several changes one after the otherand finally exit the menu

  • 7/28/2019 ABT-BA-e-0911

    39/71

    ABT-BA-e-0911 39

    12.4 Selecting the menu

    In weighing mode press the [CAL] key. First function "i-Cal" (ex.) appears.

    The various functions of the menu are passed through by pressing the [CAL]key again.

    Weighing mode

    Preset adjustment method (see ch. 8.3)

    Not documented

    Setting control display (see ch. 12.6)

    Standard mode

    P0VR 1nG

    Apportioning mode

    Anti-convection mode

    High-stability Mode

    Entry to second level

    (application functions, menu group for individual settings)

    Entry to third level (system settings menu group)

    Entry to forth level (communication settings menu group)

    Weighing mode

    Settings for stability and response

    (see chapter 13.1)

  • 7/28/2019 ABT-BA-e-0911

    40/71

    ABT-BA-e-091140

    12.5 Menu overview

    * = default setting

  • 7/28/2019 ABT-BA-e-0911

    41/71

    ABT-BA-e-0911 41

  • 7/28/2019 ABT-BA-e-0911

    42/71

    ABT-BA-e-091142

  • 7/28/2019 ABT-BA-e-0911

    43/71

    ABT-BA-e-0911 43

    12.6 Menu Lock

    The menu setting operations can be locked so that the settings cannot be inadver-tently changed. This is called Menu Lock. The menu lock is activated with the follow-ing procedure.

    Turn on the power supply for the balance and wait until "oFF" ap-pears.

    Press the [CAL] key until "LoCKEd" appears. Menu is disabled,menu symbol appears. This is followed by a reappearance of"oFF".

    If a menu item selection is attempted in locked status, the message LoCKEd ap-pears and the menu selection is not allowed. To deactivate menu lock, follow thesequence of operations below:

    Disconnect the balance from the power and connect again after awhile.

    When "oFF" is displayed, press the [CAL] key until "rELASE" ap-pears. The menu lock is deactivated.

    12.7 Resetting the menu

    This will return all the settings to default. The reference value stored in previous useof piece counting or percentage conversion will also be cleared. The default settingsare indicated with # on the Menu Map.

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key. The display shows CAL dEF.

    Press the [CAL] key repeatedly until rESET appears.

    Press the [TARE] key. The display shows confirmation enquiry"rESET?

    Confirm by pressing the [TARE] key; balance automaticallyreturns to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    44/71

    ABT-BA-e-091144

    12.8 Settings control display

    To receive a confirmation of the current settings, repeatedly actuate the [CAL] key inweighing mode until "St. b1 t" (ex.) appears.

    Stability and response setting

    St: Standard modeCo: Anti-convection modeHi: High-stability modePo: Apportioning mode

    Stability and response setting

    b1: 1 countb5: 5 countb10: 10 count

    Appears when zero

    function is turned on.

    The weight symbol is displayed when either automatic adjustment PSC or Clock-CALor both functions are turned on.

    Appears when the automatic adjustment has

    been switched on by PSC or Clock-CAL.

    Appears when the issue of the adjustment

    data is switched on.

  • 7/28/2019 ABT-BA-e-0911

    45/71

  • 7/28/2019 ABT-BA-e-0911

    46/71

    ABT-BA-e-091146

    13.1.2 Anti-convection mode

    If measurements have to be taken under unavoidable ambience conditions, (e. g.changing air condition systems) convection occurring within the weighing space mayresult in fluctuations in the display on the balance after the stability display has ap-peared. The small range (readability 0.01 mg) of series ABT is particularly suscepti-ble to this effect.Anti-convection mode adjusts the timing of appearance of the stability mark. Notethat the stability mark will then appear at a late timing.

    Repeatedly press the [CAL] key until "ConvECt" appears.

    To set anti-convection mode, press the [TARE] key.

    Setting of this mode can be confirmed only in settings checkdisplay (see ch. 12.6).

    13.1.3 High-stability Mode

    The balances of series ABT are designed to react only minimally to the effects of vi-brations and drafts. However, if it must be set up in a location with poor conditions,

    use this function to reduce the effects of vibration or air current even further. There isa slight increase in the reaction time of the balance but the display itself is stabilized.

    Repeatedly press the [CAL] key until "ConvECt" appears.

    To set high stability mode, press the [TARE] key.

    Setting of this mode can be confirmed only in settings checkdisplay (see ch. 12.6).

    Setting control

    display

    (when selectinganti-convection mode)

    Setting control

    display

    (when selectinghigh-stability mode)

  • 7/28/2019 ABT-BA-e-0911

    47/71

    ABT-BA-e-0911 47

    13.1.4 Apportioning mode

    Use this function if you wish to increase display speed,e.g. during apportioning. However, please note that the balance is very susceptible toambience conditions. You can set the degree of sensitivity according to site(quiet/unquiet).

    P0VR 1nG

    Repeatedly press the [CAL] key until "PoUr inG" appears.

    nQlnAL.Eno Press the [TARE] key; display shows current setting of sensitiv-

    ity

    Setting of this mode can be confirmed only in settings checkdisplay (see ch. 11.6).

    5TABL.Eno

    nQlnAL.Eno

    Un5TABL.Eno

    You can use the [CAL] key to select between the following set-tings:

    StAbl.Env very quiet site/sensitive and fast

    normL.Env normal site/ medium setting

    UnStAbl.Env very unquiet site/insensitive but slow

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    For mode setting in setting control display (see ch.12.6):

    PQ.S.B1T

    very quiet site/sensitive and fast

    PQ.n.B1T

    normal site/ medium setting

    PQ.U.B1T

    very unquiet site/insensitive but slow

  • 7/28/2019 ABT-BA-e-0911

    48/71

    ABT-BA-e-091148

    13.2 Standstill display

    If the display shows the stability display ( ) the balance is in a stable status. Thecondition for judging stability is user-selectable. There are three levels to select: 1count, 5 counts and 10 counts. The default setting is 1 count.

    The setting of the standstill display can be checked on the setting control display (seech.12.6).

    Setting the standstill display

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press [TARE] key

    (Example)

    Repeatedly press the [CAL] key until the current "b And:**" set-ting appears.

    Press [TARE] key

    You can use the [CAL] key to select between the following set-tings:

    "b-1" 1 count

    "b-1" 5 count

    "b-1" 10 count

    The current setting is marked by the stability symbol ( ).

    Confirm your selection by pressing the [TARE] key

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    49/71

    ABT-BA-e-0911 49

    13.3 Auto Zero

    This function is used to tare small variations in weight automatically.In the event that small quantities are removed or added to the material to beweighed, incorrect weighing results can be displayed due to the stability compensa-tion in the balance. (e.g. slow flow of liquids from a container placed on the balance,evaporating processes).When apportioning involves small variations of weight, it is advisable to switch off thisfunction.

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press [TARE] key

    (Example)

    Repeatedly press the [CAL] key until the current "trC:**" settingappears.

    Press [TARE] key

    You can use the [CAL] key to select between the following set-tings:

    trC-onFunction activated

    trC-oFFunction deactivated

    The current setting is marked by the stability symbol ( ).

    Confirm your selection by pressing the [TARE] key

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:

    Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    50/71

    ABT-BA-e-091150

    13.4 Setting date

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key.

    Repeatedly press the [CAL] key until "d-MM.DD" appears (MMand DD provide two digits each, indicating month and day).

    (15. march 2003)

    Press the [TARE] key. The date currently set appears. In the up-

    per part of the display panel, the symbol and the # symbol

    appear in order to indicate numerical input status. The leftmostdigit blinks.

    (2. nov. 2004)

    When the [UNIT] key is pressed, the numerical of the blinking digitincreases by 1 at a time. You can determine the value of the flash-ing digit, or shift the flashing digit by one position to the right, bypressing the [PRINT] key. Confirm your setting by pressing the

    [TARE] key.

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    51/71

    ABT-BA-e-0911 51

    13.5 Setting time

    Balances of the series ABT are equipped with an integrated clock. Set the clock be-fore using the functions Clock-CAL (ch. 8.2) or GLP log (ch. 10). Note that the currenttime is displayed at the stand by status (ch. 7.5.1).

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key.

    Repeatedly actuate the [CAL] key until "t-HH.MM" appears (HH

    and MM provide 2 digits each for hour and minute display).

    Press the [TARE] key. The currently set time appears. In the up-

    per part of the display panel, the symbol and the # symbolappear in order to indicate numerical input status. The leftmostdigit blinks.

    When the [UNIT] key is pressed, the numerical of the blinking digitincreases by 1 at a time. You can determine the value of the flash-ing digit, or shift the flashing digit by one position to the right, bypressing the [PRINT] key. Confirm your setting by pressing the

    [TARE] key.

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    52/71

    ABT-BA-e-091152

    13.6 Capacity display

    This function displays a bar graph representation of the load on the weighing plate.This may be used to prevent sudden appearance of oL (overload) duringmeasurement.

    It is possible to turn the display of capacity on or off.

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press [TARE] key

    Repeatedly press the [CAL] key until the current "AdiSP:**" set-ting appears.

    Press [TARE] key

    You can use the [CAL] key to select between the following set-tings:

    Ad-on Function activated

    Ad-oF Function deactivated

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode.

    Display will appear when about one third of capacity isexhausted.

  • 7/28/2019 ABT-BA-e-0911

    53/71

    ABT-BA-e-0911 53

    14 Application Functions

    14.1 Parts counting

    With parts counting you can either count parts into a container or remove parts from

    a container. To count a greater number of parts the average weight per part has tobe determined with a small quantity (reference quantity). The larger the referencequantity, the higher the counting exactness. High reference must be selected forsmall parts or parts with considerably different sizes.The process has four steps:

    Tare the weighing container

    Determine the reference unit

    Weigh in the reference unit

    Count the items

    Condition: Function "U- PSC" activated (ch. 11.3)

    (When PCS is used for thefirst time)

    Repeatedly press the [UNIT] until display is in percentagemode. The display symbol PCS pops up.

    If you are using a weighing container use the [TARE] key for

    taring.

    To determine the reference piece number, place 10, 20, 50 or100 counting parts on the balance.

    Press the [CAL] key

    Repeated pressing of the [CAL] key allows the change betweenthe following reference piece numbers "Ld 10", Ld 20,Ld 50and Ld 100.

    Important: The larger the reference quantity, the more accuratethe parts counting.

    (When loading 20 pieces)

    When standstill control ( ) is complete, confirm your referencepiece number by pressing the [TARE] key.

    The display shows "SEt" for several seconds and the stored ref-erence piece number is displayed.

  • 7/28/2019 ABT-BA-e-0911

    54/71

    ABT-BA-e-091154

    Remove reference weight.

    Now you can fill the items to be counted into the container.The respective quantity is shown in the display.

    Repeated pressing of the [UNIT] key changes the display value,

    e. g.: [g] [%] [Pcs] [ct]

    Note:Error message "Err 20" indicates that the weight for the reference piece number istoo low.

    14.2 Percent determination

    Percent weighing allows to display weight in percent, in relation to a referenceweight. The displayed weighing value is stored as a standard percent value(default setting: 100%).

    Condition: Function "U- %" is activated (ch. 11.3)

    (When PCS is used for thefirst time)

    Repeatedly press the [UNIT] key until the display is in percent-age determination mode. The display symbol % pops up.

    If you are using a weighing container use the [TARE] key fortaring.

    Put on reference weight (=100%)(minimum weight: Readability d x 100)

    When standstill control ( ) is complete, press the [CAL] key.The reference weight is saved.

    Remove reference weight.

    From now on the placed weight will be shown as %.

    Repeated pressing of the [UNIT] key changes the display value,

    e. g.: [g] [%] [Pcs] [ct]

  • 7/28/2019 ABT-BA-e-0911

    55/71

    ABT-BA-e-0911 55

    14.3 Density determination

    Density determination with the help of the flush-mounted platform facility is describedin the following.Density determination becomes even easier when an optional set for density deter-mination is applied. For further information please refer to the operating instructionsenclosed with the set for density determination.

    1. Remove the below-weigh hook cover on the floor of the balance, after remov-ing the two fixing screws.

    2. Hang the user-prepared hanging pan from the below-weigh hook and sink thathanging pan into the sample liquid in the tank.

    14.3.1 Density determination of solids

    Condition: Function ",d" (density of solids) is activated. See Chapter 11.3.

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key.

    Repeatedly actuate the [CAL] key until "LSG SEt" appears

    (Example)

    Press the [TARE] key. The currently set density for the liquid to bemeasured appears. In the upper part of the display panel, the

    symbol and the # symbol appear in order to indicate nu-merical input status. The leftmost digit blinks.

    (Example)

    Enter density for your liquid to be measured. When the [UNIT] keyis pressed, the numerical of the blinking digit increases by 1 at atime. You can determine the value of the flashing digit, or shift theflashing digit by one position to the right, by pressing the [PRINT]

    key. Confirm your setting by pressing the [TARE] key.

    Repeatedly press the [ON/OFF] key until the balance is in weigh-ing mode.

  • 7/28/2019 ABT-BA-e-0911

    56/71

    ABT-BA-e-091156

    Repeatedly press the [UNIT] key until the balance is in densitydetermination mode for solids ",d". Note that g also appears dur-ing weight measurement in air.

    Press the [TARE] key. Place the item to be measured on theweighing pan.

    When standstill control is complete, press the [CAL] key

    Place the item to be measured on the immersed hanging weighingpan. The display is showing the density of the measured item.dSP oL may be displayed when nothing is on the weighing pan,which is normal.

    14.3.2 Determining density of liquids

    Condition: Function "d" (density of liquids) is activated. See Chapter 11.3.

    Press the [CAL] key repeatedly until "SettinG" appears.

    Press the [TARE] key.

    Repeatedly actuate the [CAL] key until "Sv SEt" appears

    (Example)

    Press the [TARE] key. The currently set density for the body to beimmersed appears. In the upper part of the display panel, the

    symbol and the # symbol appear in order to indicate nu-merical input status. The leftmost digit blinks.

    (Example)

    Enter density for your body to be immersed. When the [UNIT] keyis pressed, the numerical of the blinking digit increases by 1 at atime. You can determine the value of the flashing digit, or shift the

    flashing digit by one position to the right, by pressing the [PRINT]key. Confirm your setting by pressing the [TARE] key.

    Repeatedly press the [ON/OFF] key until the balance is in weigh-ing mode.

  • 7/28/2019 ABT-BA-e-0911

    57/71

    ABT-BA-e-0911 57

    Repeatedly press the [UNIT] key until the balance is in densitydetermination mode for liquids "d". Note that g also appears dur-ing weight measurement in air.

    Press the [TARE] key. Place the body to be immersed on the

    weighing pan.

    When standstill control is complete, press the [CAL] key

    Immerse the body to be immersed without bubbles developing.The display is showing the density of the sample. dSP oL may bedisplayed when nothing is on the weighing pan, which is normal.

  • 7/28/2019 ABT-BA-e-0911

    58/71

    ABT-BA-e-091158

    14.4 Auto Print

    Using Auto Print allows measurement results to be automatically output via the RS-232C interface without pressing the [PRINT] key individually with every measure-ment. An issue will be made after standstill control ( ) is complete. The next meas-urement is done after removing the items from the weighing plate and the display is

    returning to a value within the 3-fold of the zero range.

    Important: This mode can be applied to any unit.

    The initial load of the balance must be kept within the 5-fold of the zerorange.

    Weight of sample 10 x zero range (zero range = 0.25 e)

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press [TARE] key

    (Example)

    Repeatedly press the [CAL] key until the current "AtPrt:**" set-ting appears.

    Press [TARE] key

    You can use the [CAL] key to select between the following set-tings:

    AtPrt-on Function activated

    AtPrt-oF Function deactivated

    The current setting is indicated by the standstill display ( ).

    Confirm your selection by pressing the [TARE] key

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode. Whilst the auto print function is activethe display keeps showing the [AP] symbol.

  • 7/28/2019 ABT-BA-e-0911

    59/71

    ABT-BA-e-0911 59

    14.5 Add up mode

    An optional number of single weighing processes are automatically added up to atotal sum, e.g. all single weighing processes of a batch or if you have to carry out agreat number of measurements of small samples.When the standstill control ( ) is complete the weighing value is automatically issuedto the optional printer. The displayed value is added into the total adding memory.Afterwards automatic taring will take place. This is repeated newly every subsequenttime a new sample is placed on the weighing pan. When the last single weighingprocess is finished, press the [ON/OFF] key to receive the total sum ("TOTAL=").

    Important: This mode can be applied to any unit.

    The initial load of the balance must be kept within the 5-fold of the zerorange.

    Weight of sample 10 x zero range (zero range = 0.25 e)

    If during the use of multiple range balances the smallest readability digitwas turned off by pressing the [1d/10d] key, evaluation is based on thelatest add-on value displayed before.

    Activate function:

    Repeatedly press the [CAL] key until "FUnC.SEL" appears.

    Press [TARE] key

    (Example)

    Repeatedly press the [CAL] key until the current "Addon:**" set-ting appears.

    Press [TARE] key

    You can use the [CAL] key to select between the following set-tings:

    Adon-on Function activated

    Adon-oF Function deactivated

    The current setting is indicated by the standstill display ( ).

  • 7/28/2019 ABT-BA-e-0911

    60/71

    ABT-BA-e-091160

    Confirm your selection by pressing the [TARE] key. Printout "-------- ADDON MODE ---------" takes place.

    To exit the function, press the [ON/OFF] key.

    Brief actuation of ON/OFF key:Back to previous menu.

    Long actuation of ON/OFF key:Back to weighing mode. Whilst the add-on function is active, thedisplay keeps showing the add-on symbol [ ].

    Add up:

    Operation Display

    When using a container, place this on the weighing plate and tare.

    Prior to weighing the component press the [PRINT] key 1x.

    Weigh first component.

    5.0003 g

    Wait until standstill display ( ) appears, the weighing value forthe first component (CMP001) will be issued automatically to theoptional printer. This is followed by automatic taring.

    0.0000 g

    Weigh second component 5.0017 gWait until standstill display ( ) appears, the weighing value forthe second component (CMP002) will be issued automatically tothe optional printer. This is followed by automatic taring.

    0.0000 g

    Weigh third component 5.0010 gWhen the last single weighing process is finished, press the

    [ON/OFF] key to receive the total sum ("TOTAL=").15.0030 g

  • 7/28/2019 ABT-BA-e-0911

    61/71

    ABT-BA-e-0911 61

    Printout:---ADDON Mode---

    CMP001=5,0003 g

    CMP002=5,0017 g

    CMP003=5,0010 g

    TOTAL=15,0030 g

    Note: Only one function at a time may be activated (on) when working with func-tions auto print, add-on and recipe.Whilst the function is active, automatic adjustment (PSC/Clock-CAL) does

    not take place. A flashing symbol, however, indicates that adjustment isrequired.

    14.6 Recipe mode

    The formula function allows to add on various components of a mixture. The totalweight of all components can be invoked to check (separate memory for the weightof tare vessel and recipe components).

    Activate function "FormU: on (see chap. 14.5 Add up):Whilst the recipe function is active, the display keeps showing the "FormU -on" sym-

    bol [ M].

    Operation Display

    When using a container, place this on the weighing plateand tare. 0.0000gWeigh first component. 0.5361gWhen standstill control ( ) is complete, the weighingvalue of the first component (CMP001) is issued to the

    optional printer by pressing the [PRINT] key. The dis-played value is added into the total adding memory. Af-terwards automatic taring will take place.

    0.5361g

    0.0000gWeigh second component 0.5422gWhen standstill control ( ) is complete, the weighingvalue of the second component (CMP002) is issued tothe optional printer by pressing the [PRINT] key. The dis-played value is added into the total adding memory. Af-

    terwards automatic taring will take place.

    0.54221g

    0.0000g

  • 7/28/2019 ABT-BA-e-0911

    62/71

    ABT-BA-e-091162

    Place further components 0.4488gWhen the last single weighing process is complete press

    [ON/OFF] key to receive the total sum (Total).1.5271g

    Remove weight from weighing plate. The balance is nowready for further measurements.

    Printout:---Formulation Mode---

    CMP001=0,5361 g

    CMP002=0,5422 g

    CMP003= 0,4488 gTOTAL=

    1,5271 g

    Note: Only one function at a time may be activated (on) when working with func-tions auto print, add-on and recipe.Whilst the function is active, automatic adjustment (PSC/Clock-CAL) does

    not take place. A flashing symbol, however, indicates that adjustment isrequired.

  • 7/28/2019 ABT-BA-e-0911

    63/71

    ABT-BA-e-0911 63

    15 Data output

    15.1 RS 232C interface

    Pin allocation of balance output plug:

    PIN No. Signal

    2 TXD

    3 RXD

    6 DSR

    7 SG

    20 DTR

    5 CTS

    4 RTS

    15.2 Data Formats

    Note:9 space character and DL the delimiter code.

    1. Input Data FormatCOMMAND CODE + DL (see 15.2.3)

    2. Output data format In weighing mode:

    (ex.) S -200.000g9DLstability polarity

    Polarity dataPositive: space character (()Negative: minus (-)

    Stability information (when output includes stability information)Stable : SUnstable : U

    In oL or -oL display

    (ex.) U -9999oL9999DLstability polarity

    Polarity dataPositive: space character (9)Negative: minus (-)

    Stability data (when output includes stability information)Stable : SUnstable : U

    3. Data format ASCII (JIS) code

    Baud rate, parity (and bit length), delimiter, stop bit, format, and handshakediffer by menu item selections.

  • 7/28/2019 ABT-BA-e-0911

    64/71

    ABT-BA-e-091164

    15.3 Remote control instructions

    Attention:Inputting characters and command codes not shown here into the balance may notonly alter the previous settings but may also impair proper measurement.If by mistake characters or commands not shown here are entered into the balance,immediately unplug the power supply cable and wait about ten seconds before plug-ging again.

    Command Code Function Description

    D01 Continuous output The balance continuously outputs every 110ms .

    D05 1 time output Corresponds to [PRINT] key

    D06 Auto Print See 14.4

    D071 time output with stability in-

    formation

    The status of the stability mark is appended to the head

    of the data with output.

    S: when the standstill display appears

    U: when standstill display does not appear

    D08 1 time output at stabilityAfter command input, the data are output at the first

    appearance of the standstill display.

    D09 Halt output Auto Print and continuous output halted

    Q ON/OFF switchingSwitches between standby status and measurement

    status.

    T Taring Corresponds to the [O/T] key

    TS Taring after stability waitAfter command input, taring is done at the first appear-

    ance of the stability mark.

    C18 Measurement span calibration

    + Measurements in add-up mode See ch. 14.5

    R Total reset All application measurements terminated and reset

    mg mg unit

    PERCENT Percent determination

    PCS Parts counting

    CT ct unit

    SDENCE Density of solids

    LDENCE Density of liquids

    % 100% setting

    G g, % switching

    - g g unit removal

    - mg mg unit removal

    - PERCENT

    Deleting percentage determina-

    tion t

    - PCS Piece counting removal

    - CT ct unit removal

    - SDENCE Solid specific gravity removal

    - LDENCE Liquid specific gravity removal

  • 7/28/2019 ABT-BA-e-0911

    65/71

    ABT-BA-e-0911 65

    C02 High-stability mode setting

    C13 Anti-convection mode setting

    C14 Standard mode setting

    C05

    Standstill display,

    setting to 1 count

    C06Standstill display,

    setting to 5 count

    C15Standstill display,

    setting to 10 count

    C07 Auto Zero on

    C08 Auto zero off

    C10 Autom. CAL on

    C11 Autom. CAL offC17 Display setting status

    Measurement conditions set by menu selection in abbre-

    viated form is output.

    15.4 Standard Settings

    Displayfor menuselection

    BaudRate

    DelimiterParity(Bit

    length)Stop bit

    Dataformat

    Hand-shake

    Standard set-ting 1

    iF:F1 1200 C/R None (8) 1 dF1 Hardware

    Standard set-ting 2

    iF:F2 1200 C/R None (8) 1 dF2 Hardware

    Standardsetting 3

    iF:F3 2400 C/R+L/F Even (7) 1 dF3 Hardware

    Standard set-ting 4 iF:F4 1200 C/R+L/F Odd (7) 1 dF4 Hardware

    Operatorsettings

    (see 15.5)iF:USEr

    Usersettings

    Usersettings

    Usersettings

    Usersettings

    Usersettings

    User set-tings

    KERNYKB-01N

    iF:USEr 1200 C/R None (8) 1 dF1 off

    Selecting one of standard Settings:

    Repeatedly press the [CAL] key until "intFACE" appears.

    Press [TARE] key

    Repeatedly press the [CAL] key until the desired standard settingappears.

    Confirm your selection by pressing the [TARE] key.

    Repeatedly press the [ON/OFF] key. The balance returns to toler-ance weighing mode.

  • 7/28/2019 ABT-BA-e-0911

    66/71

    ABT-BA-e-091166

    15.5 User Settings

    The user setting allows individual setting for each item in communication settings.

    Repeatedly press the [CAL] key until "intFACE" appears.

    Press [TARE] key

    Repeatedly press the [CAL] key until "iF:USEr" appears.

    (Example)Press [TARE] key

    Use the [CAL] key to select between the following settings(The **** reflect the current setting, two to four characters):

    io.b:**** Speed of communication

    io.d:**** Delimiter

    io.P:**** Parity

    io.S:**** Stop bit

    io.F:**** Data format

    io.H:**** handshake

    Confirm your selection by pressing the [TARE] key and set theparameters you require as follows:

  • 7/28/2019 ABT-BA-e-0911

    67/71

    ABT-BA-e-0911 67

    15.5.1 Communication speed settings

    1. The display changes from io.b:**** to b-300. Pressing the [CAL] key changes thedisplay. The standstill display ( ) marks the current setting.

    Display during setting b-300 b-600 b-1200 b-2400 b-4800

    Setting specifics300bps

    Baud Rate600bps

    Baud Rate1200bps

    Baud Rate2400bps

    Baud Rate4800bps

    Baud Rate

    Display during setting b-9600 b-19.2K b-38.4K

    Setting specifics9600bpsBaud Rate

    19.2KbpsBaud Rate

    38.4KbpsBaud Rate

    2. Confirm your desired setting by pressing the [TARE] key.

    3. To return to "io.S:****", press the [ON/OFF] key.

    15.5.2 Delimiter settings

    1. The display changes from io.d:**** to d-Cr. Pressing the [CAL] key changes thedisplay. The standstill display ( ) marks the current setting.

    Display during setting d-Cr d-LF d-CrLF d-Cn d-win 1 d-win -

    Setting specifics C/R L/F C/R + L/F Comma Not documented

    2. Confirm your desired setting by pressing the [TARE] key.

    3. To return to "io.d:****", press the [ON/OFF] key.

    15.5.3 Parity settings

    1. The display changes from io.P:**** to P-no. Pressing the [CAL] key changes thedisplay. The standstill display ( ) marks the current setting.

    Display during setting P-no P-odd P-EvEn

    Setting specificsNo parity

    (eight bits)

    Odd parity

    (seven bits)

    Even parity

    (seven bits)

    2. Confirm your desired setting by pressing the [TARE] key.

    3. To return to "io.P:****", press the [ON/OFF] key.

  • 7/28/2019 ABT-BA-e-0911

    68/71

    ABT-BA-e-091168

    15.5.4 Stop bit settings

    1. The display changes from io.S:**** to S-S1. Pressing the [CAL] key changesthe display. The standstill display ( ) marks the current setting.

    Display during setting S-S1 S-S2

    Setting specifics Stop bit, 1 bit Stop bit, 2bit

    2. Confirm your desired setting by pressing the [TARE] key.

    3. To return to "io.S:****", press the [ON/OFF] key.

    15.5.5 Input-output data format settings

    1. The display changes from "io.F:****" to " F-dF1". Pressing the [CAL] key changesthe display. The standstill display ( ) marks the current setting.

    Display during setting F-dF1 F-dF2 F-dF3 F-dF4

    Setting specifics

    Data format 1.

    Standard format

    Data format 2.

    Not documentedData format 3.

    Not documented

    Data format 4.

    Not documented

    2. Confirm your desired setting by pressing the [TARE] key.

    3. To return to "io.F:****", press the [ON/OFF] key.

    Note: When set to data format 2, the balance will always send a process resultagainst commands from the computer.

    15.5.6 Handshake settings

    1. The display changes from "io.H:****" to " H-oFF". Pressing the [CAL] key changesthe display. The standstill display ( ) marks the current setting.

    Display during setting H-oFF H-Soft H-HArd H-tr

    Setting specificsNo handshake

    Software

    handshake

    Hardware

    handshake

    Timer

    handshake

    2. Confirm your desired setting by pressing the [TARE] key.

    3. To return to "io.H:****", press the [ON/OFF] key.

  • 7/28/2019 ABT-BA-e-0911

    69/71

    ABT-BA-e-0911 69

    16 Service, maintenance, disposal

    16.1 Cleaning

    Before cleaning, please disconnect the appliance from the operating voltage.

    Please do not use aggressive cleaning agents (solvents or similar agents), but a clothdampened with mild soap suds. Ensure that no liquid penetrates into the device andwipe with a dry soft cloth.Loose residue sample/powder can be removed carefully with a brush or manual vac-uum cleaner.

    Spilled weighing goods must be removed immediately.

    16.2 Service, maintenance

    The appliance may only be opened by trained service technicians who are authorizedby KERN.Before opening, disconnect from power supply.

    16.3 Disposal

    Disposal of packaging and appliance must be carried out by operator according to

    valid national or regional law of the location where the appliance is used.

  • 7/28/2019 ABT-BA-e-0911

    70/71

    ABT-BA-e-091170

    17 Instant help

    In case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.

    Error code table:

    Error code display Explication Remedy

    CAL E2High zero point shift dur-ing adjustment

    Remove items from theweighing pan.

    CAL E3Great deviation of meas-uring values in PCAL.

    Use correct adjustingweight.

    CAL E4Great deviation of meas-uring values during ad-justment

    CHE X (X is a numeral)(when the display stops

    here)

    Internal malfunctionPlease contact a servicerepresentative.

    Err 0X (X is a numeral) Internal malfunctionPlease contact a servicerepresentative.

    Err 20An improper value settingwas attempted.

    Enter the correct numbersor decimal points.

    Err 24 Power voltage error Check the power voltage .

    Possible causes of errors:

    When Symptom Possible causes Remedy

    Beforemeasure-ment

    Nothing appears inthe display.

    The AC adapter is not securelyconnected. The power switchboardof the room is turned off. The powervoltage is incorrect.

    Check power supply andconnect correctly.

    Duringmeasure-ment

    The display fluctuates.

    The standstill displaydoes not appear fastenough.

    The measured resultshave poor repeatabil-ity.

    CAL d appears fre-quently.

    Vibrations or draft. Change the installation site.Change stability and reac-tion settings or standstilldisplay.

  • 7/28/2019 ABT-BA-e-0911

    71/71

    Attempting to measure volatile sub-stances

    Measure with a lid on it.

    The weighed item is electricallycharged.

    Measure in a metal con-tainer. Measure with a metalobject larger than the items.

    Sample temperature and tempera-

    ture inside weighing space differ.

    Measure at the same tem-

    perature. Leave the item inthe chamber before meas-urement. Change to high-stability mode.

    Draft in weighing space. Leave the glass doors of theweighing chamber open 1 to2 cm wide when not in use.

    Effects of electronic noise or strongelectromagnetic waves

    Move away from the noisesource.

    Internal trouble with the balance Contact a service represen-tative.

    oL or -oL is dis-played

    The load on the weighing pan is toolarge. The weighing pan is de-tached.

    Use within the weighingcapacity. Place the weighingpan on correctly.