allengers venus mammo

Upload: thiagarajan-kanagarajan

Post on 04-Nov-2015

16 views

Category:

Documents


0 download

DESCRIPTION

Mammogram

TRANSCRIPT

  • ALLENGERSMAMVENUS(Highfrequency(40KHz)mammographyunit)

    SpecialFeatures AutomaticfilterselectionwithKV.Automatic release of compression with termination ofexposureincaseofpowerfailure.DisplayofmAsinAECmodeafterexposure.

    A) XRAYGENERATOR

    Xraygenerator Highfrequency40KHz.xraygenerator

    Maximumpower 3.5KW*

    MaximummAoutput 70mA

    MaximumKVoutput 35KV

    Highvoltagecable 1No.

    B) XRAYTUBE

    Xraytube Rotatinganodexraytube.

    Focalspot Smallfocus=0.1mm Largefocus=0.3mm.

    Targetmaterial Molybdenum.Tubeangle 12

    Inherentfiltration. Ba0.5mm

    Anodeheatstoragecapacity 300KHU

    Anoderotation 2700RPM

    C) CONTROLPANEL

    Controlpanel Micro processor controlled feather touch control panelwithLCD

    KVrange 22to35KVinstepsof0.5KVeach.

    mAsrange 1mAsto700mAs

    Techniqueselection x Manualtwopointtechnique(KV,mAs).

    3.5 KW VENUS SERIES HIGH FREQUENCY MAMMOGRAPHY SYSTEM

  • Anatomicprogrammedradiography x Auto selection of exposure factors with APR forsmall,mediumandlargebreasts.

    Filmscreencombination x 5filmscreencombinationsareprovided.Automaticexposurecontrol(AEC) x InbuiltthreechamberAECforbetterexposurecontrol

    anddiagnosis

    Filmdensitycontrol x 5stepfilmdensitycontrolisprovidedFilterselection x FilterisautomaticallyselectedaspertheKVselected

    (molybdenumfilter(for29.5to35KV)andaluminiumfilter(for22to29KV)

    Digitaldisplays LCDdisplayoncontrolpanelx KVx mAsx Interlocksindicatingthefaultinthemachinex FilterselectedLCDdisplayonthestandx CompressionforceinKgx Compressedbreastthicknessx Degreescale

    Switches x Focalspotselectionswitch.x Machineon/offswitch.x Xrayswitch.x Emergency stop switches mushroom type

    emergency switch provided on both sides of C.This switch releases the breast by moving thecompressionpaddleincaseofanypowerfailure

    Indicators x XrayONindicator.D) STANDASSEMBLY

    Acompactstandonwhichcarmcontainingxraytubeandbuckyassemblyismounted.

    Verticalmovement(motoroperated) 550mm

    Angularmovementofxraytubeandbucky

    180

    Focusfilmdistance 650mm.

    Breastcompression Motorised compression upto 15Kg with the switch foractivationandrelease.Adjustablecompressionforce.

    Compressionpaddles Compressionpaddles for normal andmagnificationmode(spotcompression).Compressionscale:0to10cm.

  • Magnificationdevice 1.8X

    Bucky 18x24cmbucky,motoroperatedoscillatinggridofsize18X26cm,5:1,30lines/cmfocaldistance60to70cm.

    Collimator Light beam collimatorwith halogen lampwith auto shutofftimer

    Collimationplate. 18X24cmcollimationplate.

    Cone Coneforlocalizationandradiationprotection.

    AutomaticExposureControl(AEC) InbuiltthreechamberAECforbetterexposurecontrolanddiagnosis

    CArmmovement. Switches for up/down movement of C Arm, placedconvenientlyonbothsidesofCArm.

    Exposureswitch Hand switch with retractable cord for initiation ofexposure.

    Filmmarkingdevice. Filmmarkingdevice.Alphanumericidentificationsystem.

    Protectivebarrierforoperator Free standing fully transparent lead glass screen foroperatorprotection.

    E) POWERSUPPLYREQUIREMENT

    PowerSupply Singlephase,230volts10%,AC,50Hz,15amps.withindependentearthingonthewallsocket.

    OPTIONALACCESSORIES&DEVICES.

    x 24x30cmbuckywithgrid(motoroperated).x 24x30cmcollimationplateandcompressionpaddle.x Rhodiumfilter.x Stereotaticbiopsysystem

    ForALLENGERSMEDICALSYSTEMSLTD.

    AUTHORISEDSIGNATORY

    MEDILINES LIFECARE TECHNOLOGY