ALLENGERSMAMVENUS(Highfrequency(40KHz)mammographyunit)
SpecialFeatures AutomaticfilterselectionwithKV.Automatic release of compression with termination ofexposureincaseofpowerfailure.DisplayofmAsinAECmodeafterexposure.
A) XRAYGENERATOR
Xraygenerator Highfrequency40KHz.xraygenerator
Maximumpower 3.5KW*
MaximummAoutput 70mA
MaximumKVoutput 35KV
Highvoltagecable 1No.
B) XRAYTUBE
Xraytube Rotatinganodexraytube.
Focalspot Smallfocus=0.1mm Largefocus=0.3mm.
Targetmaterial Molybdenum.Tubeangle 12
Inherentfiltration. Ba0.5mm
Anodeheatstoragecapacity 300KHU
Anoderotation 2700RPM
C) CONTROLPANEL
Controlpanel Micro processor controlled feather touch control panelwithLCD
KVrange 22to35KVinstepsof0.5KVeach.
mAsrange 1mAsto700mAs
Techniqueselection x Manualtwopointtechnique(KV,mAs).
3.5 KW VENUS SERIES HIGH FREQUENCY MAMMOGRAPHY SYSTEM
-
Anatomicprogrammedradiography x Auto selection of exposure factors with APR forsmall,mediumandlargebreasts.
Filmscreencombination x 5filmscreencombinationsareprovided.Automaticexposurecontrol(AEC) x InbuiltthreechamberAECforbetterexposurecontrol
anddiagnosis
Filmdensitycontrol x 5stepfilmdensitycontrolisprovidedFilterselection x FilterisautomaticallyselectedaspertheKVselected
(molybdenumfilter(for29.5to35KV)andaluminiumfilter(for22to29KV)
Digitaldisplays LCDdisplayoncontrolpanelx KVx mAsx Interlocksindicatingthefaultinthemachinex FilterselectedLCDdisplayonthestandx CompressionforceinKgx Compressedbreastthicknessx Degreescale
Switches x Focalspotselectionswitch.x Machineon/offswitch.x Xrayswitch.x Emergency stop switches mushroom type
emergency switch provided on both sides of C.This switch releases the breast by moving thecompressionpaddleincaseofanypowerfailure
Indicators x XrayONindicator.D) STANDASSEMBLY
Acompactstandonwhichcarmcontainingxraytubeandbuckyassemblyismounted.
Verticalmovement(motoroperated) 550mm
Angularmovementofxraytubeandbucky
180
Focusfilmdistance 650mm.
Breastcompression Motorised compression upto 15Kg with the switch foractivationandrelease.Adjustablecompressionforce.
Compressionpaddles Compressionpaddles for normal andmagnificationmode(spotcompression).Compressionscale:0to10cm.
-
Magnificationdevice 1.8X
Bucky 18x24cmbucky,motoroperatedoscillatinggridofsize18X26cm,5:1,30lines/cmfocaldistance60to70cm.
Collimator Light beam collimatorwith halogen lampwith auto shutofftimer
Collimationplate. 18X24cmcollimationplate.
Cone Coneforlocalizationandradiationprotection.
AutomaticExposureControl(AEC) InbuiltthreechamberAECforbetterexposurecontrolanddiagnosis
CArmmovement. Switches for up/down movement of C Arm, placedconvenientlyonbothsidesofCArm.
Exposureswitch Hand switch with retractable cord for initiation ofexposure.
Filmmarkingdevice. Filmmarkingdevice.Alphanumericidentificationsystem.
Protectivebarrierforoperator Free standing fully transparent lead glass screen foroperatorprotection.
E) POWERSUPPLYREQUIREMENT
PowerSupply Singlephase,230volts10%,AC,50Hz,15amps.withindependentearthingonthewallsocket.