An Integrated Microfluidic D evice for the Continuous Sam pling and Analysis of B iological Fluids Author Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert P.Pisano and Dorian Liepmann Reporter: Wun-Hao Wu ( Wun-Hao Wu ( 吳吳吳 吳吳吳 ) ) 10/17, 2007

Post on 21-Dec-2015

219 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

An Integrated Microfluidic Device for the Continuous Sampling and Analysis

of Biological Fluids

Author: Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert P.Pisano and Dorian LiepmannReporter: Wun-Hao Wu (Wun-Hao Wu (吳文豪吳文豪 ))

10/17, 2007

Page 2: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

Abstract

First: introduce the device Second: the device work principle Third: question

Page 3: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

1.

2.

3.

4.

Packaged Microneedle

4.

4.

1. device

Page 4: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

Microneedle

It is permeable to small molecular weight (MW) molecules, yet excludes large MW compounds such as proteins.Why they devise this some holes (they call microdialysis membrane)? Because, it can reduce large molecular weight proteins adsorbing on the surface of the glucose sensor.

Page 5: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

Fabrication for Needles

back

Page 6: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

back

Gate valve

First: Left Electrolysis bubbles

are opened.

Second: Right Electrolysis bubbles are opened and Left Electrolysis bubbles are closed.

Then : Right bubbles are more

and more.

Third: Left Electrolysis bubbles are opened again.

Then : Left bubbles are more and more.

One entire cycle, from fully open to fully closed an back to fully open, takes on average around 120 seconds.

Page 7: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

back

back

Glucose sensor

Page 8: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

Multichannel Pumping

Pump 2

1.

2.

3.

1.

2.

3.

左邊十字微結構當水從左邊來時,可阻擋水通過;右邊方塊形結構當水從右邊來時可以推開方塊而通過。

Pump 2

Valve opens

work

Page 9: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

2. work principle

1. 首先從人的血管取得血液進入微探針裡,經過微結構設計過濾掉一些大分子量的蛋白質。

2. 待感測的液體進入葡萄糖感測器,量出電流的變化後。

3. 啟動 pump(1) 且 gate valve 打開,將 PBS 注入葡萄糖感測器中清洗。

4. 關閉 pump(1) 打開 pump(2) 且 gate valve 打開將葡萄糖溶液注入到葡萄糖感測器中恢復到原來的

濃度。

5. 最後 pump(3) 將胰島素經過探針進入人體血液中。

PBS(pump1)

Stock concentration of glucose(pump2)

Insulin(pump3)

Page 10: An Integrated Microfluidic Device for the Continuous Sampling and Analysis of Biological Fluids Author : Jeffrey D.Zahn, Ajay A Deshmukh,Alexandros P.Papavasiliou,Albert

3. Question

The blood is in the needle’s channel ,the proteins can adsorb on the microdialysis membrane. That proteins maybe block the microfluidic channel. It doesn’t tell us how to remove these proteins.

The Gate Valve creates the actuation bubbles. Maybe bubbles can move into the microfluidic channel. They also block it.

They don’t tell us this chip how to work. I suspect that some device can’t work. If they have some picture to show that action are step by step. I will believe it.