    HIGHLIGHTS OF MECHANICALVENTILATION UNIT 4

    Modes and initiation of ventilationBy

    Elizabeth Kelley Buzbee AAS, RRT-NPS

  • 8/12/2019 Bipap y Aprv

    2/77

    The modes of ventilation: A spontaneous breathis one that the patient

    triggers and cycles the breath, and he controlsthe VT. This breath could be assisted by theapplication of positive pressure.

    A mandatory breathis defined as one that istriggered and cycled by the machine. All

    mandatory breaths are assisted breaths.

  • 8/12/2019 Bipap y Aprv

    3/77

    The modes of ventilation: full

    support modes CMV: continuous mandatory ventilationin which

    all breaths are mandatory.

    VC-CMVvolume control also called

    Assist/Control mode Set VT, f to get VE; guaranteed VT

    Default ventilatory mode for full support with adults

    PC-CMVpressure control mode. Patient can

    trigger breaths just like with A/C Set PIP, f and TI no guaranteed VT

    Default ventilator mode for full support for infants

  • 8/12/2019 Bipap y Aprv

    4/77

    Indications for PC: the RCP selects

    pressure ventilation when:

    The adult patient who cannot be managed withVC In this case, we keep the PIP less than 30cmH20.

    PC results in better distribution of ventilation inpersons with unequal RAW, but consistentcompliances.

    There is such an airway leak so that the VTareunstable [most common with infants and smallchildren with uncuffed ET or tracheostomy tubes]

  • 8/12/2019 Bipap y Aprv

    5/77

    Compare PC to VC

    In PC, the airway pressures; mPAW and PIPwill stay the same, but the VE and VT can varybased on patients time constants

    In VC the VE and VT are basically stable[patient can increase f so VE could vary] thePIP and the mPAW can be altered by patient

    time constants

  • 8/12/2019 Bipap y Aprv

    6/77

    Compare control mode to

    Assist/control We control patients by giving them sedation

    and paralytic agents so that the VE we set onVC-CMV is exactly the same

    We can control their PaC02 thus their acidbase balance

    In A/C, the patient can trigger breaths thatwill increase the VE, so that the VE based onset VT and f could be lower than the actualmeasured VE

  • 8/12/2019 Bipap y Aprv

    7/77

    Controlling the chronic

    hypercapnic patient If your patient has a hypoxic drive,

    administrate enough Fi02to get his Pa02between 80-100 mmHg.

    This will result in apnea and works as a formof sedation in the first 24 hours.

    Must wean the Fi02 to get Pa02 between 55-65 mmHg before weaning

  • 8/12/2019 Bipap y Aprv

    8/77

    Problems with A/C

    Excessively high PAWcan cause problems withhemodynamics once patient starts to breath.

    Another problem with A/C mode is the risk ofauto-PEEP and air trapping.

  • 8/12/2019 Bipap y Aprv

    9/77

    Inverse Ratio Ventilation [with

    PC or with VC] This is a form of full support that uses

    increased Ti to raise the mPAW whenpatients compliance is so bad that PIP andPplateau are excessive

    In IRV, the expiratory time is so short that thepatient never completely exhales. This works

    like PEEP to recruit alveoli

  • 8/12/2019 Bipap y Aprv

    10/77

    Raising mPAW with IRV

    mPAW = PIP [I] + PEEP [E]

    I + E

    Because we raise the inspiratory time somuch we can decrease the PIP

    Because we create auto-PEEP with the shortTE, we can decrease the PEEP

  • 8/12/2019 Bipap y Aprv

    11/77

  • 8/12/2019 Bipap y Aprv

    12/77

    Problems with Negative pressure

    Ventilation: patient must be able to:

    protect airway

    Handle being supine all the time

    hemodynamically stable

    be comfortable in one position all the time

    handle being disconnected from vacuum for

    short time spans

  • 8/12/2019 Bipap y Aprv

    13/77

    More problems with NPV

    Patient can get skin lesions from movementof body inside the device

    Patient can get cold from wind Best 02 device is nasal cannula because 02

    can be sucked into the neck opening

  • 8/12/2019 Bipap y Aprv

    14/77

    NPV

    Classified as controllers, but newer modelscan be A/C if there is a flow sensor placed onthe patients nose

    Old metal iron lungs have a constant I:E of I:I;newer fiberglass devices can have altered I:Eratios

  • 8/12/2019 Bipap y Aprv

    15/77

    Setting parameters on NPV:

    Change level of the vacuum to increase the VT[he could use a Wrights spirometer attachedto an IPPB mask to measure exhaled V

    T

    ]

    Change the respiratory rate.

  • 8/12/2019 Bipap y Aprv

    16/77

    CSV

    continuous spontaneous ventilationin whichall breaths are spontaneous.

    patient who can completely control his VE and only needs a little help such as with

    increased baseline pressures [CPAP]

    or some application of assisted breaths suchas pressure support [PS]

    or who might require monitoring of VE

  • 8/12/2019 Bipap y Aprv

    17/77

    Pressure support ventilation

    PSV is the most common form of pressurecycled CSV.

    Although this does raise the airway pressure sothat we have a higher and lower pressure, wecall this PS rather than PIP because of thespecific characteristics of PS

    Flow triggered and flow cycled Patient controls his VT, f and inspiratory time

  • 8/12/2019 Bipap y Aprv

    18/77

    Indications for PS:

    When used with SIMV to reduce the WOB byincreasing the spontaneous VT. We generallyselect the PS that will deliver a reasonable VT

    [watch the spontaneous RR] Can be used alone during weaning. Once a

    patient is on a PS of 5-10 cmH20, he is

    considered at a level that only compensatesfor RAW of the tubing, so is consideredconsistent with spontaneous breathing.

  • 8/12/2019 Bipap y Aprv

    19/77

    PSV flow patterns

    The flow pattern is descending till it reaches5 LPM [or 25% of the peak flow] in which theflow stops abruptly.

    The flow slows down as the device attemptsto keep the PS at the preset pressure.

  • 8/12/2019 Bipap y Aprv

    20/77

    VT on PS

    There is no guaranteed VT, nor VE, but we canincrease the VT by increasing the PS pressure

    We need to set VE & high f alarms closely towarn us of problems

    The patient sends more air to Zone IIIbecause he is using his diaphragm more with

    PS

  • 8/12/2019 Bipap y Aprv

    21/77

    To choose the correct level of PSV

    there are three methods: get an appropriate VT [10-15 ml/kg] and titrate the PS level toachieve this VT

    increase the PS level till the respiratory rate is normalized [25

    bpm or less]

    increase the PSV until you decrease the work of breathingthrough the ET tube

    To select the appropriate level of PSV to overcome the R

    AWuse this

    formula

    PSV= (PIP - Pplateau) x spont insp. Flow rate [l/sec]

    Ventilator flow rate [l/sec]

  • 8/12/2019 Bipap y Aprv

    22/77

    PSmax

    or straight pressure support or stand alone PS [PS without SIMV.] In this case, the PS is not usedas a weaning modality but for initial ofmechanical ventilation.

    We generally select a PS level that will deliver 10-12 ml kg IBW.

    The RCP must remember that this mode is anassist only and the patients VTand VE will vary

    base on lung dynamics. There is no guaranteedVT.

    Patient must have an intact ventilatory drivefor this to work

  • 8/12/2019 Bipap y Aprv

    23/77

    CPAP mode

    spontaneous mode

    application of PEEP without any positivepressure breathes.

    CPAP is merely a raised baseline with a flow rate

    with adjustable Fi02 recruits alveoli which will improve diffusion of 02 CPAP can help return a low compliant lung back

    to normal once atelectasis has been resolved.The FRC should rise.

    should decrease WOB.

    proper application of CPAP should decreaseWOB- watch respiratory rates on this

  • 8/12/2019 Bipap y Aprv

    24/77

    CPAP interfaces

    CPAP via the ET tube or a trach tube is calledCPAP

    CPAP via a nose mask, face mask or full facemask is called nasal-CPAP [n-CPAP]

    Obviously we select the interface based onthe patients ability to protect his airway

  • 8/12/2019 Bipap y Aprv

    25/77

    n-CPAP indications

    The successful candidate for n-CPAP would

    be the patient who is oriented, has good ventilatory drive without excessive

    WOB

    and who has the ability to protect his airway.

  • 8/12/2019 Bipap y Aprv

    26/77

    n-CPAP contraindications

    Persons at risk for vomiting and aspiration

    persons with skin necrosis,

    claustrophobia.

  • 8/12/2019 Bipap y Aprv

    27/77

    CPAP indications

    Management of the person who is in hypoxemia respiratory failure.This patient will have refractory hypoxemia without respiratoryacidosis..

    Treatment of Congestive Heart Failure[CHF] in the patient whohas an intact ventilatory drive and can keep his PaC02down. CPAP of8-12 with Fi02 100% is suggested. [Egans pp, 1095]

    A weaning modalityThis invasive CPAP may be the last step beforeextubation. Generally a patient can be extubated from a CPAP of 5-7cmH20 [or can be extubated at a stand-alone PSV of 5-7 cmH20.

    Non-invasive management of persons with obstructive sleep apnea[OSA

  • 8/12/2019 Bipap y Aprv

    28/77

    APRV

    a spontaneous mode

    airway pressure release ventilation

    Patient is breathing on two different levels ofCPAP

  • 8/12/2019 Bipap y Aprv

    29/77

    Initial settings for APRV for

    ARDS: The higher CPAP is set with the Phigh, while the Plowsets the

    lower pressure.

    The RCP should also set the time interval [Thigh] for Phighand the time interval [Tlow]for Plow

    To initial APRV, the RCP looks to the patients PplateauonPPV and uses that figure for the Phigh.

    The Thighis started at 4 seconds for adults and can beprogressively increased to 10-15 seconds

    Set the Plow at zero and use the release time [Tlow] to keep

    the pressure from dropping to zero Set the Tlow at about.5 to .8 [one time constant] so that the

    breath ends with the expiratory flow at 50-75% of peak flow

  • 8/12/2019 Bipap y Aprv

    30/77

    What happens if the patient

    goes apnic?

    During APRV ventilation if the patient wasstop breathing, the time-cycling betweenhigh and low pressures would appear similarto PC-IRV.

    So this is a spontaneous mode that happensto have a back up of sorts

  • 8/12/2019 Bipap y Aprv

    31/77

    Contraindications to APRV

    persons with COPD or other problemsassociated with air trapping.

    persons with excessively high intracranialpressures [high ICP]

  • 8/12/2019 Bipap y Aprv

    32/77

    Bilevel ventilation

    An alternative to APRV is bilevel ventilation.The only difference between bilevelventilationand APRVis that the patientspends more time at the [Plow] lower airwaypressure than at the high airway pressure[Phigh].

  • 8/12/2019 Bipap y Aprv

    33/77

    BiPap- NIPPV

    Non-invasive positive pressure ventilation

    These BiPap breathes tend to be flow or time

    triggered, flow cycled off with the operator selecting PIP [called IPAP]

    and PEEP [called EPAP] and bleeding insupplementary 02.

    The newer Visioncan get a Fi02. http://emedicine.medscape.com/article/1417959-treatment

    http://emedicine.medscape.com/article/1417959-treatmenthttp://emedicine.medscape.com/article/1417959-treatmenthttp://emedicine.medscape.com/article/1417959-treatmenthttp://emedicine.medscape.com/article/1417959-treatment
  • 8/12/2019 Bipap y Aprv

    34/77

    contraindications/hazards of NIPPV

    do not put this device on an apnic patient because itis NOT a ventilatorit is a breath augmenter.

    Persons who cannot protect their airways

    Hemodynamically unstable patients Facial burns or trauma

    Uncooperative patients

    Persons at risk for aspiration: vomiting, nose bleeds,

    unconscious, poor gag reflex

    Copious secretions

    Anatomical problems with gas exchange

  • 8/12/2019 Bipap y Aprv

    35/77

    Indications for NIPPV: acute

    care of:

    congestive heart disease [n-CPAP or BiPap]

    COPD patient who doesnt want to be

    intubated recently extubated patient who is at risk of

    failing.

    immune-suppression for whom we may notwant to risk VAP

  • 8/12/2019 Bipap y Aprv

    36/77

    Indications for long-term NIPPV

    Long-term management of both obstructivesleep apnea and central sleep apnea

    Long-term management of patients withskeletal or neuromuscular disorders

    Long-term management of the COPD patientwho has s/s of chronic hypoventilation

    [especially at night] and who is optimallytreated with drugs and other care.

  • 8/12/2019 Bipap y Aprv

    37/77

    Initial settings for BiPap:

    IPAP at 8 cmH20 and EPAP at 4 cmH20.

    . Increase IPAP in increments of 2 cmH20 todeliver more VT.

    To hypoxemia, increase the EPAP in incrementsof 2 cm H20.

    Oddly enough, if the EPAP is raised withoutraising the IPAP, the VT might decrease becausethe VTis a function of the change in pressure orthe delta P [ P]

  • 8/12/2019 Bipap y Aprv

    38/77

    The BiPap ST/D

    EPAP/CPAP: in this mode, all you get is CPAP

    IPAP: in this mode, again, all you get is CPAP.

    Spontaneous modethis is a form of PSV in which

    you select the PS with the IPAPand the PEEP withthe EPAP. All breaths are patient triggered

    Spontaneous/timed: is their version of A/C PCwith each breath patient or time triggered. In this

    mode you select the bpm Timed mode: their version of control ventilation

    in which you now select the rate and theinspiratory time

  • 8/12/2019 Bipap y Aprv

    39/77

    What is so strange about the

    BiPAP ST/D circuit? only a single, large-bore tubing going from

    the compressor to the patients mask.

    constant leak at the Whisper swivel thiswill leak a minimal amount of gas out of thecircuit and between the very high flow rates

    and the leak, the patient doesnt rebreathehis C02. Never plug up this hole!

  • 8/12/2019 Bipap y Aprv

    40/77

    Adding extra 02to the BiPap STD

    without starting a fire

    add 02at the mask,

    start machine first before adding 02so gaswill not leak back into machine

    never exceed 15 LPM

  • 8/12/2019 Bipap y Aprv

    41/77

    Compare the BiPap STD to the Vision

    BiPap machine

    The BiPAP ST/D has no 02inlet

    The Respironics Visionplugs into 50 psig 02&can get 21% to 100% Fi0

    2

    The BiPAP ST/D has no internal alarm, youmust buy a separate alarm

    The Respironics Visioncan be used forinvasive ventilation with A/C, SIMV +PSV andCPAP as well as NIPPV [CPAP and S/T]

  • 8/12/2019 Bipap y Aprv

    42/77

    Use of critical care ventilators such as

    BiPap machines in the ICU.

    As a rule, we would operate these machinesin the PSV mode with PEEP to mimic theBiPap.

    It is important to understand that the alarmson these machines may have to be adjusted

    out of range

  • 8/12/2019 Bipap y Aprv

    43/77

    Dual modes

    combinemandatory ventilation withspontaneous ventilation

    IMV:intermittent mandatory ventilation: inwhich some breaths are mandatory and

    others are spontaneous. In this type of breath, the ventilator will give

    a PPV usually based on VC at timed intervals.The patient can breathe off a constant flowrate or from a demand valve at a VT and flowrate determined by his muscle strength,ventilatory drive and lung mechanics.

  • 8/12/2019 Bipap y Aprv

    44/77

    Advantages of IMV/ SIMV

    patient comfort

    maintains muscle coordination & musclestrength

    reduces V/Q mismatch;Zone III is beingutilized,

    [4] lower PAW and is an excellent weaning

    modality less likely to cause air-trapping

  • 8/12/2019 Bipap y Aprv

    45/77

    Disadvantage of IMV/ SIMV:

    If the patients PPV support is removed tooquickly the patient can suffer increased WOB

    We need to monitor the spontaneous VE , RRand VT, we may need to increase support by:

    increasing the SIMV rate

    adding PS

  • 8/12/2019 Bipap y Aprv

    46/77

    Indications for IMV/SIMV:

    IMV is a partial mode of ventilation thatusually includes dual modes.

    weaning from CMV when the patients

    ventilator muscles are weakened an initial ventilatorsetting when the patient is

    at risk for air trapping and is breathing on hisown,

    or if the patient who is able to breathe partiallyfor himself is at risk for decreased CO.

  • 8/12/2019 Bipap y Aprv

    47/77

    The difference between SIMV and IMV:

    SIMV stands for synchronized intermittentmandatory ventilation.

    The mandatory breath can come in sooner ifpatient triggers within the synchronizationwindowof fractions of seconds.

  • 8/12/2019 Bipap y Aprv

    48/77

    In apressure regulated volume controlmode,we are attempting to deliver the VT[becausewe are in VC mode] but we want to keep the

    airway pressures low.

    ventilator will attempt to deliver the VTat 5

    cmH20 below a preset pressure setting.

    Special modes: PRVC

    PRVC

  • 8/12/2019 Bipap y Aprv

    49/77

    Special modes: VAPS

    volume assured, pressure support,the ventilatorwill be attempting to deliver a stable VTwith PSbreaths so that the patient has the advantage of

    stable VEas well as the advantages of If a PS breath fails to reach the pre-set VT, the

    breath will continue at a constant flow until thevolume is reached. If the patient got the pre-set VT

    with the PS breath, it stays PS. Unlike normal PS, these breaths arent just flow

    triggered, but can be time triggered.

  • 8/12/2019 Bipap y Aprv

    50/77

    Special modes: MMV

    Mandatory minute ventilation gives the patient extra breaths or extra PS

    pressure to keep a predetermined minimal VE.

    This differs from apnea parameters in thatthe patient doesnt have to actually go apneic

    for 20 seconds or more for this to activate. Hemerely needs to have hypoventilation.

  • 8/12/2019 Bipap y Aprv

    51/77

    One problem with MMV

    when the patient starts the rapid, shallowbreathing associated with respiratorydistress.

    If a patient keeps the VEup with rate only, hecan be in a lot of distress

    It is suggested to keep the maximal high

    respiratory rate 10 BPM above the average

  • 8/12/2019 Bipap y Aprv

    52/77

    Special modes: ASV adaptive support ventilation: the RCP inputs

    the patients IBW and a percentage of the VE.

    The ventilator will deliver a VEbased on thepatients IBW.

    As the patient takes over more of the breathingthe VE is maintained with PS breaths.

    The level of PS changes to give the VT calculated

    by the machine, The VT will be determined bythe patients IBW and VD ventilation.

  • 8/12/2019 Bipap y Aprv

    53/77

    Special modes: PAV

    In proportional assist ventilation modesimilar to ASV in that the ventilator willcollect data about patients elasticity and

    resistance and flow or volume demands inorder to arrive at PS levels that varies.

  • 8/12/2019 Bipap y Aprv

    54/77

    High frequency ventilation

    controlled ventilation- the patient is sedatedand paralyzed

    VT of less or equal to the VDanatomical

    respiratory frequencies of 60 BPM-3600 bpm

    All HFV counts on the gas stream going downthe ET tube (inside) AT THE SAME TIME andthe gas flow existing (outside stream).

  • 8/12/2019 Bipap y Aprv

    55/77

    How does HFV work

    Penduluft action due to various timeconstants of different portions of the lungs,the gas moves from one lobe to another ,

    there is some bulk transfer

  • 8/12/2019 Bipap y Aprv

    56/77

    What are the types of HFV

    high frequency jet ventilation

    high frequency positive pressure

    ventilation

    high frequency oscillation

    combination of HFJ with CMV

  • 8/12/2019 Bipap y Aprv

    57/77

    Special modes: PRVC

    Inpressure regulated volume control, aneffort is made to maintain both a safe levelof airway pressure and delivered V

    T

    .

    In PRVC, the RCP selects a PIP that willnot be exceeded.

    To keep the VT, at this safe PIP, the

    inspiratory time and the flow rate mustvary.

  • 8/12/2019 Bipap y Aprv

    58/77

    Special modes:Auto-mode:

    in some ventilators selection of the auto-mode will allow the ventilator to decreasesupport as a patient starts to take over the

    WOB.

    The ventilator reverts between a CMV modeand a spontaneous mode based on breath by

    breath assessment of the patient

  • 8/12/2019 Bipap y Aprv

    59/77

    Special modes: ATC

    Automatic tubing compensation, in thismode the ventilator will compensate for theRAW of the ET tube.

  • 8/12/2019 Bipap y Aprv

    60/77

    INITIAL VENTILATOR SETTINGS

  • 8/12/2019 Bipap y Aprv

    61/77

    VT, set f and VE

    Full support A/C or SIMV rate 12-16 BPM

    Partial support SIMV below 10 BPM

    8-10 ml/Kg IBW normal lungs

    6-8 ml/kg IBW asthma

    5-8 ml/ kg IBW for ARDS & COPD

    VE needs to be 80-100 ml/KgIBW

  • 8/12/2019 Bipap y Aprv

    62/77

    TI and Flow rates

    Inspiratory flow ratesof 60-80 LPM for most

    If air hungry raise above 80

    COPD- 60-100 LPM

    Inspiratory times .80-1.2 seconds

  • 8/12/2019 Bipap y Aprv

    63/77

    Flow wave pattern

    Constant flowswill decrease inspiratory timeand help with I:E ratios, but can raise the PIP.

    Descendingflow curve has the advantage ofbetter distribution of gas into the lung, butwill increase the TIand increase the mPAW

    Sine wave: while considered more

    physiological, a classic sine wave may nothave enough initial flow to satisfy a patient.Like the descending flow pattern it will raise

    the TIand change the I:E ratio

  • 8/12/2019 Bipap y Aprv

    64/77

    Rise time or Ramp

    : in an effort to fine-tune flow patterns, theconstant flow can be damped by a rise timeadjustments. When set high, this almost

    mimics an ascending flow pattern.

  • 8/12/2019 Bipap y Aprv

    65/77

    Inspiratory pause

    The temporary use of the inspiratory pauseat about .5 to 1 second is generally reservedfor gathering Pplateau

  • 8/12/2019 Bipap y Aprv

    66/77

    Fi02

    100% is a good place to

    Weaning rapidly to 40-50% after ABG

    Fi02 needs to be weaned about 20% at a step.

  • 8/12/2019 Bipap y Aprv

    67/77

    PEEP

    may be started at zero, PEEP at 5 or lesscmH20 is considered physiologicaland shouldnot result in CV problems-but-- remember

    any PEEP that causes hemodynamicproblems is excessive.

    Increase or decrease by units of 2

  • 8/12/2019 Bipap y Aprv

    68/77

    Humidification by HME

    is limited to persons with good fluid balances,normal secretions and VEless than 10 LPMand normal body temperatures.

    If the patient has a gross leak so that 30% ofthe delivered VTis lost, the HME will notwork.

  • 8/12/2019 Bipap y Aprv

    69/77

    Humidification by heated

    humidifier can be used with everyone but are necessaryfor patients with secretions. Keep thetemperatures close to 330C +/- 2

  • 8/12/2019 Bipap y Aprv

    70/77

    Sighs:

    multiple sighs every hour or so. These sighvolumes were about 1.5 x the VT.

    important if VT is less than 7 ml/kg

  • 8/12/2019 Bipap y Aprv

    71/77

    s/p lung resection or lung

    transplants

    need lower VTand faster rates to protect the

    torn lung from rupture. Keep the Pplateauat or below 30 cmH20 [old

    Egans 1011]

  • 8/12/2019 Bipap y Aprv

    72/77

    lobar pneumonia:

    place patient on the good lung sideso gas goesto the bad lung

    avoid PEEP in lobar pneumonia if possible

    Try to prolong the Ti

    Consider double lumen ET tube so we can settwo ventilators on the patient

  • 8/12/2019 Bipap y Aprv

    73/77

    long-term neuromuscular

    patients more comfortable at higher than usual VT[decrease the RR] of 10-12 ml/kg. Thesepatients also tend to want higher flow rates.

    They can be managed with low Fi02-even .21as long as Sp02 is above 90-92%

    low PEEP of 3-5 to prevent atelectasis are ok

  • 8/12/2019 Bipap y Aprv

    74/77

    Persons with Congestive

    Heart failure We can start with normal settings, but if thePIP and Pplateauare excessive, we need todecrease the VT

    PEEP at 10 cmH20 and wean the

    Once the patients compliance gets better, wemust wean the PEEP

    If the patient has an intact ventilatory drive, &good VE, he could be maintained on CPAP

  • 8/12/2019 Bipap y Aprv

    75/77

    Initial parameters when High RAW

    is an issue?

    start with SIMV because this mode is lesslikely to cause air trapping.

    minimize air trapping and auto-PEEP

  • 8/12/2019 Bipap y Aprv

    76/77

    COPD SIMV rate between 10-12 BPM : decrease this to 6-8 to

    allow time to exhale start at 60 and raise to100 LPM].

    A COPD patient can be started at 40-50% Fi02

    Use of PEEP with COPD is dangerous, but if the set PEEP

    and the auto-PEEP are kept about the same, the gas ismore likely to leave the lung

    , keep Sp02 at 90-92% and keep the PaC02and pH closeto baseline so the patient will not sufferpost-hypercapnic

    alkalosis

    A th ti [AHI 2005 CPR CPG IV 141]

  • 8/12/2019 Bipap y Aprv

    77/77

    Asthmatic [AHI 2005 CPR CPG pp IV 141]

    Alert? may do well on BiPap machine

    SIMV rate 6-10 BPM

    VTof 6-8 ml/ kg IBW

    80-100 LPM with a descending flow pattern to

    get 1:4 or 1:5 Start Fi02 at 100%.

    Use of PEEP with asthmatics is dangerous, but if

    the set PEEP and the auto-PEEP are kept thesame, the gas is more likely to leave the lung.

    permissive hypercapnia,