空気圧を利用したトレッドミル歩行の評価DEVELOPMENT AND CLINICAL USE OF A FLOAT
WALKING ASSIST SYSTEM USING AIRPRESSURE
(科学研究費 萌芽 2005年度, 萌芽 2007年度,シーズ試験研究 2007年度)
In the field of physical therapy, walkingassist systems that comfortably reducebody weight are needed for patients withjoint pain or muscle weakness of lowerlimbs. Although suspension devices orunderwater walking systems reduce theweight, those do not resemble to normalwalking pattern.
Therefore, we developed a float walkingassist system using LBPP4,5. Today, the2nd generation system has beendeveloped.
4) Morinaga T (2002) the8th ACPT 2002,Bangkok
5) Kuroki H (2003) the14th WCPT 2003,Barcelona
The first generation system
Side view
The LBPP chambercontaining a treadmill,which encloses thesubjects lower body in aninflatable chamber at thewaist by a flexible seal,has been used.
A float walking assist system using air pressure(The second generation system)
Drafting picturePosterior view
A pressure gauge
A loadindicator
An air pressurecontrollerA controller to
elevatetreadmill
Four load cells aremounted under thetreadmill
Treadmill
An airpressurization device
A flexible seal at thewaist
Subjects: Thirteen females with hiposteoarthritis, including eightsubjects who underwent total hiparthroplasty
Age (mean ± SD): 55.6 ± 7.7 yearsBody height (mean ± SD) : 151.9 ± 13.2 cmBody weight (mean ± SD) : 55.6 ± 11.8 kgLength around the hip (level of ASIS): 90.0 ± 11.6 cm
Walking: 1) Before walking, written informed consent was
obtained. 2) Target air pressure for 2/3 and 1/2 partial
weight bearing was measured at standingposition.
3) Each subject randomly walked at a full-weight,2/3- and 1/2-weight bearing setting, at aspeed of 1.5 km/hour.
Four measuring parameters: 1) Maximum pressure on a foot 2) Mean pressure on a foot during the stance 3) VO2 4) Heart rate
Statistical analysis: Repeated measure ANOVAand Post-hoc Sheffe’s test
Pressure measuring systemon a foot(MP-100, AnimaLtd., Japan )
Respiratory gas analyzer(AE-280, Minato Ltd.,Japan )
Heart rate measuring system(EBP-300, Minato Ltd.,Japan)
Maximum pressure on a foot
% body weight
0
20
40
60
80
100
120
140
Full-weight
bearing
2/3-weight
bearing
1/2-weight
bearing% !!"p<0.01
!!
!!!!
Body weight
0
10
20
30
40
50
60
70
80
90
Full-weight
bearing
2/3-weight
bearing
1/2-weight
bearing
kg
!!"p<0.01
!!
!!
Average pressure was 2.3 kPa for 2/3- weight bearing setting and 3.2 kPa for 1/2-weight bearing setting, respectively. Maximum load on the foot at full-, 2/3- and1/2- weight bearing settings was 104.3 %, 75.7 % and 64.0 %, respectively.
Mean pressure on a foot during stance phase
Body weight
0
10
20
30
40
50
60
Full-weight
bearing
2/3-weight
bearing
1/2-weight
bearing
kg
!!"p<0.01
!!!!
% body weight
0
20
40
60
80
100
Full-weight
bearing
2/3-weight
bearing
1/2-weight
bearing
%
!!"p<0.01
!!
!!!!
The mean pressure on a foot was 69.1 %, 49.7 % and 41.8 %,respectively.
VO2 and Heart Rate
Heart Rate
0
20
40
60
80
100
120
140
Full-weight
bearing
2/3-weight
bearing
1/2-weight
bearingbea
ts/m
in
!"p<0.05
!!
VO2
0
2
4
6
8
10
12
14
16
Full-weight
bearing
2/3-weight
bearing
1/2-weight
bearing
ml/kg/min
!"p<0.05
! !
VO2 was 11.0, 8.4 and 7.8 mlO2/kg/min, respectively. Heartrate was 107.6, 91.8 and 88.6 beats/min, respectively.
Pressure on a foot significantly decreased atpartial weight bearing settings.The pressure pattern was two peaks thatseemed to natural walking pattern.At the 2/3-weight bearing setting, the bodyweight was reduced accurately.
0
20
40
60
80
100
120
1 21 41 61 81
time (second)
FWB 2/3 PWB 1/2 PWB
!
67%
50%
% b
od
y w
eig
ht
0.2 0.4 0.6 0.8
It may be controlled by additionalair pressure before or duringwalking.
At the 1/2-weight bearing setting with airpressure of 3.2 kPa, peak pressure was alittle bit higher than the setting.This means the air pressure was notenough to reduce the body weight.