Nuclear Instruments and Methods 198 (1982) 165-167 165 North-Holland Publishing Company COMBINED LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY FOR TRACE ANALYSIS OF PHARMACEUTICALS Lothar SCHMIDT, Harald DANIGEL, and Hartmut JUNGCLAS Kernchemie, Fachbereich Physikalische Chemie, Philipps-Universti~t, D-3550 Marburg, FRG A 252 Cf-plasma desorption mass spectrometer (PDMS) for the analysis of thin layers from nonvolatile organic samples has been set up to be combined with a liquid chromatograph. A novel interface performs the direct inlet of the liquid sample through a capillary into the vacuum system of the spectrometer. Samples of drugs are periodically collected, transferred to the ion source and analysed using a rotating disk. This on-line sample preparation has been tested for three antiarrhythmic drugs using various solvents and mixtures. 1. Introduction The detection of pharmaceuticals in biological fluids like blood and urine requires a careful sam- ple preparation and a sensitive detection system. In this sense high-performance liquid chromatog- raphy (HPLC) is a very sensitive analytical method [1]. Utilizing the proper detection for the desired substance, concentrations like 1 ng/ml and lower can be determined. In general mass spectrometry (MS) is very sensitive too but also produces specific information on many substances, however, de- pending on the ionization mode employed. A com- bination of these two methods (LC/MS) should be very powerful - similar to GC/MS - indepen- dent if one calls MS an universal detector for LC or calls LC just a sample clean-up for MS. Some commercially available LC/MS systems utilize the moving belt interface [2] in combination with electron impact and chemical ionization (El, CI). Since HPLC is suitable for nonvolatile organic samples, it is desirable to combine HPLC and a surface ionization mode with desorption, induced by fast atoms (SIMS), laser pulses (LIMS) or fission fragments (252Cf-PDMS). These three methods produce similar mass spectroscopical in- formation for thermic unstable and non-volatile molecules [3]. Applying 252Cf-PDMS [4] we developed a novel LC/MS interface [5] which couples the LC on-line to the MS using a capillary inlet. A vacuum lock allows also the off-line analysis of already pre- pared samples. For some antiarrhythmic drugs we investigated the detection limits and the influence of solvents on the relative detection efficiency. 2. Equipment The system, which was set-up with the intention to combine 2~2Cf-PDMS [4] and reversed-phase LC was described recently [5]. Section 2.1 shall explain further modifications. 2.1. The time-of-flight mass spectrometer The present set-up is schematically shown in fig. l. The 252Cf source (1 #Ci) is mounted behind ~ ) ' START e- -CONVERTER SAMPLE (-15 kV) (-+10 kV) / /, I n15keV MCP 252Cf STOP TIMING SIGNALS GRIDS ON GROUND *- 10 keY IONS MCP x~ ~ 11, e--CONVERTER (-t5 kV) I CONDENSER Fig. 1: Schematic drawing of the 252Cf-PDMS time-of-flight spectrometer displaying the ion source, flight path and ion detection via conversion electrons (MCP: micro channel plates). 0167-5087/82/0000-0000/$02.75 © 1982 North-Holland IV. APPLICATIONS

Upload: lothar-schmidt

Post on 21-Jun-2016

217 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Combined liquid chromatography-mass spectrometry for trace analysis of pharmaceuticals

Nuclear Instruments and Methods 198 (1982) 165-167 165 North-Holland Publishing Company

COMBINED LIQUID C H R O M A T O G R A P H Y - M A S S SPECTROMETRY FOR TRACE ANALYSIS OF PHARMACEUTICALS

Lothar SCHMIDT, Harald DANIGEL, and Hartmut JUNGCLAS Kernchemie, Fachbereich Physikalische Chemie, Philipps-Universti~t, D-3550 Marburg, FRG

A 252 Cf-plasma desorption mass spectrometer (PDMS) for the analysis of thin layers from nonvolatile organic samples has been set up to be combined with a liquid chromatograph. A novel interface performs the direct inlet of the liquid sample through a capillary into the vacuum system of the spectrometer. Samples of drugs are periodically collected, transferred to the ion source and analysed using a rotating disk. This on-line sample preparation has been tested for three antiarrhythmic drugs using various solvents and mixtures.

1. Introduction

The detection of pharmaceuticals in biological fluids like blood and urine requires a careful sam- ple preparation and a sensitive detection system. In this sense high-performance liquid chromatog- raphy (HPLC) is a very sensitive analytical method [1]. Utilizing the proper detection for the desired substance, concentrations like 1 ng /ml and lower can be determined. In general mass spectrometry (MS) is very sensitive too but also produces specific information on many substances, however, de- pending on the ionization mode employed. A com- bination of these two methods (LC/MS) should be very powerful - similar to G C / M S - indepen- dent if one calls MS an universal detector for LC or calls LC just a sample clean-up for MS.

Some commercially available L C / M S systems utilize the moving belt interface [2] in combination with electron impact and chemical ionization (El, CI). Since HPLC is suitable for nonvolatile organic samples, it is desirable to combine HPLC and a surface ionization mode with desorption, induced by fast atoms (SIMS), laser pulses (LIMS) or fission fragments (252Cf-PDMS). These three methods produce similar mass spectroscopical in- formation for thermic unstable and non-volatile molecules [3].

Applying 252Cf-PDMS [4] we developed a novel L C / M S interface [5] which couples the LC on-line to the MS using a capillary inlet. A vacuum lock

allows also the off-line analysis of already pre- pared samples. For some antiarrhythmic drugs we investigated the detection limits and the influence of solvents on the relative detection efficiency.

2. Equipment

The system, which was set-up with the intention to combine 2~2Cf-PDMS [4] and reversed-phase LC was described recently [5]. Section 2.1 shall explain further modifications.

2.1. The time-of-flight mass spectrometer

The present set-up is schematically shown in fig. l. The 252Cf source (1 #Ci) is mounted behind

~ ) ' START

e- -CONVERTER SAMPLE

(-15 kV) (-+10 kV)

/ /,

I n15keV

MCP 252Cf

STOP • TIMING SIGNALS

GRIDS

ON GROUND

*- 10 keY IONS

MCP x~ ~ 11,

e--CONVERTER

( - t5 kV)

I

CONDENSER

Fig. 1: Schematic drawing of the 252Cf-PDMS time-of-flight spectrometer displaying the ion source, flight path and ion detection via conversion electrons (MCP: micro channel plates).

0167-5087/82/0000-0000/$02.75 © 1982 North-Holland IV. APPLICATIONS

Page 2: Combined liquid chromatography-mass spectrometry for trace analysis of pharmaceuticals

166 L. Schmidt et al. / Trace analysis of pharmaceuticals

the sample foil. When a fission fragment passes this foil (100/s), the antiparallel fragment gen- erates secondary electrons from a converter foil, which are detected by a micro channel plate (MCP) arrangement and finally produce a "start" timing signal. Also the desorbed ions ( ± 1 0 keV) are converted to electrons at the end of the drift tube (72 cm length), which are detected by a second MCP arrangement and produce the "s top" timing signals. Since the grid in front of the converter has 80% transmission, the detection efficiency for the ions arriving at the end of the flight path is ca. 80%. This improvement will be very useful, when a time digitizer with multi stop logic (buffer mem- ory) is utilized instead of a conventional time to amplitude converter.

2.2. The L C / M S interface

This novel interface combines LC and 252Cf- PDMS in a two fold way: on-line by introducing the effluent continuously through a capillary inlet and off-line by transferrring prepared samples though a vacuum lock.

The interface is a rotating disk containing 8 discrete sample positions and controlled by a step- ping motor. Fig. 2 shows a cut through this disk displaying the capillary and the ion source posi- tions. The disk is centered in the rough vacuum chamber and extends into the high vacuum (10 6 mbar) of the MS. Running a 500 1/s Roots pump the rough vacuum (10 2 mbar) is maintained even at 300 /~l/min solvent flow through the capillary inlet.

~J// '///////////////~

.... . . . . . . . . . . . . t~ '~ ! HEATER ~ ROOTS • I ~ =~:~,~//~ =====a=- CAPILLARY PUMP ~ FROM HPLC

"~ ~ I~ ~0- mbor 2'I n V/. ~."/2~///////////~

~ i i [ 10-6mb°r [ ~ ' I - - ~ 10keY [0NS , / ' u,,,P I ............ ('v

I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I ] I I I I I I I I I I A

Fig. 2: L C / M S interface, cut through capi l lary inlet and ion

source.

Due to the rapid decompression of the effluent, it can freeze at the tip of the capillary, especially when it contains water. To prevent plugging of the capillary its tip can be heated by a coil (ca. l0 W).

3. Experiments

The system was tested with the pure anti- arrhythmic drugs Bunitrolol.HC1, Quinidine.SO4, and Verapamil.HC1. The nominal molecular weights M of the free bases are 248, 324 and 454, which are observed in the mass spectra as (M + H) + peaks due to proton transfer. Pure samples and mixtures of these three drugs were desolved in various solvents, which are suitable for reversed- phase LC: H20, CH3OH, CH3CN, and various mixtures of these. Mass spectra were recorded with acquisition times between 10 rain and 5 h, depend- ing on the count rate. The intensities for the three molecular lines are deduced from background cor- rected peak areas and normalized to 100 min.

4. Results

Usually samples for 252Cf-PDMS are prepared in advance by the electro-spray technique [6] and then transferred into the MS (off-line). Comparing the results for the pure pharmaceuticals (Bunitro- lol, Quinidine and Verapamil) no significant dif- ference in the count rates was observed in off-line and on-line samples preparation. But we found a large variation in molecular ion intensities for the three pharmaceuticals during on-line test experi- ments. The collection process of the pharmaceuti-

10 4 -

E oE 10 3 2

z o 10 2 '

101

6

o 8

i i i

I0 10 2 10 3

Drug concentrotion

A A O O

O Verapomtl

A Quinidine

i i 10 4 10Sng/ml

Fig. 3: Molecular ion in tensi ty as a funct ion of drug concentra- t ion using a C H a O H : H 2 0 = 1 : 1 solvent.

Page 3: Combined liquid chromatography-mass spectrometry for trace analysis of pharmaceuticals

L. Schmidt et al. / Trace analysis of pharmaceuticals 167

1000-

800 - _= E

"~ 6oo-

,ooi 200"

• 8unitrotol ~100 ng • Verapamil I ~ ] -

J J i 3

I0 10 2 10 ~g/ml

Sort concentration [(NHz.)H 2 POt,]

Fig. 4: Molecular ion intensity as a function of (NH4)H2PO 4 concentration, added to the CH3CN: H 2 0 = 4:1 solvent.

1000-

800- __. E

600-

z

o

~ 1,00-

2 0 0 "

Drugs (100 ng/ml)

- [- Salts ( 500 gig/roll

/ [Z] NoCI

I [NH~) HzPO ~

5 I

Bunitrolot Quinidine Verapamil

Fig. 5: Molecular ion intensity using a C H 3 C N : H 2 0 = 9:1 solvent and adding two different salts. The error bars indicate the standard deviation from 6 experiments per data point.

cals in the rough vacuum on the sample foil is very critical. If the pressure is too high, the sample molecules do not even stick to the foil. Also we observed that an already collected deposit was washed away by the following liquid jet.

Fig. 3 shows the intensities for Quinidine and Verapamil as a function of drug concentration using a C H a O H : H 2 0 = 1:1 solvent. Using this solvent Bunitrolol was hardly detectable, but it worked very well with C H 3 C N - H 2 0 mixtures. We investigated the influence of (NH4)H=PO 4 added to the solvent. Fig. 4 displays the positive effect on Bunitrolol, while Verapamil is slowly varying and Quinidine hardly detectable.

Since the effect is not uniform we used two different salts and repeated each experiment six times. Fig. 5 compares the influence of NaC1 and (NH4)H2PO 4. While Bunitrolol is enhanced by (NH4)H 2 PO 4, Quinidine and Verapamil are more intense using NaCI.

5. Conclusion

The preparation of a sample from small amounts (<~ l0 ng in 100 /~1 solvent) in the on-line mode

using the capillary inlet is still in an empirical state. The collection efficiency of the sample mole- cules in a rough vacuum chamber, i.e. the sensitiv- ity of the intented L C / M S analysis varies from pharmaceutical to pharmaceutical and depends on the solvents used. The efficiency can be improved by adding to the solvent salts, which act as con- densation and crystallization centers.

This work was supported by the BMFT, Bonn.

References

[1] GLC and HPLC Determination of Therapeutic Agents, Chromatographic Science Series, vol. 9, part 3, ed., K. Tsuji (Marcel Dekker, New York, 1979).

[2] W.H. McFadden and D.C. Bradford, 25th Conf. Mass. Spectrom. Allied Topics, Paper FP-I1, Washington DC, May (1977).

[3] Proc. of the Workshop Ion Formation from Organic Solids, Miinster (October 6-8, 1980) to be published by Springer.

[4] R.D. Macfarlane and D.F. Torgerson, Int. J. Mass Spec- trom. Ion. Phys. 21 (1976) 81.

[5] H. Jungclas, H. Danigel, and L. Schmidt, Org. Mass Spec- trom. 17 (1982) 86.

[6] C.J. McNeal, R.D. Macfarlane, and E.L. Thurston, Anal. Chem. 51 (1979) 2036.