Techniques for Rapid and Accurate Sample Analysis in the Field Aaron Baensch — Principal Geologist, International Mining Group
Dane Burkett — XRD Product Specialist
Olympus Scientific Solutions Americas (OSSA) – Monday, June 19, 2017
![Page 2: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/2.jpg)
Webinar Agenda
§ Introduction to OSSA, the mining team, and analytical products § Quick introduction to pXRF and pXRD § Focus on pXRF
§ New Olympus Vanta™ pXRF analyzer § Workflows and standard operating procedures (SOPs) § Sample preparation, CRMs, QA/QC, and data
management § References § A selection of case studies and applications
§ Focus on pXRD § Diffraction 101 § Introduction to Olympus pXRD instruments § A selection of studies and applications
OLYMPUS Scientific Solutions
![Page 3: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/3.jpg)
Olympus’ Dedicated International Mining Group (IMG) ToddHoulahan—Director
• JoinedOlympusin2008• 17yearsportableXRFexperience• Globalminingsales,marketing,andapplications• BAppSc.UniversityofNewcastle,Australia,1993(soilsmajor)• BasedinLondon,UK
MarcusLake—GlobalBusinessDevelopmentManager
• JoinedOlympusin2008• Drilling&blastingandcommoditytradingbackground• GlobalBD,distributorandsaleschanneltrainingandKAM• BachelorofCommerce,UniversityofNSW,Australia,1994• BasedinHongKong
AaronBaensch—PrincipalGeologist
• JoinedOlympusin2008• 10+yearspriorasindustrygeologist• Productandapplicationdevelopment,keyaccounts,andstrategicpartners
• ManagingDETCRCLab-at-Rig®• BachelorofMineralExplorationandMiningGeology(Hons)fromWASchoolOfMines,Kalgoorlie• BasedinSydney,Australia
DaneBurkett—XRDProductandApplicationsSpecialist
• JoinedOlympusin2015• BusinessDevelopmentandSupportingXRDApplications
• CompletedPhDusingOlympuspXRF&pXRD@UniversityNewSouthWales(UNSW)• BSc.Geology/EarthScience(Hons)fromUniversityofNSW(UNSW)• BasedinSydney,Australia
![Page 4: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/4.jpg)
Olympus and Olympus Scientific Solutions Americas (OSSA) § Olympus has a long history in optics, imaging, and sensor technology
– 40,000 employees, market cap of US$12B, founded in 1919 in Japan – Core Businesses:
– Cameras and audio – Medical, surgical, and life sciences (End, GI, Gyn, Urol…etc.) – Industrial = OSSA (HQ = Boston & Quebec)
» Nondestructive testing (NDT) » Ultrasonic » Phased array » Eddy current » Advanced NDT solutions and systems » Industrial borescopes and videoscopes
» Microscopy and optical metrology » Analytical instruments
» X-ray fluorescence (XRF) » X-ray diffraction (XRD)
OLYMPUS Scientific Solutions
![Page 5: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/5.jpg)
OLYMPUS Scientific Solutions
0 5 10 15 20 25 30
Quartz
Pentlandite
Violarite
Talc
Pyrrhotite
Pyrite
Anorthite
Chalcopyrite
Biotite
Cummingtonite
Mineral wt %
Combining XRD + XRF + Microscopy = Integrated Geoscience Solution
![Page 6: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/6.jpg)
OLYMPUS Scientific Solutions
Olympus — XRF Products Overview (Chemistry)
All brand names are trademarks or registered trademarks of Olympus Corporation
![Page 7: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/7.jpg)
OLYMPUS Scientific Solutions
Olympus — XRD Systems (Mineralogy) § TERRA® analyzer = a truly portable XRD device
§ BTX II™ analyzer = small and cost-efficient benchtop XRD device
TERRA portable XRD analyzer
BTX II benchtop XRD analyzer
Web-based GUI
![Page 8: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/8.jpg)
OLYMPUS Scientific Solutions
Olympus — Microscopes (Optical Mineralogy) § Olympus also has a full range of stereo and polarizing petrographic and
metallurgical microscopes
Stereo microscopes Polarizing microscopes
Metallurgical microscopes
![Page 9: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/9.jpg)
OLYMPUS Scientific Solutions
Physics 101 — X-ray Fluorescence (XRF) and X-ray Diffraction (XRD)
R²=0.993
45
50
55
60
65
70
75
45 50 55 60 65 70 75
LAB(%)
pXRF(%)
Fe
www.ruby.colorado.edu
www.mdpi.com
§ XRF yields elemental chemistry (energy dispersive spectroscopy)
§ XRD yields quantitative mineralogy (fundamental crystallography)
![Page 10: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/10.jpg)
OLYMPUS Scientific Solutions
30+ Elements Down to PPM Level
![Page 11: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/11.jpg)
OLYMPUS Scientific Solutions
Mining and
Geoscience
Universities, research groups,
and geological surveys
Mineral exploration
Mining and grade control
Mineral processing and geometallurgy
Environmental and mine closure
Fixed and mobile plant maintenance and construction (alloy, PMI, and
oil analysis)
Servicing the Entire Mining Value Chain
![Page 12: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/12.jpg)
OLYMPUS Scientific Solutions
Focus on pXRF
![Page 13: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/13.jpg)
§ Rugged § IP65/67 rated — dust and water resistant § High temperature rating: 122º F / 50º C (with optional fan) § Drop tested § Detector shutter
§ Revolutionary § AxonTM Technology = High Count Rate § Stable hardware and excellent resolution § Precise and accurate § Automated calibration, no cal check required § Tool-less window change
§ Productive § New detector pulse processor § Optional Wi-Fi enables powerful cloud applications § Embedded GPS § New software, methods, modes, and GUI
Handheld XRF — Vanta™ Analyzer
OLYMPUS Scientific Solutions
![Page 14: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/14.jpg)
OLYMPUS Scientific Solutions
Geoscience Week 2017 Feature: pXRF Best Practice Guide (Blog)
![Page 15: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/15.jpg)
OLYMPUS Scientific Solutions
SOPs, Chain of Custody, and QA/QC (Fisher & Gazley, 2014)
![Page 16: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/16.jpg)
OLYMPUS Scientific Solutions
Geoscience = Many Different Types of Samples
![Page 17: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/17.jpg)
OLYMPUS Scientific Solutions
Sample Selection — Horizon, Grain Size, Bags, etc.
OLYMPUS Scientific Solutions
![Page 18: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/18.jpg)
OLYMPUS Scientific Solutions
Site-Specific Calibration and Test Time Optimization (Client Samples)
XRF_Cal_ID Cu_Lab Cu_pXRF Fe_Lab Fe_pXRFSF01 0.02 0.02 11.13 12.38SF02 2.50 2.54 33.84 30.76SF03 5.09 4.91 35.42 35.24SF04 7.49 7.46 38.89 38.21SF05 10.00 9.67 28.58 27.48SF06 11.47 11.19 17.73 18.50SF07 12.45 11.88 14.50 15.73SF08 13.23 12.81 29.39 29.31SF09 15.20 14.56 16.09 17.39SF10 17.67 16.61 32.35 30.84SF11 21.72 19.40 31.19 29.41SF12 30.19 28.67 9.54 11.05SF13 32.60 30.98 10.25 11.26SF14 35.23 32.30 8.70 10.43SF15 37.51 35.79 8.52 9.78SF16 40.14 39.77 8.87 9.83SF17 42.50 39.20 7.94 8.19SF18 43.72 43.72 9.18 10.67SF19 45.02 42.54 7.33 8.50SF20 46.04 46.01 5.94 7.12SF21 47.55 47.64 7.33 8.94SF22 48.19 48.57 6.15 7.11SF23 50.10 50.25 7.27 8.02SF24 52.52 53.50 6.79 7.75SF25 53.01 52.08 5.91 7.00SF26 55.00 56.51 6.51 7.41SF27 57.04 57.59 5.87 6.75SF28 60.58 61.90 4.41 4.95SF29 64.67 65.64 5.93 5.57SF30 68.73 68.85 4.85 4.73
y=1.0062xR²=0.9961
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
80.00
0.00 10.00 20.00 30.00 40.00 50.00 60.00 70.00 80.00
LabXR
F
pXRF
Cu_pXRF
y=1.0029xR²=0.9874
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
45.00
0.00 10.00 20.00 30.00 40.00 50.00
LabXR
F
pXRF
Fe_pXRF
![Page 19: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/19.jpg)
OLYMPUS Scientific Solutions
Real-Time 3D Visualisation
http://www.leapfrog3d.com/online-resources/case-studies
![Page 20: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/20.jpg)
OLYMPUS Scientific Solutions
Reporting (eg. JORC Code) — pXRF Referenced in Table 1
OLYMPUS Scientific Solutions
![Page 21: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/21.jpg)
OLYMPUS Scientific Solutions
pXRF Benchmarking and Industry Reference (2010–2013)
OLYMPUS Scientific Solutions
![Page 22: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/22.jpg)
OLYMPUS Scientific Solutions
GEEA pXRF Thematic Set — February 2016
OLYMPUS Scientific Solutions
![Page 23: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/23.jpg)
OLYMPUS Scientific Solutions
Examples and Case Studies
![Page 24: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/24.jpg)
OLYMPUS Scientific Solutions
2016 PDAC Short Course: Case Studies (Videos on YouTube) § Olympus-IMS Channel
![Page 25: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/25.jpg)
OLYMPUS Scientific Solutions
Geological Surveys — GTK Special Publication — Novel Technologies
OLYMPUS Scientific Solutions
![Page 26: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/26.jpg)
OLYMPUS Scientific Solutions
Geological Surveys: Litho-Geochemistry (Drill Chips and Cores)
![Page 27: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/27.jpg)
OLYMPUS Scientific Solutions
Geoscience Research and Development
![Page 28: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/28.jpg)
OLYMPUS Scientific Solutions
Geoscience R&D — Automated Geology
![Page 29: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/29.jpg)
OLYMPUS Scientific Solutions
Soil and Regolith Geochemical Surveys
![Page 30: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/30.jpg)
OLYMPUS Scientific Solutions
Soil and Regolith Geochemical Surveys — Rapid Multivariate Datasets
![Page 31: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/31.jpg)
OLYMPUS Scientific Solutions
Soils — What can be achieved in one month with one instrument?
OLYMPUS Scientific Solutions Phoenix Copper, 2008.
![Page 32: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/32.jpg)
OLYMPUS Scientific Solutions
Auger Drilling — Regolith Mapping for Gold (Au) in West Africa
OLYMPUS Scientific Solutions
Robust rock classification at
surface using a Ti-Zr classification diagram
![Page 33: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/33.jpg)
OLYMPUS Scientific Solutions
Gold Exploration — Using Pathfinders and Geochemical Signatures
![Page 34: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/34.jpg)
OLYMPUS Scientific Solutions
Gold Mine Case Study Published in GEEA
§ Paper published in Journal of Geochemical Exploration (2011)
![Page 35: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/35.jpg)
OLYMPUS Scientific Solutions
Gold — Chemo-Stratigraphic Controls on Au Deposition
§ Using Au vs Cr to map and model mine stratigraphy and lode locations
![Page 36: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/36.jpg)
OLYMPUS Scientific Solutions
Gold Processing — Modelling Recovery from Arsenic in Refractory Ores
OLYMPUS Scientific Solutions
Free Au
Aspy + Au
A
B
![Page 37: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/37.jpg)
OLYMPUS Scientific Solutions
Textural and Structural Investigation Using a Camera and Collimator
OLYMPUS Scientific Solutions
§ pXRF (with in-line camera) for assessing Au in diamond drill cores
![Page 38: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/38.jpg)
OLYMPUS Scientific Solutions
Iron Ore Grade Control with a High Count Rate (15 seconds)
OLYMPUS Scientific Solutions
90 Sec Test (30s B1/60s B2) vs. 15 Sec Test (5s B1/10s B2)
![Page 39: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/39.jpg)
OLYMPUS Scientific Solutions OLYMPUS Scientific Solutions
Iron Ore Grade Control with a High Count Rate (15 seconds)
90 Sec Test (30s B1/60s B2) vs. 15 Sec Test (5s B1/10s B2)
![Page 40: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/40.jpg)
OLYMPUS Scientific Solutions OLYMPUS Scientific Solutions
Lithium Exploration Combining pXRF and pXRD
§ Current example — LCT pegmatite exploration – mineralogy is key! R²=0.9676
0.00
0.50
1.00
1.50
2.00
2.50
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7
Li2O(%
)
pXRFRb(%)
LabLivspXRFRbMineralName Li2O(%)* ChemicalFormulaSpodumene 8.03 LiAl(SiO3)2Petalite 4.50 LiAlSi4O10
Eucryptite 11.86 LiAlSiO4
Amblygonite 7.40 (Li,Na)AlPO4(F,OH)
Lepidolite 7.70 K(Li,Al)3(Al,Si)4O10(F,OH)2Lithiophilite 9.53 LiMnPO4
Zinnwaldite 3.42 KLiFeAl(AlSi3)O10(OH,F)2Holmquistite 3.98 Li2(Mg,Fe2+)3Al2Si8O22(OH)2Triphylite 9.47 Li(Fe,Mn)PO4SOURCE:webminera l .com*Ca lculatedLi 2OContent
![Page 41: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/41.jpg)
OLYMPUS Scientific Solutions OLYMPUS Scientific Solutions
Focus on XRD
![Page 42: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/42.jpg)
What is XRD?
§ X-ray diffraction (XRD)
§ Way of defining crystalline phases (minerals) – E.g. quartz, magnetite, hematite, muscovite
§ Can identify mineral phases irrespective of reflectance
§ Can be used to identify and quantify minerals
TERRA® portable XRD analyzer
BTX-II™ benchtop XRD analyzer
![Page 43: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/43.jpg)
How Does XRD Work?
§ Crystals have a defined structure that is like a fingerprint
§ X-ray diffraction characterises this crystal structure
www.ccp14.ac.uk
www.ruby.colorado.edu
www.spectraminerals.com
www.marinmineral.com www.mdpi.com
![Page 44: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/44.jpg)
Conventional XRD Instruments
§ Goniometer
§ Large lab-based equipment
§ Energy intensive
§ External water source
www.sfu.ca/
Whitefield (2012) http://www.tulane.edu/
![Page 45: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/45.jpg)
Sample Preparation
§ Pressed pellets
§ Dry, < 10 µm, approximately 300 mg
§ Need homogenous grain size
§ Assumes a random orientation
Whitfield (2012) Mdpi.com
![Page 46: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/46.jpg)
Conventional XRD Sample Preparation
§ Careful sample preparation required (15 + minutes)
§ Micronized sample is recommended (<10 µm)
A very flat surface is required
![Page 47: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/47.jpg)
XRD on Mars?
§ Conventional XRD is too large, sample preparation too onerous, and energy requirements too great
§ Need for an XRD instrument that was light, rugged, and did not require regular calibrations
Mars
![Page 48: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/48.jpg)
XRD on Mars!
Termed “the ultimate exploration tool / robotic geologist on wheels…”
Sojourney (1997) 11.5 kg
MER (2004) 185 kg
MSL Curiosity (2012) 900 kg
• Spirit • Opportunity
![Page 49: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/49.jpg)
CheMin — the Olympus Link
• Chemistry = X-ray fluorescence (XRF) • Mineralogy = X-ray diffraction (XRD)
![Page 50: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/50.jpg)
Olympus XRD Instruments
§ Fully portable XRD
§ Fast, easy sample preparation
§ Energy efficient
§ No need for regular part alignments
12.5 kg 14.5 kg TERRA® portable XRD analyzer
BTX II™ benchtop XRD analyzer
![Page 51: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/51.jpg)
Sample Preparation
§ 15 mg of sample
§ 2–4 minutes
§ < 150 µm
Crush
Sieve
15 mg of sample
![Page 52: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/52.jpg)
Why Are Grain Size Requirements Not as Small for pXRD Compared to Lab XRD?
§ Conventional XRD recommends powdering to < 80 µm (preferably < 10 µm) – Rotating sample in 1D with reflectance geometry – Alleviate grain size effects and orientation effects
§ Portable XRD < 150 µm for soils, < 75 µm for sulfides and heavy mineral concentrates
– Vibrating sample with transmission geometry
Grain-size effects Whitfield (2012)
Orientation effects prims.mit.edu pXRD debye rings
![Page 53: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/53.jpg)
Portability
§ 4-hour battery life with hot-swap capability
§ Rugged
§ Can run off A/C power
![Page 54: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/54.jpg)
Patented Sample Vibration Holder — Orientation Effects
§ Vibrates the sample rather than rotating the detector and source
§ Eliminates orientation effects
Sararazin et al. 2005
![Page 55: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/55.jpg)
Power Consumption and Maintenance
Olympus XRD Lab-based XRD Power consumption 100 W
Detector cooling 1000 W
Filtered water + tube < $15/year $5000/year
Tube power 10 W 1.2 kW Tube replacement $0/year* $2000/year
Requires skilled technician External water No Yes
* Assuming a 6-year tube life of 6000 hours a year (equal to 5 day work week at 8 hours of work per day)
![Page 56: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/56.jpg)
What can be achieved with XRD?
§ Qualitative and quantitative mineralogical analysis – For all minerals, irrespective of color – Lower level of detection ~1–2 wt %
§ Quantify the amorphous content within a sample – High chrysocolla-bearing ores
§ Crystallinity
0%
20%
40%
60%
80%
100%
Carnallite
Lazulite
Sylvite
Halite
Eberl & Velde (1989)
Geology.com
![Page 57: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/57.jpg)
Testing Procedure
§ Overall, no significant change in results with varying runtimes – Effects the limits of detection
– For > 5 wt % – Simple mineralogy (2 minutes) – Complex mineralogy (5 minutes)
– For 1–2 wt % – Simple mineralogy (7 minutes) – Complex mineralogy (15 minutes)
Lab-based XRD, 5 mins, 10 mins, 20 mins, 40 mins Burkett et al. (2015)
![Page 58: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/58.jpg)
Lab-based XRD, 5 mins, 10 mins, 20 mins, 40 mins
![Page 59: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/59.jpg)
Runtime Analysis
Lab-based XRD, 5 mins, 10 mins, 20 mins, 40 mins
R2 = 0.80 R2 = 0.87
R2 = 0.92 R2 = 0.88
![Page 60: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/60.jpg)
Data Processing
§ Semi-quantitative analysis – Reference intensity ratio (RIR) – OK when there is minimal overlap between minerals
I
2 θ
![Page 61: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/61.jpg)
Data Processing
§ Quantitative analysis – Rietveld refinement – Many software packages available
– We suggest Siroquant® software because it has a database which utilizes observed reference material (better for poorly crystalline material)
![Page 62: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/62.jpg)
RIR vs. Rietveld
§ Porphyry copper deposit
§ 4 to 9 phases per sample
y=1.0115x+0.0026R²=0.97925
0
20
40
60
80
100
0 20 40 60 80 100
pXRD
Lab-XRD
Siroquant-hkl
![Page 63: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/63.jpg)
Rietveld Processing
§ A simple refinement of the hkls (peak positions) get you most of the way there
§ Can get better results if other parameters are refined (e.g. width)
y=0.9946x+0.0846R²=0.99212
0
20
40
60
80
100
0 20 40 60 80 100
pXRD
Lab-XRD
Siroquanty=1.0115x+0.0026R²=0.97925
0
20
40
60
80
100
0 20 40 60 80 100
pXRD
Lab-XRD
Siroquant-hkl
![Page 64: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/64.jpg)
XRD for Mineral Exploration: Current Methods for Mineralogical Characterization and Quantification
§ Core-logging using a hand lens – Alteration is typically fine-grained
– Often rely on color as a guide (three-way color split) – Geologists typically use geochemical data to infer mineralogy rather
than quantify it directly
Can roughly estimate mineralogy
Fine-grained alteration Brathwaite et al. 2014
![Page 65: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/65.jpg)
Current Methods for Mineralogical Characterization and Quantification
§ Petrography – Expensive, time-intensive, and often the phases
are still too fine-grained to identify/quantify – Hard to differentiate between species within
similar mineral groups – Carbonates: calcite, dolomite, ankerite,
rhodocrosite, etc. – Phyllosilicates: illite, muscovite, smectite,
interlayered clays, etc.
Fine-grained alteration
![Page 66: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/66.jpg)
Quantitative XRD
![Page 67: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/67.jpg)
Importance of XRD in Metal Exploration
§ Alteration studies – Alteration assemblages used to guide towards orebodies – Provide evidence to help constrain the geological environment
– Temperature, pressure, fluid chemistry, etc.
Tosdal et al. 2009 Hedenquist et al. 2000
![Page 68: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/68.jpg)
Integrating Mineralogical Data into the Logging Protocol Core-logging
XRD data
Core-logs with assay data
S (ppm)
20000 0
Uvarova et al. (2014)
Mineralogical data
Uvarova et al. (2014)
![Page 69: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/69.jpg)
Gold Exploration — Wall Rock Alteration
![Page 70: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/70.jpg)
Why is Mineralogy Important?
10 g/t
8 g/t
0 g/t
0 g/t
0 g/t
0 g/t
Ore-zone
Edge of ore-zone
Barren
Assay data can take weeks to months to get back
![Page 71: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/71.jpg)
Alteration Zonation — Fluid to rock buffered
Phyllic Illite Rock buffered
Potassic Adularia-illite-pyrite Zone of mixing
Epidote Adularia-epidote Fluid buffered Higher temperature
‘Burkett 2017’
![Page 72: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/72.jpg)
20 minute dataset
![Page 73: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/73.jpg)
Case Study — Reliability (Burkett et al. 2015)
§ Key findings – Highest grades encountered in wall rock with a relative proportion
of 60–80% adularia to 40–20% clays
§ Fluid mixing
§ How do we know the results are reliable?
0.01
0.10
1.00
10.00
0.00 0.20 0.40 0.60 0.80 1.00
Adularia/(Adularia + Total clays)
Au
(g/t)
![Page 74: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/74.jpg)
Verification of Approach Quantification
![Page 75: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/75.jpg)
XRD for Mineral Exploration
§ Graphite
§ Bauxite
§ Lithium
§ Copper
§ Iron ore
§ Potash
![Page 76: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/76.jpg)
Graphite Exploration
§ Carbon cannot be analyzed with XRF
§ XRD can quantify graphite and carbonate minerals
– Work out total carbon, total graphitic carbon, and total carbon in carbonates
Investorintel.com
Comparisons between lab assay data (x-axis) and estimates from XRD (y-axis)
![Page 77: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/77.jpg)
Importance of XRD in Mineral Processing
§ XRD can aid: – Comminution – Flotation – Leaching – Pyrometallurgy (roasting and smelting)
§ Can help to optimize each of the above processes and also help improve blending strategies
http://thedictionaryprojectblog.com/
http://mining-technology.com
http://wealthdaily.com
http://metallurgist.in
![Page 78: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/78.jpg)
XRD for Metallurgy Available on www.olympus-ims.com
![Page 79: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/79.jpg)
The Future — Combining pXRF and pXRD
OLYMPUS Scientific Solutions
![Page 80: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/80.jpg)
Complete pXRF and pXRD Integration Delivers Drill Logs of the Future
OLYMPUS Scientific Solutions
![Page 81: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/81.jpg)
Further Reading:
OLYMPUS Scientific Solutions
Uvarova, Cleverley, Baensch & Verrall, Coupled XRF and XRD analyses for rapid and low-cost characterisation of geological materials in the mineral exploration and mining industry, AAG Explore Vol. 162, March 2014
![Page 82: Techniques for Rapid and Accurate Sample Analysis in the Field](https://reader034.vdocuments.pub/reader034/viewer/2022051521/5a6821437f8b9a81378b666f/html5/thumbnails/82.jpg)
Thank you for your attendance & attention
OLYMPUS Scientific Solutions