ei-l2
TRANSCRIPT
-
7/31/2019 EI-L2
1/13
Basic concepts of Measurements
Instrument
A device for determining the value or magnitude of a quantity or variable
Accurac *
Closeness with which an instrument reading approaches the true value of thevariable being measured.
Precision*
Measure of the reproducibility of the results
Measure of the degree to which successive measurements differ from one
another ens v y
Ratio of output signal or response of the instrument to a change of input ormeasured variable.
The smallest change in measured value to which the instrument will respond
Errors
Deviation from the true value of the measured variable
-
7/31/2019 EI-L2
2/13
Accuracy
Precision Degree of agreement within a group of measurements or instruments
Composed of two characteristics
Conformity
Significant Figures
-
7/31/2019 EI-L2
3/13
Accuracy Vs Precision
Example 1.
Consider an ammeter which posses high degree of precision by virtue
, ,
with mirror arrangement to remove parallax. LC=(1/100) of Ampere. But the zero adjustment is wrong
, , ,
and the readings are consistent. However, the readings are not accurate
since they do not confirm to the truth.
Example 2.
Consider a known voltage of 100V. Five reading are taken: 104V,
, , .precision is 1%.
The instrument can be calibrated so that it could be used properly.
-
7/31/2019 EI-L2
4/13
Accuracy Vs Precision
By calibration, accuracy can be improved upon but not theprecision of the instrument.
Althou h the readin s are close to ether the have a small scatter andthus have a high degree ofprecision, but the results are far fromaccurate.
When it is stated that a set of reading shows precision, itmeans that the results agree among themselves.
A reement is no uarantee ofaccurac as there ma be some
disturbing effect that cause all the measured values to be in error.
-
7/31/2019 EI-L2
5/13
Significant Figures (SF)
Convey the actual information regarding the magnitude and the
.
The more SF, the greater is the precision of the measurement
The result is only as accurate as the least accurate measurement
The number of SF in multiplication may increase rapidly
Examples of SF
-
7/31/2019 EI-L2
6/13
Direct method
Unknown quantity (Measurand) is directly compared against a standard
,
Indirect method
Measuring Instruments
Absolute Instruments (AI)
Give the magnitude of the quantity in terms of physical constant of the
instrument
Secondary Instruments (SI)
Quantity can be measured by observing the output indicated by the instrument
Calibrated by comparison with an absolute instrument
Working with AI for routine work is time consuming since every time ameasurement is made, it takes a lot of time to compute the magnitude
Therefore, SI is most commonly used.
-
7/31/2019 EI-L2
7/13
Types of Instruments
Depending upon the way the instruments present the results of measurements,they are classified into two broad categories:
Deflection Type (DT)
Value of the measurand will be inferred from deflection or mechanicaldisplacement at the point of balance
Opposing effect = cause producing the deflection
Example: PMMC
Null Type (NT)
A zero indication leads to the determination of magnitude of measurand
Attempt to maintain the deflection at zero by suitable application of an
affect opposing that generated by the measurand Example: DC Potentiometer
-
7/31/2019 EI-L2
8/13
Comparison between Deflection and Null Type
Accuracy is higher for NT
Because the o osin effect is calibrated with the hel of standards whichhave high degree of accuracy
Whereas for DT, calibration depends upon the instrument constant whichare normally not known to a high degree of accuracy
Measurand is balanced out for NT
Because the detector have to cover a small range around the balance point
Whereas for DT, it has to be large in size, more rugged and less sensitive
D namic measurement
NT is not suitable because it require many manipulation before balance
DT is suitable because it can follow the rapid variations of the measurand
-
7/31/2019 EI-L2
9/13
Characteristic of Instrument & Measurement System
Static (measurand are either constant or vary slowly with time)
Static Characteristic
Accuracy
Sensitivity
Drift
Static error
Dead zone
-
7/31/2019 EI-L2
10/13
True Value Defined as the average of an infinite number of measured value when
the average deviation due to the various factors tend to zero.
Variation between the largest and the smallest reading (or largest one)
Scale Span
Difference between the largest and the smallest reading
Frequency Range requenc es over w c measuremen s can e per orme w a
specified degree of accuracy
Static Error
Difference between the measured and true vale
Static Correction
-
7/31/2019 EI-L2
11/13
Drift
Perfect reproducibility means no drift
i.e., with a given input, the values do not vary with time
Zero Drift
If the whole calibration gradually shift due to the slippage, permanent set,
or due to the undue warming up of the electronic circuit.
Can be prevented by zero setting
Span Drift or Sensitive Drift
Zonal Drift
Occurs onl over a ortion of s an
Drift is an undesirable quantity for the industries.
-
7/31/2019 EI-L2
12/13
Static Sensitivity Ratio of the magnitude of response or output signal to the magnitude of
the measurand.
When a calibration curve is linear sensitivit can be defined as the
slope of curve Constant over the entire range
, .
Dead Time
Time re uired b a measurement s stem to be in to res ond to a
change in the measurand.
Time before the instrument begins to respond after the measurand has been
changed
Dead Zone
Largest change of input for which there is no output of the instrument
-
7/31/2019 EI-L2
13/13
Resolution
The smallest increment in the input which can be detected with
So, the resolution defines the smallest measurable input change While, the threshold defines the smallest measured input.
Loading Effect
The ori inal si nal should not be distorted b introduction of an
element in the measurement system
However, practically it is not possible.
phase shift, etc.