fpix_flip_chip_test module calibration tests november 14 th 2012

93
FPIX_flip_chip_test module calibration tests November 14 th 2012

Upload: lee-pope

Post on 12-Jan-2016

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FPIX_flip_chip_test module calibration tests November 14 th 2012

FPIX_flip_chip_test module calibration tests

November 14th 2012

Page 2: FPIX_flip_chip_test module calibration tests November 14 th 2012

2

Sensor

• 2 SINTEF 2004 1x1 bump bonded– Planar 300 μm thick pixel sensor

• Flip-chipped then reflow• Assembled on a new testboard on November

7th, and 12th 2012• No post-dicing IV was performed for the first

sensor

Page 3: FPIX_flip_chip_test module calibration tests November 14 th 2012

3

Leakage currentFirst sensor’s before BBM IV was not performed!

Page 4: FPIX_flip_chip_test module calibration tests November 14 th 2012

4

Full calibration summary – First sensor

ROC pixel alive test

Bump-bonding test

Second sensor has similar calibration

Bias = -150 V

Page 5: FPIX_flip_chip_test module calibration tests November 14 th 2012

5

Testing bum-bond with forward bias test

• Run pixel alive test • ROC on with sensor at reverse bias• ROC on with sensor at forward bias

• Idea:– Forward bias will saturate the sensor, causing pre-

amplifiers in chip pixels saturate: chip pixels disappear from the pixel alive test map if bump-bonds are in contact

Page 6: FPIX_flip_chip_test module calibration tests November 14 th 2012

6

Testing bum-bond with forward bias test

SINTEF 2004 1x1 First sensor

NO

BIA

SRE

VERS

E BI

ASFO

RWAR

D B

IAS

SINTEF 2004 1x1 Second sensor

REVERSE BIAS = -150 VFORWARD BIAS = +10 V

Page 7: FPIX_flip_chip_test module calibration tests November 14 th 2012

7

Radioactive source test (Sr-90)

Bias = -150VMean = 22 ke-MP = 18 ke-285 μm thick

Double pixel hits In one trigger

SINTEF 2004 1x1 Second sensorSINTEF 2004 1x1 First sensor

Bias = -150VMean = 23 ke-MP = 18 ke-285 μm thick

Page 8: FPIX_flip_chip_test module calibration tests November 14 th 2012

8

Charge collection Charge distribution mean values versus bias voltages Scan not performed for the second sensor

Page 9: FPIX_flip_chip_test module calibration tests November 14 th 2012

9

Summary• Two SINTEF 2004 1x1 sensors bump bonded• Both sensor ROCs work fine

– Good calibrations: Gain, SCurve, Vsf etc– Full calibration

• Forward bias current and standard methods showed that all bump-bonds are in good contact – IV suggest a good ROC to sensor contact established since

ground goes through ROC for sensor HV• Standard bump-bond algorithm implemented in PSI

DAQ software consistent with forward bias test• Source tests showed a good charge collection

Page 10: FPIX_flip_chip_test module calibration tests November 14 th 2012

10

SINTEF 2011 1x1 SENSORS

Page 11: FPIX_flip_chip_test module calibration tests November 14 th 2012

11

Page 12: FPIX_flip_chip_test module calibration tests November 14 th 2012

12

Past design 1x2 in previous production

L1, S1, S2, S3, S5

New in this batch.Maximized pixel area

S8

L2, L3, S6

New in this batch.Removed asimmetry

S4, S7Note: S1, S2 and S3 have slim edges (S1 the slimmest and increasing in S2 and S3)

SINTEF_2011 1x1 sensors

Page 13: FPIX_flip_chip_test module calibration tests November 14 th 2012

13

November 13, 2012

Page 14: FPIX_flip_chip_test module calibration tests November 14 th 2012

14

Sensor: SINTEF_2011_WA15_S5

• Planar 300 μm thick• Bump bonded to a ROC on 11-13-2012

- Flip-chipped then reflown• Assembled on a new testboard on November

13th 2012

Page 15: FPIX_flip_chip_test module calibration tests November 14 th 2012

15

A15 S5 - Leakage currentBreakdown [V]

On wafer 195Post-dicing 225Module 960

Page 16: FPIX_flip_chip_test module calibration tests November 14 th 2012

16

A15 S5 - Full calibration summary (-150V)

ROC pixel alive test

Bump-bonding test (this test found 3185 dead bumps)

Bias = -150 V

Page 17: FPIX_flip_chip_test module calibration tests November 14 th 2012

17

A15 S5 - Full calibration summary (-400V)

ROC pixel alive test

Bump-bonding test (this test found 1 dead bumps)

Bias = -400 V

Page 18: FPIX_flip_chip_test module calibration tests November 14 th 2012

18

A15 S5 - Testing bum-bond with forward bias test

SINTEF_2011_WA15_S5 pixel alive test at REVERSE BIAS

SINTEF_2011_WA15_S5 pixel alive test at NO BIAS

REVERSE BIAS = -150 VFORWARD BIAS = +10 V

pixel alive test at FORWARD BIASpixel alive test at REVERSE BIAS with LIGHT

Page 19: FPIX_flip_chip_test module calibration tests November 14 th 2012

19

A15 S5 - Radioactive source test (Sr-90)

Source above sensor center Source above sensor edge

Pixel alive map at forward bias Source placed on different locationsabove the sensor No correlation found betweenforward bias pixel alive map andsource hitmap Can we trust forward bias test?

REVERSE BIAS = -150 VFORWARD BIAS = +10 V

Page 20: FPIX_flip_chip_test module calibration tests November 14 th 2012

20

A15 S5 - Radioactive source test (Sr-90)

Bias = -150VMean = 25 ke-MP = 21.5 ke-Thickness = 300 μm

Double pixel hits In one trigger

Page 21: FPIX_flip_chip_test module calibration tests November 14 th 2012

21

Sensor: SINTEF_2011_WA15_S6

• Planar 300 μm thick• Bump bonded to a ROC on 11-13-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

13th 2012

Page 22: FPIX_flip_chip_test module calibration tests November 14 th 2012

22

A15 S6 - Leakage currentBreakdown [V]

On wafer 360Post-dicing 370Module 835

Page 23: FPIX_flip_chip_test module calibration tests November 14 th 2012

23

A15 S6 - Full calibration summary (-150V)

ROC pixel alive test

Bump-bonding test (this test found 2488 dead bumps)

Bias = -150 V

Page 24: FPIX_flip_chip_test module calibration tests November 14 th 2012

24

A15 S6 - Full calibration summary (-300V)

ROC pixel alive test

Bump-bonding test (this test found 6 dead bumps)-400 V has also identical results

Bias = -300 V

Page 25: FPIX_flip_chip_test module calibration tests November 14 th 2012

25

A15 S6 - Testing bum-bond with forward bias test

SINTEF_2011_WA15_S6 pixel alive test at REVERSE BIAS

SINTEF_2011_WA15_S6 pixel alive test at FORWARD BIAS

SINTEF_2011_WA15_S6 pixel alive test at NO BIAS

~1200 pixels

Forward bias: 1200 dead bumpsPSI software test: 2488 dead bumps

REVERSE BIAS = -150 VFORWARD BIAS = +10 V

Page 26: FPIX_flip_chip_test module calibration tests November 14 th 2012

26

A15 S6 - Radioactive source test (Sr-90)

Bias = -150VMean = 20.7 ke-MP = 17.5 ke-Thickness = 300 μm

Source hit map

Page 27: FPIX_flip_chip_test module calibration tests November 14 th 2012

27

Summary• SINTEF 2011 S5 and S6 sensors bump bonded on 11-13-2012• Both sensor ROCs work fine

– Good calibrations: Gain, SCurve, Vsf etc– Full calibration with minimum effort

• Sensor breakdown voltages increased after assembly– No breakdown observed up to 800 V

• Bump-bond tests– Forward bias method: 1000 (S5), 1200 (S6) dead bumps – PSI method: 3185 (S5), 2488 (S6) dead bumps

• Source tests showed a good charge collection– Double hits in one trigger needs to removed from analysis– S6 will be retested with source

Page 28: FPIX_flip_chip_test module calibration tests November 14 th 2012

28

November 14, 2012

Page 29: FPIX_flip_chip_test module calibration tests November 14 th 2012

29

Sensor: SINTEF_2011_WA15_S7

• Planar 300 μm thick• Bump bonded to a ROC on 11-14-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

14th 2012• IVs on wafer, before and after BBM, and on

testboard measured

Page 30: FPIX_flip_chip_test module calibration tests November 14 th 2012

30

A15 S7 - Leakage currentBreakdown [V]

On wafer 310Before BBM 320BBM 945Testboard >1100

Page 31: FPIX_flip_chip_test module calibration tests November 14 th 2012

31

A15 S7 - Full calibration summary

ROC pixel alive test

Bump-bonding test

Bias = -200 V

Page 32: FPIX_flip_chip_test module calibration tests November 14 th 2012

32

A15 S7 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA15_S6 pixel alive test at REVERSE BIAS

pixel alive test at FORWARD BIAS

SINTEF_2011_WA15_S6 pixel alive test at NO BIAS

pixel alive test at REVERSE BIAS with LIGHT

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

Page 33: FPIX_flip_chip_test module calibration tests November 14 th 2012

33

A15 S7 - Radioactive source test (Sr-90)

Bias = -200VMean = 25 ke-MP = 21.7 ke-Thickness = 300 μm

Source hit map

Page 34: FPIX_flip_chip_test module calibration tests November 14 th 2012

34

Sensor: SINTEF_2011_WA15_S8

• Planar 300 μm thick• Bump bonded to a ROC on 11-14-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

14th 2012• IVs on wafer, before and after BBM, and on

testboard measured

Page 35: FPIX_flip_chip_test module calibration tests November 14 th 2012

35

A15 S8 - Leakage currentBreakdown [V]

On wafer 310Before BBM 220BBM >1100Testboard 995

Page 36: FPIX_flip_chip_test module calibration tests November 14 th 2012

36

A15 S8 - Full calibration summary

ROC pixel alive test

Bump-bonding test

Bias = -200 V

Page 37: FPIX_flip_chip_test module calibration tests November 14 th 2012

37

A15 S8 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA15_S6 pixel alive test at FORWARD BIAS

SINTEF_2011_WA15_S6 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA15_S6 pixel alive test with LIGHT at REVERSE BIAS

Page 38: FPIX_flip_chip_test module calibration tests November 14 th 2012

38

A15 S8 - Radioactive source test (Sr-90)

Bias = -200VMean = 26 ke-MP = 22.2 ke-Thickness = 300 μm

Source hit map

Page 39: FPIX_flip_chip_test module calibration tests November 14 th 2012

39

Sensor: SINTEF_2011_WA17_S5

• Planar 300 μm thick• Bump bonded to a ROC on 11-15-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

15th 2012• IVs on wafer, before and after BBM, and on

testboard measured

Page 40: FPIX_flip_chip_test module calibration tests November 14 th 2012

40

A17 S5 - Leakage currentBreakdown [V]

On wafer 205Before BBM 185BBM >1000Testboard >1000

Page 41: FPIX_flip_chip_test module calibration tests November 14 th 2012

41

A17 S5 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 67 dead bumps)

Bias = -200 V

Page 42: FPIX_flip_chip_test module calibration tests November 14 th 2012

42

A17 S5 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA17_S5 pixel alive test at FORWARD BIAS

SINTEF_2011_WA17_S5 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA17_S5 pixel alive test with LIGHT at REVERSE BIAS

Page 43: FPIX_flip_chip_test module calibration tests November 14 th 2012

43

A17 S5 - Radioactive source test (Sr-90)

Bias = -200VMean = 22.3 ke-MP = 19.7 ke-Thickness = 300 μm

Source hit map

Page 44: FPIX_flip_chip_test module calibration tests November 14 th 2012

44

Sensor: SINTEF_2011_WA17_S6

• Planar 300 μm thick• Bump bonded to a ROC on 11-15-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

15th 2012• IVs on wafer, before and after BBM, and on

testboard measured

Page 45: FPIX_flip_chip_test module calibration tests November 14 th 2012

45

A17 S6 - Leakage currentBreakdown [V]

On wafer 210Before BBM 185BBM >1000Testboard >1000

Page 46: FPIX_flip_chip_test module calibration tests November 14 th 2012

46

A17 S6 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 6 dead bumps)

Bias = -200 V

Page 47: FPIX_flip_chip_test module calibration tests November 14 th 2012

47

A17 S6 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA17_S6 pixel alive test at FORWARD BIAS

SINTEF_2011_WA17_S6 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA17_S6 pixel alive test with LIGHT at REVERSE BIAS

Page 48: FPIX_flip_chip_test module calibration tests November 14 th 2012

48

A17 S6 - Radioactive source test (Sr-90)

Bias = -200VMean = 26 ke-MP = 21.2 ke-Thickness = 300 μm

Source hit map

Page 49: FPIX_flip_chip_test module calibration tests November 14 th 2012

49

Sensor: SINTEF_2011_WA17_S7

• Planar 300 μm thick• Bump bonded to a ROC on 11-15-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

15th 2012• IVs on wafer, before, and on testboard

measured

Page 50: FPIX_flip_chip_test module calibration tests November 14 th 2012

50

A17 S7 - Leakage currentBreakdown [V]

On wafer 405Before BBM 310Testboard 830

Page 51: FPIX_flip_chip_test module calibration tests November 14 th 2012

51

A17 S7 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 15 dead bumps)

Bias = -200 V

Page 52: FPIX_flip_chip_test module calibration tests November 14 th 2012

52

A17 S7 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA17_S7 pixel alive test at FORWARD BIAS

SINTEF_2011_WA17_S7 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA17_S7 pixel alive test with LIGHT at REVERSE BIAS

Page 53: FPIX_flip_chip_test module calibration tests November 14 th 2012

53

A17 S7 - Radioactive source test (Sr-90)

Bias = -200VMean = 27 ke-MP = 21.4 ke-Thickness = 300 μm

Source hit map

Page 54: FPIX_flip_chip_test module calibration tests November 14 th 2012

54

Sensor: SINTEF_2011_WA17_S8

• Planar 300 μm thick• Bump bonded to a ROC on 11-15-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

15th 2012• IVs on wafer, before, and on testboard

measured

Page 55: FPIX_flip_chip_test module calibration tests November 14 th 2012

55

A17 S8 - Leakage currentBreakdown [V]

On wafer 325Before BBM 330Testboard 810

Page 56: FPIX_flip_chip_test module calibration tests November 14 th 2012

56

A17 S8 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 2 dead bumps)

Bias = -200 V

Page 57: FPIX_flip_chip_test module calibration tests November 14 th 2012

57

A17 S8 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA17_S8 pixel alive test at FORWARD BIAS

SINTEF_2011_WA17_S8 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA17_S8 pixel alive test with LIGHT at REVERSE BIAS

Page 58: FPIX_flip_chip_test module calibration tests November 14 th 2012

58

A17 S8 - Radioactive source test (Sr-90)

Bias = -200VMean = 25.5 ke-MP = 21 ke-Thickness = 300 μm

Source hit map

Page 59: FPIX_flip_chip_test module calibration tests November 14 th 2012

59

Sensor: SINTEF_2011_WA16_S5

• Planar 300 μm thick• Bump bonded to a ROC on 11-16-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

16th 2012• IVs on wafer, before, and on testboard

measured

Page 60: FPIX_flip_chip_test module calibration tests November 14 th 2012

60

A16 S5 - Leakage currentBreakdown [V]

On wafer 210Before BBM 185After BBM >1000Testboard 950

Page 61: FPIX_flip_chip_test module calibration tests November 14 th 2012

61

A16 S5 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 2 dead bumps)

Bias = -200 V

Page 62: FPIX_flip_chip_test module calibration tests November 14 th 2012

62

A16 S5 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA16_S5 pixel alive test at FORWARD BIAS

SINTEF_2011_WA16_S5 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA16_S5 pixel alive test with LIGHT at REVERSE BIAS

Page 63: FPIX_flip_chip_test module calibration tests November 14 th 2012

63

A16 S5 - Radioactive source test (Sr-90)

Bias = -200VMean = 26.7 ke-MP = 21.7 ke-Thickness = 300 μm

Source hit map

Page 64: FPIX_flip_chip_test module calibration tests November 14 th 2012

64

Sensor: SINTEF_2011_WA16_S6

• Planar 300 μm thick• Bump bonded to a ROC on 11-16-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

16th 2012• IVs on wafer, before, and on testboard

measured

Page 65: FPIX_flip_chip_test module calibration tests November 14 th 2012

65

A16 S6 - Leakage currentBreakdown [V]

On wafer 345Before BBM 190After BBM 835Testboard 825

Page 66: FPIX_flip_chip_test module calibration tests November 14 th 2012

66

A16 S6 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 13 dead bumps)

Bias = -200 V

Page 67: FPIX_flip_chip_test module calibration tests November 14 th 2012

67

A16 S6 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA16_S6 pixel alive test at FORWARD BIAS

SINTEF_2011_WA16_S6 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA16_S6 pixel alive test with LIGHT at REVERSE BIAS

Page 68: FPIX_flip_chip_test module calibration tests November 14 th 2012

68

A16 S6 - Radioactive source test (Sr-90)

Bias = -200VMean = 27 ke-MP = 22.6 ke-Thickness = 300 μm

Source hit map

Page 69: FPIX_flip_chip_test module calibration tests November 14 th 2012

69

Sensor: SINTEF_2011_WA16_S7

• Planar 300 μm thick• Bump bonded to a ROC on 11-16-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

16th 2012• IVs on wafer, before, and on testboard

measured

Page 70: FPIX_flip_chip_test module calibration tests November 14 th 2012

70

A16 S7 - Leakage currentBreakdown [V]

On wafer 345Before BBM 190Testboard 980

Page 71: FPIX_flip_chip_test module calibration tests November 14 th 2012

71

A16 S7 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 5 dead bumps)

Bias = -200 V

Page 72: FPIX_flip_chip_test module calibration tests November 14 th 2012

72

A16 S7 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA16_S7 pixel alive test at FORWARD BIAS

SINTEF_2011_WA16_S7 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA16_S7 pixel alive test with LIGHT at REVERSE BIAS

Page 73: FPIX_flip_chip_test module calibration tests November 14 th 2012

73

A16 S7 - Radioactive source test (Sr-90)

Bias = -200VMean = 29 ke-MP = 25 ke-Thickness = 300 μm

Source hit map

Page 74: FPIX_flip_chip_test module calibration tests November 14 th 2012

74

Sensor: SINTEF_2011_WA16_S8

• Planar 300 μm thick• Bump bonded to a ROC on 11-16-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

16th 2012• IVs on wafer, before, and on testboard

measured

Page 75: FPIX_flip_chip_test module calibration tests November 14 th 2012

75

A16 S8 - Leakage currentBreakdown [V]

On wafer 345Before BBM 325After BBM 970Testboard >1000

Page 76: FPIX_flip_chip_test module calibration tests November 14 th 2012

76

A16 S8 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 13 dead bumps)

Bias = -200 V

Page 77: FPIX_flip_chip_test module calibration tests November 14 th 2012

77

A16 S8 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA16_S8 pixel alive test at FORWARD BIAS

SINTEF_2011_WA16_S8 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA16_S8 pixel alive test with LIGHT at REVERSE BIAS

Page 78: FPIX_flip_chip_test module calibration tests November 14 th 2012

78

A16 S8 - Radioactive source test (Sr-90)

Bias = -200VMean = 25.6 ke-MP = 21.3 ke-Thickness = 300 μm

Source hit map

Page 79: FPIX_flip_chip_test module calibration tests November 14 th 2012

79

Sensor: SINTEF_2011_WA18_S5

• Planar 300 μm thick• Bump bonded to a ROC on 11-17-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

17th 2012• IVs on wafer, before, and on testboard

measured

Page 80: FPIX_flip_chip_test module calibration tests November 14 th 2012

80

A18 S5 - Leakage currentBreakdown [V]

On waferBefore BBMTestboard

Page 81: FPIX_flip_chip_test module calibration tests November 14 th 2012

81

A18 S5 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 493 dead bumps)Not repeatable and degrades at higher biases

Bias = -200 V

Page 82: FPIX_flip_chip_test module calibration tests November 14 th 2012

82

A18 S5 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA18_S5 pixel alive test at FORWARD BIAS

SINTEF_2011_WA18_S5 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA18_S5 pixel alive test with LIGHT at REVERSE BIAS

Page 83: FPIX_flip_chip_test module calibration tests November 14 th 2012

83

A18 S5 - Radioactive source test (Sr-90)

Bias = -200VMean = 27.6 ke-MP = 22.5 ke-Thickness = 300 μm

Source hit map

Page 84: FPIX_flip_chip_test module calibration tests November 14 th 2012

84

Sensor: SINTEF_2011_WA18_S6

• Planar 300 μm thick• Bump bonded to a ROC on 11-17-2012

- Flip-chipped then reflow• Assembled on a new testboard on November

17th 2012• IVs on wafer, before, and on testboard

measured

Page 85: FPIX_flip_chip_test module calibration tests November 14 th 2012

85

A18 S6 - Leakage currentBreakdown [V]

On waferBefore BBMTestboard

Page 86: FPIX_flip_chip_test module calibration tests November 14 th 2012

86

A18 S6 - Full calibration summary

ROC pixel alive test

Bump-bonding test (found 1 dead bumps)

Bias = -200 V

Page 87: FPIX_flip_chip_test module calibration tests November 14 th 2012

87

A18 S6 - Testing bum-bond with forward bias + light test

SINTEF_2011_WA18_S6 pixel alive test at FORWARD BIAS

SINTEF_2011_WA18_S6 pixel alive test at REVERSE BIAS

REVERSE BIAS = -200 VFORWARD BIAS = +10 V

SINTEF_2011_WA18_S6 pixel alive test with LIGHT at REVERSE BIAS

Page 88: FPIX_flip_chip_test module calibration tests November 14 th 2012

88

A18 S6 - Radioactive source test (Sr-90)

Bias = -200VMean = 30 ke-MP = 25 ke-Thickness = 300 μm

Source hit map

Page 89: FPIX_flip_chip_test module calibration tests November 14 th 2012

89

November 27, 2012

Page 90: FPIX_flip_chip_test module calibration tests November 14 th 2012

90

Noise study

• SCurve tests performed at– 22 °C– Vbias [-V] = 25, 50, 75, 100, 125, 150, 175, 200,

250, and 300– Almost no noise date at Vbias = -25 V

Page 91: FPIX_flip_chip_test module calibration tests November 14 th 2012

91

SINTEF 2011 WA15

Page 92: FPIX_flip_chip_test module calibration tests November 14 th 2012

92

SINTEF 2011 WA16

Page 93: FPIX_flip_chip_test module calibration tests November 14 th 2012

93

SINTEF 2011 WA17