logiq p5 pro - imv imaging · 2016. 6. 22. · ge ultrasound - europe: +49-212-28020 p.o. box 1105...

13
GE Medical Systems – Americas: 888-202-5528 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. GE Ultrasound - Europe: +49-212-28020 P.O. Box 1105 60 D-42665 Solingen GE Ultrasound - Asia: 65-291-8528 Singapore: 298 Tiong Bahru Road, #15-01/06 Tiong Bahru Plaza, Singapore 168730 Internet: http://www.gemedicalsystems.com Rev4.0.1 September 23. 2010 LOGIQ P5 Pro TruScan Imaging Technology Product Description User Interface Operator Keyboard The LOGIQ® P5 Pro is a high performance, highly mobile and easy to use multipurpose color Doppler Ultrasound imaging system designed for Obstetrical, Gynecological, Cardiac, Musculoskeletal, Vascular, Urological, Small Parts, Superficial, Pediatric, Neonatal, Transcranial, Abdominal and other applications. Keyboard Width: 43cm Keyboard Height: 84cm Alphanumeric Keyboard Ergonomic Hard Key Operations Indicator Lights Identify Activated Keys Integrated Recording Keys for Remote Control of Up to 2 Peripheral Devices and DICOM Devices 8 TGC Pods, with Re-mapping Functionality at Any Depth System Architecture Monitor TruScan™ Architecture – GE’s innovative, software-intensive platform provides exceptional computational power, image-manipulation capabilities, workflow flexibility and product upgradeability. 15 inch TFT LCD Electrical Power XGA Format: - Display size: 1024 x 768 Voltage: 100-120Vac or 220-240Vac - Recording area: 800 x 600 Frequency: 50/60 Hz Adjustable Tilt/Swivel Power: Max. 750 VA with Built-in and On-Board Peripherals Digital Brightness/Contrast Adjustment TruAccess is the GE advanced, raw-data processing technology that has helped transform ultrasound imaging. By accessing raw data, TruScan achieves excellent image quality and ensures exceptional image management capabilities. Maximum Thermal Output: 2200 BTU/hr. System Overview Console Design Applications 2 Active Probe Ports Abdominal Integrated HDD (Capacity: 250GB) Obstetrical Integrated DVD-R/W Drive SmartScan utilizes new advances in operating algorithms and system operations to aid in improving image acquisition and patient throughput while helping to increase diagnostic confidence and exam consistency. Gynecological On-board Storage for Peripherals (Max 3 peripherals) Cardiac Musculoskeletal Wheels Vascular - Wheel diameter: 12.5cm Urological - Integrated locking mechanism that provides rolling lock and optional caster swivel lock Small Parts and Superficial ComfortScan, our advanced user- friendly designed software suite helps to maximize productivity. It simplifies every exam you perform. Pediatric and Neonatal Transcranial Probe Holders, Removable for Cleaning and Washing Scanning Methods Gel Holder, Removable for Cleaning and Washing Electronic Sector General Specification Electronic Convex Air Filters, Easily Removable Electronic Linear Dimensions and Weight Front Handle Transducer Types Height: 135/141 cm Rear Handle (Option) Sector Phased Array Width: 43cm Drawer (Option) Convex Array Depth: 64cm Probe Cable Hanger (Option) Microconvex Array Weight: approx. 75 kg (165 lb.) Urology Probe Holder (Option) LOGIQ P5 DOC0518685 Page 1 of 13

Upload: others

Post on 11-Feb-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

  • GE Medical Systems – Americas: 888-202-5528 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. GE Ultrasound - Europe: +49-212-28020 P.O. Box 1105 60 D-42665 Solingen GE Ultrasound - Asia: 65-291-8528 Singapore: 298 Tiong Bahru Road, #15-01/06 Tiong Bahru Plaza, Singapore 168730

    Internet: http://www.gemedicalsystems.com Rev4.0.1 September 23. 2010

    LOGIQ P5 Pro TruScan Imaging Technology

    Product Description User Interface

    Operator Keyboard The LOGIQ® P5 Pro is a high performance, highly mobile and easy to use multipurpose color Doppler Ultrasound imaging system designed for Obstetrical, Gynecological, Cardiac, Musculoskeletal, Vascular, Urological, Small Parts, Superficial, Pediatric, Neonatal, Transcranial, Abdominal and other applications.

    • Keyboard Width: 43cm Keyboard Height: 84cm

    • Alphanumeric Keyboard • Ergonomic Hard Key Operations • Indicator Lights Identify Activated Keys • Integrated Recording Keys for Remote

    Control of Up to 2 Peripheral Devices and DICOM Devices

    • 8 TGC Pods, with Re-mapping Functionality at Any Depth

    System Architecture

    Monitor TruScan™ Architecture – GE’s innovative, software-intensive platform provides exceptional computational power, image-manipulation capabilities, workflow flexibility and product upgradeability.

    • 15 inch TFT LCD Electrical Power • XGA Format:

    - Display size: 1024 x 768 • Voltage: 100-120Vac or 220-240Vac - Recording area: 800 x 600 • Frequency: 50/60 Hz

    • Adjustable Tilt/Swivel • Power: Max. 750 VA with Built-in and On-Board Peripherals • Digital Brightness/Contrast Adjustment TruAccess is the GE advanced, raw-data

    processing technology that has helped transform ultrasound imaging. By accessing raw data, TruScan achieves excellent image quality and ensures exceptional image management capabilities.

    • Maximum Thermal Output: 2200 BTU/hr. System Overview

    Console Design Applications • 2 Active Probe Ports • Abdominal • Integrated HDD (Capacity: 250GB) • Obstetrical • Integrated DVD-R/W Drive

    SmartScan utilizes new advances in operating algorithms and system operations to aid in improving image acquisition and patient throughput while helping to increase diagnostic confidence and exam consistency.

    • Gynecological • On-board Storage for Peripherals (Max 3 peripherals) • Cardiac

    • Musculoskeletal • Wheels • Vascular - Wheel diameter: 12.5cm • Urological - Integrated locking mechanism that

    provides rolling lock and optional caster swivel lock

    • Small Parts and Superficial ComfortScan, our advanced user-friendly designed software suite helps to maximize productivity. It simplifies every exam you perform.

    • Pediatric and Neonatal • Transcranial • Probe Holders, Removable for Cleaning

    and Washing Scanning Methods • Gel Holder, Removable for Cleaning

    and Washing • Electronic Sector

    General Specification • Electronic Convex • Air Filters, Easily Removable • Electronic Linear Dimensions and Weight • Front Handle

    Transducer Types • Height: 135/141 cm • Rear Handle (Option) • Sector Phased Array • Width: 43cm • Drawer (Option) • Convex Array • Depth: 64cm • Probe Cable Hanger (Option) • Microconvex Array • Weight: approx. 75 kg (165 lb.) • Urology Probe Holder (Option)

    LOGIQ P5 DOC0518685 Page 1 of 13

  • - B-Flow + PW (option) • MpegView • Linear Array • Dual B (B/B) • Key Macro • Doppler Pencil

    - Dual B + CFM or PDI • Network Storage • Bi-plane Microconvex Array • Selectable Alternating Modes • Auto Focus Operating Modes - B/M

    • Wide Field of View on convex - B/PW • B-Mode • Simultaneous Display on BE9CS probe - B + CFM/M • M-Mode • Quick Save - B + CFM (PDI)/PW • Anatomical M Mode (option) - 3D – Mode • B-Steer

    • Color Flow Mode (CFM) - 3D – Color Mode (option) • Virtual Convex • Power Doppler Imaging (PDI) with

    Directional Map - B/CW (option) • ED Report - B + CFM (PDI)/CW (option) • Multi Image Split Screen • PW Doppler with High PRF - Live and/or frozen System Options • Steered CW Doppler (option) - B + B/CFM or PDI • DICOM 3.0 Connectivity • Dedicated CW Doppler (option) - Independent CINE playback • InSite Capability • B-Flow - Quad screen format

    • CrossBeam • Real Time 4D • Zoom: Write/Read/Pan • B-Flow • Colorized Image System Standard Features • ECG - Colorized B • SRI - Colorized M • CW Doppler

    • Hard Disk partition of 50GB for image storage

    - Colorized PW • Anatomical M-Mode - Colorized CW (option) • LOGIQView

    • Without compression: • Time line Display • Easy 3D (Baby Face, Vascular) • Raw DICOM: HDD capacity 33.000

    images - Independent Dual B/PW

    Display • Advanced 3D, with 3D Landscape • 3-Pedal Foot Switch, with

    Programmable Functionality Display Formats: • DICOM: Top/ Bottom selectable format (Size:

    1/2:1/2; 1/3:2/3; 2/3:1/3) - Image Only: HDD capacity 71.000 images

    • Rear Handle Side/Side selectable format • Drawer

    • CINE Memory Frames (60 sec) (192MB) >1000 depending on probe, imaging depth, line density and field of view

    (1/2:1/2; 1/3:2/3; 0:1) • Probe Cable Hanger Switchable after freeze • Urology Probe Holder

    • 3rd. Probe Port Kit • Real-time Triplex Mode at any Depth

    and PRF Display Annotation • 1-Pedal Foot Switch • 1st. Patient Name: First, Last, &

    Middle name each store 27 characters. Up to 62 total characters displayed.

    • Remote Control Switch • Automatic Optimization

    • PDF report conversion - Auto Tissue Optimization: ATO

    • Real Time 4D - Auto Spectrum Optimization: ASO

    • 2nd. Patient ID: 31 characters. Up to 27 total characters displayed.

    - Auto Color Optimization: ACO • Coded harmonic Imaging Media & Peripheral Options

    • Age, Sex and Birth Date (optional) • Coded Excitation • Integrated Mounting Kits and Remote Controls Provided for

    • Hospital Name: 23 characters. • Virtual Convex • Date: 3 Types selectable

    MM/DD/YY, DD/MM/YY,

    • Patient Information Database - B/W Digital thermal printer • Image Archive on Hard Drive and DVD - Color Digital thermal printer • Easy Backup for Media - DVD Video Recorder • YY/MM/DD • Report Designer • Time: 2 types selectable • Vascular Calcs Display Modes • 24 hours, 12 hours • Cardiac Calcs • Live and Stored Display Format: Full

    size and split screen - both w/ thumbnails. For Still and CINE

    • Gestational Age from LMP/EDD/GA/BBT • OB Calcs

    • Probe Name • Fetal Trending • Gray Map names • Review Image Format: 4x4, and

    "thumbnails". For Still and CINE • Multi Gestational Calcs

    • Probe Orientation • Hip Dysplasia Calcs • Simultaneous Capability • Depth Scale Marker • Gynecological Calcs

    - B/PW • Lateral Scale Marker • Urological Calcs - B/CFM or PDI • Focal Zone Markers • Renal Calcs - B/M • Image Depth • Real-time Auto Doppler Calculations - B + CFM/M • Zoom Depth • TruAccess, Raw Data Processing - Realtime Triplex Mode • B-Mode

    LOGIQ P5 Page 2 of 13

    (B + CFM or PDI/PW) • On-board Reporting

  • - Gain • Internal Hard Drive partition for Image Storage 50 GB - Dynamic Range

    - Imaging Frequency - Edge Enhance - Frame Averaging - Gray Map - ATO On/Off

    • M-Mode - Gain - Dynamic Range - Time Scale

    • Doppler Mode - Gain - Angle - Sample Volume Depth and Width - Wall Filter - Velocity and/or Frequency Scale - Spectrum Inversion - Time Scale - PRF - Doppler Frequency

    • Color Flow Mode - Line Density - Frame Averaging - Packet Size - Color Scale - Color Velocity Range and Baseline - Color Threshold Marker - Color Gain - PDI - Color Scale Inversion - Color Doppler Frequency

    • TGC Curve • Acoustic Frame Rate • Cine Frame Number • VCR Counter • VCR Status • VCR Playback Counter • Body Pattern • Application Name • Measurement Results • Operator Message • Displayed Acoustic Output

    - TIS: Thermal Index Soft Tissue - TIC: Thermal Index Cranial (Bone) - TIB: Thermal Index Bone - MI: Mechanical Index

    • % of Power output • Biopsy Guide Line and/or Zone • Heart Rate General System Parameters System Setup • Pre-programmable Categories • User Programmable Preset

    Capability • Factory Default Preset Data • Languages: English, French, German,

    Spanish, Italian, Portuguese, Russian,

    Greek, Swedish, Danish, Dutch, Finnish, Norwegian, Japanese, Chinese

    • OB Report Format: 5 Types, Tokyo Univ., Osaka Univ., USA, Europe, and ASUM

    • External USB 2.0 Hard Drive support for Import, Export, DICOM Read, SaveAs and MPEGVue

    • EFBW: 10 types, Japan, USA and Europe (Tokyo Univ., Osaka Univ., Tokyo Shinozuka, JSUM, German, Shephard, Merz, Hadlock/Shephard, Williams, Brenner)

    • USB 1.1 / 2.0 Memory Stick support for SaveAs and MPEGVue

    • Network Storage support for Import, Export, DICOM Read, SaveAs, MPEGVue

    Connectivity • 348 Pre-defined Annotations and

    User Programmable Libraries/Annotations

    • Ethernet network connection • RS-232 serial data output (need a

    converter cable) • Body Patterns: 140 human types

    plus 14 animal types • DICOM 3.0 (option)

    - Verify • Customized Comment Home

    Position - Print - Store

    - Modality Worklist Complete User Manual available on board through Help (F1)

    - Storage Commitment - Modality Performed Procedure

    Step (MPPS) User Manual and Service Manual are included on CD with each system. A printed Manual is available upon request.

    - Media Exchange - Off network / mobile storage

    queue - Query/Retrieve; supported on

    Centricity and other compatible vendors

    CINE Memory/Image Memory

    - Structured Reporting; compatible with LOGIQworks

    • CINE Memory: 192MB • Dual Image CINE Display

    - Public SR Template • Quad Image CINE Display - Media Store of SR • CINE Gauge and CINE Image

    Number Display - iLinq capability • CINE Review Loop Scanning Parameters • CINE Review Speed: 20 steps (10, 20,

    30, 40, 50, 60, 70, 80, 90, 100, 150, 200, 300, 400, 500, 600, 700, 800, 900, 1000%)

    • Digital Beamformer • 2560 processing channels

    • Selectable CINE Sequence for CINE Review

    • Frame Rate: > 700 depending on settings

    • Measurements, Calculations and Annotations on CINE Playback

    • Displayed Imaging Depth: 0 – 30 cm • Minimum Depth of Field: 0 –2cm

    • Scrolling Timeline Memory (probe dependent) • Cine Capture Function • Maximum Depth of Field: 0 – 30 cm (probe dependent)

    • Transmission Focus Image Storage 1 – 8 Focus Points selectable (probe and application dependent)

    • On-board database of patient information from past exams

    • Continuous Dynamic Receive Focus/Aperture

    • Storage Format: DICOM/Raw Data • DICOM Still Image Storage Size:

    • Multi-Frequency/Wideband Technology

    - Gray Image: ~300K to ~1.3 MB - Color Image: ~900K to 1.9 MB

    • Frequency Range 1 to 13MHz • Multiframe • 256 Shades of Gray • Display Format: Full size, 4x4, and

    "thumbnails" • Up to 197 db processing dynamic range • Live image and stored image side-

    by-side display • 16,777,216 Hues of Color • Adjustable Field of View (FOV) up to

    170 degree depending on probe • CD-R storage: 650, 700 MB • DVD storage: -R 4,7GB

    • Image Reverse: Right/Left • Conversion to JPEG, AVI (SaveAs) and WMV (MPEGvue) file formats • Image Rotation: 4 steps

    LOGIQ P5 Page 3 of 13

  • • CF/PDI Vertical Size (mm) of ROI: default pre-settable

    Rotation: 0°, 90°, 180°, 270° Can be activated from a CINE loop from a live or stored image

    • CF/PDI Center Depth (mm) of ROI: default pre-settable

    Available with Color Flow Mode B-Mode M & A capability • B/M Acoustic Output: 0 – 100%,

    2% step • CF/PDI Frequency: Up to 3 steps, depend on probe PW/CW-Mode • Image Reverse: On/Off

    • Color Map: 20 types depend on application

    • Maximum and Minimum Velocity Scales

    • B Color: 10 types • Thermal Index: TIC, TIS, TIB • Transparent: 5 steps - Max: 10 m/sec • Softener: 4 steps • Color Threshold: 0 – 100 %, 5 % step - Min: 5 cm/sec • Focus Number: 8 steps • Auto Line Density: On/Off pre-

    settable • Gray Scale Map: 4 types • Focus Width: 3 types • Dynamic Range: 24 – 60, 4 dB step • Range Focus: On/Off • PW/CF Ratio: 1, 2, 4 • Base Line: 5 – 95 %, 11 steps

    • Compression: 0.5 – 1.5, 0.1 step • Accumulation: 8 steps • SV Gate: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12, 14, 16 mm • Line Density: 4 steps

    • Line Density Zoom: 4 steps Power Doppler Imaging • Angle Correct: +/- 90°, 1° step • Suppression: 6 steps • PDI Map: 13 types • Spectral Color: 6 types • Frame Average: 8 steps • CF/PDI Flash Suppression: 2 steps • PW Sweep Speed: 8 steps • Edge Enhance: 6 steps • CF/PDI Focus Depth: default pre-

    settable for 0 – 100 % of ROI in depth, 10 % step

    • Invert: On/Off • Scanning Size (FOV or Angle):

    probe dependent, see probe specifications

    • M/PW Display Format: V-1/3B, V-1/2B, V-2/3B, H-1/2B, H-1/4B, TL only

    • CF/PDI Acoustic Output: 0 – 100%, 10% step

    • Duplex: On/Off(PW only) • Gray Scale Map: 23 types • PW Acoustic Output: 0 – 100 %,

    10 % step • CF/PDI Angle Steer: 0, +/- 20° • Clear Map: 23 types • Packet Size: 5 – 16, dependent on

    probe/application • Spectral Averaging: 4 steps • Tint Map: 10 types • Time Resolution: 4 steps • Gain: 0 – 98 dB, 2 dB step

    • Spatial Filter: 6 steps • PW/CF Ratio: 1, 2, 4 • Dynamic Range: 30 – 120 dB, 3 dB step • Frame Average: 7 steps • Rejection: 15 steps

    • PRF: 280 Hz – 19600 Hz • Gain: 0 – 32 dB, 1 dB step • Depth: 1 – 30 cm, 1 cm step, depend on probe. • Power Threshold: 0 – 100 %, 5 %

    step • Wall Filter: 5 – 1500 Hz, 22 steps,

    depend on probe/application • Rejection: 6 steps • Gain: 0 – 40 dB, 0.5 dB step • PW Angle Steer: 0, +/- 5, 10, 15, 20° • Frequency: Up to 5 steps, depend

    on probe • Wall Filter: 7 steps depend on probe/application

    • PRF: 640 – 29800 Hz with PW, 50000 Hz with CW

    • CF/PDI Frequency: Up to 3 steps, depend on probe

    • Sample Volume Depth: 29 steps default pre-settable

    • Auto Line Density: On/Off pre-settable

    • Auto Line Density: On/Off pre-settable

    • CW-Mode (option) is Available on the Following Probes • Steered Linear: +/- 15° • Transparent: 5 steps - 3S

    Color Flow Mode • Invert: On/Off - 5S • Base Line: 0 – 100 %, 10 % step • Accumulation: 8 steps - 7S • Invert: On/Off - 3Sp • CF/PDI Focus Depth: default pre-

    settable for 0 – 100 % of ROI in depth, 10 % step

    - 5Sp M-Mode - P2D • Sweep Speed: 8 steps - P6D • M Color: 10 types • CF/PDI Flash Suppression: 2 steps • M/PW Display Format: V-1/3B, V-

    1/2B, V-2/3B, H-1/2B, H-1/4B, TL only • CF/PDI Acoustic Output: 0 – 100%, 10% step Coded Harmonic Imaging • Available on the Following Probes • B/M Acoustic Output: 0 – 100 %, 2 % step • CF/PDI Angle Steer: 0, +/- 20° - 3.5C • Packet Size: 5 – 16, dependent on

    probe/application - 3.5CS • Rejection: 6 steps - 3CRF • Dynamic Range: 30 – 120 dB, 3 dB step • Line Density: 5 steps - 4C • Line Density Zoom: 5 steps - 5CS • Edge Enhance: 6 steps • Frame Average: 7 steps - E8C • Gray Scale Map: 23 types • PRF: 280 Hz – 19600 Hz - E8CS • Clear Map: 23 types • Spatial Filter: 6 steps - 8C • M Gain: 0 – 98 dB, 2 dB step • Gain: 0 – 40 dB, 0.5 dB step - ERB • Wall Filter:4 steps depend on

    probe/application Anatomical M-Mode - BE9CS - 3S M-mode cursor adjustable at any plane

    • Scanning Size (FOV or Angle): probe dependent - 5S

    LOGIQ P5 Page 4 of 13

  • B-Flow (option) • Side by Side Display - 7S • Compatible with ALL linear, convex

    and sector transducers - 3Sp • Available on 8L, 9L, 10L, 11L, i739

    and T739 Probes - 5Sp • Compatible with ALL scanning

    mode - 8L • Background: On/Off - 9L • Sensitivity/PRI: 14 steps

    • Compatible with Side by Side Display

    - 11L • Line Density: 4 steps - i12L • Edge Enhance: 6 steps

    - i739 • Frame Average: 8 steps - T739 • Gray Scale Map: 23 types Coded Excitation

    • Softener: 4 steps • Clear Map: 23 types • Available on the following probes the Following Probes • Line Density: 4 steps • Tint Map: 10 types

    • Line Density Zoom: 4 steps • Dynamic Range: 30 – 120 dB, 3 dB step

    - 11L • Suppression: 6 steps - 12L • Edge Enhance: 6 steps • Rejection: 6 steps - 8C • Gray Scale Map: 23 types • Gain: 0 – 98 dB, 2 dB step - E8C • Clear Map: 23 types • Auto Line Density: On/Off pre-

    settable - E8CS

    • Tint Map: 10 types - BE9CS • Gain: 0 – 98 dB, 2 dB step • Accumulation: 8 steps - ERB • Dynamic Range: 30 – 120 dB, 3 dB

    step LOGIQView (option) Virtual Convex • Rejection: 6 steps • Available on all probes • Provides a Convex Field of View for

    Convex, Linear and Sector probes • Auto Line Density: On/Off pre-

    settable • Extended Field of View Imaging • For use in B-Mode • Available on the Following Probes • Frequency: Up to 5 steps, depend on probe • LOGIQView Status - 8L • Auto detection of scan direction - 9L • Pre or post-process zoom up to 10X - 10L CrossBeam (option) • Rotation - 11L

    • Provides Spatial Compounding • Auto best fit on monitor - 12L • Available on the linear and convex

    probes • Measurements in B-Mode - i12L • Up to 60 cm scan length - T739

    - 3CRF - I739 - 4C - 3S Easy 3D (option) - 5CS - 5S Colorize - 8C - 7S Threshold (Opacification) - E8C - 3Sp - E8CS Render - 5Sp - 8L Texture - ERB - Linear - 9L Gray Surface - 4C - 11L - 4D3C Scalpel - I12L - 5CS Auto Movie - I739 - 3CRF Undo - T739 Reset - BE9CS

    Automatic Optimization - 4D3C-L • Optimize B-Mode, B-Flow image to

    improve contrast resolution. - 4DE7C Advanced 3D (option) - 4D8C • Acquisition of Color data • Selectable amount of contrast

    resolution improvement (low, medium, high)

    - ERB • Automatic rendering • Provides 3, 5 or 7 angles for

    Compounding on Linear probes, 3 or 5 angles on Convex probes

    • 3D Landscape technology • 3D Movie • Auto-TGC in B-Mode and Color –

    adjusts overall and axial gain • Main Mode • Compatible with Side by Side Display

    • Optimize Spectral Waveform – adjusts baseline, invert, PRF (on live image), and angle correction

    Real Time 4D (option) • Compatible with: Color mode,

    Timeline mode, SRI-HD, Coded Harmonic Imaging, Virtual Convex

    • Acquisition Modes: - Realtime 4D Bmode • Algorithm works on focal zone/

    number and depth changes - Static 3D Bmode • Visualization Modes: • Available on stored or live image SRI-HD - 3D Rendering (diverse surface and intensity projection modes) • Available in B-Mode, B-Flow, PW Doppler, Color • High Definition Speckle Reduction Imaging - Sectional Planes (3 Section planes perpendicular to each other)

    LOGIQ P5 Page 5 of 13

    • Provides 6 levels of speckle reduction

  • • Render Mode: - Image Reverse • Angle between two lines - Surface texture, Surface Smooth,

    max-, min- and X-ray (average intensity projection), Gradient, Inversion, Glass Body, Mix Mode of two render Modes

    - Image Rotation - Gray Map General M-Mode - Colorized B and M • M-Depth - Gain • Distance - Dynamic Range • Time

    • Curved 3point Render start - TGC • Slope • 3D Movie - Compression • Heart Rate • Scalpel: 3D Cut tool - Rejection • Display Format: - Frame Averaging General Doppler

    - Quad: A-/B-/C-Plane/3D - B Softener Measurements/Calculations - Dual: A-Plane/3D - Suppression • Velocity - Single: 3D or A- or B- or C-Plane - Sweep Speed for M-Mode • Time - 4D3C, 4DE7C, 4D8C supported - Anatomical M Mode • A/B Ratio (Velocities / Frequency Ratio

    • PW/CW-Mode Pre-Processing - ASO (Auto Spectral Optimization) • PS (Peak Systole) - Base Line Shift • Write Zoom up to 8x • ED (End Diastole) - Gray Map • B/M-Mode • PS/ED (PS/ED Ratio) - Post Gain - Gain • ED/PS (ED/PS Ratio) - Compression - TGC • AT (Acceleration Time) - Rejection - Acoustic Output • ACCEL (Acceleration) - Spectral Averaging - Transmission Focus Position • TAMAX (Time Averaged Maximum

    Velocity - Colorized D - Transmission Focus Number - Display Format - Transmission Focus Width • Volume Flow (TAMEAN and Vessel

    Area) - Sweep Speed - Imaging Frequency

    • Color Flow Mode - Edge Enhancement • Heart Rate - ACO (Auto Color Optimization) - Line Density Control • PI (Pulsatility Index) - Base Line Shift - Live Anatomical M-mode • RI (Resistivity Index) - Color Map PW/CW-Mode - Post Gain - Gain Real-time Doppler Auto - Frame Averaging - Dynamic Range Measurements / Calculations - CFM Display Threshold - Acoustic Output • PS (Peak Systole) - Angle Correct (PW mode) - Doppler Frequency • ED (End Diastole) - Quick Angle Correct (PW mode) - Velocity Scale (PRF) • MD (Minimum Diastole) - Auto Angle Correct (PW mode) - PW/CF Ratio • PI (Pulsatility Index) - Spectral Invert for Color and

    Doppler - Wall Filter

    • RI (Resistivity Index) - Time Resolution • AT (Acceleration Time) • 3D reconstruction from a stored

    CINE loop - Sample Volume Gate for PW-

    Mode • ACC (Acceleration) • PS/ED (PS/ED Ratio) • Anatomical M-Mode Length • ED/PS (ED/PS Ratio) • Accumulation Depth • HR (Heart Rate) • Cine Capture • Color Flow Mode • TAMAX (Time Averaged Maximum

    Velocity) - CFM Gain

    Physiological Input Panel (Option) - CFM Velocity Scale (PRF) • PVAL (Peak Velocity Value) • Physiological Input - Acoustic Output • Volume Flow (TAMEAN and Vessel

    Area) - ECG, 1 channel - CFM Frequency

    • Dual R-Trigger - Wall Filter • Pre-settable ECG R Delay Time - Packet Size OB Measurements/Calculations • Pre-settable ECG Position - CFM Spatial Filter • Gestational Age by: • Adjustable ECG Gain Control - CFM Line Density

    - GS (Gestational Sac) • Automatic Heart Rate Display - CFM Regression Filter - CRL (Crown Rump Length) - Accumulation

    Measurements / Calculations - FL (Femur Length) - BPD (Biparietal Diameter) Post-Processing w/ TruAccess (Raw

    Data)

    - AC (Abdominal Circumference) General B-Mode - HC (Head Circumference) • SRI-HD – Selectable level of Speckle Reduction Imaging

    • Depth & Distance - APTD x TTD (Anterior/Posterior Trunk Diameter by Transverse Trunk Diameter)

    • Circumference (Ellipse / Trace) • Max Read Zoom to 16x • Area (Ellipse / Trace) • B/M-Mode • Volume (Ellipsoid) - LV (Length of Vertebra) - ATO (Auto Tissue Optimization) • % Stenosis (Area or Diameter)

    LOGIQ P5 Page 6 of 13

    - FTA (Fetal Trunk Cross-sectional

  • • DIAS MCCA (Diastolic Mid Common Carotid Artery)

    Area) Left Atrium Area (LAA(d), LAA(s)) - HL (Humerus Length) Left Atrium Volume, Single Plane,

    Method of Disk (LAEDV A2C, LAESV A2C ) (LAEDV A4C, LAESV A4C)

    - BD (Binocular Distance) • SYS PCCA (Systolic Proximal Common Carotid Artery) - FT (Foot Length)

    - OFD (Occipital Frontal Diameter) • DIAS PCCA (Diastolic Proximal Common Carotid Artery) • Left Ventricle - TAD (Transverse Abdominal

    Diameter) Left Ventricle Mass (LVPWd, LVPWs) • SYS DICA (Systolic Distal Internal Carotid Artery) - TCD (Transverse Cerebellum

    Diameter) Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI Ds) • DIAS DICA (Systolic Distal Internal

    Carotid Artery) - THD (Thorax Transverse Diameter) Left Ventricle Internal Diameter (LVIDd, LVI Ds) - TIB (Tibia Length) • SYS MICA (Systolic Mid Internal

    Carotid Artery) - ULNA (Ulna Length) Left Ventricle Length (LVLd, LVLs) • Estimated Fetal Weight (EFW) by: • DIAS MICA (Diastolic Mid Internal

    Carotid Artery) Left Ventricle Outflow Tract Diameter (LVOT Diam)

    - AC, BPD - AC, BPD, FL • SYS PICA (Systolic Proximal Internal

    Carotid Artery) Left Ventricle Posterior Wall Thickness (LVPWd, LVPWs)

    - AC, BPD, FL, HC - AC, FL • DIAS PICA (Diastolic Proximal

    Internal Carotid Artery) - AC, FL, HC Left Ventricle Length (LV Major) - AC, HC • SYS DECA (Systolic Distal External

    Carotid Artery) Left Ventricle Width (LV Minor)

    • Calculations and Ratios Left Ventricle Outflow Tract Area (LVOT) - FL/BPD • DIAS DECA (Diastolic Distal External Carotid Artery) Left Ventricle Area, Two Chamber/Four

    Chamber/Short Axis (LVA (d), LVA (s)) - FL/AC - FL/HC • SYS PECA (Systolic Proximal External

    Carotid Artery) Left Ventricle Endocardial Area, Width (LVA (d), LVA(s))

    - HC/AC - CI (Cephalic Index) • DIAS PECA (Diastolic Proximal

    External Carotid Artery) - AFI (Amniotic Fluid Index) Left Ventricle Epicardial Area, Length (LVAepi (d), LVAepi (s)) • Measurements / Calculations by:

    Jeanty, Merz, Tokyo University, Mercer, Hansmann, Erickson, Hill, Shephard, Hadlock, Hohler, Campbell

    • VERT (Systolic Vertebral Velocity) • SUBCLAV (Systolic Subclavian

    Velocity) Left Ventricle Mass Index (LVPWd, LVPWs) • Summary Reports Ejection Fraction, Teichholz/Cube (LVIDd, LVIDs)

    • Fetal Graphical Trending • Mean IMT Measurement Tools • Growth Percentiles

    • Multi-Gestational Calculations (4) Left Ventricle Posterior Wall Fractional Shortening (LVPWd, LVPWs) Cardiac • Fetal Qualitative Description

    (Anatomical survey) Measurements/Calculations Left Ventricle Stroke Index, Teichholz/Cube (LVIDd, LVIDs, and Body Surface Area)

    • Fetal Environmental Description (Biophysical profile) B-Mode Measurements

    • Aorta • Programmable OB Tables Left Ventricle Fractional Shortening (LVIDd, LVIDs)

    Aortic Root Diameter (Ao Root Diam) • Over 20 selectable OB Calcs • Expanded Worksheets Aortic Arch Diameter (Ao Arch Diam) Left Ventricle Stroke Volume,

    Teichholz/Cubic (LVIDd, LVIDs) Ascending Aortic Diameter (Ao Asc)

    GYN Measurements/Calculations Descending Aortic Diameter (Ao Desc Diam) • Right Ovary Length, Width, Height Left Ventricle Stroke Index, Single Plane,

    Two Chamber, Method of Disk (LVI Dd, LVIDs, LVSd, LVSs)

    • Left Ovary Length, Width, Height Aorta Annulus Diameter (Ao Annulus Diam)

    • Uterus Length, Width, Height • Ovarian Volume

    Left Ventricle Stroke Index, Single Plane, Four Chamber, Method of Disk (LVI Dd, LVIDs, LVSd, LVSs)

    Aorta Isthmus (Ao Isthmus) • ENDO (Endometrial thickness) Aorta *** (Ao st junct) • Ovarian RI

    • Uterine RI • Aortic Valve Left Ventricle Stroke Index, Bi-Plane, Bullet, Method of Disk (LVAd, LVAs)

    • Follicular measurements Aortic Valve Cusp Separation (AV Cusp) • Summary Reports Aortic Valve Area Planimetry (AVA Pl

    ani met ry) Interventricular Septum (IVS) Vascular Measurements/Calculations

    Left Ventricle Internal Diameter (LVI D) *** ( Trans AVA) Left Ventricle Posterior Wall Thickness (LVPW)

    • Left Atrium • SYS DCCA (Systolic Distal Common

    Carotid Artery) Left Atrium Diameter (LA Diam) • Mitral Valve LA Length (LA Major) • DIAS DCCA (Diastolic Distal Common Carotid Artery) Mitral Valve Annulus Diameter (MV Ann

    Diam) LA W idth (LA Minor) Left Atrium Diameter to AoRoot Diameter Ratio (LA/Ao Ratio)

    • SYS MCCA (Systolic Mid Common Carotid Artery) E-Point-to-Septum Separation (EPSS)

    LOGIQ P5 Page 7 of 13

  • Tricuspid Valve Area (TV Panimetry) Interventricular Septum Thickness (IVSd, IVSs)

    Mitral Valve Area by Pressure Half Time (MVA By PHT) Tricuspid Valve Annulus Diameter (TV

    Annulus Diam) Pericard Effusion (PE(d)) Mitral Valve Area Planimetry (MVA Planimetry) Interventricular Septum (IVS) Fractional

    Shortening (IVSd, IVSs) M-Mode Measurements • Pulmonic Valve • Tricuspid Valve • Aorta Pulmonic Valve Area (PV Planimetry)

    QRS complex to end of envelope (Q-to-TV close)

    Pulmonic Valve Annulus Diameter (PV Annulus Diam)

    Aortic Root Diameter (Ao Root Diam) • Aortic Valve

    Pulmonic Diameter (Pulmonic Diam) Aortic Valve Diameter (AV Diam) Doppler Mode Measurements • Right Atrium Aortic Valve Cusp Separation (AV Cusp) • Aortic Valve Right Atrium Diameter, Length (RAD

    Ma) Aortic Valve Ejection Time (LVET)

    • Left Atrium Aortic Insufficiency Mean Pressure Gradient (AR Trace) Right Atrium Diameter, Width (RAD Mi) Left Atrium Diameter to AoRoot

    Diameter Ratio (LA/Ao Ratio) Right Atrium Area (RAA) Aortic Insufficiency Peak Pressure Gradient (AR Vmax) Right Atrium Volume, Single Plane,

    Method of Disk (RAAd) Left Atrium Diameter (LA Diam)

    • Left Ventricle Aortic Insufficiency End Diastole Pressure Gradient (AR Trace) Right Atrium Volume, Systolic, Single

    Plane, Method of Disk (RAAs) Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI Ds) Aortic Insufficiency Mean Velocity (AR

    Trace) • Right Ventricle Left Ventricle Internal Diameter (LVIDd, LVI Ds) Right Ventricle Outflow Tract Area

    (RVOT Planimetry) Aortic Insufficiency Mean Square Root Velocity (AR Trace) Left Ventricle Posterior Wall Thickness

    (LVPWd, LVPWs) Left Pulmonary Artery Area (LPA Area) Aortic Insufficiency Velocity Time Integral (AR Trace) Right Pulmonary Artery Area (RPA Area) Left Ventricle Ejection Time (LVET)

    Right Ventricle Internal Diameter (RVIDd, RVIDs)

    Left Ventricle Pre-Ejection Period (LVPEP)

    Aortic Valve Mean Velocity (AV Trace) Aortic Valve Mean Square Root Velocity (AV Trace) Right Ventricle Diameter, Length (RVD

    Ma) Interventricular Septum (IVS) Left Ventricle Internal Diameter (LVI D) Aortic Valve Velocity Time Integral (AV

    Trace) Right Ventricle Diameter, Width (RVD Mi)

    Left Ventricle Posterior Wall Thickness (LVPW) Aortic Valve Mean Pressure Gradient

    (AV Trace) • Mitral Valve Right Ventricle Wall Thickness (RVAWd, RVAWs) E-Point-to-Septum Separation (EPSS) Aortic Valve Peak Pressure Gradient

    (AR Vmax) Right Ventricle Outflow Tract Diameter (RVOT Diam)

    Mitral Valve Leaflet Separation (D-E Excursion) Aortic Insufficiency Peak Velocity (AR

    Vmax) Left Pulmonary Artery (LPA) Mitral Valve Anterior Leaflet Excursion (D-E Excursion) Main Pulmonary Artery (MPA) Aortic Insufficiency End-Diastolic

    Velocity (AR Trace) Right Pulmonary Artery (RPA) Mitral Valve D-E Slope (D-E Slope) • System Mitral Valve E-F Slope (E-F Slope) Aortic Valve Peak Velocity (AV Vmax)

    • Pulmonic Valve Interventricular Septum Thickness (IVSd, IVSs)

    Aortic Valve Peak Velocity at Point E (AV Vmax) QRS complex to end of envelope (Q-to-

    PV close) Inferior Vena Cava Aorta Proximal Coarctation (Coarc Pre-Duct) • Right Ventricle Pulmonary Artery Diameter (MPA)

    Systemic Vein Diameter (Systemic Diam)

    Right Ventricle Internal Diameter (RVIDd, RVIDs)

    Aorta Distal Coarctation (Coarc Post-Duct)

    Patent Ductus Arterosis Diameter (PDA Diam)

    Right Ventricle Wall Thickness (RVAWd, RVAWs)

    Aortic Valve Insufficiency Pressure Half Time (AR PHT)

    Pericard Effusion (PEs) Right Ventricle Outflow Tract Diameter (RVOT Diam)

    Aortic Valve Flow Acceleration (AV Trace) Patent Foramen Ovale Diameter (PFO

    Diam) Right Ventricle Ejection Time (RVET) Aortic Valve Pressure Half Time (AV Trace) Ventricular Septal Defect Diameter

    (VSD Diam) Right Ventricle Pre-Ejection Period (RVPEP) Aortic Valve Acceleration Time (AV Acc

    Ti me) Interventricular Septum (IVS) Fractional Shortening (IVSd, IVSs)

    Velocity Circumferential Fiber Shortening (Vcf) Aortic Valve Deceleration TIme (AV

    Trace) • Tricuspid Valve • System

    LOGIQ P5 Page 8 of 13

  • Pulmonic Valve Mean Velocity (PV Trace)

    Mitral Valve Flow Deceleration (MV Trace)

    Aortic Valve Ejection Time (AVET) Aortic Valve Acceleration to Ejection Time Ratio (AV Acc Time, AVET) Pulmonic Valve Mean Square Root

    Velocity (PV Trace) Mitral Valve Pressure Half Time (PV PHT) Aortic Valve Area according to PHT

    • Left Ventricle Pulmonic Valve Velocity Time Integral (PV Trace)

    Mitral Valve Flow Acceleration (MV Trace) Left Ventricle Outflow Tract Peak

    Pressure Gradient (VLOT Vmax) Pulmonic Insufficiency Pressure Half Time (PR PHT)

    Mitral Valve E-Peak to A-Peak Ratio (A-C and D-E) (MV E/ARatio) Left Ventricle Outflow Tract Peak

    Velocity (LVOT Vmax) Pulmonic Valve Flow Acceleration (PV Acc Time)

    Mitral Valve Acceleration Time (MV Acc Time) Left Ventricle Outflow Tract Mean

    Pressure Gradient (LVOT Trace) Pulmonic Valve Acceleration Time (PV Acc Time)

    Mitral Valve Deceleration Time (MV Dec Time) Left Ventricle Outflow Tract Mean

    Velocity (LVOT Trace) Pulmonic Valve Ejection Time (PVET) Mitral Valve Ejection Time ((MV Trace) Pulmonic Valve Pre-Ejection Period (PVPEP)

    Mitral Valve A-Wave Duration (MV A Dur)

    Left Ventricle Outflow Tract Mean Square Root Velocity (LVOT Trace)

    QRS complex to end of envelope (Q-to-PV close)

    Mitral Valve Time to Peak (MV Trace) Left Ventricle Outflow Tract Velocity Time Integral (LVOT Trace) Mitral Valve Acceleration

    Time/Deceleration Time Ratio (MVAcc/Dec Time)

    Pulmonic Valve Acceleration to Ejection TIme Ratio (PV Acc Time, PVET)

    Left Ventricle Ejection Time (LVET) Cardiac Output by Aortic Flow (AVA Pl ani met ry, AV Trace) Pulmonic Valve Pre-Ejection to Ejection

    Time Ratio (PVPEP, PVET) Stroke Volume Index by Mitral Flow (MVA Planimetry, MVTrace) Stroke Volume Index by Aortic Flow

    (AVA Planimetry, AV Trace) • Right Ventricle Mitral Valve Area from Continuity Equation (MVAPlanimetry, LVOT Vmax, MV Vmax)

    • Mitral Valve Right Ventricle Outflow Tract Peak Pressure Gradient (RVOT Vmax) Mitral Valve Regurgitant Flow

    Acceleration (MR Trace) • Pulmonic Valve Right Ventricle Systolic Pressure (RVOT Vmax) Pulmonic Insufficiency Peak Pressure

    Gradient (PR Vmax) Mitral Valve Regurgitant Mean Velocity (MR Trace) Right Ventricle Outflow Tract Peak

    Velocity (RVOT Vmax) Pulmonic Insufficiency End-Diastolic Pressure Gradient (PRTrace)

    Mitral Regurgitant Mean Square Root Velocity (MR Trace) Right Ventricle Diastolic Pressure (RVOT

    Trace) Pulmonic Valve Peak Pressure Gradient (PV Vmax)

    Mitral Regurgitant Mean Pressure Gradient (MR Trace) Right Ventricle Outflow Tract Velocity

    Time Integral (RVOTTrace) Pulmonic End-Diastolic Pressure Gradient (PR Trace)

    Mitral Regurgitant Velocity Time Integral (MR Trace) Right Ventricle Ejection Time (RV Trace)

    Stroke Volume by Pulmonic Flow (RVOTPlanimetry, RVOTTrace)

    Pulmonic Insufficiency Peak Velocity (PR Vmax)

    Mitral Valve Mean Velocity (MR Trace) Mitral Valve Mean Square Root Velocity (MR Trace) Right Ventricle Stroke Volume Index by

    Pulmonic Flow (RVOT Planimetry, RVOT Trace)

    Pulmonic Insufficiency End-Diastolic Velocity (Prend Vmax) Mitral Valve Velocity Time Integral (MR

    Trace) Pulmonic Valve Peak Velocity (PV Vmax) • System Mitral Valve Mean Pressure Gradient

    (MR Trace) Pulmonary Artery Peak Velocity (PV Vmax)

    Pulmonic End-Diastolic Velocity (PV Trace) Mitral Regurgitant Peak Pressure

    Gradient (MR Vmax) Pulmonary Vein Velocity Peak A (reverse) (P Vein A)

    Pulmonary Artery Diastolic Pressure (PV Trace) Mitral Valve Peak Pressure Gradient

    (MR Vmax) Pulmonary Vein Peak Velocity (P Vein D, P Vein S)

    Pulmonic Insufficiency Mean Pressure Gradient (PR Trace) Mitral Regurgitant Peak Velocity (MR

    Vmax) Systemic Vein Peak Velocity (PDA Diastolic, PDA Systolic)

    Pulmonic Valve Mean Pressure Gradient (PV Trace) Mitral Valve Peak Velocity (MR Vmax)

    Ventricular Septal Defect Peak Velocity (VSD Vmax)

    Pulmonic Insufficiency Mean Velocity (PR Trace)

    Mitral Valve Velocity Peak A (MV A Velocity)

    Atrial Septal Defect (ASD Diastolic, ASD Systolic)

    Pulmonic Insufficiency Mean Square Root Velocity(PR Trace)

    Mitral Valve Velocity Peak E (MV E Velocity)

    Pulmonary Artery Velocity Time Integral (PV Trace)

    Pulmonic Insufficiency Velocity Time Integral (PR Trace)

    Mitral Valve Area according to PHT (MV PHT)

    LOGIQ P5 Page 9 of 13

  • Systemic Vein Velocity Time Integral (PDA Trace) Pulmonary Vein A-Wave Duration (P Vein A Dur) IsoVolumetric Relaxation Time (IVRT) IsoVolumetric Contraction Time (IVCT) Pulmonary Vein S/D Ratio (P Vein D, P Vein S) Ventricular Septal Defect Peak Pressure Gradient (VSD Vmax) Pulmonic-to-Systemic Flow Ratio (Qp/Qs)

    • Tricuspid Valve Tricuspid Regurgitant Peak Pressure Gradient (TR Vmax) Tricuspid Valve Peak Pressure Gradient (TV Vmax) Tricuspid Regurgitant Peak Velocity (TR Vmax) Tricuspid Valve Peak Velocity (TV Vmax) Tricuspid Valve Velocity Peak A (TV A Velocity) Tricuspid Valve Velocity Peak E (TV E Velocity) Tricuspid Regurgitant Mean Pressure Gradient (TR Trace) Tricuspid Valve Mean Pressure Gradient (TV Trace) Tricuspid Regurgitant Mean Velocity (TR Trace) Tricuspid Regurgitant Mean Square Root Velocity (TR Trace) Tricuspid Regurgitant Velocity Time Integral (TR Trace) Tricuspid Valve Mean Velocity (TV Trace) Tricuspid Valve Mean Square Root Velocity (TV Trace) Tricuspid Valve Velocity Time Integral (TV Trace) Tricuspid Valve Time to Peak (TV Acc/Dec Time) Tricuspid Valve Ejection Time (TV Acc/Dec Time) Tricuspid Valve A-Wave Duration (TV A Dur) QRS complex to end of envelope (Q-to-TV close) Tricuspid Valve Pressure Half Time (TV PHT) Stroke Volume by Tricuspid Flow (TV Planimetry, TV Trace) Tricuspid Valve E-Peak to A-Peak Ratio

    (TV E/A Velocity) Color Flow Mode Measurements • Aortic Valve

    - Proximal Isovelocity Surface Area: Regurgitant Orifice Area (AR Radius)

    - Proximal Isovelocity Surface Area: Radius of Aliased Point (AR Radius)

    - Proximal Isovelocity Surface Area: Regurgitant Flow (AR Trace)

    - Proximal Isovelocity Surface Area: Regurgitant Volume Flow (AR Trace)

    Proximal Isovelocity Surface Area: Aliased Velocity (AR Vmax)

    • Mitral Valve Proximal Isovelocity Surface Area: Regurgitant Orifice Area (MR Radius) Proximal Isovelocity Surface Area: Radius of Aliased Point (MR Radius) Proximal Isovelocity Surface Area: Regurgitant Flow (MR Trace) Proximal Isovelocity Surface Area: Regurgitant Volume Flow (MR Trace) Proximal Isovelocity Surface Area: Aliased Velocity (MR Vmax)

    Combination Mode Measurements • Aortic Valve

    - Aortic Valve Area (Ao Root Diam, LVOT Vmax, AV Vmax)

    - Aortic Valve Area by Continuity Equation by Peak Velocity (Ao Root Diam, LVOT Vmax, AV Vmax)

    Stroke Volume by Aortic Flow (AVA Pl ani met ry, AV Trace) Cardiac Output by Aortic Flow (AVA Planimetry, AV Trace, HR) Aortic Valve Area by Continuity Equation VTI (Ao Root Diam, LVOT Vmax, AV Trace)

    • Left Ventricle Cardiac Output, Teichholz/Cubic (LVIDd, LVI Ds, HR) Cardiac Output Two Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs, HR) Cardiac Output Four Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs, HR) Ejection Fraction Two Chamber, Single Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs) Ejection Fraction Four Chamber, Single

    Plane, Area-Length/ Method of Disk(Simpson) (LVAd, LVAs) Left Ventricle Stroke Volume, Single Plane, Two Chamber/Four Chamber, Area-Length (LVAd, LVAs) Left Ventricle Stroke Volume, Single Plane, Two Chamber/Four Chamber, Method of Disk(Simpson) (LVIDd, LVIDs, LVAd, LVAs) Left Ventricle Volume, Two Chamber/Four Chamber, Area-Length (LVAd, LVAs) Ejection Fraction, Bi-Plane, Method of Disk (LVAd, LVAs, 2CH, 4CH) Left Ventricle Stroke Volume, Bi-Plane, Method of Disk (LVAd, LVAs, 2CH, 4CH) Left Ventricle Volume, Bi-Plane, Method of Disk (LVAd, LVAs, 2CH, 4CH) Left Ventricle Stroke Index, Single Plane, Two Chamber/Four Chamber, Area-Length (LVSd, LVSs, and BSA) Left Ventricle Volume, Single Plane, Two Chamber/Four Chamber, Method of Disk (LVAd, LVAs) Left Ventricle Volume, Apical View, Long Axis, Method of Disk (LVAd, LVAs) Stroke Volume by Aortic Flow (AVA Planimetry, AV Trace)

    • Mitral Valve Stroke Volume by Mitral Flow (MVA Planimetry, MV Trace) Cardiac Output by Mitral Flow (MVA Planimetry, MV Trace, HR)

    • Pulmonic Valve Stroke Volume by Pulmonic Flow (PV Planimetry, PV Trace) Cardiac Output by Pulmonic Flow (PV Planimetry, PV Trace, HR)

    • Tricuspid Valve Cardiac Output by Tricuspid Flow (TV

    Planimetry, TV Trace, HR) Report Designer • On-board reporting package

    automates report writing • Formats various exam results into a

    report suitable for printing to a windows printer or reviewing on a standard PC

    • Exam results include patient info, exam info, measurements, calculations, images, comments and diagnosis

    • Standard templates provided Customizable templates

    LOGIQ P5 Page 10 of 13

  • Probes - Number of Element : 192 - FOV (Max) : 58° - FOV (Max) : 38.4 mm - Physical Foot Print : 60 x 18 mm

    - Physical Foot Print : 38 x 10mm - B-mode Imaging Frequency : 2.0, • 3.5CS Thin Wide Band Convex Probe - B-mode Imaging Frequency : 7.0, - 3.0, 4.0, 5.0 MHz

    - Applications: Abdomen, OB Gyn, Urology, Vascular

    - 10.0, 12.0 MHz - Harmonic Frequency : 4.0, 5.0, 5.5, - Harmonic Frequency : 10.0, 12.0, - 6.0 MHz - 13.0 MHz - Doppler Frequency : 2.5, 3.3 MHz - Probe Band Width: 2.0 – 5.0 MHz - Doppler Frequency : 5.0, 6.7 MHz - Biopsy Guide Available : Multi

    Angle, Reusable - Number of Element: 128

    - Steered Angle : 0, +/- 5, 10, 15, 20° - Convex Radius: 38 mmR - Biopsy Guide Available : Multi - FOV (Max): 68° - Angle, Reusable • 8L Wide Band Linear Probe - Physical Foot Print: 48 x 11 mm - Applications : Vascular, Small - B-mode Imaging Frequency: 2.0, 3.0,

    4.0, 5.0 MHz Parts • 12L Wide Band Linear Probe - Probe Band Width : 4 – 11MHz - Applications: Small Parts, Vascular,

    Neonatal, Pediatrics - Number of Element : 128 - Harmonic Frequency: 4, 5, 5.2, 5.5 MHz - FOV (Max) : 38 mm - Probe Band Width: 5 – 13.0 MHz - Physical Foot Print : 38 x 10 mm - Doppler Frequency: 2.5, 3.3 MHz - Number of Element: 192 - B-mode Imaging Frequency : 6.0,

    - Biopsy Guide Available: Single Angle, Reusable

    - FOV (Max): 39 mm - 8.0, 10.0 MHz - Physical Foot Print: 45 x 10 mm - Harmonic Frequency : 8.0, 10.0,

    - 12.0 MHz - B-mode Imaging Frequency: 7.0, 10.0, 12.0 MHz - Doppler Frequency : 5.0, 6.7 MHz • 3.5C Wide Band Convex Probe

    - Steered Angle : 0, +/- 5, 10, 15, 20° - Applications: Abdomen, OB Gyn, Urology, Vascular

    - Harmonic Frequency: 10.0, 12.0 MHz - Biopsy Guide Available : Multi - Doppler Frequency: 5.0, 6.7 MHz Angle, Reusable - Probe Band Width: 2.0 – 5.0 MHz - Steered Angle: +/- 10°, 20°

    - Number of Element: 128 Biopsy Guide Available: Single Angle, Reusable

    • 9L Wide Band Linear Probe - Convex Radius: 38 mmR - Applications: Vascular, Small Parts

    - FOV (Max): 68° - Probe Band Width : 3 – 10 MHz • 3CRF Wide Band Microconvex Probe - Number of Element : 192 - Physical Foot Print: 48 x 11 mm

    - Applications: Abdomen, Urology, Vascular

    - FOV (Max) : 44 mm - B-mode Imaging Frequency: 2.0, 3.0, 4.0, 5.0 MHz - Physical Foot Print : 44 x 6 mm - Probe Band Width : 2.0 – 4 MHz - B-mode Imaging Frequency : 5.0, - Harmonic Frequency: 4, 5, 5.2, 5.5 MHz

    - Number of Element: 128 - 7.0, 9.0 MHz - Convex Radius : 20 mmR - Harmonic Frequency : 8.0, 10.0 - FOV (Max) : 80° - Doppler Frequency: 2.5, 3.3 MHz - MHz - Physical Foot Print : 28 x 16 mm - Biopsy Guide Available: Single Angle,

    Reusable - Doppler Frequency : 4.0, 5.0 MHz

    - B-mode Imaging Frequency : 2.0, - Steered Angle : 0, +/- 5, 10, 15, 20° - 3.0, 3.5, 4.0 MHz - Biopsy Guide Available : Multi - Harmonic Frequency : 3.6, 3.8, 4.0, - Angle, Reusable • 4C Wide Band Convex Probe - 4.2MHz - Applications: Abdomen, OB Gyn,

    Urology, Vascular - Doppler Frequency : 2.5, 3.3 MHz • 10L Wide Band Linear Probe - Biopsy Guide Available : Single

    Angle, Reusable - Probe Band Width : 1.4 – 5 MHz - Applications: Vascular, Small Parts,

    Neonatal, Pediatrics - Number of Element: 128 - Convex Radius : 60 mmR - Probe Band Width: 4 – 11 MHz • E8C Wide Band Microconvex Probe - FOV (Max) : 58° - Applications : OB Gyn, Urology,

    Endocavity - Number of Element: 192 - Physical Foot Print : 60 x 18 mm - FOV (Max): 39 mm - B-mode Imaging Frequency : 2.0,

    - Probe Band Width : 4 – 11 MHz - Physical Foot Print: 39 x 10 mm - 3.0, 4.0, 5.0 MHz - Number of Element : 128 - Harmonic Frequency : 4.0, 5.0, 5.2, - B-mode Imaging Frequency: 6.0, 8.0,

    10.0 MHz - Convex Radius : 11 mmR - 5.5 MHz - FOV (Max) : 133° - Doppler Frequency : 2.5, 3.3 MHz - Doppler Frequency: 5.0, 6.7 MHz - Physical Foot Print : 26 x 10 mm - Biopsy Guide Available : Multi

    Angle, Reusable - Steered Angle: +/- 10°, 20° - B-mode Imaging Frequency : 6.0, - Biopsy Guide Available: Multi Angle,

    Reusable - 8.0, 10.0 MHz - Harmonic Frequency : 8.0, 10.0 • 5CS Convex Probe - MHz - Applications: Abdomen, OB Gyn,

    Urology - Doppler Frequency : 4.0, 5.0 MHz • 11L Wide Band Linear Probe - Biopsy Guide Available : Single

    Angle, Disposable and Reusable - Applications : Small Parts, - Probe Band Width : 2 – 6 MHz Vascular, Neonatal, Pediatrics - Number of Element : 128

    - Probe Band Width : 5 – 13 MHz

    LOGIQ P5 Page 11 of 13

    - Convex Radius : 60 mmR

  • • E8CS Wide Band Microconvex Probe - FOV (Max) : 25 mm Biopsy Guide Available: Multi Angle, Reusable - Applications : OB Gyn, Urology, - Physical Foot Print : 29 x 10 mm

    - B-mode Imaging Frequency : 6.0, Endocavity - Probe Band Width : 4 – 11 MHz • 7S Wide Band Phased Array Sector

    Probe - 8.0, 10.0 MHz

    - Number of Element : 128 - Harmonic Frequency : 10.0, - Convex Radius : 9 mmR - Applications: Cardiac, Pediatrics,

    Abdomen - 12.0MHz

    - FOV (Max) : 168° - Doppler Frequency : 5.0, 6.7 MHz - Physical Foot Print : 15 x 10 mm - Probe Band Width : 3 – 8 MHz - Steered Angle : +/- 5, 10, 15, 20° - B-mode Imaging Frequency : 6.0, - Number of Element : 96 - Biopsy Guide Available : None

    - 8.0, 10.0 MHz - FOV (Max) : 90° - Harmonic Frequency : 8.0, 10.0

    MHz - Physical Foot Print : 14 x 7 mm • t739 Intraoperative Wide Band - B-mode Imaging Frequency : 5.0, Linear Probe

    - Doppler Frequency : 4.0, 5.0 MHz - 6.0, 7.0 MHz - Applications : Intraoperative, Small - Biopsy Guide Available : Single - Harmonic Frequency : 6.0, 7.0, 8.0, Parts, Vascular, Pediatrics - Angle, Disposable and Reusable - 8.5 MHz - Probe Band Width : 4 -10 MHz - Doppler Frequency : 4.0, 5.0 MHz - Number of Element: 192

    • 8C Wide Band Microconvex Probe - CW Doppler Frequency : 4.0 MHz - FOV (Max) : 39 mm - Applications : Neonatal, Pediatrics - Biopsy Guide Available : None - Physical Foot Print : 39 x 10 mm - Probe Band Width : 4 - 11 MHz - B-mode Imaging Frequency : 6.0, - Number of Element : 128 • 3Sp Wide Band Phased Array Sector - 8.0, 10.0 MHz - Convex Radius : 11 mmR - Harmonic Frequency : 8.0, 10.0, Probe - FOV (Max) : 133° - Applications : Cardiac, Trancranial, - 12.0 MHz - Physical Foot Print : 26 x 10 mm - Doppler Frequency : 5.0, 6.7 MHz Abdomen - B-mode Imaging Frequency : 6.0, - Probe Band Width : 1.5 – 5MHz - Steered Angle : +/- 5, 10, 15, 20° - 8.0, 10.0 MHz - Number of Element : 64 - Biopsy Guide Available : Multi - Harmonic Frequency : 8.0, 10.0 - FOV (Max) : 90° - Angle, Reusable

    - Physical Foot Print : 16 x 14 mm MHz - Doppler Frequency : 4.0, 5.0 MHz - B-mode Imaging Frequency : 2.0, • i739 Intraoperative Wide Band

    Linear Probe - Biopsy Guide Available : None - 3.0, 4.0, 5.0 MHz - Harmonic Frequency : 3.0, 3.5, 4.0, - Applications : Intraoperative, Small

    • 3S Wide Band Phased Array Sector - 5.0, 5.5 MHz - Parts, Vascular, Pediatrics - Doppler Frequency : 1.8, 2.0, 2.5, - Probe Band Width : 4 – 10 MHz Probe

    - Applications : Cardiac, Trancranial, - 3.3, 4.0 MHz - Number of Element: 192 - CW Doppler Frequency : 2.0 MHz - FOV (Max) : 39 mm Abdomen

    - Probe Band Width : 1.5 – 3.5 MHz - Physical Foot Print : 39 x 10 mm Biopsy Guide Available: Multi Angle, Reusable - Number of Element : 64 - B-mode Imaging Frequency : 6.0,

    - FOV (Max) : 90° - 8.0, 10.0 MHz - Physical Foot Print : 19 x 16 mm • 5Sp Wide Band Phased Array Sector - Harmonic Frequency : 8.0, 10.0, - B-mode Imaging Frequency : 2.0, - 12.0 MHz Probe - 2.5, 3.0 MHz - Applications : Cardiac, Trancranial, - Doppler Frequency : 5.0, 6.7 MHz - Harmonic Frequency : 2.8, 3.0, 3.2, - Steered Angle : +/- 5, 10, 15, 20° Abdomen - 3.6 MHz Probe Band Width : 2.0 – 8.0 MHz Biopsy Guide Available : None - Doppler Frequency : 1.7, 2.0, 2.2 - Number of Element : 64 - MHz - FOV (Max) : 90° • BE9CS Wide Band Biplane

    Microconvex Probe - CW Doppler Frequency : 2.0 MHz - Physical Foot Print : 10 x 10 mm - Biopsy Guide Available : Multi - B-mode Imaging Frequency : 4.0, - Applications: Urology, Endocavity - Angle, Reusable - 5.0, 6.7, 8.0 MHz - Probe Band Width : 4 – 11MHz

    - Harmonic Frequency : 5.0, 6.0, 8.0, - Number of Element : 96 x 2 • 5S Wide Band Phased Array Sector - 10.0 MHz - Convex Radius : 9 mmR

    - Doppler Frequency : 2.7, 3.0, 3.3 - FOV (Max) : 127° Probe - Applications : Cardiac, Trancranial, - 4.0, 5.0 MHz - Physical Foot Print : 20 x 10 mm

    - CW Doppler Frequency : 2.5 MHz - B-mode Imaging Frequency : 6.0, Abdomen - Probe Band Width : 2 – 5 MHz - 8.0, 10.0 MHz Biopsy Guide Available: Multi Angle,

    Reusable - Number of Element : 96 - Harmonic Frequency : 8.0, 10.0 - FOV (Max) : 90° - MHz - Physical Foot Print : 19 x 11 mm • i12L Intraoperative Wide Band

    Linear Probe - Doppler Frequency : 4.0, 5.0 MHz

    - B-mode Imaging Frequency : 3.0 , - Biopsy Guide Available : Single Angle Reusable (stainless steel), Disposable

    - 4.0, 5.0 MHz - Applications : Intraoperative, Small - Harmonic Frequency : 4.0, 5.0 MHz - Parts, Vascular, Pediatrics - Doppler Frequency : 2.5, 3.3 MHz - Probe Band Width : 4 – 12 MHz - CW Doppler Frequency : 2.5 MHz - Number of Element : 96 • ERB Wide Band Biplane Probe

    LOGIQ P5 Page 12 of 13

  • - Applications: Urology - Probe Band Width: ERB Linear5 – 10 MHz ERB Convex: 5 - 9MHz

    - Number of Element: 128 Safety Conformance - FOV (Max) : The LOGIQ P5 is: GE Healthcare. - America: 888-202-5528 P.O

    Box 414, Milwaukee , Wisconsin 53201 USA ERB Linear: 51mm

    ERB Convex: 123° - Physical Foot Print: • Listed to UL 60601-1 by a Nationally

    Recognized Test Lab GE Healthcare , GE Asia Technology Park ERB Linear: 51 x 5 mm 3rd Floor, Building 1,1 Hua Tuo Road , • Certified to CAN/CSA C22.2 601.1 by an

    SCC accredited Test Lab ERB Convex: 17 x 5.5 mm

    Shanghai – 201203 - B-mode Imaging Frequency: 6.0, 8.0, 10.0 MHz

    Tel : +86 21 38774000 • CE Marked to Council Directive 93/42/EEC on Medical Devices - Harmonic Frequency: 8.0, 10.0 MHz GE Ultraschall Deutschland GmbH & Co KG

    • Conforms to the following standards for safety:

    - Doppler Frequency: 5.0, 6.7 MHz Beethovenstr. 239, D-42655 Solingen - Biopsy Guide Available: Multi

    Angle, Grid Reusable Tel.: (+49) 212-2802-0

    -IEC/EN 60601-1 Medical Electrical equipment, General requirement for Safety

    Fax: (+49) 212-2802-28 Internet : www.gehealthcare.com • P2D Non-imaging Single CW Doppler Pencil Probe -IEC/EN 60601-1-1 Safety

    requirements for Medical Electrical Systems

    - Applications : Cardiac GE, GE Monogram and LOGIQ®, CrossXBeam , TruScan , LOGIQView and Inte are trademarks of General Electric Company.

    - Frequency: 2.0 MHz

    -IEC/EN 60601-1-2 Electromagnetic compatibility

    • P6D Non-imaging Single CW Doppler Pencil Probe GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC a General Electric company doing business as GE Healthcare

    -IEC/EN 60601-1-4 Programmable medical systems

    - Applications : Cardiac, Vascular, Pediatric

    - Frequency: 5.0 MHz -IEC/EN 60601-2-37 Particular

    requirements for the safety of ultrasonic medical diagnostic and monitoring equipment

    © 2008 General Electric Company - All rights reserved. • UG7C Gastro Probe – Europe Only

    - Applications Transesophageal, Stomach - Probe Band Width : 4.5 – 10MHz -IEC 61157 Declaration of acoustic

    output - Number of Element: 128 - Convex Radius: 8.5 mmR -ISO 10993 Biological evaluation of

    medical devices - FOV (Max) : 125° - Physical Foot Print : 19 x 5 mm

    ·NEMA UD3 Acoustic output display (MI, TIS, TIB, TIC)

    - B-mode Imaging Frequency : 6.0, - 8.0, 10.0 MHz - Harmonic Frequency : 8.0, 10.0 - MHz - Doppler Frequency : 5.0, 6.7 MHz - Biopsy Guide Available: None

    * available only in Europe

    Inputs and Outputs Not all features or specifications described in this document may be available in all probes and/or modes.

    • Video In - S-Video

    • Video Out - S-Video General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product at any time without notice or obligation. Contact GE Representative for the most current information.

    • Audio Stereo In • VGA out • Connectors

    - Footswitch - USB (6) - Ethernet - Power for Peripherals

    LOGIQ P5 Page 13 of 13

    Product DescriptionDimensions and WeightElectrical PowerConsole DesignOperator KeyboardMonitor

    System OverviewApplicationsScanning MethodsTransducer TypesOperating ModesSystem Standard FeaturesSystem OptionsMedia & Peripheral OptionsDisplay ModesDisplay Annotation

    General System ParametersSystem Setup

    CINE Memory/Image MemoryImage StorageConnectivity

    Scanning ParametersB-ModeColor Flow ModePower Doppler ImagingM-ModeAnatomical M-Mode

    PW/CW-ModeCoded Harmonic ImagingCrossBeam (option)SRI-HD

    Coded ExcitationVirtual ConvexAutomatic Optimization

    B-Flow (option)LOGIQView (option)Advanced 3D (option)

    Pre-ProcessingPost-Processing w/ TruAccess (Raw Data)Physiological Input Panel (Option)General B-ModeGeneral M-ModeGeneral DopplerMeasurements/CalculationsReal-time Doppler AutoMeasurements / CalculationsOB Measurements/CalculationsGYN Measurements/CalculationsVascular Measurements/Calculations

    CardiacMeasurements/Calculations

    Inputs and OutputsSafety ConformanceTel.: (+49) 212-2802-0