measuring coriolis mass flow meter monitoring for … · tmu 1/03-2016 1 s5 malaysia, mexico,...
TRANSCRIPT
TMU
103
- 201
6
1
S5
measuring
Coriolis Mass Flow Meter bull
monitoring
for liquids and gas bull
analysing
OO Measuring range 0 - 60 kgh hellip 0 - 2200 th water
OO Accuracy plusmn 01 of reading plusmnzero point stability (from liquids up to TMU-x040)
OO p PN40 t -40 +260degC max max
OO Connection flange DN10hellip DN400 frac12hellip 14 class 150300
OO Material 14404 (316 L) 14571 (316 Ti) Hastelloy C-22
OO Options contacts analogue output with HARTreg Profibus-PAreg Fieldbusreg Foundationtrade or Modbus RTU
KOBOLD companies worldwide
ARGENTINA AUSTRALIA AUSTRIA BELGIUM BULGARIA CANADA CHILE CHINA COLOMBIA CZECHIA EGYPT FRANCE GERMANY GREAT BRITAIN HUNGARY INDIA INDONESIA ITALY MALAYSIA MEXICO NETHERLANDS PERU POLAND REPUBLIC OF KOREA ROMANIA SINGAPORE SPAIN SWITZERLAND TAIWAN THAILAND TUNISIA TURKEY USA VIETNAM
KOBOLD Messring GmbH Nordring 22-24 D-65719 HofheimTs
Head Office +49(0)6192 299-0 +49(0)6192 23398 infodekoboldcom wwwkoboldcom
2 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Description The Kobold Mass Flow Meter model TMU utilizes the Coriolis principle of operation to measure mass flow Density and temperature are simultaneously monitored and volumetric flow is additionally calculated with these parameters The TMU Series is available with a direct mounted transmitter or in a remote mounted configuration The unit can be used to meter nearly all liquid or gaseous media and was especially designed to operate in many standard applications It is applied in many different industrial branches The TMU Series is also used for precise dosing as well as in loading and unloading applications Approvals for service in custody transfer (fiscal metering) applications are also available
Application Areas OO chemical industry OO petrochemical industry OO food industry OO oil industry OO gas industry
Technical Details Sensor Measuring principle Coriolis Measurable media liquids and gases
Materials - flow tubes splitter
flanges stainless steel 14404 (316 L) 14571 (316 Ti) Hastelloy C-22
- housing stainless steel 14301 (304) up to TMU-x040 St 372 14301 (304) from TMU-x050
Process connections acc to EN 1092 ASME B165 DIN 2512 special connections on request
Nominal pressure PN 40 ASME Cl 150 300 higher pressures on request
Process temperature -40 hellip +260degC Ambient temperature -20 hellip +60degC (standard)
-20hellip+80 degC (as special version) Protection IP 6668 (EN60529) Certificates and approvals - explosion protection sensor circuits intrinsically safe
DMT 01 ATEX E 149 X II frac12 G EEx ia IIC T6ndashT2
(approval for zone 0 inside flow tubes available)
- CE-marking pressure equipment directive 97 23 EG
Transmitter UMC3
Material
- housing
- display cover
Mounting
Power supply
Outputs
Current
Binary 1
Frequency
Binary 2
Status
Input binary
Ambient temperature
Protection
Communication
Transmitter UMC4
Material
- housing
Mounting
Power supply
Output signals
All signal outputs
Analogue output
aluminium (painted)
safety glass
integrated or remote mounted (junction box or plug in connector)
19 - 36 VDC 24 VAC plusmn20 90 - 265 VAC
galvanically isolated
2 x 0 (4) - 20 mA
active potential free 24 V = max 200 mA passive optocoupler Ui = 30 V Ii = 200 mA Pi = 3 W
1 KHz
passive optocoupler
Ui = 30 V Ii = 200 mA Pi = 3 W
passive optocoupler Ui = 30 V Ii = 200 mA Pi = 3 W
counter reset
-20 hellip +60degC (standard) -20 hellip +80degC (remote mounted transmitter with appovals 5 and 6)
IP 68 (EN60529)
HARTreg
Profibusreg PA Fieldbusreg Foundationtrade Modbus RTU (RS 485)
aluminium (painted)
integrated or remote mounted (junction box or cable)
19 - 36 VDC
24 VAC + 5-20 5060 Hz 90 - 265 VAC 50 60 Hz
galvanically isolated
2 x 4 -20 mA passive
Current output 1 mass flow volume flow density temperature (when using HARTreg protocol output 1 is assigned to mass flow)
Current output 2 mass flow volume flow density temperature
No responsibility taken for errors subject to change without prior notice
3 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Technical Details (continued) Pulse output Binary 1 pulse duration standard 50 ms
adjustable from 01 2000 ms pulsepause ratio 11 (if the set pulse duration is below limit) as frequency output max 1 KHz passive optocoupler Unom 24 V Ui = 30 V Ii = 60 mA Pi = 18 W
Pulse value settable in decade increments of selected pulse unit (e g kg or msup3)
Status forward flow reverse flow Min Max flow rate Min Max density Min Max temperature alarm second pulse output (out of phase by 90deg) passive optocoupler Unom 24 V Ui = 30 V Ii = 60 mA Pi = 18 W
Ambient temperature -20 hellip +60 degC (standard)
below 0degC the readability of the LC-display is limited
Protection IP 68 (EN60529) Communication HARTreg
Accuracy Liquid plusmn01 of reading
plusmnzero point stability up to TMU-x040 plusmn015 of reading plusmnzero point stability from TMU-x050
Measuring Ranges
Gas plusmn05 of reading plusmnzero point stability
Density (liquid) plusmn0005 gcmsup3 with density calibration plusmn0001 gcmsup3 with special density calibration for TMU-x015 TMU-x025 and TMU-x040 plusmn0002 gcmsup3 with special density calibration for other models
Volume plusmn02 of reading plusmnzero point stability
Certifications and Approvals Sensor Instrisically safe sensor circuits DMT 01 ATEX E 149 X
II 12G Ex ia IIC T6 -T2 (Zone 0 is permitted in the measuring tube) NEPSI Approval Cert No GYJ06476X GOST-R
UMC-3 BVS 05 ATEX E 021 X
II (1)2G EEx de [ia] IICIIB T6ndashT3
II (1)2G EEx d [ia] IICIIB T6ndashT3 NEPSI Approval Cert No GYJ06477 GOST-R
UMC-4 BVS 10 ATEX E 110 X
II (1)2G Ex d [ia Ga] IIC T4ndashT3 Gb IEC-Ex
Model Min measuring range
kgh [lbsmin] Max measuring range
kgh [lbsmin] Nominal (Δp = 1bar)
kgh [lbsmin] Zero point stability (of range)
kgh [lbsmin]
TMU-x008 60 [22] 600 [220] 330 [121] 006 [0002]
TMU-x010 250 [92] 2500 [919] 1150 [423] 025 [001]
TMU-x015 1200 [441] 12000 [4409] 5250 [1929] 12 [004]
TMU-x025 3000 [1102] 30 000 [11023] 20 000 [7349] 3 [01]
TMU-x040 6000 [2205] 60 000 [22046] 55 000 [20209]1) 6 [02]
TMU-x050 20 000 [7349] 80 000 [29394] 74 000 [27190] 8 [03]
TMU-x080 25 000 [9186] 120 000 [44092] 118 000 [43357]2) 12 [04]
TMU-x100 30 000 [11023] 200 000 [73486] 200 000 [73486]3) 20 [07]
TMU-x150 60 000 [22046] 460 000 [16 9018] 460 000 [16 9018]3) 46 [17]
TMU-x200 150 000 [55115] 700 000 [25 7202] 700 000 [25 7202]4) 70 [26]
TMU-x250 300 000 [11 0229] 1 500 000 [55 1146] 1 350 000 [49 6032] 150 [55]
TMU-x300 400 000 [14 6972] 2 200 000 [80 8348] 1 900 000 [69 8119] 220 [81]
Reference condition according to IEC 770 Water at 20 degC 1) (Δp = 087bar) 2) (Δp = 095bar) 3) (Δp = 093bar) 4) (Δp = 066bar)
No responsibility taken for errors subject to change without prior notice
4 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
008 = DN8 0 - 600 kgh (min 0 - 60 kgh)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
201R = flange frac12 Class 150 RF 3)
241R = flange frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
010 = DN10 201R = flange frac12 Class 150 RF 3)
0 - 2500 kgh 241R = flange frac12 Class 600 RF 3)
(min 0 - 250 kgh) 202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
223R = flange 1 Class 300 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special
TMU-S stainless steel
TMU-H Hastelloy C-22
(please specify in clear text)
A = standard containment 143011)
B = standard containment with drain connection frac12 NPT female
0 = without
A = heating plate (up to TMU040)
B = heating loop (from
6030 = frac12 NPT IG 305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
321B = flange DN50 PN40 2)
201R = flange frac12 Class 150 RF 3)
015 = DN15 241R = flange frac12 Class 600 RF 3) TMU040) 0 - 12 000 kgh 202R = flange frac34 Class 150 RF 3)
(min 0 - 1200 kgh) 242R = flange frac34 Class 600 RF 3)
203R = flange 1 Class 150 RF 3)
243R = flange 1 Class 600 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special (please specify in clear text)
309B = flange DN 25 PN 40 2)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
025 = DN25 203R = flange 1 Class 150 RF 3)
0 - 30 000 kgh 243R = flange 1 Class 600 RF 3)
(min 0 - 3000 kgh) 205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
5 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
040 = DN40 0 - 60 000 kgh (min 0 - 6000 kgh)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF 206R = flange 2 Class 150 RF 3)
228R = flange 3 Class 300 RF 3)
250R = flange 4 Class 600 RF 3)
XXXX = special (please specify in clear text)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
205R = flange 1frac12 Class 150 RF 3)
050 = DN50 0 - 80 000 kgh (min 0 - 20 000 kgh)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF 4)
208R = flange 3 Class 150 RF 3)
TMU-S stainless steel
248R = flange 3 Class 600 RF 3)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
0 = without
A = heating plate (up to321B = flange DN 50 PN40 2)
TMU-H Hastelloy C-22
331B = flange DN 80 PN40 2)
335B = flange DN100 PN16 2)
340B = flange DN125 PN16 2)
345B = flange DN150 PN16 2)
B = standard containment with drain connection frac12 NPT female
TMU040)
B = heating loop (from
206R = flange 2 Class 150 RF 3)
246R = flange 2 Class 600 RF 3)
TMU040)
080 = DN80 208R = flange 3 Class 150 RF 3)
0 - 120 000 kgh 248R = flange 3 Class 600 RF 3)
(min 0 - 25 000 kgh) 210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
211R = flange 5 Class 150 RF 3)
231R = flange 5 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
208R = flange 3 Class 150 RF 3)
100 = DN100 248R = flange 3 Class 600 RF 3)
0 - 200 000 kgh 210R = flange 4 Class 150 RF 3)
(min 0 - 30 000 kgh) 230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
6 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
150 = DN150 0 - 460 000 kgh (min 0 - 60 000 kgh)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
XXXX = special (please specify in clear text)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
212R = flange 6 Class 150 RF 3)
200 = DN200 0 - 700 000 kgh (min 0 - 150 000 kgh)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
214R = flange 10 Class 150 RF 3)
TMU-S stainless steel
TMU-H Hastelloy
234R = flange 10 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
B = standard containment with drain connection
0 = without
A = heating plate (up to TMU040)
B = heating loop
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
C-22
250 = DN250
363B = flange DN 300 PN16 2)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
frac12 NPT female (from TMU040)
0 - 1 500 000 kgh 214R = flange 10 Class 150 RF 3)
(min 0 - 300 000 kgh) 234R = flange 10 Class 300 RF 3)
215R = flange 12 Class 150 RF 3)
235R = flange 12 Class 300 RF 3)
XXXX = special (please specify in clear text)
355B = flange DN 250 PN10 2)
362B = flange DN 300 PN10 2)
369B = flange DN 350 PN10 2)
375B = flange DN 400 PN10 2)
214R = flange 10 Class 150 RF 3)
300 = DN300 234R = flange 10 Class 300 RF 3)
0 - 2 200 000 kgh 215R = flange 12 Class 150 RF 3)
(min 0 - 400 000 kgh) 235R = flange 12 Class 300 RF 3)
216R = flange 14 Class 150 RF 3)
236R = flange 14 Class 300 RF 3)
217R = flange 16 Class 150 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
2 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Description The Kobold Mass Flow Meter model TMU utilizes the Coriolis principle of operation to measure mass flow Density and temperature are simultaneously monitored and volumetric flow is additionally calculated with these parameters The TMU Series is available with a direct mounted transmitter or in a remote mounted configuration The unit can be used to meter nearly all liquid or gaseous media and was especially designed to operate in many standard applications It is applied in many different industrial branches The TMU Series is also used for precise dosing as well as in loading and unloading applications Approvals for service in custody transfer (fiscal metering) applications are also available
Application Areas OO chemical industry OO petrochemical industry OO food industry OO oil industry OO gas industry
Technical Details Sensor Measuring principle Coriolis Measurable media liquids and gases
Materials - flow tubes splitter
flanges stainless steel 14404 (316 L) 14571 (316 Ti) Hastelloy C-22
- housing stainless steel 14301 (304) up to TMU-x040 St 372 14301 (304) from TMU-x050
Process connections acc to EN 1092 ASME B165 DIN 2512 special connections on request
Nominal pressure PN 40 ASME Cl 150 300 higher pressures on request
Process temperature -40 hellip +260degC Ambient temperature -20 hellip +60degC (standard)
-20hellip+80 degC (as special version) Protection IP 6668 (EN60529) Certificates and approvals - explosion protection sensor circuits intrinsically safe
DMT 01 ATEX E 149 X II frac12 G EEx ia IIC T6ndashT2
(approval for zone 0 inside flow tubes available)
- CE-marking pressure equipment directive 97 23 EG
Transmitter UMC3
Material
- housing
- display cover
Mounting
Power supply
Outputs
Current
Binary 1
Frequency
Binary 2
Status
Input binary
Ambient temperature
Protection
Communication
Transmitter UMC4
Material
- housing
Mounting
Power supply
Output signals
All signal outputs
Analogue output
aluminium (painted)
safety glass
integrated or remote mounted (junction box or plug in connector)
19 - 36 VDC 24 VAC plusmn20 90 - 265 VAC
galvanically isolated
2 x 0 (4) - 20 mA
active potential free 24 V = max 200 mA passive optocoupler Ui = 30 V Ii = 200 mA Pi = 3 W
1 KHz
passive optocoupler
Ui = 30 V Ii = 200 mA Pi = 3 W
passive optocoupler Ui = 30 V Ii = 200 mA Pi = 3 W
counter reset
-20 hellip +60degC (standard) -20 hellip +80degC (remote mounted transmitter with appovals 5 and 6)
IP 68 (EN60529)
HARTreg
Profibusreg PA Fieldbusreg Foundationtrade Modbus RTU (RS 485)
aluminium (painted)
integrated or remote mounted (junction box or cable)
19 - 36 VDC
24 VAC + 5-20 5060 Hz 90 - 265 VAC 50 60 Hz
galvanically isolated
2 x 4 -20 mA passive
Current output 1 mass flow volume flow density temperature (when using HARTreg protocol output 1 is assigned to mass flow)
Current output 2 mass flow volume flow density temperature
No responsibility taken for errors subject to change without prior notice
3 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Technical Details (continued) Pulse output Binary 1 pulse duration standard 50 ms
adjustable from 01 2000 ms pulsepause ratio 11 (if the set pulse duration is below limit) as frequency output max 1 KHz passive optocoupler Unom 24 V Ui = 30 V Ii = 60 mA Pi = 18 W
Pulse value settable in decade increments of selected pulse unit (e g kg or msup3)
Status forward flow reverse flow Min Max flow rate Min Max density Min Max temperature alarm second pulse output (out of phase by 90deg) passive optocoupler Unom 24 V Ui = 30 V Ii = 60 mA Pi = 18 W
Ambient temperature -20 hellip +60 degC (standard)
below 0degC the readability of the LC-display is limited
Protection IP 68 (EN60529) Communication HARTreg
Accuracy Liquid plusmn01 of reading
plusmnzero point stability up to TMU-x040 plusmn015 of reading plusmnzero point stability from TMU-x050
Measuring Ranges
Gas plusmn05 of reading plusmnzero point stability
Density (liquid) plusmn0005 gcmsup3 with density calibration plusmn0001 gcmsup3 with special density calibration for TMU-x015 TMU-x025 and TMU-x040 plusmn0002 gcmsup3 with special density calibration for other models
Volume plusmn02 of reading plusmnzero point stability
Certifications and Approvals Sensor Instrisically safe sensor circuits DMT 01 ATEX E 149 X
II 12G Ex ia IIC T6 -T2 (Zone 0 is permitted in the measuring tube) NEPSI Approval Cert No GYJ06476X GOST-R
UMC-3 BVS 05 ATEX E 021 X
II (1)2G EEx de [ia] IICIIB T6ndashT3
II (1)2G EEx d [ia] IICIIB T6ndashT3 NEPSI Approval Cert No GYJ06477 GOST-R
UMC-4 BVS 10 ATEX E 110 X
II (1)2G Ex d [ia Ga] IIC T4ndashT3 Gb IEC-Ex
Model Min measuring range
kgh [lbsmin] Max measuring range
kgh [lbsmin] Nominal (Δp = 1bar)
kgh [lbsmin] Zero point stability (of range)
kgh [lbsmin]
TMU-x008 60 [22] 600 [220] 330 [121] 006 [0002]
TMU-x010 250 [92] 2500 [919] 1150 [423] 025 [001]
TMU-x015 1200 [441] 12000 [4409] 5250 [1929] 12 [004]
TMU-x025 3000 [1102] 30 000 [11023] 20 000 [7349] 3 [01]
TMU-x040 6000 [2205] 60 000 [22046] 55 000 [20209]1) 6 [02]
TMU-x050 20 000 [7349] 80 000 [29394] 74 000 [27190] 8 [03]
TMU-x080 25 000 [9186] 120 000 [44092] 118 000 [43357]2) 12 [04]
TMU-x100 30 000 [11023] 200 000 [73486] 200 000 [73486]3) 20 [07]
TMU-x150 60 000 [22046] 460 000 [16 9018] 460 000 [16 9018]3) 46 [17]
TMU-x200 150 000 [55115] 700 000 [25 7202] 700 000 [25 7202]4) 70 [26]
TMU-x250 300 000 [11 0229] 1 500 000 [55 1146] 1 350 000 [49 6032] 150 [55]
TMU-x300 400 000 [14 6972] 2 200 000 [80 8348] 1 900 000 [69 8119] 220 [81]
Reference condition according to IEC 770 Water at 20 degC 1) (Δp = 087bar) 2) (Δp = 095bar) 3) (Δp = 093bar) 4) (Δp = 066bar)
No responsibility taken for errors subject to change without prior notice
4 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
008 = DN8 0 - 600 kgh (min 0 - 60 kgh)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
201R = flange frac12 Class 150 RF 3)
241R = flange frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
010 = DN10 201R = flange frac12 Class 150 RF 3)
0 - 2500 kgh 241R = flange frac12 Class 600 RF 3)
(min 0 - 250 kgh) 202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
223R = flange 1 Class 300 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special
TMU-S stainless steel
TMU-H Hastelloy C-22
(please specify in clear text)
A = standard containment 143011)
B = standard containment with drain connection frac12 NPT female
0 = without
A = heating plate (up to TMU040)
B = heating loop (from
6030 = frac12 NPT IG 305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
321B = flange DN50 PN40 2)
201R = flange frac12 Class 150 RF 3)
015 = DN15 241R = flange frac12 Class 600 RF 3) TMU040) 0 - 12 000 kgh 202R = flange frac34 Class 150 RF 3)
(min 0 - 1200 kgh) 242R = flange frac34 Class 600 RF 3)
203R = flange 1 Class 150 RF 3)
243R = flange 1 Class 600 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special (please specify in clear text)
309B = flange DN 25 PN 40 2)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
025 = DN25 203R = flange 1 Class 150 RF 3)
0 - 30 000 kgh 243R = flange 1 Class 600 RF 3)
(min 0 - 3000 kgh) 205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
5 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
040 = DN40 0 - 60 000 kgh (min 0 - 6000 kgh)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF 206R = flange 2 Class 150 RF 3)
228R = flange 3 Class 300 RF 3)
250R = flange 4 Class 600 RF 3)
XXXX = special (please specify in clear text)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
205R = flange 1frac12 Class 150 RF 3)
050 = DN50 0 - 80 000 kgh (min 0 - 20 000 kgh)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF 4)
208R = flange 3 Class 150 RF 3)
TMU-S stainless steel
248R = flange 3 Class 600 RF 3)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
0 = without
A = heating plate (up to321B = flange DN 50 PN40 2)
TMU-H Hastelloy C-22
331B = flange DN 80 PN40 2)
335B = flange DN100 PN16 2)
340B = flange DN125 PN16 2)
345B = flange DN150 PN16 2)
B = standard containment with drain connection frac12 NPT female
TMU040)
B = heating loop (from
206R = flange 2 Class 150 RF 3)
246R = flange 2 Class 600 RF 3)
TMU040)
080 = DN80 208R = flange 3 Class 150 RF 3)
0 - 120 000 kgh 248R = flange 3 Class 600 RF 3)
(min 0 - 25 000 kgh) 210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
211R = flange 5 Class 150 RF 3)
231R = flange 5 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
208R = flange 3 Class 150 RF 3)
100 = DN100 248R = flange 3 Class 600 RF 3)
0 - 200 000 kgh 210R = flange 4 Class 150 RF 3)
(min 0 - 30 000 kgh) 230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
6 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
150 = DN150 0 - 460 000 kgh (min 0 - 60 000 kgh)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
XXXX = special (please specify in clear text)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
212R = flange 6 Class 150 RF 3)
200 = DN200 0 - 700 000 kgh (min 0 - 150 000 kgh)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
214R = flange 10 Class 150 RF 3)
TMU-S stainless steel
TMU-H Hastelloy
234R = flange 10 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
B = standard containment with drain connection
0 = without
A = heating plate (up to TMU040)
B = heating loop
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
C-22
250 = DN250
363B = flange DN 300 PN16 2)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
frac12 NPT female (from TMU040)
0 - 1 500 000 kgh 214R = flange 10 Class 150 RF 3)
(min 0 - 300 000 kgh) 234R = flange 10 Class 300 RF 3)
215R = flange 12 Class 150 RF 3)
235R = flange 12 Class 300 RF 3)
XXXX = special (please specify in clear text)
355B = flange DN 250 PN10 2)
362B = flange DN 300 PN10 2)
369B = flange DN 350 PN10 2)
375B = flange DN 400 PN10 2)
214R = flange 10 Class 150 RF 3)
300 = DN300 234R = flange 10 Class 300 RF 3)
0 - 2 200 000 kgh 215R = flange 12 Class 150 RF 3)
(min 0 - 400 000 kgh) 235R = flange 12 Class 300 RF 3)
216R = flange 14 Class 150 RF 3)
236R = flange 14 Class 300 RF 3)
217R = flange 16 Class 150 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
3 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Technical Details (continued) Pulse output Binary 1 pulse duration standard 50 ms
adjustable from 01 2000 ms pulsepause ratio 11 (if the set pulse duration is below limit) as frequency output max 1 KHz passive optocoupler Unom 24 V Ui = 30 V Ii = 60 mA Pi = 18 W
Pulse value settable in decade increments of selected pulse unit (e g kg or msup3)
Status forward flow reverse flow Min Max flow rate Min Max density Min Max temperature alarm second pulse output (out of phase by 90deg) passive optocoupler Unom 24 V Ui = 30 V Ii = 60 mA Pi = 18 W
Ambient temperature -20 hellip +60 degC (standard)
below 0degC the readability of the LC-display is limited
Protection IP 68 (EN60529) Communication HARTreg
Accuracy Liquid plusmn01 of reading
plusmnzero point stability up to TMU-x040 plusmn015 of reading plusmnzero point stability from TMU-x050
Measuring Ranges
Gas plusmn05 of reading plusmnzero point stability
Density (liquid) plusmn0005 gcmsup3 with density calibration plusmn0001 gcmsup3 with special density calibration for TMU-x015 TMU-x025 and TMU-x040 plusmn0002 gcmsup3 with special density calibration for other models
Volume plusmn02 of reading plusmnzero point stability
Certifications and Approvals Sensor Instrisically safe sensor circuits DMT 01 ATEX E 149 X
II 12G Ex ia IIC T6 -T2 (Zone 0 is permitted in the measuring tube) NEPSI Approval Cert No GYJ06476X GOST-R
UMC-3 BVS 05 ATEX E 021 X
II (1)2G EEx de [ia] IICIIB T6ndashT3
II (1)2G EEx d [ia] IICIIB T6ndashT3 NEPSI Approval Cert No GYJ06477 GOST-R
UMC-4 BVS 10 ATEX E 110 X
II (1)2G Ex d [ia Ga] IIC T4ndashT3 Gb IEC-Ex
Model Min measuring range
kgh [lbsmin] Max measuring range
kgh [lbsmin] Nominal (Δp = 1bar)
kgh [lbsmin] Zero point stability (of range)
kgh [lbsmin]
TMU-x008 60 [22] 600 [220] 330 [121] 006 [0002]
TMU-x010 250 [92] 2500 [919] 1150 [423] 025 [001]
TMU-x015 1200 [441] 12000 [4409] 5250 [1929] 12 [004]
TMU-x025 3000 [1102] 30 000 [11023] 20 000 [7349] 3 [01]
TMU-x040 6000 [2205] 60 000 [22046] 55 000 [20209]1) 6 [02]
TMU-x050 20 000 [7349] 80 000 [29394] 74 000 [27190] 8 [03]
TMU-x080 25 000 [9186] 120 000 [44092] 118 000 [43357]2) 12 [04]
TMU-x100 30 000 [11023] 200 000 [73486] 200 000 [73486]3) 20 [07]
TMU-x150 60 000 [22046] 460 000 [16 9018] 460 000 [16 9018]3) 46 [17]
TMU-x200 150 000 [55115] 700 000 [25 7202] 700 000 [25 7202]4) 70 [26]
TMU-x250 300 000 [11 0229] 1 500 000 [55 1146] 1 350 000 [49 6032] 150 [55]
TMU-x300 400 000 [14 6972] 2 200 000 [80 8348] 1 900 000 [69 8119] 220 [81]
Reference condition according to IEC 770 Water at 20 degC 1) (Δp = 087bar) 2) (Δp = 095bar) 3) (Δp = 093bar) 4) (Δp = 066bar)
No responsibility taken for errors subject to change without prior notice
4 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
008 = DN8 0 - 600 kgh (min 0 - 60 kgh)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
201R = flange frac12 Class 150 RF 3)
241R = flange frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
010 = DN10 201R = flange frac12 Class 150 RF 3)
0 - 2500 kgh 241R = flange frac12 Class 600 RF 3)
(min 0 - 250 kgh) 202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
223R = flange 1 Class 300 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special
TMU-S stainless steel
TMU-H Hastelloy C-22
(please specify in clear text)
A = standard containment 143011)
B = standard containment with drain connection frac12 NPT female
0 = without
A = heating plate (up to TMU040)
B = heating loop (from
6030 = frac12 NPT IG 305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
321B = flange DN50 PN40 2)
201R = flange frac12 Class 150 RF 3)
015 = DN15 241R = flange frac12 Class 600 RF 3) TMU040) 0 - 12 000 kgh 202R = flange frac34 Class 150 RF 3)
(min 0 - 1200 kgh) 242R = flange frac34 Class 600 RF 3)
203R = flange 1 Class 150 RF 3)
243R = flange 1 Class 600 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special (please specify in clear text)
309B = flange DN 25 PN 40 2)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
025 = DN25 203R = flange 1 Class 150 RF 3)
0 - 30 000 kgh 243R = flange 1 Class 600 RF 3)
(min 0 - 3000 kgh) 205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
5 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
040 = DN40 0 - 60 000 kgh (min 0 - 6000 kgh)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF 206R = flange 2 Class 150 RF 3)
228R = flange 3 Class 300 RF 3)
250R = flange 4 Class 600 RF 3)
XXXX = special (please specify in clear text)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
205R = flange 1frac12 Class 150 RF 3)
050 = DN50 0 - 80 000 kgh (min 0 - 20 000 kgh)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF 4)
208R = flange 3 Class 150 RF 3)
TMU-S stainless steel
248R = flange 3 Class 600 RF 3)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
0 = without
A = heating plate (up to321B = flange DN 50 PN40 2)
TMU-H Hastelloy C-22
331B = flange DN 80 PN40 2)
335B = flange DN100 PN16 2)
340B = flange DN125 PN16 2)
345B = flange DN150 PN16 2)
B = standard containment with drain connection frac12 NPT female
TMU040)
B = heating loop (from
206R = flange 2 Class 150 RF 3)
246R = flange 2 Class 600 RF 3)
TMU040)
080 = DN80 208R = flange 3 Class 150 RF 3)
0 - 120 000 kgh 248R = flange 3 Class 600 RF 3)
(min 0 - 25 000 kgh) 210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
211R = flange 5 Class 150 RF 3)
231R = flange 5 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
208R = flange 3 Class 150 RF 3)
100 = DN100 248R = flange 3 Class 600 RF 3)
0 - 200 000 kgh 210R = flange 4 Class 150 RF 3)
(min 0 - 30 000 kgh) 230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
6 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
150 = DN150 0 - 460 000 kgh (min 0 - 60 000 kgh)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
XXXX = special (please specify in clear text)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
212R = flange 6 Class 150 RF 3)
200 = DN200 0 - 700 000 kgh (min 0 - 150 000 kgh)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
214R = flange 10 Class 150 RF 3)
TMU-S stainless steel
TMU-H Hastelloy
234R = flange 10 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
B = standard containment with drain connection
0 = without
A = heating plate (up to TMU040)
B = heating loop
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
C-22
250 = DN250
363B = flange DN 300 PN16 2)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
frac12 NPT female (from TMU040)
0 - 1 500 000 kgh 214R = flange 10 Class 150 RF 3)
(min 0 - 300 000 kgh) 234R = flange 10 Class 300 RF 3)
215R = flange 12 Class 150 RF 3)
235R = flange 12 Class 300 RF 3)
XXXX = special (please specify in clear text)
355B = flange DN 250 PN10 2)
362B = flange DN 300 PN10 2)
369B = flange DN 350 PN10 2)
375B = flange DN 400 PN10 2)
214R = flange 10 Class 150 RF 3)
300 = DN300 234R = flange 10 Class 300 RF 3)
0 - 2 200 000 kgh 215R = flange 12 Class 150 RF 3)
(min 0 - 400 000 kgh) 235R = flange 12 Class 300 RF 3)
216R = flange 14 Class 150 RF 3)
236R = flange 14 Class 300 RF 3)
217R = flange 16 Class 150 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
4 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
008 = DN8 0 - 600 kgh (min 0 - 60 kgh)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
201R = flange frac12 Class 150 RF 3)
241R = flange frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
6010 = frac14 NPT IG 6030 = frac12 NPT IG 301B = flange DN10 PN40 2)
305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
010 = DN10 201R = flange frac12 Class 150 RF 3)
0 - 2500 kgh 241R = flange frac12 Class 600 RF 3)
(min 0 - 250 kgh) 202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
223R = flange 1 Class 300 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special
TMU-S stainless steel
TMU-H Hastelloy C-22
(please specify in clear text)
A = standard containment 143011)
B = standard containment with drain connection frac12 NPT female
0 = without
A = heating plate (up to TMU040)
B = heating loop (from
6030 = frac12 NPT IG 305B = flange DN15 PN40 2)
309B = flange DN 25 PN 40 2)
321B = flange DN50 PN40 2)
201R = flange frac12 Class 150 RF 3)
015 = DN15 241R = flange frac12 Class 600 RF 3) TMU040) 0 - 12 000 kgh 202R = flange frac34 Class 150 RF 3)
(min 0 - 1200 kgh) 242R = flange frac34 Class 600 RF 3)
203R = flange 1 Class 150 RF 3)
243R = flange 1 Class 600 RF 3)
225R = flange 1frac12 Class 300 RF 3)
XXXX = special (please specify in clear text)
309B = flange DN 25 PN 40 2)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
202R = flange frac34 Class 150 RF 3)
242R = flange frac34 Class 600 RF 3)
025 = DN25 203R = flange 1 Class 150 RF 3)
0 - 30 000 kgh 243R = flange 1 Class 600 RF 3)
(min 0 - 3000 kgh) 205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF XXXX = special
(please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
5 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
040 = DN40 0 - 60 000 kgh (min 0 - 6000 kgh)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF 206R = flange 2 Class 150 RF 3)
228R = flange 3 Class 300 RF 3)
250R = flange 4 Class 600 RF 3)
XXXX = special (please specify in clear text)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
205R = flange 1frac12 Class 150 RF 3)
050 = DN50 0 - 80 000 kgh (min 0 - 20 000 kgh)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF 4)
208R = flange 3 Class 150 RF 3)
TMU-S stainless steel
248R = flange 3 Class 600 RF 3)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
0 = without
A = heating plate (up to321B = flange DN 50 PN40 2)
TMU-H Hastelloy C-22
331B = flange DN 80 PN40 2)
335B = flange DN100 PN16 2)
340B = flange DN125 PN16 2)
345B = flange DN150 PN16 2)
B = standard containment with drain connection frac12 NPT female
TMU040)
B = heating loop (from
206R = flange 2 Class 150 RF 3)
246R = flange 2 Class 600 RF 3)
TMU040)
080 = DN80 208R = flange 3 Class 150 RF 3)
0 - 120 000 kgh 248R = flange 3 Class 600 RF 3)
(min 0 - 25 000 kgh) 210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
211R = flange 5 Class 150 RF 3)
231R = flange 5 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
208R = flange 3 Class 150 RF 3)
100 = DN100 248R = flange 3 Class 600 RF 3)
0 - 200 000 kgh 210R = flange 4 Class 150 RF 3)
(min 0 - 30 000 kgh) 230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
6 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
150 = DN150 0 - 460 000 kgh (min 0 - 60 000 kgh)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
XXXX = special (please specify in clear text)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
212R = flange 6 Class 150 RF 3)
200 = DN200 0 - 700 000 kgh (min 0 - 150 000 kgh)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
214R = flange 10 Class 150 RF 3)
TMU-S stainless steel
TMU-H Hastelloy
234R = flange 10 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
B = standard containment with drain connection
0 = without
A = heating plate (up to TMU040)
B = heating loop
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
C-22
250 = DN250
363B = flange DN 300 PN16 2)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
frac12 NPT female (from TMU040)
0 - 1 500 000 kgh 214R = flange 10 Class 150 RF 3)
(min 0 - 300 000 kgh) 234R = flange 10 Class 300 RF 3)
215R = flange 12 Class 150 RF 3)
235R = flange 12 Class 300 RF 3)
XXXX = special (please specify in clear text)
355B = flange DN 250 PN10 2)
362B = flange DN 300 PN10 2)
369B = flange DN 350 PN10 2)
375B = flange DN 400 PN10 2)
214R = flange 10 Class 150 RF 3)
300 = DN300 234R = flange 10 Class 300 RF 3)
0 - 2 200 000 kgh 215R = flange 12 Class 150 RF 3)
(min 0 - 400 000 kgh) 235R = flange 12 Class 300 RF 3)
216R = flange 14 Class 150 RF 3)
236R = flange 14 Class 300 RF 3)
217R = flange 16 Class 150 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
5 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
040 = DN40 0 - 60 000 kgh (min 0 - 6000 kgh)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
205R = flange 1frac12 Class 150 RF 3)
245R = flange 1frac12 Class 600 RF 3)
226R = flange 2 Class 300 RF 206R = flange 2 Class 150 RF 3)
228R = flange 3 Class 300 RF 3)
250R = flange 4 Class 600 RF 3)
XXXX = special (please specify in clear text)
317B = flange DN 40 PN 40 2)
321B = flange DN 50 PN 40 2)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
205R = flange 1frac12 Class 150 RF 3)
050 = DN50 0 - 80 000 kgh (min 0 - 20 000 kgh)
245R = flange 1frac12 Class 600 RF 3)
206R = flange 2 Class 150 RF 3)
226R = flange 2 Class 300 RF 4)
208R = flange 3 Class 150 RF 3)
TMU-S stainless steel
248R = flange 3 Class 600 RF 3)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
0 = without
A = heating plate (up to321B = flange DN 50 PN40 2)
TMU-H Hastelloy C-22
331B = flange DN 80 PN40 2)
335B = flange DN100 PN16 2)
340B = flange DN125 PN16 2)
345B = flange DN150 PN16 2)
B = standard containment with drain connection frac12 NPT female
TMU040)
B = heating loop (from
206R = flange 2 Class 150 RF 3)
246R = flange 2 Class 600 RF 3)
TMU040)
080 = DN80 208R = flange 3 Class 150 RF 3)
0 - 120 000 kgh 248R = flange 3 Class 600 RF 3)
(min 0 - 25 000 kgh) 210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
211R = flange 5 Class 150 RF 3)
231R = flange 5 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
331B = flange DN 80 PN 40 2)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
208R = flange 3 Class 150 RF 3)
100 = DN100 248R = flange 3 Class 600 RF 3)
0 - 200 000 kgh 210R = flange 4 Class 150 RF 3)
(min 0 - 30 000 kgh) 230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
6 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
150 = DN150 0 - 460 000 kgh (min 0 - 60 000 kgh)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
XXXX = special (please specify in clear text)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
212R = flange 6 Class 150 RF 3)
200 = DN200 0 - 700 000 kgh (min 0 - 150 000 kgh)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
214R = flange 10 Class 150 RF 3)
TMU-S stainless steel
TMU-H Hastelloy
234R = flange 10 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
B = standard containment with drain connection
0 = without
A = heating plate (up to TMU040)
B = heating loop
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
C-22
250 = DN250
363B = flange DN 300 PN16 2)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
frac12 NPT female (from TMU040)
0 - 1 500 000 kgh 214R = flange 10 Class 150 RF 3)
(min 0 - 300 000 kgh) 234R = flange 10 Class 300 RF 3)
215R = flange 12 Class 150 RF 3)
235R = flange 12 Class 300 RF 3)
XXXX = special (please specify in clear text)
355B = flange DN 250 PN10 2)
362B = flange DN 300 PN10 2)
369B = flange DN 350 PN10 2)
375B = flange DN 400 PN10 2)
214R = flange 10 Class 150 RF 3)
300 = DN300 234R = flange 10 Class 300 RF 3)
0 - 2 200 000 kgh 215R = flange 12 Class 150 RF 3)
(min 0 - 400 000 kgh) 235R = flange 12 Class 300 RF 3)
216R = flange 14 Class 150 RF 3)
236R = flange 14 Class 300 RF 3)
217R = flange 16 Class 150 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
6 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (Example TMU-S008 6010 A00 A 0 1 0) (continued)
Model material
Sensor measuring range Process connection1)
Containment Heating cooling
150 = DN150 0 - 460 000 kgh (min 0 - 60 000 kgh)
335B = flange DN100 PN16 2)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
210R = flange 4 Class 150 RF 3)
230R = flange 4 Class 300 RF 3)
212R = flange 6 Class 150 RF 3)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
XXXX = special (please specify in clear text)
345B = flange DN150 PN16 2)
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
212R = flange 6 Class 150 RF 3)
200 = DN200 0 - 700 000 kgh (min 0 - 150 000 kgh)
232R = flange 6 Class 300 RF 3)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
214R = flange 10 Class 150 RF 3)
TMU-S stainless steel
TMU-H Hastelloy
234R = flange 10 Class 300 RF 3)
XXXX = special (please specify in clear text) A = standard containment
143011)
B = standard containment with drain connection
0 = without
A = heating plate (up to TMU040)
B = heating loop
350B = flange DN200 PN16 2)
356B = flange DN 250 PN16 2)
C-22
250 = DN250
363B = flange DN 300 PN16 2)
213R = flange 8 Class 150 RF 3)
233R = flange 8 Class 300 RF 3)
frac12 NPT female (from TMU040)
0 - 1 500 000 kgh 214R = flange 10 Class 150 RF 3)
(min 0 - 300 000 kgh) 234R = flange 10 Class 300 RF 3)
215R = flange 12 Class 150 RF 3)
235R = flange 12 Class 300 RF 3)
XXXX = special (please specify in clear text)
355B = flange DN 250 PN10 2)
362B = flange DN 300 PN10 2)
369B = flange DN 350 PN10 2)
375B = flange DN 400 PN10 2)
214R = flange 10 Class 150 RF 3)
300 = DN300 234R = flange 10 Class 300 RF 3)
0 - 2 200 000 kgh 215R = flange 12 Class 150 RF 3)
(min 0 - 400 000 kgh) 235R = flange 12 Class 300 RF 3)
216R = flange 14 Class 150 RF 3)
236R = flange 14 Class 300 RF 3)
217R = flange 16 Class 150 RF 3)
XXXX = special (please specify in clear text)
1) Stainless steel 14301 from sensor 040 steel for sensor 050 and bigger 2) Flange according to DIN EN1092-1 form B1 3) ASME B165-2003
No responsibility taken for errors subject to change without prior notice
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
7 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Sensor (continued)
Connection for heating cooling
Sensor process temperature thread junction box
Approvals Calibration massflow
Calibration density
Additional options
0 = without
A = with Ermeto EO12 A = integrated transmitter
B = with Swagelok12 mm -20+100 degC
C = with DN15 B = integrated transmitter PN40 acc to DIN -20+150 degC
EN 1092-1 form B1 C2) = remote mounted transmitter 0 = without
D = with flange frac12 class 150 RF ASME B165
E = with frac12 NPT female
-40+100 degC frac12 NPT
D2) = remote mounted transmitter -40+180 degC frac12 NPT
E2) = remote mounted transmitter -40+260 degC frac12 NPT
0 =
A =
B =
without
II 12 G EEx ia IIC T6 - T2
NEPSI
1 = standard 3-point
2 = 10-point
3 = outside lab
1 = standard 3-point (not for range 008)
0 = without
X = with (separate specification necessary)
F = with DN25 F2) = remote mounted transmitter 2 = special PN40 acc to DIN -40+100 degC M 20x15 5-point
EN 1092-1 form B1 G2) = remote mounted transmitter -40+180 degC M 20x15
G = with flange 1 H2) = remote mounted transmitter class 150 RF -40+260 degC M 20x15 ASME B 165
H = with 1 NPT female
1) Other flange-form or ANSI-flange on request 2) Please order cable glands separately see accessories
Necessary Details for Dimensioning the TMU Instrument OO medium OO operating pressure
OO process temperature minmax OO viscosity
OO ambient temperature minmax OO density
OO measuring range
Order Details Transmitter UMC3 (Example UMC3 - A 0 1 A 0 0K)
Model Mountingconnection to
sensor thread cable connection Display Interface
board Power Power supply Outputs
A = integrated transmitter - frac12 NPT
B = integrated transmitter - M 20x15 0 = without
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
B3) = analogue output 0 (4) - 20 mA
C1) = remote mounted transmitter terminal block with frac12 NPT frac12 NPT
1 = integrated in transmitter housing ambient
1 = 90 - 265 VAC 5060 Hz
withwithout HARTreg pulse output active 24 VDC status output passive Um = 30 VDC
UMC3shy D1) = remote mounted transmitter terminal block with M20x15 M20x15
E1) = remote mounted transmitter plug-in connector Harting Hanreg R23 frac12 NPT
F1) = remote mounted transmitter plug-in connector Harting Hanreg R23 M20x15
temperature up to 60degC
22) = remotable separate board plus panel mounting adapter set
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
D4) = Profibus PA (EEx ia IIC) without analogue and binary outputs
E = Profibusreg DP V1 without analogue and binary outputs
F5) = ModbusRTU (RS485 analogue output 0 (4) - 20 mA
J = Fieldbusreg Foundationtrade
No responsibility taken for errors subject to change without prior notice
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
8 wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Order Details Transmitter UMC3 (continued)
Approvals Protection
signal output
0 = without 0K = without
1 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 60degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60degC
4 = NEPSI ambient temperature up to 60degC
5 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
6 = II(1)2G Eex de [ia] IIBIIC T3-T6 ambient temperature up to 80degC
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
2) Connection cable must be ordered separately 3) Signal output in EEx ia not possible 4) Not available with approval 3 and 4 5) Not available with approval 3 4 5 or 6 not with signal output protection 2
Order Details Transmitter UMC4 (Example UMC4 - B 1 1 A 0 0K)
Model Mounting
thread power supply output Display interface
board Power supply Outputs
UMC4shy
B = integrated transmitter M20x15
D1) = remote mounted transmitter terminal block mountedM20x15 (specify cable length in clear text)
E1) = remote mounted transmitter transmitter with 5 m cable M20x15
1 = integrated in transmitter housing ambient temperature up to 60degC
1 = 90 - 265 VAC 5060 Hz
2 = 19 - 36 VDC 24 VAC (plusmn 20) 5060 Hz
A = analogue output 0 (4) - 20 mA withwithout HARTreg pulse output passive Um = 30 VDC status output passive Um = 30 VDC
Order Details Transmitter UMC4 (continued)
Approvals Protection
signal output
0 = without 0K = without
2 = II(1)2G Eex d [ia] IIBIIC T3-T6 ambient temperature up to 60 degC
1K = EEX ia
2K = EEx e (not intrinsically safe)
1) - Incl wall mounting bracket pipe mounting bracket must be ordered separately (see accessories)
- Connection cable (sensor to transmitter) and cable gland must be ordered separately (see accessories)
Order Details Accessories (Example TMK - BL KK 005)
Order number
Model Version Cable length application area
TMKshy
BL = connection cable
KK = sensor-transmitter with connection cable SK = sensor-transmitter
cable end 1 plug (Harting Hanreg R23) cable end 2 cable connect
SS = plug connection on both sides (Harting Hanreg R23)
UB = transmitter-interface board plug connection
Cable length 005 = 5 metre 010 = 10 metre 015 = 15 metre 030 = 30 metre 075 = 75 metre 150 = 150 metre 300 = 300 metre XXX = special length
V = cable gland set AU = integrated transmitter GU = remote mounted transmitter
Application area NEM 20 = not Ex M 20 x15 NENPT = not Ex frac12 NPT DEIAM 20 = EEx de - EEx ia M 20 x15 DEIANPT = EEx de - EEx ia frac12 NPT DEEM 20 = EEx de - EEx e M 20 x15 DEENPT = EEx de - EEx e frac12 NPT
TMshy ROHRMONT = accessory for 2 pipe mounting
No responsibility taken for errors subject to change without prior notice
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
9 wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Installation Length
Model A (Installation length)
end connection
mm [inch] end
connection mm [inch]
end connection
mm [inch]
TMU-x008 DN10 360 [142] frac14 NPT (f) 300 [118] frac12 Cl150 366 [144]
- - - frac12 NPT (f) 300 [118] frac12 Cl600 375 [148]
TMU-x010
DN10 390 [154] DN15 396 [156] DN25 424 [167]
frac14 NPT (f) 300 [118] frac12 Cl150 416 [164] frac34 Cl150 350 [138]
frac12 NPT (f) 300 [118] frac12 Cl600 425 [167] frac34 Cl600 360 [142]
TMU-x015
DN15 515 [203] DN25 520 [205] DN50 586 [231]
frac12 NPT (f) 450 [177] frac12 Cl150 535 [211] frac34 Cl150 546 [215]
- - - frac12 Cl600 546 [215] frac34 Cl600 556 [219]
TMU-x025
DN25 632 [249] DN40 642 [253] DN50 500 [197]
frac34 Cl150 657 [259] 1 Cl150 664 [261] 1frac12 Cl150 676 [266]
frac34 Cl600 667 [263] 1 Cl600 676 [266] 1frac12 Cl600 692 [272]
TMU-x040
DN40 770 [303] DN50 776 [306] DN80 600 [236]
1frac12 Cl150 804 [317] 2 Cl150 810 [319] - - -
1frac12 Cl600 820 [323] 2 Cl600 828 [326] - -
TMU-x050
DN40 1018 [401] DN50 1024 [403] DN80 1025 [411]
DN100 804 [317] - - - - - -
1frac12 Cl150 1050 [413] 2 Cl150 1053 [415] 3 Cl150 1066 [420]
1frac12 Cl600 1066 [420] 2 Cl600 1072 [422] 3 Cl600 1091 [430]
TMU-x080
DN50 1176 [463] DN80 1196 [471] DN100 1184 [466]
DN125 925 [431] DN150 on request - - -
2 Cl150 1207 [475] 3 Cl150 1218 [480] 4 Cl150 1230 [484]
2 Cl600 1226 [483] 3 Cl600 1243 [489] 4 Cl300 1250 [492]
TMU-x100
DN80 1370 [539] DN100 1358 [535] DN150 1090 [429]
3 Cl150 1388 [546] 4 Cl150 1400 [551] 6 Cl150 1154 [454]
3 Cl300 on request 4 Cl300 1420 [559] 6 Cl300 1173 [462]
3 Cl600 1413 [556] 4 Cl600 on request 6 Cl600 on request
TMU-x150
DN100 1726 [680] DN150 1732 [682] DN200 1448 [570]
4 Cl150 1770 [697] 6 Cl150 1796 [707] 8 Cl150 1525 [600]
4 Cl300 1790 [705] 6 Cl300 1815 [715] 8 Cl300 1545 [608]
4 Cl600 on request 6 Cl600 on request 8 Cl600 on request
TMU-x200
DN150 2184 [860] DN200 2198 [865] DN250 1850 [734]
6 Cl150 2250 [886] 8 Cl150 2270 [894] 10 Cl150 1925 [758]
6 Cl300 2270 [894] 8 Cl300 2287 [900] 10 Cl300 1957 [771]
6 Cl600 on request 8 Cl600 on request 10 Cl600 on request
TMU-x250
DN200 2268 [893] DN250 2284 [899] DN300 1900 [748]
8 Cl150 2348 [924] 10 Cl150 2348 [924] 12 Cl150 1945 [766]
8 Cl300 2363 [930] 10 Cl300 2375 [935] 12 Cl300 1977 [778]
8 Cl600 on request 10 Cl600 on request 12 Cl600 on request
TMU-x300
DN250 2913 [1147] DN300 2925 [1152] DN350 2933 [1155]
DN400 on request - - - - - -
10 Cl150 2976 [1172] 12 Cl150 2995 [1179] 14 Cl150 3020 [1189]
10 Cl300 3008 [1184] 12 Cl300 3030 [1193] 14 Cl300 3050 [1201]
10 Cl600 on request 12 Cl600 on request 14 Cl600 on request
No responsibility taken for errors subject to change without prior notice
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions Dimensions with Transmitter UMC3
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212degF) (-40hellip+302degF) (-40hellip+212 degF) (-40hellip+356degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 328 [129] 430 [169] 228 [90] 330 [13] 430 [169] 85 [33] 40 [16]
TMU-x010 343 [135] 445 [175] 243 [96] 345 [136] 445 [175] 100 [39] 40 [16]
TMU-x015 395 [156] 497 [196] 295 [116] 397 [156] 497 [196] 148 [58] 48 [19]
TMU-x025 460 [181] 562 [221] 360 [142] 462 [182] 562 [221] 200 [79] 74 [29]
TMU-x040 528 [208] 630 [248] 428 [169] 530 [209] 630 [248] 255 [100] 101 [40]
TMU-x050 1010 [398] 1112 [438] 910 [358] 1012 [398] 1112 [438] 615 [242] 230 [91]
TMU-x080 1210 [476] 1312 [517] 1110 [437] 1212 [477] 1312 [517] 800 [315] 250 [98]
TMU-x100 1230 [484] 1332 [524] 1130 [445] 1232 [485] 1332 [524] 815 [321] 270 [106]
TMU-x150 1560 [614] 1662 [654] 1460 [575] 1562 [615] 1662 [654] 1070 [421] 380 [150]
TMU-x200 1720 [677] 1822 [717] 1620 [638] 1722 [678] 1822 [717] 1210 [476] 400 [157]
TMU-x250 1860 [732] 1962 [772] 1760 [693] 1862 [733] 1962 [772] 1300 [512] 550 [217]
TMU-x300 1865 [734] 1967 [774] 1765 [695] 1867 [735] 1967 [774] 1400 [551] 510 [201]
Standardversion Models TMU-008 - TMU-040 206
Integrated transmitter UMC3 89
B
C
A G
Remote mounted transmitter 125 80
B
C
A G
No responsibility taken for errors subject to change without prior notice
10
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 Integrated transmitter UMC3
Remote mounted transmitter
No responsibility taken for errors subject to change without prior notice
11
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
wwwkoboldcom 10
3- 2
016
Coriolis Mass Flow Meter Model TMU
Dimensions with Transmitter UMC4
C
Standardversion Models TMU-008 - TMU-040
Integrated transmitter UMC4
Remote mounted transmitter
B C G
Integrated transmitter Remote mounted transmitter Model -40+100degC -40+150degC -40+100degC -40+180degC -40+260degC
(-40hellip+212 degF) (-40hellip+302 degF) (-40hellip+212 degF) (-40hellip+356 degF) (-40hellip+500degF)
mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch] mm [inch]
TMU-x008 359 [141] 461 [181] 256 [101] 358 [141] 458 [180] 85 [33] 40 [16]
TMU-x010 374 [147] 476 [187] 271 [107] 373 [147] 473 [186] 100 [39] 40 [16]
TMU-x015 426 [168] 528 [208] 323 [127] 425 [167] 525 [207] 148 [58] 48 [19]
TMU-x025 491 [193] 593 [233] 388 [153] 490 [193] 590 [232] 200 [79] 74 [29]
TMU-x040 559 [220] 661 [260] 456 [180] 558 [220] 658 [259] 255 [100] 101 [40]
TMU-x050 1041 [410] 1143 [450] 938 [369] 1040 [409] 1140 [449] 615 [242] 230 [91]
TMU-x080 1241 [489] 1343 [529] 1138 [448] 1240 [488] 1340 [528] 800 [315] 250 [98]
TMU-x100 1261 [496] 1363 [537] 1158 [456] 1260 [496] 1360 [535] 815 [321] 270 [106]
TMU-x150 1591 [626] 1693 [667] 1488 [586] 1590 [626] 1690 [665] 1070 [421] 380 [150]
TMU-x200 1751 [689] 1853 [730] 1648 [649] 1750 [689] 1850 [728] 1210 [476] 400 [157]
TMU-x250 1891 [744] 1993 [785] 1788 [704] 1890 [744] 1990 [783] 1300 [512] 550 [217]
TMU-x300 1896 [746] 1998 [787] 1793 [706] 1895 [746] 1995 [785] 1400 [551] 510 [201]
BB
C
205
108
A
125 80
A G
No responsibility taken for errors subject to change without prior notice
12
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13
wwwkoboldcom 10
3 - 2
016
Coriolis Mass Flow Meter Model TMU
Standard version Models TMU-050 - TMU-300 C
Integrated transmitter UMC4
Remote mounted transmitter
B
205
108
GA
Dimensions with heating jacketheating coil on request
No responsibility taken for errors subject to change without prior notice
13