Page 1: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Micro Heater with PT-CR thin film Thermo-Couple for high speed

Biological Measurements

Fujita Laboratory, IIS, cIRMM, the University of

Tokyo ,JAPAN F. Gillot, H. F. Araa, F. O. Morin, D. Collard

and H. Fujita

Tranducer 2007

報告人: 蘇聖欽

Page 2: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Outline

• Introduction

• Manufacturing Process

• Biological Application

• Conclusion

• References

Page 3: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Outline

• Introduction

• Manufacturing Process

• Biological Application

• Conclusion

• References

Page 4: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Thermo-Couple

• Thermoelectric voltage :

TmVAB

m : number of couples legs

ΔT : temperature between hot and cold junctions

αAB : seebeck coefficient

Page 5: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Outline

• Introduction

• Manufacturing Process

• Biological Application

• Conclusion

• References

Page 6: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Fabrication of PT-CR

Hot junction has 2.5-um * 2.5-um area

PT-CR couple offers a high Seebeck coefficient(17.4uV/K)

Page 7: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Device Calibration

Thermal device behavior using finite element analysis (FemLab 3.1)

Micro TFTC coupled with a 471Ω CMS resistor ramping the voltage by steps of 0.1V (waiting time of at least 30s)

Page 8: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Output voltage of Thermocouple

4-noctyl-4-cyanobiphenyl(8CB)

Separation line between the nematic and the smectic phase on 1.2V applied

(a)-(c) pictures taken by a BX 51 Olympus(20x)

Page 9: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Outline

• Introduction

• Manufacturing Process

• Biological Application

• Conclusion

• References

Page 10: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Unzipping dsDNA into ssDNA

Syber green binds to dsDNA but not to ssDNA

The final quick decrease corresponded to unzipping of a DNA that completed within 52 ms(stage 3)

16um long λDNA together with fluorecent on the thermodevice

Page 11: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Outline

• Introduction

• Manufacturing Process

• Biological Application

• Conclusion

• References

Page 12: Micro Heater with PT-CR thin film Thermo-Couple for high speed Biological Measurements Fujita Laboratory, IIS, cIRMM, the University of Tokyo,JAPAN F

Conclusion

• Showing fast thermal shifting as well as thermal sensing in the ms range.

• Thermocouple can use for other application (actuator).