User’s Manual 개시 매뉴얼(한국어판)은 CD-ROM에 붙이고 있습니다 사용자 매뉴얼(영어판, 일본어판)은 CD-ROM에 붙이고 있습니 자세하는 곳은 사용자 매뉴얼을 봐 주십시요 开机手册 ( 中文版 ) 请参考 CD-ROM 用户手册 ( 英语版,日语版 ) 请参考 CD-ROM 其他详细内容请参考用户手册 (CD-ROM) Inbetriebnahme Anleitung (Deutsche Version) ist auf CD-ROM enthalten. Bedienungsanleitung (Englische Version) ist auf CD-ROM enthalten. Nutzen Sie die Bedienungsanleitung, wenn Sie Details benötigen. Start-up Manual (except English version) is included in CD-ROM. User's Manual (English version) is included in CD-ROM. Refer to User's Manual (CD-ROM) when you need the details. Model FLXA21 Two-wire Analyzer Start-up Manual IM 12A01A02-12E IM 12A01A02-12E 2nd Edition

Upload: others

Post on 10-Oct-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

User’sManual

개시 매뉴얼(한국어판)은 CD-ROM에 붙이고 있습니다

사용자 매뉴얼(영어판, 일본어판)은 CD-ROM에 붙이고 있습니

자세하는 곳은 사용자 매뉴얼을 봐 주십시요

开机手册 (中文版 )请参考 CD-ROM

用户手册 (英语版,日语版 )请参考 CD-ROM

其他详细内容请参考用户手册 (CD-ROM)

Inbetriebnahme Anleitung (Deutsche Version) ist auf CD-ROM enthalten.

Bedienungsanleitung (Englische Version) ist auf CD-ROM enthalten.Nutzen Sie die Bedienungsanleitung, wenn Sie Details benötigen.

Start-up Manual (except English version) is included in CD-ROM.

User's Manual (English version) is included in CD-ROM.Refer to User's Manual (CD-ROM) when you need the details.

Model FLXA21Two-wire AnalyzerStart-up Manual

IM 12A01A02-12E

IM 12A01A02-12E2nd Edition

Page 2: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

i

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

u IntroductionThank you for purchasing the FLXA21 Two-wire Analyzer.Please read the following respective documents before installing and using the FLXA21.

n Notes on Handling User’s Manuals• Please hand over the user’s manuals to your end users so that they can keep the user’s

manuals on hand for convenient reference.• Please read the information thoroughly before using the product.• The purpose of these user’s manuals is not to warrant that the product is well suited to any

particular purpose but rather to describe the functional details of the product.• No part of the user’s manuals may be transferred or reproduced without prior written

consent from YOKOGAWA.• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at

any time, without notice or obligation.• Ifyouhaveanyquestions,oryoufindmistakesoromissionsintheuser’smanuals,please

contact our sales representative or your local distributor.

n Drawing ConventionsSomedrawingsmaybepartiallyemphasized,simplified,oromitted,fortheconvenienceofdescription.Some screen images depicted in the user’s manual may have different display positions or character types (e.g., the upper / lower case). Also note that some of the images contained in this user’s manual are display examples.

Media No. IM 12A01A02-12E 2nd Edition : Aug. 2010 (YK)All Rights Reserved Copyright © 2010, Yokogawa Electric Corporation

Page 3: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

ii

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

u Safety Precautions

n Safety,Protection,andModificationoftheProduct• In order to protect the system controlled by the product and the product itself and ensure

safe operation, observe the safety precautions described in this user’s manual. We assume no liability for safety if users fail to observe these instructions when operating the product.

• Ifthisinstrumentisusedinamannernotspecifiedinthisuser’smanual,theprotectionprovided by this instrument may be impaired.

• If any protection or safety circuit is required for the system controlled by the product or for the product itself, prepare it separately.

• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to as YOKOGAWA) when replacing parts or consumables.

• Modificationoftheproductisstrictlyprohibited.• The following safety symbols are used on the product as well as in this manual.

DANGERThis symbol indicates that an operator must follow the instructions laid out in this manual in order to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes what special care the operator must take to avoid such risks.

WARNINGThis symbol indicates that the operator must refer to the instructions in this manual in order to prevent the instrument (hardware) or software from being damaged, or a system failure from occurring.

CAUTIONThis symbol gives information essential for understanding the operations and functions.

Note!This symbol indicates information that complements the present topic.

This symbol indicates Protective Ground Terminal

This symbol indicates Function Ground Terminal (Do not use this terminal as the protective ground terminal.)

n Warning and DisclaimerThe product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any person or entity with respect to any direct or indirect loss or damage arising from using the product or any defect of the product that YOKOGAWA can not predict in advance.

Page 4: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

iii

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

n FLXA21l Regulatory Compliance

Safety: EN61010-1 UL 61010-1 (Pending) CSA C22.2 No.61010-1 (Pending)EMC: EN61326-1 Class A,Table 2 (For use in industrial locations) EN61326-2-3 AS/NZS CISPR11Installation altitude: 2000 m or lessCategory based on IEC 61010: I (Note 1)Pollution degree based on IEC 61010: 2 (Note 2)Note1: Installationcategory,calledover-voltagecategory,specifiesimpulsewithstandvoltage. Equipment with "Category I" (ex. 2-Wire transmitter) is used for connection to circuits in which measures are taken to limit

transient over-voltages to an appropriately low level.Note 2: Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric

strength. Degree 2 is the normal indoor environment.

DANGER

Installation and wiringThe FLXA21 should only be used with equipment that meets the relevant IEC, American or Canadian standards. Yokogawa accepts no responsibility for the misuse of this unit.

DANGER

Don’t install “general purpose type” instruments in the hazardous area.

CAUTION

The Instrument is packed carefully with shock absorbing materials, nevertheless, the instrument may be damaged or broken if subjected to strong shock, such as if the instrument is dropped. Handle with care.

CAUTION

This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the industrial environment only.

CAUTION

TheHARTcommunicationmaybeinfluencedbystrongelectromagneticfield.In this case another trial of the HART communication and/or operation with FLXA21 touch screen can be carried out.

WARNING• Do not use an abrasive or organic solvent in cleaning the instrument.• Do not modify the FLXA21.• Substitution of components may impair suitability for Division 2. Do not remove or replace while circuit is live unless area is know to be non-hazardous. Explosion Hazard – Do not disconnect equipment unless area is known to be

nonhazardous. Do not reset circuit breaker unless power has been removed from the equipment or the area

is known to be non-hazardous.

Page 5: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

iv

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

WARNINGElectrostaticdischargeThe FLXA21 contains devices that can be damaged by electrostatic discharge.When servicing this equipment, please observe proper procedures to prevent such damage.Replacement components should be shipped in conductive packaging. Repair work should be done at grounded workstations using grounded soldering irons and wrist straps to avoid electrostatic discharge.

l Howtodisposethebatteries:This is an explanation about the new EU Battery Directive(DIRECTIVE 2006/66/EC).This directive is only valid in the EU.Batteries are included in this product.Batteries incorporated into this product cannot be removed by yourself.Dispose them together with this product.WhenyoudisposethisproductintheEU,contactyourlocalYokogawaEuropeB.V.office.Donotdispose them as domestic household waste.Battery type: Manganese dioxide lithium batteryNotice: The symbol means they shall be sorted out and collected as ordained in ANNEX

II in DIRECTIVE 2006/66/EC.

n Warranty and serviceYokogawa products and parts are guaranteed free from defects in workmanship and material under normal use and service for a period of (typically) 12 months from the date of shipment from the manufacturer.Individual sales organisations can deviate from the typical warranty period, and the conditions of sale relating to the original purchase order should be consulted. Damage caused by wear and tear, inadequate maintenance, corrosion, or by the effects of chemical processes are excluded from this warranty coverage.In the event of warranty claim, the defective goods should be sent (freight paid) to the service department of the relevant sales organisation for repair or replacement (at Yokogawa discretion). The following information must be included in the letter accompanying the returned goods:

• Part number, model code and serial number• Original purchase order and date• Length of time in service and a description of the process• Description of the fault, and the circumstances of failure• Process/environmental conditions that may be related to the failure of the device.• A statement whether warranty or nonwarranty service is requested• Complete shipping and billing instructions for return of material, plus the name and phone

number of a contact person who can be reached for further information.Returned goods that have been in contact with process fluids must be decontaminated/disinfected before shipment. Goods should carry a certificate to this effect, for the health and safety of our employees.

Material safety data sheets should also be included for all components of the processes to which the equipment has been exposed.

Page 6: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

v

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

n CopyrightandTrademarkNoticesThe copyrights of online manual contained in the CD-ROM are reserved.TheonlinemanualisprotectedbythePDFsecurityfrommodification,however,itcanbeoutputvia a printer. Printing out the online manual is only allowed for the purpose of using the product. When using the printed information of the online manual, check if the version is the most recent one by referring to the CD-ROM’s version.No part of the online manual may be transferred, sold, distributed (including delivery via a commercial PC network or the like), or registered or recorded on video tapes.

Adobe, Acrobat and Acrobat Reader are either registered trademarks or trademarks of Adobe Systems Incorporated in the United States and/or other countries.All other company and product names mentioned in this user’s manual are trademarks or registered trademarks of their respective companies.We do not use TM or ® mark to indicate those trademarks or registered trademarks in this user’s manual.

l FLXA21’s fonts(c) Copyright 2000-2001 /efont/ The Electronic Font Open Laboratory. All rights reserved.Redistributionanduseinsourceandbinaryforms,withorwithoutmodification,arepermittedprovided that the following conditions are met:

1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.

2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.

3. Neither the name of the team nor the names of its contributors may be used to endorse or promoteproductsderivedfromthisfontwithoutspecificpriorwrittenpermission.

THIS FONT IS PROVIDED BY THE TEAM AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE TEAM OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS FONT, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.

Page 7: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

Toc-1

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Model FLXA212-Wire AnalyzerStart-up Manual

CONTENTS

IM 12A01A02-12E 2nd Edition

u Introduction ....................................................................................................iu Safety Precautions .......................................................................................ii1. Instrumentcheck ..........................................................................................12. Wiring and Installation .................................................................................3

2.1 Installation site ..................................................................................................... 42.2 Installingthecableglands .................................................................................. 42.3 Wiringthepowersupply ..................................................................................... 6

2.3.1 Grounding ............................................................................................. 7

2.3.2 Connection of the power supply ........................................................... 7

2.3.3 Wiring cover .......................................................................................... 8

2.4 Sensor wiring ....................................................................................................... 82.4.1 pH/ORP sensor wiring ........................................................................ 10

2.4.2 Wiring the conductivity (SC) sensor .................................................... 16

2.4.3 Wiring the inductive conductivity (ISC) sensor ................................... 17

2.4.4 Connecting the dissolved oxygen (DO) sensor .................................. 18

2.5 Mountingmethods ............................................................................................. 192.5.1 Plastic housing .................................................................................... 19

2.5.2 Stainless steel housing ....................................................................... 21

3. Operation .................................................................................................... 253.1 Changelanguage ............................................................................................... 263.2 Screen operation ................................................................................................ 273.3 Passwords .......................................................................................................... 31

Appendix1 UsersettingtableofPH/ORP ............................................App.1-1Appendix2 UsersettingtableofSC .....................................................App.2-1Appendix3 UsersettingtableofISC ....................................................App.3-1Appendix4 UsersettingtableofDO ....................................................App.4-1Revision Record .......................................................................................................i

Page 8: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<1.Instrumentcheck> 1

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

1. InstrumentcheckUpon delivery, unpack the instrument carefully and inspect it to ensure that it was not damaged during shipment. If damage is found, retain the original packing materials (including the outer box)andthenimmediatelynotifythecarrierandtherelevantYokogawasalesoffice.

l CheckingthemodelandsuffixcodeLoosen four front panel screws of the FLXA21, open the panel to the left, and make sure the modelandsuffixcodeonthenameplateaffixedtothebacksideofthefrontpanelagreeswithyour order.

24VDC

FLXA21

STYLE

2-WIRE ANALYZER

Made in Japan

-20 to +55°C4-20mA DC

NO.

TEMP. RANGEOUTPUTSUPPLY

SUFFIXMODEL

-D-P-S-AA-P1-NN-AN-L1-N-NN/UM/SCT/H6/CA1/K

Figure1-1. InsideoftheFLXA21andexampleofnameplate

Note!

Be sure to apply correct power to the unit, as detailed on the nameplate.

CAUTION

When you open the front panel, make sure the screws are completely out of the screw holes, and then open the front panel slowly in order not to damage the threaded parts on the housing. If the threaded parts are damaged and the screws cannot be tightened, the waterproof performance will deteriorate.

CAUTION

Be careful not to lose the four front panel screws.

Page 9: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<1.Instrumentcheck> 2

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

l CheckingtheaccessoriesMake sure the accessories in Table 1.1 are included.

Table1.1 Accessories

Product Name Quantity RemarkCable glands 3 or 4 sets In the case of the plastic housing. 4 sets

when 2 sensors are used.With a rubber plug.

7 sets In the case of the stainless steel housing.With rubber plugs (5pcs).

PH analyzer Jumper 2 pcs/sensorGrommet set 1 set/sensor

Option Bracket (1 set) Option code /UM, /U, /PMSun shade hood (1 set) Option code /H6, /H7, /H8

Tag plate (1 set) Option code /SCT

Adapter for conduit work (1 set) Option code /CB4, /CD4, /CF4Instruction Manual (CD) 1 copyStartup Manual 1 copy This document

Page 10: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 3

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2. Wiring and InstallationInstall the cable glands into the FLXA21 before installing it. Upon delivery, the cable glands are not installed. Refer to section 2.2. and install the M20 cable glands.

(Plastic housing)(Stainless steel housing)

*2 *2

*3

*4

*5

*5

n n n n n n n n n

n n n n n n n n n

+–

*1: Use a 2-conductor shielded cable with an outside diameter of 6 to 12 mm.*2: Ground FLXA21 ( Class D ground: 100 ohm or less)

The way of connecting the grounding cable varies depending on the plastic housing and stainless steel housing.

In the case of the plastic housing, connect the grounding cable to the terminal of the power module inside, and in the case of the stainless steel housing, connect the grounding cable to the terminal of the housing.

Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing.

*3: Refer to module*4: Two modules can be connected to the same object. When measuring

inductive conductivity, only one module can be connected. *5: The terminal box may need to be connected depending on the object under

test or the sensor selected.*6: This line is connected to a distributor or 24V DC power supply.*7: Two outputs of PH201G or SDBT are same.

*1

WTB10 or BA10Terminal box FLXA21 2-Wire Analyzer

Sen

sor

WTB10 or BA10Terminal box

Sen

sor

161519171318141211

PH

NC1817141315161211

DO

15161417131211

ISC

NC161514131211

SC

PH Module

SC Module

ISC Module

DO Module

Power supply

Output 1 (1-5V DC)

20 to 130V DCor80 to 138V AC, 47 to 63Hz

Power supply20 to 130V DCor80 to 138V AC, 47 to 63Hz

HOLD WASHFAIL

Output 2 (1-5V DC)

Output 1 (1-5V DC)

Output 2 (1-5V DC)

Case of DistributorPH201G (Style B)

Case of DistributorSDBT

(*6)

*6

*7

*7

(*6)

A+

– CMN CMNB

CD

FH

L N

A+

–+

–B

1+

–2

FH

L N

Figure 2.1 Wiring diagrams

Page 11: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 4

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.1 Installation siteThe FLXA21 is weatherproof and can be installed both inside and outside. It should, however, be installed as close as possible to the sensor to avoid long cable runs between the instrument and sensor. When the pH sensor is used, the cable length including the sensor cable should not exceed 20 meters (65.6 feet).

Select an installation site that meets the following conditions.

• Mechanical vibrations and shocks are negligible• No relay switch and power switch are installed close to the converter• There is space for cable connection beneath the cable glands• Not exposed to direct sunlight or severe weather conditions• Maintenance is possible• No corrosive atmosphere

Select an installation site where the ambient temperature and humidity are within the limits of the instrumentspecifications(seeIM12A01A02-01Echapter2).Iftheinstrumentisinstalledoutsideand exposed to direct sunlight, a sun shade hood should be used.

2.2 InstallingthecableglandsBefore installing this instrument, open the front panel and remove the wiring cover, and then install the cable glands.

CAUTION

Be careful not to lose the wiring cover screws.

The supplied cable glands are for cables with an outside diameter of 6 to 12 mm (0.24 to 0.47 inches). Unused cable entry holes must be sealed with cable glands including the supplied close up plugs.

(A hole is drilled, if specified.)

Plastic Housing Stainless Steel HousingF0202.ai

(For sensor 2 cable)For sensor 1 cable

(For sensor 2 cable)For sensor 1 cable

For power supply

(Either terminal is provided depending on the option code.)

For power supply

For grounding cable

Figure2.2 Cableglanddiagram

Page 12: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 5

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

CAUTION

Be careful not to be injured by the sharp hole edges on the housing.

Follow the procedure below to install the supplied cable glands or an adapter for conduit work if the latter is ordered.

Cable gland nut

O-ring

Gable glandSleeve(for grounding cable line of the plastic housing)Cable gland cap

Close up plug(rubber) F0203.ai

Figure2.3 Cableglands

The unused cable glands should be sealed with the supplied close up plug.

DANGER

When installing cable glands, hold cable glands and tighten cable gland nuts to a torque of 6 N•m. If cable glands, not gland nuts, are tightened, O-rings may be come out from the proper positions.

l AdapterforconduitworkWhen protecting the cable with a conduit, use an adapter (option codes: /CB4, /CD4, or /CF4). Settheadapterasshowninfigure2.4.insteadofusingthesuppliedM20cablegland.

Adapter

49(1.93")

G1/2 screw (/CB4), 1/2 NPT screw (/CD4)M20x1.5 screw (/CF4)

Approx.55(2.2")

Packing

Unit: mm(inch)Nut

F0204.ai

Figure2.4 Adapterforconduitwork(option)

DANGER

Whenusingacableconduit,useaflexibleconduittoavoidstressontheconduitadapter.The stress on the conduit adapter may damage the housing.

Page 13: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 6

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.3 WiringthepowersupplyFirstmakesurethatthepowersupplyisinaccordancewiththegivenspecifications.

CAUTION

The FLXA21 is used with a DC power supply. Do not use an alternating current or 100 volt mains power supply.

The cable leading to the distributor (power supply) supplies the power to and transmits the output signal from the FLXA21.Use a two conductor shielded cable with a minimum cross sectional area of 1.25 mm2 and an outside diameter of 6 to 12 mm. The cable gland supplied with the instrument is compatible with these diameters.Terminal screw size is M4, and torque of screw up is 1.2 N•m.Pin terminal, ring terminal and spade terminal can be used for the power supply terminals in the plastic housing and the stainless housing, and the grounding terminal in the plastic housing. For the grounding terminal on the stainless steel case, ring terminal should be used. Pin terminal: pin diameter: max. 1.9 mm Ring and spade terminal: width: max. 7.8 mm

Figure 2.5 Power supply and ground wiring cover

WARNING

Cables that withstand temperatures of at least 60°C should be used for wiring.

WARNING

Always place the shield cover over the power supply and contact terminals for safety reasons and to avoid interference.

Page 14: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 7

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.3.1 GroundingThe way of connecting the grounding terminal varies depending on whether the housing is plastic or stainless steel.

For the safety of the user and to protect the instrument against interference, the converter housing must always be connected to a class D ground (a ground resistance of 100 ohm or less).

WARNING

The minimum cross sectional area of the protective grounding cable should be 0.75 mm2.

l GroundingtheplastichousingA single separate cable line for grounding should be connected to the terminal inside the housing. (Figure 2.6A)The cable with an outside diameter of 3.4 to 7 mm should be used. For this cable size, use the sleeve delivered with cable glands. (Figure 2.3)The shield of a two conductor shielded cable connecting the distributor and transmitter should be connected to the terminal on the terminal block inside the transmitter.This shield should not be connected to the ground terminal on the distributor side.

l GroundingthestainlesssteelhousingThe terminal outside the transmitter housing should be used for grounding. (Figure 2.6B or C)The position of grounding terminal of stainless steel housing with the option code of panel mountinghardware(/PM)oruniversalmountingkit(/UM)isshownonthefigure2.6C.For this terminal, ring terminal should be used.The shield of a two conductor shielded cable connecting the distributor and transmitter should be connected to the terminal on the terminal block inside the transmitter. This shield should not be connected to the ground terminal on the distributor side.

A B CPlastichousing(Internalgrounding) Stainlesssteelhousing(Externalgrounding)Figure 2.6 Grounding

2.3.2 ConnectionofthepowersupplyOpen the front panel and remove the wiring cover to make the terminal block accessible.Passthesupply/outputcablethroughthepowercableholeindicatedinfigure2.2tothetransmitter and connect it to the terminals marked +, - and according to the power terminal marks.

Page 15: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 8

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.3.3 Wiring coverWhenattachingthewiringcover,alignthecoverwiththeareasmarkedwithinthefigureandfastenitwiththescrewintheareamarkedwith∆.

WARNING

The screw in the area marked with ∆ should be tightened to a torque of 0.35 to 0.45 N•m.

If a second module is used, wire the sensor for the second module and attach the wiring cover (refertofigure2.8).Then,wirethesensorforthefirstmoduleandattachthewiringcover(refertofigure2.9).The sensor wiring is shown in section 2.4.

CAUTION

Be careful not to lose the wiring cover screws.

2.4 Sensor wiringThe FLXA21 can be used with a wide range of commercially available sensor types, both from Yokogawa and other manufacturers.Terminal screw size is M4, and torque of screw up is 1.2 N•m.Pin terminal, ring terminal and spade terminal can be used. Pin terminal: pin diameter: max. 1.9 mm Ring and spade terminal: width: max. 7.8 mmFor details on the sensors, refer to the respective instruction manuals of the sensors.

CAUTION

Whenthewiringofsensorsisfinished,besuretoattachthewiringcovers.Furthermore,makesure the module is locked with the mechanisms on the left and right of the module.

Unlock LockFigure2.7 Lockingthemodule

CAUTION

Confirmthatalllocking-tabs(includingforBLANKslots)are“Locked”positionwhenclosethefront panel.If locking-tabs are “Unlock” position, the front panel may be interfered with locking-tabs.

Page 16: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 9

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Figure2.8 Modulewiringcover(lowerlevel)andjumperholder

Note: Jumpers are supplied only with the pH sensor.

Figure2.9 Modulewiringcover(upperlevel)

CAUTIONWhen input 2 is used, the upper-level module is for input 1. Be careful not to make a mistake when wiring the modules.

Wheninput2isused,firstwireinput2whichisonthelowersecondmodule,andattachthewiringcover.Evenwheninput2isnotused,firstattachthepowersupply/groundingwiringcoverandthenthewiringcoverforthesecondmodule,andfinallywireinput1.

When all sensor wiring is completed, close the front cover of the FLXA21, and turn on the power. Make sure that the display is normal.

WARNINGDo not tighten up four front panel screws one by one.Each front panel screw should be tightened up in two times of screwing. And, firstly the screw at the upper left should be screwed a bit, the next is at the lower right, third is at the upper right, and fourth is at the lower left. The second round is the same sequence again to tighten up four screws.Do not use an electric screwdriver with high revolutions. If an electric screwdriver is used for these front panel screws, the revolutions of the electric screwdriver should be less than 400 rpm.Four screws should be tightened to the following torque; 0.8 to 0.9 N•m (for the plastic housing) 1.5 to 1.6 N•m (for the stainless steel housing)

Page 17: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 10

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.4.1 pH/ORPsensorwiring161519171318141211

PH

Figure 2.10 PH module terminal diagram

n ImpedancemeasurementjumpersettingsImpedance measurement is a powerful diagnostic tool. In order to perform impedance measurementsitisimportanttohaveagoodjumpersetting.Thetableandfigurebelowwillguideyou to make the right setting.

Note!

Itisimportanttodecidefirstwhichapplicationandwhichsettingsareappropriatefortheinstallation.Thisdecisionis best done before the jumpers are installed, because the cables will rest on top of the jumpers in their installed positions.

Figure 2.11 shows the jumper positions.For Low impedance 18 or 19 should be shorted by a jumper.When shipped, no jumper is placed on any terminals. Two jumpers are placed in a plastic bag and supplied with the product.Typical setting for pH measurement, 13 and 18 are shorted to become a low impedance input. Unused jumpers should be stored in jumper holders in the cover, as shown in Figure 2.8.Special electrodes using 2 glass sensors (e.g. Pfaudler, SC24V) doesn’t need jumper(s).

161519171318141211

PH

161519171318141211

PH

Input 1Input 2

Normal pH sensors (including FU20)

Glass sensor on Input 1.Reference sensor on Input 2.

Input 1Input 2

ORP (Redox measurement)

Metal sensor on Input 1.Normal reference on Input 2.

Figure 2.11 Jumper placements for low impedance setting

After placing jumper(s), set values of Input 1 and Input 2 impedance by [Impedance setting] menu.Refer IM12A01A02-01 section 5.2.5 for details.Initial setting value of [Impedance setting] is same as Figure2.11 (Normal pH sensors).

n Sensor wiringRefertofigure2.13to2.15,whichincludesdrawingsthatoutlinesensorwiring.The FLXA21 can be used with a wide range of commercially available sensor types, both from Yokogawa and other manufacturers. The sensor systems from Yokogawa fall into two categories; theonesthatuseafixedcableandtheoneswithseparatecables.Toconnectsensorswithfixedcables,simplymatchtheterminalnumbersintheinstrumentwiththeidentificationnumbersonthecableends.The separate sensors and cables are not numbered, but instead use a color-coding system. The electrodes have a colored band incorporated in the label on the connection cap:

Page 18: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 11

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

• Red for measuring electrodes (both pH and ORP)• Yellow for reference electrodes• Blue for combined sensors with both measuring and reference elements in the same

body• Green for temperature sensors

The recommended procedure is to color-code each end of the cables to match the sensors with the color strips provided with each cable. This provides a quick way to identify the ends of the cables belonging to a particular sensor when they are installed.

n ConnectioncableThe coaxial cable has two connections.

coreshield

redblue

Coaxial

redbluewhite

Triaxial

Figure 2.12

• Red to measuring element• Blue to screen (shield)

The triaxial cable has three connections, (it has an extra white wire termination) these wires are connected:

• Red to measuring element• Blue to reference• White to screen (shield)

To connect the other sensor systems, follow the general pattern of the terminal connections as listed below:

TerminalSingle Measurement

(pHorRedox)Combinedmeasurement

(pHandRedox)pH ORP pH and ORP pH and rH

11 Temperature 1 – Temperature 1 Temperature 112 Temperature 2 – Temperature 2 Temperature 213 Reference Reference Reference Reference14 Liquid Earth Liquid Earth ORP ORP15 pH ORP pH pH16 Shield of no. 15 Shield of no. 15 Shield of no. 15 Shield of no. 1517 Shield of no. 13 Shield of no. 13 Shield of no. 13 Shield of no. 13

Note!

For ORP measurement, temperature is not required.

Page 19: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 12

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

*

*

*

*

Green

Yellow

Red

Red

Blue

BlueBlue

RedBlackRed

Cable markers

11 Temperature12 Temperature13 Reference14 Solution ground15 Glass (measure)16 Shield17 Shield

FLXA21 FLXA21

FLXA21

FLXA21

Separate electrodes for pH measurement

11 Temperature12 Temperature13 Reference14 Solution ground15 Glass (measure)16 Shield17

111213141516

FU20/FU24 four-in one sensor for pH measurement

*

*

*

16 Shield15 Glass (measure)14 Solution ground13 Reference12 Temperature11 Temperature

Red

Blue

Green

RedBlue

BlueBlack

White 17 Shield

Cable markers

Combined glass/reference electrode for pH measurement

15 Glass (measure)14 Solution ground13 Reference12 Temperature11 Temperaturered

blueblackyellowtransparentblack (coax) 16 Shield

17

Retractable sensor assembly PR10(Also PD20, PF20 & PS20)

Note:Connect cables to similaly marked terminals: 11 to 11, 12 to 12, etc.This configuration also enables combined pH+ORP measurements

Note:Connect cables to similaly marked terminals: 11 to 11, 12 to 12, etc.

FLXA21SC24V

16 Shield15 Glass (measure)14 Solution ground13 Reference12 Temperature11 TemperatureRed

Blue

Brown

WhiteYellow

17 Shield

REFTC PHLiquidEarth

CombPH/REF

pH

Figure 2.13 Sensor wiring for pH measurement

Separete electrodes for pH and Redox(ORP/rH) measurement

FU20/FU24 Four-in-one sensor for pH and Redox measurement

***

Cable markersThe blue cable can also be disconnected

Note:Connect cables to similaly marked terminals: 11 to 11, 12 to 12, etc.This configuration also enables combinedpH+ORP measurements.

11 Temperature12 Temperature13 Reference14 Metal ORP sensor15 Glass (measure)16 Shield17 Shield

11 Temperature12 Temperature13 Reference14 Solution ground15 Glass (measure)16 Shield17

111213141516

*

*

***

Green

Yellow

Red

Red

Blue

BlueBlue

RedRed & Blue

Red

REFTC PHORP

pH+ORPFLXA21 FLXA21

Figure2.14 Sensorwiringforcombined(pH+Redox)measurement

Page 20: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 13

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

11 Temperature

12 Temperature

13 Reference

14 Solution ground (ORP)

15 Metal

16 Shield17

FU20/FU24 four-in-one sensor for ORPmeasurement with normal reference

Separate electrodes for ORPmeasurements

* Cable Markers

* Cable Markers

RED

BLUE

RED

BLACK

RED

BLUE

BLUE

Red*

Yellow*

Green*11 Temperature

12 Temperature

13 Reference

14 Solution ground

15 Metal (measure)

16 Shield

17 Shield

11 Temperature

12 Temperature

13 Reference

14 Solution ground

15 Metal (measure)

16 Shield

17 Shield

Combined metal/reference electrode for ORP measurement

RED

BLUE

BLUE

BLACK

RED

WHITE

Blue*

Green*

11 Temperature

12 Temperature

13 Glass

14 Solution ground

15 Metal

16 Shield

17 Shield

Combined metal/glass electrodefor sensor pH compensated Redox (ORP/rH) measurement

RED

BLUE

BLUE

BLACK

RED

WHITE

Blue*

Green*

16 15

Single electrodes for rH measurement

RED

BLUE

RED

BLACK

RED

BLUE

BLUE

Red*

Yellow*

Green*11 Temperature

12 Temperature

13 Glass

14 Solution ground

15 Metal

16 Shield

17 Shield

* Cable Markers

* Cable Markers

111213

14

11 Temperature

12 Temperature

13 Reference

14 Solution ground (ORP)

15 Metal

16 Shield17 Shield

FU20/FU24 four-in-one sensor for ORPmeasurement with pH sensor as reference

16

111215

14

13(not used) (not used)

ORP

REFTC LiquidEarth

CombPH/ORP

ORP CombORP/REF

FLXA21 FLXA21

FLXA21 FLXA21

FLXA21 FLXA21

Figure 2.15 Sensor wiring for redox measurement

n SensorcableconnectionwithspecialgrommetIn order to seal multiple sensor cables into FLXA21, a special grommet is provided that is designed to accommodate one, two or three sensor cables (5 mm dia.) plus a liquid earth cable (2.5 mm dia.). In the pack with the grommet are blanking pieces to close any unused holes. When correctly assembled, the grommet maintains the IP66 and NEMA 4X rating of the FLXA21 housing.

Note!ThespecialgrommetisintendedtobeusedtosealthemultiplecablesfromtheYokogawaflowfittingssuchasFF20.The designated cables are WU20 sensor cables, which are approximately 5 mm (0.2”) in diameter, and K1500FV liquid earth cables, which are approximately 2.5 mm (0.1”) in diameter.

Page 21: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 14

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

For sensor systems using a single cable, like the FU20/FU24 and the PR10, PD20, PF20 and PS20, the standard gland will accommodate the cable adequately. Single cables between approximately 6 mm and 12 mm (0.24” and 0.47”) can be sealed properly with these glands and the standard tule.

Removestandard tule

Figure 2.16 Grommet set use Content of grommet set

n Sensorcableconnectionsusingjunctionbox(BA10)andextensioncable(WF10)

Where a convenient installation is not possible using the standard cables between sensors and converter, a junction box and extension cable may be used. The Yokogawa BA10 junction box and the WF10 extension cable should be used. These items are manufactured to a very high standardandarenecessarytoensurethatthespecificationsofthesystemcanbemet.Thetotalcablelengthshouldnotexceed60metres(e.g.5mfixedcableand55mextensioncable).

1112

12

1314

1416

15

1314

1416

15

17

1117

11 Red12 Blue

14 Overall Screen

15 Core 16 ScreenWhite Co-axial cable

13 Core 17 ScreenBrown Co-axial Cable

11

12

13

14

15

16

17

11

12

13

14

15

16

17

BA10

WF10

FLXA21

17 (screen)

14 (overall screen)

12 (blue)

11 (red)

13 (core)

16 (screen)

15 (core)Co-axial cable

(white)Co-axial cable

(brown)

Figure2.17 ConnectionofWF10extensioncableandBA10junctionbox

Page 22: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 15

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Anextensioncablemaybepurchasedinbulkquantitiesorinpre-finishedlengths.Inthecaseofbulk quantities cut to length, then it is necessary to terminate the cable as shown below.Termination procedure for WF10 cable.

1. Slide 3 cm of heat shrink tube (9 x 1.5) over the cable end to be terminated.2. Strip 9 cm of the outer (black) insulating material, taking care not to cut or damage internal

cores.3. Remove loose copper screening, and cut off the cotton packing threads as short as

possible.4. Strip insulation from the last 3 cm of the brown, and the white coaxial cores.5. Extract the coaxial cores from the braid, and trim off the black (low-noise) screening material

as short as possible.6. Insulate the overall screen and the 2 coaxial screens with suitable plastic tubing.7. Strip and terminate all ends with suitable (crimp) terminals and identify with numbers as

shown.8. Finally shrink the overall heat shrink tube into position.

3 cmheat shrink

9 cmremove insulation

Cotton threads

3 cm

111217

15161314

Figure 2.18

Page 23: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 16

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.4.2 Wiringtheconductivity(SC)sensorNC161514131211

SC

Figure 2.19 SC module terminal diagram

Generally, signals from sensors are at low voltage and low current level. Thus a lot of care must be taken to avoid interference. Before connecting sensor cables to the converter make sure that following conditions are met:

– the sensor cables are not mounted in tracks together with high voltage and or power switching cables

– only standard sensor cable or extension cable is used– the converter is mounted within the distance of the sensor cables (max. 10 m) + up to 60m

WF10 extension cable.– thesetupiskeptflexibleatthesensorsendforeasyinsertionandretractionofthesensorin

thefitting.

Refertofigure2.19,whichincludesdrawingsthatoutlinesensorwiring.The FLXA21can be used with a wide range of sensor types. The sensor system fall into two categories,theonesthatusefixedcablesandtheoneswithseparatecables.Toconnectsensorswithfixedcables,simplymatchtheterminalnumbersintheinstrumentwiththeidentificationnumbersonthecableends.

FLXA21

11 Temperature

12 Temperature13 1st inner cell

14 1st outer cell15 2nd inner cell16 2nd outer cell

FLXA21

SX42-SX** -* F sensors

Red Yellow/Green

Brown

11 Temperature

12 Temperature

13 Inner electrode

14 Outer electrode

15 Inner electrode

16 Outer electrode

BA10 Connection BoxWF10 Extension Cable

15

16

14

14

17

13

12

11

15

16

14

14

17

13

12

11

Figure 2.19 Sensor wiring diagrams

n Sensorcableconnectionsusingjunctionbox(BA10)andextensioncable(WF10)

Refer pH sensor

Page 24: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 17

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.4.3 Wiringtheinductiveconductivity(ISC)sensor15161417131211

ISC

Figure 2.20 ISC module terminal diagram

Thermistor (Temperature sensor)

ground (shield)

Secondary Coil

Primary Coil

11 12 17 13 15 16 14

Figure 2.21 ISC40G Sensor wiring diagrams

n Sensorcableconnectionsusingjunctionbox(BA10)andextensioncable(WF10)

Refer pH sensor

Page 25: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 18

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.4.4 Connectingthedissolvedoxygen(DO)sensorNC1817141315161211

DO

Figure 2.22 DO module terminal diagram

111213141516

Galvanic sensor

1112141718

Polarographic sensor

Figure 2.23 Connections DO

CAUTION

When connecting pin-terminal 13 of the sensor cable, make sure that the connected cables are not disconnected. Also make sure that the sensor cable is not disconnected even when it is pulled slightly. If the cable is not connected tightly enough, measurement values may become unstable and measurement errors may occur.

n WiringthestandardgalvanicsensorThe individual wires of the sensor cable have markings on them. These markings match the markings on the terminals. Note: “The membrane rupture detection function is only available for the DO30G with and external liquid earth. Like in for example the

PB350G/PB30/PB360G.

n WiringothergalvanicsensorsConsulttheuser’smanualforthecoloridentificationofthesensorcableandconnectthetemperature sensor, the cathode, and the anode to the terminals (for terminals 11, 12, 13, and 15). If only one shielded cable is available, connect it to terminal 14. In this case, sensor diagnosis cannot usually be carried out.

n WiringpolarographicsensorsConsulttheuser’smanualforthecoloridentificationofthesensorcableandconnectthetemperature sensor, the anode, and the cathode to the terminals in the order of terminals 11, 12, 17, and 18. If a shielded cable is available, connect it to terminal 14.

n Sensorcableconnectionsusingjunctionbox(BA10)andextensioncable(WF10)

Refer pH sensor

Page 26: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 19

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

2.5 MountingmethodsRefertofigures2.25to27and2.32to34.TheFLXA21hasavariouskindsofmountingcapabilities.

• Panel mounting using optional mounting hardware (/PM or /UM)• Wall mounting using optional mounting hardware (/U or /UM) (for example, on a solid wall)• Pipe mounting using optional mounting hardware (/U or /UM) on a horizontal or vertical pipe

(Nominal size 50A)(Note) The universal mounting kit (/UM) contains the pipe and wall mounting hardware (/U) and the panel mounting hardware (/PM).

2.5.1 Plastichousing

144 80

80144

22.5141

151

137

20.244.9

26.5

24.7

26.5

1

(For sensor 2 cable)For sensor 1 cable

For power supply

For grounding cable

4-M6 depth 5

FB4_01.ai

Uit: mm

Figure2.24 Housingdimensionsandlayoutofglands(Plastichousing)

Page 27: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 20

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

135 0

+1138

0+113

8

195

185

17810

0

2-M5 length 35P

anel

thic

knes

s

Spacing panel cutout

4-M6 *

1 to

12

FB4_03.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Figure2.25 Panelmounting(Option/PM,/UM)(Plastichousing)

224.

8

144

13144

164

10070

15

200

50

3-ø10 holesFor wall mountig

4-M6 *

FB4_02.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Figure2.26 Wallmounting(Option/U,/UM)(Plastichousing)

Page 28: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 21

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

204.7151

5017

4.1

144

100

200

FB4_04.ai

Unit: mm

Pipe mounting (Horizontal) Pipe mounting (Vertical)

M8 U-bolt

Pipe 50A(ø60.5)

4-M6 *

*: Tighten the four screws to a torque of 2 N•m.

Figure2.27 Pipemounting(Option/U,/UM)(Plastichousing)

(100)(70)

(50)

10220184

199

72

FB4_05.ai

Unit: mm

57

Figure2.28 Housingwithstainlesssteelhood(Option/H6,/H7,/H8)(Plastichousing)

2.5.2 StainlesssteelhousingThe stainless housing with urethane coating (-U) or epoxy coating (-E) has an external plate as shownonfigure2.29A.Whenthesecoatedtypehousinghavetheoptionalmountinghardwareof the panel mounting hardware (/PM) or the universal mounting kit (/UM), an external plate is on therearsideasshownonfigure2.29B.

A BFigure2.29 Externalsofthecoatedtypehousing

Page 29: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 22

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

80

80

137

160.1

32.6150.1

165

165

144.9 22.2

137

23.7

26.5

26.5

FB4_06.ai

Unit: mm

(For sensor 2 cable)For sensor 1 cable

For power supply

4-M6 depth 5

Figure2.30 Housingdimensionsandlayoutofglands(Stainlesssteelhousing)80

80

137

160.1

32.6150.1

165

165

144.9 22.2

137

23.7

26.5

26.5

(Only with coating)

FB4_06EN.ai

Unit: mm

(For sensor 2 cable)For sensor 1 cable

For power supply

4-M6 depth 5

Figure2.31 Housingdimensionsandlayoutofglands(Stainlesssteelhousing)(Option:/PMor/UM)

Page 30: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<2.Wiringandinstallation> 23

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

100

134

178

0+1138

0+113

8

195

185

FB4_08.ai

1 to

12

4-M6 *

Spacing panel cutout

Unit: mm

2-M5 length 35

Pan

el th

ickn

ess

*: Tighten the four screws to a torque of 2 N•m.

Figure2.32 Panelmounting(Option/PM,/UM)(Stainlesssteelhousing)

234.

5

165

165

173.113

200

5015

10070

3-ø10 holesFor wall mounting

4-M6 *

FB4_07.ai

Unit: mm

*: Tighten the four screws to a torque of 2 N•m.

Figure2.33 Wallmounting(Option/U,/UM)(Stainlesssteelhousing)

Page 31: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 24

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

213.9

184.

550

160.1

165

100

200

FB4_09.ai

Unit: mm

4-M6 *

Pipe mounting (Horizontal) Pipe mounting (Vertical)

M8 U-bolt

Pipe 50A(ø60.5)

*: Tighten the four screws to a torque of 2 N•m.

Figure2.34 Pipemounting(Option/U,/UM)(Stainlesssteelhousing)

199

220184 10165

165

(100)(70)

(50)

72

FB4_10.ai

Unit: mm

57

Figure2.35 Housingwithstainlesssteelhood(Option/H6,/H7,/H8)(Stainlesssteelhousing)

Page 32: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 25

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

3. OperationWhen all wiring is completed, turn on the power of the instrument. Make sure that the LCD screen turns on. Then, the screen changes to the Quick Setup display. If the instrument is not configured,anerrorindicatormaybedisplayed,orthemeasurementvaluesdisplayedmaybeincorrect. Check the initial settings and change them to suit your purpose. For details, refer to the IM 12A01A02-01E User’s manual.

If you are not using any display unit, refer to Appendix 4 “HART communication” of IM 12A01A02-01E.

Note!

After turning on the power, it takes occasionally more than 1 minute to change to the Quick Setup display.

Page 33: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 26

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

3.1 ChangelanguageWhen the power is turned on, the Quick setup display appears.Upon delivery, the display language is set to English. If you want to use a language other than English, select your desired language.

ChineseEnglishGermanJapaneseKoreanPortugueseRussian

Change language

The instrument will restart

Are you sure? No

Warning

The instrument will restart

Are you sure? No

Warning

Start quick setup?

Yes No

Change language

Quick setup

Start quick setup?

Yes No

Change language

Quick setup

NoYes

ChineseEnglishGermanJapaneseKoreanPortugueseRussian

Figure3.1 Changelanguageflow

When the language is changed, the instrument restarts automatically.Besuretocheckandchangetheparameterstosuityourspecificationsbeforeusingtheinstrument. For details on the quick setup method and other information, please read the instruction manual (IM 12A01A02-01E). The initial values (factory settings) of parameters and a table for writing your setting values are attached to the end of this instruction manual. Please use themwhenyouconfigurethesettings.

Page 34: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 27

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

3.2 Screen operation

WARNING

Press the touch screen only with your finger. Do not use a tool with a sharp tip (ex. pencil, ballpoint pen), a thin stick, a tool with a hard tip etc. to avoid scratches on the touch screen.

The main screen operations are as follows.When you press any of the following icons on the screen, the screen changes to the respective display or executes the respective function.

Returns to the main display (or home display).

Returns to the previous display.

Moves the cursor for selecting the menu item. Pressing the icon moves the cursor down to the next item. When the cursor is on the item at the very bottom, the cursor jumps back to the item at the top.

Confirmstheinput;forexample,theselectedmenuitemorsettingvalue.

Note!

When no operation is performed for 10 minutes (or the time set for automatic recovery elapses), the screen changes to the main display (or home display) automatically. (With the exception of trend display)

ThefiguresinthissectionshowthepHmeasurementforexample.Each display is different in details due to settings. Each display is only an example for some settings.For details, refer to the IM 12A01A02-01E User's manual.

n Main display

10.38Tag:FLXA21–PH

TEXT_PH

TEXT_ORPTEXT_TEMP

19.0025.0

mV4mA 20mAPH1

°CpH

HOLDWASH

Go to Home (Fig. 3.3)

Go to Zoom (Fig. 3.5)

Go to Infomation (Fig. 3.6)

Go to Execute & Setup (Fig. 3.7)BC

A

D

J

GEF

K H NM

Figure 3.2 Example of main display

Main displayAppears upon startup when one sensor is connected.When the number of the sensor 1 or sensor 2 on the home display is pressed when two sensors are connected, the main display of the selected sensor appears.

Page 35: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 28

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

A: Measurement value: Primary value (large characters/user selectable)B, C: Measurement value: Second and tertiary values (small characters)D: UnitE: Tag No. (user programmable)F: Sensor No. *G:Sensorwellnessindicator(Moreindicatethebettercondition.)H: Hold/Wash status button (indicated only during the Hold/Wash status)J: Analog output display and parameter * (ex.: PH1…PH=Parameter, 1=sensor number)K: Additional text (set in alphanumeric characters/user programmable)M: Function buttons (Home, Zoom, Status, Execute & Setup)N: Warn/Fail indicators (indicated in blinking only during Warn/Fail status)

* When the parameter of which measurement value is indicated is selected as a process parameter, that is, mA output set on the Output setup (refer to the section 5.3, 10.2, 13.2, and 16.3), its Sensor No. is indicated as a white number in the black box, for example , and its mA output is indicated as a bar at the bottom.

Note!

Measurement values on the display can be set independent of the process parameter.

0.034mA 20mADiff-pH

pH

C

Sensor2

Sensor1

Calculated data

Differential10.38

4mA 20mAPH1

pH

Sensor2

Sensor1

R(1) Redundant sys.

Figure 3.3 Example of calculated date and redundant system

n Homedisplay(whentwosensorsareconnected)

6.35Tag:FLXA21–PH24.9

4mA 20mApH1

°C24.4 mV

pH

Tag:FLXA21–PH25.0 °C-1.8 mV

pH

10.38 ABC

F

K

H

ED

G

J

Figure3.4 Exampleofhomedisplay

Home displayAppears upon startup when two sensors are connected.(Home display is not available when only one sensor is connected)

A: Measurement value: Primary value (large characters/user selectable)B, C: Measurement value: Secondary and tertiary values (small characters)D: UnitE: Tag No. (user programmable)F: Sensor No.G:Sensorwellnessindicator(Moreindicatethebettercondition.)H: Hold/Wash indicators (indicated only during Hold/Wash status)J: Analog output display and parameter (ex.: PH1…PH=Parameter, 1=sensor number)K: Warn/Fail indicators (indicated in blinking only during Warn/Fail status)

Page 36: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 29

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

4mA 20mADiff-pH

pH

0.03C

6.3524.9 °C24 mV

pHTag:FLXA21–PH

Calculated data

4mA 20mAPH1

pH

10.38R(1)

25.0 °C19 mV

pH

10.38Tag:FLXA21–PH

Redundant system

Figure3.5 Exampleofhomedisplayofcalculateddataandredundantsystem

n Trend display

12.00

8.40

5.60

2.50

12:00 12:20 12:40

10.38pH

Minimum

MaximumAverage

Current process value

Time

Tag:

Mea

sure

men

t val

ue Maximum value on this display

Minimum value on this display

FLXA21-PH pH1

A

D

B

G

E

FC

Figure 3.6 Example of trend display

Trend displayAppears when the primary value on the main display is pressed, orwhen the Trend button on the Zoom display is pressed.

A: X axis; Time axis (user programmable from 15 minutes to 14 days)B: Y zxis; Measurement value axis (user programmable)C: Tag No.D: Current measurement value with unitE: Trend (maximum, minimum and average values during the display update time)F: Icon (current measurement value, and maximum and minimum values until the display

update)G: Warn/Fail indicators (indicated in blinking only during Warn/Fail status)

n Zoom display

Next

20

12

4

mA

15.00

HOLD

FAIL

WASH

Next

Figure 3.7 Example of zoom displayAppears when the Zoom button on the main display is pressed.The Zoom display shows an easy-to-read graphical display of the output status. Pressing Next displays the contact state if communication is set to PH201G distributor. Pressing Next again displays the status information of the transmitter.

Page 37: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 30

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

n InformationbuttonDiagnosis information on converters and sensors appears.

No malfunction detected.

Warning

When a warning indicator appears, maintenance is required. Pressing this key displays the detected malfunction message, and pressing the malfunction message displays troubleshooting guidelines for resolving the malfunction.

FailIndicates a malfunction. Pressing this key displays the detected malfunction message, and pressing the malfunction message displays troubleshooting guidelines for resolving the malfunction.

l Example of (Fail)

Fail:Temperature too high

Warn: No warn.

Measured processtemperature exceedsmaximum limit. RemedyCheck process temperature.Check programmed sensortype. Check connections andcable.

Figure 3.8 Example of Fail

If no error is detected, “No fail” and “No warn” are displayed.If an error is detected and warn is indicated, up to 3 messages appear.

n Execute & Setup Allowsyoutocalibrateandconfiguretheinstrument.Theseoperationscanbeprotectedwithapassword.

Execute: Calibration/Wash Hold Temporary output

Setup: Commissioning Change language Start Quick Setup

Execute & Setup

Figure 3.9 Example of Execute & Setup display

Page 38: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<3.Operation> 31

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

3.3 PasswordsPasswords can protect the FLXA21 from other persons’ access except a right operator’s.

Execute: Calibration/Wash HOLD Temporary output

Setup: Comissioning Change language Start Quick Setup

Execute &Setup Sensor setup Measurement setup Output setup Error configuration Logbook configuration Advanced setup Display setup Calculated data setup

Commissioning Settings Tag Passwords Date/Time Communication Factory setup

Advanced setupExecute:

Commissioning:

Entering an empty passwordresults in disabling thepassword check.

Passwords

Figure 3.10 Example of passwords setting

Execute operation and Commissioning operation can be separately protected by each password. When the Execute password is set, execution of Calibration(/Wash), HOLD or Temporary output needs to enter the password to start each execution. When the Commissioning password is set, Commissioning and Start Quick Setup needs to enter the password to start the setting-up.

For details on the password, refer to the IM 12A01A02-01E User's manual.

Page 39: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix1UsersettingtableofPH/ORP> App.1-1

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Appendix1 UsersettingtableofPH/ORP* Reference section of User’s Manual, IM 12A01A02-01E

Parameter Default values Applicablevalues User

settings*Ref. sect.

Sensor setup

Sensor type pH PH, ORP, PH+ORP 5.1

Measure-ment setup

(Measurement) pH+ORP pH, ORP, pH+ORP, pH+rH, rH

5.2.1

Tempera-ture setting

Temp. element Pt1000 Pt1000, Pt100, 3kBalco, 8k55, PTC10k,6k8,500Ω

5.2.2

Unit ºC ºC, ºF

Temp. compensa-tion

Compensation Automatic Manual, Automatic 5.2.3

Manual Manual temp. 25.0 ºC (77.0 ºF)

-30 to 140 ºC (-22.0 to 284.0 ºF)

Reference temp. 25.0 ºC (77.0 ºF)

0.0 to 100.0 ºC (32.0 to 212.0 ºF)

Process Temp. Compen-sation

pH None None, TC, Matrix, NEN6411

T.C.pH 0.000 pH/ºC(0.000 pH/ºF)

-0.100 to 0.100 pH/ºC(-0.056 to 0.056pH/ºF)

Matrix — —

ORP None None, TC

T.C.ORP 0.000 mV/ºC(0.000 mV/ºF)

-10.00 to 10.00 mV/ºC(-5.556 to 5.556mV/ºF)

PH

Page 40: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix1UsersettingtableofPH/ORP> App.1-2

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Measure-ment setup

Calibration settings

pH settings Zero and Slope units

Zero unit mV mV, pH 5.2.4

Slope unit % %, mV/pH

Limits and timig

Zero High 120.0 mV (2.03 pH)

0.00 to 500 mV (0.00 to 8.45 pH)

Zero Low -120.0 mV (-2.03 pH)

-500.0 to 0.00 mV (-8.45 to 0.00 pH)

Slope High 110.0 % (65.08mV/pH)

100.0 to 110.0 % (59.16 to 65.08mV/pH)

Slope Low 70.00 % (41.41mV/pH)

70.0 to 100.0 % (41.41 to 59.16mV/pH)

Step Range 0.01 pH 0.01 to 1.00 pH

Stabilization time 5 s 2 to 30 s

Calibr. interval 250 days 1 to 250 days

Buffers (select set)

Select buffer set

NIST/DIN 19266

NIST/DIN 19266, DIN 19267, US, Free programmable

Zero/Slope/ITP

Zero 0.000 mV (7.00 pH)

mV: (zero low) to (zero high)pH: (zeo low + ITP) to (zero high + ITP)

Slope 100.0 % (59.16mV/pH)

(slope low) to (slope high)

ITP 7.00 pH 0.00 to 14.00 pH

Auto correct(Zero, Slope)

pH Auto correct Disable Disable, Enable

ORP settingsrH settings

Limit and timing

Zero High 120.0 mV 0.00 to 500 mV

Zero Low -120.0 mV -500.0 to 0.00 mV

Slope High 110.0 % 100.0 to 110.0 %

Slope Low 70.00 % 70.0 to 100.0 %

Step Range 1 mV (0.05 rH)

1 to 100 mV (0.01 to 10.00 rH)

Stabilization time 5 s 2 to 30 s

Calibr. interval 250 days 1 to 250 days

Zero/Slope Zero 0.000 mV mV: (zero low) to (zero high)pH: (zeo low + ITP) to (zero high + ITP)

Slope 100.0 % (slope low) to (slope high)

Impedance settings

Input 1 impedance: pH, pH+ORP: HighORP: Low

Low, High 5.2.5

Low High limit 200.0 kΩ 1.000 kΩ to 1.000 MΩ

Low limit 1.000 kΩ 1.000kΩto1.000MΩ

Input 2 impedance: Low Low, High

Low High limit 200.0kΩ 1.000kΩto1.000MΩ

Low limit 1.000kΩ 1.000kΩto1.000MΩ

Page 41: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix1UsersettingtableofPH/ORP> App.1-3

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Measure-ment setup

Concentra-tion

Unit % %, mg/L, g/L, ppm 5.2.6

(pH1) 0.00 pH 0.00 to 14.00 pH

(pH2) 14.00 pH 0.00 to 14.00 pH

(Concentration1) 0.00 %, mg/L, g/L, ppm

%: 0.000 to 100.0mg/L, g/L, ppm: 0.000 to 9999(Concentration2) 100.0 %, mg/L,

g/L, ppm

Sensor diag. settings

Input 1 imp.: Disable Disable, Enable 5.2.7

FINE 10.0MΩ 1 to 10MΩ

Input 2 imp.: Disable Disable, Enable

Process time: Disable Disable, Enable

BAD limit 2000 days 1 to 10000 days

Heat cycle: Disable Disable, Enable

BAD limit 500 cycles 10 to 1000 cycles

Defineheatcycle

Heat cycle tempe 50 ºC 1 to 250 ºC

Heat cycle time: 10.0 min. 0.1 to 60.0 min.

Output setup

mA Output Output, Simulate 5.3

Output Process parameter pH1 pH1, Temperature1, ORP1, pH2, Tem-perature2, Calculated, Redundant

Setup Linear Linear, Table

Linear 0 % value: pHTemperature

ORPrH

Calculate/DifferentialCalculate/

Average

0.00 pH, 0 ºC, -500 mV, 0 rH,-1pH, -500 mV, 0 pH, -500 mV

-2 to 20 pH, -30 to 140 ºC, -1500 to 1500 mV, 0 to 100 rH, -20 to 20 pH, -1500 to 1500 mV, -2 to 20 pH, -1500 to 1500 mV

100% value:pHTemperature

ORPrH

Calculate/DifferentialCalculate/

Average

14.00 pH, 100 ºC, 500 mV, 50 rH, 1pH, 500 mV, 14 pH, 500 mV

-2 to 20 pH, -30 to 140 ºC, -1500 to 1500 mV, 0 to 100 rH, -20 to 20 pH, -1500 to 1500 mV, -2 to 20 pH, -1500 to 1500 mV

Table — —

Burn Off Off, Low, High

Damping time 0.0 s 0.0 to 3600 s

Simulate Simulation perc. 50.00 % -2.500 to 112.5 %

ConfigureHold

Lastorfixed Last Last, Fixed

Fixed value mA 12.00 mA 3.60 to 22.00 mA

Hold during Calibration(/Wash) Enable Disable, Enable

PH

Page 42: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix1UsersettingtableofPH/ORP> App.1-4

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Error configura-tion

pH too high Warn Off, Warn, Fail 5.4

pH too low Warn Off, Warn, Fail

Temperature too high Warn Off, Warn, Fail

Temperature too low Warn Off, Warn, Fail

ORP too high Off Off, Warn, Fail

ORP too low Off Off, Warn, Fail

rH too high Off Off, Warn, Fail

rH too low Off Off, Warn, Fail

Matrixconfig.error Fail Off, Warn, Fail

Calib. time exceeded Off Off, Warn, Fail

Wash half-time error Off Off, Warn, Fail

Impedance 1 too high Warn Off, Warn, Fail

Impedance 1 too low Warn Off, Warn, Fail

Impedance 2 too high Warn Off, Warn, Fail

Impedance 2 too low Off Off, Warn, Fail

Logbook configura-tion

Setting logbook

Sensor: Error on 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2) 5.5

Error off 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2)

T.C. Off Off, 1-1 (2-1), 1-2 (2-2)

Zero low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Zero high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Slope low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Slope high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Reference temp. Off Off, 1-1 (2-1), 1-2 (2-2)

Stabilization Off Off, 1-1 (2-1), 1-2 (2-2)

Imp.1 low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Imp.1 high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Imp.2 low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Imp.2 high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Wash setup Off Off, 1-1 (2-1), 1-2 (2-2)

mA: mA setup Off Off, 1-1 (2-1), 1-2 (2-2)

Holdfixedvalue Off Off, 1-1 (2-1), 1-2 (2-2)

Burn Off Off, 1-1 (2-1), 1-2 (2-2)

Linear output % Off Off, 1-1 (2-1), 1-2 (2-2)

Damping time Off Off, 1-1 (2-1), 1-2 (2-2)

Erase logbook: 1-1 (2-1) 1-1 (2-1), 1-2 (2-2)

Erase? No No, Yes

Warn if logbook “full”: No No, Yes

Page 43: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix1UsersettingtableofPH/ORP> App.1-5

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Advanced setup

Settings No action No action, Load factory settings, Save user settings, Load user settings

5.6.1

Tag FLXA21-PH (Max. 12 characters) 5.6.2

Passwords Execute: (blank) (Max. 8 characters) 5.6.3

Commissioning: (blank) (Max. 8 characters)

Date/Time Format YYYY/MM/DD

YYYY/MM/DD, MM/DD/YYYY, DD/MM/YYYY

5.6.4

Date — — —

Time — — —

Communication HART None, HART, PH201G 5.6.5

HART Network address 0 0 to 15

PV (pH1) — —

SV Temperature1 pH1 (2), Tempera-ture1 (2), Ref.imped-ance1 (2), ORP1 (2), rH1 (2), Calculated, Redundant, Empty

TV Empty, pH2

FV Empty, Temperature2

PH201G Hold contact: Enabled Disabled, Enabled

Fail contact: Fail+Warn Fail+Warn, Fail only, Disabled

Wash contact Enabled Disabled, Enabled

Interval time 6.0 h. 0.1 to 36.0 h.

Wash (Meas-ure) time

0.5 min. 0.1 to 10.0 min.

Recovery time 0.5 min. 0.1 to 10.0 min.

Manual Wash Enabled Disabled, Enabled

Imp2 Wash Disabled Disabled, Enabled

Continuous Wash

Disabled Disabled, Enabled

Factory setup — — — 5.6.6

PH

Page 44: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix1UsersettingtableofPH/ORP> App.1-6

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Display setup

Dual display

1st row PH1 PH1, PH2, Calculated. Redundant

5.7.1

2nd row PH2 PH1, PH2, Calculated. Redundant, Empty

Individual display

Main display

PH1 (2) 1st line pH1 (2) pH1 (2), Temperature1 (2), ORP1 (2), rH1 (2), Concentration1 (2)

2nd line Temperature1 (2)

pH1 (2), Tempera-ture1 (2), ORP1 (2), rH1 (2), Concentra-tion1 (2), Empty3rd line Empty,

(ORP1)

Additional text (blank) (Max. 12 characters)

Trend Trend Graph Screen:

1st Trend pH1 pH1 (2), Temperature1 (2), ORP1 (2), rH1 (2), Concentration1 (2), Calculated, Redun-dant, Empty

5.7.2

2nd Trend Temperature1

3rd Trend Empty, (ORP1)

X-axis: Timing

TIme Span (X)= 1 hour 15 min., 30 min., 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, 7 days, 14 days

Y-axis: Limits

pH low 0.00 pH -2 to 20 pH

high 14.00 pH

Tempera-ture

low 0.0 ºC -30 to 140 ºC

high 100.0 ºC

ORP low -500 mV -1500 to 1500 mV

high 500 mV

rH low 0 rH 0 to 100 rH

high 50 rH

Calculate (Differential)

low -1pH, -500mV -20 to 20 pH, -1500 to 1500 mVhigh 1 pH, 500 mV

Calculate (Average)

low 0 pH, -500mV -20 to 20 pH, -1500 to 1500 mVhigh 14pH, 500mV

Concentra-tion

low 0 %, mg/L, g/L, ppm

0 to 100 %, 0.000 to 9999 mg/L, g/L, ppmhigh 100.0 %,

mg/L, g/L, ppm

Auto Return

Auto Return 10min Dissable, 10min, 60min

5.7.3

Adjust contrast

Contrast: 0 -5 to +5 5.7.4

Calculated data setup

Function Differential Differential, Average 5.8

Page 45: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix2UsersettingtableofSC> App.2-1

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Appendix2 UsersettingtableofSC* Reference section of User’s Manual, IM 12A01A02-01E

Parameter Default values Applicablevalues User

settings*Ref. sect.

Measure-ment setup

Measurement Conductivity Conductivity, Resistiv-ity, Concentration, Conduct.+Concentr.

10.1.1

Configuresensor

Sensor type 2-electrode 2-electrode, 4-electrode

10.1.2

Measuring unit /cm /cm, /m

Cell constant (factory) 0.1000 /cm10.00 /m

0.0050 to 50.00 /cm0.500 to 5000 /m

Tem-perature settings

Temp. element Pt1000 Pt1000, Pt100, Ni100, 8k55, Pb36(JIS6k)

10.1.3

Unit ºC ºC, ºF

Temp. compensa-tion

Compensation Automatic Manual, Automatic 10.1.4

Manual Manual temp. 25.0 ºC (77.0 ºF)

-20 to 140 ºC (-4.0 to 284.0 ºF)

Reference temp. 25.0 ºC (77.0 ºF)

0.0 to 100.0 ºC (32.0 to 212.0 ºF)

Method: Temp.comp.1 (Temp.comp.2)

NaCl(None)

None, TC, NaCl, Matrix

TC T.C.method1(2) 2.10 %/ºC(1.17 %/ºF)

0.00 to 10.0 %/ºC (0.00 to 5.56 %/ºF)

Matrix Select matrix Ammonia 0..50ppb

Ammonia 0..50ppb, Morpholine 0..500ppb, Ammonia 15..30%, Sulfuric acid 0..27%, Sulfuric acid 39..85%, Sulfuric acid 93..100%, NaOH 0..15%, NaOH 25..50%, HCl 0..200ppb, HCl 0..18%, HCl 24..44%, HNO3 0..25%, HNO3 35..80%,Userdefined1,Userdefined2

Definematrix: — —

Calibration settings

Limits: Air adjust 10.00 µS 0.000 to 20.00 µS 10.1.5

c.c. high 120.0 % 100.0 to 120.0 %

low 80.00 % 80.00 to 100.0 %

Timing: Step Range 1 % 0.1 to 20.0 %

Stabilization time 5 s 2 to 30 s

Calibr.interval 250 days 1 to 250 days

Concentra-tion

Additional table Disabled Disabled, Enabled 10.1.6

Unit for table % %, ppt, ppm, ppb

Sensor diag. settings

Process time: Disable Disable, Enable 10.1.7

BAD limit 2000 days 1 to 10000 days

Heat cycle: Disable Disable, Enable

BAD limit 500 cycles 10 to 1000 cycles

Defineheatcycle

Heat cycle tempe 50 ºC 1 to 250 ºC

Hear cycle time: 10.0 min. 0.1 to 60.0 min.

SC

Page 46: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix2UsersettingtableofSC> App.2-2

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Output setup

mA Output Output, Simulate 10.2

Output Process parameter Conduct1-TC1(Resist1-TC1)

Conduct1-TC1, Conduct1-TC2, Temperature1, Concent1-TC1, Concent1-TC2, Conduct2-TC1, Temperature2, Conduct2-TC2, Resist1-TC1, Resist1-TC2, Calculated, Redundant

Setup Linear Linear, Table

Linear 0 % valueConductivity

Resistivity

TemperatureConcentration

Calculated (Differential)

Calculated (Average)

Calculated (Ratio)Calculated (Passage)Calculated (Rejection)Calculated (Deviation)Calculated (pH

calc. (VGB))

0.000 nS/cm, µS/m

0.000mΩ•cm,mΩ•m

0.0 ºC0.000 %, ppm, ppb, 0 ppt-1.000 mS/cm, -100.0 mS/m, -20.00Ω•cm,-0.200Ω•m0.000 nS/cm, µS/m,mΩ•cm,mΩ•m

0.0000.000 %0.000 %0.000 %0.00 pH

0.000 nS/cm to 5.000 to 5.0005.000 S/cm, 0.000 µS/m to to 500.0 S/m0.000mΩ•cm to 1.000 to 1.0001.000 GΩ•cm,0.000mΩ•m to10.00 MΩ10.00MΩ•m-20 to 250 to 250250 ºC0.000 to 999.9 %, to 999.9 %,999.9 %, 0.000 to 9999 ppm, to 9999 ppm,9999 ppm, ppb, 0 to 9999 ppt to 9999 ppt9999 ppt-5.000 to 5.000 S/cm, to 5.000 S/cm,5.000 S/cm, -500.0 to 500.0 S/m, to 500.0 S/m,500.0 S/m, -1.000 to 1.000 Ωto1.000 Ω1.000Ω•cm, -10.00 to 10.00 MΩto10.00 MΩ10.00MΩ•m0 µS/cm to 5.000 S/cm, to 5.000 S/cm,5.000 S/cm, 0 µS/m to 500.0 S/m, to 500.0 S/m,500.0 S/m, 0.000mΩ•cm to 1.000 GΩto1.000 GΩ1.000GΩ•cm, 0Ω•m to 10.00 MΩto10.00 MΩ10.00MΩ•m0 to 20.0 to 20.020.00 to 200 % to 200 %200 %-200 to 200 % to 200 %200 %-200 to 200 % to 200 %200 %0 to14 pH to14 pH14 pH

100 % valueConductivity

Resistivity

TemperatureConcentration

Calculated (Differential)

Calculated (Average)

Calculated (Ratio)Calculated (Passage)Calculated (Rejection)Calculated (Deviation)Calculated (pH

calc. (VGB))

500.0 µS/cm, 50.00 mS/m20.00Ω•cm, 0.200Ω•m

100 ºC100.0 %, 9999 ppm, ppb, ppt1.000 mS/cm, 100.0 mS/m, 20.00Ω•cm, 0.200Ω•m500.0 µS/cm, 50.00 mS/m, 20.00Ω•cm, 0.200Ω•m20.00100.0 %100.0 %100.0 %14.00 pH

0 µS/cm to 5.000 S/cm, to 5.000 S/cm,5.000 S/cm, 0 µS/m to 500.0 S/m to 500.0 S/m500.0 S/m0.000mΩ•cm to 1.000 to 1.0001.000 GΩ•cm,0Ω•m to to 10.00MΩ•m-20 to 250 to 250250 ºC0.000 to 999.9 %, to 999.9 %,999.9 %, 0.000 to 9999 ppm, to 9999 ppm,9999 ppm, ppb, 0 to 9999 ppt to 9999 ppt9999 ppt-5.000 to 5.000 S/cm, to 5.000 S/cm,5.000 S/cm, -500.0 to 500.0 S/m, to 500.0 S/m,500.0 S/m, -1.000 to 1.000 Ωto1.000 Ω1.000Ω•cm, -10.00 to 10.00 MΩto10.00 MΩ10.00MΩ•m0 µS/cm to 5.000 S/cm, to 5.000 S/cm,5.000 S/cm, 0 µS/m to 500.0 S/m, to 500.0 S/m,500.0 S/m, 0.000mΩ•cm to 1.000 GΩto1.000 GΩ1.000GΩ•cm, 0Ω•m to 10.00 MΩto10.00 MΩ10.00MΩ•m0 to 20.0 to 20.020.00 to 200 % to 200 %200 %-200 to 200 % to 200 %200 %-200 to 200 % to 200 %200 %0 to 14 pH to 14 pH14 pH

Table — —

Burn Off Off, Low, High

Damping time 0.0 s 0.0 to 3600 s

Simulate Simulation perc. 50.00 % -2.5 to 112.5 %

ConfigureHold

Lastorfixed Last Last, Fixed

Fixed value mA 12.00 mA 3.60 to 22.00 mA

Hold during Calibration Enabled Disabled, Enabled

Page 47: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix2UsersettingtableofSC> App.2-3

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Error configura-tion

Conductivity too high (or Concentration) Warn Off, Warn, Fail 10.3

High limit 250.0 mS 0.00 nS to 500.0 S

Conductivity too low (or Concentration) Warn Off, Warn, Fail

Low limit 1.000 µS 0.000 nS to 500.0 S

Temperature too high Warn Off, Warn, Fail

Temperature too low Warn Off, Warn, Fail

Polarization detect Warn Off, Warn, Fail

Calibr.time exceeded Off Off, Warn, Fail

USP Off Off, Warn, Fail

USP safety margin 0.000 % 0.000 to 100.000 %

Configura-tion error:

1st comp. matrix Fail Off, Warn, Fail

2nd comp. matrix Fail Off, Warn, Fail

Cioncentration table Fail Off, Warn, Fail

Resisitivity too low Warn Off, Warn, Fail

Low limit 4.000Ω 0.000mΩto1.000GΩ

Resisitivity too high Warn Off, Warn, Fail

High limit 1.000 MΩ 0.000 mΩto1.000GΩ

Logbook configura-tion

Setting logbook

Sensor: Error on 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2) 10.4

Error off 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2)

Temp.coeff. Off Off, 1-1 (2-1), 1-2 (2-2)

Air adjust limit Off Off, 1-1 (2-1), 1-2 (2-2)

Adj c.c.low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Adj c.c.high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Reference temp. Off Off, 1-1 (2-1), 1-2 (2-2)

Stabilization time Off Off, 1-1 (2-1), 1-2 (2-2)

mA: mA setup Off Off, 1-1 (2-1), 1-2 (2-2)

Holdfixedvalue Off Off, 1-1 (2-1), 1-2 (2-2)

Burn Off Off, 1-1 (2-1), 1-2 (2-2)

Linear output % Off Off, 1-1 (2-1), 1-2 (2-2)

Damping time Off Off, 1-1 (2-1), 1-2 (2-2)

Erase logbook: 1-1 (2-1) 1-1 (2-1), 1-2 (2-2)

Erase? No No, Yes

Warn if logbook “full”: No No, Yes

SC

Page 48: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix2UsersettingtableofSC> App.2-4

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Advanced setup

Settings No action No action, Load factory settings, Save user settings, Load user settings

10.5.1

Tag FLXA21-SC (Max. 12 characters) 10.5.2

Passwords Execute: (blank) (Max. 8 characters) 10.5.3

Commissioning: (blank) (Max. 8 characters)

Date/Time Format YYYY/MM/DD

YYYY/MM/DD, MM/DD/YYYY, DD/MM/YYYY

10.5.4

Date — — —

Time — — —

Communication HART None, HART, PH201G 10.5.5

HART Network address 0 0 to 15

PV (Conduct1-TC1)(Resist1-TC1)

— —

SV Temperature1 Conduct1-TC1, Temperature1, Concent1-TC1, USP margin1, Conduct2-TC1, Temperature2, Conduct2-TC2, USP margin2, Resist1-TC1, Resist1-TC2, Calculated, Redundant, Empty

TV Empty, (Concent1-TC1)

FV Empty, (Temperature 2)

PH201G Hold contact: Enabled Disabled, Enabled

Fail contact: Fail+Warn Fail+Warn, Fail only, Disabled

Factory setup — — — 10.5.6

Display setup

Dual display

1st row SC1 SC1, SC2, Calculated. Redundant

10.6.1

2nd row SC2 SC1, SC2, Calculated. Redundant, Empty

Individual display

Main display

SC1 (2) Conductivity (Resistivity) unit Auto Auto, S/cm, mS/cm, µS/cm, nS/cmResistivity:Auto, mΩ•cm,Ω•cm,kΩ•cm,MΩ•cm,GΩ•cm

1st line Conduct1(2)-TC1, (Concent1-TC1, Resist1-TC1)

Conduct1(2)-TC1, Temperature1(2), Concent1(2)-TC1(2), Resist1-TC1(2)

2nd line Temperature1 (2)

Conduct1(2)-TC1, Temperature1(2), Concent1(2)-TC1(2), Resist1-TC1(2), Empty

3rd line Empty, (Concent1-TC1)

Additional text (blank) (Max. 12 characters)

Page 49: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix2UsersettingtableofSC> App.2-5

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Display setup

Trend Trend Graph Screen:

1st Trend Conduct1-TC1, (Concent1-TC1), (Resist1-TC1)

Conduct1(2)-TC1(2), Temperature1(2), Concent1(2)-TC1(2), Resist1-TC1(2), Calculated, Redun-dant, Empty

10.6.2

2nd Trend Temperature1

3rd Trend Empty, (Concent1-TC1)

X-axis: Timing

TIme Span (X)= 1 hour 15 min., 30 min., 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, 7 days, 14 days

Y-axis: Limits

Conduct*-TC*

low 0.000 nS/cm, µS/m

0 µS/cm to 5.000 S/cm, 0 µS/m to 500.0 S/mhigh 500.0 µS/cm,

50.00 mS/m

Tempera-ture

low 0.0 ºC -20 to 250 ºC

high 100.0 ºC

Resist*-TC*

low 0.000 mΩ•cm,mΩ•m

0.000mΩ•cm to 1.000 GΩ•cm,0Ω•m to 10.00MΩ•m

high 20.00 Ω•cm,0.200Ω•m

Concent*-TC*

low 0.000 %, ppm, ppb, 0 ppt

0.000 to 999.9 %, 0.000 to 9999 ppm, ppt, 0 to 9999 ppt

high 100.0 %, 9999 ppm, ppb, ppt

Calculated

Differ-ential

low -1.000 mS/cm, -100.0 mS/m, -20.00Ω•cm,-0.200Ω•m

-5.000 to 5.000 S/cm, -500.0 to 500.0 S/m, -1.000to1.000GΩ•cm, -10.00to10.00MΩ•m

high 1.000 mS/cm, 100.0 mS/m, 20.00Ω•cm,0.200Ω•m

Aver-age

low 0.000 nS/cm, µS/m, 0.000 mΩ•cm,mΩ•m

0 to 5.000 S/cm, 0 to 500.0 S/m, 0.000mΩ•cmto1.000GΩ•cm,0Ω•mto10.00MΩ•m

high 500.0 µS/cm, 50.00 mS/m, 20.00Ω•cm,0.200Ω•m

Ratio low 0.000 0 to 20.0

high 20.00

Pas-sage

low 0.000 % 0 to 200 %

high 100.0 %

Rejec-tion

low 0.000 % -200 to 200 %

high 100.0 %

Devia-tion

low 0.000 % -200 to 200 %

high 100.0 %

pH calc. (VGB)

low 0.00 pH 0 to 14.0 pH

high 14.00 pH

Auto Return

Auto Return 10min Dissable, 10min, 60min

10.6.3

Adjust contrast

Contrast: 0 -5 to +5 10.6.4

SC

Page 50: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix2UsersettingtableofSC> App.2-6

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Calculated data setup

Function Differential Differential, Average, Ratio, Passage(%), Rejection(%), Deviation(%), pH calc.(VGB)

10.7

Page 51: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix3UsersettingtableofISC> App.3-1

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Appendix3 UsersettingtableofISC* Reference section of User’s Manual, IM 12A01A02-01E

Parameter Default values Applicablevalues User

settings*Ref. sect.

Measure-ment setup

Measurement Conductivity Conductivity, Concentration, Conduct.+Concentr.

13.1.1

Configuresensor

Measuring unit /cm /cm, /m 13.1.2

Cell constant (factory) 1.880 /cm188.0 /m

0.2000 to 50.00 /cm20.00 to 5000 /m

Tem-perature settings

Temp. element Pt1000 Pt1000, NTC30k 13.1.3

Unit ºC ºC, ºF

Temp. compensa-tion

Compensation Automatic Manual, Automatic 13.1.4

Manual Manual temp. 25.0 ºC (77.0 ºF)

-20 to 140 ºC (-4.0 to 284.0 ºF)

Reference temp. 25.0 ºC (77.0 ºF)

0.0 to 100.0 ºC (32.0 to 212.0 ºF)

Method: Temp.comp.1 (Temp.comp.2)

NaCl(None)

None, TC, NaCl, Matrix

TC T.C.method1(2) 2.10 %/ºC(1.17 %/ºF)

0.00 to -10.0 %/ºC (0.00 to 5.56 %/ºF)

Matrix Select matrix Sulfuric acid 1..5%

Sulfuric acid 1..5%, Sulfuric acid 0..27%, Sulfuric acid 39..85%, Sulfuric acid 93..100%, HCl 1..5%, HCl 0..18%, HCl 24..44%, HNO3 1..5%, HNO3 0..25%, HNO3 35..80%, NaOH 1..5%, NaOH 0..15%, NaOH 25..50%,Userdefined1,Userdefined2

Definematrix: — —

Calibration settings

Limits: Air adjust 100.0 µS 0.000 to 100.0 µS 13.1.5

c.c. high 120.0 % 100.0 to 120.0 %

low 80.00 % 80.00 to 100.0 %

Timing: Step Range 1 % 0.1 to 20.0 %

Stabilization time 5 s 2 to 30 s

Calib.interval 250 days 1 to 250 days

Concentra-tion

Additional table Disabled Disabled, Enabled 13.1.6

Unit for table % %, ppt, ppm, ppb

Sensor diag. settings

Process time: Disable Disable, Enable 13.1.7

BAD limit 2000 days 1 to 10000 days

Heat cycle: Disable Disable, Enable

BAD limit 500 cycles 10 to 1000 cycles

Defineheatcycle

Heat cycle tempe 50 ºC 1 to 250 ºC

Hear cycle time: 10.0 min. 0.1 to 60.0 min.

ISC

Page 52: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix3UsersettingtableofISC> App.3-2

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Output setup

mA Output Output, Simulate 13.2

Output Process parameter Conduct1-TC1, (Concent1-TC1)

Conduct1-TC1, Temperature1, Conduct1-TC2, Concent1-TC1, Concent1-TC2

Setup Linear Linear, Table

Linear 0 % valueConductivityTemperature

Concentration

0.000 µS/cm, mS/m0.0 ºC0.000 %, ppm, ppb, 0 ppt

0.000 to 5.000 S/cm, 0.000 to500.0 S/mto500.0 S/m500.0 S/m-20 to 140 to 140140 ºC0.000 to 999.9 %, to 999.9 %,999.9 %, 0.000 to 9999 ppm, to 9999 ppm,9999 ppm, ppb, 0 to 9999 ppt to 9999 ppt9999 ppt

100 % valueConductivityTemperature

Concentration

1.000 S/cm, 100.0 S/m100 ºC100.0 %, 9999 ppm, ppb, ppt

0 to 5.000 S/cm, 0 to to 5.000 S/cm, 0 to5.000 S/cm, 0 to to 500.0 S/m-20 to 140 to 140140 ºC0.000 to 999.9 %, to 999.9 %,999.9 %, 0.000 to 9999 ppm, to 9999 ppm,9999 ppm, ppb, 0 to 9999 ppt to 9999 ppt9999 ppt

Table — —

Burn Off Off, Low, High

Damping time 0.0 s 0.0 to 3600 s

Simulate Simulation perc. 50.00 % -2.5 to 112.5 %

ConfigureHold

Lastorfixed Last Last, Fixed

Fixed value mA 12.00 mA 3.60 to 22.00 mA

Hold during Calibration Enabled Disabled, Enabled

Error configura-tion

Conductivity too high (or Concentration) Warn Off, Warn, Fail 13.3

High limit 1.000 S 0.00 nS to 500.0 S

Conductivity too low (or Concentration) Warn Off, Warn, Fail

Low limit 5.000 µS 0.000 nS to 500.0 S

Temperature too high Warn Off, Warn, Fail

Temperature too low Warn Off, Warn, Fail

Calibr.time exceeded Off Off, Warn, Fail

Configura-tion error:

1st comp. matrix Fail Off, Warn, Fail

2nd comp. matrix Fail Off, Warn, Fail

Concentration table Fail Off, Warn, Fail

Logbook configura-tion

Settings logbook

Sensor: Error on 1-1 Off, 1-1, 1-2 13.4

Error off 1-1 Off, 1-1, 1-2

Temp.coeff. Off Off, 1-1, 1-2

Air adjust limit Off Off, 1-1, 1-2

Adj c.c.low limit Off Off, 1-1, 1-2

Adj c.c.high limit Off Off, 1-1, 1-2

Reference temp. Off Off, 1-1, 1-2

Stabilization time Off Off, 1-1, 1-2

mA: mA setup Off Off, 1-1, 1-2

HoldfixedValue Off Off, 1-1, 1-2

Burn Off Off, 1-1, 1-2

Linear output % Off Off, 1-1, 1-2

Damping time Off Off, 1-1, 1-2

Erase logbook: 1-1 1-1, 1-2

Erase? No No, Yes

Warn if logbook “full”: No No, Yes

Page 53: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix3UsersettingtableofISC> App.3-3

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Advanced setup

Settings No action No action, Load factory settings, Save user settings, Load user settings

13.5.1

Tag FLXA21-ISC (Max. 12 characters) 13.5.2

Passwords Execute: (blank) (Max. 8 characters) 13.5.3

Commissioning: (blank) (Max. 8 characters)

Date/Time Format YYYY/MM/DD

YYYY/MM/DD, MM/DD/YYYY, DD/MM/YYYY

13.5.4

Date — — —

Time — — —

Communication HART None, HART, PH201G 13.5.5

HART Network address 0 0 to 15

PV (Conduct1-TC1), (Concent1-TC1)

— —

SV Temperature1 Conduct1-TC1, Temperature1, Concent1-TC1, Conduct1-TC2, Empty

TV Empty, (Concent1-TC1)

FV Empty

PH201G Hold contact: Enabled Disabled, Enabled

Fail contact: Fail+Warn Fail+Warn, Fail only, Disabled

Factory setup — — — 13.5.6

ISC

Page 54: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix3UsersettingtableofISC> App.3-4

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Display setup

Main display

Conductivity unit Auto Auto, S/cm, mS/cm, µS/cm

13.6.1

1st line Conduct1-TC1(Concent1-TC1)

Conduct1-TC1, Temperature1, Concent1-TC1, Conduct1-TC2

2nd line Temperature1 Conduct1-TC1, Temperature1, Conduct1-TC2, Concent1-TC1, Empty

3rd line Empty, (Concent1-TC1)

Additional text (blank) (Max. 12 characters)

Trend Trend Graph Screen:

1st Trend Conduct1-TC1(Concent1-TC1)

Conduct1-TC1, Temperature1, Concent1-TC1, Conduct1-TC2, Empty

13.6.2

2nd Trend Temperature1

3rd Trend Empty, (Concent1-TC1)

X-axis: Timing

TIme Span (X)= 1 hour 15 min., 30 min., 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, 7 days, 14 days

Y-axis: Limits

Conduct1-TC1(2)

low 0.000 µS/cm, mS/m

0 to 5.000 S/cm, 0 to 500.0 S/m

high 1.000 S/cm, 100.0 S/m

Tempera-ture1

low 0.0 ºC -20 to 140 ºC

high 100.0 ºC

Concent1-TC1(2)

low 0.000 %, ppm, ppb, 0 ppt

0.000 to 999.9 %, 0.000 to 9999 ppm, ppt, 0 to 9999 ppthigh 100.0 %,

9999 ppm, ppb, ppt

Auto Return

Auto Return 10min Dissable, 10min, 60min

13.6.3

Adjust contrast

Contrast: 0 -5 to +5 13.6.4

Page 55: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix4UsersettingtableofDO> App.4-1

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Appendix4 UsersettingtableofDO* Reference section of User’s Manual, IM 12A01A02-01E

Parameter Default values Applicablevalues User

settings*Ref. sect.

Sensor setup

Sensor type Galvanic Galvanic, Polaro-graphic

16.1

Measure-ment setup

Configuresensor

Unit mg/L mg/L, ppm, ppb, %SAT

16.2.1

Sensor sensitivity Gal: 0.45 µA/ppm(50 µm)

Gal: 0.45 µA/ppm(50 µm), 0.90 µA/ppm(25 µm), Others (0.10 to 20.00 µA/ppm)

Pol: 7.5 nA/ppm

Pol: 1.0 to 200.0 nA/ppm

Polarization Voltage 0.675 V 0.000 to 1.000 V

Tem-perature settings

Temp. element Pt1000 Pt1000, NTC22k 16.2.2

Unit ºC ºC, ºF

Temp. compensa-tion

Compensation Automatic Manual, Automatic 16.2.3

Manual Manual temp. 25.0 ºC(77.0 ºF)

-20 to 150 ºC

Salinity compensa-tion

Compensation Disabled Disabled, Enabled 16.2.4

Input Salinity 0.0 g/kg 0.0 to 100.0 g/kg

Pressure comp. (Measure)

Input Pressure Level (Process) 101.3 kPa 0.0 to 999.9 kPa 16.2.5

Calibration settings

Limits and timing

Zero Low Gal: -0.50 µAPol: -5.0 nA

Gal: -20.00 to 0 µAPol: -200.0 to 0 nA

16.2.6

High Gal: 0.50 µAPol: 5.0 nA

Gal 0 to 20.00 µAPol: 0 to 200.0 nA

Slope Low 40.00 % 0.1 to 99.9 %

High 200.0 % 100.0 to 600.0 %

Stabilization time 60 s 5 to 600 s

Step Range

Zero 0.10mg/L, ppm1.0 ppb, %SAT

ppm, mg/L: 0.00 to 50.00ppb: 0 to 2000%SAT: 0.0 to 600.0Span 0.10mg/L, ppm

1.0 ppb, %SAT

Cal.Interval 250 days 1 to 250 days

zero/slope Zero Current Gal: 0.00 µAPol: 0.0 nA

Gal: -20.00 to 20.00 µAPol: -200.0 to 200.0 nA

Slope 100.0 % 0.1 to 600.0 %

Pressure Comp. (Cal.)

Input Pressure Level (Cal.) 101.3 kPa 0.0 to 999.9 kPa

Zero Calibration Disabled Disabled, Enabled

Sensor diag. settings

Progress time: Disable Disable, Enable 16.2.7

BAD limit 2000 days 1 to 10000 days

Heat cycle: Disable Disable, Enable

BAD limit 500 cycles 10 to 1000 cycles

Defineheatcycle

Heat cycle tempe 50 ºC 1 to 250 ºC

Hear cycle time: 10.0 min. 0.1 to 60.0 min.

DO

Page 56: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix4UsersettingtableofDO> App.4-2

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Output setup

mA Output Output, Simulate 16.3

Output Process parameter Oxygen1 Oxygen1, Tem-perature1, Oxygen2, Temperature2, Calculated, Redun-dant

Setup Linear Linear, Table

Linear 0 % value: Oxygen

Temperature

0.00 mg/L, ppm, 0.0 ppb, %SAT0.0 ºC

mg/L, ppm: 0.00 to 50.00ppb: 0.0 to 9999%SAT: 0.0 to 600.0ºC: -20.0 to 150.0100 % value:

Oxygen

Temperature

20.00 mg/L, ppm100.0 ppb, %SAT50.0 ºC

Table — —

Burn Off Off, Low, High

Damping time 0.0 s 0.0 to 3600 s

Simulate Simulation perc. 50.00 % -2.5 to 112.5 %

ConfigureHold

Lastorfixed Last Last, Fixed

Fixed value mA 12.00 mA 3.60 to 22.00 mA

Hold during Calibration(/Wash) Enabled Disabled, Enabled

Error configura-tion

DO too high Warn Off, Warn, Fail 16.4

DO too low Warn Off, Warn, Fail

Temperature too high Warn Off, Warn, Fail

Temperature too low Warn Off, Warn, Fail

Sensor membarane Off Off, Warn, Fail

Calib. time exceeded Off Off, Warn, Fail

Logbook configura-tion

Settings logbook

Sensor: Error on 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2) 16.5

Error off 1-1 (2-1) Off, 1-1 (2-1), 1-2 (2-2)

Zero low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Zero high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Slope low limit Off Off, 1-1 (2-1), 1-2 (2-2)

Slope high limit Off Off, 1-1 (2-1), 1-2 (2-2)

Stabilization time Off Off, 1-1 (2-1), 1-2 (2-2)

Pressure(meas.) Off Off, 1-1 (2-1), 1-2 (2-2)

Salinity comp. Off Off, 1-1 (2-1), 1-2 (2-2)

Wash setup Off Off, 1-1 (2-1), 1-2 (2-2)

mA: mA setup Off Off, 1-1 (2-1), 1-2 (2-2)

HoldfixedValue Off Off, 1-1 (2-1), 1-2 (2-2)

Burn Off Off, 1-1 (2-1), 1-2 (2-2)

Linear output % Off Off, 1-1 (2-1), 1-2 (2-2)

Damping time Off Off, 1-1 (2-1), 1-2 (2-2)

Erase logbook: 1-1 (2-1) 1-1 (2-1), 1-2 (2-2)

Erase? No No, Yes

Warn if logbook “full”: No No, Yes

Page 57: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix4UsersettingtableofDO> App.4-3

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Advanced setup

Settings No action No action, Load factory settings, Save user settings, Load user settings

16.6.1

Tag FLXA21-DO (Max. 12 characters) 16.6.2

Passwords Execute: (blank) (Max. 8 characters) 16.6.3

Commissioning: (blank) (Max. 8 characters)

Date/Time Format YYYY/MM/DD

YYYY/MM/DD, MM/DD/YYYY, DD/MM/YYYY

16.6.4

Date — — —

Time — — —

Communication HART None, HART, PH201G 16.6.5

HART Network address 0 0 to 15

PV (Oxygen1) — —

SV Temperature1 Oxygen1 (2), Temperature1 (2), Calculated, Redun-dant, Empty

TV Empty, (Oxygen2)

FV Empty, (Tempera-ture2)

PH201G Hold contact: Enabled Disabled, Enabled

Fail contact: Fail+Warn Fail+Warn, Fail only, Disabled

Wash contact: Enabled Disabled, Enabled

Interval time 0.6 h. 0.1 to 36.0 h.

Wash (Meas-ure) time

0.5 min. 0.1 to 10.0 min.

Recovery time 0.5 min. 0.1 to 10.0 min.

Manual Wash Enabled Disabled, Enabled

Continuous Wash

Disabled Disabled, Enabled

Factory setup — — — 16.6.6

DO

Page 58: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

<Appendix4UsersettingtableofDO> App.4-4

IM 12A01A02-12E 2nd Edition : Aug. 18, 2010-00

Parameter Default values Applicablevalues User

settings*Ref. sect.

Display setup

Dual display

1st row DO1 DO1, DO2, Calcu-lated. Redundant

16.7.1

2nd row DO2 DO1, DO2, Calcu-lated. Redundant, Empty

Individual display

Main display

DO1 (2) 1st line Oxygen1 (2) Oxygen1 (2), Temperature1 (2)

2nd line Temperature1 (2) Oxygen1 (2), Temperature1 (2), Empty3rd line Empty

Additional text (blank) (Max. 12 characters)

Trend Trend Graph Screen:

1st Trend Oxygen1 (2) Oxygen1 (2), Temperature1 (2), (Calculated, Redun-dant), Empty

16.7.2

2nd Trend Temperature1 (2)

3rd Trend Empty

X-axis: Timing

Time Span (X)= 1 hour 15 min., 30 min., 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, 7 days, 14 days

Y-axis: Limits

Oxygen low 0.00 mg/L, ppm0.0 ppb, %SAT

mg/L, ppm: 0.00 to 50.00ppb: 0.0 to 9999%SAT: 0.0 to 600.0high 20.00 mg/L, ppm

100.0 ppb, %SAT

Tempera-ture

low 0.0 ºC -20.0 to 150.0 ºC

high 50.0 ºC

Auto Return

Auto Return 10min Dissable, 10min, 60min

16.7.3

Adjust contrast

Contrast: 0 -5 to +5 16.7.4

Calculated data setup

Function Differential Differential, Average 16.8

Page 59: Model FLXA21 Two-wire Analyzer Start-up Manual• If you have any questions, or you find mistakes or omissions in the user’s manuals, please contact our sales representative or your

i

IM 12A01A02-12E

Revision Recordl Manual Title : Model FLXA21 Two-wire Analyzer Start-up Manuall Manual No. : IM 12A01A02-12E

May2010/1stEditionNewly published

Aug.2010/2ndEditionFollowings are mainly revised; Additionofgroundingterminalpositiononstainlesshousingwithspecificmountings Additionofplatepositiononstainlesshousingwithspecificcoatings Addition of explanation of sleeve for grounding wire for plastic housing Correction of torques Addition of Note, Warning etc. Addition of detail descriptions for wire terminals Addition of drawings of housing with hood Addition of example displays for calculated data and redundant system Addition of explanation of passwords Correction of errors on the User setting tables

2nd Edition : Aug. 18, 2010-00