MRI Conditional Pacemaker

核磁共振相容節律器

Sep 07, 2013

振興醫院 張鴻猷醫師

2013 Taiwan Heart Rhythm Society –

Allied Professional Education Program

Page 2: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MR Conditional Labeling*

MR Safe

MR Conditional

MR Unsafe

*ASTM standard F2503: Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment

An item that has been demonstrated to pose no known hazards in a specified MRI environment with specified conditions of use.

Page 3: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MRI 基本步驟

•1. 把人放進去MRI裡

•2. 給能量去激發人體內的氫原子

•3. 把能量關掉

•4. 接收氫原子激發後回傳的能量波

•5. 轉換此能量波訊號成為影像

Page 4: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

設備圖

Magnet Gradient Coil RF Coil

RF Coil

4T magnet

gradient coil

(inside)

B0

Page 5: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Main Components of a MRI

Scanner

• Static Magnetic Field Coils

• Gradient Magnetic Field Coils

• Magnetic shim coils

• Radiofrequency Coil

• Control computer

1.5 or 3.0 telsa

1 telsa (T) = 10000 gauss

地球磁場 = 0.5 gauss

Page 6: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Transmit Receive

RF

coil

RF

coil

Main

magnet Main

magnet

Gradient Shimming

Control

Computer

Page 7: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

質子 (Proton)

•帶正電

•會自旋

Page 8: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

The Effect of Irradiation to the Spin System

Lower

Higher

Spin System After Irradiation

Page 9: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MRI 原理

•利用磁場原理,把人體置於強大且均勻的靜磁場 (static magnet) 中,再利用特定的射頻無線電波脈衝,激發人體組織內的氫原子核

•由於人體內的水分子和脂肪都含有氫原子核,這些氫原子核本身又具有磁場特性,如同一個小小的磁鐵

•若使儀器改變體內氫原子核的旋轉排列方向,原子核就會釋放吸收的能量,能量激發後放出電磁波信號,再經由電腦分析組合成影像,這就是一般所看到的 MRI 影像

•由於不同的組織受到刺激後,釋出不同的回波,因此在影像上便會產生非常良好的對比

Page 10: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MRI 優點

•非侵入性、無傷害的診斷系統,沒有使用 X 光及放射線物質

•安全、無痛苦,能呈現具體的多方向切面影像

•可以很精細的提供腦部、脊椎、軟組織的狀況

•所測量的不僅是組織的解剖影像,更重要的是組織的各種功能性影像,具備對各種組織器官功能性與新陳代謝性障礙的診斷能力

•檢查前病患不需要特別做什麼準備,例如在檢查前不需要空腹、可以進食

Page 11: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MRI 缺點

•設備費用昂貴,檢查費用較高

•掃描時間需時較長

•對肺部的檢查不優於X射線或CT檢查

•對身體移動非常敏感,若移動易產生偽影,故不適合急診病患及危急病人,也容易引發禁閉恐懼症

•心律調節器、金屬性人工心臟瓣膜、各類電子傳導器、腦動脈瘤手術夾等裝置會受到磁場干擾,患者不適合 MRI 檢查

•無法鑑別鈣化點

Page 12: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

裝置節律器的病人常需要MRI檢查

It is estimated that 50 to 75% of pacemaker patients will have a medical need for an MRI over the lifetime of their device

Average Age of Pacemaker Patient

13% Ages 25-64

86% Age 65+

The Prevalence of Common

Comorbidities Increases Rapidly

Over Age 65

Page 13: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

慢性病多、需要多專科同時照護

Number of Comorbidities in

Pacemaker Patients

85% of all pacemaker patients have one or more comorbidities

15% 0 comorbidities

30% 1 comorbidity

28% 2 comorbidities

27% 3 or more

comorbidities

Opthalmology

Otolaryngology

Neurosurgery

Neurology

Cardiothoracic Surgery

Surgical Oncology

Nephrology

Gynecologic Oncology

Urology

Rheumatology

Oncology

Radiation Oncology

Interventional Radiology

Gastrointestinal Surgery

Orthopedic Surgery

Vascular Surgery

85%1 1 or more

comorbidities

1 Kalin R, Stanton MS. PACE. 2005;28:326-328.

Page 14: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MRI: Three Powerful Fields

Static Field

Gradient Field

RF Field

MP – 14

Page 15: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

背景磁力永遠存在

Page 16: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Static Gradient RF

Device Interactions

Stimulation (UCS)

Lead Heating

UCS = unintended cardiac stimulation - Inhibitor pacemaker function

- Trigger rapid pacing

- Deliver inappropriate shocks

Page 17: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Lead Heating Clinical Impact

• PCT = Pacing Capture

Threshold

• PCT is lowest at implant

• Healing produces scar

• Increased distance

increases PCT

• Significant heating causes tissue damage

• Increased scar volume increases PCT

Page 18: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Stimulation hazard gradient-induced high

rate pacing

Start of Scan

EKG

Pulse Ox

Canine Test

Unintended Stimulation Clinical

Impact

The MRI scanner is pacing the heart

Page 19: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Device Interactions Clinical

Impact

Failures can be temporary or permanent

MP – 53

Page 20: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Pacemaker System Design

Solutions

Page 21: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Device Design Solutions

• Minimize ferromagnetic content

Unintended cardiac stimulation Device interactions Force and Torque

• Hybrid-case connection

• Hall sensor

• Optimize input circuitry

Page 22: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Lead Heating Design Solution • Lead inner conductor coil design mitigates lead heating

• 4 filar to 2 filar increases inductance and reduces

heating

• 5086 provides 3:1 reduction in lead tip heating

0

10

20

30

40

50

60

70

80

90

100

0 5 10 15 20 25 30 35 40 45 50

Perc

en

t o

f M

axim

um

Lead Path

5076 vs. 5086MRI Comparison

5076 - 52

5086 - 52

Page 23: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Clinical Confirmation of

Safety

Page 24: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Clinical Study Experience (EnRhythm MRI)

• Prospective, randomized, multi-center – Implantation: 464 subjects

– 41 international centers • 21 Europe, 13 U.S., 6 Canada, 1 Middle East

• Enrollment: – February 2007 through July 2008

• Primary End Points: – Consistent Capture Thresholds

– Consistent Sensed Amplitudes

Page 25: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

MRI Group

MRI Scans

9-12 Week Visit

Post-MRI

Pre-MRI

Control Group

No MRI

1-Week Post-MRI

Visit

1-Month Post-MRI

Visit

Enrollment

2 Month

Visit

Implant

and Randomization

Clinical Visit Schedule

6-Month

Post-Implant Visit

and Every Six

Months Thereafter

Page 26: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Primary Effectiveness Endpoint

No measurable difference between MRI and control groups

Atrial & Ventricular Capture Threshold: Pre-visit to 1-month post-visit

0

20

40

60

80

-1 -0.5 0 0.5 1

Change in Atrial

Threshold (V)

% o

f P

ati

en

ts

MRI

(n=165)

Control

(n=164)

0

20

40

60

80

-1 -0.5 0 0.5 1

Change in Ventricular

Threshold (V)%

of

Pati

en

ts

MRI

(n=190)

Control

(n=184)

Page 27: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Clinical Confirmed Safety and

Effectiveness

• Safety demonstrated – No MRI-related complications

– No sustained ventricular arrhythmias or

asystole

• Effectiveness demonstrated – Consistent capture thresholds

– Consistent sensed amplitudes

Page 28: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

• Complete system: MRI compatible

pacemaker with two MRI compatible leads

• Implanted for at least 6 weeks

• No abandoned leads

• Pacing capture threshold ≤ 2.0 V at 0.4 ms

• Lead impedance of 200 Ω to 1500 Ω

• SureScan programming

US Conditions of Use – Cardiology

Page 29: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

• 1.5 Tesla Scanner

• Gradient field:

– Maximum Gradient Slew Rate ≤ 200 T/m/s (per axis)

• RF field: Normal Operating Mode

– Whole body SAR ≤ 2.0 W/kg

– Head SAR < 3.2 W/kg

• Patient monitoring

US Conditions of Use – Radiology

Page 30: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Patient Care Pathway

1. Pre-Screening and Scheduling

2. Pre-scan

3. Scan

4. Post-Scan

Page 31: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Pre-screening

• Ways to ID

pacemaker system:

1. Patient ID Card

2. X-Ray

3. Call cardiology

1

PRESCREENING

2 PRE-SCAN

3 SCAN

4 POST SCAN

Page 32: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

X-Ray

Radiopaque identifies the system via unique radiopaque visible under x-ray

1 PRE-

SCREENING

2 PRE-SCAN

3 SCAN

4 POST SCAN

1 Location of the device radiopaque symbol 2 Device radiopaque MRI symbol 3 Lead radiopaque MRI symbol

Page 33: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

Programmer print out

Programmer

Printout

Provides

confirmation that

the system is MR-

Conditional, model

number and

checklist for

relevant conditions

1 PRE-

SCREENING

2 PRE-SCAN

3 SCAN

4 POST SCAN

Page 34: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

SureScan Programming Steps

1 PRE-

SCREENING

2 PRE-SCAN

3 SCAN

4 POST SCAN

Page 35: New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130907北區

During the Scan – Monitoring

• Patient must be monitored during the MRI

scan:

– Visual and verbal contact

– Pulse oximetry

– Blood pressure

– Electrocardiography (rate only)

• There is no requirement for the monitoring

person to be from cardiology.

1 PRE-

SCREENING

2 PRE-SCAN

3 SCAN

4 POST SCAN