KROMATOGRAFI• Kromatografi adalah suatu cara pemisahan senyawa dari campurannya melalui elusi
sepanjang fasa lain yang tak bergerak.
• Eluen dapat berupa fasa gas atau cair yang selanjutnya disebut fasa gerak (mobile phase), sedangkan fasa tak bergerak atau fasa stasioner (fasionery phase) dapat berupa padatan atau cairan yang diselaputkan pada permukaan padatan pendukung (supporting solid) yang innert.
• Chromatography involves a sample (or sample extract) being dissolved in a mobile phase (which may be a gas, a liquid or a supercritical fluid).
• The mobile phase is then forced through an immobile, immiscible stationary phase. The phases are chosen such that components of the sample have differing solubilities in each phase.
• A component which is quite soluble in the stationary phase will take longer to travel through it than a component which is not very soluble in the stationary phase but very soluble in the mobile phase. As a result of these differences in mobilities, sample components will become separated from each other as they travel through the stationary phase.
![Page 3: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/3.jpg)
SEJARAH KROMATOGRAFI
• 1834 Runge F F :pemisahan campuran zat warna dan ekstrak tanaman
• 1906 Tswett M. : pemisahan pigmen kloropil pada CaCO3 dengan eluen
petroleum eter → kromatografi
• 1931 Kuhn, R et al. : kromatografi cair-padat, pemisahan xanthophyl kuning telur
• 1941 Tiselius A.: kromatografi cair, frontal analisis
• 1941 Martin & Synge : model efisiensi kolom
• 1944 Consden, Gordon, Martin : kromatografi kertas
• 1947 Boyd et al.: kromatografi penukar ion
• 1951 Kirchner : Kromatografi lapis tipis (TLC)
• 1952 James & Martin : Kromatografi gas
• 1956 van Dempter : fungsi distribusi gausian
• 1964 J C. Moore : kromatografi gel permeation
![Page 4: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/4.jpg)
Klasifikasi kromatografi
Fasa gerak Fasa diam Mekanisme Istilah/nama
Gas padat adsorbsi GSC
Cair partisi GLC
Cair cair partisi LLC
padat adsorbsi LSC
Penukar ion Penukar ion IEC
Gel Exclusion EC
Cairan superkritik
Spesi organik pada padatan
partisi SFC
![Page 5: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/5.jpg)
Kromatografi
Kromatografigas
Kromatografi cair
Gas - Cair Gas - padat Liqiud-liquid Lquid-solid Ion-exchange Exclusion
![Page 6: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/6.jpg)
Klasifikasi : peralatan
• Kromatografi kolom
• Kromatografi lapis tipis (TLC = Thin Layer Chromatography)
• Kromatografi kertas (paper chromatography)
• Kromatografi gas (GC= gas chromatography)
• Kromatografi cair kinerja tinggi (HPLC = high performance liquid chromatography)
![Page 7: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/7.jpg)
KROMATOGRAFI KOLOM
![Page 8: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/8.jpg)
Kromatografi Kolom
![Page 9: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/9.jpg)
Kesetimbangan distribusi
Fasa gerak
Fasa diam
K = -----------
Cd
CgK = koefisien distribusiCg = Konsentrasi dalam fasa gerakCd = konsentrasi zat dalam fasa diam
Fraksi waktu tinggal dalam fasa gerak =------------------------------------------------- Jumlah molekul dalam fasa gerak
Total molekul yang ada
= ------------------------Cg Vg
Cg Vg + Cd Vd
= ------------------------ 1
1 + K Vd/Vg
= ----------- 1
1 + k’ k’ = faktor kapasitas
Vg =volume fasa gerak
![Page 10: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/10.jpg)
Laju pergerakan
Fraksi = ----------- 1
1 + k’
Laju linier =
Laju pergerakan = ----------- 1
1 + k’
Laju pergerakan rata-rata ,molekul ditentukan oleh:
1.Laju fasa gerak2.Rasio volume fasa diam terhadap volume fasa gerak3.Koefisien distribusi: khas untuk setiap komponen
![Page 11: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/11.jpg)
Waktu retensi, tR
Waktu yang diperlukan oleh komponen menempuh perjalanan sepanjang kolom
tR = ------------------ = -------- (1+k’) panjang lintasan
Laju pergerakan
L
= tM (1+k’)
tM = waktu yang diperlukan oleh molekul fasa gerak sepanjang lintasan/kolom
![Page 12: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/12.jpg)
Volume retensi
Volume fasa gerak yang diperlukan untuk mengelusi komponen sampel
Vol = waktu x laju alir
VR = tR x F
VR = VM (1+k’) = VM + KVd
Retensi relatif:
![Page 13: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/13.jpg)
Kromatografi lapis tipis TLC
Thin-layer chromatography (TLC) is a chromatographic technique that is useful for separating organic compounds. Because of the simplicity and rapidity of TLC, it is often used to monitor the progress of organic reactions and to check the purity of products.
![Page 14: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/14.jpg)
Teknik Elusi KLT/KK
Elusi Ascending : dari bawah ke atasElusi Discending : dari atas ke bawah
Elusi dua dimensi
![Page 15: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/15.jpg)
Gambar teknik elusi
![Page 16: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/16.jpg)
Elusi dua dimensi
Two-dimensional TLC uses the TLC method twice to separate spots that are unresolved by only one solvent. After running a sample in one solvent, the TLC plate is removed, dried, rotated 90o, and run in another solvent. Any of the spots from the first run that contain mixtures can now be separated. The finished chromatogram is a two-dimensional array of spots.
![Page 17: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/17.jpg)
Identifikasi spot hasil elusi
• Spot berwarna• Spot tak berwarna
Spot tak berwarna:
- Diuapi dengan amoniak
- Diuapi dengan iodium (I2)
- Disemprot dengan larutan pengembang (pewarna)
- Disemprot dengan asam sulfat pekat
![Page 18: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/18.jpg)
Nilai Rf
![Page 19: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/19.jpg)
Aplikasi pemisahan TLC
No Senyawa yang dipisahkan
Solven (Fasa gerak) Sorben (fasa diam) Reagent deteksi
1. Kloroplast pigmens
Isooktan-aceton-eter (3:1:1) -Silika gel-Polyamida
-
Petroleum eter-aseton (4:6) Alumina
2. Alkaloid Bensen-etanol (9:1) Silika gel Bromocresol greenDragendorffIodineCHCl3-aseton-dietilamin (5:4:1) Silika gel
3. Amina Etanol 90% - NH3 25% (4:1) Silika gel AsetoasetilfenolAlizarin
Aseton-H2O (99:1) Kielsguhr G
4. Gula Benzena-as.asetat-metanol (1:1:3)
Silika gel dg bufer as. borat
Butanol-piridin-H2O (6:4:3) Selulosa
5. Food Dyes Metil etil keton-as asetat-metanol (40:5:5)
Silika gel G
Butanol-etanol-H2O (9:1:1) Alumina
![Page 20: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/20.jpg)
No Senyawa yang dipisahkan Solven (Fasa gerak) Sorben (fasa diam)
6. Tserpenoid Heksana-etil asetat (85:15) -pati-asam asalisilat
Benzena petroleum eter Alumina
7. Vitamin Metanol, CCl4, atau petroleum eter Alumina
Metanol, propanol atau CHCl3 Silika gel G
8. Fenol Benzena, atau dietil eter alumina
Heksana-etilasetat (4:1 atau 3:2) Silkika gel + as oksalat
9. Asam amino Butanol-asamasetat-H2O (4:1:1) Silika gel G
N-butanol-aseton-NH3-H2O (10:10:5:2) Selulosa
Iso propanol-asam format-H2O (20:1:5) selulosa
10. Flavonoid dan coumarin
Metanol-H2O (8:2 atau 6:4) Polyamida
Petroleum eter – etil asetat (2:1) Silika gel G
11. Steroid dan sterol Benzene-etil asetat (9:1) Silika gel G
CHCl3 – etanol (96:4) alumina
Zweig G & Joseph S, 1972, Handbook of Chromatography, Vol II. CRC Press USA
![Page 21: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/21.jpg)
Pereaksi pengembang (detection Reagent)
![Page 22: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/22.jpg)
![Page 23: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/23.jpg)
Aplikasi pemisahan
![Page 24: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/24.jpg)
Kromatografi Kertas
![Page 25: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/25.jpg)
Aplikasi Pemisahan
![Page 26: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/26.jpg)
Kromatografi Gas
![Page 27: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/27.jpg)
Gas Chomatography
GC
![Page 28: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/28.jpg)
![Page 29: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/29.jpg)
![Page 30: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/30.jpg)
![Page 31: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/31.jpg)
PRINSIP GAS-KROMATOGRAFI
• Gas chromatography - specifically gas-liquid chromatography - involves a sample being vapourised and injected onto the head of the chromatographic column. The sample is transported through the column by the flow of inert, gaseous mobile phase. The column itself contains a liquid stationary phase which is adsorbed onto the surface of an inert solid.
![Page 32: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/32.jpg)
Ekstraksi
![Page 33: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/33.jpg)
![Page 34: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/34.jpg)
Skema peralatan GC
Plotter → kromatogram
![Page 35: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/35.jpg)
![Page 36: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/36.jpg)
1
2
3
45
![Page 37: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/37.jpg)
Carrier gas
- The carrier gas must be chemically inert.
- Commonly used gases include nitrogen, helium, argon, and carbon dioxide.
- The choice of carrier gas is often dependant upon the type of detector which is used.
- The carrier gas system also contains a molecular sieve to remove water and other impurities.
![Page 38: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/38.jpg)
Sample injection port
For optimum column efficiency, the sample should not be too large, and should be introduced onto the column as a "plug" of vapour - slow injection of large samples causes band broadening and loss of resolution.
Microsyringe is used to inject sample through a rubber septum into a flash vapouriser port at the head of the column.
The temperature of the sample port is usually about 50°C higher than the boiling point of the least volatile component of the sample.
For packed columns, sample size ranges from tenths of a microliter up to 20 microliters. Capillary columns, on the other hand, need much less sample, typically around 10-3 mL. For capillary GC, split/splitless injection is used. Have a look at this diagram of a split/splitless injector;
![Page 39: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/39.jpg)
A small amount of liquid (microliters) is injected through a silicon rubber septum (using a special microliter syringe) into the hot (usually 200+ degrees C) GC injector that is lined with an inert glass tube. The injector is kept hot by a relatively large, metal heater block that is thermostatically controlled.
The injector can be used in one of two modes; split or splitless. The injector contains a heated chamber containing a glass liner into which the sample is injected through the septum. The carrier gas enters the chamber and can leave by three routes (when the injector is in split mode). The sample vapourises to form a mixture of carrier gas, vapourised solvent and vapourised solutes. A proportion of this mixture passes onto the column, but most exits through the split outlet. The septum purge outlet prevents septum bleed components from entering the column.
![Page 40: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/40.jpg)
InjectorDiagram skematik
![Page 41: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/41.jpg)
Columns• There are two general types of column, packed and capillary (also known as open tubular). • Packed columns contain a finely divided, inert, solid support material (commonly based on
diatomaceous earth) coated with liquid stationary phase. Most packed columns are 1.5 - 10m in length and have an internal diameter of 2 - 4mm.
• Capillary columns have an internal diameter of a few tenths of a millimeter. They can be one of two types; wall-coated open tubular (WCOT) or support-coated open tubular (SCOT). Wall-coated columns consist of a capillary tube whose walls are coated with liquid stationary phase. In support-coated columns, the inner wall of the capillary is lined with a thin layer of support material such as diatomaceous earth, onto which the stationary phase has been adsorbed. SCOT columns are generally less efficient than WCOT columns. Both types of capillary column are more efficient than packed columns.
![Page 42: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/42.jpg)
Kolom Kapiler dalam oven
Kolom packing Kolom kapiler
![Page 43: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/43.jpg)
Column temperature
For precise work, column temperature must be controlled to within tenths of a degree.
The optimum column temperature is dependant upon the boiling point of the sample. As a rule of thumb, a temperature slightly above the average boiling point of the sample results in an elution time of 2 - 30 minutes. Minimal temperatures give good resolution, but increase elution times. If a sample has a wide boiling range, then temperature programming can be useful. The column temperature is increased (either continuously or in steps) as separation proceeds.
![Page 44: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/44.jpg)
capillary gas chromatography : high resolution. Capillary columns : 1) more theoretical plates per meter as compared to packed columns and 2) since they have less resistance to flow they can be longer than packed
columns. This means that the average capillary column (30 meters long) has approximately 100,000 theoretical plates while the average packed column (3 meters) has only 2500 plates
But with this separation power comes some limitations: 1) Capillary columns, because they have smaller diameters (0.05 to 0.53 mm) than packed columns (2 to 4 mm), require relatively specialized injectors and ancillary flow and pressure controllers and 2) capillary column require a smaller amount of sample than packed columns. While the average sample mass of each component in a mixture that is separable by packed column GC can be in the microgram range (10-6 grams) per injection, capillary columns routinely only handle 50 nanograms (10-9 grams) of a particular component or less.
![Page 45: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/45.jpg)
DETEKTOR
FID = Flame ionisation detector
ECD = Electron chapture
FPD = Flame photometry detector
AED = Atomic Emission Detector
PID = Photo ionisation detector …
![Page 46: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/46.jpg)
There are many detectors which can be used in gas chromatography. Different detectors will give different types of selectivity. A non-selective detector responds to all compounds except the carrier gas, a selective detector responds to a range of compounds with a common physical or chemical property and a specific detector responds to a single chemical compound.
Detectors can also be grouped into concentration dependant detectors and mass flow dependant detectors. The signal from a concentration dependant detector is related to the concentration of solute in the detector, and does not usually destroy the sample Dilution of with make-up gas will lower the detectors response. Mass flow dependant detectors usually destroy the sample, and the signal is related to the rate at which solute molecules enter the detector. The response of a mass flow dependant detector is unaffected by make-up gas. Have a look at this tabular summary of common GC detectors:
Detectors
![Page 47: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/47.jpg)
Detector
Type Support gases
Selectivity Detectability Dynamic range
FID Mass flow Hydrogen and air
Most organic cpds. 100 pg 107
TCD Concentration Reference Universal 1 ng 107
ECP Concentration Make-up Halides, nitrates, nitriles, peroxides, anhydrides, organometallics
50 fg 105
FPD Mass flow Hydrogen and air possibly oxygen
Sulphur, phosphorus, tin, boron, arsenic, germanium, selenium, chromium
100 pg 103
PID Concentration Make-up Aliphatics, aromatics, ketones, esters, aldehydes, amines, heterocyclics, organosulphurs, some organometallics
2 pg 107
NPD Mass flow Hydrogen and air
Nitrogen, phosphor 10 pg
![Page 48: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/48.jpg)
FID
![Page 49: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/49.jpg)
ECD
![Page 50: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/50.jpg)
FPD
![Page 51: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/51.jpg)
Photodiode dan photovoltaic diode
![Page 52: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/52.jpg)
PID
![Page 53: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/53.jpg)
AED
![Page 54: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/54.jpg)
Kondisi Operasional analisis
Jenis kolom
Bahan kolom (fasa diam)
Laju alir gas pembawa (carrier gas)
Suhu injector
Suhu kolom
Mode pemanasan kolom (isotermal atau temperatur terprogram)
Jenis detektor
Suhu detektor
![Page 55: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/55.jpg)
Hasil Kromatografi Gas
• Berupa kromatogram• Aspek kualitatif (nilai Rt)• Aspek kuantitatif
Metoda Standar:
Metoda standar eksternal Internal addisi
![Page 56: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/56.jpg)
Standar eksternal
• Sampel dibuat kromatogram diinjeksikan sendiri)
• Standar dibuat kromatogram• Dibandingkan antara nilai Rt standar dan Rt
sampel. Peak atau puncak sampel yang memiliki nilai Rt yang sama berarti senyawa sampel sama dengan senyawa standar
![Page 57: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/57.jpg)
Sandar internal
• Sampel dibuat kromatogram diinjeksikan sendiri)
• Standar + sampel dicampur dan dibuat kromatogram
• Puncak/peak sampel dan peak campuran diperbandingkan. Peak yang semakin tinggi menunjukkan senyawa sampel + sandar.
![Page 58: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/58.jpg)
![Page 59: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/59.jpg)
Liquid Chromatography (LC) Liquid chromatography (LC) is an analytical
chromatographic technique that is useful for separating ions or molecules that are dissolved in a solvent. If the sample solution is in contact with a second solid or liquid phase, the different solutes will interact with the other phase to differing degrees due to differences in adsorption, ion-exchange, partitioning, or size. These differences allow the mixture components to be separated from each other by using these differences to determine the transit time of the solutes through a column.
![Page 60: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/60.jpg)
![Page 61: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/61.jpg)
HPLC
![Page 62: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/62.jpg)
Kolom HPLC
![Page 63: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/63.jpg)
![Page 64: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/64.jpg)
![Page 65: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/65.jpg)
Waktu Retensi
Retention factor, k', is often used to describe the migration rate of an analyte on a column. You may also find it called the capacity factor. The retention factor for analyte A is defined as;
k'A = t R - tM / tM
t R and tM are easily obtained from a chromatogram.
The time taken for the mobile phase to pass through the column is called tM.
![Page 66: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/66.jpg)
TEORI PLAT
HETP = L / N
![Page 67: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/67.jpg)
Kromatogram
![Page 68: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/68.jpg)
Resolusi dan Selektivitas
selectivity factor, a , which describes the separation of two species (A and B) on the column;
a = k 'B / k 'A
![Page 69: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/69.jpg)
Jumlah plat
![Page 70: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/70.jpg)
TEORI LAJU ALIR
HETP = A + B / u + C u
where u is the average velocity of the mobile phase. A, B, and C are factors which contribute to band broadening.
![Page 71: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/71.jpg)
A - Eddy diffusionThe mobile phase moves through the column which is packed with stationary phase. Solute molecules will take different paths through the stationary phase at random. This will cause broadening of the solute band, because different paths are of different lengths.
B - Longitudinal diffusionThe concentration of analyte is less at the edges of the band than at the center. Analyte diffuses out from the center to the edges. This causes band broadening. If the velocity of the mobile phase is high then the analyte spends less time on the column, which decreases the effects of longitudinal diffusion.
C - Resistance to mass transferThe analyte takes a certain amount of time to equilibrate between the stationary and mobile phase. If the velocity of the mobile phase is high, and the analyte has a strong affinity for the stationary phase, then the analyte in the mobile phase will move ahead of the analyte in the stationary phase. The band of analyte is broadened. The higher the velocity of mobile phase, the worse the broadening becomes.
![Page 72: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/72.jpg)
![Page 73: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/73.jpg)
Metoda Analisis
• Kondisi operasional alat• Aspek kuantitaif: standar
- Standar ekxternal- Standar internal- Standar adisi
![Page 74: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/74.jpg)
Hasil Analisis Kromatografi
Aspek Kualitatif
Aspek kuantitatif
![Page 75: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/75.jpg)
![Page 76: Pert Ke 9 Kromatografi](https://reader035.vdocuments.pub/reader035/viewer/2022081421/577cc9fa1a28aba711a51aa3/html5/thumbnails/76.jpg)
Aplikasi Analisis kromatografi