qc ii lecture handout combined 10-11
DESCRIPTION
Lecture on quality control in the clinical laboratoryTRANSCRIPT
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 1
Quality Control II
CLS 431 Clinical Laboratory Management IIy gQuality Management II
QC Basics: What You Know
QC is a statistical tool used to ensure reliability of laboratory measurements Reliability = precision and accuracy
QC value must fall in its acceptable range Acceptable range = 95% confidence interval Within acceptable range: accept results Outside of acceptable range: reject results
2
Why are we here today? Clarify terminology
Control materials Control programs Internal controls External controls Instrument Performance Controls Batch and Random Access Assays
Discuss troubleshooting strategies3
Quality Control Procedure for monitoring work processes,
detecting problems, and making corrections prior to delivery of products and services
The CONTEXT QC is used in is very The CONTEXT QC is used in is very important, and can be misinterpreted
Must differentiate: Internal/External Quality Control Material Internal/External Quality Control Programs
4
What is CLSI?
CLSI provides guidelines for quality control The Clinical and Laboratory Standards
Institute (CLSI) is a global, nonprofit, standards-developing organization that p g gpromotes the development and use of voluntary consensus standards and guidelines within the health care community
5
Quality Control Material Clinical and Laboratory Standards Institute
(CLSI) Definition: A device, solution or lyophilized
preparation intended for use in thepreparation intended for use in the quality control process to monitor the reliability of a test system and to maintain its performance within established limits.
6
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 2
Components of a QC Program
7
How are QC Ranges Established? Assay new lot number of QC through
parallel testing to collect data
Current lot number of QC validates the run Current lot number of QC validates the run Do not use new lot QC values to accept/reject
a run
Do not accept new lot data point if current QC failed
8
How are QC Ranges Established? If the current lot number of QC is within
acceptable limits, the new lot value is valid and saved for statistical analysis
If the current lot number of QC is not within acceptable limits, the new lot value is invalid (rejected) and not saved for statistical analysis
9
Establish QC Acceptable Range The laboratory establishes its own means
and acceptable ranges based on the saved data of the new lot QC
Acceptable ranges should be establishedAcceptable ranges should be established using N=30 (tentative limits N=10 or N=20) Ideal: one data point per day
The laboratory should not rely solely on the means and ranges from the manufacturers package insert (these must be verified)
10
Is Documentation Really all that Important?
11
Document Changes to the Test System
Document any changes or processes to help troubleshoot future issues New lot # reagents opened/put into use
New lot # standards opened/put into use New lot # standards opened/put into use New calibration curve put into use New lot # of QC material opened Maintenance or when instrument repairs
occur
12
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 3
Documenting QC Results All technologists are responsible for
identifying, correcting and documenting QC problems as they occur
13
Failed QC results are never deleted/discarded
Documentation of ALL QC is reviewed by supervisor or designee according to SOP
Documentation of QC Failures All unacceptable control results should be
documented by indicating: If run is not acceptable If patient results cannot be reportedp p Corrective action(s) taken
If QC failure cannot be resolved in a timely manner, you may need to notify your customers (ER, oncology, OR, etc.) of the potential delay in reporting patient results
14
Leaving a trail: QC Documentation #1
CLSI provides recommendations for when quality control should be performed on commercially prepared microbiological media
Due to high failure rate in some commercially prepared media, a higher frequency of testing is performed (not needed for media that have low failure rates)
15
Leaving a trail: QC Documentation #1 Upon receipt in the laboratory, QC is
performed on the commercially prepared media just received We want to ensure that media after shipment
will support growth of the suspected organismg g
Growth of Campylobacter species is used to QC the CVA agar Campylobacter species also require the correct
gas mixture and temperature to support growth on the CVA agar
16
Leaving a trail: QC Documentation #1Incubation Technique Protocol A fresh subculture of a known Campylobacter
species is subbed from frozen on Sunday On Tuesday a new subculture is made from
Sundays culture growth On Thursday a new subculture is made from
Tuesdays growth
All subcultures are incubated in the same manner as patient samples and are checked for growth the following day
17
Example of the Campylobacter Quality control documentation:
S M T W Th FIndicate day C. jejuni was
Leaving a trail: QC Documentation #1
j jsubbed
Check for growth
18
Does anything appear wrong with documentation?
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 4
This organism had actually failed to grow on Monday The technologist suspected a problem with the
organism and subbed another from frozen On Tuesday: failed to grow subbed from frozen
Leaving a trail: QC Documentation #1
On Tuesday: failed to grow subbed from frozen On Wednesday: failed to grow subbed from frozen On Thursday On Friday..
Failure of the organism to grow (M F) was never documented, nor was it investigated
19
On Saturday the technologist questioned why the new lot of media had not been QCd with the subbed Campylobacterspecies Found a sticky note saying:
Leaving a trail: QC Documentation #1
Found a sticky note saying: QC organism not growing
A considerable amount of time was spent investigating the issue After piecing together verbal reports from
several other technologists, the problem was identified 20
Problem Identified 7 days earlier the wrong gas mixture was
delivered to the laboratory
Consequence of inadequate documentation
Leaving a trail: QC Documentation #1
Consequence of inadequate documentation Patient samples from the past 6-7 days
were compromised Medical director instructed technologist to
notify providers of the problem and the need to resubmit samples if clinically indicated 21
Documentation is Important Several technologists had been on the bench
No clear communication between technologists Lack of documentation delayed identification of
problem
Leaving a trail: QC Documentation #1
Time and resources wasted = $$$ lost Angry physician(s) Patient diagnosis delayed Patient inconvenience (recollect sample )
22
TB Quantiferon Testing Protocol: Manual test with strict QC requirements for
run to pass (std in duplicate; duplicates must agree; ABS value of STD cannot vary)
Leaving a trail: QC Documentation #2
Run failed because the standards did not meet these strict requirements QC failure not documented The technologist tossed the worksheet (thought
he/she made a pipetting error and did not want anyone to know)
23
What are the potential issues with not documenting the failed run? Next run, standards again failed requirements
Run was repeated and again the standards failed
Leaving a trail: QC Documentation #2
requirements After two failed runs, the issue was investigated:
Pipette used to dilute the standards appeared to be sticking, the pipette was checked and failed QC
After the pipette was cleaned, the pipette passed QC
24
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 5
Important to document If the original run had been documented as a
failure, troubleshooting would have been done after the second run failed, rather than the 3rd
V i t t t ( $500) b t
Leaving a trail: QC Documentation #2
Very expensive test to run (~$500), but even more so since it was run four times before results could be reported
Time, resources, money, patient care!!
25
Troubleshooting Strategies
26
Troubleshooting QC for Manual vs. Automated Tests
Usually there are fewer potential sources of error in an automated method, as the instrument does not vary the analytic y yprocess from specimen to specimen
With manual methods, there is variation between CLSs, and possibly between runs performed by the same CLS
27
Troubleshooting for Manual Tests Begin with the processing steps that are
more prone to human error: Was the proper SOP followed? Were proper techniques employed? Were the proper control materials selected?Were the proper control materials selected? Had any of the materials used expired? Was there a calculation/transcription error?
Manual methods are more frequently corrected by simply repeating the test
28
Basic Troubleshooting Steps Was the proper procedure followed?
Check the controls: Expired? Correct control? (level) Current/new lot number?
Check the reagents: Expired? Contaminated? (cloudy, bacteria) Correct reagents? Properly prepared? Current/new lot number? 29
Basic Troubleshooting Steps Check the instrument:
Mechanical malfunction? Required maintenance performed? Calibration current?
Check for calculation or transcription errors
30
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 6
QC Troubleshooting Flow Chart
Flow charts may be used by some laboratories in an effort to streamline the troubleshooting process
It i i t d d t di t t f It is intended to direct you to perform a series of defined steps in response to a QC failure
This way, all technologists follow this same series of steps
31
QC Troubleshooting Flow Chart Intended to minimize use of controls,
recalibration procedures, tech time and frustration
Often helps inexperienced technologists work through the troubleshooting process What should I do in response to a QC failure?
32
33
Using QC Failure Flowchart
You have just completed the daily maintenance on the urinalysis dipstick reader. Control results follow:
Level 1: within 2SD of meanLevel 2: outside 3SD of mean
You have already rerun the same level 2 control and results remain outside 3SD
What is your next step?34
How QC Policies May Vary
35
How is QC Used in the Lab? Contingent on the type of testing performed
Qualitative assay Quantitative assay Batch assay Batch assay Random Access assay Laboratorys SOP
36
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 7
Qualitative Assay Positive or negative results
Reactive or nonreactive resultsGrowth or no growth
Control(s) must yield the expected results before reporting patient results
Can be manual or automated assay If QC result(s) fails for manual assay, suspect
human error first37
Qualitative Assay Examples Screening procedures
Pregnancy Strep throat Antigen typing Hemoglobin S Monospot
Identification tests: catalase, oxidase Media checks: growth/no growth Test reactivity: Rh-hr, Coombs control cells
38
Using 1 Control: Control Fails Blood Bank Antibody Screen
SC AHG CCI 0 3+II 0 0III 0 3+
QC for Screen Cell (SC) II failed Coombs Control Cells did not yield the expected positive result
Patient results can not be reported Test must be repeated on SCII
39
Using 2 Controls: 1 FailsBlood Bank Antigen Typing
Anti-Fya
AHG CC
Positive Control Cells 0 3+
Negative Control Cells 0 3+
Positive Control Cells failed Patient had positive reactivity Coombs Control Cells worked
g
Patient Cells 2+
40
Using 2 Controls: Both FailBlood Bank Antigen Typing
Neither control gave expected result Look for problems with:
Reagent Reagent Method used Cells selected for controls
41
Quantitative Assay Numerical results
Glucose = 185 mg/dl Potassium = 3.2 mmol/L
Si i l l bt i d Since numerical values are obtained, we can calculate QC statistics (mean, SD, CV) Control acceptable ranges can be stored in
computer memory or we can manually prepare Levey-Jennings charts for visual review of QC data
42
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 8
Quantitative Assay Most often 2 levels of control are analyzed
Westgards rules are often used as minimal acceptable standards to accept/reject QC
Some laboratories and/or departments may choose to use stricter guidelineschoose to use stricter guidelines
Even one QC value 2-3SD from mean = QC failure
Examples: Batch assay (instruments or manual testing) Random access instruments
43
What is a Batch Assay? Can be a manual or automated test
Can analyze multiple specimens at one time (per run) but can only analyze ONE analyte(per run) but can only analyze ONE analyte per run
Each run requires validation of calibration using standards and/or controls
44
Example of a Manual Batch Assay
45
Example of Automated Batch Assay
46
QC Evaluation: Batch Assay Before patient results are released, QC is
evaluated: If QC fails, the entire run fails:
Reject all test results (controls, patients)
If you reject the run, troubleshooting should occur. After the problem has been corrected, the entire run will need to be re-analyzed
47
What is a Random Access Assay? Automated instrument that is capable of
analyzing any test at any time Mary Jones: TSP, BUN, Creatinine Joe Smith: Glucose, Na, K, Cl, BUN, TSP, , , , ,
Capable of interrupting the instruments current work list to analyze STAT specimens out of sequence
48
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 9
Current work list Mary Jones: TSP, BUN, Creatinine Joe Smith: Glucose, Na, K, Cl, BUN, TSP
What is a Random Access Assay?
STAT order John Doe: Amylase, Lipase
Work list continued after STAT testing completed 49
Examples of Random Access
Vitros Beckman Centaur Sysmex Sysmex Provue Galileo Stago
50
QC Evaluation: Random Access Usually QC for all analytes on the
instruments test menu are run at the start of the shift or once per day (follow SOP)
Before you can analyze and/or reportBefore you can analyze and/or report patient test results, QC is evaluated If QC is acceptable, patient samples can be
analyzed, evaluated and reported
If QC fails, patient samples cannot be analyzed until troubleshooting is performed and QC is acceptable 51
Evaluation of QC Failures: Random vs Batch Assays,
RAE vs SAE
52
Review of RAE vs SAE
Determine if the QC failure is caused by a RAE or SAE
Only one control >3SD: RAEy Both controls >3SD: SAE Shift: SAE Trend: SAE Second consecutive control value 2-3SD
from the mean: SAE
53
Chemistry Random Access Assay #1
You are running the required two controls for a random access chemistry assay. The level 1 control value is within 2SD of the mean and the level 2 control is >3SD from the mean
What is the next appropriate action?
A. Report both controls and patientsB. Rerun a new vial of the level 2 control onlyC. Rerun the same vial of both controlsD. Recalibrate the assay
54
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 10
You are running the required two controls for a chemistry batch assay. Both levels of control are >3SD from the mean
What is the next appropriate action?
Chemistry Batch Assay #2
A. Report all control and patient resultsB. Rerun all controls and patients againC. Rerun only the controls, but use fresh vialsD. Recalibrate the assay, then rerun controls and all
patients55
Types of Control Materials
56
Internal Control Materials Instrument Performance Controls
Checks that are performed to ensure instrument is operating as expected
Electronics Voltage Voltage Optics Fluidics Pneumatics/hydraulics Pumps Pipette alignment
57
Internal Control Materials
Controls that are built into the test platform Control line: pregnancy, Rapid-Strep pH indicator : Rapid-Strep, EluKit
Background clearing: pregnancy Rapid Strep Background clearing: pregnancy, Rapid-Strep Color change: back of glucose strip turns
blue, Low Ionic Strength Solution (LISS) changes color when patient plasma is added
58
Internal Control Materials
Example: pregnancy screen
59
External Control Materials
Control material that is tested in a manner identical to patient testing
Thi t i th b lk f QC f d This category is the bulk of QC performed in laboratory testing Daily controls Level 1, Level 2 Abnormal, Normal
60
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 11
One Level of Control
Not common practice to use only one control
May run only one level of control to verify an instrument problem has developed (seeing multiple bad patient CBC results) Troubleshooting can begin sooner
Alternatively may rerun a previously reported sample, and compare to see if results match Patient sample acts as a free control Used in conjunction with, not instead of low, normal or
high liquid control material61
Two Levels of Control:Coagulation Instrument
Normal and Abnormal controls are used
Labs SOP states: troubleshooting required for any control value exceeding 2SD of the mean PRIOR TO REPORTING ti t ltPRIOR TO REPORTING patient results. Differentiate a control problem from an
instrument/reagent problem, and once problem is corrected.
Determine if control material, patient samples or both need to be repeated
62
Two Levels of Control:Coagulation #1
When entering QC, the Laboratory Information System (LIS) shows the following SDs from the mean
What test(s) are reportable? What test(s) require troubleshooting prior to reporting patient results?
63
Do you suspect a control vial problem or instrument/reagent problem?
Two Levels of Control:Coagulation #1
64
Once the APTT controls are in range, do the patient samples need to be repeated as well as the controls?
Two Levels of Control:Coagulation #1
65
When entering QC, the LIS shows the following SDs from the mean
What test(s) are reportable? What test(s) require troubleshooting prior to reporting
Two Levels of Control:Coagulation #2
require troubleshooting prior to reporting patient results?
66
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 12
Do you suspect a control vial problem or instrument/reagent problem?
Two Levels of Control:Coagulation #2
67
Once the control is in range for PTs and APTTs, do the patient samples need to be repeated?
Two Levels of Control:Coagulation #2
repeated?
68
Causes of Control/Reagent Problems
Manufacturers instructions should alwaysbe followed Most reconstituted control or reagents need
an equilibration time at room temperature prior to use (e g 30 minutes)prior to use (e.g., 30 minutes)
Frozen controls should be thawed by following manufacturers instructions
Room temperature 37 C water bath or heat block
69
Recommended stability times should be strictly followed
Always write a date and time on controls and
Causes of Control/Reagent Problems
yreagents when they are reconstituted, thawed or opened
Remember, documentation is invaluable when troubleshooting an out of range control result
70
Low, normal, high controls
Manufacturer instructions should always be followed:
Three Levels of Control:Cell Counter
followed: Warm at room temp for 15 minutes Mix thoroughlybut GENTLY (these are cells!)
Proper documentation is essential Date opened Be aware of expiration date (cell stability) 71
Labs SOP states: All directly measured parameters must be 2SD
from the mean on at least two of three levels of control
Three Levels of Control: Cell Counter
A result on a directly measured parameter can be between 2-3SD IF the same parameter is 2SD on the other two levels
Calculated parameters can be between 2-3SD if the directly measured parameters used in their calculation are 2SD from the mean 72
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 13
When entering QC results, your LIS shows the following SD from the mean. On this instrument, the RBC and Hgb are directly measured; others are calculated.
What action should be taken?
Three Levels of Control: Cell Counter #1
What action should be taken?
73
Analyze the following control results from the same analyzer the next day. On this instrument, the RBC and Hgb are directly measured; others are calculated.Wh t ti h ld b t k ?
Three Levels of Control: Cell Counter #2
What action should be taken?
74
The manufacturer of reagents and instruments often have controls that are formulated to match the reagent, standard and/or a specific instrument
Manufacturer Controls
How do you compensate for a possible bias caused by the manufacturers formulation?
75
Manufactured by an independent company other than the instruments manufacturer Provides an independent, unbiased
assessment of your laboratory method
What Are Third Party Controls?
These can be the controls that are usually run with your assay every day
Biorad: Lyphocheck I, Lyphocheck II Dade: Level 1, Dade Level 2 Streck: Low, Normal, High
76
Advantages: Long-term QC monitoring: manufacturer controls
expire when standards or reagents expire
Different formulation than standards or reagents
Use of Third Party Controls
Same control can be used over multiple reagent or calibration changes
Some controls are used on multiple instruments
Disadvantage: costs money77
Use of Third Party Controls:Detecting a Problem
Labs SOP: After instrument preventative maintenance
(PM) is performed, the manufacturers controls AND third party controls must be
78
controls AND third party controls must be analyzed
Results of both controls (manufacturer and third party) must be within acceptable limits for the system to be in control
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 14
QC Results after PMTest Units
3rd party Level 1
Expected Range
3rd party Level 2
Expected Range
Triglyceride mg/dl 214 147-209 109 74-108
Cholesterol mg/dl 267 200-265 113 84-111
79
Test UnitsManufacture Level 1
Expected Range
Manufacture Level 2
Expected Range
Triglyceride mg/dl 137 109-143 267 229-283
Cholesterol mg/dl 143 128-154 260 233-265
QC Results after PM Both levels of the manufacturers controls
were within the expected range, but both levels of the 3rd party control were out high
Additional troubleshooting was performed on
80
Additional troubleshooting was performed on the instrument and it was discovered that the sampling probe was out of alignment and needed adjustment
After the probe adjustment was made, another set of both controls (manufacturer and 3rd party) were run
QC Results after Re-alignmentTest Units 3rd party
Level 1Expected Range
3rd party Level 2
Expected Range
Triglyceride mg/dl 198 147-209 100 74-108
Cholesterol mg/dl 248 200-265 103 84-111
81
Test Units Manufacture Level 1
Expected Range
Manufacture Level 2
Expected Range
Triglyceride mg/dl 130 109-143 255 229-283
Cholesterol mg/dl 139 128-154 245 233-265
QC Results After Re-alignment Both the manufacturers controls and the 3rd
party controls were within the expected limits
The 3rd party control was able to detect the
82
problem that occurred after the PM
The manufacturers controls did not detect the sampling probes misalignment
The sampling probes shift out of alignment could have affected patient results
Types of Control Programs
83
External/Interlaboratory QC Program
Example:Mandatory Proficiency Testingy y g
84
-
CLS 431 Clinical Laboratory Management II - Quality Management IIQuality Control II Lecture Handout 15
Proficiency Testing CLIA regulations require clinical laboratories
participate in a proficiency testing program for regulated analytes
Testing an unknown using the same methods Testing an unknown using the same methods, reagents, and personnel as used to perform patient testing
Samples usually provided by an outside vendor College of American Pathologists (CAP)
85
Proficiency Testing Comparison is made between the results the lab
obtained and the referee results, as well as with the results of other laboratories in the nation
Allows the laboratory to monitor its performance Allows the laboratory to monitor its performance in relation to that of its peers using the same test method, and those using different test methods
Proficiency Testing will be discussed further as a component of the CLS431 course, in the Compliance and Regulatory Issues unit
86
87
Resources Used www.biorad.com www.cap.org www.jointcommission.org www.clsi.org
88
McKenzie, S.B. & Williams, J.L., Clinical Laboratory Hematology, 2nd edition, Pearson Education, Inc., 2010.
Westgard, James O., Basic QC Practices, 3RDEdition, Westgard QC, Inc., 2010
-
Policy 6.001 Attachment 1
Quality Control Failure Troubleshooting Flowchart
Control Within Limit Control Not within Limit Control Within Limit Control Not within Limit Control Within Limit
Control Not within Limit Control Within Limit Control Still Not within Limit Document all remedial action on appropriate log sheet.
Control falls outside the satisfactory control limit
Repeat the control
Use control from another bottle
Open another reagent with the same and/or new lot number and repeat with the first vial
Using the new reagent, repeat with the new vial of control
CANNOT USE FOR PATIENT TESTING. If instrumentation, call the Customer Service Hotline of the Manufacturer for assistance. If they are unable to help correct the problem, have it checked and repaired if necessary by Biomedical Instrumentation. If kit, discontinue using kit and open another kit.
PROCEED
WITH PATIENT
TESTING
Contact Lab Consultant
QCII Lecture Handout 10-11.pdfQuality Control Failure Flowchart.pdf