software updates to support 6-dye chemistry implementation

44
1 Life Technologies™ | 1 Life Technologies™ | Software Updates to Support 6-dye Chemistry Implementation

Upload: human-identification

Post on 20-Aug-2015

519 views

Category:

Business


5 download

TRANSCRIPT

Page 1: Software Updates to Support 6-dye Chemistry Implementation

1 Life Technologies™ | 1 Life Technologies™ |

Software Updates to Support 6-dye Chemistry Implementation

Page 2: Software Updates to Support 6-dye Chemistry Implementation

2 Life Technologies™ |

GlobalFilerTM Express Workflow Solutions

Page 3: Software Updates to Support 6-dye Chemistry Implementation

3 Life Technologies™ |

6-Dye Summary

Chemistry/  Standards  

     

GeneMapper®  ID-­‐X  So5ware  

v1.4  

Service  

3730  DNA  Analyzer  /  

31xx  GeneEc  Analyzers  

3500  GeneEc  Analyzer  

Data  CollecEon  So5ware  

1   2   3  

31xx/3730  only    All  Instruments  

 

Page 4: Software Updates to Support 6-dye Chemistry Implementation

4 Life Technologies™ |

What  CE  instrument  are  you  currently  using  for  casework?  

A.  310 B.  3100 C.  3130 D.  3500

Page 5: Software Updates to Support 6-dye Chemistry Implementation

5 Life Technologies™ |

Product  Overview  3500  Data  CollecEon  v2  Updates  

§  Windows® 7 Professional SP1 utilizes a newer, better, fully supported operating system designed for newer systems

§  New Dell Optiplex XE - 4 GB RAM, 500 GB RAID Disk replaces  exis8ng,  no  longer  supported  computer  systems

§  Modern Licensing Technology with a new license scheme that utilizes on-line registration §  RFID warnings have been added for (HID/RUO version only) §  Data migration from Data Collection v1 -> v2 §  Software issues addressed include:

−  Improved spatial calibration QC checks

−  Updated instrument run modules to improve polymer leak detection during array fill operations

−  Consumable Usage Statistics Report added

−  Consumables Dashboard improvements

−  Target EP voltage displayed correctly during run on Sensor Details page

−  Hovering tool in the Fragment Results view displays the quality value for OffScale data

−  Improved Instrument Error Dialog when reporting fatal instrument errors

§  GeneMapper® ID-X Software Compatibility: .hid sample files generated by the 3500 Data Collection software v2 can be viewed and analyzed with GeneMapper® ID-X v1.4 – v1.2 Software.  

Reagent and consumable lifetime minimum performance values are based on developmental validation studies and promote the production of high quality data and help minimize instrument downtime.

Page 6: Software Updates to Support 6-dye Chemistry Implementation

6 Life Technologies™ |

3500 System Workflow and Feature Highlights

Instrument   Run  Set-­‐Up   Data  CollecEon   GMID-­‐X  v1.2  QC  Analysis  

Redesigned  laser  and  polymer  pump  block  

Pre-­‐packaged  consumables  with  RFID  Tracking

Intuitive Run Set-Up and system monitoring

Predefined protocols and HID templates

New oven design Shorter run times

Optional signal normalization

Data analysis QC tools and flags

Simplified reinjection scheme

Upgrade to GeneMapper® ID-X 1.4 for

Window ® 7 comp

Analyzes .fsa & new .hid file format

Instrument

Run Set-Up

Data Collection

QC Analysis

GeneMapper® ID-X v1.4

Page 7: Software Updates to Support 6-dye Chemistry Implementation

7 Life Technologies™ |

Dashboard

Key      OperaEons  

Maintenance  Prompts  

Instrument  Status  

Consumable  Gauges  

Page 8: Software Updates to Support 6-dye Chemistry Implementation

8 Life Technologies™ |

Spatial Calibration Improvements

• Evaluates  each  peak  to  confirm  that  is  above  the  threshold.  • Capillary  spacing  enhanced  to  look  at  the  spacing  difference  between  cap  n  and  n+1  (which  is  why  the  spacing  of  the  last  capillary  is  always  0).    

Added  the  peak  threshold  to  help  detect  this  issue.

Page 9: Software Updates to Support 6-dye Chemistry Implementation

9 Life Technologies™ |

Redesigned Consumable Packaging

§  Pre-filled, quality-controlled reagents

§  Radio Frequency Identification Tags −  Fixed, passive tags −  Limited read range −  Memory only, cannot be read by the

computer and cannot execute code

§  Comparable per sample running cost to 31xx instruments

Page 10: Software Updates to Support 6-dye Chemistry Implementation

10 Life Technologies™ |

Radio Frequency Identification (RFID) §  Three independent RFID readers track and record

information on four primary consumables (with RFID labels)

§  Consumables tracked −  Polymer −  Cathode Buffer −  Anode Buffer −  Array

§  Information tracked −  Lot numbers −  Part numbers −  Serial numbers −  Dates (expiration and installation) −  Capacity/Usage

Page 11: Software Updates to Support 6-dye Chemistry Implementation

11 Life Technologies™ |

Consumables: Polymer Pouch Volume

Excess volume for installation, wizards and proper pouch

functionality

Purchased volume

- 960 or 384 samples

Provides: -  Predictable cost per sample

-  Ease of use -  Data tracking (via RFID)

Page 12: Software Updates to Support 6-dye Chemistry Implementation

12 Life Technologies™ |

3500  Data  CollecEon  v2  New  RFID  SeZngs  

The  use  of  consumable  beyond  the  manufacturer’s  expiry  date  or  on-­‐instrument  life  can  impact  data  quality  and  cause  damage  to  the  instrument.    The  cost  of  repairing  such  damage  is  NOT  covered  by  

Life  Technologies’  warranty  or  service  plans.  

2On-instrument life/sample usage limit is based on whichever limit is reached first. The primary indicator of pouch depletion is the Sample Count. Injection count is present as a fail safe mechanism for low sample count reinjection scenarios. 1Polymer validated for 7 days on instrument and the array for 160 injections

Page 13: Software Updates to Support 6-dye Chemistry Implementation

13 Life Technologies™ |

Calibration and Maintenance Calendar

FR or F = Applied Biosystems recommended tasks, user editable - specify dates

R = User-specified repeating tasks

Page 14: Software Updates to Support 6-dye Chemistry Implementation

14 Life Technologies™ |

Calibration and Maintenance Reports and Export Files

Page 15: Software Updates to Support 6-dye Chemistry Implementation

15 Life Technologies™ |

QC Analysis: QC Protocol

•  QC protocol is the required primary analysis protocol for HID applications

•  It defines peak detection, sizing, and quality values

•  Factory-provided QC protocols are available in the software or can be created by the user

G5_LS(80-400)+Normalization

Page 16: Software Updates to Support 6-dye Chemistry Implementation

16 Life Technologies™ |

QC Analysis: Result Review

   Review  QC  Flag  Details  –  Set  Up  Reinjec8ons    

   Sizing  Table  

   Plot  view  with  plot  op8ons:  by  dye,  overlay,  off  scale  

Page 17: Software Updates to Support 6-dye Chemistry Implementation

17 Life Technologies™ |

Improved  Signal  Consistency:                                                                                  NormalizaEon  Methods  

§  Factory  StandardizaEon:  Hardware-­‐based  calibra8on  for  more  consistent  instrument  to  instrument  performance      

§  Internal  Standard  (IS)  NormalizaEon:    Chemistry  and  So[ware  based  method  for  more  consistent  signal  across  capillaries,  injec8ons  and  instruments    

•  User  enabled  •  U8lizes  re-­‐designed  GeneScan™  600  LIZ®  Size  Standard    •  Sample  level  peak  heights  are  scaled  rela8ve  to  the  intensity  of  the  co-­‐injected  size  standard  compared  to  an  op8mized  average  size  standard  peak  height  (Normaliza8on  Target)  

Page 18: Software Updates to Support 6-dye Chemistry Implementation

18 Life Technologies™ |

Signal Normalization Tools

§  Forensic laboratories with multiple instruments have reported that signal variation between instruments can impact data interpretation

§  Factory standardization and user-selectable internal standard normalization minimize signal variation between instruments and from injection to injection

%CV

%C

V

Injection to Injection Peak Height Consistency

%C

V

Instrument to Instrument Peak Height Consistency

Page 19: Software Updates to Support 6-dye Chemistry Implementation

19 Life Technologies™ |

Are  you  interested  in  using  the  normalizaEon  feature  of  the  3500?  

A.  Yes, definitely! B.  We are looking into it, but are unsure C.  No D.  I have never heard of this before!

Page 20: Software Updates to Support 6-dye Chemistry Implementation

20 Life Technologies™ |

Data Scaling Considerations

§  Peak heights on the 3500 ~4x greater than 31XX platforms due to differences in data scaling

§  Higher peak amplitude threshold required for 3500 analysis if using a static threshold −  Use of the Global Cut-off function in the analysis method may allow for fine-tuning of analysis

thresholds

3130xl

3500xL

3000

9000

Page 21: Software Updates to Support 6-dye Chemistry Implementation

21 Life Technologies™ |

Comparison of 3130xl and 3500xL Data

PAT = 50 RFU PAT = 175 RFU 3130xl 3500xL

Identification of appropriate analysis thresholds promotes consistency of analysis outcome

Page 22: Software Updates to Support 6-dye Chemistry Implementation

22 Life Technologies™ |

PAT Definition for 3500 Data PAT: 50rfu PAT: 175rfu

Page 23: Software Updates to Support 6-dye Chemistry Implementation

23 Life Technologies™ |

Security, Audit and E-Sig

§ User Configurable according to laboratory’s needs § Security controls user access to system functions and data § Auditing tracks changes and provides history reports § E-signature requires user-authentication before changes are

saved

Page 24: Software Updates to Support 6-dye Chemistry Implementation

24 Life Technologies™ |

System Preferences

Default application, polymer and array

Decrease # of spectral borrowing events

Page 25: Software Updates to Support 6-dye Chemistry Implementation

25 Life Technologies™ |

GeneMapper® ID-X Software v1.4 Sample Info Tab

Consumable information tracked via RFID

Normalization Factor Value

Page 26: Software Updates to Support 6-dye Chemistry Implementation

26 Life Technologies™ |

Product  Overview  3130/3730  Data  CollecEon  So5ware  v4  Updates  

§     Windows®  7  Professional  SP1  u8lizes  a  newer,  becer,  fully  supported  opera8ng  system  §     New  Dell  OpEplex  XE  -­‐    4  GB  RAM,  500  GB  RAID  Disk  replaces  exis8ng,  no  longer  supported  computer  systems  §   6-­‐dye  capability  view  6  dye  data  in  the  Capillaries  Viewer,  Cap/Array  Viewer,  and  Spectral  Viewer  windows        (+  6-­‐dye  Module  to  run  and  analyze  6-­‐dye  data)  §     Updated  Sapphire  pump  wizard  seZngs  updated  pump  control  values  to  be  in  line  with  system  specifica8ons  §     Modern  Licensing  Technology  u8lizing  on-­‐line  so[ware  registra8on    §     Data  migraEon  from  UDC3.0,  3.1  -­‐>  UDC4.0  §     Oracle  180  day  password  expira8on  fix  §     MAC  address  Holix  integrated  §     BigDye®  Direct  mobility  files    §     GeneMapper®  ID-­‐X  So5ware  CompaEbility:    .fsa  sample  files  generated  by  this  so[ware  can  be  viewed  and  analyzed  with  all  versions  of  GeneMapper®  ID-­‐X  or  GeneMapper®  ID  So[ware.  

Page 27: Software Updates to Support 6-dye Chemistry Implementation

27 Life Technologies™ |

Data  CollecEon  AddiEonal  InformaEon

§  3500 Series Software 2 GETTING STARTED GUIDE PN 4476988 Rev. A. Contains instrument calibration, run setup data review, software licensing information.

§  3130 Series Data Collection Software 4 GETTING STARTED GUIDE PN 4477796 Rev. A. Contains instrument calibration, run setup data review, software licensing information.

§  3730/3730xl DNA Analyzer Data Collection Software 4 GETTING STARTED GUIDE PN 4478016 Rev A. Contains instrument calibration, run setup data review, software licensing information.

§  Online 3500 FAQs www.appliedbiosystems.com/3500hid

•  3500 Webinar Series Considerations for Efficient Implementation of the 3500 Series Genetic Analyzer for Human Identification www.3500Webinar1.com

Page 28: Software Updates to Support 6-dye Chemistry Implementation

28 Life Technologies™ | 28 Life Technologies™ |

GeneMapper® ID-X v1.4 Updates

Page 29: Software Updates to Support 6-dye Chemistry Implementation

29 Life Technologies™ |

Product  Overview  GeneMapper®  ID-­‐X  v1.4  Value  ProposiEon  

 

§  A  single  data  analysis  tool    that  significantly  reduces  the  amount  of  8me  required  to  perform  data  review,  mixture  analysis  and  staEsEcal  calculaEons  on  rou8ne  and  casework  forensic  samples.  

§  Expert  system  and  mixture  interpretaEon  capabiliEes  introduced  specifically  to  handle  expanded  core  loci  set  of  GlobalFilerTM  Kit,  minimizes  data  review  8me  and  provides  an  addi8onal  layer  of  quality  assurance  to  the  review  and  interpreta8on  process.    

 

§  Long-­‐standing,  trusted  approach  to  fragment  analysis  (same  validated  algorithms  from  GeneMapper  ID)  minimizes  valida8on  and  implementa8on  burdens.  

§  CriEcal  component  of  a  complete,  validated  soluEon  providing  op8mized  and  validated  workflow  solu8on  –  to  address  expanding  data  processing  and  interpreta8on  needs.  

Page 30: Software Updates to Support 6-dye Chemistry Implementation

30 Life Technologies™ |

Product  Overview  GeneMapper®  ID-­‐X  v1.4  So5ware  Updates  

•  Fragment  and  mixture  analysis  to  support  4,  5  and  6  dye  analysis  •  Expert  system  funcEonality  (PQVs)  to  evaluate  Amelogenin  and  Y  marker  data  quality  –  

ACC  (Amelogenin  Cross  Check)  PQV  and  Y  marker  AN  PQV  •  Updated  allele  labeling  funcEonality  to  handle  both  homozygote  genotypes  and  

haplotypes  in  a  single  sample  •  Enabled  spike  detecEon  switch  to  turn  on/off  automated  spike  detec8on  for  sample  type  

=  ‘’sample’’  (similar  to  the  allelic  ladder  spike  detec8on  switch  from  v1.1.1)  •  Latest  analysis  files  (including  GlobalFiler™  panels  and  bins  and  GlobalFiler™  Example  

Data)  •  Windows®  7  and  XP  compa8ble  •  Issues  Addressed:  Updated  ‘redefine  sample  file  path’  for  .hid  (3500)  files,  minor  update  

to  exis8ng  popula8on  data  

Validated  with  data  generated  from  the  GlobalFiler™  and  Legacy  AmpfℓSTR®  Kits  on  the  3130,  3730  and  3500  CE  plalorms.    Concordant  with  data  analyzed  using  GeneMapper®  ID-­‐X  

v1.3.  

Page 31: Software Updates to Support 6-dye Chemistry Implementation

31 Life Technologies™ |

31  

Y  Marker  Handling    Panel  and  Bin  Func4onality  

Y  Marker  Checkbox  is  used  by  the  so[ware  to  apply  Y-­‐marker  handling  rules  for        ACC  PQV,  AN  PQV,  Mixture  Analysis  and  the  Duplicate  Homozygous  Allele  Label  

Page 32: Software Updates to Support 6-dye Chemistry Implementation

32 Life Technologies™ |

32  

Y Marker Handling Amelogenin Cross Check (ACC) PQV

New  Marker  Level  AMEL/Y  Markers  PQV:        -­‐   Cross  checks  Amelogenin  results  with  Y  marker  (selected  in  the  Panel                                          Manager)  results    

-­‐   The  ACC  PQV  flag  will  fire  when  Amelogenin  and  Y  marker  results  are  discordant  

Y

Page 33: Software Updates to Support 6-dye Chemistry Implementation

33 Life Technologies™ |

33  

Y  Marker  Handling    Amelogenin  Cross  Check  (ACC)  PQV  

Page 34: Software Updates to Support 6-dye Chemistry Implementation

34 Life Technologies™ |

34  

Y  Marker  Handling    ACC  PQV  Results  for  GlobalFiler  

Page 35: Software Updates to Support 6-dye Chemistry Implementation

35 Life Technologies™ |

35  

Y Marker Handling ACC PQV – GQ/CGQ

 New  ACC  GQ  SeZng:    By  adjus8ng  the  weight  of  the  ACC,  the  user  can  be  alerted  to  discordant  results  between  Amelogenin/Y  Marker(s)  with  a  Genotype  Quality  flag  (GQ)      ACC  weigh8ng  contributes  to  the  status  of  Quality  Value  flags  for  GQ  and  in-­‐turn  CGQ.  

Page 36: Software Updates to Support 6-dye Chemistry Implementation

36 Life Technologies™ |

36  

When  the  ACC  Fails:  •   Marker  header  bars  (Genotype  Quality  flag)  may  fire  as  yellow  or  red  depending  on  the  weight  ACC  has  for  the  GQ  flag  •   Quality  Value  Details  Window  =  Amelogenin/Y  marker  discordance  

Y  Marker  Handling    ACC  PQV  

Page 37: Software Updates to Support 6-dye Chemistry Implementation

37 Life Technologies™ |

37  

Y Marker Handling Y Marker AN PQV

Allele  Number  PQV  Threshold:    -­‐   A  new  AN  Threshold  field  has  been  added  for  Y  Markers  

-­‐   The  AN  flag  will  fire  if  the  number  of  alleles  exceeds  the  user  defined  threshold  

Page 38: Software Updates to Support 6-dye Chemistry Implementation

38 Life Technologies™ |

38  

Y Marker Handling Mixture Analysis

A  warning  message  will  be  displayed  when  a  sample  that  contains  Y  markers  is  imported  into  the  MA  Tool  reminding  the  user  that  Mixture  Analysis  does  not  analyze  the  Y  markers.  

Page 39: Software Updates to Support 6-dye Chemistry Implementation

39 Life Technologies™ |

Are  you  interested  in  using  the  Mixture  Tool?  

A.  Yes, we are already using it B.  We plan to start using it soon C.  We are looking into it, but unsure D.  No, we are not interested at this time E.  Mixture tool? You have one of those?????

Page 40: Software Updates to Support 6-dye Chemistry Implementation

40 Life Technologies™ |

40  

Y Marker Handling Allele Labeling Option

Existing

Project  Op8on>  Duplicate  Homozygous  Alleles  op8on      –  Y  marker  alleles  (designated  in  the  Panel  Manager)  are  displayed  with  a  single  label  

Page 41: Software Updates to Support 6-dye Chemistry Implementation

41 Life Technologies™ |

41  

Spike Detection Switch Sample & Controls

Enhanced  to  allow  you  to  disable  spike  detec8on  for  samples  and  controls.      Expands  disable/enable  spike  detec7on  func7onality  introduced  for  allelic  ladders  in  v1.1.1  

Page 42: Software Updates to Support 6-dye Chemistry Implementation

42 Life Technologies™ |

Updated  Spike  DetecEon    Rules  

Page 43: Software Updates to Support 6-dye Chemistry Implementation

43 Life Technologies™ |

§  GeneMapper®  ID-­‐X  So5ware  Version  1.4  User  BulleEn  PN4477684  Rev  B  Contains  informa8on  on  the  new  features,  so[ware  valida8on  and  installa8on  instruc8ons  

 

§  Updates  to  ID-­‐X  So5ware  Help  -­‐  informa8on  on  6  dye  specific  features,  func8onality  and  troubleshoo8ng  

 

§  So5ware  Release  Notes  includes  overview  of  new  features  and  system  requirements  

 §  So5ware  ValidaEon  CerEficate  covers  valida8on  of  GlobalFiler™  and  Legacy  

AmpfℓSTR®  Kits  on  the  3130,  3730  and  3500  CE  plalorms                                For  more  informaEon  go  to  http://idx.appliedbiosystems.com      

GeneMapper®  ID-­‐X                                                                                                                            User  DocumentaEon      

Page 44: Software Updates to Support 6-dye Chemistry Implementation

44 Life Technologies™ |

Thank  You  

©  2013  Life  Technologies  Corpora8on.  All  rights  reserved.  The  trademarks  men8oned  herein  are  the  property  of  Life  Technologies  Corpora8on  and/or its affiliate(s) or  their  respec8ve  owners.  Microsoft and Windows are registered trademarks of Microsoft Corporation. Dell is a registered trademark of Dell Inc.  The  content  provided  herein  may  relate  to  products  that  have  not  been  officially  released  and  is  subject  to  change  without  no8ce.