the accuracy of electrical method for microleakage evaluation by a three-dimensional analysis
TRANSCRIPT
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4
The accuracy of electrical method for microleakageevaluation by a three-dimensional analysis
Yukiteru Iwami *, Mikako Hayashi, Fumio Takeshige, Shigeyuki Ebisu
Department of Restorative Dentistry and Endodontology, Osaka University Graduate School of Dentistry, 1-8 Yamadaoka,
Suita, Osaka 565-0871, Japan
a r t i c l e i n f o
Article history:
Received 13 June 2006
Received in revised form
25 August 2006
Accepted 6 September 2006
Keywords:
Electrical method
3D
Analysis
Microleakage
Coronal restoration
a b s t r a c t
Objectives: This in vitro study aimed to investigate the accuracy of an electrical method for
the evaluation of microleakage by a three-dimensional analysis of dye penetration.
Methods: Coronal cavities were prepared on buccal, palatal or lingual surfaces in extracted
human molars. The cavities were then filled with resin composites and were subjected to
10,000 load cycles (425 g). Before cavity preparation and after load cycling, physiological
saline was applied and wiped off, and the change in conductance was measured across the
margin of the restoration in each specimen. After dye penetration, the specimens were
reduced by 100 mm in a direction parallel to the cavity floor, from the surface of the
restoration to the cavity floor. The sequence of reducing the sections by 100 mm and image
taking was repeated. Three-dimensional images of dye penetration were made and the
proportions of the interface showing penetration were calculated.
Results: Pearson’s correlation coefficients between changes in conductance and the surface
area of dye penetration, between these and the rate of dye penetration, and between these
and the depth of dye penetration were 0.932, 0.920 and 0.732, respectively. The correlations
were significant ( p < 0.05).
Conclusions: The results of this electrical method for microleakage evaluation showed
stronger correlations with the three-dimensional amount of marginal leakage (surface area
of dye penetration and rate of dye penetration) than the two-dimensional amount (depth of
dye penetration).
# 2006 Elsevier Ltd. All rights reserved.
avai lable at www.sc iencedi rec t .com
journal homepage: www. int l .e lsev ierhea l th .com/ journals / jden
1. Introduction
Microleakage of cariogenic bacteria along the cavity wall of
restorations is reported to be an important cause of pulpal
inflammation, pulp necrosis1–3 and secondary caries4,5 in
restored teeth. The symptoms of microleakage include
postoperative sensitivity, chronic sensitivity and marginal
discoloration.6,7 However, objective judgement of whether
the cause of these symptoms is microleakage is difficult in
the clinical settings. Previously, in vitro electrical methods for
* Corresponding author. Tel.: +81 6 6879 2927; fax: +81 6 6879 2929.E-mail address: [email protected] (Y. Iwami).
0300-5712/$ – see front matter # 2006 Elsevier Ltd. All rights reservedoi:10.1016/j.jdent.2006.09.003
the evaluation of microleakage were reported and their
clinical use in the future was suggested.8–12 However, these
electrical methods evaluated the margins of restorations
after application of an electrolyte that was not wiped off, so
measurements of conductance or impedance depended on
the area of applied electrolyte.13,14 In addition, these
methods cannot be applied to restorations on dentin or
cementum surfaces, because the influence of the applied
area or amount of electrolyte leads to increased measure-
ment errors.9
d.
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4 269
Fig. 1 – Schema of load cycling. The coronal enamel was
reduced with a standardized cavity preparation device. In
load cycling, the plunger was moved repeatedly down
5 mm and the molar was moved repeatedly in a buccal
lingual direction 1.5 mm. The load was 425 g (plumb:
400 g; plunger only: 25 g) and the cycle was 60 times/min.
To resolve these problems, a new electrical method was
reported in our previous study.15 In brief, after application of
electrolyte to the margin of a restoration and wiping it away to
leave only that penetrating into the marginal gap, microleak-
age of the restoration can be detected by measurement of the
change in conductance across the margin. Using this electrical
method, microleakage in both coronal and root surface
cavities could be detected, and the results of the evaluation
correlated significantly with those of the conventional dye
penetration test.15 Moreover, this method can detect exten-
sions of microleakage in the direction of the pulp chamber of
restorations, and the size of the restorative margin did not
influence the results.16 In our previous in vitro studies,15,16
however, it was difficult to evaluate the actual three-
dimensional microleakage of restorations for comparison
with the results by the electrical method, so dentin bonding
systems were not used in composite filling and the amount of
marginal leakage was determined using the depth of the
prepared cavities. Therefore, previous studies have estab-
lished the principle of this electrical method, but the accuracy
of the results cannot be evaluated clearly when resin
composite restorations are filled with a dentin bonding system
as in the clinical settings. For future clinical use of this method
with composite restorations, it is necessary to evaluate its
results with a dentin bonding system.
Recently, a quantitative three-dimensional method for the
evaluation of contraction gap formation was reported.17 This
three-dimensional method can assess the condition and
amount of microleakage of composite restorations more
accurately than the conventional two-dimensional method.15
Using the three-dimensional method, smaller microleakages
can be detected from composite restorations with a dentin
bonding system than without the bonding system. Therefore,
the purpose of this in vitro study was to investigate the
accuracy of the electrical method for the evaluation of
marginal leakage by the three-dimensional method using
dye penetration, and to evaluate the value of this electrical
method in clinical settings.
Fig. 2 – The prepared cavity and measurement of
conductance. Coronal cavity was prepared on the buccal,
palatal or lingual surface. Each cavity was around 3 mm in
diameter and 2 mm in depth. Conductance was measured
consecutively from the starting point of measurement to
the end point, across the margin.
2. Materials and methods
2.1. Preparation of specimens
In this study, 18 extracted non-carious human molars, which
had been stored in physiological saline at 4 8C, were used
within 1 year of extraction. The molars were obtained from
Osaka University Dental Hospital in accordance with the
protocol approved by the Ethics Committee of Osaka Uni-
versity Graduate School of Dentistry. The molars were divided
into three groups (bonding group, non-bonding group and
non-restorative group, each n = 6). The roots of the molars
were reduced from the apex with a polishing machine (Ecomet
III, Buehler Ltd., Lake Bluff, IL, USA). The pulp tissues of the
molars were then removed using a round bur and a fissure bur.
The coronal enamel was reduced as shown in Figs. 1 and 2
with a standardized cavity preparation device15,16 (Itoh
Engineering Co., Kyoto, Japan) after the molars had been fixed
on the device with self-curing acrylic resin (Uni-Fast II, GC
Corporation, Tokyo, Japan). The device has three micrometers
for controlling the movement of an air-turbine handpiece with
a diamond point. A circular cavity (3 mm in diameter, 2 mm in
depth, Fig. 2) was prepared on the buccal, palatal, or lingual
surface of each molar (bonding group and non-bonding group)
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4270
using the standardized cavity preparation device after
measuring the baseline conductance (the method of measure-
ment is described in the next section). The cavities of the
molars in the bonding group were treated with a dentin
bonding system (Clearfil Liner Bond II S, Kuraray Co. Ltd.,
Osaka, Japan). This binding system included a self-etching
primer, and the procedure was as follows. One drop of Primer
A (Batch No. 00114A) and one drop of Primer B (Batch No.
00111B) were mixed, applied to the cavities for 30 s and air-
dried for 5 s. Bonding Agent (Batch No. 00186B) was applied,
spread with a gentle air-blow and cured for 20 s with a visible
light-curing unit (Optilux 401, Kerr GmbH, Karlsruhe, Ger-
many). The unit was tested with a radiometer attached to the
unit before curing. Then, a hybrid-type resin composite
(Clearfil AP-X, A3, Kuraray, Batch No. 00924A) was filled into
the cavities and cured in bulk for 40 s. The cavities of the
molars in the non-bonding group were filled with the resin
composite without treatment with the bonding system. After
the molars had been stored in physiological saline at 37 8C for
24 h, the surfaces of the restorations were finished and
polished with polishing disks (Soflex, 3M ESPE, St. Paul, MT,
USA). In the non-restorative group, the cavities were not
prepared, were not treated with the bonding system, and were
not filled with the resin composite.
2.2. Load cycling and measurement of conductance
After composite filling, all molars were subjected to mechan-
ical load cycling. The root of the molar was mounted in a metal
tube in a load cycling machine (Yuyama Irika Co., Osaka,
Japan) with self-curing acrylic resin. In this machine, the
plunger was moved repeatedly down 5 mm from the hor-
izontal plane of the molar in the direction of the tooth axis and
the molar was moved repeatedly in a buccal lingual direction
1.5 mm against the plunger (Fig. 1). The load was 425 g (plumb:
400 g; plunger only: 25 g) and the cycle was 60 times/min.
After 10,000 load cycles, the conductance of the margin of
the molar was measured by the following method. An
experimental electrode that can measure conductance under
constant pressure (tip 0.6 mm in diameter at the pointed head)
was used in this experiment; the structure of this electrode is
described in our previous study.15,16 The pulp chamber of each
molar was filled with physiological saline and a copper wire
electrode was put into it. Both of the electrodes were
connected to an impedance meter (HP4263B, Hewlett-Packard
Japan Ltd., Tokyo, Japan) with a personal computer (Endeavor,
NT2500, Epson Direct Corp., Tokyo, Japan) (Fig. 2). The method
of conductance measurement is described in our previous
studies.15,16 In brief, the starting and end points were located
on each composite surface or coronal enamel surface 1.5 mm
from the margin of the restoration. After 10 ml of physiological
saline had been applied on the margin of the restoration with a
micropipette (Varipette 4710, Eppendorf, Hamburg, Germany)
and the excess wiped off, the conductance across the margin
from the starting point to the end point was measured
continuously with the experimental electrode. When this
process was carried out, the surface of the margin was not wet
to the naked eye. The electrical frequency was 100 kHz and the
voltage was 1 V. The change in conductance was defined as the
difference between the conductance at the starting point or
the end point, whichever was larger, and the maximum
conductance excluding the starting and end points. The
change in conductance for each margin was measured 10
times and the mean (CC value) was calculated for data
analysis. As described in the preceding section, before cavity
preparation, the CC value of the enamel surface (at the
location of the cavity) was measured and calculated using the
same method (including measurement after load cycling in
the non-restorative group).
2.3. Three-dimensional analysis
After measurement of conductance and load cycling, the pulp
chambers of the molars were blocked with self-curing acrylic
resin and the molar surfaces were coated with nail varnish,
except for an area 0.5 mm around the margin. The molars
were immersed in 2 wt% methylene blue solution at 37 8C for
24 h, then taken out and the solution wiped off. The molars
were then placed on the standardized cavity preparation
device and were reduced in thickness by 100 mm in a direction
parallel to the cavity floor of the composite restorations, from
the surface of the restoration to the cavity floor. An image of
each reduced surface was taken using an operation micro-
scope (Universal S3, Carl Zeiss, Oberkochen, Germany) with a
CCD camera (Hitachi, Tokyo, Japan). The magnification was
20�. The sequence of reducing the sections by 100 mm and
image taking was repeated for all molars in the bonding group
and the non-bonding group (14–20 images of reduced sections
were taken for each cavity).
After image taking, three-dimensional analysis was carried
out. The method of this analysis has been described
previously.17 The image files were manipulated using a photo
retouching software package (Adobe Photoshop ver.5.0, Adobe
Systems, San Jose, CA, USA). Dye penetration of the sections
was marked over the images of the margins (black lines:
margin penetrated with dye; gray lines: margin not penetrated
with dye). Three-dimensional images were then created, with
black and gray lines marked over the images, using analytical
software (NIH image, National Institutes of Health, USA). The
length of the cavity wall (CL), the area of the cavity floor (CF)
and the length of the margin penetrated by dye (DL) on the
images were measured using the analytical software, and the
surface area of the cavity wall (CA), the surface area of dye
penetration into the cavity wall (DA) and the rate of dye
penetration (DR) were calculated using the following formu-
lae. In addition, the maximal depth of dye penetration (DD)
was obtained by image analysis:
CA ðmm2Þ ¼XðCL� 100 mmÞ þ CF;
DA ðmm2Þ ¼XðDL� 100 mmÞ; DR ð%Þ ¼ DA
CA� 100
2.4. Statistical analysis
For comparison of the three groups after load cycling, CC values
were analyzed by one-way ANOVA and the Games–Howell test.
For comparison between before cavity preparation and after
load cycling, CC values were analyzed by the t-test. In addition,
the relationships between CC values after load cycling and DA,
between these and DR, and between these and DD were
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4 271
Fig. 3 – Example of three-dimensional image of dye penetration after load cycling (bonding group: sample 5). Black portion
(line): cavity wall with dye penetration. Gray portion (line): cavity wall without dye penetration. Gothic line: cavity margin
and line angles. The image data of each section, which consisted of black and gray lines, are shown by the same six black
and gray lines.
analyzed by Pearson’s correlation coefficient and Fisher’s Z
transformation. The significance level was set at 5% in all
analyses.
3. Results
Examples of three-dimensional images of dye penetration in
the molars in the bonding group after load cycling are shown
in Fig. 3. Dye penetration in the bonding group was observed
on the lateral walls only, and not the cavity floor. In the
bonding group, four specimens did not show dye penetration
into dentin, while two specimens did show penetration into
dentin. However, because the direction of section was almost
parallel to the dentino-enamel junction of the specimens, it
Fig. 4 – CC values of the three groups after load cycling and
before cavity preparation. There were significant
differences in the CC values after load cycling among the
three groups ( p < 0.05). In the non-bonding group and the
bonding group, the CC values after load cycling were
significantly larger than those before cavity preparation
( p < 0.05), but in the non-restorative group, there were no
significant difference between them.
was difficult to determine accurately whether the wall was
enamel or dentin. In the non-bonding group, the dye
penetrated all cavity walls, including the cavity floor.
Fig. 4 shows the CC values of the three groups after load
cycling. These increased in the following order: non-restora-
tive group, bonding group, and non-bonding group. There
were significant differences in the CC values after load cycling
among each of the three groups (p < 0.05). In the non-bonding
group and the bonding group, the CC values after load cycling
were significantly larger than those before cavity preparation
(p < 0.05), but in the non-restorative group, there was no
significant difference between the CC values after load cycling
and those before cavity preparation.
Fig. 5 shows the relationship between CC values after load
cycling and DA. The correlation coefficient was 0.932 and there
was a significant correlation (p < 0.05). Fig. 6 shows the
relationship between CC values after load cycling and DR. The
correlation coefficient was 0.920 and there was a significant
correlation (p < 0.05). Fig. 7 shows the relationship between
CC values after load cycling and DD. The correlation coefficient
was 0.732 and there was a significant correlation (p < 0.05).
4. Discussion
In this study, the actual amount of microleakage was
determined by a three-dimensional method and the results
of the electrical evaluation of microleakage were compared
with this amount. In general, the actual amount of micro-
leakage has been determined by tracer penetration methods.
In these methods, after immersion of an extracted tooth with
restorative into a tracer solution (dye,18,19 chemical tracer,20,21
radioactive tracer22,23), the tooth has been sectioned and the
extent of penetration of tracer on one or a few sections
observed. Thus, these methods are not suitable for three-
dimensional evaluation. However, the electrical method for
the evaluation of microleakage is thought to detect three-
dimensional electrolyte penetration into the marginal gap of
composite restoratives. So, the three-dimensional method is
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4272
Fig. 5 – The relationship between CC values after load
cycling and DA (correlation coefficient: 0.932, p < 0.05).
Fig. 7 – The relationship between CC values after load
cycling and DD (correlation coefficient: 0.732, p < 0.05).
suitable for determination of the actual amount of micro-
leakage when the accuracy of electrical method is being
evaluated. In this study, the accuracy of the electrical method
was evaluated by three-dimensional parameters (CA and DR)
using restorations filled with a bonding system, as in the
clinical settings. In addition, for comparison, evaluation using
a two-dimensional parameter (DD) was undertaken. Use of the
three-dimensional method and of the bonding system to
achieve a clearer evaluation of the accuracy of the electrical
method are new in this study, compared with our previous
study.15,16
The three-dimensional method for the evaluation of
microleakage reported in our previous study was also used
in this study.17 We did not previously use dye penetration, but
it was used in this study, as in Gale et al.24 By this method, it
was difficult to evaluate the interface between the cavity wall
and the filling materials, which was parallel to the sectioning
surface.17 Therefore, in this study, when the surface area of
the cavity wall was calculated, the area of the cavity floor,
which was calculated from the actual diameter of the cavity,
was added toP
(CL � 100 mm) to overcome this weak point.
Load cycling has a significant effect on microleakage.25 In
particular, flexural load cycling has a significant effect on the
microleakage of cervical resin composite restorations.26
Flexural load cycling was used in this study before the
Fig. 6 – The relationship between CC values after load
cycling and DR (correlation coefficient: 0.920, p < 0.05).
evaluation of microleakage to increase the amount of
microleakage in each specimen as far as possible and thereby
obtain clear data. However, flexural load cycling may lead to
microfractures on the cervical region of specimens near the
margin of the composite restoration, due to the concentration
of load stresses in this region. It is therefore possible that, in
this experiment, electrolyte penetrating into microfractures
may have been detected as microleakage. As this is a weak
point of this experiment, future investigations addressing this
are necessary.
The CC values of the non-restorative group after load cycling
were significantly smaller than those of the bonding group and
the non-bonding group (p < 0.05, Fig. 4). In addition, in the non-
bonding group and the bonding group, the CC values after load
cycling were significantly larger than those before cavity
preparation (p < 0.05, Fig. 4). From these results, the electrical
method can be seen to detect microleakage. There was a
significant correlation between CC values after load cycling and
DA, DR and DD (p < 0.05), and the correlation between them
was strong (Figs. 5–7). It is therefore considered that the two- or
three-dimensional amount of microleakage can be detected by
measurement of change in conductance with the electrical
method. In some previous electrical methods,13,14 electrolyte
was applied to the whole margin of restorations and was not
wiped off, and the conductance or impedance was then
measured. Consequently, electrolyte other than that which
had penetrated into the marginal gap could have influenced the
results. In this study, because the excess electrolyte on the
margin of the restorations was wiped off, the results were not
influenced by excess electrolyte and showed microleakage
more clearly than previous methods.13,14 The results showed
that this electrical method can detect the degree of micro-
leakage into the interface betweenrestorationsand the dentinal
cavitywall,and that the sensitivity of themethod isveryhigh. In
addition, the correlation coefficient between CC values and the
surface area of dye penetration (DA) and between CC values and
the rate of dye penetration (DR) (three-dimensional amount)
were larger than that between CC values and the depth of dye
penetration (DD) (two-dimensional amount). Accordingly, the
results are thought to depend on three-dimensional penetra-
tion of electrolyte. However, because the depth of dye
penetration in this experiment was measured from many
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4 273
sectional images, compared with one or a few in previous
studies,18,19 the values were maximal on many images. Thus,
there was a significant correlation even between CC values and
the two-dimensional amount of dye penetration (DD) in this
experiment.
The demonstration that this electrical method can detect
microleakage of composite restorations with a dentin bonding
system has possible clinical relevance. In future clinical
applications, it is thought that a buccal clip might be placed
into the oral cavity as an electrode, and the CC values of the
margin of composite restorations might be measured. If this
electrical method is applied clinically, the amount of micro-
leakage of restorations will be determined without their
destruction. In addition, the CC values of margins exposed to
microleakage were correlated with the three-dimensional
amount of dye penetration in this experiment, but the CC
values of margins not exposed to microleakage are thought to
be smaller than those of margins exposed to microleakage. It
may be possible to detect the location of microleakage by this
electrical method in future clinical use. Moreover, as this
method does not involve the destruction of restorations, it is
possible that the progress of microleakage could be monitored
over time. It is thought that this method can be applied to
ceramic inlay restorations and glass-ionomer restorations in
addition to composite restorations, because these are highly
insulated. However, this method cannot be applied to metal
restorations, because the electric current would flow in the
metal itself in addition to the electrolyte.
Before this electrical method can be applied clinically, it
will be necessary to further examine the optimal frequency
and voltage for minimization of external noise. Higher voltage
minimizes external noise,9 but if the voltage is very high,
irreversible or reversible pulpitis may occur.27 Other factors
such as pulp tissue, pulpal pressure, dentin condition, the
moving speed of the electrode, and the optimal electrolyte and
application method must be investigated.
5. Conclusion
In this study, the accuracy of an electrical method for the
evaluation of microleakage was investigated by three-dimen-
sional analysis of dye penetration into the cavity wall of
restorations. The correlation between changes in conductance
(CC values) after load cycling and the surface area of dye
penetration into the cavity wall (DA), and between CC values
and the rate of dye penetration (DR), were stronger than the
correlation between CC values and the depth of dye penetra-
tion (DD). Therefore, the results of this electrical method for
microleakage evaluation were correlated more strongly with
the three-dimensional amount of marginal leakage (DA and
DR) than the two-dimensional amount (DD).
Acknowledgments
This work formed part of the 21st Century COE entitled
‘Origination of Frontier BioDentistry’ held at Osaka University
Graduate School of Dentistry and supported by the Ministry of
Education, Culture, Sports, Science and Technology. This
investigation was also supported, in part, by a Grant-in-Aid for
Scientific Research (17591990) from the Japan Society for the
Promotion of Science.
r e f e r e n c e s
1. Schmeiser R, Gulzow HJ. The influence of luting cements onthe fermentation of sucrose by oral microorganisms. Journalof Dentistry 1999;27:37–41.
2. Camps J, Dejou J, Remusat M, About I. Factors influencingpulpal response to cavity restorations. Dental Materials2000;16:432–40.
3. Murray PE, About I, Franquin JC, Remusat M, Smith AJ.Restorative pulpal and repair responses. Journal of AmericanDental Association 2001;132:482–91.
4. Ben-Amer A, Cardash HS, Judes H. The sealing of the tooth/amalgam interface by corrosion products. Journal of OralRehabilitation 1995;22:101–4.
5. Moreira Jr G, Sobrinho APR, Nicoli JR, Bambirra EA, FariasLM, Carvalho MAR, et al. Evaluation of microbial infiltrationin restored cavities—an alternative method. Journal ofEndodontics 1999;25:605–8.
6. Going RE. Microleakage around dental restorations: asummarizing review. Journal of American Dental Association1972;84:1349–57.
7. Triadan H. When is microleakage a real clinical problem?Operative Dentistry 1987;12:153–7.
8. Martinez CR, Greener EH. Utilization of electricalconductivity as an alternative method of assessing marginalleakage of pit and fissure sealants. Journal of OralRehabilitation 1976;3:69–74.
9. Nakano Y. A new electrical testing method on marginalleakage of composite resin restorations. Japan Journal ofConservative Dentistry 1985;28:1183–98.
10. Momoi Y, Iwase H, Nakano Y, Kohno A, Asanuma A,Yanagisawa K. Gradual increases in marginal leakage ofresin composite restorations with thermal stress. Journal ofDental Research 1990;69:1659–63.
11. Verdonschot EH, Rondel P, Huysmans MCDNJM. Validity ofelectrical conductance measurements in evaluating themarginal integrity of sealant restorations. Caries Research1995;29:100–6.
12. Von Fraunhofer JA, Adachi EI, Barnes DM, Romberg E. Theeffect of tooth preparation on microleakage behavior.Operative Dentistry 2000;25:526–33.
13. Hoppenbrouwers PMM, Scholberg HPF, Borggreven JMPM.Measurement of the permeability of dental enamel and itsvariation with depth using an electrical method. Journal ofDental Research 1986;65:154–7.
14. Huysmans MCDNJM, Verdonschot EH, Rondel P. Electricalconductance and electrode area on sound smooth enamelin extracted teeth. Caries Research 1995;29:88–93.
15. Iwami Y, Yamamoto H, Ebisu S. A new electrical method fordetecting marginal leakage of in vitro resin restorations.Journal of Dentistry 2000;28:241–7.
16. Iwami Y, Yamamoto H, Hayashi M, Takeshige F, Ebisu S.Validity of electrical conductance measurements inevaluating marginal leakage around resin compositerestorations. Operative Dentistry 2002;27:606–12.
17. Iwami Y, Shimizu A, Hayashi M, Takeshige F, Ebisu S. Three-dimensional evaluation of gap formation of cervicalrestorations. Journal of Dentistry 2005;33:325–33.
18. Kubo S, Yokota H, Yokota H, Hayashi Y. The effect oflight-curing modes on the microleakage of cervicalresin composite restorations. Journal of Dentistry2004;32:247–54.
j o u r n a l o f d e n t i s t r y 3 5 ( 2 0 0 7 ) 2 6 8 – 2 7 4274
19. Owens BM, Johnson WW, Harris EF. Marginal permeabilityof self-etch and total-etch adhesive systems. OperativeDentistry 2006;31:60–7.
20. Ramos RP, Chimello DT, Chinelatti MA, Dibb RG, Mondelli J.Effect of three surface sealants on marginal sealing of classV composite resin restorations. Operative Dentistry2000;25:448–53.
21. Tulunoglu o. uctash M, Alacam A, omurlu H. Microleakageof light-cured resin and resin-modified glass-ionomerdentin bonding agents applied with co-cure vs.pre-cure technique. Operative Dentistry 2000;25:292–8.
22. Powis DR, Prosser HJ, Shortall AC, Wilson AD. Long-termmonitoring of microleakage of composites. Part I:Radiochemical diffusion technique. Journal of ProstheticDentistry 1988;60:304–7.
23. Puckett AD, Fitchie JG, Bennett B, Hembree JH. Microleakageand thermal properties of hybrid ionomer restoratives.Quintessence International 1995;26:577–81.
24. Gale MS, Darvell BW, Cheung GSP. Three-dimensionalreconstruction of microleakage pattern using a sequentialgrinding technique. Journal of Dentistry 1994;22:370–5.
25. Jang KT, Chung DH, Shin D, Garcıa-Godoy F. Effect ofeccentric load cycling on microleakage of class V flowableand packable composite resin restorations. OperativeDentistry 2001;26:603–8.
26. Kubo S, Yokota H, Sata Y, Hayashi Y. The effect of flexuralload cycling on the microleakage of cervical resincomposites. Operative Dentistry 2001;26:451–9.
27. Ricketts DNJ, Kidd EAM, Wilson RF. A re-evaluation ofelectric resistance measurements for the diagnosis ofocclusal caries. British Dental Journal 1995;178:11–7.