The Effect of Ratio of Meldola’s Blue to Carbon Powders and Carbon Pas te on Detection of Hydrogen Peroxide for the Carbon Paste Electrode Mo dified with Meldola’s Blue Chi-Wen Lo ( 羅羅羅 ) , Chih-Ying Wu ( 羅羅羅 ) , Hau Lin ( 羅羅 ) Department of Chemical and Materials Engineering, Southern Taiwan University 南南南南南南南南南南南南南南南南 ABSTRACT . The glucose and oxygen can be catalyzed by the glucose oxidase to produce the gluconic acid and hydrogen peroxide, and therefore, as the concentration of hydrogen peroxide can be determined, the concentration of glucose can also be determined. A study was conducted to use the Meldola’s Blue to modify the carbon paste electrode. Because the Meldola’s Blue possesses the excellent catalytic characteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to el evate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solut ion(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be d etermined. At 30℃, -300 mV operating potential, 500 rpm stirring rate and in 0.05 M phosphate buffer solution (PBS), the TB (Time Bas e) graphs for the carbon paste electrode at different ratios of Meldola’s Blue to carbon powders and carbon paste were plotted to ev aluate the effect of the ratio of Meldola’s Blue to carbon powders and carbon paste on the responding current of detection of hyd rogen peroxide. At the optimum operating conditions -300 mV operating potential, 500 rpm stirring rate and in 0.05 M PBS buffer soluti on ( pH = 7.4 ) , when the carbon paste electrode was modified with Meldola’s Blue [Meldola’s Blue : carbon powders : carbon paste = 2 15 15 ( weight ratio )] , the detection limit was 0.02 mM H 2 O 2 , the linear range was 0.02 1.4 mM H 2 O 2 , R 2 = 0.9994 , and the sensitiv ity was 159.39 µA/cm 2 mM H 2 O 2 . Both the detecting techniques of hydrogen peroxide and glucose ar e important. The glucose and oxygen can be catalyzed by the glucose oxid ase to produce the gluconic acid and hydrogen peroxide, and therefore, a s the concentration of hydrogen peroxide can be determined, the concentr ation of glucose can also be determined. Because the Meldola’s Blue poss esses the excellent catalytic characteristic it can be used with the car bon paste and carbon powders which possess the high conductivity to make the carbon paste electrode and to elevate the responding current of hydr ogen peroxide. The responding current of hydrogen peroxide is detected i n the phosphate buffer solution (PBS), and then the concentration of hyd rogen peroxide can be determined. At 30℃, -300 mV operating potential, 500 rpm stirring rate and in 0.05 M phosphate buffer solution (PBS), the TB (Time Base) graphs for the carbon paste electrode at different ratios of Meldola’s Blue to carbon powders and carbon paste were plotted to evaluate the effect of ratio of Meldola’s Blue to carbon powders and carbon paste on the responding current of detection of hydrogen peroxi de. The optimum operating conditions were used to detect the detection l imit , linear range and sensitivity of detection of hydrogen peroxide fo r the carbon paste electrode modified with Meldola’s Blue. INTRODUCTION EXPERIMENTAL 1. Equipment Electrochemical Analyzer (CHI 401A, CH Instruments, Inc) was used to measure the activity of electrode by Cyclic Voltammetry ( CV ) and Time Base ( TB ) mode ; Oven (DENG YNG) ; Electric Stirrer(Fa rgo) ; pH meter (Metrohm 731); Constant Temperature Thermal Bath (Wi sdom BC-2DT 10L); Carbon Paste Electrode was used as the working ele ctrodes, Coiled Platinum Wire was used as the counter electrode and Ag / AgCl was used as the reference electrode. 2. Chemicals and Reagents Meldola’s Blue ; Hydrochloric Acid (HCl); Sodium Hydroxide (N aOH) ; Hydrogen Peroxide (H 2 O 2 ); Graphite Carbon Powder( C ); Carbon Paste; Cyclohexanone(C 6 H 10 O); Potassium Dihydrogenphosphate (KH 2 PO 4 ); Potassium Chloride (KCl). 3. Preparation of the Carbon Paste Electrode (1) Take one section of 7 cm electric wire with 0.05 cm inside diameter. After depriving the coating 0.5 cm length from both ends, the nake-end ed wire was washed, dried and ready for use. 7 cm 0.05 cm 0.5 cm (2) Then the Meldola’s Blue powders, graphite carbon powders and c arbon paste were mixed with the appropriate ratio [Meldola’s Blue : carbon powders : carbon paste = 2 15 15 ( weight ratio )]. Aft er the mixing was complete, the mixture was evenly coated on the na ke-ended electric wire and dried in the oven and then we obtained the carbon paste electrode. Meldola’s Blue Carbon Powders Mixing with Carbon Paste DEPARTMENT OF CHEMICAL AND MATERIALS ENGINEERING, SOUTHERN TAIWAN UNIVERSITY RESULTS AND DISCUSSION Table 1 The sensitivities, respond ing currents and R 2 values of differen t ratios of Meldola’s Blue to carbon powders a nd carbon paste for the carbon pas te electrode modified with Meldola ’s Blue Fig 1. CV graphs for (A) unmodi fied carbon paste electrode( the range of scanning potential: -0. 8 +0.8 V) (B) carbon paste ele ctrode modified with Meldola’s Blue ( the range of scanning pot ential: -0.8 +0.8 V) Fig. 2 The TB graphs of carbo n paste electrodes for detectio n of H 2 O 2 at different ratios of Meldola’s Blue to carbon pow ders and carbon paste ; [Meldola’s Blue : carbon powder s : carbon paste = (A) 0 15 15 (B) 1 15 15 (C) 2 15 15 (D) 4 15 15 (E) 6 15 15 ] Fig. 3 The calibration curves of diffe rent ratios of Meldola’s Blue to ca rbon powders and carbon paste ; [Meldola’s Blue : carbon powders : car bon paste = [ (A) 0 15 15 (B) 1 15 15 (C) 2 15 15 (D) 4 15 15 (E) 6 15 15 ] Fig. 5 The TB graphs of carb on paste electrode for determ ining the linear range of H 2 O 2 [Meldola’s Blue : carbon pow ders : carbon paste = 2 15 15 ( weight ratio )] CONCLUSIONS The results showed that the responding current for the carbon paste e lectrode modified with Meldola’s Blue was elevated significantly. The TB (Time Base ) graphs at different ratios of Meldola’s Blue to carbon powders and ca rbon paste were plotted to evaluate the effect of the ratio of Meldola’s Blue to carbon powders and carbon paste on the responding current of detection of hydrogen peroxide and determine the optimum operating conditions. The results showed that at the optimum operating conditions –300mV operating potential, 500 rpm stirring rate and in 0.05M phosphate buffer solution(pH=7.4) [Meldola’s Bl ue : carbon powders : carbon paste = 2 15 15 ( weight ratio )] , the detection l imit was 0.02 mM H 2 O 2 , the linear range was 0.02~1.4 mM H 2 O 2 , R 2 =0.9994, and the sensitivity was 159.39μA/cm 2 羅mM H 2 O 2 . REFERENCES 1.Kulys, J., Biosensors and Bioelectronics, 1999, 14, 473-479. 2. Mao, L. ; Yamamoto, K., Talanta, 2000, 51, 187-195. 3. Singh, S. ; Chaubey, A. ; Malhotra, B. D., Analytica Chimica Acta, 2004, 502, 229-234. 4. Liu, Y. ; Lei, J. ; Ju, H., Talanta, 2008, 74, 965-970. 5. Lai, G.-S. ; Zhang, H.-L. ; Han, D.-Y., Sensors and Actuators B, 20 08, 129, 497-503. Fig. 4 The TB graphs of carbon paste electrode electrodes for determining the detection limit of H 2 O 2 [Meldola’s Blue : carbon powders : carbon paste = 2 15 15 ( weight ratio )]

Upload: nancy-doyle

Post on 19-Jan-2016

217 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: The Effect of Ratio of Meldola’s Blue to Carbon Powders and Carbon Paste on Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with

The Effect of Ratio of Meldola’s Blue to Carbon Powders and Carbon Paste on Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Meldola’s Blue

Chi-Wen Lo ( 羅濟玟 ) , Chih-Ying Wu ( 巫致穎 ) , Hau Lin ( 林浩 )

Department of Chemical and Materials Engineering, Southern Taiwan University 南台科技大學化學工程與材料工程系

ABSTRACT.

The glucose and oxygen can be catalyzed by the glucose oxidase to produce the gluconic acid and hydrogen peroxide, and therefore, as the concentration of hydrogen peroxide can be determined, the concentration of glucose can also be determined. A study was conducted to use the Meldola’s Blue to modify the carbon paste electrode. Because the Meldola’s Blue possesses the excellent catalytic characteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be determined. At 30℃ , -300 mV operating potential, 500 rpm stirring rate and in 0.05 M phosphate buffer solution (PBS), the TB (Time Base) graphs for the carbon paste electrode at different ratios of Meldola’s Blue to carbon powders and carbon paste were plotted to evaluate the effect of the ratio of Meldola’s Blue to carbon powders and carbon paste on the responding current of detection of hydrogen peroxide. At the optimum operating conditions -300 mV operating potential, 500 rpm stirring rate and in 0.05 M PBS buffer solution ( pH = 7.4 ) , when the carbon paste electrode was modified with Meldola’s Blue [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )] , the detection limit was 0.02 mM H2O2 , the linear range was 0.02 ~ 1.4 mM H2O2 , R2 = 0.9994 , and the sensitivity was 159.39 µA/cm2 . mM H2O2 .

Both the detecting techniques of hydrogen peroxide and glucose are important. The glucose and oxygen can be catalyzed by the glucose oxidase to produce the gluconic acid and hydrogen peroxide, and therefore, as the concentration of hydrogen peroxide can be determined, the concentration of glucose can also be determined. Because the Meldola’s Blue possesses the excellent catalytic characteristic it can be used with the carbon paste and carbon powders which possess the high conductivity to make the carbon paste electrode and to elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution (PBS), and then the concentration of hydrogen peroxide can be determined. At 30 , -300 mV operating potential, 500 rpm stir℃ring rate and in 0.05 M phosphate buffer solution (PBS), the TB (Time Base) graphs for the carbon paste electrode at different ratios of Meldola’s Blue to carbon powders and carbon paste were plotted to evaluate the effect of ratio of Meldola’s Blue to carbon powders and carbon paste on the responding current of detection of hydrogen peroxide. The optimum operating conditions were used to detect the detection limit , linear range and sensitivity of detection of hydrogen peroxide for the carbon paste electrode modified with Meldola’s Blue.

INTRODUCTION

EXPERIMENTAL

1. Equipment Electrochemical Analyzer (CHI 401A, CH Instruments, Inc) was used to measure the activity of electrode by Cyclic Voltammetry ( CV ) and Time Base ( TB ) mode ; Oven (DENG YNG) ; Electric Stirrer(Fargo) ; pH meter (Metrohm 731); Constant Temperature Thermal Bath (Wisdom BC-2DT 10L); Carbon Paste Electrode was used as the working electrodes, Coiled Platinum Wire was used as the counter electrode and Ag / AgCl was used as the reference electrode.

2. Chemicals and Reagents

Meldola’s Blue ; Hydrochloric Acid (HCl); Sodium Hydroxide (NaOH) ; Hydrogen Peroxide (H2O2); Graphite Carbon Powder( C ); Carbon Paste; Cyclohexanone(C6H10O); Potassium Dihydrogenphosphate (KH2PO4); Potassium Chloride (KCl).

3. Preparation of the Carbon Paste Electrode

(1) Take one section of 7 cm electric wire with 0.05 cm inside diameter. After depriving the coating 0.5 cm length from both ends, the nake-ended wire was washed, dried and ready for use.

7 cm

0.05 cm

0.5 cm

(2) Then the Meldola’s Blue powders, graphite carbon powders and carbon paste were mixed with the appropriate ratio [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )]. After the mixing was complete, the mixture was evenly coated on the nake-ended electric wire and dried in the oven and then we obtained the carbon paste electrode.

Meldola’s Blue

Carbon Powders

Mixing with Carbon Paste

DEPARTMENT OF CHEMICAL AND MATERIALS ENGINEERING, SOUTHERN TAIWAN UNIVERSITY

RESULTS AND DISCUSSION

Table 1 The sensitivities, responding currents and R2 values of different ratios of Meldola’s Blue to carbon powders and carbon paste for the carbon paste electrode modified with Meldola’s Blue

Fig 1. CV graphs for (A) unmodified carbon paste electrode( the range of scanning potential: -0.8 ~ +0.8 V) (B) carbon paste electrode modified with Meldola’s Blue ( the range of scanning potential: -0.8 ~ +0.8 V)

Fig. 2 The TB graphs of carbon paste electrodes for detection of H2O2 at different ratios of Meldola’s Blue to carbon powders and carbon paste ;[Meldola’s Blue : carbon powders : carbon paste = (A) 0 : 15 : 15 (B) 1 : 15 : 15 (C) 2 : 15 : 15 (D) 4 : 15 : 15 (E) 6 :15 : 15 ]

Fig. 3 The calibration curves of different ratios of Meldola’s Blue to carbon powders and carbon paste ;[Meldola’s Blue : carbon powders : carbon paste = [ (A) 0 : 15 : 15 (B) 1 : 15 : 15 (C) 2 : 15 : 15 (D) 4 : 15 : 15 (E) 6 : 15 : 15 ]

Fig. 5 The TB graphs of carbon paste electrode for determining the linear range of H2O2 [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )]

CONCLUSIONS

The results showed that the responding current for the carbon paste electrode modified with Meldola’s Blue was elevated significantly. The TB (Time Base ) graphs at different ratios of Meldola’s Blue to carbon powders and carbon paste were plotted to evaluate the effect of the ratio of Meldola’s Blue to carbon powders and carbon paste on the responding current of detection of hydrogen peroxide and determine the optimum operating conditions. The results showed that at the optimum operating conditions –300mV operating potential, 500rpm stirring rate and in 0.05M phosphate buffer solution(pH=7.4) [Meldola’s Blue :

carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )] , the detection limit was 0.02 mM H2O2, the linear range was 0.02~1.4 mM H2O2 , R2=0.9994, and the sensitivity was 159.39μA/cm2 ּmM H2O2 .

REFERENCES

1. Kulys, J., Biosensors and Bioelectronics, 1999, 14, 473-479.

2. Mao, L. ; Yamamoto, K., Talanta, 2000, 51, 187-195.

3. Singh, S. ; Chaubey, A. ; Malhotra, B. D., Analytica Chimica Acta, 2004, 502, 229-234.

4. Liu, Y. ; Lei, J. ; Ju, H., Talanta, 2008, 74, 965-970.

5. Lai, G.-S. ; Zhang, H.-L. ; Han, D.-Y., Sensors and Actuators B, 2008, 129, 497-503.

Fig. 4 The TB graphs of carbon paste electrode electrodes for determining the detection limit of H2O2 [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )]