the result of goldendom calibration hiroko miyamoto t.noda, s.yoshida, k.mase k.hoshina, m.krasberg,...

40
The result of GoldenDOM The result of GoldenDOM Calibration Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Post on 19-Dec-2015

218 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

The result of GoldenDOM The result of GoldenDOM CalibrationCalibration

Hiroko MiyamotoT.Noda, S.Yoshida, K.Mase

K.Hoshina, M.Krasberg, K.Hanson, C.Rott

Chiba University, UW Madison, PSU

Page 2: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Who is doing what?

Chiba University Hiroko, Keiichi, Tsuyoshi, Shigeru

UW-MadisonKael, Mark, Chris and FAT people…

PSU

Carsten, Martin

Page 3: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Contents

Golden PMT calibration @ Chiba   Golden DOM calibration @ Chiba (useful LC circuit!) DOM vs PMT? Schedule

Page 4: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Golden PMT/DOM production and calibration

2004,Nov

Hamamatsu

Chiba

Madison

SouthPole

Hamamatsu PMT QE calibration

2005,Jun. Jul. Aug. Sep. Oct. Nov. Dec.

Chiba PMT calibration

FAT#8

Chiba DOM calibration

2005/06 deployment

Page 5: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

IceCube Golden PMT/DOM calibrations at Chiba

[PMT] Low Temperature (-32C) Gain Waveform Noise Rate

Room Temperature Relative Collection Efficiency (2π) Absolute QExCE Efficiency (1 slice)

http://www.ppl.phys.chiba-u.jp/research/IceCube/pmt/GoldenPMT/for GoldenPMT calibration 2005

[DOM]Room Temperature Relative Collection Efficiency (4π) Absolute Efficiency (1 point)

Page 6: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Hakkenden (“eight” dog) Warriors….? DOM name PMT #Sosuke_Inukawa : TA1181 --failedDosetsu_Inuyama : TA1069 Shino_Inuzuka : TA1052Genpachi_Inukai : TA1167Kobungo_Inuta : TA1062Keno_Inuzaka : TA1056 --failedShinbei_Inue : TA1059Daikaku_Inuyama : TA1176 --failed

Page 7: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2D PMT Scan

Data library available Implemented in the detector MC

Page 8: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2D CE map ITA1052

TA1059 TA1062

Page 9: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2D CE map II

TA1069 TA1167

Page 10: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

PMT absolute calibration

Page 11: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Chamber and DAQ

ATTENUATOR

PressuremeterTemp

monitor

Page 12: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Energy meter Numerical calculation

Typically,

~ 4x109 x 5x10-11 ~ 0.2

QExCE(absolute efficiency) = photoelectron # / cathode hit photon #

cathode hit photon # ~ initial photon# x aperture x Cross Section

Estimate incidental photon #.

Page 13: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Pedestal subtraction

Page 14: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2DCE combined with 2D absolute QExCE

Page 15: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

HAMAMATSU QE vs Chiba QExCE

Page 16: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Golden DOM calibration

Page 17: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

How useful external trigger is? SPE-based measurement SPE-based study is very important and fundamental, but we can’t see SPE signals from light source because of the bunch of noise Control trigger rate You don’t need to take 10 kHz dark noises anymore but signals at a rate just as

you like Standalone operation DOM can take the data from the light source by itself alternatively (you don’t need other DOMs or MB anymore) After pulse measurement You can trigger anytime you like, which enables you to see after pulses

Page 18: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Absolute efficiency measurement for DOM

Same setup as for PMT (with 337nm N2 laser), except:

2000 incidental photons (increase the laser luminosity)

N2 laserN2

Page 19: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Shino_InuzukaShino_Inuzuka POS.1

POS.2 POS.3

POS.1

POS.2 POS.3

DOM surface

~2cm

Page 20: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

PMT- ID PMT- Qecenter[%] HPK- QE[%] DOM- name QE- pos1[%] QE- pos2[%] QE- pos3[%] DOM- QEcenter[%]TA1052 21.1 20.97 Shino_Inuzuka 7.9 8.3 8.3 8.2TA1059 19.2 18.86 Shinbei_Inue 6.9 6.9 6.9 6.9TA1062 20.5 20.8 Kobungo_Inuta 9.7 9.6 9.5 9.6TA1069 20.0 20.3 Dosetsu_Inuyama 8.3 8.4 8.5 8.4TA1167 19.4 18.89 Genpachi_Inukai 7.7 7.7 8.0 7.8

HPQ(340nm) vs CHIBA(337nm)

y = 1.2448x - 4.9814

18.5

19

19.5

20

20.5

21

21.5

19 20 21 22

HPK- QE[%]

CHIB

A-PM

T-QE

cen

ter[%

]

1系列 ( 1)線形 系列

PMT- Eff. vs DOM- Eff.

0.0

2.0

4.0

6.0

8.0

10.0

12.0

19 19.5 20 20.5 21 21.5

CHIBA PMT Eff.[%]

DOM

Eff.[%

]

Shino_InuzukaShinbei_InueKobungo_InutaGenpachi_InukaiGenpachi_Inukai

PMT-Eff. vs DOM-Eff.

0.0

2.0

4.0

6.0

8.0

10.0

12.0

19 19.5 20 20.5 21 21.5

CHIBA PMT Eff.[%]

DO

M E

ff.[%

] Shino_InuzukaShinbei_InueKobungo_InutaGenpachi_InukaiGenpachi_Inukai

PMTs vs DOMs (Preliminary)

Page 21: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

DOM efficiency

+

Acceptance()

Glass/gel transmittance

Page 22: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

DOM 2D 4 scan

Map out the detectionefficiency

Combine DOM absolute efficiency and DOM 2D scanning map!

Page 23: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2D Scanning System for DOM

Page 24: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Dosetsu_Inuyama

Very very preliminary!

Page 25: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

By the way…

Quad-LC signal works very well as a event trigger.

Cosmic μ(bkg)

Rayleigh scatteredPhoton signal peak

Externally triggeredLED signal,no bkg

Page 26: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

How about next year?

We are providing another set of 8 Golden DOMs next year.Chiba calibrate and ship them to UW via Hamamatsu in May 2006 for FAT in May or June.

Page 27: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Any idea for this year?(with 5 Golden DOMs…)

For example….

Discussion

Ice Layer

Page 28: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU
Page 29: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Genpachi_Inukai

Page 30: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Shinbei_Inue

Page 31: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU
Page 32: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Kobungo_Inuta

Page 33: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Dosetsu_Inuyama

Page 34: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Kobungo_Inuta

Page 35: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Shinbei_Inue

Page 36: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Shino_InuzukaShino_Inuzuka

Page 37: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

Genpachi_Inukai

Page 38: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2D CE map ITA1052

TA1059 TA1062

Page 39: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

2D CE map IITA1069 TA1167

Page 40: The result of GoldenDOM Calibration Hiroko Miyamoto T.Noda, S.Yoshida, K.Mase K.Hoshina, M.Krasberg, K.Hanson, C.Rott Chiba University, UW Madison, PSU

HPQ(340nm) vs CHIBA(337nm)

y = 1.2448x - 4.9814

16.5

17.5

18.5

19.5

20.5

21.5

22.5

23.5

19 19.5 20 20.5 21 21.5

HPK-QE[%]

CH

IBA

-PM

T-Q

E c

ente

r[%]

HPK vs CHIBA