The result of GoldenDOM The result of GoldenDOM CalibrationCalibration
Hiroko MiyamotoT.Noda, S.Yoshida, K.Mase
K.Hoshina, M.Krasberg, K.Hanson, C.Rott
Chiba University, UW Madison, PSU
Who is doing what?
Chiba University Hiroko, Keiichi, Tsuyoshi, Shigeru
UW-MadisonKael, Mark, Chris and FAT people…
PSU
Carsten, Martin
Contents
Golden PMT calibration @ Chiba Golden DOM calibration @ Chiba (useful LC circuit!) DOM vs PMT? Schedule
Golden PMT/DOM production and calibration
2004,Nov
Hamamatsu
Chiba
Madison
SouthPole
Hamamatsu PMT QE calibration
2005,Jun. Jul. Aug. Sep. Oct. Nov. Dec.
Chiba PMT calibration
FAT#8
Chiba DOM calibration
2005/06 deployment
IceCube Golden PMT/DOM calibrations at Chiba
[PMT] Low Temperature (-32C) Gain Waveform Noise Rate
Room Temperature Relative Collection Efficiency (2π) Absolute QExCE Efficiency (1 slice)
http://www.ppl.phys.chiba-u.jp/research/IceCube/pmt/GoldenPMT/for GoldenPMT calibration 2005
[DOM]Room Temperature Relative Collection Efficiency (4π) Absolute Efficiency (1 point)
Hakkenden (“eight” dog) Warriors….? DOM name PMT #Sosuke_Inukawa : TA1181 --failedDosetsu_Inuyama : TA1069 Shino_Inuzuka : TA1052Genpachi_Inukai : TA1167Kobungo_Inuta : TA1062Keno_Inuzaka : TA1056 --failedShinbei_Inue : TA1059Daikaku_Inuyama : TA1176 --failed
2D PMT Scan
Data library available Implemented in the detector MC
2D CE map ITA1052
TA1059 TA1062
2D CE map II
TA1069 TA1167
PMT absolute calibration
Chamber and DAQ
ATTENUATOR
PressuremeterTemp
monitor
Energy meter Numerical calculation
Typically,
~ 4x109 x 5x10-11 ~ 0.2
QExCE(absolute efficiency) = photoelectron # / cathode hit photon #
cathode hit photon # ~ initial photon# x aperture x Cross Section
Estimate incidental photon #.
Pedestal subtraction
2DCE combined with 2D absolute QExCE
HAMAMATSU QE vs Chiba QExCE
Golden DOM calibration
How useful external trigger is? SPE-based measurement SPE-based study is very important and fundamental, but we can’t see SPE signals from light source because of the bunch of noise Control trigger rate You don’t need to take 10 kHz dark noises anymore but signals at a rate just as
you like Standalone operation DOM can take the data from the light source by itself alternatively (you don’t need other DOMs or MB anymore) After pulse measurement You can trigger anytime you like, which enables you to see after pulses
Absolute efficiency measurement for DOM
Same setup as for PMT (with 337nm N2 laser), except:
2000 incidental photons (increase the laser luminosity)
N2 laserN2
Shino_InuzukaShino_Inuzuka POS.1
POS.2 POS.3
POS.1
POS.2 POS.3
DOM surface
~2cm
PMT- ID PMT- Qecenter[%] HPK- QE[%] DOM- name QE- pos1[%] QE- pos2[%] QE- pos3[%] DOM- QEcenter[%]TA1052 21.1 20.97 Shino_Inuzuka 7.9 8.3 8.3 8.2TA1059 19.2 18.86 Shinbei_Inue 6.9 6.9 6.9 6.9TA1062 20.5 20.8 Kobungo_Inuta 9.7 9.6 9.5 9.6TA1069 20.0 20.3 Dosetsu_Inuyama 8.3 8.4 8.5 8.4TA1167 19.4 18.89 Genpachi_Inukai 7.7 7.7 8.0 7.8
HPQ(340nm) vs CHIBA(337nm)
y = 1.2448x - 4.9814
18.5
19
19.5
20
20.5
21
21.5
19 20 21 22
HPK- QE[%]
CHIB
A-PM
T-QE
cen
ter[%
]
1系列 ( 1)線形 系列
PMT- Eff. vs DOM- Eff.
0.0
2.0
4.0
6.0
8.0
10.0
12.0
19 19.5 20 20.5 21 21.5
CHIBA PMT Eff.[%]
DOM
Eff.[%
]
Shino_InuzukaShinbei_InueKobungo_InutaGenpachi_InukaiGenpachi_Inukai
PMT-Eff. vs DOM-Eff.
0.0
2.0
4.0
6.0
8.0
10.0
12.0
19 19.5 20 20.5 21 21.5
CHIBA PMT Eff.[%]
DO
M E
ff.[%
] Shino_InuzukaShinbei_InueKobungo_InutaGenpachi_InukaiGenpachi_Inukai
PMTs vs DOMs (Preliminary)
DOM efficiency
+
Acceptance()
Glass/gel transmittance
DOM 2D 4 scan
Map out the detectionefficiency
Combine DOM absolute efficiency and DOM 2D scanning map!
2D Scanning System for DOM
Dosetsu_Inuyama
Very very preliminary!
By the way…
Quad-LC signal works very well as a event trigger.
Cosmic μ(bkg)
Rayleigh scatteredPhoton signal peak
Externally triggeredLED signal,no bkg
How about next year?
We are providing another set of 8 Golden DOMs next year.Chiba calibrate and ship them to UW via Hamamatsu in May 2006 for FAT in May or June.
Any idea for this year?(with 5 Golden DOMs…)
For example….
Discussion
Ice Layer
Genpachi_Inukai
Shinbei_Inue
Kobungo_Inuta
Dosetsu_Inuyama
Kobungo_Inuta
Shinbei_Inue
Shino_InuzukaShino_Inuzuka
Genpachi_Inukai
2D CE map ITA1052
TA1059 TA1062
2D CE map IITA1069 TA1167
HPQ(340nm) vs CHIBA(337nm)
y = 1.2448x - 4.9814
16.5
17.5
18.5
19.5
20.5
21.5
22.5
23.5
19 19.5 20 20.5 21 21.5
HPK-QE[%]
CH
IBA
-PM
T-Q
E c
ente
r[%]
HPK vs CHIBA