commissioning lhc beam instrumentation for nominal bunch intensities beam commissioning working...

Post on 14-Dec-2015

220 Views

Category:

Documents

1 Downloads

Preview:

Click to see full reader

TRANSCRIPT

Commissioning LHC Beam Instrumentation for Nominal

Bunch Intensities

Beam Commissioning Working Group

June 15th 2010

Rhodri Jones

(CERN Beam Instrumentation Group)

● Report from 2 MD periods● May 28th : 4 hours

● 1st go at● BPM reading v intensity studies● Studies on DCCT dependence on fill pattern● Calibration of various Fast BCT modes

● June 8th : 7 hours● 2nd go at

● BPM reading v intensity studies● Studies on DCCT dependence on fill patterns● Calibration of various Fact BCT modes

● 1st go at● Wire scanner timing calibration● BGI commissioning● PLL setting-up

● Current Status of Abort Gap Monitor

High Intensity BI LHC Commissioning

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● BPMs work on bunch to bunch basis● Only depend on bunch intensity

● BPMs can be used in 2 sensitivity modes● High sensitivity

● From ~ 1×109 to ~5×1010

● Low sensitivity● From ~ 5×1010 to ~2×1011

● Only changes threshold for bunch detection● No gains changed● Required to make system immune from reflections generated by

imperfect cabling, connections & BPMs.

● 1st MD showed that● B2 behaved as expected● B1 had a grey zone between 3×1010 and 5×1010 where neither

sensitivity gave required results

Dependence of BPM Readings on Bunch Intensity

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● One nominal bunch of 1×1011 slowly scraped away using a primary collimator

● Sensitivity constantly changed from high to low● Outliers due to acquisition overlapping two sensitivity ranges

● Sensitivity ranges seen to overlap as expected at around 5×1010

BPM Dependence on Intensity - Beam 2

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

02:21 02:35 02:48 03:01 03:15 03:280

1

2

3

4

5

6

7

8

9

10x 10

10

UTC Time (2010-05-28)

Inte

nsity

(p/b

unch

)

B2 high sens

B2 low sens

5.1

5.2

5.3

5.4

5.5

5.6

5.7

5.8

5.9

x 1010

UTC Time (2010-05-28)

Inte

nsity

(p/b

unch

)

B2 high sens

B2 low sens

● One nominal bunch of 1×1011 slowly scraped away using a primary collimator● 2 fills – one for low sensitivity and

one for high sensitivity

BPM Dependence on Intensity - Beam 1

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Low Sensitivity

High Sensitivity

Dead zone where neither setting works well

0 2 4 6 8 10 12×1010

0 2 4 6 8 10 12×1010

B1 Arc BPM variation - 5×1010 to 1×1011

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Mean = 55mmStdev = 80mm

Mean = 53mmStdev = 76mm

B1 Arc BPM variation - 6×1010 to 1×1011

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Mean = 11mmStdev = 50mm

Mean = 8mmStdev = 44mm

B1 LSS BPM variation - 6×1010 to 1×1011

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Mean = 20mmStdev = 57mm

Mean = 15mmStdev = 74mm

B1 v B2 Comparison - 6×1010 to 1×1011

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Mean = 8mmStdev = 44mm

Mean = 15mmStdev = 74mm

Arc BPMs

LSS BPMs

Mean = 21mmStdev = 60mm

Mean = 2mmStdev = 48mm

● BPM System● Max variation with intensity <200mm for 6×1010 to 1×1011

● Beam 1 behaviour in high sensitivity still to be understood for 3.5×1010 to 5×1010

● Hypothesis that intensity card is influencing B1 power supplies

● Oustanding Issues● Temperature variations (~50mm / °C)

● New software being tested to correct for this on-line● Influence of other beam on directional BPMs in the IRs

● New firmware & software using synchronous mode & bunch selection being tested to overcome this

● May also help overcome B1 issues in high sensitivity

BPM System Conclusions

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● At these intensities the fast BCTs behave as expected

● As in SPS small % is measured in neighbouring 25ns slot

● No impact observed from bunch length variations● 250ps to 150 ps sigma

Fast BCTs at 2×1010 protons per bunch

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● Saturation observed during first MD● Technical stop used to install attenuators for system B● Response tested during second MD

Saturation of the Fast BCT

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

System A Bunch Sum Intensity

(High Bandwidth ) loses track at

~6×1010

System B with added attenuators

Bunch Sum Intensity stays in agreement with the DCCT

Time

Tota

l Int

ensi

ty

● No issues observed with 2×1010 protons per bunch

● At 1×1011 the DCCT measurement using its highest gain range becomes dependent on the filling pattern

DCCT Performance (1/2)

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

2×1010 per injection 1×1011 per injection

Bunch at 1001

Bunch at 3001

Bunch at 11001

Grey = DCCTBlue = Fast LowBWRed = Fast HighBW (suffering from saturation)

● At 1×1011 the DCCT measuring in gain range 3 seems insensitive to the filling pattern

DCCT Performance (2/2)

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Dashed Blue = What it should beGreen = BCTDC A toggling from range 3 to 4Gray = BCTDC B toggling from range 3 to 4

● Fast BCT System● Add attenuators to operational systems suring next technical stop

● Ensure that this only affects the high bandwidth channel and not the low bandwidth providing the beam presence flag

● Need to understand the remaining calibration issues● Why raw calibration still needs tweaking to align fast BCT with DCCT● The cause and effect of the signal tail in the trailing 25ns slot

● Fast BCT dependence on bunch length● Look carefully at response during ramp when no longitudinal blow-up is applied to verify

that observed intensity variations with bunch length were due to saturation

● DCCT System● Switch to gain range 3 just above 1×1011 protons to avoid fill pattern

dependence● Currently investigating possibility to eliminate Range 4 for SMP data

● Investigations continuing for source of this problem

● Correcting the intensity stored in the logging database● Requested by experiments● Currently working with CO to store corrected data in logging

● Correction of known errors – e.g. DC offset & fast BCT signal tail

BCT System Conclusions

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● test

Wirescanner StudiesSignal Dependence on Acquisition Delay

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

B2 Horizontal

B2 Vertical1

50

100

800

50

1

1000

800Slot number

Need to space bunches by ~900 slots to measure individual sizes

Slot selection now available from OP app.

● Started testing without beam● Effect of emergency

HV shutdown on pressure

● Various scenarios tested

● One set of parameters led to a pressure rise to

10-6 mbar● Resulted in vacuum

valve closure

Rest Gas Ionisation Monitor (BGI) Commissioning

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● Conclusion● Emergency HV shutdown procedure will be modified to ensure

that it does not lead to a rise in pressure which could close the valves & dump the beam

● First images seen with 2 bunches of 1×1011 protons

● Gas pressure of 10-8 mbar

● Tests then carried out on signal level with respect to main parameters● MCP voltage, gas pressure etc.● Not possible to vary the BGI magnet current (interlock issue to be followed-up)

First Images from the BGI

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● BBQ sensitivity is so good PLL has been put on back burner but it does have advantages

● Less likely to jump onto spurious peaks● Can track tune without need for predefined windows● May be easier to make compatible with damper● Has built-in demodulation for chromaticity

measurement

● Beam 2:● Only one scan was possible in time allowed ● The scan was performed with debris of other tests

(very low intensity)!● Measurement indicates SNR better in H than V

● Would have been easy to obtain stable lock on either

● Poor look of V plane phase an artifact of wraparound due to incorrect phase offset correction

● Beam 1:● Just few minutes available for tests on this beam● Not possible to excite with the PLL system

● Hardware being investigated

Tune PLL Commissioning

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

● A bug in the software of the LHC abort gap monitor combined with insufficient hardware interlocks led to irreparable damage to both of the installed photomultiplier tubes (PMTs)

● 2 PMTs with 2 new amplifiers have been installed● A CERN spare and one kindly provided by the LARP collaboration

(Berkeley)● Spare PMT has lower sensitivity probably due to ageing

● Not an issue for operation

● 3 new PMTs have been ordered to arrive this summer

● New HV power supplies with output limitation are at CERN● SW to control them currently being developed & tested

● The abort gap monitor is operational again for both beams● It will be only be possible/safe to put the AG system into an automated

mode after replacing the power supplies (next tech stop)● Working at fixed Voltage only has small impact on system operation

Status of the LHC Abort Gap Monitors

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Status of the LHC Abort Gap Monitors

Beam Commissioning Working Group – 15th June 2010 Rhodri Jones

Little difference between expected signal at 450GeV and at 3.5TeV

top related