    Sartorius Economy SeriesSartorius Express Series

    EA and EB Models

    Electronic Precision Scales

    Installation and Operating Instructions

    98648-007-09

    IndustrialWeighi

    ngTechnology

  • 8/2/2019 Bascula Verficadora

    2/90

    2

    Intended Use

    The Economy/Express Series from Sartorius offers precision scales for weighing,

    with capacities ranging from 0.1 g to 150 kg.Economy/Express scales meet the highest requirements on the accuracy andreliability of weighing results through the following features: Efficient filtering-out of unfavorable ambient conditions, such as vibrations, drafts etc. Stable and reproducible results Excellent readability under any lighting conditions Rugged, durable weighing system

    Type of scale housing protection The weighing platforms are IP54-protected

    The display and control units are IP40-protectedEconomy/Express scales save work and speed up simple routineapplications through: Ultrafast response times Easy operation Computer connectivity through a built-in RS-232 interface Two lines in the printout are configurable to show, e.g., your company name

    Additional Features of the EB Models:With the EB model, you can choose from the following extra functions

    for simple applications: Counting Toggling between weight units (second weight unit) Net-total formulation (tare memory) Second tare memory (incl. automatic container taring function) Weighing in percent (incl. display of the difference between current weight

    and stored reference weight) Averaging Calculation by a factor

    Totalizing Gross/net toggling

    Note to Users in the U.S.This equipment has been tested and found to comply with the limits for a Class Adigital device, pursuant to Part 15 of the FCC Rules. These limits are designedto provide reasonable protection against harmful interference when the equipmentis operated in a commercial environment. This equipment generates, uses, and canradiate radio frequency energy and, if not installed and used in accordance withthe instruction manual, may cause harmful interference to radio communications.

    Operation of this equipment in a residential area is likely to cause harmfulinterference in which case the user will be required to correct the interference at hisown expense. Changes or modifications not expressly approved by Sartorius AGcould void the users authority to operate the equipment.

  • 8/2/2019 Bascula Verficadora

    3/90

    3

    Contents

    Intended Use................................................. 2

    Contents....................................................... 3

    Warnings and Safety Precautions ..................... 4

    Operating Design .......................................... 5

    Getting Started .............................................10

    Operating the ScaleBasic Weighing Function.............................14Calibration/Adjustment ...............................17

    Application Programs in the EB ModelsCounting ................................................. 20Toggle between Weight Units .......................23Second Tare Memory .................................27Weighing in Percent ...................................30Calculation by a Factor ...............................33Totalizing ..................................................36Gross/Net Toggling ...................................39Averaging.................................................41

    Data Output Functions ...................................45Data Interface............................................50Pin Assignment Chart ..................................59Cabling Diagram .......................................60

    Configuring the Scale (Menu)Setting Parameters in the Operating Menu ......61Scale Operating Menu (Overview) ................64

    Error Codes .................................................68

    Care and Maintenance .................................70Instructions for Recycling.................................71

    OverviewGeneral View of the Scale .......................... 72Description of the Keys ................................73Specifications ............................................74Dimensions (Scale Drawings)........................79Accessories (Options)..................................80Information on EC Verification ......................82Declarations of Conformity...........................85EC Type Approval ......................................88Index .......................................................89

  • 8/2/2019 Bascula Verficadora

    4/90

    4

    Warnings and Safety Precautions

    This scale has been constructed in accordance with

    the European Directives as well as internationalregulations and standards for operation ofelectrical equipment, electromagnetic compatibility,and stipulated safety requirements. Improper useor handling, however, can result in damageand/or injury.

    Read these operating instructions thoroughly beforeusing your scale to prevent damage to theequipment. Keep these instructions in a safe place.

    Follow the instructions below to ensure safe andtrouble-free operation of your scale:

    !Do not use this scale in a hazardousarea/location

    !Make sure that the voltage rating printed on theAC adapter is identical to your local line voltage

    The only way to turn the power off completely is

    to disconnect the AC adapter The scale housing is IP54-protected against dust

    deposits and water splashes

    The display and control unit is IP40-protectedagainst penetration of solid foreign objects

    Connect only Sartorius accessories and options,as these are optimally designed for use with yourEconomy/Express scale

    Protect the AC adapter and the display andcontrol unit from contact with liquids

    When cleaning your scale, make sure that noliquids enters the scale housing; use only a slightlymoistened cloth to clean the scale.

    Do not open the scale housing. If the seal isbroken, this will result in forfeiture of all claims underthe manufacturers warranty.

    In case you have any problems with your scale:

    $ Contact your local Sartorius office, dealer orservice center

  • 8/2/2019 Bascula Verficadora

    5/90

    5

    Operating Design

    The scales in the Economy/Express Series consist of

    a weighing cell and a display and control unit.In addition to the choice of power supply (viaAC adapter or external rechargeable battery pack),

    your scale also has an interface port for connectingperipheral devices, such as a printer, computer,universal remote control switch, etc.

    The display and control unit is affixed to theweighing platform. Operation of the Economy

    scales is simple and uniform.The EB models include all the features of theEA models plus application programs for togglingbetween weight units, counting, weighing inpercent, etc.

    Where not expressly stated otherwise, theinformation contained here also applies to theverified model versions in the European Union

    and European Economic Area (indicated by the-L0CE suffix to the model name).

    Keys

    The Economy/Express Series scales can beoperated using either the keys on the display andcontrol unit or a connected computer (PC).Some of the keys activate different functions,

    depending on how long they are pressed (pressbriefly = < 2 sec.; press and hold = 2 sec.).

    The equipment supplied includes adhesive labelsprinted in different languages for identifying thefunctions assigned to the keys.

    EA Model

    The EA model has 4 keys. The function activated

    by each key will differ according to whetherthe scale is in the weighing mode or setup mode.

  • 8/2/2019 Bascula Verficadora

    6/90

    6

    EB Model

    The EB model has 3 additional keys (gray), forcontrolling the application programs.

    Display

    The display is divided into 5 sections:

    Basis-Einheiten +Zusatzkennungen

    Hinweise f rAnwendungs-programme

    Einheiten-pfeile

    Mewert

    Busy-ZeichenVorzeichenStillstand

    Busy Symbol, Plus/Minus Sign, Stability Symbol

    If the symbol displayed here is a trianglev, this indicates that the scale is

    performing a function (busy symbol) a plus or minus sign (+ or ), this applies

    to the value displayed (on the EB models, thisalso applies to calculated values; e.g.,during counting)

    a small S, this indicates that the verified scale

    has been zeroed/tared a flashing triangle w, this indicates that thecalculated values are not valid for legal metrology

    Measured Value

    When the scale is in the weighing mode, theweight is displayed here. In the setup mode, themenu code numbers are shown here.On EB models, calculated values and applicationparameter settings are also displayed here.

    Busy symbolPlus/minus signStability symbol

    Applicationinformation

    Unitindicator

    Basic unit +additional information

    Measured value

  • 8/2/2019 Bascula Verficadora

    7/90

    7

    Basic Unit and Additional Information

    When the scale has stabilized, the weight unitis displayed here (g or kg). On the EB model,the calculated values (e.g., % or pcs) are alsodisplayed.

    Application Information

    On EB models, this section shows informationconcerning the current application program; e.g.,

    the reference quantity for counting.

    Unit Indicator

    You can affix one of several adhesive overlays to theright of this section. These overlays show theoptional weight units selected in the scale Setupmenu. The arrows indicate the selected units.

    Input FunctionsWhen the scale is in the setup mode, you can selectparameters from lists. The lists are contained in amenu, which is divided into 3 levels.

    To set parameters: Turn the scale off and backon again. While all segments are displayed, press)/[TARE] briefly.

    To navigate within a menu level: Press )/[TARE]briefly; when you reach the last menu item, the first isshown again.

    To change menu levels: Press p/[PRINT] briefly.

    To confirm the selected parameters on all 3 levels:Press and hold the p/[PRINT] key.

    o indicates the selected parameter option.

  • 8/2/2019 Bascula Verficadora

    8/90

    8

    All parameter options are listed in detail under

    Configuring the Scale.To store parameter settings and exit the menu:Press and hold the )/[TARE] keyParameter settings are stored in the non-volatilememory. When you turn on the scale, thelast parameters used are active.

    To exit the menu without storing any new settings:Press e/[ON/OFF]

    Output Functions

    Your Economy/Express scale is equippedwith a data interface for connecting your choiceof the following:

    Printer

    PC

    Universal remote control switch

    Printer

    You can configure a variety of options for dataoutput to a printer.

    You can have a printout generated automatically oronly when the p/[PRINT] key is pressed. You canmake the print function dependent on or indepen-dent of stability parameters, and define whetherdata ID codes will be included in the printout or not.

    The print functions are described in detail underData Output Functions in the chapter entitledOperating the Scale.

  • 8/2/2019 Bascula Verficadora

    9/90

    9

    Data Interface

    Instead of a printer, you can connect a differentperipheral device, such as a computer, to the datainterface.

    You can use a connected computer to control theEconomy/Express scale. Communication betweenthe scales data interface and a PC is in the formof request and response messages. Not all requestmessages will activate response messages.

    See Data Output Functions in the chapter entitledOperating the Scale for a detailed description.

    Error Messages

    Error messages are displayed for 2 seconds.The format is as follows:

    Sequence error: E + 2 digits Hardware error: E + 3 digits

    The section entitled Error Messages containsdetailed information.

  • 8/2/2019 Bascula Verficadora

    10/90

    10

    Getting Started

    Warranty

    Do not miss out on the benefits of our full warranty.Complete the warranty registration card, indicatingthe date of installation, and return the card to

    your Sartorius office or dealer.

    Storage and Shipping Conditions

    Do not expose the scale to shocks, vibrations,

    moisture or extreme temperatures.

    Unpacking the Scale

    After unpacking the scale, check it immediatelyfor any visible damage as a resultof rough handling during shipment.

    $ Note:The display and control unit is attached to theweighing platform via a cable.

    $ If you see any sign of damage, proceedas directed in the chapter entitled Care andMaintenance, under the section on SafetyInspection.

    Save the box and all parts of the packaging untilyou have successfully installed your scale. Only the

    original packaging provides the best protection forshipment. Before packing your scale,unplug all connected cables to prevent damage.

    Equipment Supplied

    The equipment supplied includes the componentslisted below:

    Scale with attached display and control unit

    Load pan

    AC adapter

    Installation and operating instructions

  • 8/2/2019 Bascula Verficadora

    11/90

    11

    Installation Instructions

    The Sartorius Economy/Express scales are designedto provide reliable weighing results under normalambient conditions. When choosing a location toset up your scale, observe the following sothat you will be able to work with added speedand accuracy:

    Set up the scale on a stable, even surface Avoid placing the scale in close proximity

    to a heater or otherwise exposing the scaleto heat or direct sunlight Protect the scale from drafts that come from

    open windows or doors Avoid exposing the scale to extreme vibrations

    during weighing Protect the scale from aggressive

    chemical vapors Do not expose the scale to extreme moisture

    over long periodsConditioning the Scale:Moisture in the air can condense on the surfaces ofa cold scale whenever it is brought into asubstantially warmer place. If you transfer the scaleto a warmer area, make sure to condition it forabout 2 hours at room temperature, leaving itunplugged from AC power. Afterwards, keep thescale continuously connected to AC power.

    Seal on Scales Verified for Use in Legal Metrologyin the EU*:

    EU legislation requires that a control seal beaffixed to the verified scale. The control sealconsists of a sticker with the Sartorius logo. Thisseal will be irreparably damaged if you attemptto remove it. If the seal is broken, the validity

    of the verification will become void, and you musthave your scale re-verified.

    * including the Signatories of the Agreementon the European Economic Area

  • 8/2/2019 Bascula Verficadora

    12/90

    12

    Place the load plate on the scale

    Connecting the Scale to AC Power Check the voltage rating and the plug design

    If they do not match the rating or standard youuse, contact your Sartorius office or dealer

    Use only

    Original Sartorius AC adapters

    AC adapters with a registered approval ratingfrom a national testing laboratory

    $ See the chapter entitled Accessories forinformation on using an external rechargeablebattery pack with your scale

    Insert the right-angle plug into the jack and thentighten the screws

    Then insert the plug of the AC adapter intoa wall outlet (mains)

    Safety Precautions

    The AC adapter rated to Class 2 can be pluggedinto any wall outlet without requiring any additionalsafety precautions. The pole of the output voltage

    is connected to the scale housing, which can begrounded for operation. The data interface isalso electrically connected to the scale housing(ground).

    EA/EBDCE/EDE:

    EA/EBFEG:

  • 8/2/2019 Bascula Verficadora

    13/90

    13

    Connecting Electronic Peripheral Devices

    Make absolutely sure to unplug the scalefrom AC power before you connect ordisconnect a peripheral device (printer or PC)to or from the interface port.

    Warmup Time

    To deliver exact results, the scale must warm up for

    at least 30 minutes after initial connection to ACpower. Only after this time will the scale havereached the required operating temperature.

    Leveling the Scale

    Remove the load plate

    Loosen the lock nuts on all leveling feet(use a 13 mm open-end wrench if necessary)

    Turn the leveling feet until the scale is level

    Avoid unintentional changes in the adjustment:Tighten the lock nuts with an open-end wrench

    and then replace the load plate

    Leveling the Scale (for EA/EB-I, -L0CEmodels only)

    Using the 4 footscrews, level the scale so that theair bubble is centered within the level indicator

  • 8/2/2019 Bascula Verficadora

    14/90

    14

    Operating the Scale

    Basic Weighing Function

    Purpose

    The basic weighing function is always accessible in theEA and EB scales.

    In the EB model, this function can be used aloneor in combination with an application program(Toggle between Weight Units, Counting, Weighingin Percent, etc.).

    Features

    Zeroing the scaleDepending on prevailing ambient conditions,the display may not show a zero readout eventhough there is no load on the scale. If theweight shown is less than 2% of the scales

    maximum weighing capacity, you can zerothe scale. Taring the scale

    Tare the scale with an empty container onthe weighing pan to obtain a readout of the netweight after the container is filled.

    Printing weights

    Factory SettingsWeight unit 1: Kilogram (1 7 3)

    Manual/automatic printing:Manual after stability (6 1 2)

    Print format:Gross, tare and net values with data ID codes (7 1 3)

  • 8/2/2019 Bascula Verficadora

    15/90

    15

    Preparation

    Turn on the scale: Press e/[ON/OFF]

    > All segments light up briefly

    $ To change configurations: See the chapterentitled Configuring the Scale

    $ To load factory-set configurations: See Configuringthe Scale, parameter 9 - 1

    > TheS

    symbol is displayed when the scale iszeroed or tared (only on scales verified for use inlegal metrology)

    Additional Functions in EB models:

    Start an application

    Toggle the display between weight and

    calculated value Delete an application

  • 8/2/2019 Bascula Verficadora

    16/90

    16

    Practical Examples

    Simple weighing

    Menu code settings: Factory-set codes

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF] 0.000 kg

    2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)

    3. Place the container on the scale + 0.015 kg(here: 0.015 kg)

    4. Tare the scale )/[TARE] 0.000 kg

    5. Place a sample in the container + 0.125 kgon the scale (here: 0.125 kg)

    6. Print weight* p/[PRINT] EISENMUELLERGOETTINGEN

    N + 0.125 kgT + 0.015 kgG# + 0.140 kg

    * The Sartorius Service Centeror your Sartorius dealer canconfigure the scale to include2 customer-specific lines on theprintout. Software is available forsetting this configuration with a PC.

  • 8/2/2019 Bascula Verficadora

    17/90

    17

    Calibration/Adjustment

    Purpose

    Calibration is the determination of the differencebetween the weight readout and the trueweight (mass) of a sample. Calibration does notentail making any changes within the scale.Adjustment is the correction of this difference betweenthe measured value displayed and the true weight(mass) of the sample, or the reduction of the differenceto an allowable level within the maximum permissibleerror limits.

    Features

    External adjustment can only be performed when there is no load on the scale, the scale is set to zero, and the internal signal is stable.

    If these conditions are not met, an error messageis displayed.

    The weight on the scale must not differ from thenominal weight by more than 2%.You can use g, kg or lb as the weight unit forcalibration (1 4)You can block calibration of the scale (1 5)

    Calibration/Adjustment for Verified Scales The calibration function is blocked in verified

    scales (sealed switch cover)> You can only perform calibration by removing

    this seal. In this case, verification is no longervalid and you must have your scale re-verified.

    Factory Settings

    Weight unit for calibration: kg (1 4 2)Calibration function: accessible (1 5 1)

  • 8/2/2019 Bascula Verficadora

    18/90

    18

    Example

    Calibrate the scale

    Menu code settings: Factory-set codes

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF] 0.000 kg

    2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)

    3. Begin calibration )/[TARE] + 5.000Calibration weight is > 2 sec.displayed without wt. unit

    4. Place the indicated calibration 5.000

    weight on the scale(here: 5,000 g)

    After calibration, the calibration + 5.000 kgweight is displayed with wt. unit

    5. Remove the calibration weight 0.000 kg

  • 8/2/2019 Bascula Verficadora

    19/90

    19

    Application Programs (EB Models Only)

    Note Concerning Verified Scales Used as LegalMeasuring Instruments in the EU*:

    All application programs can also be used inlegal metrology. Calculated weight values areindicated by one of the following symbols,displayed to the right of the numerical value:

    Percentage = %

    Piece count = pcs

    Other calculated value = o

    A flashing triangle w displayed to the left of thenumerical value indicates the value is notvalid for legal metrology (e.g., in averagingor calculation).

    * including the Signatories of the Agreement on theEuropean Economic Area

  • 8/2/2019 Bascula Verficadora

    20/90

    20

    Counting (EB Models Only)

    Purpose

    With the Counting program you can determinethe number of parts that have approximately thesame average piece weight.

    Available Features

    The reference sample quantity can be changedeither in the operating menu or during weighing

    Store the current weight value to have it loadedas the preset reference sample quantity the nexttime you initialize the Counting application

    Re-initialize without quitting the program

    The average piece weight is automatically outputvia the data interface port after initialization,

    if the menu code for Printout with data ID codesis set

    Press w/[TOGGLE] to toggle the displaybetween piece count and weight

    Factory Settings

    Display accuracy 2: Standard resolution (3 2 1)

    Reference sample quantity: 10 (3 3 2)

    Preparation

    Configure the Counting application in theoperating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; whileall segments are lit, press )/[TARE]

  • 8/2/2019 Bascula Verficadora

    21/90

    21

    Select the Counting menu item:

    Press )/[TARE], p/[PRINT],and p/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 4 Counting

    Confirm the selection of Counting:Press p/[PRINT] for > 2 sec.

    > o indicates that this is the current menucode setting

    Set the next parameters: Press p/[PRINT]

    Select and confirm: Reference sample quantity:

    3 3 1 5 pcs3 3 2 10 pcs3 3 3 20 pcs3 3 4 50 pcs3 3 5 100 pcs

    see also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale

    Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:

    Initialize application (v key) Delete initialization values (c key) Change reference sample quantity after deleting

    initialization value (w/[TOGGLE] key, press for

    at least 2 sec.) Display weight (w/[TOGGLE] key) Calibrate/adjust scale ()/[TARE], press for

    at least 2 sec.)

  • 8/2/2019 Bascula Verficadora

    22/90

    22

    Example

    Determine an unknown piece count; store the weight on the scaleas a reference sample quantity

    Settings (changes in the factory settings required for this example):Menu: Application program: Counting (2 1 4)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. If necessary, zero the scale (/[ZERO] 0.000 kg10

    (S symbol on the left:verified scale is zeroed)

    3. Select the reference sample w/[TOGGLE] ref 20 (briefly)quantity (here: 50 pcs) w/[TOGGLE] ref 50 (briefly)

    4. Place the reference sample + 0.930 kg50quantity (50 pcs) on the scale(here: 0.930 kg)

    5. Start the application; if the print v + 50 pcs50format is set to include dataID codes, the following is printed wRef + 0.186 kg

    6. Display weight w/[TOGGLE] + 0.930 kg50

    7. Display quantity w/[TOGGLE] + 50 pcs50

    8. Weigh uncounted parts + 174 pcs50(here: 174 pcs)

    9. Print total piece count p/[PRINT] EISENMUELLERGOETTINGEN

    Qnt + 174 pcs

    10. Unload the scale0

    pcs50

    11. Repeat the procedure startingfrom Step 8, if desired.

  • 8/2/2019 Bascula Verficadora

    23/90

    23

    Toggle between Weight Units (EB Models only)

    Purpose

    With this application program you can toggle thedisplay of a weight value back and forth between twoweight units.

    Available Features

    Toggling the displayed weight Other features as for the basic weighing function

    Factory Settings

    Weight unit 1: kg (1 7 3)

    Weight unit 2: kg (3 1 3)

    Preparation

    Configure the Toggle weight units application in theoperating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; while

    all segments are lit, press )/[TARE]

    Select the Toggle weight units menu item:Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 2 Toggle weight units

    Confirm selection: Press p/[PRINT]for at least 2 sec.

    > o indicates that this is the current menucode setting

  • 8/2/2019 Bascula Verficadora

    24/90

    24

    Set the next parameters: Press p/[PRINT]

    Select and confirm:

    Weight unit 1: see next page (1 7 x)

    Display resolution 1:1 8 1 Standard resolution1 8 2 10x higher resolution**1 8 3 2x higher resolution (PolyRange)*

    Weight unit 2: see next page (3 1 x)

    Display resolution 2:3 2 1 Standard resolution3 2 2 10x higher resolution**3 2 3 2x higher resolution (PolyRange)*

    see also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale

    Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the functions for:

    power off (e/[ON/OFF] key), zeroing ((/[ZERO] key),

    taring ()/[TARE] key), and printing (p/[PRINT] key),

    you can also access the following functions fromthis application:

    toggle between weight units/and 2(w/[TOGGLE] p/[PRINT] or v key),

    calibration/adjustment ()/[TARE] key; press forat least 2 sec.)

    ** = not applicable for verified balances** = only for EB-0UR models

  • 8/2/2019 Bascula Verficadora

    25/90

    25

    You can affix one of several adhesive overlays to the right of this section.

    These overlays show the optional weight units selected in the scale Setup menu.The arrows indicate the selected units.

    These settings cannot be changed on verified scales.

    Menu code Unit Conversion factor Printout(1 7 2) (3 1 2) Grams 1.00000000000 g(1 7 3) (3 1 3) Kilograms 0.00100000000 kg

    (1 7 4) (3 1 4) Carats 5.00000000000 ct(1 7 5) (3 1 5) Pounds 0.00220462260 lb(1 7 6) (3 1 6) Ounces 0.03527396200 oz(1 7 7) (3 1 7) Troy ounces 0.03215074700 ozt(1 7 8) (3 1 8) Hong Kong taels 0.02671725000 tlh(1 7 9) (3 1 9) Singapore taels 0.02645544638 tls(1 7 10) (3 1 10) Taiwanese taels 0.02666666000 tlt(1 7 11) (3 1 11) Grains 15.43235835000 GN

    (1 7 12) (3 1 12) Pennyweights 0.64301493100 dwt(1 7 13) (3 1 13) Milligrams 1000.00000000000 mg(1 7 14) (3 1 14) Parts per pound 1.12876677120 /lb(1 7 15) (3 1 15) Chinese taels 0.02645547175 tlc(1 7 16) (3 1 16) Mommes 0.26670000000 mom(1 7 17) (3 1 17) Austrian carats 5.00000000000 K(1 7 18) (3 1 18) Tola 0.08573333810 tol(1 7 19) (3 1 19) Baht 0.06578947437 bat(1 7 20) (3 1 20) Mesghal 0.21700000000 MS(1 7 21) (3 1 21) Tons 0.00000100000 T(1 7 22) (3 1 22) lb/oz 0.03527396200 o

    If you select menu code 1 7 22 or 3 1 22, you can affix the label with the lbweight unit under the left number, and the label with the oz weight unit underthe number on the right.

  • 8/2/2019 Bascula Verficadora

    26/90

    26

    Example

    Toggle the display from kilograms [kg] (1st unit) to pounds [lb] (2nd unit)

    Settings (changes in the factory settings required for this example):Menu: Application program: Toggle between weight units (2 1 2)Menu: Weight unit 2: pounds (3 1 5)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. Place weight on scale(here: 2.295 kg)

    3. Toggle to weight unit w/[TOGGLE]2: Pounds [lb]

    4. Print weight p/[PRINT] EISENMUELLERGOETTINGEN

    G + 5.060 lb

    5. Toggle to weight unit w/[TOGGLE]1: kilograms [kg]

  • 8/2/2019 Bascula Verficadora

    27/90

    27

    Second Tare Memory (EB Models Only)

    Purpose

    With this application program you can store the weighton the scale as a tare compensation weight.

    Available Features

    Press v to store the weight on the scale in thesecond tare memory

    NET is displayed next to the net weight readout ifthe scale has been tared or if a tare value isstored in the second tare memory (N in theprintout indicates that the scale was tared bypressing the )/[TARE] key; N1 indicates thatthere is data in the second tare memory)

    You can have a series of containers tared in

    sequence, if the scale is unloaded to under 30%of the previous weight after measurement andthen loaded with 70% 130% of the previousweight (i.e., the next container)

    Press c to delete the value from the secondtare memory

    Press w/[TOGGLE] to toggle betweenweight unit 1 and weight unit 2

    Preparation

    Configure the Second tare memory application inthe operating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; while

    all segments are lit, press )/[TARE]

  • 8/2/2019 Bascula Verficadora

    28/90

    28

    Select the Second tare memory menu item:

    Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 3 Second tare memory2 1 10 Second tare memory with automatic

    container taring

    Confirm selection of Second tare memory:Press p/[PRINT] for at least 2 sec.

    > o indicates that this is the current menucode setting

    Set the next parameters: Press p/[PRINT]

    Select and confirm: Weight unit 1: (see also Toggle between

    Weight Units)1 7 2 Grams... ...

    1 7 22 lb/oz Weight unit 2: (see also Toggle between

    Weight Units)3 1 2 Grams... ...3 1 22 lb/oz

    see also the Scale Operating Menu(Overview) in the chapter entitled Configuring

    the Scale Save settings and exit the Setup menu:

    Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:

    Store weight in second tare memory (v key) Delete tare memory (c key) Toggle weight unit (w/[TOGGLE] key)

  • 8/2/2019 Bascula Verficadora

    29/90

    29

    Example

    Fill containers with nearly identical weights ( 30 %)

    Settings (changes in the factory settings required for this example):Menu: Application program: Second tare memory with automatic containertaring (2 1 10)Menu: Weight unit 2: Grams (3 1 2)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)

    3. Place an empty container on + 0.065 kgthe scale (here: 65 g)

    4. Store the container weight in v 0.000 kgNET

    the second tare memory.If the print format is set to includedata ID codes, the following N1 + 0.065 kgis printed

    5. Fill the container + 0.500 kgNET

    (here: to 500 g)

    6. Display weight unit 2 w/[TOGGLE] + 500 gNET

    7. Display weight unit 1 w/[TOGGLE] + 0.500 kgNET

    8. Remove the filled container 0.065 kgNET

    from the scale

    9. Place the next empty container 0.000 kgNET

    on the scale (here: 75 g) (automatic container

    taring is performed)10. Repeat the procedure starting

    from Step 5.

  • 8/2/2019 Bascula Verficadora

    30/90

    30

    Weighing in Percent(EB Models Only)

    Purpose

    This application program allows you to obtainweight readouts in percent which are in proportionto a reference weight.

    Available Features

    Reference percentage can be changed either inthe operating menu or during weighing

    Store the current weight value to have it loadedas the preset reference percentage the next time

    you initialize the Weighing in Percent application

    Optional auto-zero function

    The reference sample weight is automatically

    output via the data interface port afterinitialization, if the menu code for Printout withdata ID codes is set

    Press w/[TOGGLE] to toggle the displaybetween percentage and weight

    Factory Settings

    Display accuracy 2: Standard resolution (3 2 1)

    Reference percentage: 10 (3 3 2)

    Preparation

    Configure the Weighing in Percent application inthe operating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF];while all segments are lit, press )/[TARE]

  • 8/2/2019 Bascula Verficadora

    31/90

    31

    Select the Weighing in Percent menu item:

    Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 5 Weighing in percent without taring2 1 11 Weighing in percent with taring

    Confirm selection of Weighing in Percent:Press p/[PRINT] for at least 2 sec.

    > o indicates that this is the current menu

    code setting

    Set the next parameters: Press p/[PRINT]

    Select and confirm: Reference percentage at power-on:

    3 3 1 5 %3 3 2 10 %3 3 3 20 %3 3 4 50 %3 3 5 100 %

    see also the Scale Operating Menu (Overview)in the chapter entitled Configuring the Scale

    Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:

    Initialize application (v key) Delete initialization values (c key) Change reference percentage after deleting

    initialization value (w/[TOGGLE] key, press forat least 2 sec.)

    Display weight (w/[TOGGLE] key) Calibrate/adjust scale ()/[TARE], press forat least 2 sec.)

  • 8/2/2019 Bascula Verficadora

    32/90

    32

    Example

    Determine an unknown percentage; store the weight on the scaleas a reference percentage

    Settings (changes in the factory settings required for this example):Menu: Application program: Weighing in percent (2 1 5)Menu: Reference percentage 100 % (3 3 5)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)

    3. Place the reference weight + 2.295 kgon the scale(here: 2.295 kg = 100%)

    4. Start the application; if the print v + 100.00 %format is set to include dataID codes, the following is printed Wxx% + 2.295 kg

    5. Display weight w/[TOGGLE] + 2.295 kg

    6. Display percentage w/[TOGGLE] + 100.00 %

    7. Display the reference v ref100percentage (at least 2 sec.)

    8. Place an unknown weight + 140.41 %on the scale(here: 3.225 kg)

    9. Print percentage p/[PRINT] Prc + 140.41 %

    10. Unload the scale 0.00 %

    11. Repeat the procedure startingwith Step 8, if desired.

  • 8/2/2019 Bascula Verficadora

    33/90

    33

    Calculation by a Factor (EB Models Only)

    Purpose

    With this application program you can calculate theweight of a sample by a given factor.

    Available Features

    Calculate a weight value using one of the

    following factors: 0.25; 0.50. 0.75; 1.0; 1.5;2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0

    A flashing triangle in the display indicatesa calculated value

    Press c to delete the calculation factor

    Press w/[TOGGLE] to toggle between weightunit 1 and weight unit 2

    Factory Settings

    Weight unit 1: kg (1 7 3)

    Weight unit 2: kg (3 1 3)

    Display accuracy 2: Standard resolution (3 2 1)

    Preparation

    Configure the Calculation application in the

    operating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; whileall segments are lit, press )/[TARE]

    Select the Calculation menu item:Press )/[TARE], p/[PRINT], and

    p/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 7 Calculation by a factor

  • 8/2/2019 Bascula Verficadora

    34/90

    34

    Confirm selection of Calculation:

    Press p/[PRINT] for at least 2 sec.> o indicates that this is the current menu

    code setting

    Set the next parameters: Press p/[PRINT]

    Select and confirm: Weight unit 1: (see also Toggle between

    Weight Units)1 7 2

    Grams... ...1 7 22 lb/oz

    Weight unit 2: (see also Toggle betweenWeight Units)3 1 2 Grams... ...3 1 22 lb/oz

    see also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale

    Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the basic functions (power off,

    zeroing, taring and printing), you can also accessthe following functions from this application:

    Initialize application (v key) Change the calculation factor after the

    application is initialized (v key) Delete initialization values (c key) Toggle weight unit (w/[TOGGLE] key) Calibration/adjustment ()/[TARE] key;

    press for at least 2 sec.)

  • 8/2/2019 Bascula Verficadora

    35/90

    35

    Example

    You have a formula requiring several different components for a total amount of1,000 g. You want to use this recipe to make a 500 g batch (factor: 0.5).The scale shows the amounts called for in the formula, although only half theamounts are placed on the scale. This saves you having to recalculate theamount of each component.

    Settings (changes in the factory settings required for this example):Menu: Application program: Calculation (2 1 7)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed

    verified scales only)3. Place an empty container on + 0.065 kg

    the scale (here: 65 g)

    4. Tare the scale )/[TARE] 0.000 kgNET

    5. Start calculation v 0.000 kgNET

    .25

    6. Select the calculation factor v 0.000 kgNET

    .50

    7. Add first component 0.240 kgNET

    .50

    (amount called for in the formula:240 g; actual amount: 120 g)

    8. Add other components 1.000 kgNET

    .50

    (until the display indicates1,000 g)

    9. Print weight of components p/[PRINT] Res + 1.000 kg

  • 8/2/2019 Bascula Verficadora

    36/90

    36

    Totalizing (EB Models Only)

    Purpose

    This application program acts as a cumulativememory function.

    Available Features

    The current weight is stored as a net value when

    the scale stabilizes A value indicated as negative () in the

    display is stored as a positive (+) value in thetotalizing memory

    The data stored in the totalizing memory isautomatically output via the data interface portafter initialization, if the menu code forPrintout with data ID codes is set

    The number of values in the totalizing memoryis displayed (transaction counter)

    Press v for > 2 sec. to display the contentsof the totalizing memory

    Press c to clear the totalizing memory. Thetotal is displayed for 1.5 seconds and printed

    You cannot toggle between weight units when

    there is data in the totalizing memory

    Factory Settings

    Weight unit 1: kg (1 7 3)

    Display accuracy 2: Standard resolution (3 2 1)

  • 8/2/2019 Bascula Verficadora

    37/90

    37

    Preparation

    Configure the Totalizing application in theoperating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF];while all segments are lit, press )/[TARE]

    Select the Totalizing menu item:Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 8 Totalizing

    Confirm selection of Totalizing:Press p/[PRINT] for at least 2 sec.

    > o indicates that this is the current menucode setting

    Set the next parameters: Press p/[PRINT]

    Select and confirm: Weight unit 1: (see also Toggle between

    Weight Units)1 7 2 Grams... ...1 7 22 lb/oz

    see also the Scale Operating Menu(Overview) in the chapter entitled Configuring

    the Scale Save settings and exit the Setup menu:

    Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:

    Start application (v key) Display value in totalizing memory (v key;press for at least 2 sec.)

    Clear totalizing memory (c key)

  • 8/2/2019 Bascula Verficadora

    38/90

    38

    Example

    Totalize weight values

    Settings (changes in the factory settings required for this example):Menu: Application program: Totalizing (2 1 8)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)

    3. Place sample on the scale + 0.380 kg(here: 380 g)

    4. Store value in memory; if the v + 0.380 kg1print format is set to include dataID codes, the following is printed G + 0.380 kg

    5. Place the next sample on + 0.575 kg1the scale (here: 575 g)

    6. Store value in memory v + 0.575 kg2

    7. Display the total in memory v for at least 0.955 kg2 (briefly)(total is also printed) 2 sec. Sum + 0.955 kg

    8. Clear totalizing memory and c Sum + 0.955 kgprint total

    9. Repeat the procedure startingfrom Step 4, if desired.

  • 8/2/2019 Bascula Verficadora

    39/90

    39

    Gross/Net Toggling (EB Models Only)

    Purpose

    With this application program you can togglebetween net and gross values.

    Available Features

    Toggle the display between gross and net weightswhen there is data in the tare memory

    Press w/[TOGGLE] to toggle between weightunit 1 and weight unit 2

    Factory Settings

    Weight unit 1: kg (1 7 3)

    Weight unit 2: kg (3 1 3)

    Display accuracy 2: Standard resolution (3 2 1)

    Preparation

    Configure the Gross/net toggling application in theoperating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; while

    all segments are lit, press )/[TARE] briefly Select the Gross/net toggling menu item:

    Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 9 Gross/net toggling

    Confirm selection of Gross/net toggling:Press p/[PRINT] for at least 2 sec.

    > o indicates that this is the current menucode setting

    Set the next parameters: Press p/[PRINT]

  • 8/2/2019 Bascula Verficadora

    40/90

    40

    Select and confirm:

    Weight unit 1: (see also Toggle betweenWeight Units)1 7 2 Grams... ...1 7 22 lb/oz

    Weight unit 2: (see also Toggle betweenWeight Units)3 1 2 Grams...

    ...3 1 22 lb/oz

    Display accuracy 2:3 2 1 Standard resolution3 2 2 10x higher resolution**3 2 3 2x higher resolution (PolyRange)*

    see also the Scale Operating Menu (Overview)in the chapter entitled Configuring the Scale

    Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.

    Additional Functions

    In addition to the basic functions (power off, zeroing,taring and printing), you can also access thefollowing functions from this application:

    Toggle display between gross and net values

    (v key) Toggle weight unit (w/[TOGGLE] key)

    Example of Printout

    Net display(p/[PRINT] key): N + 0.125 kg

    T + 0.015 kgG# + 0.140 kg

    Gross display(p/[PRINT] key): G + 0.140 kg

    ** = not applicable for verified scales** = only for EA/EB-0UR models

  • 8/2/2019 Bascula Verficadora

    41/90

    41

    Averaging (EB Models Only)

    Purpose

    Use this program to determine the weights ofunstable samples (e.g., live animals) or to determineweights under unstable ambient conditions. In thisprogram, the scale calculates the weight as theaverage value from a defined number of individualweighing operations. These weighing operationsare also known as subweighing operationsor subweighs.

    Available Features

    You can set the number of subweighingoperations performed in the operating menu,or before starting the Averaging application(press w/[TOGGLE])

    The number of subweighs remainingto be performed is indicated in the displayduring weighing

    The measured result displayed is the arithmeticmean shown in the selected weight unit;a triangle under the plus or minus sign indicatesthat this is a calculated value

    Press v for at least 2 sec. to display the pre-set

    number of subweighing operations Press w/[TOGGLE] to toggle the display

    between the measured result and the weight

    The measurement result is automatically output viathe data interface port after initialization, if themenu code for Printout with data ID codes is set

    Factory SettingsDisplay accuracy 2: Standard resolution (3 2 1)

    Number of subweighs for averaging: 10 (3 3 2)

  • 8/2/2019 Bascula Verficadora

    42/90

    42

    Preparation

    Configure the Averaging application in theoperating menu:

    $ If the scale is on, turn it off:Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; whileall segments are lit, press )/[TARE]

    Select the Averaging menu item:

    Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 12 Averaging

    Confirm selection of Averaging: Press p/[PRINT]for at least 2 sec.

    > o indicates that this is the current menucode setting

    Set the next parameters: Press p/[PRINT]

    Select and confirm: Number of subweighs for averaging:

    3 3 1 5 subweighs3 3 2 10 subweighs3 3 3 20 subweighs3 3 4 50 subweighs3 3 5

    100 subweighssee also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale

    Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.

  • 8/2/2019 Bascula Verficadora

    43/90

    43

    Additional Functions

    In addition to the functions for:

    power off (e/[ON/OFF] key), zeroing ((/[ZERO] key), taring ()/[TARE] key), and printing (p/[PRINT] key),

    you can also access the following functions fromthis application:

    Start measuring (v key) Clear display of results (c key) Change number of measurements after clearing

    the display (w/[TOGGLE] key) Display weight (w/[TOGGLE] key) Calibration/adjustment ()/[TARE] key; press

    for at least 2 sec.)

  • 8/2/2019 Bascula Verficadora

    44/90

    44

    Example

    Determine the weight of a sample in extremely unstable ambient conditions.Enter the number of subweighs while in the weighing mode.

    Settings (changes in the factory settings required for this example):Menu: Application program: Averaging (2 1 12)

    Step Key (or instruction) Display/Output

    1. Turn on the scale e/[ON/OFF]

    2. If necessary, zero the scale (/[ZERO] 0.000 kg10

    (S symbol: scale is zeroed verified scales only)

    3. Set the number of subweighs w/[TOGGLE] ref 20 (briefly)for averaging w/[TOGGLE] ref 50 (briefly)(here: 5 subweighs) w/[TOGGLE] ref100 (briefly)

    w/[TOGGLE] ref1 (briefly)w/[TOGGLE] ref 2 (briefly)w/[TOGGLE] ref 5 (briefly)

    0.000 kg 5

    4. Place sample on the scale + 8888 5(weight readout fluctuates;here: around 275 g)

    5. Start measurement; if the v 8888 5 . . . 1print format is set to include 0.275 kg 5

    data ID codes, the following Res + 0.275 kgis printed

    6. Unload the scale 0.275 kg 5

    (stable display)

    7. Repeat the procedure starting

    from Step 4.

  • 8/2/2019 Bascula Verficadora

    45/90

    45

    Data Output Functions

    There are 3 options for data output:

    Output to the display and control unit

    Output to a printer (generate a printout)

    Output to a peripheral device (e.g., computer)via the interface port

    Output to the Display and Control Unit

    The display is divided into 5 sections. Informationabout the scale, the application being usedand the sample weighed is output in the followingsections:

    Busy symbol, plus/minus sign, stability symbol

    Measured value line

    Basic units and additional identifiers

    Application parameter

    Weight unit indicator

    Busy Symbol, Plus/Minus Sign, Stability Symbol

    This section contains:

    Busy symbol

    Plus/minus sign

    Zero indicator on verified scales

    Calculated value indicator (flashing triangle;in the examples shown in this manual, it appearsas a white triangle with a black border)

    Basis-Einheiten +Zusatzkennungen

    Hinweise f rAnwendungs-programme

    Einheiten-pfeile

    Mewert

    Busy-ZeichenVorzeichenStillstand

    Busy symbolPlus/minus signStability symbol

    Measured valueApplicationinformation

    Basic unit +additional information

    Unitindicator

  • 8/2/2019 Bascula Verficadora

    46/90

    46

    Measured Value Line

    This section shows:

    Current weight (digits bordered with a broken lineare invalid in use for legal metrology)

    Calculated values (e.g., piece counts)

    Application parameters during configuration(e.g., reference sample quantity, etc.)

    Software version (r = release; in this example:19 = scale program; 01 = software version)

    Weight Units, Additional Identifiers

    This section shows:

    The current weight unit (e.g., kg)

    Other unit of measurement (e.g., piece count)

    Indicates that there is data in the tare memory

    Application Parameter (EB models only)

    This section shows application parameters(e.g., reference sample quantity, no. of subweighs

    for averaging, etc.).

    Weight Unit Indicator (EB models only)

    The arrows in this section indicate the configuredweight units (upper arrow: weight unit 1;lower arrow: weight unit 2)

  • 8/2/2019 Bascula Verficadora

    47/90

    47

    Printing a Data Record

    Purpose

    You can generate a printout of weights as well asother measured values and identifiers fordocumentation purposes. You can format theprintout to meet individual requirements.

    Available Features

    The printout can be configured to include 2 user-defined header lines of 14 characters each. Thesetext lines are entered either at the Sartorius factoryor on location with a special program that uses theBPI interface command sequence.

    Print an individual weight, or net, gross andtare weights.

    Line format: You can configure a data ID code ofup to 6 characters for each of the values printed; thisdata ID code is printed at the beginning of the line

    Printouts generated automatically or by pressingthe p/[PRINT] key, dependent on or independentof scale stability

    You can have the following values output automaticallywhen using the application programs (EB models only)

    if menu code 7 1 2 or 7 1 3 is configured (printoutwith data ID codes):

    Second tare memory: last net value

    Counting: Reference weight for one piece(average piece weight)

    Weighing in percent: Reference weight for thepercentage selected

    Totalizing: Current weight, total weight (net) Averaging: Result of measurement

  • 8/2/2019 Bascula Verficadora

    48/90

    48

    Factory Settings

    Header:The default header lines do not contain any text

    Print manual/automatic:Individual printout or automatic printout dependenton scale stability:Manual after stability (6 1 2)

    Line format:

    Up to 6 characters at the beginning of each line toidentify the weight or calculated value:Print net, tare, gross value with data ID (7 1 3)

    See the chapter entitled Configuring the Scalefor instructions on setting parameters

    Examples

    You can configure the EISENMUELLER User-definedprintout to include 2 user- GOETTINGEN User-defineddefined header lines.The software required forconfiguring these lines via

    your PC is available fromSartorius in Goettingen

    or your local Sartoriusoffice or dealer.

    Printout without Data ID Codes:

    The value currently + 1530.0 g Weight in gramsdisplayed is printed + 58.562 ozt Weight in Troy ounces(weight or calculated + 253 pcs Piece countvalue with unit) + 88.2 % Percentage

    + 105.8 o Calculated value

    Headers:

  • 8/2/2019 Bascula Verficadora

    49/90

    49

    Printout with Data ID Codes:

    The current value N + 1.530 kg Current net weightdisplayed can be printed N1 + 1.530 kg Current net weightwith a data ID code T + 0.234 kg Value in tare memoryof up to 6 characters T1 + 0.102 kg Value in 2nd tare memoryat the beginning of the G + 1.553 kg Current gross weightline. You can use this G# + 1.630 kg Calculated gross weightdata ID code, e.g., to Qnt + 253 pcs Calculated quantitydesignate a weight Prc + 88.2 % Calculated percentage

    readout as a net weight Sum + 1.279 kg Total value (net)(N) or a calculated Res + 1.530 kg Calculated resultvalue as a piece count(QNT).

    Print Application Parameters:

    You can generate a print- wRef + 0.014 kg Counting:out of one or more of the Average piece weightvalues configured for Wxx% + 1.200 kg Weighing in percent:initialization of an appli- reference weight for thecation as soon as you selected percentageinitialize the scale.

    Auto Print:

    You can have the weight N + 1.530 kg Net weightreadout printed auto- Stat Display blankmatically. The display Stat L Display underload

    update interval depends Stat H Display overloadon the operating statusof the scale and on thescale model.

  • 8/2/2019 Bascula Verficadora

    50/90

    50

    Interface Description

    Purpose

    Your Economy/Express scale comes equippedwith an interface port for connection to a computeror other peripheral device.

    You can use an on-line computer to change, startand/or monitor the functions of the scale and theapplication programs (EB models only).

    Available Features

    Type of interface: Serial interface

    Operating mode: Full duplex

    Standard: RS-232

    Transmission rates:

    150; 300; 600; 1,200; 2,400; 4,800;9,600 baud

    Parity: Mark, space, odd, even

    Character format:1 start bit, 7-bit ASCII, parity, 1 or 2 stop bits

    Handshake:2-wire interface: via software (XON/XOFF)4-wire interface: via hardware handshake lines(CTS/DTR)

    Operating mode: SBI

    Data output format of the scale:16 or 22 characters

  • 8/2/2019 Bascula Verficadora

    51/90

    51

    Factory Settings:

    Transmission rate: 1,200 baud (5 1 4)

    Parity: Odd (5 2 3)

    Stop bits: 1 stop bit (5 3 1)

    Handshake:Hardware, 1 character after CTS (5 4 3)

    Operating mode: Standard SBI (5 5 1)

    Print manually/automatically:Manual after stability (6 1 2)

    Preparation

    See the chapter entitled Overview for the pinassignment chart

    Line Format (Data Output Format)

    You can output the values displayed in the measuredvalue line and the weight unit with or without a dataID code.

    Example: Without data ID code+ 253 pcs

    Example: With data ID code

    Qnt + 253 pcs

    Configure this parameter in the Setup menu(Menu: Printout format 7 1 1 or 7 1 2 or 7 1 3).

    The output without data ID code has 16 characters;with data ID code, 22 characters.

    !Note on Operation of the Scale with theYDP03-0CE Printer

    When data output with 22 characters per lineis configured (menu code 7 .1. 2), the GLP/GMPprintout setting in the printer must be set to off(-A1- 0 = off)

  • 8/2/2019 Bascula Verficadora

    52/90

    52

    Output Format With 16 Characters

    Display segments that are not activated are output as spaces. Characterswithout a decimal point are output without a decimal point.

    The following characters can be output, depending on the characters displayedon the scale:

    Normal Operation

    Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16+ D D D D D D * U U U CR LF

    or . . . . . . . . * * *

    or * * * * * * * * *

    or 0 0 0 0 0 0

    *: Space CR: Carriage returnD: Digit or letter LF: Line feedU: Unit symbol

    Special Codes

    Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16* * * * * * * * CR LF

    or H *

    or L *

    or C *

    *: Space H: Overload : Weight L: Underload

    C: Calibration/adjustment

  • 8/2/2019 Bascula Verficadora

    53/90

    53

    Error Codes

    Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16* * E * # # # * * * * CR LF

    *: Space# # #: Error code numberExample: Data output example: + 1255.7 g

    Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

    + * * * 1 2 5 5 . 7 * g * * CR LF

    Position 1: Plus or minus sign or spacePosition 2: SpacePosition 310: Weight with a decimal point; leading zeros = spacePosition 11: SpacePosition 1214: Unit symbol or spacePosition 15: Carriage returnPosition 16: Line feed

  • 8/2/2019 Bascula Verficadora

    54/90

    54

    Output Format with 22 Characters

    When data with an ID code is output, the ID code consisting of 6 charactersprecedes the data with the 16-character format. These 6 characters identify thefollowing value.

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22I I I I I I + * D D D D D D D D * U U U CR LF

    * * * * * . . . . . . . . * * *

    * * * * * * * * *

    0 0 0 0 0 0

    I: ID code character1) U: Unit symbol1)

    *: Space CR: Carriage returnD: Digit or letter LF: Line feed

    1) depends on scale type; e.g., not all units and characters are availableon scales verified for use in legal metrology (see also next page)

    Special Codes

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22S t a t * * * * * * * * * * * * * * CR LF

    H *

    L *

    *: Space H: Overload : Weight L: Underload

    Error Codes

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22S t a t * * * * * E r r * # # # * * * * CR LF

    *: Space # # #: Error code number

  • 8/2/2019 Bascula Verficadora

    55/90

    55

    ID code

    characters I MeaningStat Status

    G GrossG# Gross calculated valueT Tare T

    T1 Tare T1N Net N

    N1 Net N1

    Qnt Counting: quantityPrc Weighing in percent: percentageRes Calculation, averaging: resultSum Totalizing: net total

    wRef Automatic printout: average piece weightWxx% Automatic printout: reference percentage weight

  • 8/2/2019 Bascula Verficadora

    56/90

    56

    Data Input Format

    You can connect a computer to your scale to sendcommands via the scale interface port to controlscale functions and applications (EB models only).

    The commands sent are control commands andmay have different formats; e.g., control commandscan have up to 26 characters. Each charactermust be transmitted according to the settingsconfigured in the Setup menu for data transmission.

    Format for Control Commands

    Format : Esc ! CR LF

    Esc: Escape CR: Carriage return (optional)!: Command LF: Line feed (optional)

    character

    Command character ! MeaningK Weighing mode 1L Weighing mode 2

    M Weighing mode 3N Weighing mode 4

    O Block keysP PrintR Release keysS RestartT Tare and zero (combined)U Tare (Tare only)V Zero

    W External calibration/adjustment

  • 8/2/2019 Bascula Verficadora

    57/90

    57

    Synchronization

    During data communication between the scale andan on-line device (computer), messages consisting ofASCII characters are transmitted viathe interface. For error-free data communication,the parameters for baud rate, parity, handshakemode and character format must be the samefor both units.

    You can set these parameters in the Setup menu so

    that they match those of the on-line device. You canalso define parameters in the scale to makedata output dependent on various conditions. Theconditions that can be configured are describedunder each of the application program descriptions.

    If you do not plug a peripheral device into the scaleinterface port, this will not generate an errormessage.

    Handshake

    The scale interface (Sartorius Scale Interface = SBI)has transmit and receive buffers. You can define thehandshake parameter in the Setup menu:

    Hardware handshake (CTS/DTR) Software handshake (XON, XOFF)

    Hardware Handshake

    With a 4-wire interface, 1 more character can betransmitted after CTS (Clear to Send).

    Software Handshake

    The software handshake is controlled via XONand XOFF. When a device is switched on,XON must be transmitted to enable any connecteddevice to communicate.

  • 8/2/2019 Bascula Verficadora

    58/90

    58

    Activating Data Output

    You can define the data output parameter so thatoutput is activated either when a print command isreceived or automatically and synchronous with thescale display or at defined intervals (see applicationprogram descriptions (EB models only) and auto-printsetting).

    Data Output by Print CommandThe print command can be transmitted by pressingp/[PRINT] or by a software command (Esc P).

    Automatic Data Output

    In the auto print operating mode, data are outputto the interface port without a print command.

    You can choose to have data output automatically atdefined print intervals with or without the stabilityparameter. Whichever parameter you select, thedata will be output as the readouts appear onthe scale display. The display update frequencydepends on the setting for Adapting the Filter(1.1.x.).

    If you select the auto print setting, data will be

    transmitted immediately the moment you turn onthe scale.

  • 8/2/2019 Bascula Verficadora

    59/90

    59

    Pin Assignment Chart

    Female Interface Connector:25-position D-Submini, DB25S, with screw lock hardware for cable gland

    Male Connector Used: (please use connectors with the same specifications):25-pin D-Submini, DB25S, with integrated shielded cable clamp assembly(Amp type 826 985-1C) and fastening screws (Amp type 164 868-1)

    ! Warning When Using Pre-wired RS-232 Connecting Cables!RS-232 cables purchased from other manufacturers often have incorrect pinassignments for use with Sartorius scales. Be sure to check the pin assignment

    against the chart below before connecting the cable, and disconnect any linesmarked Internally Connected (e.g., pin 6). Failure to do so may damage or evencompletely ruin your scale and/or peripheral device.

    Pin Assignment Chart:

    Pin 1: Signal groundPin 2: Data Output (TxD)Pin 3: Data Input (RxD)Pin 4: Internal Ground (GND)

    Pin 5: Clear to Send (CTS)Pin 6: Internally ConnectedPin 7: Internal Ground (GND)Pin 8: Internal Ground (GND)Pin 9: Internally ConnectedPin 10: Not ConnectedPin 11: Not ConnectedPin 12: Reset _ Out *) For remote switchPin 13: Internally ConnectedPin 14: Internal Ground (GND)

    Pin 15: Universal Remote SwitchPin 16: Not ConnectedPin 17: Not ConnectedPin 18: Not ConnectedPin 19: Not ConnectedPin 20: Data Terminal Ready (DTR)Pin 21: Ground input for external power supply (GND _ V in)Pin 22: Internally ConnectedPin 23: Internally ConnectedPin 24: Power supply input +12 ... 30 V

    Pin 25: Not Connected*) = Hardware restart

  • 8/2/2019 Bascula Verficadora

    60/90

    60

    Cabling Diagram

    Diagram for interfacing a computer or different peripheral deviceto the scale using the RS-232/V24 standard and cables up to 15 m(50 ft.) long

    V24

    Scale Peripheral Device

    Type of cable: AWG 24 specification

    (PC)25 pin 9 pin

    2 3

    20 4

    3 2

    5 8

    6 6

    7 5

  • 8/2/2019 Bascula Verficadora

    61/90

    61

    Configuring the Scale

    Setting the Parameters (Menu)

    Purpose

    You can configure your Economy/Express scaleto meet individual requirements by selecting from theparameters available in the menu.

    Parameters that are not permitted in legal metrologyare not shown in the menu on verified scales.

    Features

    The parameters are divided into the followinggroups (1st menu level):

    1 Scale functions

    2 Application programs

    3 Application parameters

    5 Data interface6 Print for weighing

    7 Print for application

    8 Extra functions

    9 Reset menu to factory-set parameters

  • 8/2/2019 Bascula Verficadora

    62/90

    62

    Factory-Set Parameters

    The factory-set parameters are marked by an oin the list starting on page 64.

    Preparation

    Turn off the scale: Press e/[ON/OFF]

    Turn on the scale: Press e/[ON/OFF]; whileall segments are displayed, press)/[TARE] briefly

    > Measured value line: 1 (1st menu level)$ To navigate within a menu level: Press

    )/[TARE]; when you reach the last menu item,the first is shown again

    $ To select the next sublevel within a group:Press p/[PRINT] briefly.

    $ To return to the higher menu level:Press p/[PRINT]

    To confirm selected parameters: Press and holdp/[PRINT] for more than two seconds

    > o indicates the selected menu code

    To store parameter settings and exit the menu:Press and hold the )/[TARE] key

    To exit the menu without storing any new settings:Press e/[ON/OFF]

    > Restart the application

  • 8/2/2019 Bascula Verficadora

    63/90

    63

    Example

    Adapt the scale to unstable ambient conditions: Menu code 1.1. 4

    Step Key Display

    1. Turn the scale off e/[ON/OFF]

    2. Turn the scale back on; e/[ON/OFF]

    while all segments are )/[TARE]displayed briefly

    3. Confirm 1st menu level(scale functions) p/[PRINT]

    4. Confirm Adapt filter group(2nd menu level) p/[PRINT]

    5. Menu level 3: )/[TARE]Select the desired menu item repeatedly

    6. Confirm selected item p/[PRINT]for 2 seconds

    7. Set other parameters, p/[PRINT],if desired )/[TARE]

    8. Saves changes and )/[TARE]

    exit menu for 2 seconds

  • 8/2/2019 Bascula Verficadora

    64/90

    64

    Scale Operating Menu (Overview)

    Factory setting User setting

    Menu items that are only available in the EB and not in the EA model are marked by shading.

    Menu 1 Scale 1.1. Adapt filter 1.1.1 Minimum vibrationfunctions 1.1.2 o Normal vibration

    1.1.3 Strong vibration1.1.4 Extreme vibration

    1.2. Application filter 1.2.1 o Final readout1.2.2 Filling mode

    1.3. Stability range 1.3.1 1/4 digit1.3.2 1/2 digit1.3.3 1 digit*1.3.4 o 2 digits*1.3.5 4 digits*

    1.4. Weight unit for 1.4.1 Gramscalibration weight* 1.4.2 o Kilograms1.4.3 Pounds

    1.5. Adjustment* 1.5.1 o Accessible1.5.2 Blocked

    1.6. Auto zero 1.6.1 o On1.6.2 Off

    1.7. through 1.8. see next page

    2. through 9. see next pages

    Menu

    level1

    Menu

    level2

    Menu

    level3

    Factor

    ysetti

    ng

    Menu

    item

    * = These settings cannot be changed on verified scales

  • 8/2/2019 Bascula Verficadora

    65/90

    65

    Menu 1. Scale 1.1. through 1.6. see previous pagefunctions

    1.7. Weight unit 1* 1.7.2 Grams1.7.3 o Kilograms1.7.4 Carats1.7.5 Pounds1.7.6 Ounces1.7.7 Troy ounces1.7.8 Hong Kong taels1.7.9 Singapore taels1.7.10 Taiwanese taels1.7.11 Grains1.7.12 Pennyweights1.7.13 Milligrams1.7.14 Parts per pound1.7.15 Chinese taels1.7.16 Mommes1.7.17 Austrian carats1.7.18 Tola1.7.19 Baht1.7.20 Mesghal1.7.21 Tons1.7.22 Pounds+ounces

    (non-decimal display)

    1.8. Display accuracy 1* 1.8.1 o Standard resolution1.8.2 10x higher resolution**1.8.3 2x higher resolution

    (PolyRange)

    2. Application 2.1. Program selection 2.1.1 Weighing

    programs 2.1.2 Toggle weight units2.1.3 Net-total (tare memory)2.1.4 o Counting2.1.5 Weighing in percent2.1.7 Calculation by a factor2.1.8 Totalizing2.1.9 Gross/net toggling2.1.10 Tare memory with autom.

    container taring2.1.11 Weighing in percent with

    difference display

    2.1.12 Averaging3. through 9. see following pages

    ** = These settings cannot be changed on verified scales** = Only for EB-0UR models

    Menulev

    el1

    Menulev

    el2

    Menulev

    el3

    Factorys

    etting

    Menuite

    m

  • 8/2/2019 Bascula Verficadora

    66/90

    66

    Menu 1. through 2. see previous pages

    3. Application 3.1. Weight unit 2* 3.1.2 Gramsparameters 3.1.3 o Kilograms

    3.1.4 Carats3.1.5 # Pounds3.1.6 Ounces3.1.7 Troy ounces3.1.8 Hong Kong taels3.1.9 Singapore taels3.1.10 Taiwanese taels3.1.11 Grains3.1.12 Pennyweights3.1.13 Milligrams3.1.14 Parts per pound3.1.15 Chinese taels3.1.16 Mommes3.1.17 Austrian carats3.1.18 Tola3.1.19 Baht3.1.20 Mesghal3.1.21 Tons3.1.22 Pounds+ounces

    (non-decimal display)

    3.2. Display accuracy 2* 3.2.1 o Standard resolution3.2.2 10x higher resolution**3.2.3 2x higher resolution (PolyRange)

    3.3. Preset reference 3.3.1 5quantity, reference 3.3.2 o 10

    percentage or no. 3.3.3 20of subweighs 3.3.4 50for averaging 3.3.5 100

    5. Data 5.1. Baud rate 5.1.1 150interface 5.1.2 300

    5.1.3 6005.1.4 o 1,2005.1.5 2,4005.1.6 4,8005.1.7 9,600

    5.2. through 5.5. see next page6. through 9. see next page

    Menulev

    el1

    Menulev

    el2

    Menulev

    el3

    Factorys

    etting

    Menuite

    m

    # = Factory setting on scale model EA/EB ...UR** = These settings cannot be changed on verified scales** = Only for EB-0UR models

  • 8/2/2019 Bascula Verficadora

    67/90

    67

    Menu 5. Interface 5.1. through 5.3. see previous page

    5.2. Parity 5.2.1 Mark5.2.2 Space5.2.3 o Odd5.2.4 Even

    5.3. No. of stop bits 5.3.1 o 15.3.2 2

    5.4. Handshake mode 5.4.1 Software5.4.2 Hardware,

    1 char. after CTS5.4.3 o Hardware,

    2 char. after CTS

    5.5. Communication 5.5.1 o PC, YDP03 printer (SBI)mode 5.5.2 YDP01IS printer*

    6. Print for 6.1. Manual/auto print 6.1.1 Manual without stabilityweighing mode 6.1.2 o Manual after stability

    6.1.3 Automatic without stability6.1.4 Automatic at stability

    7. Print for 7.1. Printout format 7.1.1 No data ID codesapplication 7.1.2 With data ID codes

    7.1.3 o With data ID codes,N, T and G-values and2 header lines

    8. Extra 8.1. Menu** 8.1.1 o Parameter settingsfunctions alterable

    8.1.2 Parameter settings

    read only8.2. External switch 8.2.1 Zero setting and taring

    function combined8.2.3 w/[TOGGLE] key func.8.2.4 o p/[PRINT] key function8.2.5 (/[ZERO] key function8.2.6 c key function8.2.7 v key function8.2.8 )/[TARE] key function

    8.3. Power-on mode 8.3.1 Off/on

    8.3.2 o Stand-by/on

    9. Reset menu** 9. Factory settings 9. 1 Reset9. 2 o Do not reset

    Menulev

    el1

    Menulev

    el2

    Menulev

    el3

    Factorys

    etting

    Menuite

    m

    ** = To operate the YDP01IS, also set menu item 5.1.7 9,600 baud** = Not for EB-0UR

  • 8/2/2019 Bascula Verficadora

    68/90

    68

    Error Codes

    Error codes are shown in the main display for approx. 2 seconds, after which

    the program automatically returns to the weighing mode.

    Display Cause Solution

    No segments appear on No AC power is Check the ACthe display available power supply

    The AC adapter is not Plug in the AC adapterplugged in

    Automatic shutoff Turn on the scaleis configured

    Rechargeable battery Recharge batteryhas run down (see battery instructions)

    h The load exceeds the Unload the scalescale capacity

    l The load plate is Place the load platenot in place on the scale

    Something is touching Move the object that isthe load plate touching the load place

    e 01 Data output not Change the configurationcompatible with in the Setup menuinput format

    e 02 Calibration parameter Calibrate only whennot met; e.g.: zero is displayed scale not zeroed Press (/[ZERO]

    to zero the scale scale is loaded Unload the scale

    e 08 The scale was zeroed Only zero the scaleoutside the zero range when in the zero range;

    i.e.: ( 2% of themaximum capacity

    e 09 Taring is not possible Press (/[ZERO]when the gross weight to zero the scale

    is zero

  • 8/2/2019 Bascula Verficadora

    69/90

    69

    Display Cause Solution

    e 10 The )/[TARE] key Press c to clear theis blocked when there tare memory and releaseis data in the second the tare keytare memory (net-total) only 1 tare function canbe used at a time

    e 11 Tare memory not Press (/[ZERO]accessible

    e 22 Weight is too light or Increase the referencethere is no sample on quantitythe scale

    e 30 Interface port for printer Perform Reset menuoutput is blocked or

    Contact your localSartorius Service Center

    The maximum The scale was turned on Press the e/

    weighing capacity without the load plate [ON/OFF] key to turnis less than is indicated in place the scale off and backin Specifications on again

    The weight readout Unstable ambient Set up the scalechanges constantly conditions in another area

    Too much vibration, Change setupor the scale is exposed configuration to adaptto a draft the scale to the

    ambient conditionsA foreign object is caught Remove the foreignbetween the load plate objectand the scale frame

    The weight readout The scale has not Calibrate/adjustis obviously wrong been calibrated/adjusted the scale

    The scale was not Tare before weighingtared before weighing

    If any other errors occur, contact your local Sartorius Service Center!

  • 8/2/2019 Bascula Verficadora

    70/90

    70

    Care and Maintenance

    Service

    Regular servicing by a Sartorius technician willextend the service life of your scale and ensureits continued weighing accuracy. Sartorius canoffer you service contracts, with your choice ofregular maintenance intervals ranging from1 month to 2 years.The frequency of these maintenance intervalsdepends on the operating conditions and yourrequirements concerning the tolerances for the scale.

    Repairs

    Repair work must be performed by trained servicetechnicians. Any attempt by untrained personsto perform repairs may lead to hazards for the user.

    Cleaning

    !Disconnect the scale from the AC adapterand unplug any data cables that are connectedto the scale

    !Make sure that no liquids enter the scale housing

    !Do not use any aggressive cleaning agents(solvents or similar agents)

    Unplug the AC adapter from the wall outlet(mains supply)

    If you have a data cable connected to theinterface, unplug it from the scale

    Clean the scale using a piece of cloth which hasbeen wet with a mild detergent (soap)

  • 8/2/2019 Bascula Verficadora

    71/90

    71

    After cleaning, wipe down the scale with

    a soft, dry clothSafety Inspection

    If there is any indication that safe operationof the scale with the AC adapter is no longerwarranted:

    Turn off the power and disconnect the equipmentfrom AC power immediately

    > Lock the equipment in a secure place to ensurethat it cannot be used for the time being

    Safe operation of the scale with the AC adapter isno longer ensured when:

    there is visible damage to the AC adapter

    the AC adapter no longer functions properly

    the AC adapter has been stored for a relativelylong period under unfavorable conditions

    In this case, notify your nearest Sartorius ServiceCenter or the International Technical SupportUnit based in Goettingen, Germany. Maintenanceand repair work may only be performed byservice technicians who are authorized by Sartoriusand who

    have access to the required maintenance manuals

    have attended the relevant service training courses

    Instructions for Recycling the Packaging

    To ensure safe shipment, your scale has beenpackaged using environmentally friendly materials.After successful installation of the scale,

    you should return this packaging for recycling.For information on recycling options, includingrecycling of old weighing equipment, contact yourmunicipal waste disposal center or local recyclingdepot.

  • 8/2/2019 Bascula Verficadora

    72/90

    72

    Overview

    General View of the Scale

    No. Designation Order no.forreplacement

    1 Load plate EA/EB ... DCE: 69EA0011 EA/EB ... EDE: 69EA0004 EA/EB ... FEG: 69EA0017

    2 Display and control unit3 Toggle key

    (EB models only)4 Print key5 Function key

    (EB models only)6 Tare key

    No. Designation Order no.forreplacement

    7 Display8 Zero key9 Clear function key

    (EB models only)10 On/off key11 DC jack12 Leveling feet Set: 69EA002013 Data interface port

    Not shown:

    Protective cap forinterface port 69LC0084

    4*

    5*6*

    7*7

    89

    10

    3

    * = As shown in this illustration, there are some differences between the standardand the U.S./Canadian keypad versions.

    10

    9*8*

    4

    56

    1*

    2*

    3*

    1112

    12

    1213

    Standard keypad version: U.S./Canadian keypad version:

    *

  • 8/2/2019 Bascula Verficadora

    73/90

    73

    Description of the Keys

    w/[TOGGLE] key:Toggle

    For toggling the displaybetween weighed andcalculated values(counting result,

    percentage, calculatedresult)

    For changing the refer-ence quantity/percen-tage or numbers ofsubweighs for averaging

    v key:

    Start ApplicationProgram

    The procedure thatfollows after this key ispressed depends on theapplication programselected; for a completedescription, see the

    chapter entitledOperating the Scalein the correspondingsection.

    )/[TARE] key: Tare

    For taring the weightof containers so thatthe scale readoutshows the net weightof samples.

    p/[PRINT] key:Data Output

    Press this key to outputthe displayed values tovia the built-in interfaceto a Sartorius Data

    Printer or a computer.

    e/[ON/OFF] key:On/off switch

    Switches the displayon/off. The scaleremains in the standbymode.

    (/[ZERO] key: Zero

    Sets the readout to zero.The scale can onlybe zeroed if the load is 2% of the maximumcapacity.

    c key: Clear Function

    This key is generallyused to interrupt/cancelfunctions:

    Clear memory anddelete applicationdata

    Interrupt calibration/adjustment routines

  • 8/2/2019 Bascula Verficadora

    74/90

    74

    Specifications

    Standard modelsModel EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB

    3DCE-L 6DCE-L 15DCE-L35EDE-L 60EDE-L 60FEG-L 150FEG-L

    Weighing capacity kg 3 6 15 35 60 60 150lb 6.6 13.2 33 77 132 132 330

    Readability g 1 2 5 10 20 20 50

    Maximum load kg 6 12 30 70 120 120 300

    Tare range (by subtraction) kg 3 6 15 35 60 60 150

    Reproducibility g 1 2 5 10 20 20 50

    Linearity g 1 2 5 10 20 20 50

    Sensitivity drift within ppm/10+40C K 100 100 100 100 100 100 125

    External calibration weight kg 2 (M1) 2 (M1) 5 (M1) 10 (M1) 20 (M1) 20 (M1) 50 (M1)(of at least accuracy class...) lb 5 (M1) 5 (M1) 10 (M1) 20 (M1) 50 (M1) 50 (M1) 100 (M1)

    Net weight, approx. kg 5 5 5 8 8 24 24

    Dust and water protection ratingfor the weighing platformaccording to EN 60529 IP54

    Dust protection rating for thedisplay and control unitaccording to EN 60529 IP40

    AC power source/power requirements AC adapter, 230 or 115 V, +15% ... 20%

    Frequency 48 60 Hz

    Power consumption maximum 16 VA; average 8 VA

    Operating temperature range 10 ... +40C (263 ... 313 K, 14F ... 104F)

    Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels

    Display update (dependson the filter level selected) 0.1 0.4

    Hours of operation with fullycharged YRB 05 Z externalbattery pack, approx. 48 h

    Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces,Hong Kong taels, Singapore taels, Taiwanese taels, grains,pennyweights, milligrams, parts per pound, Chinese taels,mommes, Austrian carats, tola, baht, mesghal and tons

    Built-in interface RS-232

    Format: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or spaceTransmission rates: 150 to 9,600 baudHandshake mode: Software or hardware

  • 8/2/2019 Bascula Verficadora

    75/90

    75

    Model EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB

    6DCE-L 12DCE-L 30EDE-L 60EDE-L 60FEG-L 120FEG-L0UR 0UR 0UR 0UR 0UR 0UR

    Weighing capacity kg 6 12 30 60 60 120lb 13.2 26.4 66 132 132 264.5

    Readability g 1 2 5 10 10 20

    Maximum load kg 12 30 70 120 120 300

    Tare range (by subtraction) kg 6 12 30 60 60 120

    Reproducibility g 1 2 5 10 10 20

    Linearity g 1 2 5 10 10 20

    Sensitivity drift within ppm/+5+35C K 50 50 50 50 50 50

    External calibration weight kg 2 (M1) 5 (M1) 10 (M1) 20 (M1) 20 (M1) 50 (M1)(of at least accuracy class...) lb 5 (M1) 10 (M1) 20 (M1) 50 (M1) 50 (M1) 100 (M1)

    Net weight, approx. kg 5 5 8 8 24 24

    Dust and water protection ratingfor the weighing platformaccording to EN 60529 IP54

    Dust protection rating for thedisplay and control unit

    according to EN 60529 IP40AC power source/power requirements AC adapter, 230 or 115 V, +15% ... 20%

    Frequency 48 60 Hz

    Power consumption maximum 16 VA; average 8 VA

    Operating temperature range 10 ... +40C (263 ... 313 K, 14F ... 104F)

    Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels

    Display update(depends on the filter level selected) 0.1 0.4

    Hours of operation with fullycharged YRB 05 Z externalbattery pack, approx. 48 h

    Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces,Hong Kong taels, Singapore taels, Taiwanese taels, grains,pennyweights, milligrams, parts per pound, Chinese taels,mommes, Austrian carats, tola, baht, mesghal and tons

    Built-in interface RS-232Format: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or space

    Transmission rates: 150 to 9,600 baudHandshake mode: Software or hardware

  • 8/2/2019 Bascula Verficadora

    76/90

    76

    Model EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB

    3DCE-I 6DCE-I 15DCE-I 35EDE-I 60EDE-I 60FEG-I 150FEG-I(0UR) (0UR) (0UR) (0UR) (0UR) (0UR)

    Weighing capacity kg 3 6 15 35 60 60 150lb 6.6 13.2 33 77 132 132 330

    Readability g 0.1 0.2 0.5 1 2 2 5

    Maximum load kg 6 12 30 70 120 120 300

    Tare range (by subtraction) kg 3 6 15 35 60 60 150

    Reproducibility g 0.2 0.2 0.5 1 2 2 5

    Linearity g 0.2 0.4 1 2 4 4 10

    Sensitivity drift within ppm/+5+35C K 10 10 10 10 10 10 10

    External calibration weight kg 2 (F2) 2 (F2) 5 (F2) 10 (F2) 20 (F1) 20 (F2) 50 (F2)(of at least accuracy class...) lb 5 (F2) 5 (F2) 10 (F2) 20 (F2) 50 (F1) 50 (F2) 100 (F2)

    Net weight, approx. kg 5 5 5 8 8 24 24

    Dust and water protection ratingfor the weighing platformaccording to EN 60529 IP54

    Dust protection rating for thedisplay and control unit

    according to EN 60529 IP40AC power source/powerrequirements AC adapter, 230 or 115 V, +15% ... 20%

    Frequency 48 60 Hz

    Power consumption maximum 16 VA; average 8 VA

    Operating temperature range 10 ... +40C (263 ... 313 K, 14F ... 104F)

    Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels

    Display update(depends on the filter level selected) 0.1 0.4

    Hours of operation with fullycharged YRB 05 Z externalbattery pack, approx. 48 h

    Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces,Hong Kong taels, Singapore taels, Taiwanese taels, grains,pennyweights, milligrams, parts per pound, Chinese taels,mommes, Austrian carats, tola, baht, mesghal and tons

    Built-in interface RS-232Format: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or space

    Transmission rates: 150 to 9,600 baudHandshake mode: Software or hardware

  • 8/2/2019 Bascula Verficadora

    77/90

    77

    Models Verified by the Manufacturer with EC Type Approval

    Model EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB3DCE-L 6DCE-L 15DCE-L 30EDE-L 35EDE-L 60EDE-L0CE 0CE 0CE 0CE 0CE 0CE

    Type designation DS BH 300 DN BH 300 DN BH 300

    Accuracy class* l l l l l l

    Weighing capacity, max.* kg 3 6 15 30 35 60

    Scale interval d* (Readability) g 1 2 5 10 10 20

    Verification scale interval e* g 1 2 5 10 10 20

    No. of scale intervals* e 3,000 3,000 3,000 3,000 3,500 3,000

    Minimum capacity, min.* g 20 40 100 200 200 400

    Application range acc. to CD* kg 0.023 0.046 0.115 0.230 0.235 0.460

    Maximum load kg 6 12 30 70 70 120

    Tare range (by subtraction)) 100% of maximum weighing capacity

    Net weight, approx. kg 5 5 5 8 8 8

    Dust and water protection ratingfor the weighing platform IP54

    Dust protection rating for thedisplay and control unit IP40

    AC power source/power AC adapter, 230 or 115 V, +15% ... 20% requirements

    Frequency 48 60 Hz

    Power consumption maximum 16 VA; average 8 VA

    Operating temperature range 10 ... +40 C

    Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels

    Display update (dependson the filter level selected) 0.1 0.4

    Hours of operation with fullycharged YRB05Zexternal battery pack, approx. 48 h

    Weight unit Kilograms

    Built-in interface RS232CFormat: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or spaceTransmission rates: 150 to 9,600 baudHandshake mode: Software or hardware

    * CD = Council Directive 90/384/EEC on non-automatic weighing instruments used in the EU

    and the Signatories of the Agreement on the European Economic Area

  • 8/2/2019 Bascula Verficadora

    78/90

    78

    Model EA/EB EA/EB

    60FEG-L 150FEG-L0CE 0CE

    Type designation DQ BH 300 DQ BH 300

    Accuracy class* l l

    Weighing capacity, max.* kg 60 150

    Scale interval d* (Readability) g 20 50

    Verification scale interval e* g 20 50

    No. of scale intervals* e 3,000 3,000

    Minimum capacity, min.* g 400 1,000

    Application range acc.