Sartorius Economy SeriesSartorius Express Series
EA and EB Models
Electronic Precision Scales
Installation and Operating Instructions
98648-007-09
IndustrialWeighi
ngTechnology
-
8/2/2019 Bascula Verficadora
2/90
2
Intended Use
The Economy/Express Series from Sartorius offers precision scales for weighing,
with capacities ranging from 0.1 g to 150 kg.Economy/Express scales meet the highest requirements on the accuracy andreliability of weighing results through the following features: Efficient filtering-out of unfavorable ambient conditions, such as vibrations, drafts etc. Stable and reproducible results Excellent readability under any lighting conditions Rugged, durable weighing system
Type of scale housing protection The weighing platforms are IP54-protected
The display and control units are IP40-protectedEconomy/Express scales save work and speed up simple routineapplications through: Ultrafast response times Easy operation Computer connectivity through a built-in RS-232 interface Two lines in the printout are configurable to show, e.g., your company name
Additional Features of the EB Models:With the EB model, you can choose from the following extra functions
for simple applications: Counting Toggling between weight units (second weight unit) Net-total formulation (tare memory) Second tare memory (incl. automatic container taring function) Weighing in percent (incl. display of the difference between current weight
and stored reference weight) Averaging Calculation by a factor
Totalizing Gross/net toggling
Note to Users in the U.S.This equipment has been tested and found to comply with the limits for a Class Adigital device, pursuant to Part 15 of the FCC Rules. These limits are designedto provide reasonable protection against harmful interference when the equipmentis operated in a commercial environment. This equipment generates, uses, and canradiate radio frequency energy and, if not installed and used in accordance withthe instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmfulinterference in which case the user will be required to correct the interference at hisown expense. Changes or modifications not expressly approved by Sartorius AGcould void the users authority to operate the equipment.
-
8/2/2019 Bascula Verficadora
3/90
3
Contents
Intended Use................................................. 2
Contents....................................................... 3
Warnings and Safety Precautions ..................... 4
Operating Design .......................................... 5
Getting Started .............................................10
Operating the ScaleBasic Weighing Function.............................14Calibration/Adjustment ...............................17
Application Programs in the EB ModelsCounting ................................................. 20Toggle between Weight Units .......................23Second Tare Memory .................................27Weighing in Percent ...................................30Calculation by a Factor ...............................33Totalizing ..................................................36Gross/Net Toggling ...................................39Averaging.................................................41
Data Output Functions ...................................45Data Interface............................................50Pin Assignment Chart ..................................59Cabling Diagram .......................................60
Configuring the Scale (Menu)Setting Parameters in the Operating Menu ......61Scale Operating Menu (Overview) ................64
Error Codes .................................................68
Care and Maintenance .................................70Instructions for Recycling.................................71
OverviewGeneral View of the Scale .......................... 72Description of the Keys ................................73Specifications ............................................74Dimensions (Scale Drawings)........................79Accessories (Options)..................................80Information on EC Verification ......................82Declarations of Conformity...........................85EC Type Approval ......................................88Index .......................................................89
-
8/2/2019 Bascula Verficadora
4/90
4
Warnings and Safety Precautions
This scale has been constructed in accordance with
the European Directives as well as internationalregulations and standards for operation ofelectrical equipment, electromagnetic compatibility,and stipulated safety requirements. Improper useor handling, however, can result in damageand/or injury.
Read these operating instructions thoroughly beforeusing your scale to prevent damage to theequipment. Keep these instructions in a safe place.
Follow the instructions below to ensure safe andtrouble-free operation of your scale:
!Do not use this scale in a hazardousarea/location
!Make sure that the voltage rating printed on theAC adapter is identical to your local line voltage
The only way to turn the power off completely is
to disconnect the AC adapter The scale housing is IP54-protected against dust
deposits and water splashes
The display and control unit is IP40-protectedagainst penetration of solid foreign objects
Connect only Sartorius accessories and options,as these are optimally designed for use with yourEconomy/Express scale
Protect the AC adapter and the display andcontrol unit from contact with liquids
When cleaning your scale, make sure that noliquids enters the scale housing; use only a slightlymoistened cloth to clean the scale.
Do not open the scale housing. If the seal isbroken, this will result in forfeiture of all claims underthe manufacturers warranty.
In case you have any problems with your scale:
$ Contact your local Sartorius office, dealer orservice center
-
8/2/2019 Bascula Verficadora
5/90
5
Operating Design
The scales in the Economy/Express Series consist of
a weighing cell and a display and control unit.In addition to the choice of power supply (viaAC adapter or external rechargeable battery pack),
your scale also has an interface port for connectingperipheral devices, such as a printer, computer,universal remote control switch, etc.
The display and control unit is affixed to theweighing platform. Operation of the Economy
scales is simple and uniform.The EB models include all the features of theEA models plus application programs for togglingbetween weight units, counting, weighing inpercent, etc.
Where not expressly stated otherwise, theinformation contained here also applies to theverified model versions in the European Union
and European Economic Area (indicated by the-L0CE suffix to the model name).
Keys
The Economy/Express Series scales can beoperated using either the keys on the display andcontrol unit or a connected computer (PC).Some of the keys activate different functions,
depending on how long they are pressed (pressbriefly = < 2 sec.; press and hold = 2 sec.).
The equipment supplied includes adhesive labelsprinted in different languages for identifying thefunctions assigned to the keys.
EA Model
The EA model has 4 keys. The function activated
by each key will differ according to whetherthe scale is in the weighing mode or setup mode.
-
8/2/2019 Bascula Verficadora
6/90
6
EB Model
The EB model has 3 additional keys (gray), forcontrolling the application programs.
Display
The display is divided into 5 sections:
Basis-Einheiten +Zusatzkennungen
Hinweise f rAnwendungs-programme
Einheiten-pfeile
Mewert
Busy-ZeichenVorzeichenStillstand
Busy Symbol, Plus/Minus Sign, Stability Symbol
If the symbol displayed here is a trianglev, this indicates that the scale is
performing a function (busy symbol) a plus or minus sign (+ or ), this applies
to the value displayed (on the EB models, thisalso applies to calculated values; e.g.,during counting)
a small S, this indicates that the verified scale
has been zeroed/tared a flashing triangle w, this indicates that thecalculated values are not valid for legal metrology
Measured Value
When the scale is in the weighing mode, theweight is displayed here. In the setup mode, themenu code numbers are shown here.On EB models, calculated values and applicationparameter settings are also displayed here.
Busy symbolPlus/minus signStability symbol
Applicationinformation
Unitindicator
Basic unit +additional information
Measured value
-
8/2/2019 Bascula Verficadora
7/90
7
Basic Unit and Additional Information
When the scale has stabilized, the weight unitis displayed here (g or kg). On the EB model,the calculated values (e.g., % or pcs) are alsodisplayed.
Application Information
On EB models, this section shows informationconcerning the current application program; e.g.,
the reference quantity for counting.
Unit Indicator
You can affix one of several adhesive overlays to theright of this section. These overlays show theoptional weight units selected in the scale Setupmenu. The arrows indicate the selected units.
Input FunctionsWhen the scale is in the setup mode, you can selectparameters from lists. The lists are contained in amenu, which is divided into 3 levels.
To set parameters: Turn the scale off and backon again. While all segments are displayed, press)/[TARE] briefly.
To navigate within a menu level: Press )/[TARE]briefly; when you reach the last menu item, the first isshown again.
To change menu levels: Press p/[PRINT] briefly.
To confirm the selected parameters on all 3 levels:Press and hold the p/[PRINT] key.
o indicates the selected parameter option.
-
8/2/2019 Bascula Verficadora
8/90
8
All parameter options are listed in detail under
Configuring the Scale.To store parameter settings and exit the menu:Press and hold the )/[TARE] keyParameter settings are stored in the non-volatilememory. When you turn on the scale, thelast parameters used are active.
To exit the menu without storing any new settings:Press e/[ON/OFF]
Output Functions
Your Economy/Express scale is equippedwith a data interface for connecting your choiceof the following:
Printer
PC
Universal remote control switch
Printer
You can configure a variety of options for dataoutput to a printer.
You can have a printout generated automatically oronly when the p/[PRINT] key is pressed. You canmake the print function dependent on or indepen-dent of stability parameters, and define whetherdata ID codes will be included in the printout or not.
The print functions are described in detail underData Output Functions in the chapter entitledOperating the Scale.
-
8/2/2019 Bascula Verficadora
9/90
9
Data Interface
Instead of a printer, you can connect a differentperipheral device, such as a computer, to the datainterface.
You can use a connected computer to control theEconomy/Express scale. Communication betweenthe scales data interface and a PC is in the formof request and response messages. Not all requestmessages will activate response messages.
See Data Output Functions in the chapter entitledOperating the Scale for a detailed description.
Error Messages
Error messages are displayed for 2 seconds.The format is as follows:
Sequence error: E + 2 digits Hardware error: E + 3 digits
The section entitled Error Messages containsdetailed information.
-
8/2/2019 Bascula Verficadora
10/90
10
Getting Started
Warranty
Do not miss out on the benefits of our full warranty.Complete the warranty registration card, indicatingthe date of installation, and return the card to
your Sartorius office or dealer.
Storage and Shipping Conditions
Do not expose the scale to shocks, vibrations,
moisture or extreme temperatures.
Unpacking the Scale
After unpacking the scale, check it immediatelyfor any visible damage as a resultof rough handling during shipment.
$ Note:The display and control unit is attached to theweighing platform via a cable.
$ If you see any sign of damage, proceedas directed in the chapter entitled Care andMaintenance, under the section on SafetyInspection.
Save the box and all parts of the packaging untilyou have successfully installed your scale. Only the
original packaging provides the best protection forshipment. Before packing your scale,unplug all connected cables to prevent damage.
Equipment Supplied
The equipment supplied includes the componentslisted below:
Scale with attached display and control unit
Load pan
AC adapter
Installation and operating instructions
-
8/2/2019 Bascula Verficadora
11/90
11
Installation Instructions
The Sartorius Economy/Express scales are designedto provide reliable weighing results under normalambient conditions. When choosing a location toset up your scale, observe the following sothat you will be able to work with added speedand accuracy:
Set up the scale on a stable, even surface Avoid placing the scale in close proximity
to a heater or otherwise exposing the scaleto heat or direct sunlight Protect the scale from drafts that come from
open windows or doors Avoid exposing the scale to extreme vibrations
during weighing Protect the scale from aggressive
chemical vapors Do not expose the scale to extreme moisture
over long periodsConditioning the Scale:Moisture in the air can condense on the surfaces ofa cold scale whenever it is brought into asubstantially warmer place. If you transfer the scaleto a warmer area, make sure to condition it forabout 2 hours at room temperature, leaving itunplugged from AC power. Afterwards, keep thescale continuously connected to AC power.
Seal on Scales Verified for Use in Legal Metrologyin the EU*:
EU legislation requires that a control seal beaffixed to the verified scale. The control sealconsists of a sticker with the Sartorius logo. Thisseal will be irreparably damaged if you attemptto remove it. If the seal is broken, the validity
of the verification will become void, and you musthave your scale re-verified.
* including the Signatories of the Agreementon the European Economic Area
-
8/2/2019 Bascula Verficadora
12/90
12
Place the load plate on the scale
Connecting the Scale to AC Power Check the voltage rating and the plug design
If they do not match the rating or standard youuse, contact your Sartorius office or dealer
Use only
Original Sartorius AC adapters
AC adapters with a registered approval ratingfrom a national testing laboratory
$ See the chapter entitled Accessories forinformation on using an external rechargeablebattery pack with your scale
Insert the right-angle plug into the jack and thentighten the screws
Then insert the plug of the AC adapter intoa wall outlet (mains)
Safety Precautions
The AC adapter rated to Class 2 can be pluggedinto any wall outlet without requiring any additionalsafety precautions. The pole of the output voltage
is connected to the scale housing, which can begrounded for operation. The data interface isalso electrically connected to the scale housing(ground).
EA/EBDCE/EDE:
EA/EBFEG:
-
8/2/2019 Bascula Verficadora
13/90
13
Connecting Electronic Peripheral Devices
Make absolutely sure to unplug the scalefrom AC power before you connect ordisconnect a peripheral device (printer or PC)to or from the interface port.
Warmup Time
To deliver exact results, the scale must warm up for
at least 30 minutes after initial connection to ACpower. Only after this time will the scale havereached the required operating temperature.
Leveling the Scale
Remove the load plate
Loosen the lock nuts on all leveling feet(use a 13 mm open-end wrench if necessary)
Turn the leveling feet until the scale is level
Avoid unintentional changes in the adjustment:Tighten the lock nuts with an open-end wrench
and then replace the load plate
Leveling the Scale (for EA/EB-I, -L0CEmodels only)
Using the 4 footscrews, level the scale so that theair bubble is centered within the level indicator
-
8/2/2019 Bascula Verficadora
14/90
14
Operating the Scale
Basic Weighing Function
Purpose
The basic weighing function is always accessible in theEA and EB scales.
In the EB model, this function can be used aloneor in combination with an application program(Toggle between Weight Units, Counting, Weighingin Percent, etc.).
Features
Zeroing the scaleDepending on prevailing ambient conditions,the display may not show a zero readout eventhough there is no load on the scale. If theweight shown is less than 2% of the scales
maximum weighing capacity, you can zerothe scale. Taring the scale
Tare the scale with an empty container onthe weighing pan to obtain a readout of the netweight after the container is filled.
Printing weights
Factory SettingsWeight unit 1: Kilogram (1 7 3)
Manual/automatic printing:Manual after stability (6 1 2)
Print format:Gross, tare and net values with data ID codes (7 1 3)
-
8/2/2019 Bascula Verficadora
15/90
15
Preparation
Turn on the scale: Press e/[ON/OFF]
> All segments light up briefly
$ To change configurations: See the chapterentitled Configuring the Scale
$ To load factory-set configurations: See Configuringthe Scale, parameter 9 - 1
> TheS
symbol is displayed when the scale iszeroed or tared (only on scales verified for use inlegal metrology)
Additional Functions in EB models:
Start an application
Toggle the display between weight and
calculated value Delete an application
-
8/2/2019 Bascula Verficadora
16/90
16
Practical Examples
Simple weighing
Menu code settings: Factory-set codes
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF] 0.000 kg
2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)
3. Place the container on the scale + 0.015 kg(here: 0.015 kg)
4. Tare the scale )/[TARE] 0.000 kg
5. Place a sample in the container + 0.125 kgon the scale (here: 0.125 kg)
6. Print weight* p/[PRINT] EISENMUELLERGOETTINGEN
N + 0.125 kgT + 0.015 kgG# + 0.140 kg
* The Sartorius Service Centeror your Sartorius dealer canconfigure the scale to include2 customer-specific lines on theprintout. Software is available forsetting this configuration with a PC.
-
8/2/2019 Bascula Verficadora
17/90
17
Calibration/Adjustment
Purpose
Calibration is the determination of the differencebetween the weight readout and the trueweight (mass) of a sample. Calibration does notentail making any changes within the scale.Adjustment is the correction of this difference betweenthe measured value displayed and the true weight(mass) of the sample, or the reduction of the differenceto an allowable level within the maximum permissibleerror limits.
Features
External adjustment can only be performed when there is no load on the scale, the scale is set to zero, and the internal signal is stable.
If these conditions are not met, an error messageis displayed.
The weight on the scale must not differ from thenominal weight by more than 2%.You can use g, kg or lb as the weight unit forcalibration (1 4)You can block calibration of the scale (1 5)
Calibration/Adjustment for Verified Scales The calibration function is blocked in verified
scales (sealed switch cover)> You can only perform calibration by removing
this seal. In this case, verification is no longervalid and you must have your scale re-verified.
Factory Settings
Weight unit for calibration: kg (1 4 2)Calibration function: accessible (1 5 1)
-
8/2/2019 Bascula Verficadora
18/90
18
Example
Calibrate the scale
Menu code settings: Factory-set codes
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF] 0.000 kg
2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)
3. Begin calibration )/[TARE] + 5.000Calibration weight is > 2 sec.displayed without wt. unit
4. Place the indicated calibration 5.000
weight on the scale(here: 5,000 g)
After calibration, the calibration + 5.000 kgweight is displayed with wt. unit
5. Remove the calibration weight 0.000 kg
-
8/2/2019 Bascula Verficadora
19/90
19
Application Programs (EB Models Only)
Note Concerning Verified Scales Used as LegalMeasuring Instruments in the EU*:
All application programs can also be used inlegal metrology. Calculated weight values areindicated by one of the following symbols,displayed to the right of the numerical value:
Percentage = %
Piece count = pcs
Other calculated value = o
A flashing triangle w displayed to the left of thenumerical value indicates the value is notvalid for legal metrology (e.g., in averagingor calculation).
* including the Signatories of the Agreement on theEuropean Economic Area
-
8/2/2019 Bascula Verficadora
20/90
20
Counting (EB Models Only)
Purpose
With the Counting program you can determinethe number of parts that have approximately thesame average piece weight.
Available Features
The reference sample quantity can be changedeither in the operating menu or during weighing
Store the current weight value to have it loadedas the preset reference sample quantity the nexttime you initialize the Counting application
Re-initialize without quitting the program
The average piece weight is automatically outputvia the data interface port after initialization,
if the menu code for Printout with data ID codesis set
Press w/[TOGGLE] to toggle the displaybetween piece count and weight
Factory Settings
Display accuracy 2: Standard resolution (3 2 1)
Reference sample quantity: 10 (3 3 2)
Preparation
Configure the Counting application in theoperating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; whileall segments are lit, press )/[TARE]
-
8/2/2019 Bascula Verficadora
21/90
21
Select the Counting menu item:
Press )/[TARE], p/[PRINT],and p/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 4 Counting
Confirm the selection of Counting:Press p/[PRINT] for > 2 sec.
> o indicates that this is the current menucode setting
Set the next parameters: Press p/[PRINT]
Select and confirm: Reference sample quantity:
3 3 1 5 pcs3 3 2 10 pcs3 3 3 20 pcs3 3 4 50 pcs3 3 5 100 pcs
see also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale
Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:
Initialize application (v key) Delete initialization values (c key) Change reference sample quantity after deleting
initialization value (w/[TOGGLE] key, press for
at least 2 sec.) Display weight (w/[TOGGLE] key) Calibrate/adjust scale ()/[TARE], press for
at least 2 sec.)
-
8/2/2019 Bascula Verficadora
22/90
22
Example
Determine an unknown piece count; store the weight on the scaleas a reference sample quantity
Settings (changes in the factory settings required for this example):Menu: Application program: Counting (2 1 4)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. If necessary, zero the scale (/[ZERO] 0.000 kg10
(S symbol on the left:verified scale is zeroed)
3. Select the reference sample w/[TOGGLE] ref 20 (briefly)quantity (here: 50 pcs) w/[TOGGLE] ref 50 (briefly)
4. Place the reference sample + 0.930 kg50quantity (50 pcs) on the scale(here: 0.930 kg)
5. Start the application; if the print v + 50 pcs50format is set to include dataID codes, the following is printed wRef + 0.186 kg
6. Display weight w/[TOGGLE] + 0.930 kg50
7. Display quantity w/[TOGGLE] + 50 pcs50
8. Weigh uncounted parts + 174 pcs50(here: 174 pcs)
9. Print total piece count p/[PRINT] EISENMUELLERGOETTINGEN
Qnt + 174 pcs
10. Unload the scale0
pcs50
11. Repeat the procedure startingfrom Step 8, if desired.
-
8/2/2019 Bascula Verficadora
23/90
23
Toggle between Weight Units (EB Models only)
Purpose
With this application program you can toggle thedisplay of a weight value back and forth between twoweight units.
Available Features
Toggling the displayed weight Other features as for the basic weighing function
Factory Settings
Weight unit 1: kg (1 7 3)
Weight unit 2: kg (3 1 3)
Preparation
Configure the Toggle weight units application in theoperating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; while
all segments are lit, press )/[TARE]
Select the Toggle weight units menu item:Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 2 Toggle weight units
Confirm selection: Press p/[PRINT]for at least 2 sec.
> o indicates that this is the current menucode setting
-
8/2/2019 Bascula Verficadora
24/90
24
Set the next parameters: Press p/[PRINT]
Select and confirm:
Weight unit 1: see next page (1 7 x)
Display resolution 1:1 8 1 Standard resolution1 8 2 10x higher resolution**1 8 3 2x higher resolution (PolyRange)*
Weight unit 2: see next page (3 1 x)
Display resolution 2:3 2 1 Standard resolution3 2 2 10x higher resolution**3 2 3 2x higher resolution (PolyRange)*
see also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale
Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the functions for:
power off (e/[ON/OFF] key), zeroing ((/[ZERO] key),
taring ()/[TARE] key), and printing (p/[PRINT] key),
you can also access the following functions fromthis application:
toggle between weight units/and 2(w/[TOGGLE] p/[PRINT] or v key),
calibration/adjustment ()/[TARE] key; press forat least 2 sec.)
** = not applicable for verified balances** = only for EB-0UR models
-
8/2/2019 Bascula Verficadora
25/90
25
You can affix one of several adhesive overlays to the right of this section.
These overlays show the optional weight units selected in the scale Setup menu.The arrows indicate the selected units.
These settings cannot be changed on verified scales.
Menu code Unit Conversion factor Printout(1 7 2) (3 1 2) Grams 1.00000000000 g(1 7 3) (3 1 3) Kilograms 0.00100000000 kg
(1 7 4) (3 1 4) Carats 5.00000000000 ct(1 7 5) (3 1 5) Pounds 0.00220462260 lb(1 7 6) (3 1 6) Ounces 0.03527396200 oz(1 7 7) (3 1 7) Troy ounces 0.03215074700 ozt(1 7 8) (3 1 8) Hong Kong taels 0.02671725000 tlh(1 7 9) (3 1 9) Singapore taels 0.02645544638 tls(1 7 10) (3 1 10) Taiwanese taels 0.02666666000 tlt(1 7 11) (3 1 11) Grains 15.43235835000 GN
(1 7 12) (3 1 12) Pennyweights 0.64301493100 dwt(1 7 13) (3 1 13) Milligrams 1000.00000000000 mg(1 7 14) (3 1 14) Parts per pound 1.12876677120 /lb(1 7 15) (3 1 15) Chinese taels 0.02645547175 tlc(1 7 16) (3 1 16) Mommes 0.26670000000 mom(1 7 17) (3 1 17) Austrian carats 5.00000000000 K(1 7 18) (3 1 18) Tola 0.08573333810 tol(1 7 19) (3 1 19) Baht 0.06578947437 bat(1 7 20) (3 1 20) Mesghal 0.21700000000 MS(1 7 21) (3 1 21) Tons 0.00000100000 T(1 7 22) (3 1 22) lb/oz 0.03527396200 o
If you select menu code 1 7 22 or 3 1 22, you can affix the label with the lbweight unit under the left number, and the label with the oz weight unit underthe number on the right.
-
8/2/2019 Bascula Verficadora
26/90
26
Example
Toggle the display from kilograms [kg] (1st unit) to pounds [lb] (2nd unit)
Settings (changes in the factory settings required for this example):Menu: Application program: Toggle between weight units (2 1 2)Menu: Weight unit 2: pounds (3 1 5)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. Place weight on scale(here: 2.295 kg)
3. Toggle to weight unit w/[TOGGLE]2: Pounds [lb]
4. Print weight p/[PRINT] EISENMUELLERGOETTINGEN
G + 5.060 lb
5. Toggle to weight unit w/[TOGGLE]1: kilograms [kg]
-
8/2/2019 Bascula Verficadora
27/90
27
Second Tare Memory (EB Models Only)
Purpose
With this application program you can store the weighton the scale as a tare compensation weight.
Available Features
Press v to store the weight on the scale in thesecond tare memory
NET is displayed next to the net weight readout ifthe scale has been tared or if a tare value isstored in the second tare memory (N in theprintout indicates that the scale was tared bypressing the )/[TARE] key; N1 indicates thatthere is data in the second tare memory)
You can have a series of containers tared in
sequence, if the scale is unloaded to under 30%of the previous weight after measurement andthen loaded with 70% 130% of the previousweight (i.e., the next container)
Press c to delete the value from the secondtare memory
Press w/[TOGGLE] to toggle betweenweight unit 1 and weight unit 2
Preparation
Configure the Second tare memory application inthe operating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; while
all segments are lit, press )/[TARE]
-
8/2/2019 Bascula Verficadora
28/90
28
Select the Second tare memory menu item:
Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 3 Second tare memory2 1 10 Second tare memory with automatic
container taring
Confirm selection of Second tare memory:Press p/[PRINT] for at least 2 sec.
> o indicates that this is the current menucode setting
Set the next parameters: Press p/[PRINT]
Select and confirm: Weight unit 1: (see also Toggle between
Weight Units)1 7 2 Grams... ...
1 7 22 lb/oz Weight unit 2: (see also Toggle between
Weight Units)3 1 2 Grams... ...3 1 22 lb/oz
see also the Scale Operating Menu(Overview) in the chapter entitled Configuring
the Scale Save settings and exit the Setup menu:
Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:
Store weight in second tare memory (v key) Delete tare memory (c key) Toggle weight unit (w/[TOGGLE] key)
-
8/2/2019 Bascula Verficadora
29/90
29
Example
Fill containers with nearly identical weights ( 30 %)
Settings (changes in the factory settings required for this example):Menu: Application program: Second tare memory with automatic containertaring (2 1 10)Menu: Weight unit 2: Grams (3 1 2)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)
3. Place an empty container on + 0.065 kgthe scale (here: 65 g)
4. Store the container weight in v 0.000 kgNET
the second tare memory.If the print format is set to includedata ID codes, the following N1 + 0.065 kgis printed
5. Fill the container + 0.500 kgNET
(here: to 500 g)
6. Display weight unit 2 w/[TOGGLE] + 500 gNET
7. Display weight unit 1 w/[TOGGLE] + 0.500 kgNET
8. Remove the filled container 0.065 kgNET
from the scale
9. Place the next empty container 0.000 kgNET
on the scale (here: 75 g) (automatic container
taring is performed)10. Repeat the procedure starting
from Step 5.
-
8/2/2019 Bascula Verficadora
30/90
30
Weighing in Percent(EB Models Only)
Purpose
This application program allows you to obtainweight readouts in percent which are in proportionto a reference weight.
Available Features
Reference percentage can be changed either inthe operating menu or during weighing
Store the current weight value to have it loadedas the preset reference percentage the next time
you initialize the Weighing in Percent application
Optional auto-zero function
The reference sample weight is automatically
output via the data interface port afterinitialization, if the menu code for Printout withdata ID codes is set
Press w/[TOGGLE] to toggle the displaybetween percentage and weight
Factory Settings
Display accuracy 2: Standard resolution (3 2 1)
Reference percentage: 10 (3 3 2)
Preparation
Configure the Weighing in Percent application inthe operating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF];while all segments are lit, press )/[TARE]
-
8/2/2019 Bascula Verficadora
31/90
31
Select the Weighing in Percent menu item:
Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 5 Weighing in percent without taring2 1 11 Weighing in percent with taring
Confirm selection of Weighing in Percent:Press p/[PRINT] for at least 2 sec.
> o indicates that this is the current menu
code setting
Set the next parameters: Press p/[PRINT]
Select and confirm: Reference percentage at power-on:
3 3 1 5 %3 3 2 10 %3 3 3 20 %3 3 4 50 %3 3 5 100 %
see also the Scale Operating Menu (Overview)in the chapter entitled Configuring the Scale
Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:
Initialize application (v key) Delete initialization values (c key) Change reference percentage after deleting
initialization value (w/[TOGGLE] key, press forat least 2 sec.)
Display weight (w/[TOGGLE] key) Calibrate/adjust scale ()/[TARE], press forat least 2 sec.)
-
8/2/2019 Bascula Verficadora
32/90
32
Example
Determine an unknown percentage; store the weight on the scaleas a reference percentage
Settings (changes in the factory settings required for this example):Menu: Application program: Weighing in percent (2 1 5)Menu: Reference percentage 100 % (3 3 5)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)
3. Place the reference weight + 2.295 kgon the scale(here: 2.295 kg = 100%)
4. Start the application; if the print v + 100.00 %format is set to include dataID codes, the following is printed Wxx% + 2.295 kg
5. Display weight w/[TOGGLE] + 2.295 kg
6. Display percentage w/[TOGGLE] + 100.00 %
7. Display the reference v ref100percentage (at least 2 sec.)
8. Place an unknown weight + 140.41 %on the scale(here: 3.225 kg)
9. Print percentage p/[PRINT] Prc + 140.41 %
10. Unload the scale 0.00 %
11. Repeat the procedure startingwith Step 8, if desired.
-
8/2/2019 Bascula Verficadora
33/90
33
Calculation by a Factor (EB Models Only)
Purpose
With this application program you can calculate theweight of a sample by a given factor.
Available Features
Calculate a weight value using one of the
following factors: 0.25; 0.50. 0.75; 1.0; 1.5;2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0
A flashing triangle in the display indicatesa calculated value
Press c to delete the calculation factor
Press w/[TOGGLE] to toggle between weightunit 1 and weight unit 2
Factory Settings
Weight unit 1: kg (1 7 3)
Weight unit 2: kg (3 1 3)
Display accuracy 2: Standard resolution (3 2 1)
Preparation
Configure the Calculation application in the
operating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; whileall segments are lit, press )/[TARE]
Select the Calculation menu item:Press )/[TARE], p/[PRINT], and
p/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 7 Calculation by a factor
-
8/2/2019 Bascula Verficadora
34/90
34
Confirm selection of Calculation:
Press p/[PRINT] for at least 2 sec.> o indicates that this is the current menu
code setting
Set the next parameters: Press p/[PRINT]
Select and confirm: Weight unit 1: (see also Toggle between
Weight Units)1 7 2
Grams... ...1 7 22 lb/oz
Weight unit 2: (see also Toggle betweenWeight Units)3 1 2 Grams... ...3 1 22 lb/oz
see also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale
Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the basic functions (power off,
zeroing, taring and printing), you can also accessthe following functions from this application:
Initialize application (v key) Change the calculation factor after the
application is initialized (v key) Delete initialization values (c key) Toggle weight unit (w/[TOGGLE] key) Calibration/adjustment ()/[TARE] key;
press for at least 2 sec.)
-
8/2/2019 Bascula Verficadora
35/90
35
Example
You have a formula requiring several different components for a total amount of1,000 g. You want to use this recipe to make a 500 g batch (factor: 0.5).The scale shows the amounts called for in the formula, although only half theamounts are placed on the scale. This saves you having to recalculate theamount of each component.
Settings (changes in the factory settings required for this example):Menu: Application program: Calculation (2 1 7)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed
verified scales only)3. Place an empty container on + 0.065 kg
the scale (here: 65 g)
4. Tare the scale )/[TARE] 0.000 kgNET
5. Start calculation v 0.000 kgNET
.25
6. Select the calculation factor v 0.000 kgNET
.50
7. Add first component 0.240 kgNET
.50
(amount called for in the formula:240 g; actual amount: 120 g)
8. Add other components 1.000 kgNET
.50
(until the display indicates1,000 g)
9. Print weight of components p/[PRINT] Res + 1.000 kg
-
8/2/2019 Bascula Verficadora
36/90
36
Totalizing (EB Models Only)
Purpose
This application program acts as a cumulativememory function.
Available Features
The current weight is stored as a net value when
the scale stabilizes A value indicated as negative () in the
display is stored as a positive (+) value in thetotalizing memory
The data stored in the totalizing memory isautomatically output via the data interface portafter initialization, if the menu code forPrintout with data ID codes is set
The number of values in the totalizing memoryis displayed (transaction counter)
Press v for > 2 sec. to display the contentsof the totalizing memory
Press c to clear the totalizing memory. Thetotal is displayed for 1.5 seconds and printed
You cannot toggle between weight units when
there is data in the totalizing memory
Factory Settings
Weight unit 1: kg (1 7 3)
Display accuracy 2: Standard resolution (3 2 1)
-
8/2/2019 Bascula Verficadora
37/90
37
Preparation
Configure the Totalizing application in theoperating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF];while all segments are lit, press )/[TARE]
Select the Totalizing menu item:Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 8 Totalizing
Confirm selection of Totalizing:Press p/[PRINT] for at least 2 sec.
> o indicates that this is the current menucode setting
Set the next parameters: Press p/[PRINT]
Select and confirm: Weight unit 1: (see also Toggle between
Weight Units)1 7 2 Grams... ...1 7 22 lb/oz
see also the Scale Operating Menu(Overview) in the chapter entitled Configuring
the Scale Save settings and exit the Setup menu:
Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the basic functions (power off,zeroing, taring and printing), you can also accessthe following functions from this application:
Start application (v key) Display value in totalizing memory (v key;press for at least 2 sec.)
Clear totalizing memory (c key)
-
8/2/2019 Bascula Verficadora
38/90
38
Example
Totalize weight values
Settings (changes in the factory settings required for this example):Menu: Application program: Totalizing (2 1 8)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. If necessary, zero the scale (/[ZERO] 0.000 kg(S symbol: scale is zeroed verified scales only)
3. Place sample on the scale + 0.380 kg(here: 380 g)
4. Store value in memory; if the v + 0.380 kg1print format is set to include dataID codes, the following is printed G + 0.380 kg
5. Place the next sample on + 0.575 kg1the scale (here: 575 g)
6. Store value in memory v + 0.575 kg2
7. Display the total in memory v for at least 0.955 kg2 (briefly)(total is also printed) 2 sec. Sum + 0.955 kg
8. Clear totalizing memory and c Sum + 0.955 kgprint total
9. Repeat the procedure startingfrom Step 4, if desired.
-
8/2/2019 Bascula Verficadora
39/90
39
Gross/Net Toggling (EB Models Only)
Purpose
With this application program you can togglebetween net and gross values.
Available Features
Toggle the display between gross and net weightswhen there is data in the tare memory
Press w/[TOGGLE] to toggle between weightunit 1 and weight unit 2
Factory Settings
Weight unit 1: kg (1 7 3)
Weight unit 2: kg (3 1 3)
Display accuracy 2: Standard resolution (3 2 1)
Preparation
Configure the Gross/net toggling application in theoperating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; while
all segments are lit, press )/[TARE] briefly Select the Gross/net toggling menu item:
Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 9 Gross/net toggling
Confirm selection of Gross/net toggling:Press p/[PRINT] for at least 2 sec.
> o indicates that this is the current menucode setting
Set the next parameters: Press p/[PRINT]
-
8/2/2019 Bascula Verficadora
40/90
40
Select and confirm:
Weight unit 1: (see also Toggle betweenWeight Units)1 7 2 Grams... ...1 7 22 lb/oz
Weight unit 2: (see also Toggle betweenWeight Units)3 1 2 Grams...
...3 1 22 lb/oz
Display accuracy 2:3 2 1 Standard resolution3 2 2 10x higher resolution**3 2 3 2x higher resolution (PolyRange)*
see also the Scale Operating Menu (Overview)in the chapter entitled Configuring the Scale
Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.
Additional Functions
In addition to the basic functions (power off, zeroing,taring and printing), you can also access thefollowing functions from this application:
Toggle display between gross and net values
(v key) Toggle weight unit (w/[TOGGLE] key)
Example of Printout
Net display(p/[PRINT] key): N + 0.125 kg
T + 0.015 kgG# + 0.140 kg
Gross display(p/[PRINT] key): G + 0.140 kg
** = not applicable for verified scales** = only for EA/EB-0UR models
-
8/2/2019 Bascula Verficadora
41/90
41
Averaging (EB Models Only)
Purpose
Use this program to determine the weights ofunstable samples (e.g., live animals) or to determineweights under unstable ambient conditions. In thisprogram, the scale calculates the weight as theaverage value from a defined number of individualweighing operations. These weighing operationsare also known as subweighing operationsor subweighs.
Available Features
You can set the number of subweighingoperations performed in the operating menu,or before starting the Averaging application(press w/[TOGGLE])
The number of subweighs remainingto be performed is indicated in the displayduring weighing
The measured result displayed is the arithmeticmean shown in the selected weight unit;a triangle under the plus or minus sign indicatesthat this is a calculated value
Press v for at least 2 sec. to display the pre-set
number of subweighing operations Press w/[TOGGLE] to toggle the display
between the measured result and the weight
The measurement result is automatically output viathe data interface port after initialization, if themenu code for Printout with data ID codes is set
Factory SettingsDisplay accuracy 2: Standard resolution (3 2 1)
Number of subweighs for averaging: 10 (3 3 2)
-
8/2/2019 Bascula Verficadora
42/90
42
Preparation
Configure the Averaging application in theoperating menu:
$ If the scale is on, turn it off:Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; whileall segments are lit, press )/[TARE]
Select the Averaging menu item:
Press )/[TARE], p/[PRINT], andp/[PRINT]; press )/[TARE] repeatedly,if necessary2 1 12 Averaging
Confirm selection of Averaging: Press p/[PRINT]for at least 2 sec.
> o indicates that this is the current menucode setting
Set the next parameters: Press p/[PRINT]
Select and confirm: Number of subweighs for averaging:
3 3 1 5 subweighs3 3 2 10 subweighs3 3 3 20 subweighs3 3 4 50 subweighs3 3 5
100 subweighssee also the Scale Operating Menu(Overview) in the chapter entitled Configuringthe Scale
Save settings and exit the Setup menu:Press )/[TARE] for at least 2 sec.
-
8/2/2019 Bascula Verficadora
43/90
43
Additional Functions
In addition to the functions for:
power off (e/[ON/OFF] key), zeroing ((/[ZERO] key), taring ()/[TARE] key), and printing (p/[PRINT] key),
you can also access the following functions fromthis application:
Start measuring (v key) Clear display of results (c key) Change number of measurements after clearing
the display (w/[TOGGLE] key) Display weight (w/[TOGGLE] key) Calibration/adjustment ()/[TARE] key; press
for at least 2 sec.)
-
8/2/2019 Bascula Verficadora
44/90
44
Example
Determine the weight of a sample in extremely unstable ambient conditions.Enter the number of subweighs while in the weighing mode.
Settings (changes in the factory settings required for this example):Menu: Application program: Averaging (2 1 12)
Step Key (or instruction) Display/Output
1. Turn on the scale e/[ON/OFF]
2. If necessary, zero the scale (/[ZERO] 0.000 kg10
(S symbol: scale is zeroed verified scales only)
3. Set the number of subweighs w/[TOGGLE] ref 20 (briefly)for averaging w/[TOGGLE] ref 50 (briefly)(here: 5 subweighs) w/[TOGGLE] ref100 (briefly)
w/[TOGGLE] ref1 (briefly)w/[TOGGLE] ref 2 (briefly)w/[TOGGLE] ref 5 (briefly)
0.000 kg 5
4. Place sample on the scale + 8888 5(weight readout fluctuates;here: around 275 g)
5. Start measurement; if the v 8888 5 . . . 1print format is set to include 0.275 kg 5
data ID codes, the following Res + 0.275 kgis printed
6. Unload the scale 0.275 kg 5
(stable display)
7. Repeat the procedure starting
from Step 4.
-
8/2/2019 Bascula Verficadora
45/90
45
Data Output Functions
There are 3 options for data output:
Output to the display and control unit
Output to a printer (generate a printout)
Output to a peripheral device (e.g., computer)via the interface port
Output to the Display and Control Unit
The display is divided into 5 sections. Informationabout the scale, the application being usedand the sample weighed is output in the followingsections:
Busy symbol, plus/minus sign, stability symbol
Measured value line
Basic units and additional identifiers
Application parameter
Weight unit indicator
Busy Symbol, Plus/Minus Sign, Stability Symbol
This section contains:
Busy symbol
Plus/minus sign
Zero indicator on verified scales
Calculated value indicator (flashing triangle;in the examples shown in this manual, it appearsas a white triangle with a black border)
Basis-Einheiten +Zusatzkennungen
Hinweise f rAnwendungs-programme
Einheiten-pfeile
Mewert
Busy-ZeichenVorzeichenStillstand
Busy symbolPlus/minus signStability symbol
Measured valueApplicationinformation
Basic unit +additional information
Unitindicator
-
8/2/2019 Bascula Verficadora
46/90
46
Measured Value Line
This section shows:
Current weight (digits bordered with a broken lineare invalid in use for legal metrology)
Calculated values (e.g., piece counts)
Application parameters during configuration(e.g., reference sample quantity, etc.)
Software version (r = release; in this example:19 = scale program; 01 = software version)
Weight Units, Additional Identifiers
This section shows:
The current weight unit (e.g., kg)
Other unit of measurement (e.g., piece count)
Indicates that there is data in the tare memory
Application Parameter (EB models only)
This section shows application parameters(e.g., reference sample quantity, no. of subweighs
for averaging, etc.).
Weight Unit Indicator (EB models only)
The arrows in this section indicate the configuredweight units (upper arrow: weight unit 1;lower arrow: weight unit 2)
-
8/2/2019 Bascula Verficadora
47/90
47
Printing a Data Record
Purpose
You can generate a printout of weights as well asother measured values and identifiers fordocumentation purposes. You can format theprintout to meet individual requirements.
Available Features
The printout can be configured to include 2 user-defined header lines of 14 characters each. Thesetext lines are entered either at the Sartorius factoryor on location with a special program that uses theBPI interface command sequence.
Print an individual weight, or net, gross andtare weights.
Line format: You can configure a data ID code ofup to 6 characters for each of the values printed; thisdata ID code is printed at the beginning of the line
Printouts generated automatically or by pressingthe p/[PRINT] key, dependent on or independentof scale stability
You can have the following values output automaticallywhen using the application programs (EB models only)
if menu code 7 1 2 or 7 1 3 is configured (printoutwith data ID codes):
Second tare memory: last net value
Counting: Reference weight for one piece(average piece weight)
Weighing in percent: Reference weight for thepercentage selected
Totalizing: Current weight, total weight (net) Averaging: Result of measurement
-
8/2/2019 Bascula Verficadora
48/90
48
Factory Settings
Header:The default header lines do not contain any text
Print manual/automatic:Individual printout or automatic printout dependenton scale stability:Manual after stability (6 1 2)
Line format:
Up to 6 characters at the beginning of each line toidentify the weight or calculated value:Print net, tare, gross value with data ID (7 1 3)
See the chapter entitled Configuring the Scalefor instructions on setting parameters
Examples
You can configure the EISENMUELLER User-definedprintout to include 2 user- GOETTINGEN User-defineddefined header lines.The software required forconfiguring these lines via
your PC is available fromSartorius in Goettingen
or your local Sartoriusoffice or dealer.
Printout without Data ID Codes:
The value currently + 1530.0 g Weight in gramsdisplayed is printed + 58.562 ozt Weight in Troy ounces(weight or calculated + 253 pcs Piece countvalue with unit) + 88.2 % Percentage
+ 105.8 o Calculated value
Headers:
-
8/2/2019 Bascula Verficadora
49/90
49
Printout with Data ID Codes:
The current value N + 1.530 kg Current net weightdisplayed can be printed N1 + 1.530 kg Current net weightwith a data ID code T + 0.234 kg Value in tare memoryof up to 6 characters T1 + 0.102 kg Value in 2nd tare memoryat the beginning of the G + 1.553 kg Current gross weightline. You can use this G# + 1.630 kg Calculated gross weightdata ID code, e.g., to Qnt + 253 pcs Calculated quantitydesignate a weight Prc + 88.2 % Calculated percentage
readout as a net weight Sum + 1.279 kg Total value (net)(N) or a calculated Res + 1.530 kg Calculated resultvalue as a piece count(QNT).
Print Application Parameters:
You can generate a print- wRef + 0.014 kg Counting:out of one or more of the Average piece weightvalues configured for Wxx% + 1.200 kg Weighing in percent:initialization of an appli- reference weight for thecation as soon as you selected percentageinitialize the scale.
Auto Print:
You can have the weight N + 1.530 kg Net weightreadout printed auto- Stat Display blankmatically. The display Stat L Display underload
update interval depends Stat H Display overloadon the operating statusof the scale and on thescale model.
-
8/2/2019 Bascula Verficadora
50/90
50
Interface Description
Purpose
Your Economy/Express scale comes equippedwith an interface port for connection to a computeror other peripheral device.
You can use an on-line computer to change, startand/or monitor the functions of the scale and theapplication programs (EB models only).
Available Features
Type of interface: Serial interface
Operating mode: Full duplex
Standard: RS-232
Transmission rates:
150; 300; 600; 1,200; 2,400; 4,800;9,600 baud
Parity: Mark, space, odd, even
Character format:1 start bit, 7-bit ASCII, parity, 1 or 2 stop bits
Handshake:2-wire interface: via software (XON/XOFF)4-wire interface: via hardware handshake lines(CTS/DTR)
Operating mode: SBI
Data output format of the scale:16 or 22 characters
-
8/2/2019 Bascula Verficadora
51/90
51
Factory Settings:
Transmission rate: 1,200 baud (5 1 4)
Parity: Odd (5 2 3)
Stop bits: 1 stop bit (5 3 1)
Handshake:Hardware, 1 character after CTS (5 4 3)
Operating mode: Standard SBI (5 5 1)
Print manually/automatically:Manual after stability (6 1 2)
Preparation
See the chapter entitled Overview for the pinassignment chart
Line Format (Data Output Format)
You can output the values displayed in the measuredvalue line and the weight unit with or without a dataID code.
Example: Without data ID code+ 253 pcs
Example: With data ID code
Qnt + 253 pcs
Configure this parameter in the Setup menu(Menu: Printout format 7 1 1 or 7 1 2 or 7 1 3).
The output without data ID code has 16 characters;with data ID code, 22 characters.
!Note on Operation of the Scale with theYDP03-0CE Printer
When data output with 22 characters per lineis configured (menu code 7 .1. 2), the GLP/GMPprintout setting in the printer must be set to off(-A1- 0 = off)
-
8/2/2019 Bascula Verficadora
52/90
52
Output Format With 16 Characters
Display segments that are not activated are output as spaces. Characterswithout a decimal point are output without a decimal point.
The following characters can be output, depending on the characters displayedon the scale:
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16+ D D D D D D * U U U CR LF
or . . . . . . . . * * *
or * * * * * * * * *
or 0 0 0 0 0 0
*: Space CR: Carriage returnD: Digit or letter LF: Line feedU: Unit symbol
Special Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16* * * * * * * * CR LF
or H *
or L *
or C *
*: Space H: Overload : Weight L: Underload
C: Calibration/adjustment
-
8/2/2019 Bascula Verficadora
53/90
53
Error Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16* * E * # # # * * * * CR LF
*: Space# # #: Error code numberExample: Data output example: + 1255.7 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * * * 1 2 5 5 . 7 * g * * CR LF
Position 1: Plus or minus sign or spacePosition 2: SpacePosition 310: Weight with a decimal point; leading zeros = spacePosition 11: SpacePosition 1214: Unit symbol or spacePosition 15: Carriage returnPosition 16: Line feed
-
8/2/2019 Bascula Verficadora
54/90
54
Output Format with 22 Characters
When data with an ID code is output, the ID code consisting of 6 charactersprecedes the data with the 16-character format. These 6 characters identify thefollowing value.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22I I I I I I + * D D D D D D D D * U U U CR LF
* * * * * . . . . . . . . * * *
* * * * * * * * *
0 0 0 0 0 0
I: ID code character1) U: Unit symbol1)
*: Space CR: Carriage returnD: Digit or letter LF: Line feed
1) depends on scale type; e.g., not all units and characters are availableon scales verified for use in legal metrology (see also next page)
Special Codes
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22S t a t * * * * * * * * * * * * * * CR LF
H *
L *
*: Space H: Overload : Weight L: Underload
Error Codes
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22S t a t * * * * * E r r * # # # * * * * CR LF
*: Space # # #: Error code number
-
8/2/2019 Bascula Verficadora
55/90
55
ID code
characters I MeaningStat Status
G GrossG# Gross calculated valueT Tare T
T1 Tare T1N Net N
N1 Net N1
Qnt Counting: quantityPrc Weighing in percent: percentageRes Calculation, averaging: resultSum Totalizing: net total
wRef Automatic printout: average piece weightWxx% Automatic printout: reference percentage weight
-
8/2/2019 Bascula Verficadora
56/90
56
Data Input Format
You can connect a computer to your scale to sendcommands via the scale interface port to controlscale functions and applications (EB models only).
The commands sent are control commands andmay have different formats; e.g., control commandscan have up to 26 characters. Each charactermust be transmitted according to the settingsconfigured in the Setup menu for data transmission.
Format for Control Commands
Format : Esc ! CR LF
Esc: Escape CR: Carriage return (optional)!: Command LF: Line feed (optional)
character
Command character ! MeaningK Weighing mode 1L Weighing mode 2
M Weighing mode 3N Weighing mode 4
O Block keysP PrintR Release keysS RestartT Tare and zero (combined)U Tare (Tare only)V Zero
W External calibration/adjustment
-
8/2/2019 Bascula Verficadora
57/90
57
Synchronization
During data communication between the scale andan on-line device (computer), messages consisting ofASCII characters are transmitted viathe interface. For error-free data communication,the parameters for baud rate, parity, handshakemode and character format must be the samefor both units.
You can set these parameters in the Setup menu so
that they match those of the on-line device. You canalso define parameters in the scale to makedata output dependent on various conditions. Theconditions that can be configured are describedunder each of the application program descriptions.
If you do not plug a peripheral device into the scaleinterface port, this will not generate an errormessage.
Handshake
The scale interface (Sartorius Scale Interface = SBI)has transmit and receive buffers. You can define thehandshake parameter in the Setup menu:
Hardware handshake (CTS/DTR) Software handshake (XON, XOFF)
Hardware Handshake
With a 4-wire interface, 1 more character can betransmitted after CTS (Clear to Send).
Software Handshake
The software handshake is controlled via XONand XOFF. When a device is switched on,XON must be transmitted to enable any connecteddevice to communicate.
-
8/2/2019 Bascula Verficadora
58/90
58
Activating Data Output
You can define the data output parameter so thatoutput is activated either when a print command isreceived or automatically and synchronous with thescale display or at defined intervals (see applicationprogram descriptions (EB models only) and auto-printsetting).
Data Output by Print CommandThe print command can be transmitted by pressingp/[PRINT] or by a software command (Esc P).
Automatic Data Output
In the auto print operating mode, data are outputto the interface port without a print command.
You can choose to have data output automatically atdefined print intervals with or without the stabilityparameter. Whichever parameter you select, thedata will be output as the readouts appear onthe scale display. The display update frequencydepends on the setting for Adapting the Filter(1.1.x.).
If you select the auto print setting, data will be
transmitted immediately the moment you turn onthe scale.
-
8/2/2019 Bascula Verficadora
59/90
59
Pin Assignment Chart
Female Interface Connector:25-position D-Submini, DB25S, with screw lock hardware for cable gland
Male Connector Used: (please use connectors with the same specifications):25-pin D-Submini, DB25S, with integrated shielded cable clamp assembly(Amp type 826 985-1C) and fastening screws (Amp type 164 868-1)
! Warning When Using Pre-wired RS-232 Connecting Cables!RS-232 cables purchased from other manufacturers often have incorrect pinassignments for use with Sartorius scales. Be sure to check the pin assignment
against the chart below before connecting the cable, and disconnect any linesmarked Internally Connected (e.g., pin 6). Failure to do so may damage or evencompletely ruin your scale and/or peripheral device.
Pin Assignment Chart:
Pin 1: Signal groundPin 2: Data Output (TxD)Pin 3: Data Input (RxD)Pin 4: Internal Ground (GND)
Pin 5: Clear to Send (CTS)Pin 6: Internally ConnectedPin 7: Internal Ground (GND)Pin 8: Internal Ground (GND)Pin 9: Internally ConnectedPin 10: Not ConnectedPin 11: Not ConnectedPin 12: Reset _ Out *) For remote switchPin 13: Internally ConnectedPin 14: Internal Ground (GND)
Pin 15: Universal Remote SwitchPin 16: Not ConnectedPin 17: Not ConnectedPin 18: Not ConnectedPin 19: Not ConnectedPin 20: Data Terminal Ready (DTR)Pin 21: Ground input for external power supply (GND _ V in)Pin 22: Internally ConnectedPin 23: Internally ConnectedPin 24: Power supply input +12 ... 30 V
Pin 25: Not Connected*) = Hardware restart
-
8/2/2019 Bascula Verficadora
60/90
60
Cabling Diagram
Diagram for interfacing a computer or different peripheral deviceto the scale using the RS-232/V24 standard and cables up to 15 m(50 ft.) long
V24
Scale Peripheral Device
Type of cable: AWG 24 specification
(PC)25 pin 9 pin
2 3
20 4
3 2
5 8
6 6
7 5
-
8/2/2019 Bascula Verficadora
61/90
61
Configuring the Scale
Setting the Parameters (Menu)
Purpose
You can configure your Economy/Express scaleto meet individual requirements by selecting from theparameters available in the menu.
Parameters that are not permitted in legal metrologyare not shown in the menu on verified scales.
Features
The parameters are divided into the followinggroups (1st menu level):
1 Scale functions
2 Application programs
3 Application parameters
5 Data interface6 Print for weighing
7 Print for application
8 Extra functions
9 Reset menu to factory-set parameters
-
8/2/2019 Bascula Verficadora
62/90
62
Factory-Set Parameters
The factory-set parameters are marked by an oin the list starting on page 64.
Preparation
Turn off the scale: Press e/[ON/OFF]
Turn on the scale: Press e/[ON/OFF]; whileall segments are displayed, press)/[TARE] briefly
> Measured value line: 1 (1st menu level)$ To navigate within a menu level: Press
)/[TARE]; when you reach the last menu item,the first is shown again
$ To select the next sublevel within a group:Press p/[PRINT] briefly.
$ To return to the higher menu level:Press p/[PRINT]
To confirm selected parameters: Press and holdp/[PRINT] for more than two seconds
> o indicates the selected menu code
To store parameter settings and exit the menu:Press and hold the )/[TARE] key
To exit the menu without storing any new settings:Press e/[ON/OFF]
> Restart the application
-
8/2/2019 Bascula Verficadora
63/90
63
Example
Adapt the scale to unstable ambient conditions: Menu code 1.1. 4
Step Key Display
1. Turn the scale off e/[ON/OFF]
2. Turn the scale back on; e/[ON/OFF]
while all segments are )/[TARE]displayed briefly
3. Confirm 1st menu level(scale functions) p/[PRINT]
4. Confirm Adapt filter group(2nd menu level) p/[PRINT]
5. Menu level 3: )/[TARE]Select the desired menu item repeatedly
6. Confirm selected item p/[PRINT]for 2 seconds
7. Set other parameters, p/[PRINT],if desired )/[TARE]
8. Saves changes and )/[TARE]
exit menu for 2 seconds
-
8/2/2019 Bascula Verficadora
64/90
64
Scale Operating Menu (Overview)
Factory setting User setting
Menu items that are only available in the EB and not in the EA model are marked by shading.
Menu 1 Scale 1.1. Adapt filter 1.1.1 Minimum vibrationfunctions 1.1.2 o Normal vibration
1.1.3 Strong vibration1.1.4 Extreme vibration
1.2. Application filter 1.2.1 o Final readout1.2.2 Filling mode
1.3. Stability range 1.3.1 1/4 digit1.3.2 1/2 digit1.3.3 1 digit*1.3.4 o 2 digits*1.3.5 4 digits*
1.4. Weight unit for 1.4.1 Gramscalibration weight* 1.4.2 o Kilograms1.4.3 Pounds
1.5. Adjustment* 1.5.1 o Accessible1.5.2 Blocked
1.6. Auto zero 1.6.1 o On1.6.2 Off
1.7. through 1.8. see next page
2. through 9. see next pages
Menu
level1
Menu
level2
Menu
level3
Factor
ysetti
ng
Menu
item
* = These settings cannot be changed on verified scales
-
8/2/2019 Bascula Verficadora
65/90
65
Menu 1. Scale 1.1. through 1.6. see previous pagefunctions
1.7. Weight unit 1* 1.7.2 Grams1.7.3 o Kilograms1.7.4 Carats1.7.5 Pounds1.7.6 Ounces1.7.7 Troy ounces1.7.8 Hong Kong taels1.7.9 Singapore taels1.7.10 Taiwanese taels1.7.11 Grains1.7.12 Pennyweights1.7.13 Milligrams1.7.14 Parts per pound1.7.15 Chinese taels1.7.16 Mommes1.7.17 Austrian carats1.7.18 Tola1.7.19 Baht1.7.20 Mesghal1.7.21 Tons1.7.22 Pounds+ounces
(non-decimal display)
1.8. Display accuracy 1* 1.8.1 o Standard resolution1.8.2 10x higher resolution**1.8.3 2x higher resolution
(PolyRange)
2. Application 2.1. Program selection 2.1.1 Weighing
programs 2.1.2 Toggle weight units2.1.3 Net-total (tare memory)2.1.4 o Counting2.1.5 Weighing in percent2.1.7 Calculation by a factor2.1.8 Totalizing2.1.9 Gross/net toggling2.1.10 Tare memory with autom.
container taring2.1.11 Weighing in percent with
difference display
2.1.12 Averaging3. through 9. see following pages
** = These settings cannot be changed on verified scales** = Only for EB-0UR models
Menulev
el1
Menulev
el2
Menulev
el3
Factorys
etting
Menuite
m
-
8/2/2019 Bascula Verficadora
66/90
66
Menu 1. through 2. see previous pages
3. Application 3.1. Weight unit 2* 3.1.2 Gramsparameters 3.1.3 o Kilograms
3.1.4 Carats3.1.5 # Pounds3.1.6 Ounces3.1.7 Troy ounces3.1.8 Hong Kong taels3.1.9 Singapore taels3.1.10 Taiwanese taels3.1.11 Grains3.1.12 Pennyweights3.1.13 Milligrams3.1.14 Parts per pound3.1.15 Chinese taels3.1.16 Mommes3.1.17 Austrian carats3.1.18 Tola3.1.19 Baht3.1.20 Mesghal3.1.21 Tons3.1.22 Pounds+ounces
(non-decimal display)
3.2. Display accuracy 2* 3.2.1 o Standard resolution3.2.2 10x higher resolution**3.2.3 2x higher resolution (PolyRange)
3.3. Preset reference 3.3.1 5quantity, reference 3.3.2 o 10
percentage or no. 3.3.3 20of subweighs 3.3.4 50for averaging 3.3.5 100
5. Data 5.1. Baud rate 5.1.1 150interface 5.1.2 300
5.1.3 6005.1.4 o 1,2005.1.5 2,4005.1.6 4,8005.1.7 9,600
5.2. through 5.5. see next page6. through 9. see next page
Menulev
el1
Menulev
el2
Menulev
el3
Factorys
etting
Menuite
m
# = Factory setting on scale model EA/EB ...UR** = These settings cannot be changed on verified scales** = Only for EB-0UR models
-
8/2/2019 Bascula Verficadora
67/90
67
Menu 5. Interface 5.1. through 5.3. see previous page
5.2. Parity 5.2.1 Mark5.2.2 Space5.2.3 o Odd5.2.4 Even
5.3. No. of stop bits 5.3.1 o 15.3.2 2
5.4. Handshake mode 5.4.1 Software5.4.2 Hardware,
1 char. after CTS5.4.3 o Hardware,
2 char. after CTS
5.5. Communication 5.5.1 o PC, YDP03 printer (SBI)mode 5.5.2 YDP01IS printer*
6. Print for 6.1. Manual/auto print 6.1.1 Manual without stabilityweighing mode 6.1.2 o Manual after stability
6.1.3 Automatic without stability6.1.4 Automatic at stability
7. Print for 7.1. Printout format 7.1.1 No data ID codesapplication 7.1.2 With data ID codes
7.1.3 o With data ID codes,N, T and G-values and2 header lines
8. Extra 8.1. Menu** 8.1.1 o Parameter settingsfunctions alterable
8.1.2 Parameter settings
read only8.2. External switch 8.2.1 Zero setting and taring
function combined8.2.3 w/[TOGGLE] key func.8.2.4 o p/[PRINT] key function8.2.5 (/[ZERO] key function8.2.6 c key function8.2.7 v key function8.2.8 )/[TARE] key function
8.3. Power-on mode 8.3.1 Off/on
8.3.2 o Stand-by/on
9. Reset menu** 9. Factory settings 9. 1 Reset9. 2 o Do not reset
Menulev
el1
Menulev
el2
Menulev
el3
Factorys
etting
Menuite
m
** = To operate the YDP01IS, also set menu item 5.1.7 9,600 baud** = Not for EB-0UR
-
8/2/2019 Bascula Verficadora
68/90
68
Error Codes
Error codes are shown in the main display for approx. 2 seconds, after which
the program automatically returns to the weighing mode.
Display Cause Solution
No segments appear on No AC power is Check the ACthe display available power supply
The AC adapter is not Plug in the AC adapterplugged in
Automatic shutoff Turn on the scaleis configured
Rechargeable battery Recharge batteryhas run down (see battery instructions)
h The load exceeds the Unload the scalescale capacity
l The load plate is Place the load platenot in place on the scale
Something is touching Move the object that isthe load plate touching the load place
e 01 Data output not Change the configurationcompatible with in the Setup menuinput format
e 02 Calibration parameter Calibrate only whennot met; e.g.: zero is displayed scale not zeroed Press (/[ZERO]
to zero the scale scale is loaded Unload the scale
e 08 The scale was zeroed Only zero the scaleoutside the zero range when in the zero range;
i.e.: ( 2% of themaximum capacity
e 09 Taring is not possible Press (/[ZERO]when the gross weight to zero the scale
is zero
-
8/2/2019 Bascula Verficadora
69/90
69
Display Cause Solution
e 10 The )/[TARE] key Press c to clear theis blocked when there tare memory and releaseis data in the second the tare keytare memory (net-total) only 1 tare function canbe used at a time
e 11 Tare memory not Press (/[ZERO]accessible
e 22 Weight is too light or Increase the referencethere is no sample on quantitythe scale
e 30 Interface port for printer Perform Reset menuoutput is blocked or
Contact your localSartorius Service Center
The maximum The scale was turned on Press the e/
weighing capacity without the load plate [ON/OFF] key to turnis less than is indicated in place the scale off and backin Specifications on again
The weight readout Unstable ambient Set up the scalechanges constantly conditions in another area
Too much vibration, Change setupor the scale is exposed configuration to adaptto a draft the scale to the
ambient conditionsA foreign object is caught Remove the foreignbetween the load plate objectand the scale frame
The weight readout The scale has not Calibrate/adjustis obviously wrong been calibrated/adjusted the scale
The scale was not Tare before weighingtared before weighing
If any other errors occur, contact your local Sartorius Service Center!
-
8/2/2019 Bascula Verficadora
70/90
70
Care and Maintenance
Service
Regular servicing by a Sartorius technician willextend the service life of your scale and ensureits continued weighing accuracy. Sartorius canoffer you service contracts, with your choice ofregular maintenance intervals ranging from1 month to 2 years.The frequency of these maintenance intervalsdepends on the operating conditions and yourrequirements concerning the tolerances for the scale.
Repairs
Repair work must be performed by trained servicetechnicians. Any attempt by untrained personsto perform repairs may lead to hazards for the user.
Cleaning
!Disconnect the scale from the AC adapterand unplug any data cables that are connectedto the scale
!Make sure that no liquids enter the scale housing
!Do not use any aggressive cleaning agents(solvents or similar agents)
Unplug the AC adapter from the wall outlet(mains supply)
If you have a data cable connected to theinterface, unplug it from the scale
Clean the scale using a piece of cloth which hasbeen wet with a mild detergent (soap)
-
8/2/2019 Bascula Verficadora
71/90
71
After cleaning, wipe down the scale with
a soft, dry clothSafety Inspection
If there is any indication that safe operationof the scale with the AC adapter is no longerwarranted:
Turn off the power and disconnect the equipmentfrom AC power immediately
> Lock the equipment in a secure place to ensurethat it cannot be used for the time being
Safe operation of the scale with the AC adapter isno longer ensured when:
there is visible damage to the AC adapter
the AC adapter no longer functions properly
the AC adapter has been stored for a relativelylong period under unfavorable conditions
In this case, notify your nearest Sartorius ServiceCenter or the International Technical SupportUnit based in Goettingen, Germany. Maintenanceand repair work may only be performed byservice technicians who are authorized by Sartoriusand who
have access to the required maintenance manuals
have attended the relevant service training courses
Instructions for Recycling the Packaging
To ensure safe shipment, your scale has beenpackaged using environmentally friendly materials.After successful installation of the scale,
you should return this packaging for recycling.For information on recycling options, includingrecycling of old weighing equipment, contact yourmunicipal waste disposal center or local recyclingdepot.
-
8/2/2019 Bascula Verficadora
72/90
72
Overview
General View of the Scale
No. Designation Order no.forreplacement
1 Load plate EA/EB ... DCE: 69EA0011 EA/EB ... EDE: 69EA0004 EA/EB ... FEG: 69EA0017
2 Display and control unit3 Toggle key
(EB models only)4 Print key5 Function key
(EB models only)6 Tare key
No. Designation Order no.forreplacement
7 Display8 Zero key9 Clear function key
(EB models only)10 On/off key11 DC jack12 Leveling feet Set: 69EA002013 Data interface port
Not shown:
Protective cap forinterface port 69LC0084
4*
5*6*
7*7
89
10
3
* = As shown in this illustration, there are some differences between the standardand the U.S./Canadian keypad versions.
10
9*8*
4
56
1*
2*
3*
1112
12
1213
Standard keypad version: U.S./Canadian keypad version:
*
-
8/2/2019 Bascula Verficadora
73/90
73
Description of the Keys
w/[TOGGLE] key:Toggle
For toggling the displaybetween weighed andcalculated values(counting result,
percentage, calculatedresult)
For changing the refer-ence quantity/percen-tage or numbers ofsubweighs for averaging
v key:
Start ApplicationProgram
The procedure thatfollows after this key ispressed depends on theapplication programselected; for a completedescription, see the
chapter entitledOperating the Scalein the correspondingsection.
)/[TARE] key: Tare
For taring the weightof containers so thatthe scale readoutshows the net weightof samples.
p/[PRINT] key:Data Output
Press this key to outputthe displayed values tovia the built-in interfaceto a Sartorius Data
Printer or a computer.
e/[ON/OFF] key:On/off switch
Switches the displayon/off. The scaleremains in the standbymode.
(/[ZERO] key: Zero
Sets the readout to zero.The scale can onlybe zeroed if the load is 2% of the maximumcapacity.
c key: Clear Function
This key is generallyused to interrupt/cancelfunctions:
Clear memory anddelete applicationdata
Interrupt calibration/adjustment routines
-
8/2/2019 Bascula Verficadora
74/90
74
Specifications
Standard modelsModel EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB
3DCE-L 6DCE-L 15DCE-L35EDE-L 60EDE-L 60FEG-L 150FEG-L
Weighing capacity kg 3 6 15 35 60 60 150lb 6.6 13.2 33 77 132 132 330
Readability g 1 2 5 10 20 20 50
Maximum load kg 6 12 30 70 120 120 300
Tare range (by subtraction) kg 3 6 15 35 60 60 150
Reproducibility g 1 2 5 10 20 20 50
Linearity g 1 2 5 10 20 20 50
Sensitivity drift within ppm/10+40C K 100 100 100 100 100 100 125
External calibration weight kg 2 (M1) 2 (M1) 5 (M1) 10 (M1) 20 (M1) 20 (M1) 50 (M1)(of at least accuracy class...) lb 5 (M1) 5 (M1) 10 (M1) 20 (M1) 50 (M1) 50 (M1) 100 (M1)
Net weight, approx. kg 5 5 5 8 8 24 24
Dust and water protection ratingfor the weighing platformaccording to EN 60529 IP54
Dust protection rating for thedisplay and control unitaccording to EN 60529 IP40
AC power source/power requirements AC adapter, 230 or 115 V, +15% ... 20%
Frequency 48 60 Hz
Power consumption maximum 16 VA; average 8 VA
Operating temperature range 10 ... +40C (263 ... 313 K, 14F ... 104F)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update (dependson the filter level selected) 0.1 0.4
Hours of operation with fullycharged YRB 05 Z externalbattery pack, approx. 48 h
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces,Hong Kong taels, Singapore taels, Taiwanese taels, grains,pennyweights, milligrams, parts per pound, Chinese taels,mommes, Austrian carats, tola, baht, mesghal and tons
Built-in interface RS-232
Format: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or spaceTransmission rates: 150 to 9,600 baudHandshake mode: Software or hardware
-
8/2/2019 Bascula Verficadora
75/90
75
Model EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB
6DCE-L 12DCE-L 30EDE-L 60EDE-L 60FEG-L 120FEG-L0UR 0UR 0UR 0UR 0UR 0UR
Weighing capacity kg 6 12 30 60 60 120lb 13.2 26.4 66 132 132 264.5
Readability g 1 2 5 10 10 20
Maximum load kg 12 30 70 120 120 300
Tare range (by subtraction) kg 6 12 30 60 60 120
Reproducibility g 1 2 5 10 10 20
Linearity g 1 2 5 10 10 20
Sensitivity drift within ppm/+5+35C K 50 50 50 50 50 50
External calibration weight kg 2 (M1) 5 (M1) 10 (M1) 20 (M1) 20 (M1) 50 (M1)(of at least accuracy class...) lb 5 (M1) 10 (M1) 20 (M1) 50 (M1) 50 (M1) 100 (M1)
Net weight, approx. kg 5 5 8 8 24 24
Dust and water protection ratingfor the weighing platformaccording to EN 60529 IP54
Dust protection rating for thedisplay and control unit
according to EN 60529 IP40AC power source/power requirements AC adapter, 230 or 115 V, +15% ... 20%
Frequency 48 60 Hz
Power consumption maximum 16 VA; average 8 VA
Operating temperature range 10 ... +40C (263 ... 313 K, 14F ... 104F)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update(depends on the filter level selected) 0.1 0.4
Hours of operation with fullycharged YRB 05 Z externalbattery pack, approx. 48 h
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces,Hong Kong taels, Singapore taels, Taiwanese taels, grains,pennyweights, milligrams, parts per pound, Chinese taels,mommes, Austrian carats, tola, baht, mesghal and tons
Built-in interface RS-232Format: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or space
Transmission rates: 150 to 9,600 baudHandshake mode: Software or hardware
-
8/2/2019 Bascula Verficadora
76/90
76
Model EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB
3DCE-I 6DCE-I 15DCE-I 35EDE-I 60EDE-I 60FEG-I 150FEG-I(0UR) (0UR) (0UR) (0UR) (0UR) (0UR)
Weighing capacity kg 3 6 15 35 60 60 150lb 6.6 13.2 33 77 132 132 330
Readability g 0.1 0.2 0.5 1 2 2 5
Maximum load kg 6 12 30 70 120 120 300
Tare range (by subtraction) kg 3 6 15 35 60 60 150
Reproducibility g 0.2 0.2 0.5 1 2 2 5
Linearity g 0.2 0.4 1 2 4 4 10
Sensitivity drift within ppm/+5+35C K 10 10 10 10 10 10 10
External calibration weight kg 2 (F2) 2 (F2) 5 (F2) 10 (F2) 20 (F1) 20 (F2) 50 (F2)(of at least accuracy class...) lb 5 (F2) 5 (F2) 10 (F2) 20 (F2) 50 (F1) 50 (F2) 100 (F2)
Net weight, approx. kg 5 5 5 8 8 24 24
Dust and water protection ratingfor the weighing platformaccording to EN 60529 IP54
Dust protection rating for thedisplay and control unit
according to EN 60529 IP40AC power source/powerrequirements AC adapter, 230 or 115 V, +15% ... 20%
Frequency 48 60 Hz
Power consumption maximum 16 VA; average 8 VA
Operating temperature range 10 ... +40C (263 ... 313 K, 14F ... 104F)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update(depends on the filter level selected) 0.1 0.4
Hours of operation with fullycharged YRB 05 Z externalbattery pack, approx. 48 h
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces,Hong Kong taels, Singapore taels, Taiwanese taels, grains,pennyweights, milligrams, parts per pound, Chinese taels,mommes, Austrian carats, tola, baht, mesghal and tons
Built-in interface RS-232Format: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or space
Transmission rates: 150 to 9,600 baudHandshake mode: Software or hardware
-
8/2/2019 Bascula Verficadora
77/90
77
Models Verified by the Manufacturer with EC Type Approval
Model EA/EB EA/EB EA/EB EA/EB EA/EB EA/EB3DCE-L 6DCE-L 15DCE-L 30EDE-L 35EDE-L 60EDE-L0CE 0CE 0CE 0CE 0CE 0CE
Type designation DS BH 300 DN BH 300 DN BH 300
Accuracy class* l l l l l l
Weighing capacity, max.* kg 3 6 15 30 35 60
Scale interval d* (Readability) g 1 2 5 10 10 20
Verification scale interval e* g 1 2 5 10 10 20
No. of scale intervals* e 3,000 3,000 3,000 3,000 3,500 3,000
Minimum capacity, min.* g 20 40 100 200 200 400
Application range acc. to CD* kg 0.023 0.046 0.115 0.230 0.235 0.460
Maximum load kg 6 12 30 70 70 120
Tare range (by subtraction)) 100% of maximum weighing capacity
Net weight, approx. kg 5 5 5 8 8 8
Dust and water protection ratingfor the weighing platform IP54
Dust protection rating for thedisplay and control unit IP40
AC power source/power AC adapter, 230 or 115 V, +15% ... 20% requirements
Frequency 48 60 Hz
Power consumption maximum 16 VA; average 8 VA
Operating temperature range 10 ... +40 C
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update (dependson the filter level selected) 0.1 0.4
Hours of operation with fullycharged YRB05Zexternal battery pack, approx. 48 h
Weight unit Kilograms
Built-in interface RS232CFormat: 7-bit ASCII, 1 start bit, 1 or 2 stop bitsParity: mark, odd, even or spaceTransmission rates: 150 to 9,600 baudHandshake mode: Software or hardware
* CD = Council Directive 90/384/EEC on non-automatic weighing instruments used in the EU
and the Signatories of the Agreement on the European Economic Area
-
8/2/2019 Bascula Verficadora
78/90
78
Model EA/EB EA/EB
60FEG-L 150FEG-L0CE 0CE
Type designation DQ BH 300 DQ BH 300
Accuracy class* l l
Weighing capacity, max.* kg 60 150
Scale interval d* (Readability) g 20 50
Verification scale interval e* g 20 50
No. of scale intervals* e 3,000 3,000
Minimum capacity, min.* g 400 1,000
Application range acc.