1 DOI: 10.5797/jnet.oa.2017-0010 Development of Simulator System Using Endovascular Evaluator for Catheter Intervention Training Jun-ichi Koyama, 1 Yoshiki Hanaoka, 2 Takafumi Kiuchi, 2 Tetsuo Sasaki, 3 Takahiro Murata, 4 Atsushi Sato, 3 and Kazuhiro Hongo 2 Objective: We formulated a simulator system for catheter intervention training by developing a sliding table and an electrically controlled camera-holding C-arm and combining them with the silicone vessel model EndoVascular Evaluator (EVE). We performed simulation of mechanical thrombectomy using this system and evaluated its usability for simulation training. Methods: After three experts in neuroendovascular treatment were given instructions for the use of this system, they performed mechanical thrombectomy by a procedure as close as possible to that in clinical situations using an artificial thrombus model simulating left middle cerebral artery occlusion, and the procedural times and maneuvers were studied. The time required for guiding catheter placement in the cervical internal carotid artery (guiding time), time required for stent placement (stent placement time), and time required for retrieval of thrombus together with the stent (stent retrieval time) were measured, and the number of movements of the sliding table and C-arm during the procedure were counted. Results: The intended procedure could be executed faithfully by all physicians. The mean guiding, stent placement, and stent retrieval times were 185 ± 18, 387 ± 33, and 616 ± 27 seconds, respectively. The mean numbers of table and C-arm movements during the procedure were 14 ± 1.7 and 8.3 ± 0.5, respectively. Conclusion: This system allows operators to faithfully reproduce the mechanical thrombectomy procedure and is considered to have functions necessary for simulation training of catheter intervention and performance assessment. Keywordscatheter intervention, simulation training, simulator system, fidelity, mechanical thrombectomy Introduction Recently, various medical training simulations are prac- ticed. The objective of simulation training is to fill the clinical gap between knowledge and practice. As the train- ing environment is closer to the clinical reality, training can be more effective. The simulator is important for securing the fidelity required for such training, and a practical simu- lator improves the quality of simulation training. 1–4) In this study, we prepared a simulator system for training in neu- roendovascular treatment and performed simulation of mechanical thrombectomy. Subjects and Methods Preparation of simulator system EndoVascular Evaluator (FAIN-Biomedical, Nagoya, Japan) We prepared a catheter simulator system using the sili- cone artificial vessel model EndoVascular Evaluator (EVE: FAIN-Biomedical, Nagoya, Japan). 5) EVE consists of the brain, trunk, heart, and peripheral extremity parts, through which channels of transparent silicone tubes mimicking the cardiovascular system including the heart and major arteries are arranged. A motor connected to the channels 1 Neurointervention Center, Shinshu University Hospital, Matsumoto, Nagano, Japan 2 Department of Neurosurgery, Shinshu University School of Medicine, Matsumoto, Nagano, Japan 3 Division of Neurosurgery, Ina Central Hospital, Ina, Nagano, Japan 4 Division of Neurosurgery, Shinonoi General Hospital, Nagano, Nagano, Japan Received: January 25, 2017; Accepted: July 21, 2017 Corresponding author: Jun-ichi Koyama. Neurointervention Center, Shinshu University Hospital, 3-1-1 Asahi, Matsumoto, Nagano 390-8621, Japan Email: [email protected] This work is licensed under a Creative Commons Attribution-NonCommercial- NoDerivatives International License. ©2018 The Japanese Society for Neuroendovascular Therapy Journal of Neuroendovascular Therapy Vol. 12, No. 1 (2018) Journal of Neuroendovascular Therapy 2018; 12: 1–5 Online August 24 2017

Upload: others

Post on 29-May-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of Simulator System Using Endovascular Evaluator for Catheter Intervention ... · 2019-08-11 · 1. 15n211. Development of Simulator System Using . Endovascular Evaluator

1

DOI: 10.5797/jnet.oa.2017-0010

Development of Simulator System Using Endovascular Evaluator for Catheter Intervention Training

Jun-ichi Koyama,1 Yoshiki Hanaoka,2 Takafumi Kiuchi,2 Tetsuo Sasaki,3 Takahiro Murata,4 Atsushi Sato,3 and Kazuhiro Hongo2

Objective: We formulated a simulator system for catheter intervention training by developing a sliding table and an electrically controlled camera-holding C-arm and combining them with the silicone vessel model EndoVascular Evaluator (EVE). We performed simulation of mechanical thrombectomy using this system and evaluated its usability for simulation training.Methods: After three experts in neuroendovascular treatment were given instructions for the use of this system, they performed mechanical thrombectomy by a procedure as close as possible to that in clinical situations using an artificial thrombus model simulating left middle cerebral artery occlusion, and the procedural times and maneuvers were studied. The time required for guiding catheter placement in the cervical internal carotid artery (guiding time), time required for stent placement (stent placement time), and time required for retrieval of thrombus together with the stent (stent retrieval time) were measured, and the number of movements of the sliding table and C-arm during the procedure were counted.Results: The intended procedure could be executed faithfully by all physicians. The mean guiding, stent placement, and stent retrieval times were 185 ± 18, 387 ± 33, and 616 ± 27 seconds, respectively. The mean numbers of table and C-arm movements during the procedure were 14 ± 1.7 and 8.3 ± 0.5, respectively.Conclusion: This system allows operators to faithfully reproduce the mechanical thrombectomy procedure and is considered to have functions necessary for simulation training of catheter intervention and performance assessment.

Keywords▶ catheter intervention, simulation training, simulator system, fidelity, mechanical thrombectomy

Introduction

Recently, various medical training simulations are prac-ticed. The objective of simulation training is to fill the

clinical gap between knowledge and practice. As the train-ing environment is closer to the clinical reality, training can be more effective. The simulator is important for securing the fidelity required for such training, and a practical simu-lator improves the quality of simulation training.1–4) In this study, we prepared a simulator system for training in neu-roendovascular treatment and performed simulation of mechanical thrombectomy.

Subjects and Methods

Preparation of simulator system EndoVascular Evaluator (FAIN-Biomedical, Nagoya, Japan)We prepared a catheter simulator system using the sili-cone artificial vessel model EndoVascular Evaluator (EVE: FAIN-Biomedical, Nagoya, Japan).5) EVE consists of the brain, trunk, heart, and peripheral extremity parts, through which channels of transparent silicone tubes mimicking the cardiovascular system including the heart and major arteries are arranged. A motor connected to the channels

1Neurointervention Center, Shinshu University Hospital, Matsumoto, Nagano, Japan2Department of Neurosurgery, Shinshu University School of Medicine, Matsumoto, Nagano, Japan3Division of Neurosurgery, Ina Central Hospital, Ina, Nagano, Japan4Division of Neurosurgery, Shinonoi General Hospital, Nagano, Nagano, Japan

Received: January 25, 2017; Accepted: July 21, 2017Corresponding author: Jun-ichi Koyama. Neurointervention Center, Shinshu University Hospital, 3-1-1 Asahi, Matsumoto, Nagano 390-8621, JapanEmail: [email protected]

This work is licensed under a Creative Commons Attribution-NonCommercial- NoDerivatives International License.

©2018 The Japanese Society for Neuroendovascular Therapy

Journal of Neuroendovascular Therapy Vol. 12, No. 1 (2018)

Journal of Neuroendovascular Therapy 2018; 12: 1–5

Online August 24 2017

Page 2: Development of Simulator System Using Endovascular Evaluator for Catheter Intervention ... · 2019-08-11 · 1. 15n211. Development of Simulator System Using . Endovascular Evaluator

2

Koyama J, et al.

drives liquid into the heart in a pulsed manner and dynam-ically reproduces the systemic circulation of blood from the heart. Each part can be freely designed and replaced (Fig. 1A).

Development of freely sliding tableWe developed a freely sliding table that can be moved horizontally on the basis of the beds of actual angiography systems. By mounting EVE on this table, it could be moved horizontally over a distance of 100 cm in the craniocaudal direction and 60 cm in the lateral direction. A screen can be fixed on the table to restrict the trainees’ direct view of the vessel model. In addition, the height of EVE could be changed by installing the table on a medical stretcher (Fig. 1B and 1C).

Development of electrically controlled camera-holding C-armWe developed a camera holder that could be rotated in craniocaudal and lateral arcs by mimicking the C-arm of the angiography system. The movements of the camera holder were electrically controlled with button switches. For the safety of EVE itself and trainees, the rotation range in the craniocaudal direction was restricted to 90° (45° cranially and 45° caudally), and that in the lateral direction to 120° (60° to each side). A remote-controlled video camera with autofocusing and zooming functions, which was connected to a TV monitor, was mounted on the camera holder (Fig. 1D and 1E).

A simulator system that would support catheter treat-ment training was structured by combining EVE, a sliding table, a stretcher, a camera-holding arm, and a liquid crystal monitor (Fig. 1F).

Simulation of mechanical thrombectomyTo examine whether or not catheter treatment training is possible using this system, three experts of the Japanese Society for Neuroendovascular Therapy including authors (Y.H., T.S., and T.M.) carried out simulation of mechanical thrombectomy. First, they were given instructions about the use of this simulator system. Devices including an 8 Fr. OPTIMO temporary occlusion balloon (Tokai Medical Products, Aichi, Japan), 5 Fr. Excellent EN Catheter Hanafy Type (Hanako Medical, Saitama, Japan), 0.035-inch Radi-focus Guidewire M Angle type (Terumo Corporation, Tokyo, Japan), Marksman microcatheter (Medtronic, Dublin, Ire-land), 0.014-inch microwire Asahi Chikai (Asahi Intecc Co., Ltd, Aichi, Japan), and thrombectomy stent Trevo XP Provue Retriever 4 × 20 mm (Stryker, Kalamazoo, MI, USA) were prepared. In advance, the left middle cerebral artery was occluded with an artificial thrombus, a 9 Fr. sheath was inserted into the right femoral artery, and the artificial vascular channels of EVE were filled with 10 L of tap water dyed with BLOODACT (FAIN-Biomedical) Blue. To the water, 20 mL of Surfactant A and 100 mL of Surfactant B of BIOACT (FAIN-Biomedical) were added to facilitate sliding of the devices, and the channels were perfused with a pulsed flow. Information from the camera

Fig. 1 EVE, artificial silicone vessel model (A) developed freely sliding table (B) table is moved to left side; (C) table is moved to cranial side, developed electric-controlled camera-holding C-arm (D) the arm is vertically posi-tioned; (E) the arm is tilted 30° and total system for simulative catheter intervention training (F).

Journal of Neuroendovascular Therapy Vol. 12, No. 1 (2018)

Page 3: Development of Simulator System Using Endovascular Evaluator for Catheter Intervention ... · 2019-08-11 · 1. 15n211. Development of Simulator System Using . Endovascular Evaluator

3

Simulator System for Catheter Inter vention Training

without directly watching the vessel model even once during the simulation. The guiding time was 183, 163, and 209 seconds, stent placement time was 343, 397, and 423 seconds, and stent retrieval time was 652, 586, and 611 seconds. The number of table movements was 14, 18, and 15. The number of camera-holder arm movements was 9, 8, and 8 (Table 1).

Discussion

In the field of clinical medicine, surgery, in particular, the ability to eventually put knowledge into practice is of vital importance. The objective of simulation training is to fill this gap between knowledge and practice. In the field of neurosurgery, also, there have been reports on simula-tion training of neuroendoscopy, microscopic surgery, and catheter therapy. In conducting simulation training, it is important to create an environment that makes trainees feel as if they were in an actual clinical situation and allow them to concentrate on the training, and an excellent simu-lator is necessary to faithfully reproduce clinical settings in the training environment. Resusci Anne (Laerdal Medical Japan, Tokyo, Japan) and Sim Man (Laerdal) used in Basic Life Support and Advanced Life Support6) and LapVR (CAE Healthcare FL, USA) used for simulation training in endoscopic surgery7) have improved the quality of

was displayed on the monitor screen, and the angiogra-phy-guided intraprocedural environment was mimicked by placing the monitor between the operator and EVE to pre-vent direct observation of the artificial vascular channels (Fig. 2A). The operators performed thrombectomy by a pro-cedure as close as possible to their usual clinical practice. The 8 Fr. OPTIMO was placed in the cervical internal carotid artery using a coaxial system via the aorta (Fig. 2B and 2C). Then, the TREVO XP was deployed at the thrombus using the Marksman (Fig. 2D), and the stent was retrieved after 5 minutes. On this simulation, the time from the insertion of the 8 Fr. OPTIMO into the sheath to its placement in the internal carotid artery (guiding time), time until the deployment of TREVO XP at the site of occlusion (stent placement time), and time until stent retrieval (stent retrieval time) were measured. In addition, the numbers of movements of the table and camera-holding arm during the procedure were recorded. The procedure times were measured by dis-regarding whether the thrombus could be retrieved or not.

Results

The three physicians could perform the entire procedure of mechanical thrombectomy using this simulator system. They all could complete the simulation by totally depend-ing on the information displayed on the monitor screen

Fig. 2 Perspective view of simulative mechanical thrombectomy (A). Monitor view during simulative mechan-ical thrombectomy (B–D). Anterior-posterior view of aortic arch and left common carotid artery can-nulated with 8 French Fr. Optimo balloon guiding catheter (arrow) and preceding 6 French Fr. inner-catheter (arrow head) (B), left oblique view of cervical internal carotid artery cannulated with Optimo catheter (arrow head) and preceding inner-catheter (arrow head) (C) and magnified right oblique view of artificial embolus (arrow head) captured with TREVO XP stent retriever (arrow) in the left middle cerebral artery (D).

Journal of Neuroendovascular Therapy Vol. 12, No. 1 (2018)

Page 4: Development of Simulator System Using Endovascular Evaluator for Catheter Intervention ... · 2019-08-11 · 1. 15n211. Development of Simulator System Using . Endovascular Evaluator

4

Koyama J, et al.

simulation training due to their excellent functions. In the field of neurosurgery, various simulators for training in bypass surgery and endoscopic surgery have been reported.8–11)

To make simulation training in neuroendovascular treat-ment more practical, it is necessary to make the training environment closer to clinical settings. Trainees are required not only to learn the procedure of catheter treatment and the usage of instruments, but also to experience compre-hensive simulation including rinsing inside of the catheter, imaging, catheter and wire manipulation, and handling of various other devices. In the field of neuroendovascular treatment, a few simulators have been developed, and VIST (Mentice AB, Gothenburg, Sweden) is one of them.12) Since VIST is a digital system, various devices can be manipulated virtually, but they are limited to dedicated instruments, most of the devices that are actually used in clinical settings including catheters cannot be used. On the other hand, the silicone vessel model EVE can be perfused with a liquid through the artificial vascular channels and is superior as a simulator in that various maneuvers that are performed in actual treatment, such as rinsing and imaging, can be simu-lated and that most catheter-related devices in clinical use can be used. However, in actual training using EVE, time was needed for moving the manually manipulated camera, and it often became necessary to directly observe the artificial ves-sels, distracting trainees from simulation. We, therefore, developed a simulator system that would prepare an environ-ment more faithful to the clinical settings by combining a sliding table and an electrically controlled camera-holding arm with EVE.

In this study, simulation of mechanical thrombectomy was performed among neuroendovascular procedures because it includes many important technical elements of catheter treatment. As this procedure requires trainees to manipulate the microcatheter and microwire and eventu-ally deploy stents in intracranial vessels in addition to the placement of a guiding catheter, perfusion, and imaging, they can learn basic techniques of endovascular treatment. In addition, mechanical thrombectomy has been reported to be effective as a treatment for acute-phase cerebral

infarction,13–16) and shortening of the time until recanaliza-tion is a condition for improving the outcome. Therefore, simulation training in thrombectomy is considered to con-tribute to improvements in the clinical effectiveness and safety of catheter treatment.

The simulator system that we have developed cannot reproduce all aspects of clinical practice. In clinical situa-tions, patients may not be sufficiently sedated, and the surgeon is placed under psychological stress. The fidelity of simulation training has four aspects, that is, physical, technical, environmental, and psychological, but the fidel-ity pursued by the present system is limited to the technical aspect.17) The three neuroendovascular surgeons completed the entire procedure of mechanical thrombectomy in about 20 minutes, which is considered shorter than the time required for actual treatment. In addition, as the sliding table and C-arm were operated under monitoring with a high-resolution camera, the numbers of their movements were fewer. These are the results of simulation performed by neuroendovascular surgeons under specific conditions and cannot be directly compared with actual clinical data. Therefore, the results of this study are expected to serve as reference data for the assessment of the achievement level in future simulation training.

This study disclosed some points that need to be evalu-ated or improved in this system. Dyeing the water in the channel improves the visibility of blood vessels monitored with a video camera, but devices in the vessels were occa-sionally difficult to observe due to refraction of light by the silicone vascular wall. However, this refraction is expected to be reduced considerably by modifying the lighting method. It must also be noted that, in this system not using radiation, the position of a device inserted into a catheter cannot be confirmed until it is pushed out of the catheter unlike actual clinical practice. Moreover, the ves-sel models of EVE used in this system had standard shapes, but the degree of meandering and furcation angles of actual blood vessels vary among patients. Therefore, simulation training can be made more practical by prepar-ing a wide variety of vessel models and making them alterable.

Table 1 Result of simulative mechanical thrombectomy

Interventionists A B C

Time required for guiding catheter placement (sec) 183 163 209Time required for stent placement (sec) 343 397 423Time required for stent retrieval (sec) 652 586 611Number of moving table 14 18 15Number of moving camera 9 8 8

Journal of Neuroendovascular Therapy Vol. 12, No. 1 (2018)

Page 5: Development of Simulator System Using Endovascular Evaluator for Catheter Intervention ... · 2019-08-11 · 1. 15n211. Development of Simulator System Using . Endovascular Evaluator

5

Simulator System for Catheter Inter vention Training

6) Hesselfeldt R, Kristensen MS, Rasmussen LS: Evaluation of the airway of the SimMan full-scale patient simulator. Acta Anaesthesiol Scand 2005; 49: 1339–1345.

7) Iwata N, Fujiwara M, Kodera Y, et al: Construct validity of the LapVR virtual-reality surgical simulator. Surg Endosc 2011; 25: 423–428.

8) Ribeiro de Oliveira MM, Nicolato A, Santos M, et al: Face, content, and construct validity of human placenta as a hap-tic training tool in neurointerventional surgery. J Neurosurg 2016; 124: 1238–1244.

9) Tai BL, Wang AC, Joseph JR, et al: A physical simulator for endoscopic endonasal drilling techniques: technical note. J Neurosurg 2016; 124: 811–816.

10) Tai BL, Rooney D, Stephenson F, et al: Development of a 3D-printed external ventricular drain placement simulator: technical note. J Neurosurg 2015; 123: 1070–1076.

11) Benet A, Rincon-Torroella J, Lawton MT, et al: Novel embalming solution for neurosurgical simulation in cadavers. J Neurosurg 2014; 120: 1229–1237.

12) Aggarwal R, Black SA, Hance JR, et al: Virtual reality simulation training can improve inexperienced surgeons’ endovascular skills. Eur J Vasc Endovasc Surg 2006; 31: 588–593.

13) Sardar P, Chatterjee S, Giri J, et al: Endovascular ther-apy for acute ischaemic stroke: a systematic review and meta-analysis of randomized trials. Eur Heart J 2015; 36: 2373–2380.

14) Balami JS, Sutherland BA, Edmunds LD, et al: A system-atic review and meta-analysis of randomized controlled trials of endovascular thrombectomy compared with best medical treatment for acute ischemic stroke. Int J Stroke 2015; 10: 1168–1178.

15) Chen CJ, Ding D, Starke RM, et al: Endovascular vs medi-cal management of acute ischemic stroke. Neurology 2015; 85: 1980–1990.

16) Phan K, Zhao DF, Phan S, et al: Endovascular therapy including thrombectomy for acute ischemic stroke: a sys-tematic review and meta-analysis with trial sequential analysis. J Clin Neurosci 2016; 29: 38–45.

17) Shiga T, Takeda S, Mandai Y, et al: [Practical approach to simulation-based education first edition]. Medical sciences international, Ltd., Tokyo, 2014, 16–21. (in Japanese)

Conclusion

We developed a sliding table and an electrically controlled camera-holding C-arm and formulated a simulation system for catheter treatment training by combining them with the sil-icone vessel model EVE. Three experts in neuroendovascular treatment attempted simulation of mechanical thrombectomy using this simulator and could complete the entire procedure. This system is considered to have basic functions necessary for simulation training in catheter treatment.

Acknowledgments

We would like to express our gratitude to Mr. Koji Ushiyama, Mr. Satohiko Kubota, and Shinapex Co., Ltd. for their cooperation in the development of this system.

Disclosure Statement

Neither the first author nor any of the coauthors have any conflicts of interest.

References

1) Wayne DB, Didwania A, Feinglass J, et al: Simulation- based education improves quality of care during cardiac arrest team responses at an academic teaching hospital: a case-control study. Chest 2008; 133: 56–61.

2) Banasik Z, Sledziński Z, Arciszewska D, et al: The useful-ness of Resusci-Anne manikin in teaching modern methods of resuscitation. Anaesth Resusc Intensive Ther 1976; 4: 131–137.

3) Issenberg SB, McGaghie WC, Petrusa ER, et al: Features and uses of high-fidelity medical simulations that lead to effective learning: a BEME systematic review. Med Teach 2005; 27: 10–28.

4) Kim J, Park JH, Shin S: Effectiveness of simulation-based nursing education depending on fidelity: a meta-analysis. BMC Med Educ 2016; 16: 152.

5) FAIN-Biomedical, Inc.: Endovascular surgery training- evaluation-simulation EVE endoVascular evaluator. http://fain-biomedical.com/fbm_wp/wp-content/themes/fbm_ns/images/pdf/eve_fbm_e.pdf. (Accessed January 05, 2017).

Journal of Neuroendovascular Therapy Vol. 12, No. 1 (2018)