f6150sv ts us lr

Upload: irfan-akhtar

Post on 04-Jun-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/13/2019 f6150sv Ts Us Lr

    1/4

    * Maximum power delivered at high source range

    CONVERTIBLE AMPLIFIERS

    Current Mode

    Convertible Source AC Ranges Convertible Source DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    6 X 0.5, 1.0 A RMS (L-N) 6 X 0.354, 0.707 A DC 6 X 75VA / 75 W 0.0001 A 0.0001 A

    3 X 0.5, 1.0, 2.0 A RMS (L-N) 3 X .354, 0.707, 1.41 A DC 3 X 150VA / 150 W 0.0001 A 0.0001 A

    1 X 1.5, 3.0, 6.0 A RMS (L-N)S1 S2 S3

    1 X 1.06, 2.12, 4.24 A DCS1 S2 S3

    1 X 450VA / 450 WS1 S2 S3 0.001 A 0.001 A

    Current Transient Mode

    Convertible Source AC Ranges Convertible Source DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    6 X 0.75, 1.5 A RMS (L-N) 6 X 0.53, 1.06 A DC 6 X 97.5VA / 97.5 W 0.0001 A 0.0001 A

    3 X 0.75, 1.5, 3.0 A RMS (L-N) 3 X 0.53, 1.06, 2.12 A DC 3 X 195VA / 195 W 0.0001 A (0.0001 A) @ 0.53, 1.06 A DC(0.001 A) @ 2.12 A DC

    1 X 2.25, 4.5, 9.0 A RMS (L-N)S1 S2 S3

    1 X 1.59, 3.18, 6.36 A DCS1 S2 S3

    1 X 585VA / 585 WS1 S2 S3 0.001 A 0.001 A

    Voltage Mode

    Convertible Source AC Ranges Convertible Source DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    6 X 75, 150V RMS (L-N) 6 X 106, 212V DC 6 X 75VA/ 75W 0.01 V (0.01 V) @ 106V DC(0.1 V) @ 212V DC

    3 X 75, 150, 300V RMS (L-N) 3 X 106, 212, 424V DC 3 X 150VA / 150 W 0.01 V (0.01 V) @ 106V DC(0.1 V) @ 212, 424V DC

    1 X 150, 300, 600V RMS (L-L)S1 & S2

    1 X212, 424, 848V DCS1 & S2

    1 X 300VA / 300 WS1 & S2 0.01 V 0.1 V DC

    ENHANCED CURRENT AMPLIFIERS

    AC Ranges DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    6 X 8.75, 17.5, (35)* A RMS (L-N) 6 X 5.83, 11.7, (23.3)* A DC 6X 87.5VA / 87.5 W;(6 X 131.25VA /131.25 W)* 0.001 A (0.001) A @ 5.83 A DC

    (0.01) A @ 11.6, (23.3)* A DC

    3 X 8.75, 17.5, 35,(70)* A RMS (L-N) 3 X 5.83, 11.7, 23.3,(46.6)* A DC

    3 X 175VA / 175 W;(3 X 262.5VA /262.5 W)*

    (0.001) A @ 8.75, 17.5,35 A RMS

    (0.01) @ (70)* A RMS

    (0.001) A @ 5.83 A DC(0.01) A @ 11.7, 23.3, (46.6)*

    A DC

    1 X 8.75, 17.5, 26.25,52.5, 105,

    (210)* A RMS (L-N)S1 S2 S3

    1 X 5.83, 11.7, 17.5, 35,70, (140)*

    A DCS1 S2 S3

    1 X 525VA / 525 W;

    (1X787.5VA/787.5 W)*S1 S2 S3

    (0.001) A @ 8.75,17.5, 26.25 A RMS

    (0.01) A @ 52.5, 105,(210)* A RMS

    (0.001) A @ 5.83 A DC

    (0.01) A @ 11.7, 23.3, 35, 70,(140)* A DC

    The F6150sv li e f P we System Simulat s is speci cally esi e t p vi e IEC 61850 c mplia t testi ,sta a Wi-Fi a e ha ce ati s. The e a e th ee isti ct m els f y u speci c testi ee s.

    The F6150sv premier model features 12 analog sources and can run in a mixed mode of operation with acombination of analog and sampled values sources.

  • 8/13/2019 f6150sv Ts Us Lr

    2/4

    sv-SGD

    CONVERTIBLE AMPLIFIERS

    Current Mode

    Convertible Source AC Ranges Convertible Source DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    4 X 0.5, 1.0 A RMS (L-N) 4 X 0.354, 0.707 A DC 4 X 75VA / 75 W 0.0001 A 0.0001 A

    4 X 0.5, 1.0 A RMS (L-N) 2 X .354, 0.707, 1.41 A DC 2 X 150VA / 150 W (0.0001 A) @0.5, 1.0 A RMS (0.001 A) @ 2.0 A RMS

    (0.0001 A) @0.354, 0.70 A DC (0.001 A) @ 1.4

    A DC

    1 X 1.0, 2.0, 4.0 A RMS (L-N)S1 S2

    1 X 0.707, 1.41, 2.82 A DCS1 S2

    1 X 300VA / 300 WS1 S2

    (0.0001 A) @1.0 A RMS(0.001 A) @ 2.0, 4.0 A RMS

    (0.0001 A) @0.707 A D(0.001 A) @ 1.41, 2.82

    A DC

    Current Transient Mode

    Convertible Source AC Ranges Convertible Source DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    4 X 0.75, 1.5 A RMS (L-N) 4 X 0.354, 0.707 A DC 4 X 97.5VA / 97.5 W 0.0001 A 0.0001 A

    2 X 0.75, 1.5, 3.0 A RMS (L-N) 2 X .354, 0.707, 1.41 A DC 2 X 195VA / 195 W (0.0001 A) @0.75 A RMS(0.001 A) @ 1.5, 3.0 A RMS

    (0.0001 A) @ 0.53 A D(0.001 A) @ 1.06, 2.12

    A DC

    1 X 1.5, 3.0, 6.0 A RMS (L-N)S1 S2

    1 X 0.707, 1.41, 2.82 A dcS1 S2

    1 X390VA / 390 WS1 S2 0.001 A 0.001 A

    Voltage Mode

    Convertible Source AC Ranges Convertible Source DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    4 X 75, 150V RMS (L-N) 4 X 106, 212V DC 4 X 75VA/ 75W 0.01 V (0.01 V) @ 106V DC(0.1 V) @ 212V DC

    2 X 75, 150, 300V RMS (L-N) 2 X 106, 212, 424V DC 2 X 150VA / 150 W 0.01 V (0.01 V) @ 106V, 212DC (0.1 V) @ 424V D

    1 X 1.5, 3.0, 6.0 A RMS (L-N)S1 & S2

    1 X212, 424, 848V DCS1 & S2

    1 X 300VA / 300 WS1 & S2 0.01 V 0.1 V DC

    Enhanced Mode

    AC Ranges DC Ranges Power Resolution AC Ranges Resolution DC Ranges

    4 X 8.75, 17.5, (35)** A RMS (L-N) 4 X 5.83, 11.7, (23.3)** A DC 4X 87.5VA / 87.5 W;(4 X 131.25VA /131.25 W)** 0.001 A (0.001) A @ 5.83 A D

    (0.01) A @ 11.6, (23.3) A DC

    2 X 8.75, 17.5, 35,(70)** A RMS(L-N) 2 X 5.83, 11.7, 23.3,(46 .7)** A DC

    2 X 175VA / 175 W;(2 X 262.5VA /262.5 W)**

    (0.001) A @ 8.75, 17.5 A RMS(0.01) @ 35, (70)** A RMS

    (0.001) A @ 5.83 A D(0.01) A @ 11.7, 23.3,

    (46.7)** A DC

    1 X 8.75, 17.5, 35, 70, (140)* A RMS (L-N)

    S1 S2

    1 X 5.83, 11.7, 23.3, 46.7 (93.2)** A DC

    S1 S2

    1 X 350VA /350W;( 1X525VA/525 W)**

    S1 S2

    (0.001) A @ 8.75, 17.5 A RMS(0.01) A @ 35, 70, (140)** A RMS

    (0.001) A @ 5.83 A DC(0.01) A @ 11.7, 23.3, 46.7

    (93.3*)* A DC** Maximum power delivered at high source range

    The F6150sv Smart Grid Distribution (SGD) model features 8 analog sources and can run in a mixed mode ofoperation. It can be used for IEC 61850 Station and Process Bus testing compliant with 9-2LE.

  • 8/13/2019 f6150sv Ts Us Lr

    3/4

    sv-IEC

    The F6150sv IEC model has no analog source outputs. It can be used for IEC 61850 Station and ProcessBus testing compliant with 9-2LE. Its low level source mode is used to control the F5850 Intellirupter testinginterface.

    LOW LEVEL SOURCE PARAMETER RANGES

    Voltage Mode Current Mode Transient Mode

    C ve tible Ampli e S u ces 6.7 VRMS 4.5 VRMS 6.7 VRMS

    Cu e t Ampli e S u ces N/A 3.38 VRMS(3.48 VRMS for enhanced ratings)6.7 VRMS

    (6.96 VRMS for enhanced ratings)

    LOW LEVEL SOURCE ACCURACIES

    Output signal accuracy 0.25% oF rEAdIng

    Source impedance 50

    Please refer to the F6150sv user guide for output scaling and range details.

  • 8/13/2019 f6150sv Ts Us Lr

    4/4

    Metering Functions

    dC Mete I puts

    Input Range: 0 - 10V DC / 0 - 20mA DC

    Typical: