DESCRIPTION

JUMO CTI-500. PL20 Analytical measurment systems. 1 Target groups / Application. General water technology Monitoring of coolingwater Cleaning baths (e.g. galvanic) Intake to internal waste water plants Monitoring of drinking water Bottle cleaning plants - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: JUMO CTI-500

1 Reinhard Manns / PL20

JUMO CTI-500

PL20PL20

Analytical measurment systemsAnalytical measurment systems

Page 2: JUMO CTI-500

2 Reinhard Manns / PL20

1 Target groups / Application

• General water technology

• Monitoring of coolingwater

• Cleaning baths (e.g. galvanic)

• Intake to internal waste water plants

• Monitoring of drinking water

• Bottle cleaning plants

• CIP-cleaning (up to max. 100°C; Dairies)

Page 3: JUMO CTI-500

3 Reinhard Manns / PL20

2. Concept JUMO CTI - 500

JUMO CTI - 500

Page 4: JUMO CTI-500

4 Reinhard Manns / PL20

2. Concept JUMO CTI - 500

JUMO CTI - 500

Page 5: JUMO CTI-500

5 Reinhard Manns / PL20

2.1 Gerätekonzept Blockschaltbild

Block diagram JUMO CTI - 500

Signal preparation:

•Using temperature compensation at conductivity signal

output conductivity: temperature compensated conductivity signal

0(4)... 20mA 0(2)... 10V 20... 0(4)mA 10... 0(2)V

SETUP-Connection:Configuration/Parametrierung

Power supply:DC 24V (DC19..30V)

Measuring cell: inductive cell Temperature sensor

Binary inputs: 2

relay: 2 X Halbleiterschalter

AC/DC 15..50V; 200mA

output temperature: 0(4)... 20mA 0(2)... 10V 20... 0(4)mA 10... 0(2)V

Operation: LC-Grafic display 4-Keys 2-LED´s

Page 6: JUMO CTI-500

6 Reinhard Manns / PL20

2.2 Concept applications

Features:

• cost effective devise

• compacte size

• Devices

• Head transmitter

• stand alone version

Page 7: JUMO CTI-500

7 Reinhard Manns / PL20

2.2 Concept

Features:

• Operating / Programming

• Keyboard and Display

• Setup-program

• comfortable programming

• System documentation

• Power supply

• DC 19…31V

Page 8: JUMO CTI-500

8 Reinhard Manns / PL20

2.3 Measuring areas

Features:

• Measuring area freely configurable

• Measuring areas

• 0 … 500µS/cm

up to

• 0 … 2000mS/cm

• up to 4 areas activated

Page 9: JUMO CTI-500

9 Reinhard Manns / PL20

2.4 Measuring units

Features:

• Units

• µS/cm

• mS/cm

• %

• 5 signs freely editable

Page 10: JUMO CTI-500

10 Reinhard Manns / PL20

2.4 Measuring units

Features:

• Concentration measuring

• Soda lye NaOH

• Nitric acid HNO3

• 1 X free definable graph (entering 20 pairs of variates with setup-programm)

Page 11: JUMO CTI-500

11 Reinhard Manns / PL20

2.5 Temperature input

Features:

• Units (Temperature)

• °C

• °F

• Temperature sensor

• fast response

resp..

• Protected Fitting inside the measuring cell

Page 12: JUMO CTI-500

12 Reinhard Manns / PL20

2.6 Compensation

Features:

• Form of temperature compensation

• linear

• natural water (EN 27888)

• own characteristic curve (Learnfunction)

• activate up to 4 values

Page 13: JUMO CTI-500

13 Reinhard Manns / PL20

2.7 Outputs

Features:

• 2 analog continuous outputs for conductivity / resitance and parallel for temperature (electrical isolated)

• 0/4..20mA

• 0..10V

• 20..4/0mA

• 10..0V

• freely programmable in measuring areas

Page 14: JUMO CTI-500

14 Reinhard Manns / PL20

2.7 Outputs

Features:

• 2 integrated potential free contacts (semi conductor switches)

• 0,2 A, AC 14…50V• 0,2 A, AC 14…50V

Page 15: JUMO CTI-500

15 Reinhard Manns / PL20

2.8 General concept

Features:

• 2 binary inputs (the function is aktivated by a potential free contact)

• keyboard locking

• changing measuring areas

• changing temperature coefficent

• desalinisation function

Page 16: JUMO CTI-500

16 Reinhard Manns / PL20

2.8 General concept

Features:

• Connections

• PVC T-part DN32

• PVC T-part DN40

• Clamp 2 ½”

• screwed pipe fitting DN50, DIN 11 851(MK DN50, milk cone)

• SMS 2”

Page 17: JUMO CTI-500

17 Reinhard Manns / PL20

2.8 General concept

Features:

• Operating

• Keyboard

• Graphic display

Page 18: JUMO CTI-500

18 Reinhard Manns / PL20

2.8 General concept

Features:

• Display

Output-ModusHAND (Hand-Mode)HOLD (Hold-Mode)

K1 K2 E1 E2 ALARM HAND

MESSUNG 10.24 18.5 °C mg/l

contactsIf visible contact is activ

Binary inputsIf visible input is activ

InformationsALARM (e.g.Overrange)KALIB flashing (kalibration timer ended)KALIB (kalibration aktiv)

Activitymeasurement (normal)

Temperature and unit value and unit

Informations

Page 19: JUMO CTI-500

19 Reinhard Manns / PL20

2.8 General concept

Features:

• languages

• german

• english

• french

• spanish

• polish

Page 20: JUMO CTI-500

20 Reinhard Manns / PL20

3. Application

Desalinisation in cooling towers

During the process you have to consider the following points: • water has to be refilled (by permanent evaporation water is getting lost)

• change water in sporadic intervals (Desalinisation: by permanent evaporation the liquid is going away, but the ingredients (salt) are increasing. The conductivity is getting higher; deposits in the tubes.

• You have to prevebt the growth of algae (Addition of biocide; in closed systems is a permanent growth of algae possible; that contaminated the cooling tower and could make problems in process e.g. the degree of efficiency is getting down. Therefor you have to add biocides in periodic lags. A secondary effect is the growing of the conductivity.)

Page 21: JUMO CTI-500

21 Reinhard Manns / PL20

3. Application

Conductivity measurement on cooling towers(max. conductivity app. 2000…3000µS/cm)

JUMO CTI - 500

Page 22: JUMO CTI-500

22 Reinhard Manns / PL20

3.1 Possible application

• General water technology

• Monitoring of cooling water

• Air conditions

• Cleaning baths (e.g. controlling of galvanic bath)

• betriebliche Endkontrollen

Page 23: JUMO CTI-500

23 Reinhard Manns / PL20

3.1 Possible application

• Monitoring of boilers

• Drinking water control

• Reverse Osmosis (RO)

• Cooling of industrial heaters (e.g. casting houses)