mechanical ventilation in acute respiratory failure ... · mode of mechanical ventilation. common...
TRANSCRIPT
Ventilator Graphics
高雄長庚醫院
胸腔內科 呼吸治療科
吳沼漧醫師
Introduction
Understanding ventilator graphics is an
integral part of adequately treating
patients on mechanical ventilators.
Just as pulmonary functions tests are
used to better understand the lung
pathophysiology in nonmechanically
ventilated patients, ventilator graphics
are an important part of understanding
the pathophysiology in mechanically
ventilated patients.
Understanding airway pressures The respiratory system can be thought of as a mechanical
system consisting of resistive (airways +ET tube) and elastic
(lungs and chest wall) elements in series
Diaphragm
ET Tube
airways
Chest wall
PPL Pleural pressure
Paw Airway pressure
Palv Alveolar pressure
ET tube + Airways
(resistive element)
Resistive pressure varies with airflow
and the diameter of ETT and airways.
Flow resistance
The elastic pressure varies with volume and
stiffness of lungs and chest wall.
Pel = Volume x 1/Compliance
Paw = Flow X Resistance + Volume x 1/Compliance THUS
Lungs + Chest wall
(elastic element)
Airways + ET tube
(resistive element)
Lungs + Chest wall
(elastic element)
Understanding the basic
ventilator circuit diagram
ventilator
Diaphragm
Essentially the circuit diagram of a
mechanically ventilated patient can be
broken down into two parts…..
The ventilator makes up the first part
of the circuit. Its pump like action is
depicted simplistically as a piston
that moves in a reciprocating fashion
during the respiratory cycle.
The patient’s own respiratory system
makes up the 2nd part of the circuit.
The diaphragm is also shown as a
2nd piston; causing air to be drawn into
the lungs during contraction.
These two systems are connected by
an endotracheal tube which we can
consider as an extension of the
patient’s airways.
ET Tube
airways
Chest wall
Let us now understand how the respiratory systems’
inherent elastance and resistance to airflow
determines the pressures generated within a
mechanically ventilated system.
Ventilator
Diaphragm
RET tube
Rairways
Raw
Understanding basic respiratory mechanics
The total ‘airway’ resistance (Raw) in the mechanically ventilated patient
is equal to the sum of the resistances offered
by the endotracheal tube (R ET tube)
and the patient’s airways ( R airways)
The total ‘elastic’ resistance (Ers) offered by the
respiratory system is equal to the sum of
elastic resistances offered by the
Lung E lungs and the
chest wall E chest wall
Elungs
Echest wall
Thus to move air into the lungs at any given time (t),
the ventilator has to generate sufficient
pressure (Paw(t)) to overcome the combined
elastic (Pel (t)) and resistance (Pres(t)) properties
of the respiratory system
Ers
ET Tube
airways
Thus the equation of motion for the respiratory system
is
Paw (t) = Pres (t) + Pel (t)
ventilator
Diaphragm
Ppeak
Pres
RET tube
Rairways
Pres
Pplat
Understanding the pressure-time waveform
using a ‘square wave’ flow pattern
time
pre
ssure
The pressure-time waveform is a reflection
of the pressures generated within the
airways during each phase of the
ventilatory cycle.
At the beginning of the inspiratory cycle,
the ventilator has to generate a pressure Pres
to overcome the airway resistance.
Note: No volume is delivered at this time.
After this, the pressure rises in a linear fashion
to finally reach Ppeak. Again at end inspiration,
air flow is zero and the pressure drops by an
amount equal to Pres to reach the plateau
pressure Pplat. The pressure returns to
baseline during passive expiration.
Pres
Now let’s look at some different pressure-time
waveforms using a ‘square wave’ flow pattern
This is a normal pressure-time waveform
With normal peak pressures ( Ppeak) ;
plateau pressures (Pplat )and
airway resistance pressures (Pres)
time
pre
ssure
Pres
Pplat
Pres
Scenario # 1
Paw(peak) = Flow x Resistance + Volume x 1/ Compliance
time
flo
w
‘Square wave’
flow pattern
Paw(peak)
Waveform showing high airways resistance
This is an abnormal pressure-time waveform
time
pre
ssure
Ppeak
Pres
Pplat
Pres
Scenario # 2
The increase in the peak airway pressure is driven
entirely by an increase in the airways resistance
pressure. Note the normal plateau pressure.
e.g. ET tube
blockage
Paw(peak) = Flow x Resistance + Volume x 1/ Compliance + PEEP
time
flo
w
‘Square wave’
flow pattern
Normal
Three areas of clinical
interest in ventilator graphics
Dynamic characteristics of the
respiratory system.
Static compliance of the respiratory
system.
Patient-ventilator interaction--
Asynchrony, synchrony.
Using Ventilator Graphics to
Identify Patient-Ventilator
Asynchrony
Patient-ventilator interaction can be
described as the relationship between 2
respiratory pumps: (1) the patient’s
pulmonary system, which is controlled by
the neuromuscular system and influenced
by the mechanical characteristics of the
lungs and thorax, and (2) the ventilator,
which is controlled by the ventilator
settings.
Clinical applications of ventilator
waveform analysis
• Ventilator waveforms can be very useful in many
different situations including:
– Diagnosing a ventilator that is ‘alarming’
– Detecting obstructive flow patterns on the ventilator
– Detecting air trapping and dynamic hyperinflation
– Detecting lung overdistention
– Detecting respiratory circuit secretion build-up
– Detecting patient-ventilator interactions
• Dyssynchrony
• Double triggering
• Wasted efforts
• Flow starvation
When the 2 pumps function in
synchrony, every phase of the breath is
perfectly matched.
Anything that upsets the harmony
between the 2 pumps results in
asynchrony and causes patient
discomfort and unnecessarily increases
work of breathing.
Deleterious Effects of Patient-
Ventilator Asynchrony Patient fights the ventilator
More sedation required
Higher work of breathing
Muscle damage
Ventilation-perfusion problems
Dynamic hyperinflation
Delayed or prolonged weaning
Longer stay
Higher costs
Factors That Affect Patient-Ventilator
Synchrony Ventilator Factors
Trigger variables
Sensitivity setting
Rise-time capability
Design, mode, and
settings of the flow
delivery system
Flow pattern selected
Design of the exhalation
valve
How PEEP is generated
by the software
Extraneous flow (eg, from
a nebulizer, oxygen)
Patient Factors
Sedation level: pain,
splinting
Inspiratory
effort/respiratory drive;
neural timing
Pathology of the
respiratory system or
abdomen; secretions
Intrinsic PEEP
Size and type of airway
Presence of leaks
Graphical displays
Graphical displays are common in the
ICU. The following are examples of
graphical displays:
Ventilator waveforms
Arterial waveforms
ECG waveforms
Venous waveforms (central venous
pressure)
Intracranial pressure waveforms
……………
Types of Ventilator
Waveforms Scalar:
Pressure – time
Volume – time
Flow – time
Loops:
Pressure – volume
Volume – Flow
Flow – pressure
Trends:
4 components of a breath
Dividing a mechanical breath into the
following 4 components helps to better
identify and correct asynchronies…
Trigger phase
Inspiratory phase
Cycle phase
Expiratory phase
Dyssynchrony = asynchrony
Definition: Uncoupling of the
mechanical delivered breath and neural
respiratory effort.
Dysynchrony can occur in each phase
of the breath
Trigger phase
Inspiratory phase
Cycle phase
Expiratory phase
Trigger Phase
The mechanical breath may be initiated
by the patient (patient trigger) or as a
function of time (time trigger).
Patient trigger-- The 2 common types of
triggering available are pressure and
flow.
Patient Triggering
Trigger phase: 4 components
Trigger pressure (TP)
Inspiratory trigger time (ITT)
Rise time to baseline pressure (RTBP)
Inspiratory delay time (IDT)
The trigger phase can be
divided into 4 components
Trigger pressure (TP) is the pressure that
must be attained by patient effort to trigger
the mechanical breath.
Inspiratory trigger time (ITT) is the time
elapsed from the patient’s effort to reach
the TP set on the mechanical pressure.
in patients with a low respiratory drive, this time
could be prolonged and vice versa.
Rise time to baseline
pressure (RTBP)
RTBP is the time elapsed from the
mechanical breath trigger to attain the
baseline pressure (the PEEP).
The patient does not receive any support until
the circuit is pressurized to the baseline
pressure, and considerable work could be
spent during this time if not appropriately set.
Inappropriate rise times can also affect the
time for the pressure to rise to the peak
airway pressure.
Inspiratory delay time (IDT)
Inspiratory delay time (IDT) is the total
time elapsed from the initial patient
effort to the pressurization of the circuit
to baseline pressure.
In other words, the IDT can be expressed
by the following equation: IDT = ITT +
RTBP.
Trigger-phase asynchrony
Trigger asynchrony can occur with any
mode of mechanical ventilation.
Common trigger problems include
autotriggering, missed triggering, and
double triggering.
Appropriate valve sensitivity settings
are required to avoid overtly sensitive
settings that can lead to autotriggering
and insensitive settings that can lead to
missed triggering.
Auto-triggering
Auto-triggering
Ventilator show “assisted mode”, but no
patient trigger waveform.
Double triggering
Double triggering is the delivery of 2
consecutive ventilator cycles separated
by a very short expiratory time, with the
first cycle being patient triggered.
Double triggering occurs when the
patient’s ventilatory demand is high and
the inspiratory time set on the ventilator
is too short.
Double triggering
Double triggering
Double triggering occurs more
commonly in patients whose
PaO2/FiO2 ratio is lower, such acute
lung injury or ARDS.
Management: Increasing the inspiratory
time or increasing the tidal volumes.
Volume mode changing to a variable flow
(eg, pressure-control ventilation) or a dual-
control mode.
Sedation adjustments
Triggering delay
Trigger delay indicated by a large
negative deflection in pressure
waveform prior to delivery of breath.
ineffective efforts
(Trigger failure)
Trigger failure patient factors
low respiratory drive: Alkaline pH
weak inspiratory muscles
partially blocked ETT or tracheostomy
dynamic hyperinflation resulting in intrinsic
PEEP, eg COPD
ventilator factors
high level of pressure support or high tidal
volume causing intrinsic PEEP
expiratory asynchrony with delayed opening
of exhalation valve
in-line nebulizers: flow >> pressure triggering
Air Trapping
Inspiration
Expiration
Normal Abnormal
Time (sec)
Flo
w (
L/m
in)
Air Trapping Auto-PEEP
}
※ Causes of air trapping or auto PEEP?
※ How to measure the magnitude of auto PEEP?
1T 2T 3T 1T 2T 3T
Trial & error for
applied PEEP
If measured PEEPi is not available, trial and error can be used to provide applied PEEP.
One would add small increments (e.g., 2 cm H20) and watch the patient response. If successful, the patient effort to trigger the breaths should become less as the appropriate level of PEEP is provided.
Increasing dyspnea and other signs of increasing intrathoracic pressure (e.g., decreasing blood pressure, worsening of clinical signs of hyperinflation:↓BS, chest hyperresonant to percussion) suggest that excessive PEEP is being provided.
NAVA ( Neurally Adjusted
Ventilatory Assist ) in COPD
synchronization with NAVA is perfect.
Inspiratory Phase
Flow may be inadequate or excessive,
both of which may contribute to patient-
ventilator asynchrony.
inappropriate flow rates and patterns
may lead to a number of secondary
asynchronies.
For example, inappropriate flow may
reduce the expiratory time, leading to
auto-PEEP, which may lead to
ineffective triggering.
Flow can be delivered in the
following 3 forms Fixed flow (eg, CPPV, SIMS)
Variable flow (eg, PCPPV)
Combined fixed and variable flows (dual
modes, eg, volume-assured pressure
support and pressure augmentation)
Asynchrony in
inspiration phase
Inadequate Inspiratory Flow
Induce pt-ventilator dys-synchrony
Adequate Flow
Time (sec)
Inadequate Flow
Pa
w
(cm
H2O
) F
low
Flow starvation
The Stress Index
• In AC volume ventilation using a
constant flow waveform observe
the pressure time scalar.
• Normal, linear change in airway
pressure
Stress index =1
• Upward concavity indicates
decreased compliance and lung
overdistension
Stress index > 1
• Downward concavity indicates
increased compliance and
potential alveolar recruitment
Stress index < 1
flow
time
Paw
Note: Patient effort must be absent
STRESS INDEX (SI) and
prevention of VILI Atelectrauma : SI < 0.9
Barotrauma: SI > 1.1
The SI window of the SERVO-i : your
guide to lung protection
A peak ? B peak ? What’s the cycle-off in figure?
How to adjust rise-time or adjustable flow-cycle?
What’s the cycle-off criteria for PSV ?
Pressure
Flow
Volume
(L/min)
(cm H2O)
(ml)
Time (sec)
Normal inadequate rise time
A primary asynchrony may lead to
multiple secondary asynchronies.
Premature inspiratory-
expiratory cycling inspiratory muscle contraction
continues into mechanical expiratory
phase
associated with modes of ventilation
with short inspiratory time
in pressure support mode:
low levels of pressure support
short respiratory time constant (eg ARDS)
relatively high cycling off threshold
dynamic hyperinflation
Premature inspiratory-
expiratory cycling may result in a characteristic change in
flow waveform
Flow先高後低
Delayed opening of
exhalation valve mechanical inspiration continues into
neural expiration
may result in dynamic hyperinflation,
particularly in patients with COPD which
may increase trigger dyssynchrony
associated with long set inspiratory time
in pressure support mode:
long respiratory time constant (eg COPD)
high pressure support level
low cycling off threshold
The subject’s expiratory effort begins just
prior to the end of the mechanical inspiratory
time, which causes pressure spikes (arrows).
Also note the small inspiratory zero flow
plateau at the end of the inspiratory flow
waveform, which indicates that flow into the
lung has stopped just prior to mechanical
expiration.
Recognizing active expiration (pressure support)
Look at the flow-time
& pressure-time
waveform
Patient’s active expiration
during the inspiratory
phase causes a pressure spike.
Notice the high and variable
expiratory flow rates due to
varying expiratory muscle effort
This is a high drive state where increased sedation/paralysis and mode change may be appropriate for lung protection.
Pressure
Flow
Volume
(L/min)
(cm H2O)
(ml)
Time (sec)
BIPAP set too long Ti
Flow先低後高
Expiratory Phase--Shortened expiatory time
may lead to auto-PEEP
Inspiration
Expiration
Normal Abnormal
Time (sec)
Flo
w (
L/m
in)
Air Trapping Auto-PEEP
} 1T 2T 3T
1T 2T 3T
Understanding how flow rates affect I/E
ratios and the development of auto PEEP
Lluis Blanch MD, PhD et al: Respiratory Care Jan 2005 Vol 50 No 1
Decreasing the flow rate
Increase the inspiratory time
and consequently decrease the
expiratory time
(decreased I/E ratio)
Thus allowing incomplete emptying
of the lung and the development
of air trapping and auto-PEEP
A Higher Expiratory Flow Rate and a Decreased Expiratory Time Denote a
Lower Expiratory Resistance
1 2 3 4 5 6
SEC
120
120
V .
LPM
Pressue-Volume loop-- overdistension
B
A
0 20 40 60 -20 -40 -60
0.2
0.4
0.6
LITERS
Paw
cmH2O
C
A = inspiratory pressure B = upper inflection point C = lower inflection point
VT
”beaking”
Airway Resistance
Increased expiratory resistance:
secretions, bronchospasms….
Increased inspiratory resistance: kinked
ET tube, patient biting tube….
Flow-Volume Loop
Normal airway obstruction
PIP vs Pplat
Normal High Raw
High Flow Low Compliance
Time (sec)
Paw
(c
m H
2O
)
PIP
PPlat
PIP
PIP PIP
PPlat PPlat
PPlat
Active exhalation and volume gain
Airway Secretions/ Water in the Circuit
Inspiration
Expiration
Volume (ml)
Flow (L/min)
Normal: Abnormal:
“saw-tooth” pattern, on the expiratory part. on the
inspiratory curve in severe case.
Flow (L/min)
Time (sec)
What happen to PSV ?
Time (sec)
Flow (L/m)
Pressure (cm H2O)
Volume (mL)
The most common
cause ?
Inspiration
Expiration
Time (sec)
Flo
w (
L/m
in)
Paw (cm H2O)
Paw
(c
m H
2O
)