nmr relaxation methods in biological systems - bruker · 2018-11-12 · nmr relaxation methods in...

46
Innovation with Integrity NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation Bruker NMR Benutzertagung Frankfurt am Main, 5.11.2018

Upload: others

Post on 18-Jun-2020

21 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Innovation with Integrity

NMR Relaxation Methods in Biological SystemsDaniel MathieuNMR ApplikationBruker NMR BenutzertagungFrankfurt am Main, 5.11.2018

Page 2: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Proteins aren‘t rock solid…

5.11.218 2

Page 3: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Timescale of dynamics

ps ns µs ms s min hours days

R1, R2, hetNOE R1ρ HD-exchange

real-time NMRCT-CPMG

zz-exch.

RDCs

chemical exchange

5.11.218 3

Page 4: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Exchanging Proteins

E

state A

state B�� � ���

������� ≫ � � ����������

A B

5.11.218 4

Page 5: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Kay et al 2008

J. Biomol. NMR, 41, 113–120

CPMG relaxation dispersion

The HEROINE experiment

Ban, Lee, Griesinger et al 2013

J. Biomol. NMR, 57, 73-82

5.11.218 5

Page 6: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Relaxation dispersionWhat and Why?

• The dependency of effective transverse relaxation on an applied spin-lock field which averages exchange contributions

• Relaxation dispersion is a method to characterize exchange processes by NMR

• Can be used to characterize invisible states

• Can yield dynamic and thermodynamic parameters

5.11.218 6

Page 7: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Measurement of CT CPMG relaxation dispersion

• One reference experiment is recorded without a CPMG train

• Multiple CPMG fields for a constant relaxation period T are recorded

• Relaxation rates are determined by comparison to the reference

• Saves a lot of time compared to individual R1ρ

relaxation measurements at every field.

• CPMG pulse trains deposit a lot of power into the sample which leads to heating

• Scan wise interleaved measurement

• Dummy pulses in every scan to achieve temperature compensation

5.11.218 7

Page 8: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Outcome in the presence of exchange

• Exchange rates Rex contribute tothe effective rate R2,eff

• Exchange contributions can bechanged by variation of the CPMGfield

• In case of a two-site exchange,this can be described analytically

• The lowest possible field strengthis limited by the length of T(longer means lower fields)

• The highest possible fieldstrength is limited by thehardware capabilities for thesame time T

Kay et al J. Biomol. NMR 2008, 41,113–120

5.11.218 8

Page 9: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Sounds simple, what could go wrong?

• Data fully made up

• In this case 12 CPMG field strengths, data at 200 and 1250Hz using threereplicates

• Exchange occuring in the slow to medium time regime

5.11.218 9

Page 10: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Error case 1:

• At a first look some data points seem to be completely off

• Any given constant time delays only allows for certain field strengths

• Any field strength that can not be realized is rounded to the closestpossible value, leaving the constant time as it‘s supposed to be

5.11.218 10

Page 11: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Error case 2:

• Effective relaxation rates seem to go up towards higher field strengths

• Occuring only for off-resonance peaks? Might be an off-resonance effect, on newer probes try shorter CPMG pulses (e.g. 80 µs 180° pulses) orphase cycled CPMG pulses

• Is it occuring for all peaks?

• Possibly due to miscalibration or detuning of the respective channel

5.11.218 11

Page 12: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Error case 3:

• The replicate measurements do not match, however the general shape ofthe curve does not indicate poor signal to noise.

• Heating effects: the intensity of one measurement depends on the amountof power used in the previous scan.

• Use temperature compensation instead of scan wise interleavedmeasurements

5.11.218 12

Page 13: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

(Error) case 4:

• The measured R2 rates do not reach a plateau but instead decay even upto the highest utilized CPMG field

• If this is the case for all peaks that do show exchange, reduce the constanttime in favour of higher CPMG fields (or combine with the HEROINE experiment)

5.11.218 13

Page 14: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Error case 5:

• Basically everything happens in between point one and two

• Increase the constant time to be able to user smaller steps for possible RF fields (and reduce RFmax)

5.11.218 14

Page 15: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Scan wised interleaved measurement

��

� ���

� ���

� ���

� �

/nbl

• Important: nbl = td1 = # of list entries

5.11.218 15

Page 16: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Temperature compensation

• Dummy pulses to ensure every scan applies the same amount of power

• Increases the overall dutycycle

• Usually scan wised interleaved acquisition is no longer needed

5.11.218 16

Page 17: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Scan wised interleaved acquisition

define list<delay> RF_field=<$VDLIST>

1 ze

d11 pl16:f3 st0

2 6m do:f3

3 6m

[…]

goscnp ph31 cpd3:f3

3m do:f3

3m st RF_field.inc

lo to 3 times nbl

3m RF_field.res

3m ipp3 ipp4 ipp5 ipp6 ipp7 ipp8 ipp9 ipp11 ipp12 ipp31

lo to 4 times ns

d1 mc #0 to 4

F1QF()

F2EA(cal…

5.11.218 17

Page 18: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Temperature compensation

"l3=td1"

aqseq 312

1 ze

d11 pl16:f3

"RF_max=0“

9 20u if "RF_field > RF_max“

{

20u "RF_max=RF_field“

}

3m RF_field.inc

lo to 9 times l3

3m RF_field.res

"d31=RF_field[l11]"

[…]

if "RF_max > d31„

{

20u "cnst31=sqrt((RF_max*RF_max) - (d31*d31))„

20u "TAU1=(1 / (cnst31*4) ) - p30/2000000„

20u "COUNTER2=d21*cnst31*4„

}

[…]

if "abs(RF_max-d31) < 0.1“

{

d21*2

}

else

{

8 TAU1

(p30 ph2):f3

TAU1

lo to 8 times COUNTER2

}

[…]

go=2 ph31 cpd3:f3

d11 do:f3 mc #0 to 2

F1QF(calclc(l11, 1))

F2EA(cal

5.11.218 18

Page 19: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Relaxation dispersionusing Proton decoupling

• Proton decoupling improves sensitivity and relaxational properties of15N

• Proton CW field strength is varied in order to suppress NH magnetization transfer: ��� � 2� · �����

Hansen, Vallurupalli & Kay, J. Phys. Chem. B 2008, 112, 5898-5904

5.11.218 19

Page 20: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Compensating off-resonance effects

• 160µs ϖ-pulse

• 15N projection @ 800MHz

5.11.218 20

Page 21: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Compensating off-resonance effects

• Phase cycled CPMG

• Improves Offset dependency

• Minimum CPMG field for a given constant time increased by a factor of 2

• Can be used as “single train CPMG” (recuces minmum field by a factor of 2)

Jiang, Yu, Zhang, Liu & Yang, J. Magn. Reson. 2015, 257, 1-7

5.11.218 21

Page 22: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Kay et al 2008

J. Biomol. NMR, 41, 113–120

CPMG relaxation dispersion

The HEROINE experiment

Ban, Lee, Griesinger et al 2013

J. Biomol. NMR, 57, 73-82

5.11.218 22

Page 23: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Why even higher power?

• Higher power for the individual pulses reduces off-resonance effects

• One of the fitting parameters is R20 (exchange-free R2)

• This parameter is extracted from the plateau towards higher field strength

• In a lot of cases the plateau is not yet reached

R20?R20?R20?R20?R20?R20?

Really High power!Really High power!Really High power!

R20!R20!

• So why not just measure the end point?

R20!

• So why not just measure the end point?

5.11.218 23

Page 24: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Experimental conditions

• All measurements were performed on u-13C,15N Ubiquitin (1.5 mM in 90% H2O 10% D2O)

• 500 MHz CP-TCI

• 293 K

• 80 µs π-pulses during CPMG trains

5.11.218 24

Page 25: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

The HEROINE experiment

• Heteronuclear Rotating Overhauser Invaded Exchange

• High power T1ρ measurement

• Recorded for different spinlock offsets to get as close to the on resonance condition as possible

• The corresponding CPMG experiment with lower power is temperature compensated with respect to the HEROINE experiment

5.11.218 25

Page 26: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Setup

• Both experiments in one sequence

• Flag decides whether the HEROINE or CT CPMG experiment is executed

• Two lists (one for CPMG frequencies, one for T1ρ spin lock periods)

• All calculations e.g. temperature compensation are done in the pulse program

-DLABEL_R20

5.11.218 26

Page 27: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Results: 5 kHz spin-lock up to 125 ms

• T1 type analysis (e.g. when using PDC)

• Clear dependency on the offset (even @5kHz !)

5.11.218 27

Page 28: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Temperature compensated CT-CPMG measurements

• Spin-lock fields up to 1 kHz for 80 ms (using 6.25 kHz π-pulses)

• Temperature compensated to match the HEROINE experiment

5.11.218 28

Page 29: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Chemical Exchange Saturation Transfer

Vallurupalli et al.

J. Am. Chem. Soc. 2012, 134, 8148

5.11.218 29

Page 30: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

How CEST works

A

B

5.11.218 30

Page 31: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Pulse program

• (pseudo)3D Experiment with a varying 15N B1 field offset

Vallurupalli et al.

J. Am. Chem. Soc. 2012, 134, 8148

5.11.218 31

Page 32: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

What does this look like?

• Basically CW absorption spectra

0

0,2

0,4

0,6

0,8

1

1051151251350

0,2

0,4

0,6

0,8

1

105115125135

5.11.218 32

Page 33: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Experimental setup

5.11.218 33

Page 34: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Experimental parameters

5.11.218 34

Page 35: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Processed result

5.11.218 35

Page 36: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

CEST Analysis

5.11.218 36

Page 37: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Sample Information

5.11.218 37

Page 38: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Data Selection

5.11.218 38

Page 39: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Data Analysis

5.11.218 39

Page 40: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Result

5.11.218 40

Page 41: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Lipari-Szabo type order parameters

TROSY based T1, T1ρ and hetNOE measurements(for perdeuterated proteins)

Lakomek N.A., Ying J. & Bax A.

J. Biomol. NMR, 2012, 53, 209–221.

5.11.218 41

Page 42: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

What’s wrong with the current approach?

• Backbone amide detected Relaxation experiments are very sensitive using current hardware

but…

• Systematic are errors often much larger than random errors due to signal-to-noise limitations. (Especially when using a TROSY read-out)

• Cross correlated relaxation (H-N Dipole - 15N CSA)

• Water cross relaxation (due to poor water saturation or radiation damping)

• Water exchange

5.11.218 42

Page 43: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

New in Topspin 4.0.x / 3.6.x

T1

T1ρ hetNOE

5.11.218 43

Page 44: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

• New sequences:• trt1etf3gpsitc3d.3

• trtretf3gpsitc3d.3

• trnoeetf3gpsi3d.3

• T1 temperature compensated with respect to T1ρ

• No parametersets (yet), use non .3 ones, nbl = 1

• No integrated analysis in DynamicsCenter (yet)

New in Topspin 4.0.x / 3.6.x

5.11.218 44

Page 45: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Thanks…

• Donghan Lee

• Frank Löhr

• Wolfgang Bermel

• Peter Neidig

• Helena Kovacs

• Maxim Mayzel

• You for your attention

5.11.218 45

Page 46: NMR Relaxation Methods in Biological Systems - Bruker · 2018-11-12 · NMR Relaxation Methods in Biological Systems Daniel Mathieu NMR Applikation ... lo to 3 times nbl 3m RF_field.res

Innovation with Integrity

Innovation with Integrity