Page 1: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

NIDEK

OPHTHALMIC YAG LASER SYSTEM

Model YC-1800

SERVICE MANUAL

XYC16*RDA001A/ETotal pages: 86

2005.9

Page 2: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan(Manufacturer) Telephone: (0533) 67-6611

Facsimile: (0533) 67-6610NIDEK CO., LTD : 6th Floor, Takahashi Bldg., No.2, 3-chome, Kanda-jinboucho(Tokyo Office) Chiyoda-ku, Tokyo 101-0051, Japan

Telephone: (03) 3288-0571Facsimile: (03) 3288-0570Telex: 2226647 NIDEK J

NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U. S. A.(United States Agent) Telephone: (510) 226-5700

Facsimile: (510) 226-5750NIDEK SOCIETE ANONYME : Europarc 13, rue Auguste Perret, 94042 CRETEIL, France(Authorized Representative) Telephone: (01) 49 80 97 97

Facsimile: (01) 49 80 32 08

Page 3: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

Table of Contents

§1 INTRODUCTIONS ...................................................................................................... 1-1

§2 CAUTIONS ................................................................................................................... 2-12.1 General ..................................................................................................................... 2-12.2 During Service .......................................................................................................... 2-12.3 During Laser Emission .............................................................................................. 2-12.4 Handling the Laser .................................................................................................... 2-22.5 Adjustment ................................................................................................................ 2-22.6 Using the Energy Meter for the YAG laser ................................................................ 2-2

§3 TROUBLESHOOTING ............................................................................................... 3-1

§4 SUB TROUBLESHOOTING ...................................................................................... 4-14.1 The Pilot Lamp does not Light Up. ............................................................................ 4-14.2 Only the Fixation Lamp does not Light Up. ............................................................... 4-24.3 Only the Backlight Panel does not Light Up. ............................................................. 4-24.4 The Error Display Appears. ...................................................................................... 4-3

4.4.1 Error 1 appears. .............................................................................................. 4-34.4.2 Error 5 appears. .............................................................................................. 4-34.4.3 Error 7 appears. .............................................................................................. 4-44.4.4 Error 10 appears. ............................................................................................ 4-54.4.5 Error 12 appears. ............................................................................................ 4-54.4.6 Error 13 appears. ............................................................................................ 4-54.4.7 Error 14 appears. ............................................................................................ 4-64.4.8 Error 15 appears. ............................................................................................ 4-74.4.9 Error 30 appears. ............................................................................................ 4-74.4.10 Error 31 appears. .......................................................................................... 4-84.4.11 Error 50 appears. .......................................................................................... 4-84.4.12 Error 51 appears. .......................................................................................... 4-84.4.13 Error 52 appears. .......................................................................................... 4-94.4.14 Error 53 appears. ........................................................................................ 4-10

4.4.14.1 In the case of the hand switch ......................................................... 4-104.4.14.2 In the case of the foot switch .......................................................... 4-10

4.4.15 Error 54 appears. ........................................................................................ 4-114.4.16 Error 90 appears. ........................................................................................ 4-114.4.17 Error 91 appears. ........................................................................................ 4-114.4.18 Error 93 appears. ........................................................................................ 4-114.4.19 Error 94 appears. ........................................................................................ 4-11

4.5 The Illumination Light is not Projected. .................................................................. 4-124.6 The Focusing Rod Appears Moved when the PD is Changed. ................................ 4-12

Page

Page 4: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

4.7 The Illumination Light is not in the Center of the Visual Field when the Magnification is 12.5×. ............................................................................ 4-134.8 The Focusing Rod cannot be Seen Clearly. ............................................................. 4-134.9 There are Some Spots in the Illumination Light. ..................................................... 4-134.10 The Illumination Light does not Go Out. ............................................................... 4-134.11 The Base does not Move Smoothly. ...................................................................... 4-144.12 The Base does not Go Up or Down Smoothly. ...................................................... 4-144.13 The Aiming Beam is not Emitted. .......................................................................... 4-154.14 The READY Indicator does not Light Up. ............................................................. 4-154.15 The Burn Pattern is not Aligned with the Aiming Beam. ....................................... 4-164.16 The Burn Pattern is Uneven. .................................................................................. 4-164.17 The Number of Pulses does not Change. ............................................................... 4-174.18 The COUTNER Display does not Return to 0. ...................................................... 4-174.19 The Energy Output Measured with the YAG Energy Meter is Abnormal. .............. 4-184.20 The YC-1800 does not Stop with the Emergency Stop Switch. ............................. 4-18

§5 COVER REMOVEMENT ........................................................................................... 5-15.1 Right and Left Covers ............................................................................................... 5-15.2 Base Cover A............................................................................................................ 5-15.3 Lower Cover ............................................................................................................. 5-15.4 Vertical Movement Cover ......................................................................................... 5-25.5 Base Cover C ............................................................................................................ 5-35.6 Removing the Arm Covers ........................................................................................ 5-4

§6 PARTS REPLACEMENT ............................................................................................ 6-16.1 Boards ...................................................................................................................... 6-1

6.1.1 Master board ................................................................................................... 6-16.1.2 Relay board ..................................................................................................... 6-16.1.3 Power supply board ........................................................................................ 6-26.1.4 Hand switch board .......................................................................................... 6-2

6.2 Cable Assemblies ..................................................................................................... 6-36.2.1 YAG shutter ..................................................................................................... 6-36.2.2 Aiming shutter ................................................................................................. 6-46.2.3 Focus shift assy. .............................................................................................. 6-56.2.4 Energy monitor assy. ....................................................................................... 6-56.2.5 Energy motor assy. .......................................................................................... 6-66.2.6 Aiming LD assy. .............................................................................................. 6-66.2.7 Key switch assy. .............................................................................................. 6-76.2.8 Emergency switch assy. ................................................................................... 6-76.2.9 Illumination control assy. ................................................................................ 6-7

Page

Page 5: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

6.2.10 Base plate assy. ............................................................................................. 6-86.3 Fixation Lamp (Green) .............................................................................................. 6-86.4 Energy Adjustment Assy. ........................................................................................... 6-96.5 YAG Shutter Assy. ..................................................................................................... 6-96.6 Focus Assy. ............................................................................................................. 6-106.7 D Mirror Assy. ........................................................................................................ 6-106.8 LD Assy. ................................................................................................................. 6-106.9 Aiming Shutter Assy. ............................................................................................... 6-116.10 Control Board Assy. .............................................................................................. 6-116.11 Display Panel Assy. .............................................................................................. 6-126.12 Microscope ........................................................................................................... 6-126.13 Middle Gear ......................................................................................................... 6-136.14 Wheel Shaft ........................................................................................................... 6-136.15 Vertical Movement Screw..................................................................................... 6-146.16 Brake Lever .......................................................................................................... 6-146.17 Joystick ................................................................................................................. 6-146.18 Base Plate ............................................................................................................. 6-156.19 Head Rest ............................................................................................................. 6-156.20 Lamp House .......................................................................................................... 6-166.21 MYLA Laser Head ................................................................................................ 6-166.22 Slide Plate ............................................................................................................ 6-176.23 Rack ...................................................................................................................... 6-176.24 Pinion .................................................................................................................... 6-186.25 Gear A .................................................................................................................. 6-186.26 Gear B .................................................................................................................. 6-18

§7 ADJUSTMENT ............................................................................................................. 7-17.1 Microscope Optical Axis .......................................................................................... 7-1

7.1.1 Binocular microscope optical axis .................................................................. 7-17.1.2 Microscope visual field .................................................................................. 7-1

7.2 Illumination Rotation Center ..................................................................................... 7-27.3 Aiming Spots ............................................................................................................. 7-3

7.3.1 LD assy. ........................................................................................................... 7-37.3.2 Aiming spot focus ............................................................................................ 7-37.3.3 Aiming spot optical axis .................................................................................. 7-4

7.4 YAG Laser Optical Axis ........................................................................................... 7-57.4.1 Mirror holder 3 and YAG laser head ............................................................... 7-57.4.2 Focus assy. position ........................................................................................ 7-67.4.3 D mirror assy. .................................................................................................. 7-67.4.4 Focus adjustment ............................................................................................. 7-7

Page

Page 6: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

7.4.5 Focus shift indication ...................................................................................... 7-87.5 Electric Adjustments ................................................................................................. 7-9

7.5.1 YAG laser wiring check ................................................................................... 7-97.5.2 Master board initial setting .............................................................................. 7-97.5.3 Aiming beam power output ............................................................................ 7-107.5.4 Agreement between the indicated and actual PFN voltages ........................... 7-117.5.5 PFN voltage ................................................................................................... 7-11

7.6 YAG Laser Beam Energy Output ............................................................................. 7-127.6.1 Maximum energy output ................................................................................. 7-127.6.2 Minimum energy output .................................................................................. 7-137.6.3 Energy Output Display ................................................................................... 7-13

7.6.3.1 Display when the energy output is 10 mJ and 1 mJ ............................ 7-137.6.3.2 Energy Output display and preset data setting .................................... 7-14

7.7 Energy Output Check ............................................................................................... 7-157.8 Air Plasma Check ................................................................................................... 7-157.9 Parameter Setting .................................................................................................... 7-157.10 Service Mode Indication ....................................................................................... 7-157.11 Output Adjustment Mode ....................................................................................... 7-157.12 Vertical Movement Base ....................................................................................... 7-167.13 Smart Switch Adjustment ...................................................................................... 7-16

§8 INFORMATION ........................................................................................................... 8-18.1 Wiring Diagram ........................................................................................................ 8-18.2 Connector Cable ....................................................................................................... 8-28.3 Appearance ............................................................................................................... 8-98.4 Error Message Table .............................................................................................. 8-118.5 Dip Switch Setting Table ........................................................................................ 8-118.6 Parameter Setting Tables ........................................................................................ 8-12

Page

Page 7: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§1 INTRODUCTION

1 - 1

• This service manual is for the OPHTHALMIC YAG LASER SYSTEM YC-1800.

• Be sure to use this manual after understanding the operator’s manual for the OPHTHALMICYAG LASER SYSTEM YC-1800 thoroughly.

• Refer to the OPHTHALMIC YAG LASER SYSTEM YC-1800 Operator’s Manual and PartsList.

• Specifications are subject to change without notice for improvement. Refer to the TechnicalBulletin for important changes.

• In case the system can not be repaired according to the procedures described in this manual,please tell us the serial number of the system and details of the symptom or symptoms.

Page 8: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

1 - 2

Page 9: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§2 CAUTIONS

2 - 1

2.1 General

• Servicing should be performed by NIDEK service persons or other qualified personnel.

• Perform all servicing according to the procedures described in this manual. Failure to do somay cause unexpected accident or malfunction.

• Be sure to turn the power off during servicing unless there is an instruction to turn it on.

2.2 During Service

• Take countermeasures against static electricity before the service.

• If malfunction occurs, check the symptom and turn the power off.

• Do not lose or drop removed parts inside the YC-1800.Prepare a case beforehand to store loose parts.

• Take care not to bump the YC-1800.

• Use suitable tools to screw or unscrew the screws.

• Apply a threadlocking adhesive to the screws which have been loosened and tightened again.

• If the YC-1800 produces smoke or strange odors, immediately turn off the system and disconnectthe electrical plug from the outlet.Usage of the system under such abnormal conditions may cause fire or electric shock.

• Do not perform servicing with wet hands.Doing so could cause electric shock or malfunction.

• In §3 TROUBLESHOOTING, when checking a break in a cable, also check the followingaccording to the wiring diagram and connector cable.1. Proper connections and crimping2. Proper contact by connecting and disconnecting connectors3. Proper soldering* Do not yank the cables with excessive force. Doing so could cause breakage.

2.3 During Laser Emission

• Never emit the laser beam directly on personnel or things.

Page 10: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

2 - 2

2.4 Handling the Laser Head

• Never disassemble or repair the laser head.Return the laser head to NIDEK if the repair is necessary. NIDEK will not be responsible foraccidents or malfunction cause by the YC-1800 repaired by other than NIDEK.

2.5 Adjustment

• Adjust the YC-1800 on a vibration free, level, and stable surface.Failure to do so could prevent adjustment from being performed correctly.

• Do not use adjustment jigs for other than the intended purpose.

• Since the Nd:YAG laser used in the YC-1800 is an invisible infrared ray, improper operationmay cause the hazardous exposure to radiation.

• Never face to the laser beam aperture in the laser beam path.

• Wear the safety goggles specified for the Nd:YAG laser.

• Never gaze directly the Nd:YAG laser beam even when wearing the safety goggles.

• Do not touch optical parts such as mirror or lens with bare hands.

• Never expose the laser beam to your skin.

• Take care not to receive electric shock when opening the cover to check or repair the electricalsystem.

• The YAG laser uses the high voltage.

2.6 Using the Energy Meter for the YAG laser

• Depending on the energy meter to be used, the necessity of the dispersion lens, which is attachedto the objective lens at the time of energy measurement, is varied.Refer to the table below when measuring the energy.

• The detector of the energy meter may be damaged if the excessive energy density is applied tothe photoreceptor of the detector.Maximize the spot size of the aiming beam on the photoreceptor as possible so that the YAGlaser beam never come into focus on the photoreceptor.

Energy meter Dispersion lens

MODEL-365 Necessary

MODEL PHDX50 Unnecessary

Page 11: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§3 TROUBLESHOOTING

3 - 1

Plug the power cord to the wall outlet.

Does the pilot lamp light up?

Insert the key and turn it to the ON position.

Do the fixation lamp and backlight panel light up?

Does an error display appear?

Remove the hole cap and insert the focusing rod.Is the illumination light projected on the rod?

Adjust the diopter of the eyepieces.

Adjust the pupil distance (PD) of the microscope.

Is the optical system of the microscope normal?

4.1 The pilot lamp does not light up.

4.4 The error display appears.

4.5 The illumination light is not projected.

4.6 The focusing rod appears moved when the PD ischanged.

4.7 The illumination light is not in the center of thevisual field when the magnification is 12.5x.

4.8 The focusing rod can not be seen clearly.

Yes

Yes

Yes

No

Yes

No

Yes

No

No

Is the illumination condition normal? 4.9 There are some spots in the illumination light.

Yes

No

4.2 Only the fixation lamp does not light up.4.3 Only the backlight panel does not light up.

No

Page 12: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

3 - 2

Does the slit lamp function normally?

Reduce the light amount of the slit lamp.

Is the aiming beam visible?

Does pressing the STATUS switch turn off theSTANDBY indicator and turn on the READYindicator?

Does pressing the trigger switch emit the YAG laserbeam normally?

Does pressing the burst switch change the number ofpulses?

Does the pressing the reset switch resets theCOUNTER display to 0?

Is the emitted YAG laser beam energy normal?

4.11 The base does not move smoothly.4.12 The base does not go up or down smoothly.

4.14 The READY indicator does not light up.

4.4 The error display appears.4.15 The burn pattern is not aligned with the aiming

beam.4.16 The burn pattern is uneven.

4.17 The number of pulses does not change.

4.18 The COUNTER display does not return to 0.

4.19 The energy output measured with the YAGenergy meter is abnormal.

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

No

No

4.13 The aiming beam is not emitted.No

Does turning fully the i l lumination controlcounterclockwise turn off the illumination light?

No4.10 The illumination light does not go out.

Page 13: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

3 - 3

End

Does the pressing the emergency stop switch stopthe YC-1800?

4.20 The YC-1800 does not stop with the emergencystop switch.

Yes

No

Turn the key switch to the OFF position and removeit.

Page 14: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

3 - 4

Page 15: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§4 SUB TROUBLESHOOTING

4 - 1

4.1 The Pilot Lamp does not Light Up.

Secured the required voltage.Is the required voltage applied to the wall outlet?

Is the voltage between the P10 (J10) 1st and 2nd pins onthe switching power supply (13704-E020) DC + 12 V?

Is the voltage between the J20 1st and 3rd and 2nd and 4thpins on the base plate assy. (13706-EA17) DC + 12 V?

Is the voltage between the P401 (J401) 1st and 3rd and2nd and 4th pins on the relay board (13706-BA04) DC+ 12 V?

Is the voltage between the J403 4th and 5th pins on therelay board DC + 2.0 V?

Replace the key switch assy. (13706-EA11).(Refer to 6.2.7.)

Replace the switching power supply.

Replace the base plate assy.(Refer to 6.2.10.)

Replace the relay board.(Refer to 6.1.2.)

Yes

Yes

Yes

Yes

No

Yes

No

No

No

Replace the cable assy. (13706-EA16).No

Page 16: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 2

4.2 Only the Fixation Lamp does not Light Up.

4.3 Only the Backlight Panel does not Light Up.

Replace the master board.(Refer to 6.1.1.)

Is the voltage between the J101 16th and 18th and 17th and19th pins on the master board (13706-BA01) DC + 5 V?

Is there a break in the flexible flat cable (13706-E020)?

Replace the display panel assy. (13706-1400).(Refer to 6.11.)

Replace the flexible flat cable.

No

No

Yes

Yes

Replace the master board.(Refer to 6.1.1.)

Is the voltage between the P113 (J113) 10th and 13thpins on the master board (13706-BA01) DC + 12 V?

Is the voltage between the P20 12th and 15th pins onthe cable assy. (13706-EA16) DC + 12 V?

Is the voltage between the J40 1st and 2nd pins on thebase plate assy. (13706-EA17) DC + 12 V?

Replace the green fixation lamp (34405-0000).(Refer to 6.3.)

Replace or repair the cable assy.

Replace the base plate assy.(Refer to 6.2.10.)

Yes

Yes

No

Yes

No

No

Are there any breaks between the following connectorsin the head rest assy. (13706-EA14)?

P40 - JP11st pin - 1st pin (resistance: about 1.2 KΩ)2nd pin - 2nd pin (continuity)

Repair the head rest assy. or replace the head rest(13706-4100). (Refer to 6.19.)

No

Yes

Page 17: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 3

4.4 The Error Display Appears.

4.4.1 Error 1 appears.

4.4.2 Error 5 appears.

Turn the key switch to the OFF position and removethe right and left covers. Let the YC-1800 sit for 30minutes to cool it down.

Is the voltage between the J105 1st pin and GND onthe master board (13706-BA01) DC + 12 V when thekey switch is turned to the ON position.?

Return the YC-1800 to the factory.

Replace the master board.(Refer to 6.1.1.)

Yes

No

Replace the master board. (Refer to 6.1.1.)Is the voltage between the P103 (J103) 1st and 2ndpins on the master board (13706-BA01) DC + 12 Vwhen the YC-1800 is in READY?

Does the solenoid of the YAG shutter function whenthe YC-1800 is in READY?

Is the voltage between the P103 (J103) 3rd and 7th and4th and 7th pins on the master board DC + 0 V or 5 V?

Is the voltage of P103 (J103) on the master board asfollows when the YAG shutter functions?Shutter opened: 0 V between 5th and 7th pins, 5 V

between 6th and 7th pinsShutter closed: 5 V between 5th and 7th pins, 0 V

between 6th and 7th pins

Replace the master board. (Refer to 6.1.1.)

Replace the YAG shutter assy. (13706-EA03).(Refer to 6.2.1.)

Replace the YAG shutter assy. (Refer to 6.2.1.)

Yes

Yes

No

No

Yes

No

Yes

Replace or adjust the YAG shutter assy.(Refer to 6.2.1.)

No

Page 18: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 4

4.4.3 Error 7 appears.

Replace the master board. (Refer to 6.1.1.)Is the voltage between the P104 (J104) 1st and 2ndpins on the master board (13706-BA01) DC + 12 Vwhen the YC-1800 is in READY?

Does the solenoid of the aiming shutter function whenthe YC-1800 is in READY?

Is the voltage between the P104 (J104) 3rd and 5th and6th and 5th pins on the master board DC + 0 V or 5 V?

Is the voltage of P104 (J104) on the master board asfollows when the aiming shutter functions?Shutter opened: 0 V between 4th and 5th pins, 5 V

between 7th and 5th pinsShutter closed: 5 V between 4th and 5th pins, 0 V

between 7th and 5th pins

Replace the master board. (Refer to 6.1.1.)

Replace the aiming shutter (13706-EA04).(Refer to 6.2.2.)

Replace the aiming shutter. (Refer to 6.2.2.)

Yes

Yes

No

No

Yes

No

Yes

Replace or adjust the aiming shutter.(Refer to 6.2.2.)

No

Page 19: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 5

4.4.4 Error 10 appears.

4.4.5 Error 12 appears.

Replace the power supply board (13733-BA01).(Refer to 6.1.3.)

4.4.6 Error 13 appears.

Use the YC-1800 away from the instruments which may make noises.

Set DIP SW1-7 on the display board (13704-BA02)to the OFF (upper) position. The number of pulsesappears on the COUNTER display.

Emit ten shots of the YAG laser beam.

Is 4 (number of pulses) displayed on the COUNTERdisplay in eight or more of ten shots?

Adjust the PFN voltage. (Refer to 7.5.4 and 7.5.5.)

Does Error10 occur?

End

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Replace the MYLA laser head (13739-0000).(Refer to 6.21.)

No

No

Yes

Yes

Page 20: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 6

4.4.7 Error 14 appears.

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Does the flash lamp light up in the trigger switch onor test fire?

Is the PFN voltage normal?

Is there a break in the YL board assy. (13706-EA05)?

Is the TP3 (TRIG OUT) signal of the power supplyboard (13733-BA01) normal when the trigger is on?

Is there a break in the YL cable assy. (13706-EA06)?

Replace the power supply board.(Refer to 6.1.3.)

Is the error resolved?

End

Adjust the PFN voltage. (Refer to 7.5.4 and 7.5.5.)

Replace the YL board assy.

Replace the YL cable assy.

Replace the MYLA laser head (13739-0000).(Refer to 6.21.)

Yes

Yes

No

Yes

No

Yes

No

No

Yes

Yes

No

Replace the master board. (Refer to 6.1.1.)No

Page 21: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 7

4.4.8 Error 15 appears.

4.4.9 Error 30 appears.

Replace the energy monitor assy. (13706-EA08).(Refer to 6.2.4.)

End

Replace the master board (13706-BA01).(Refer to 6.1.1.)

No

Yes

Is the error resolved?

Replace the YL board assy.Is there a break in the YL board assy (13706-EA05)?

Is the voltage between the P107 (J107) 6th and 7thpins on the master board (13706-BA01) 0 V whenthe trigger switch is pressed?

Is the voltage between TP5 and GND on the powersupply board (13733-BA01) DC + 12 V when thetrigger switch is pressed?

Is the voltage between TP4 and GND on the powersupply board DC + 5 V when the trigger switch ispressed?

Replace the master board. (Refer to 6.1.1.)

Replace the master board. (Refer to 6.1.1.)

Replace the power supply board.(Refer to 6.1.3.)

Yes

Yes

Yes

Yes

No

No

No

Replace the power supply board.(Refer to 6.1.3.)

No

Page 22: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 8

4.4.10 Error 31 appears.

4.4.11 Error 50 appears.

4.4.12 Error 51 appears.

Replace the energy monitor assy. (13706-EA08).(Refer to 6.2.4.)

Is the error resolved?

End

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Yes

No

Replace the YL board assy.Is there a break in the YL board assy. (13706-EA05)?

Compare the PFN voltage setting value of Parameter10 with the PFN voltage setting value (between theJ2 1st (+) and 5th (-) pins on the power supplyboard). Is the difference within 20 V?Be aware that the voltage becomes 440 ± 10 V atmaximum.

Replace the power supply board (13733-BA01).(Refer to 6.1.3.)

Replace the master board (13706-BA01).(Refer to 6.1.1.)

No

Yes

No

Yes

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Is the TP3 voltage on the master board between 0.2V (min) and 4.5 V (max) when the energy control isturned?

Replace the energy monitor assy. (13706-EA08).(Refer to 6.2.4.)

Yes

No

Page 23: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 9

4.4.13 Error 52 appears.

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Is the TP4 voltage on the master board betweenabout 0.5 V (ANT500) and 4.5 V (POST500) whenthe focus shift control is turned?

Replace the focus shift assy. (13706-EA07).(Refer to 6.2.3.)

Yes

No

Page 24: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 10

4.4.14 Error 53 appears.

4.4.14.1 In the case of the hand switch

4.4.14.2 In the case of the foot switch

Replace the master board. (Refer to 6.1.1.)Is the voltage between the P102 (J102) 1st pin andGND on the master board (13706-BA01) DC + 12 V?

Is there a break in the hand switch relay (13706-EA01)?

Is the voltage between P102 (J102) 2nd (or 3rd) pinand GND on the master board DC + 12 V when thehand switch is pressed?

Replace the master board. (Refer to 6.1.1.)

Replace the hand switch relay.

Replace the hand switch board (13706-BA07).(Refer to 6.1.4.)

No

Yes

No

Yes

Yes

No

Replace the foot switch (FB-10).Is there continuity between the 1st and 2nd and 2ndand 3rd pins of the foot switch connector when thefoot switch is on?

Is there a break in the base plate assy. (13706-EA17)?

Is there a break in the cable assy. (13706-EA16)?

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Replace the base plate assy. (Refer to 6.2.10.)

Replace the cable assy.

No

No

No

Yes

Yes

Yes

Page 25: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 11

4.4.15 Error 54 appears.

4.4.16 Error 90 appears.

Replace the master board (13706-BA01).(Refer to 6.1.1.)

4.4.17 Error 91 appears.

Cancel the error and enter the data again.

4.4.18 Error 93 appears.

Cancel the error and enter the data again.

4.4.19 Error 94 appears.

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Replace the energy monitor assy. (Refer to 6.2.4.)Is there a break in the energy monitor assy. (13706-EA08)?

Does the waveform between the P110 (J110) 6th pinand GND on the master board (13706-BA01) changein order of Low, High, and Low when the handswitch is pressed? (High: DC + 5 V, Low: DC 0 V)

Replace the energy monitor assy. (Refer to 6.2.4.)

Replace the master board. (Refer to 6.1.1.)

Yes

Yes

No

No

Page 26: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 12

4.5 The Illumination Light is not Projected.

4.6 The Focusing Rod Appears Moved when the PD is Changed.

Change the pupil distance while looking through the binocular tube for each eye.When the illumination light shifts from the center of the visual field through the eyepiece, adjustthe optical axis of the binocular microscope. (See 7.1.1.)

EndIs the illumination light projected by turning theillumination control clockwise?

Is the illumination lamp (34530-E003) blown?

Is the voltage between the P405 (J405) 3rd and 4thpins on the relay board (13706-BA04) DC 0 V whenthe illumination control is turned?

Disconnect P405 (J405) on the relay board. Turn theillumination control. At this time, does the resistancebetween the P405 1st and 2nd pins on theillumination control assy. change?

Does the voltage between TP1 and GND on therelay board change in the range from DC 0 V to 6 Vwhen the illumination control is turned?

Is there a poor contact in the contact pins of P406 inthe lamp house (13706-5100)?

Replace the lamp house. (Refer to 6.20.)

Replace the illumination lamp.

Replace the illumination control assy. (13706-EA13). (Refer to 6.2.9.)

Replace the relay board. (Refer to 6.1.2.)

Repair P406 in the lamp house.

No

No

Yes

Yes

Yes

Yes

No

Yes

No

No

Yes

Replace the illumination control assy.(Refer to 6.2.9.)

No

Page 27: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 13

4.7 The Illumination Light is not in the Center of the Visual Field

when the Magnification is 12.5×.

Adjust the optical axis of the microscope. (Refer to 7.1.)

4.8 The Focusing Rod cannot be Seen Clearly.

4.9 There are Some Spots in the Illumination Light.

Return the YC-1800 to the factory.

4.10 The Illumination Light does not Go Out.

Wipe the objective lenses and eyepieces with themixed solvent of alcohol.

Can the focus be adjusted by turning the diopteradjustment of eyepiece for each eye?

Replace the eyepieces.

Yes

No

Replace the illumination control assy. (13706-EA13). (Refer to 6.2.9.)

Is the voltage between the P405 (J405) 3rd and 4th pinson the relay board (13706-BA04) DC + 12 V when theillumination control is turned fully counterclockwise?

Replace the relay board. (Refer to 6.1.2.)

No

Yes

Page 28: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 14

4.11 The Base does not Move Smoothly.

4.12 The Base does not Go Up or Down Smoothly.

Clean the slide plate.Is the slide plate (13704-M373) dirty?

Is the slide plate broken?

Is the rack (34348-M009) broken?

Is the pinion (32903-M138) broken?

Replace the wheel shaft (13706-3120).(Refer to 6.14.)

Replace the slide plate. (Refer to 6.22.)

Replace the rack. (Refer to 6.23.)

No

No

No

Yes

No

Yes

Yes

Replace the pinion. (Refer to 6.24.)Yes

Replace gear A. (Refer to 6.25.)Is gear A (13704-M304) broken?

Is the middle gear (13704-M320) in the middle gearassy. (13704-3020) broken?

Is gear B (13706-M312) in the vertical movementscrew (13706-3130) broken?

Adjust the vertical movement base.(Refer to 7.12.)

Replace the middle gear. (Refer to 6.13.)

Replace gear B. (Refer to 6.26.)

No

No

Yes

No

Yes

Yes

Page 29: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 15

4.13 The Aiming Beam is not Emitted.

4.14 The READY Indicator does not Light Up.

Pressing the aiming switch lights up the aimingbeam.

Does the aiming off indicator light up?

Can the voltage between TP12 and GND on themaster board (13706-BA01) adjusted in the rangefrom DC 0 V to 1.2 V?

It the aiming beam emitted when the voltagebetween TP12 and GND on the master board?

Adjust the aiming beam power output.(Refer to 7.5.3.)

Replace the master board. (Refer to 6.1.1.)

Replace the aiming LD assy. (13706-EA10).(Refer to 6.2.6.)

Yes

Yes

Yes

No

No

No

Pressing the aiming switch lights up the aimingbeam.

Does the aiming off indicator light up?

Wipe the side without the blue mark of the flexibleflat cable (13706-E020) with alcohol.

Is the trouble resolved?

Replace the display panel assy. (13706-1400).(Refer to 6.11.)

Is the trouble resolved?

END

END

Replace the master board (13706-BA01).(Refer to 6.1.1.)

Yes

No

Yes

No

Yes

No

Page 30: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 16

4.15 The Burn Pattern is not Aligned with the Aiming Beam.

Adjust the optical axis of the YAG laser beam. (Refer to 7.4.)

4.16 The Burn Pattern is Uneven.

Be sure to tape up a burn paper so that the optical system is not grimed with soot.

Blow off the dirt on the beam expander with theblower.

Is the burn pattern uneven at the front of the beamexpander (expander assy.)? (D in the figure below)

Is the burn pattern uneven at the front of the dichroicmirror (D mirror assy.)? (C in the figure below)

Is the burn pattern uneven at the front of the mirror(mirror holder 3)? (B in the figure below)

Is the burn pattern uneven between the MYLA laserhead and energy control assy? (A in the figurebelow)

Replace the power supply board and MYLA laserhead unit. (Refer to 6.1.3 and 6.21.)

Blow off the dirt on the dichroic mirror with theblower.

Blow off the dirt on the mirror with the blower.

Yes

Yes

Yes

No

Yes

No

No

Blow off the dirt on the energy control assy. with theblower.

No

Page 31: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 17

4.18 The COUTNER Display does not Return to 0.

4.17 The Number of Pulses does not Change.

Wipe the side without the blue mark of the flexibleflat cable (13706-E020) with alcohol.

Can the number of pulses set?

Replace the display panel assy. (13706-1400).(Refer to 6.11.)

Can the number of pulses set?

End

End

No

Yes

Yes

Replace the master board (13706-BA01).(Refer to 6.1.1.)

No

Wipe the side without the blue mark of the flexibleflat cable (13706-E020) with alcohol.

Can the number on the COUNTER display become 0?

Replace the display panel assy. (13706-1400).(Refer to 6.11.)

Can the number on the COUNTER display become 0?

End

End

No

Yes

Yes

Replace the master board (13706-BA01).(Refer to 6.1.1.)

No

Page 32: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

4 - 18

4.19 The Energy Output Measured with the YAG Energy Meter is

Abnormal.

4.20 The YC-1800 does not Stop with the Emergency Stop Switch.

Adjust the YAG laser energy output.(Refer to 7.6.)

Is the trouble resolved?

End

Replace the power supply board (13733-BA01) andMYLA laser head. (Refer to 6.1.3 and 6.21.)

Yes

No

Replace the emergency switch assy.(Refer to 6.2.8.)

Disconnect P404 (J4) on the relay board (13706-BA04).

Replace the relay board. (Refer to 6.1.2.)

No

Yes

Does the continuity status between the P404 1st and2nd pins on the emergency switch assy. (13706-EA12) become as follows?When it is pressed: No continuityWhen it is released: Continuity

Page 33: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§5 COVER REMOVEMENT

5 - 1

5.1 Right and Left Covers

1. Loosen SB3 × 53 (shown with the arrows inthe right figure) with a hexagonal wrench.

2. Remove the left cover (13706-M004).

3. Remove the right cover (13706-M003) in thesame manner.

5.2 Base Cover A

1. Unscrew CK2.8 × 62.

2. Remove base cover A (13706-M321).

5.3 Lower Cover

1. Unscrew CK2.6 × 42.

2. Remove the lower cover (13706-M007).

Page 34: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

5 - 2

5.4 Vertical Movement Cover

1. Remove base cover A (13706-M321). (Referto 5.2.)

2. Unscrew SB3 × 62 and rotate base cover C(13706-M323) together with the cover mount(13706-M303).

3. Pull the brake lever to the operator’s side.

4. Unscrew CK2 × 42 to remove the verticalmovement cover (13706-M332).

Page 35: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

5 - 3

5.5 Base Cover C

1. Remove base cover A (13706-M321). (Referto 5.2.)

2. Remove the vertical movement cover (13706-M332). (Refer to 5.4.)

3. Remove P404 (J4) and P405 (J5) on the relayboard (13706-BA04).

4. Remove base cover C (13706-M323)together with the cover mount (13706-M303).

5. Remove the cover (13706-M318).

6. Remove the control cap (82046-0703D).

7. Remove the control (82045-0912B).

8. Unscrew PT2 × 52 to remove the cover mount.

9. Remove the emergency switch assy. (13706-EA12).

Page 36: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

5 - 4

5.6 Removing the Arm Covers

1. Rotate the lamp house (13706-5100).

2. Unscrew CK2 × 52 to remove the left armcover (13706-M005) and right one (13706-M006).

Page 37: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§6 PARTS REPLACEMENT

6 - 1

6.1 Boards6.1.1 Master board

Replacement part: 13706-BA011. Remove the right cover (13706-M003).

(Refer to 5.1.)2. Disconnect all connectors from the master

board (13706-BA01).3. Unscrew PC3 × 84 to remove the master

board.4. Remove the spacers (13611-M076).5. Replace the master board.6. Reassemble all parts in reverse order.

Mount the EEPROM attached to the old masterboard on the new one.

7. Adjust the following.1) Adjust the aiming beam power output. (Refer

to 7.5.3.)2) Adjust the PFN voltage. (Refer to 7.5.5.)3) Adjust the YAG laser beam energy output.

(Refer to 7.6.)

6.1.2 Relay board

Replacement part: 13706-BA041. Remove the vertical movement cover. (Refer

to 5.4.)2. Disconnect all connectors from the relay

board (13706-BA04).3. Unscrew PC3 × 5 to remove TR1 on the relay

board.4. Unscrew PC3 × 54 to remove the relay board.5. Replace the relay board.6. Reassemble all parts in reverse order.

Page 38: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 2

6.1.3 Power supply board

Replacement part: 13733-BA011. Remove the arm covers. (Refer to 5.6.)2. Disconnect all connectors from the power

supply board (13733-BA01).3. Unscrew PC3 × 64 to remove the power supply

board.4. Replace the power supply board.5. Reassemble all parts in reverse order.6. Adjust the actual PFN voltage. (Refer to

7.5.4.)

6.1.4 Hand switch board

Replacement part: 13706-BA071. Turn the top of the grip (13706-M360).2. Loosen HH3 × 3, rotate the grip (13706-

M354), and unscrew FK2 × 43.3. Remove the trigger switch (13706-3210).

4. Unscrew CK2 × 123 to remove the hand switchboard (13706-BA07).

5. Remove the cables soldered to the handswitch board with a soldering iron.

6. Replace the hand switch board.7. Reassemble all parts in reverse order.8. Adjust the smart switch. (Refer to 7.13.)

Page 39: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 3

6.2 Cable Assemblies6.2.1 YAG shutter

Replacement part: 13706-EA031. Unscrew SB3 × 5 to remove the YAG shutter

assy. (13706-1120). (Refer to 6.5.)2. Unscrew HH3 × 32 to remove the shutter plate

(13706-M116).3. Unscrew 1N2.62 and SW2.62 to remove the

YAG shutter (13706-EA03).4. Unscrew CK2 × 32 to remove the YAG shutter.5. Replace the YAG shutter.6. Reassemble all parts in reverse order.

7. After reassembling the YAG shutter, adjustits sensor position so that the shutter plate(13706-M116) is as shown in the right figures.

When the shutter closes When the shutter opens

ShieldedShieldedNot shielded Not shielded

Page 40: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 4

6.2.2 Aiming shutter

Replacement part: 13706-EA041. Unscrew SB3 × 52 to remove the aiming

shutter assy. (13706-1220). (Refer to 6.9.)2. Uncrew PC2 × 32 to remove the sensor

together with the board mounting plate(13706-M126).

3. Unscrew CK2 × 32 to remove the sensor ofthe aiming shutter (13706-EA04).

4. Loosen HH3 × 3 to remove the shaft (13706-M124).

5. Unscrew CK2 × 32 and SW22 to remove thesolenoid of the aiming shutter.

6. Replace the aiming shutter.7. Reassemble all parts in reverse order.

8. After reassembling the aiming shutter, adjustits sensor and solenoid positions so that theshutter plate (13706-M125) is as shown in theright figures.

Shielded

Not shielded Shielded

Not shielded

Page 41: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 5

6.2.3 Focus shift assy.

Replacement part: 13706-EA071. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Disconnect P109 (J9) from the master board

(13706-BA01).3. Unscrew HH3 × 32 and nut attached to 13706-

EA07 to remove the focus shift assy. (13706-EA07).

4. Replace the focus shift assy.5. Reassemble all parts in reverse order.6. Adjust the focus shift indication. (Refer to

7.4.5.)

6.2.4 Energy monitor assy.

Replacement part: 13706-EA081. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Unscrew SB2 × 4 2 t o r e m o v e t h e

potentiometer of the energy monitor assy.(13706-EA08) together with the bearing(13706-M103).

3. Unscrew HH3 × 3 to remove the screw gear(13702-M063).

4. Remove the attached nut and potentiometerof the energy monitor assy.

5. Unscrew CK2 × 42 to remove the energymonitor assy.

6. Replace the the energy monitor assy.7. Reassemble all parts in reverse order.8. Set the energy output display and preset data.

(Refer to 7.6.3.2.)

Page 42: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 6

6.2.5 Energy motor assy.

Replacement part: 13706-EA091. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Disconnect P111 (J111) on the master board

(13706-BA01).3. Unscrew SB2 × 82 to remove the energy

motor assy. (13706-EA09) together with thespacer collar (13706-M105).

4. Loosen HH2 × 3 to remove the screw gear(13706-M110) together with the shaft (13706-M109).

5. Replace the energy motor assy.6. Reassemble all parts in reverse order.7. Set the energy output display and preset data.

(Refer to 7.6.3.2.)

6.2.6 Aiming LD assy.

Replacement part: 13706-EA101. Remove the LD assy. (13706-1210). (Refer

to 6.8.)2. Unscrew CK2 × 62 to remove the LD clamp

plate (13706-M121).3. Disconnect the pin of P112 on the aiming LD

assy. (13706-EA10) from the connectorhousing.

4. Remove the O ring (82001-RG003) andaiming LD assy. from the LD holder (13706-M120).

5. Replace the aiming LD assy.6. Reassemble all parts in reverse order.7. Adjust the aiming spots. (Refer to 7.3.)

Page 43: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 7

6.2.7 Key switch assy.

Replacement part: 13706-EA111. Remove the vertical movement cover (13706-

M332). (Refer to 5.4.)2. Disconnect P403 (J3) on the relay board

(13706-BA04).3. Unscrew PC3 × 53 to remove the key switch

assy. (13706-EA11).4. Replace the key switch assy.5. Reassemble all parts in reverse order.

6.2.8 Emergency switch assy.

Replacement part: 13706-EA121. Remove the vertical movement cover (13706-

M332). (Refer to 5.4.)2. Disconnect P404 (J4) on the relay board

(13706-BA04).3. Replace the emergency switch assy. (13706-

EA12).4. Reassemble all parts in reverse order.

6.2.9 Illumination control assy.

Replacement part: 13706-EA131. Remove base cover C (13706-M323). (Refer

to 5.5.)2. Remove the control cap (82046-0703D).3. Remove the control (82045-0912B).4. Unscrew PT2 × 52 to remove the cover mount

(13706-M303). Unscrew PT2 × 62.5. Remove the nut to remove the illumination

control assy. (13706-EA13).6. Replace the illumination control assy.7. Reassemble all parts in reverse order.

Page 44: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 8

6.2.10 Base plate assy.

Replacement part: 13706-EA171. Remove the head rest (13706-4100). (Refer

to 6.19.)2. Disconnect P20 on the cable assy. (13706-

EA16).3. Remove the grip (13702-M412).4. Unscrew SB4 × 62 to remove the base plate

assy. (13706-EA17).5. Replace the base plate assy.6. Reassemble all parts in reverse order.

6.3 Fixation Lamp (Green)

Replacement part: 34405-00001. Remove and replace the green fixation lamp

(34405-0000).2. Reassemble all parts in reverse order.

Page 45: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 9

6.4 Energy Adjustment Assy.

Replacement part: 13706-11101. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Remove the right and left caps (82046-1705A)

and two controls (82045-2118B).3. Unscrew HH3 × 32 to remove two stopper B

(13706-M016).4. Unscrew SB3 × 63 to remove the energy

adjustment assy. (13706-1101).5. Replace the energy adjustment assy.6. Reassemble all parts in reverse order.7. Adjust the YAG laser beam energy output.

(Refer to 7.6.)

6.5 YAG Shutter Assy.

Replacement part: 13706-11201. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Remove the control cap (82046-1705A) and

control (82045-2118B).3. Unscrew HH3 × 3 to remove stopper B

(13706-M016).4. Disconnnect P111 (J111) on the master board

(13706-BA01).5. Unscrew SB2 × 82 to remove the energy

motor assy. (13706-EA09) together with thespacer collar (13706-M105).

6. Unscrew CK2 × 42 to remove the sampleboard (13706-BA03).

7. Disconnnect P103 (J103) on the master board.8. Unscrew SB3 × 5 to remove the YAG shutter

assy. (13706-1120).9. Replace the the YAG shutter assy.

10. Reassemble all parts in reverse order.11. Set the energy output display and preset data.

(Refer to 7.6.3.2.)

Page 46: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 10

6.6 Focus Assy.

Replacement part: 13706-11301. Remove the right cover (13706-M003) and left cover (13706-M004). (Refer to 5.1.)2. Remove the energy ajustment assy. (13706-1110). (Refer to 6.4.)3. Remove the control cap (82046-1705A) and

control (82045-2118A).4. Unscrew HH3 × 3 to remove stopper B

(13706-M016).5. Disconnnect P111 (J111) on the master board

(13706-BA01).6. Adjust the focus assy. position. (Refer to

7.4.2.)7. Adjust the D mirror assy. (Refer to 7.4.3.)8. Adjust the focus. (Refer to 7.4.4.)9. Adjust the focus shist indication. (Refer to

7.4.5.)

6.7 D Mirror Assy.

Replacement part: 13704-11401. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Unscrew PC3 × 63 and PW33 to remove the

D mirror assy. (13704-1140).3. Replace the D mirror assy.4. Reassemble all parts in reverse order.5. Adjust the D mirror assy. (Refer to 7.4.3.)

6.8 LD Assy.

Replacement part: 13706-12101. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Remove the lower cover. (Refer to 5.3.)3. Unscrew SB3 × 82 and remove the etwo

ccentric shafts (13611-M117) and LD assy.(13706-1210).

4. Replace the LD assy.5. Reassemble all parts in reverse order.6. Adjust the aiming spots. (Refer to 7.3.)

Page 47: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 11

6.9 Aiming Shutter Assy.

Replacement part: 13706-12201. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Remove the lower cover (13706-M007).

(Refer to 5.3.)3. Disconnect P104 (J4) on the master board

(13706-BA01).4. Unscrew SB3 × 52 to remove the aiming

shutter assy. (13706-1220).5. Replace the aiming shutter assy.6. Reassemble all parts in reverse order.7. Adjust the following.

1) When the shutter opens: The shutter plateshall not shield the aiming beam.

2) When the shutter closes: The shutter plateshall shield the aiming beam.

6.10 Control Board Assy.

Replacement part: 13706-13001. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Disconnect all connectors from the master

board (13706-BA01).3. Unscrew PC3 × 84 and PC3 × 54 to remove

the control board assy. (13706-1300).4. Replace the control board assy.

Mount the EEPROM attached to the old masterboard on the new one.

5. Reassemble all parts in reverse order.

Page 48: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 12

6.11 Display Panel Assy.

Replacement part: 13706-14001. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Disconnect J201 from the display board

(13704-BA02).3. Unscrew SB3 × 6 and SB3 × 122 to remove

the display panel assy. (13706-1400).4. Replace the display panel assy.5. Reassemble all parts in reverse order.

6.12 Microscope

Replacement part: 13706-20001. Remove the right cover (13706-M003) and

left cover (13706-M004). (Refer to 5.1.)2. Unscrew SB4 × 122 to remove the microscope

(13706-2000).3. Replace the microscope.4. Reassemble all parts in reverse order.5. Adjust themicroscope visual field. (Refer to

7.1.2.)

Page 49: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 13

6.13 Middle Gear

Replacement part: 13704-30201. Remove base cover C (13706-M323). (Refer

to 5.5.)2. Unscrew SB3 × 62 to rotate base cover C

together with the cover mount (13706-M303). (Refer to 5.4.)

3. Unscrew SB3 × 64 and 1W34 to remove themiddle gear (13704-3020).

4. Replace the middle gear.5. Reassemble all parts in reverse order.

6.14 Wheel Shaft

Replacement part: 13706-31201. Remove the base plate (13706-4000). (Refer

to 6.18.)2. Remove base cover A (13706-M321). (Refer

to 5.2.)3. Remove the two pinions (32903-M138).

(Refer to 6.24.)4. Uncrew FK2 × 42 to remove the spacer

(34134-M038).5. Remove the shaft (13706-M337).6. Unscrew SB3 × 44 and the fixing screw

(34322-M043) to remove base cover B(13706-M322).

7. Remove the axle holder (13706-M302).8. Loosen HH3 × 44 to remove the wheel shaft

(13706-3120).9. Replace the wheel shaft.

10. Reassemble all parts in reverse order.

Page 50: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 14

6.15 Vertical Movement Screw

Replacement part: 13706-31301. Remove the brake lever (13706-3140). (Refer

to 6.16.)2. Unscrew SB3 × 42 to remove the vertical

movement screw (13706-3130).3. Replace the vertical movement screw.4. Reassemble all parts in reverse order.

6.16 Brake Lever

Replacement part: 13706-31401. Remove base cover C (13706-M323). (Refer

to 5.5.)2. Unscrew SB3 × 62 to rotate base cover C

together with the cover mount (13706-M303). (Refer to 5.4.)

3. Remove the two button caps with flange(17162-M392) and brake lever (13706-3140).

4. Replace the brake lever.5. Reassemble all parts in reverse order.

6.17 Joystick

Replacement part: 13706-32001. Remove base cover C (13706-M323). (Refer

to 5.5.)2. Disconnect P30 from the hand switch assy.

(13706-EA02).3. Remove the spring compressor (13706-

M305) and joystick (13706-3200).4. Replace the joystick.5. Reassemble all parts in reverse order.

Page 51: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 15

6.18 Base Plate

Replacement part: 13706-40001. Unscrew PC3 × 64 to remove the two wheel

covers (13706-M408).2. Remove and replace the base plate (13706-

4000).3. Reassemble all parts in reverse order.

6.19 Head Rest

Replacement part: 13706-41001. Remove the green fixation lamp (34405-

0000). (Refer to 6.3.)2. Unscrew PC3 × 44 to remove the head rest

(13706-4100).3. Replace the head rest.4. Reassemble all parts in reverse order.

Page 52: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 16

6.20 Lamp House

Replacement part: 13706-51001. Remove the vertical movement cover (13706-

M332). (Refer to 5.4.)2. Disconnect P406 (J6) on the relay board

(13706-BA04).3. Remove the cap (13706-M503) and unscrew

SB3 × 84 and HH3 × 84.4. Remove the adjustment shaft (13706-M013).5. Remove the nut retainer (34500-M238).6. Remove the lock ring (13706-M502) and

lamp house (13706-5100).7. Remove the sleeve (13706-M501) from the

lamp house (13706-5100).8. Replace the lamp house.9. Reassemble all parts in reverse order.

10. Adjust the illumination rotation center.(Refer to 7.2.)

6.21 MYLA Laser Head

Replacement part: 13739-00001. Unscrew PC3 × 84 and PC3 × 54 to remove

the control board assy. (13706-1300). (Referto 6.10.)

2. Unscrew SB3 × 43 to remove the MYLA laserhead (13739-0000).Cut the tie wrap and heat shrinkable tube with anipper. Remove the laser head while retainingthe inside plug with a long-nose plier.

3. Replace the MYLA laser head.4. Reassemble all parts in reverse order.5. Adjust the following.

1) Adjust the YAG laser optical axis. (Refer to7.4.)

2) Perform electric adjustments. (Refer to 7.5.)Perform the initial setting of the master boardas necessary.

3) Adjust the YAG laser beam energy output.(Refer to 7.6.)

Page 53: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 17

6.22 Slide Plate

Replacement part: 13704-M3731. Remove the base plate (13706-4000). (Refer

to 6.18.)2. Unscrew FC2 × 44 to remove the slide plate

(13704-M373).3. Replace the slide plate.4. Reassemble all parts in reverse order.

6.23 Rack

Replacement part: 34348-M0091. Remove the base plate (13706-4000). (Refer

to 6.18.)2. Unscrew PC4 × 124 to remove the two racks

(34348-M009).3. Replace the racks.4. Reassemble all parts in reverse order.

Page 54: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

6 - 18

6.24 Pinion

Replacement part: 32903-M1381. Remove the base plate (13706-4000). (Refer

to 6.18.)2. Remove and replace the pinions (32903-

M138).3. Reassemble all parts in reverse order.

6.25 Gear A

Replacement part: 13706-M3041. Remove the joystick (13706-3200). (Refer

to 6.17.)2. Remove the retainer (13704-M306), left

bearing (82011-4052B), and gear A (13704-M304).

3. Replace gear A.4. Reassemble all parts in reverse order.

Apply an adhesive at the tip of SP3 × 8, insert itinto gear A, apply the adhesive around SP3 × 8.Use non-chamfered SP3 × 8.

6.26 Gear B

Replacement part: 13706-M3121. Remove the vertical screw movement (13706-

3130). (Refer to 6.15.)2. Unscrew SB2 × 64 to remove the gear B

(13706-M312).3. Replace gear B.4. Reassemble all parts in reverse order.

Page 55: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§7 ADJUSTMENT

7 - 1

7.1 Microscope Optical Axis

7.1.1 Binocular microscope optical axis

1. Mount the slit lamp (SL) optical axis jig andreticle attached eyepiece.Check that the scale of the jig and reticle arelevel and plumb.

2. Set the microscope magnification to 12.5× andmaximize the slit width.

3. Adjust SB3 × 102 and HH3 × 44 so that thereticle is aligned to the center (within thescale of 0.3).

4. Adjust SB3 × 102 and HH3 × 44 so that thereticle is within the scale of 0.3 when thepupillary distance of the binocular microscope(13706-2200) is changed.

5. Check that the shift between the scale centerand reticle is within the scale of 0.3.

6. When there is a shift, tighten HH3 × 44.The reticle moves in the direction correspondingto the tightened HH3 × 4.Perform this adjustment for the right and left eyesrespectively.The reticle should be within the scale of 0.3and the same in both eye sides.

7.1.2 Microscope visual field

1. Mount the slit lamp (SL) optical axis jig andreticle attached eyepiece.

2. Adjust the jig and eyepiece so that they arealigned in center.1) Up and down: Tighten or loosen SB4 × 64

and HH4 × 64.2) Right and left: Loosen SB4 × 64 to move the

microscope to the right or left.

Page 56: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 2

7.2 Illumination Rotation Center

1. Set the microscope magnification to 12.5×.2. Remove the hole cap (13706-M503) to insert

the focusing rod (34131-M551).Insert the focusing rod so that its flat surfacefaces the microscope.

3. Minimize the slit width and position the slitimage vertically. Adjust the focus on the flatsurface of the focusing rod.

4. Perform the following so that the slit imageon the flat surface of the focusing rod doesnot move horizontally even though the lamphouse (13706-5000) is moved horizontally.

1) When the slit image moves opposite to thelamp house movement: Tighten HH3 × 82 onthe patient’s side so that the focusing rod tiltsto the operator’s side.

2) When the slit image moves along with thelamp house movement: Loosen HH3 × 82 onthe operator’s side so that the focusing rodtilts to the patient’s side.

Slit image

Slit image

In the case of 1) In the case of 2)

Page 57: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 3

7.3 Aiming Spots

7.3.1 LD assy.

1. Remove the microscope (13706-2000).(Refer to 6.26.)

2. Mount the LD optical axis adjustment jig onthe expander assy. (13706-1150).

3. Turn the aiming rotation control so that thetwo aiming spots are aligned horizontally tothe X axis of the jig scale.

4. Adjust the following so that the jig center isaligned to the center between the aimingspots.1) Right and left: SB3 × 82 and HH3 × 64

2) Up and down: lower eccentric shaft (13611-M117)

5. Adjust the upper eccentric shaft so that theintensity of the aiming spots is the same.

6. Turn the aiming rotation control so that thetwo aiming spots are aligned on the X axis ofthe jig scale. Adjust the spots referring toSteps 4 to 5.

7.3.2 Aiming spot focus

1. Set the microscope magnification to 32× andinsert the focusing rod (34131-M551).

2. Loosen FG2 × 2. Rotate the lens holder(13704-M126) of the aperture assy. (13704-1230) so that the aiming spots projected onthe focusing rod merge into one.

3. Tighten FG2 × 2 and check that the lensholder does not rotate.

Aiming spots

Page 58: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 4

7.3.3 Aiming spot optical axis

1. TC3 × 62 to remove the cover (13706-M205).2. Insert the focusing rod (34131-M551).

3. Minimize the slit width and position the slitimage horizontally. Adjust the focus on theflat surface of the focusing rod.

4. Adjust HH3 × 63 so that the aiming spot isaligned to the slit image.

5. Position the slit image vertically.6. Adjust the eccentric shaft so that the aiming

spot is aligned to the slit image.7. Repeat Steps 2 to 5 so that aiming spot is

aligned to the slit image when the slit imageis rotated horizontally and vertically again.

Slit images

Aiming spots

Page 59: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 5

7.4 YAG Laser Optical Axis

* Remove the focusing rod (34131-M551) beforehand.

7.4.1 Mirror holder 3 and YAG laser head

1. Remove the control cap (82046-1705A) andcontrol (82045-2118B).

2. Disconnect P111 (J111) on the master board(13706-BA01).

3. Unscrew SB2 × 82 to remove the energymotor assy. (13706-EA09) together with themotor mount (13706-M105).

4. Disconnect P109 (J109) on the master board(13706-BA01).

5. Remove the focus assy. (13706-1130). (Referto 6.6.)Do not remove the energy adjustment assy.(13706-1110).

6. Remove the display panel assy. (13706-1400)and mount it as shown in the right figure.

7. Remove the D mirror assy. (13704-1140).(Refer to 6.27.)

8. Mount the YAG laser optical axis adjustmentjig in the position from which the D mirrorassy. was removed.

9. Maximize the energy output with the controlon the energy adjustment assy.

10. Place a burn paper behind the jig. Emit theYAG laser beam and check the burn patternposition.

11. Adjust the following so that the grid centerof the jig is aligned to the burn pattern.1) Upward shift: Loosen HH3 × 8 and tighten

SB2 × 8.2) Downward shift: Loosen SB2 × 8 and tighten

HH × 8.3) Right or left shift: Loosen SB3 × 63 and

rotate the two eccentric shafts (13755-M052).

Apply a threadlocking adhesive to PC3 × 63,SB3 × 82, and HH3 × 8 of the mirror holder(13755-M3438) after adjustment.

12. Reassemble the parts removed in Steps 1 to5.

13. Adjust the focus assy. position. (Refer to7.4.2.)

Page 60: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 6

7.4.2 Focus assy. position

1. Remove the D mirror assy. (13704-1140). (Refer to 6.7.)2. Mount the YAG laser optical axis adjustment jig in the position from which the D mirror

assy. was removed. (Refer to 7.4.1.)3. Maximize the energy output with the control

on the energy adjustment assy. (13706-1110).4. Place a burn paper behind the jig. Emit the

YAG laser beam and check the burn patternposition and that the laser beam is notobscured.

5. Adjust the following so that the grid centerof the jig aligned to the burn pattern and thatthe laser beam is not obscured.Loosen SB3 × 64 to adjust the focus assy. (13706-1130) position.Loosen SB3 × 63 and rotate the two eccentricshafts (13755-M052) to adjust the laser headhorizontally. (Refer to 7.4.1.)* Apply a threadlocking adhesive to SB3 × 64 on the focus assy.

6. Reassemble the D mirror assy. (13704-1140). (Refer to 6.7.)

7.4.3 D mirror assy.

1. Mount the focal point jig on the chinrest with SB3 × 152.2. Minimize the energy output with the control on the energy adjustment assy. (13706-1110).3. Focus the aiming spot on the focal point jig

and emit the YAG laser beam. Check theburn pattern position.

4. Adjust the D mirror assy. (13706-1140) whenthe aiming spot is not aligned to the burnpattern position.1) Right or left shift: Loosen PC3 × 63 and

rotate the D mirror assy.2) Upward or downward shift: Tighten or

loosen SB3 × 8 and HH3 × 82.

Page 61: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 7

7.4.4 Focus adjustment

1. Mount the focal point jig on the chinrest with SB3 × 152.2. Minimize the energy output with the control on the energy adjustment assy. (13706-1110).3. Set the focal point to 0 µm.

1) Focus the aiming spot on the focal point jig.2) Turn the knob so that the plunger (82001-

PR006) edge of the attachment plate (13706-M114) is pointing down.

3) Loosen HH3 × 3 in the attachment plate.4) Turn the control on the focus assy. (13706-

1130) so that the burn pattern on the focalpoint jig is at the minimum.

5) Tighten HH3 × 3 so that the plunger edge ofthe attachment plate is pointing down andthat the control on the focus assy. is fixed.

6) When the burn pattern on the focal point jigis not at the minimum, perform the adjustmentagain.

4. Check that the YAG laser beam is not obscured.1) Maximize the energy output with the control on the energy adjustment assy.2) Put a burn paper beyond the aiming focus point and emit several shots of the YAG laser

beam.3) Check the burn pattern. If the laser is obscured, perform the following:

a. Mirror holder 3 and YAG laser head adjustment (Refer to 7.4.1.)b. Focus assy. position adjustment (Refer to 7.4.2.)c. D mirror assy. adjustment (Refer to 7.4.3.)

5. Adjust the focal point.1) Minimize the energy output with the control on the energy adjustment assy.2) Set the focal point jig to 0.3) Move the slit so that the aiming spot is focused on the jig.4) Set the focus shift to 0 µm with the focus shift control. Emit the YAG laser beam and check

the burn pattern.5) Turn the jig scale one rotation toward you and set it to 0 again.6) Set the focus shift to ATN500 µm with the focus shift control. Emit the YAG laser beam and

check the burn pattern.7) Turn the jig scale two rotations away from

you and set it to 0 again.8) Set the focus shift to POST500 µm with the

focus shift control. Emit the YAG laser beamand check the burn pattern.

9) Check that the burn pattern of each focus shiftof 0, ANT500, and POST500 µm is the samesize.If the burn pattern is within ±50 µm of the 0 µm burn pattern for ANT and POST500 µm, thenit is within specifications.

Page 62: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 8

7.4.5 Focus shift indication

1. Set the focus shift control to 0 µm.2. Set DIP SW1-6 on the display board (13704-BA02) in the display panel assy. (13706-1400)

to the OFF (upper) position.3. Press the aiming switches to display either of Address 40 to 48 on the COUNTER display.4. Loosen HH3 × 3 and rotate the shaft of the

potentiometer (13702-E080) in the focus shiftassy. (13706-EA07) so that the value on theENERGY display is 130±10 when the focusshift is 0 µm.

5. Set the focus shift control to ANT500 µm.6. Press the aiming switches to display Address 40 on the COUNTER display.7. Press the reset switch to set the parameter address.8. As the same manner, set the parameters as shown in the table below (ANT375 to POST500).

9. Set DIP SW1-6 on the display board in the display panel assy. to the ON (lower) position.10. Turn the focus shift control and check that the YAG FOCUS SHIFT display is proper.

* When the display is abnormal, perform this adjustment again.

500 375 250 125 0 125 250 375 500Address 40 41 42 43 44 45 46 47 48

POSTANTFocusshift

Page 63: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 9

7.5 Electric Adjustments

7.5.1 YAG laser wiring check

1. Check visually that the wiring around thelaser head is wired properly and that anycable is not exposed.

2. Check that there is continuity between theground points of the laser head and groundededge of the power outlet with a digitalmultimeter.

7.5.2 Master board initial setting

1. Disconnect P107 (J107) on the master board (13706-BA01).2. Set DIP SW 1-1 and 1-6 on the display board (13704-BA02) in the display panel assy.

(13706-1400) to the OFF (upper) position.3. Turn the key switch to the ON position.4. Check that the aiming beam lights up.5. Press the aiming switches to display Address 199 on the COUNTER display.6. Press the STATUS switch so that the indication on the ENERGY display blinks.7. Press the aiming switches so that the value on the ENERGY display is 99.8. Press the reset switch to lock in the settings.

* Two shot beeps are audible.9. Set DIP SW 1-1 and 1-6 on the display board to the ON (lower) position. Turn the key

switch to the OFF position.10. Turn the key switch to the ON position again.11. The countdown starts, in about ten seconds, all indications light up on the all displays of the

control panel, and the YC-1800 goes into STANDBY.

Page 64: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 10

7.5.3 Aiming beam power output

1. Mount a laser powermeter on the strut of the forehead rest.2. Set the measurement wavelength of the laser powermeter to 635 nm.3. Set the YC-1800 to prevent it from interference lights.4. Tilt the lamp house to the right by the angle of 45º5. Turn the aiming rotation control so that the aiming spots are vertical.6. Adjust the aiming beam maximum output.

1) Press the aiming switches to maximize the aiming level.2) Turn VR1 on the master board (13706-BA01) so that the aiming beam output is 25 ±5 µW.

* Turning the VR clockwise increases the output.7. Check the aiming beam minimum output.

1) Press the aiming switches so that the aiming level becomes 1.2) Check that the aiming output is 0.5 ±0.3 µW.

8. When the aiming output is out of standards, adjust as follows:1) Set DIP SW 1-6 on the display board (13704-BA02) in the display panel assy. (13706-1400)

to the OFF (upper) position.2) Press the aiming switches to display Address 19 on the COUNTER display.3) Press the STATUS switch so that the indication on the ENERGY display blinks.4) Press the aiming switches to adjust the value (initial: 100) on the ENERGY display.

* Increasing the values increases the power output.5) Press the reset switch to lock in the settings.6) Set DIP SW 1-1 and 1-6 on the display board to the ON (lower) position.

9. Check that there is no ghost image in the aiming spots when the aiming output is maximumand illumination intensity is middle.

10. Check the aiming beam output when it is set to maximum.

Page 65: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 11

7.5.4 Agreement between the indicated and actual PFN voltages

1. Remove the focusing rod (34131-M551).2. Disconnect P107 (J107) on the master board (13706-BA01).3. Set DIP SW 1-1 and 1-6 on the display board (13704-BA02) in the display panel assy.

(13706-1400) to the OFF (upper) position.4. Press the aiming switches to display Address 10 on the COUNTER display.5. Press the STATUS switch so that the indication on the ENERGY display blinks.6. Press the aiming switches so that the value on the ENERGY display is 440.7. Press the reset switch to lock in the settings.8. Set DIP SW 1-6 to the ON (lower) position and DIP SW1-7 to the OFF (upper) position.

* The PFN voltage (monitored) is displayed on the ENERGY display.9. Connect a digital multimeter between P2 1st (+) and 5th (-) pins on the power supply board

(13733-BA01).10. Check that the actual PFN value is 440 ±10 V.11. When the value is out of the specifications, turn VR1 on the power supply board.

* Turning the VR clockwise increases the voltage.12. Set the value on the ENERGY display to 350 with the aiming switches.13. Set DIP SW1-7 to the ON (lower) position.

7.5.5 PFN voltage

1. Remove the focusing rod (34131-M551).2. Set DIP SW 1-1 and 1-6 on the display board (13704-BA02) in the display panel assy.

(13706-1400) to the OFF (upper) position.3. Press the aiming switches to display Address 10 on the COUNTER display.4. Check that the value on the ENERGY display is 350.5. Press the aiming switches to display Address 11 on the COUNTER display.6. Press the STATUS switch so that the indication on the ENERGY display blinks.7. Press the aiming switches so that the value on the ENERGY display is 02.8. Press the reset switch to lock in the settings.9. Set DIP SW 1-1 and 1-6 on the display board to the ON (lower) position.

10. The indication CAL appears on the YAG FOCUS SHIFT display and the PFN voltage isadjusted automatically.The PFN voltage is displayed on the ENERGY display. The actual pulse number is displayedon the COUNTER display. The PFN voltage is adjusted automatically until the number ofpulses becomes four in eight or more of ten shots.

11. The indication CAL disappears and the PFN voltage adjustment is completed.12. Adjust the PFN voltage depending on the room temperature.

1) Set DIP SW 1-6 on the display board to the OFF (upper) position.2) Press the aiming switches to display Address 10 on the COUNTER display.3) Press the STATUS switch so that the indication on the ENERGY display blinks.4) Change the value on the ENERGY display with the aiming switches as follows:

a. 25ºC or more: Automatic set PFN voltage + 5b. 20ºC or less: Automatic set PFN voltage - 5c. 20ºC < temperature < 25ºC: No additional adjustment

5) Set DIP SW 1-6 on the display board to the ON (lower) position.

Page 66: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 12

7.6 YAG Laser Beam Energy Output

7.6.1 Maximum energy output

1. Mount the detector of the powermeter on the chinrest.2. Turn the illumination off.3. Mount the dispersion lens on the objective lens at the edge of the microscope.

This step is unnecessary when using S200 (PHDX50).4. Align the two aiming spots to the center of the detector.5. Turn the energy control to maximize the YAG laser energy.6. Emit the YAG laser beam on the detector.

Adjust right HH3 × 12 of the energyadjustment assy. (13706-1110) so that theactual energy output becomes 11 to 12 mJ.* Tightening HH3 × 12 decreases the value and

loosening it increases the value.7. Turn VR2 on the master board (13706-BA01)

counterclockwise fully.8. Emit several shots of the YAG laser beam

and turn the energy control so that the actualenergy output becomes 11 mJ.Check that the indication on the ENERGYdisplay shows 2.0 to 3.0 mJ.When it is out of the specifications, replace oradd the NF filter (13456-G009) of the energycontrol assy. to change the sensitivity of theenergy monitor.Normally, the attenuator consists of one 13456-G009A and two 13456-G009E.

9. Change the attenuator (13456-G009) or applya threadlocking adhesive around the diffuser(13456-M077) of the added energyadjustment assy.

Name Part No. ND value13456-G009A 213456-G009B 0.113456-G009C 0.213456-G009D 0.313456-G009E 0.5

Attenuator

Page 67: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 13

7.6.2 Minimum energy output

1. Mount the detector of the powermeter on the chinrest.2. Turn the illumination off.3. Mount the dispersion lens on the objective lens at the edge of the microscope.

This step is unnecessary when using S200 (PHDX50).4. Align the two aiming spots to the center of the detector.5. Turn the energy control to minimize the YAG laser energy.6. Emit the YAG laser beam on the detector.

Adjust left HH3 × 12 of the energyadjustment assy. (13706-1110) so that theactual energy output becomes 0.2 mJ (0.5 mJor less).* Tightening HH3 × 12 increases the value and

loosening it decreases the value.

7.6.3 Energy Output Display

7.6.3.1 Display when the energy output is 10 mJ and 1 mJ

1. Mount the detector of the powermeter on the chinrest.2. Turn the illumination off.3. Mount the dispersion lens on the objective lens at the edge of the microscope.

This step is unnecessary when using S200 (PHDX50).4. Align the two aiming spots to the center of the detector.5. Adjust the energy output display of 10 mJ.

1) Turn the energy control so that the YAG laser actual energy output is 10 mJ.2) Adjust VR2 on the master board so that the displayed energy output is 10 mJ.

Standard: 10 mJ ±15%6. Adjust the energy output display of 1 mJ.

1) Turn the energy control so that the YAG laser actual energy output is 1 mJ.2) Check that the displayed energy output is 1 mJ.

Standard: 1 mJ ±15% ±0.1 mJ3) When the displayed energy output is out of the standard, adjust the following:

a. Set DIP SW 1-6 on the display board (13704-BA02) to the OFF (upper) position.b. Press the aiming switches to display Address 20 on the COUNTER display.c. Press the STATUS switch so that the indication on the ENERGY display blinks.d. Press the aiming switches to adjust the displayed energy output (initial: 10). * When the actual energy output is larger than the displayed one, increase the setting. * When the actual energy output is lower than the displayed one, decrease the setting.e. Press the reset switch to lock in the settings.f. Set DIP SW 1-6 to the ON (lower) position and check that the displayed output is 1 mJ. * Repeat the adjustment until the displayed output satisfies the standard.g. Perform the adjustment for 10 mJ again.

Page 68: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 14

7.6.3.2 Energy Output display and preset data setting

1. Set DIP SW 1-1, 1-5, and 1-6 on the display board (13704-BA02) to the OFF (upper) position.2. Turn the energy control to minimize the YAG laser energy.3. Turn the master switch on while pressing the STATUS switch.4. Emit five shots of the YAG laser beam and press the reset switch.5. Perform the same adjustment for the energy settings of 1, 3, 5, 7.5 mJ and the maximum.6. Turn the master switch off.7. Set DIP SW 1-1, 1-5, and 1-6 to the ON (lower) position.8. Turn the energy control to maximize the YAG laser energy. Turn the master switch on.9. Press the burst switch to set the number of pulses to 1, emit five shots of the YAG laser

beam, and press the reset switch.10. Press the burst switch to set the number of pulses to 2, emit five shots of the YAG laser

beam, and press the reset switch.11. Press the burst switch to set the number of pulses to 3, emit five shots of the YAG laser

beam, and press the reset switch.12. Turn the master switch off and on.13. Turn the energy control to minimize the YAG laser energy.14. Mount the detector of the powermeter on the chinrest.15. Turn the illumination off.16. Mount the dispersion lens on the objective lens at the edge of the microscope.

This step is unnecessary when using S200 (PHDX50).17. Align the two aiming spots to the center of the detector.18. Emit five shots of the YAG laser beam and check that the fifth actual energy output is

within the following standard.[Standard]Actual energy output less than 1.5 mJ: Displayed energy output ±15% ±0.1 mJActual energy output of 1.5 mJ or more: Displayed energy output ±15%Measure the actual energy output with each pulse number of 1, 2, and 3. Check that the differencebetween the actual and displayed energy outputs is within the standard.

19. When the actual energy output is out ofthe standard, change the parameters in theright table and repeat Steps 1 to 18.

20. Check that the actual energy output when the displayed one is 0.5 mJ satisfies the followingstandard.1) Set the YAG laser energy output to 0.5 mJ.2) Press the burst switch to set the pulse number to 1.3) Emit five shots of the YAG laser beam and check that the fifth actual energy output is within

the following standard.[Standard]Actual energy output of 0.5 mJ: Displayed energy output ±15% ±0.1 mJ

Address Display Contents Initial value30 PPL Preset potentiel Min. 3531 PP1 Preset potentiel 1 mJ 5032 PP3 Preset potentiel 3 mJ 7033 PP5 Preset potentiel 5 mJ 8534 PP7 Preset potentiel 7.5 mJ 10035 PPH Preset potentiel Max. 125

Page 69: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 15

7.7 Energy Output Check

1. Press the burst switch to set the pulse number to 1.2. Measure the maximum and minimum energy outputs of the YAG laser beam. Check that

they are within the following standard.3. Perform the same as Step 2 each time the

pulse number is set to 2 and 3.

7.8 Air Plasma Check

Set the energy to 4 mJ and emit the YAG laser beam. Check that the air plasma occurs.

7.9 Parameter Setting

1. Set DIP SW1-6 on the master board to the OFF position.The following displays show the data.1) Address: COUNTER display2) Displayed characters: YAG FOCUS SHIFT display3) Parameter (initial) value: ENERGY display

2. Display Address to be changed with the aiming switches.3. Press the STATUS switch.4. Change the parameter value with the aiming switches.5. Press the reset switch until two short beeps are audible.

The change is completed.6. Set DIP SW1-6 on the master board to the ON position.

7.10 Service Mode Indication

1. Start the YC-1800 while pressing the STATUS (READY mode) and burst switches.2. Press the + aiming switch to change the following.

1) Total shot number (displayed number × 1000 shots)2) Historical error number3) Program version number

7.11 Output Adjustment Mode

Start the YC-1800 while pressing the STATUS (READY mode) and two aiming switchessimultaneously. So, the energy display setting and aiming beam output adjustment are possiblewith the control panel.

Pulse number 1 2 3Max. output 10 mJ 9.1 mJ 7.8 mJMin. output 0.3 mJ 0.3 mJ 0.3 mJ

≥ ≥ ≥≤ ≤

Page 70: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

7 - 16

7.12 Vertical Movement Base

1. Remove the vertical movement cover.(Refer to 5.4.)

2. Loosen SB3 × 102.3. Adjust the vertical movement nut (34500-

M234) so that the base moves vertically andsmoothly when the joystick is rotated.

4. Tighten SB3 × 102 to fix the nut.5. Reassemble the vertical movement cover.

(Refer to 5.4.)

7.13 Smart Switch Adjustment

1. Adjust parameter for Address 109 so that the energy control is rotated and the value onthe ENERGY display is changed slightly when the smart switch is moved to the right andleft.

2. Move the smart switch to either of the right and left and immediately move it to anotherside. At that time, adjust parameter for Address 108 so that the energy control is rotatedand the value on the ENERGY display is changed slightly.

3. Perform Step 1 again.

Page 71: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

§8 INFORMATION

8 - 1

8.1 Wiring Diagram

Page 72: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 2

8.2 Connector Cable

13706-EA01

13706-EA02

13706-EA03

Page 73: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 3

13706-EA04

13706-EA05

13706-EA06

Page 74: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 4

13706-EA07

13706-EA08

13706-EA09

Page 75: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 5

13706-EA10

13706-EA11

13706-EA12

Page 76: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 6

13706-EA13

13706-EA14

13706-EA15

Page 77: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 7

13706-EA16

Page 78: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 8

17306-EA17

Page 79: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 9

Chinrest elevation controlEye level marker

Base fixing knob

Forehead rest

Eyepieces

Control panel

Fixation lamp

Grips

Laser head connector

Tonometer mount

Main body: Top view

8.3 Appearance

Main body: Operator’s side view

Page 80: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 10

Main body: Patient’s side - Right side view

AC adapter

Magnification

change control

Focus shift control

Aiming rotation control

Slit width control

Energy control

Chinrest

Hole for mounting the fo-

cusing rod

Illumination arm fixing knob

Microscope arm fixing knob

Slit rotation control

Grip

Smart switch

Hand switch

Power cable Connecting cable

Page 81: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 11

8.4 Error Message Table

8.5 Dip Switch Setting Table

DIP switch No. Contents

1 For canceling error detection

2 Unassigned

3 Unassigned

4 Unassigned

5 Unassigned

6 For permitting access to the EEPROM

7 For adjustment

8 For running the system

Error No. Symptom Cause 1 Internal structure overheat Internal structure overheat

5 Safety shutter malfunction Safety shutter malfunction or detection sensor failure

7 Aiming shutter malfunction Aiming shutter malfunction or detection sensor failure

10 Insufficient number of laser emissions

Failure of the YAG laser system or detection circuit

12 Excessive number of laser emissions

Failure of the YAG laser system or detection circuit

13 Accidental laser emission Failure of the YAG laser system or trigger circuit

14 Laser system malfunction Failure of the YAG laser system or trigger circuit

15 Unstable laser output Failure of the YAG laser system or detection circuit

30 YAG laser power supply malfunction

YAG laser power supply malfunction

31 PFN voltage failure YAG laser power supply malfunction

50 Break in the photodiode Malfunction of the photodiode for YAG laser output detection

51 Break in the potentiometer for YAG laser output adjustment

Break in the potentiometer for YAG laser output adjustment or its malfunction

52 Break in the potentiometer for focus shift adjustment

Break in the potentiometer for focus shift adjustment or its malfunction

53 Trigger input signal failure Trigger input signal circuit malfunction

54 Pulse number input signal failure Pulse number input signal circuit malfunction

90 Program data failure Contents of the ROM are rewritten. 91 EEPROM malfunction EEPROM memory data failure 93 Data failure Important RAM data failure 94 Keylock error Program malfunction

Page 82: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 12

8.6 Parameter Setting Tables

Address

Abbreviation

Default Contents Setting

0: Normal mode 1 AdJ 0 Adjustment mode control 1: Adjustment mode 0: Beeps 2 AL 0 Control of electric beeps at laser emission 1: No beeps

3 dEL 0 Unassigned at present 4 cnt 10 Countdown timer

0: Function 5 toc 0 Total counter function stop 1: Stop

6 n-A 0 Unassigned 0: Disabled

7 rdy 2 READY auto-off From 1: Value × (minute) 0: Memory

8 Cc1 1 Counter value memory 1: Clear 0: Memory

9 Ac1 0 Aiming data memory 1: Clear

10 PFn 350 PFN voltage 0: OFF 1: After Err10 11 CAL 0 PFN voltage adjustment 2: Forced 0: OFF 1: ON (with beeps) 12 FSc 0 Focus shift check at start 2: ON (without beeps) 0: Prohibited

13 trS 1 Prohibition of switching the trigger switch from outside 1: Permitted

0: HAND SW 14 HF 0 Switching of the trigger switch

1 : FOOT SW 0: Two circuits

15 tr2 1 Input switch of the trigger switch 1: One circuit

16 2dl 200 Latency time after the READY switch is pressed

Value × (10 msec)

17 AH 100 Aiming soft gain 18 yH 100 Energy soft gain 19 AL 100 Aiming offset 20 yL 10 Energy offset 10 = ±0 mJ 21 n-A 100 Correction factor (reserved) 22 n-A 100 Correction factor (reserved) 23 n-A 100 Correction factor (reserved) 24 n-A 100 Correction factor (reserved) 25 n-A 100 Correction factor (reserved) 26 n-A - Correction factor (reserved) 27 n-A - Correction factor (reserved) 28 n-A - Correction factor (reserved) 29 n-A - Correction factor (reserved) 30 PPL 35 Preset potentio Min. 31 PP1 50 Preset potentio 1 mJ 32 PP3 70 Preset potentio 3 mJ 33 PP5 85 Preset potentio 5 mJ 34 PP7 100 Preset potentio 7.5 mJ 35 PPH 125 Preset potentio Max.

Page 83: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 13

Address

Abbreviation

Default Contents Setting

36 n-A 100 Unassigned — 37 n-A 100 Unassigned — 38 n-A 100 Unassigned — 39 n-A 100 Unassigned — 40 F-5 70 Focus shift value –500 µm 41 F-3 85 Focus shift value –375 µm 42 F-2 105 Focus shift value –250 µm 43 F-1 125 Focus shift value –125 µm 44 F0 145 Focus shift value 0 µm 45 F1 165 Focus shift value +125 µm 46 F2 185 Focus shift value +250 µm 47 F3 205 Focus shift value +375 µm 48 F5 225 Focus shift value +500 µm 49 n-A 0 Unassigned — 50 brt 30 Soft blinking time

51 ruS 250 Time that the shutter functions during running

52 ruF 10 Number of shots during running ×100 53 n-A 0 Unassigned — 54 n-A 0 Unassigned — 55 ErG 100 Err15 gain adjustment (%) 56 ErO 3 Err15 offset adjustment (mJ) 57 n-A 100 Unassigned 58 StG 20 Gain adjustment of stability (%) 59 StO 1 Offset adjustment of stability (mJ) 60 PEL 3 Preset energy Min. 61 PE1 10 Preset energy 1 mJ 62 PE3 30 Preset energy 3 mJ 63 PE5 50 Preset energy 5 mJ 64 PE7 75 Preset energy 7.5 mJ 65 PEH 100 Preset energy Max. 66 n-A 0 Unassigned 67 n-A 0 Unassigned 68 n-A 0 Unassigned 69 n-A 0 Unassigned

70 PF 0 Time required for resetting the protective filter

Unassigned at present

71 StO 50 Stand-by time required for opening each shutter

Value × (10 msec)

72 StC 50 Stand-by time required for closing each shutter

Value × (10 msec)

73 n-A 10 Unassigned 74 n-A 10 Unassigned 75 n-A 0 Unassigned

Page 84: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 14

Address

Abbreviation

Default Contents Setting

76 ESb 1 Detection bit of the energy control 77 FSb 1 Detection bit of the focus shift control 78 n-A 0 Unassigned 79 n-A 0 Unassigned 80 Sti 33 Latency time of single (×10 msec) Value × (10 msec) 81 bti 66 Latency time of burst (×10 msec) Value × (10 msec) 82 Uti 200 Latency time of others (×10 msec) Value × (10 msec) 83 n-A 100 Unassigned 84 n-A 100 Unassigned 85 n-A 100 Unassigned 86 n-A 100 Unassigned

162: 4800 87 bAr 162 Baud rate

80: 9600 88 n-A 100 Unassigned 89 n-A 100 Unassigned

90 PuL 0 Pulse number for automatic PFN adjustment

0: Fixed 1: Weight average 91 End 1 Display change 2: Actual indication

92 n-A 0 Unassigned — 93 Std con 0 Setting depending on destination Unassigned at present 94 bSt 0 Burst (reserved) 95 bno 3 Burst (reserved) 96 n-A 0 Unassigned

0: Display enabled

97 tOE 0 Total energy display From 1: Enabled (×10 msec)

98 n-A 0 Unassigned — 0: Disabled 1: Foot SW enabled 99 FPr 2 Motion control of the sub-switches 2: Hand SW enabled

100 n-A 0 Unassigned — 0: Function disabled 1: ENERGY DOWN 2: ENERGY UP 3: AIMING DOWN 4: AIMING UP 5: READY 6: BURST

101 F01(R) 0 SW functions

7: RESET 0: Function disabled 1: ENERGY DOWN 2: ENERGY UP 3: AIMING DOWN 4: AIMING UP 5: READY 6: BURST

102 F02(L) 0 SW functions

7: RESET

Page 85: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 15

Address

Abbreviation

Default Contents Setting

0: Function disabled 1: ENERGY DOWN 2: ENERGY UP 3: AIMING DOWN 4: AIMING UP 5: READY 6: BURST

103 H01(R) 2 SW functions

7: RESET 0: Function disabled 1: ENERGY DOWN 2: ENERGY UP 3: AIMING DOWN 4: AIMING UP 5: READY 6: BURST

104 H02(L) 1 SW functions

7: RESET

105 H03

(press) 5 SW functions

106 to

108 n-A 0 Unassigned

109 Ont 4 Motor ON time Value × (10 msec) 110 to

197 n-A 0 Unassigned

198 PuP 0 Update of the ROM 199 EEP 0 Default setting of the EEPROM 99: Default setting

Page 86: OPHTHALMIC YAG LASER SYSTEM - Frank's Hospital …frankshospitalworkshop.com/equipment/documents/ophthalmology... · nidek ophthalmic yag laser system model yc-1800 service manual

XYC16*RDA001A/E

8 - 16