Development of a Read-Out System Using CMOS ASICs for the μ-PIC, Micro-Pixel Gaseous Chamber S. Iwaki a , H. Kubo a , M. Tanaka b , Y. Fujita b , H. Ohwada c , N. Higashi a , S. Kabuki a ,S. Kurosawa a , K. Miuchi a , K. Nakamura a , J. D. Parker a , T. Sawano a , A. Takada d , M. Takahashi a , T. Tanimori a , K. Taniue a , and K. Ueno e a Department of Physics, Kyoto University, b IPNS,KEK, c BeeBeans Technologies Co. Ltd d Scientific Balloon Laboratory, JAXA/ISAS, e Cosmic Radiation Laboratory, RIKEN * e-mail address: [email protected] 1, Our Detector ~ Electron Tracking Compton Camera ~ References 4, TPC Performance 400μm Abstract For astronomical observations, we have been developing an Electron Tracking Compton Camera (ETCC) based on a gaseous time projection chamber (µ-TPC) with a micro pixel gaseous chamber (µ-PIC), and scintillation cameras. With this camera, we have the balloon born experiment named SMILE. In a balloon-borne experiment, the power consumption is seriously restricted by capacity of batteries. To detect celestial gamma-rays, our Compton camera needs large detection volume and the number of readout channels of µ-PIC reaches more than 3000. Thus the power consumption has become a serious problem. We have developed low power CMOS Front-End ASICs and read-out systems with these chips for a gaseous detector in collaboration with KEK. The chip integrated 16 channels of a charge sensitive front-end amplifier, a shaper, and a discriminator has been designed and fabricated with the TSMC 0.5 µm process. The power consumption is about 18mW/ch. The performance of this chip coupled with a 10×10×15cm μ-TPC has been tested. In this poster, we introduce this new ASIC chip and report the fundamental performance of the TPC with this chips . Schematic view of the µ-PIC (left) photographs of 10cm 2 µ-PIC, (bottom right) and its microphotograph (upper right). For the sub MeV to MeV gamma-ray observation in astronomy, we have been developing an Electron Tracking Compton Camera (ETCC). The Camera consists of a gaseous time projection chamber (µ-TPC) [1] with µ-PIC [2][3], which detects the 3D-track and the energy of the recoil electron, and a scintillation camera, which detects the absorption point and the energy of the scattered gamma ray. By using these four pieces of information, we can completely reconstruct the Compton scattering event by event, and obtain a fully ray-traced gamma-ray image[1]. With this camera, we have planned the balloon experiments, SMILE. Schematic view of the ETCC. Observed tracks of cosmic-ray muon. ATLAS TGC ASD 64ch board 4ch/chip 2, Circuit Design & SPICE Simulation of New ASIC Chip : FE2009bal Specification • TSMC 0.5μm CMOS process • Preamplifier with a gain of 0.6V/pC •16ch sum amp with a gain of 0.8V/pC • Comparator with CMOS 2.5V output • 6bit DAC for calibrate channel variety. • ENC ~ 6000 electrons at C D = 100pF • Required voltage +/-2.5V,GND •16 channels in 1 chip •Power Consumption : 18mW/ch 4mm Microphotograph of theFE2009bal. PZC Preamp Shaper ABuf Comp arator analog ch0 analog ch15 Input ch0 digital ch0 ch7 Amp ×30 6bit DAC digital ch15 CAL Sum amp Vth CAL/DAC Control AOUT DOUT1 (CMOS 2.5V) DOUT16 C=1.1pF Block diagram of the FE2009bal We have developed CMOS ASIC for μ-TPC with KEK and improved the power consumption. The new ASIC chips, named “FE2009bal” , use “TSMC 0.5μm CMOS process ”. The IC contains 16 channels in a LQFP100-pin package. In this chip, the power consumption is reduced to 18mW/ch, and other parameters are all satisfy our requirement. Energy Spectrum of X-ray from 109 Cd . 3, Chip Performance 6, Summary & Future Works 22keV Res.=20%(FWHM) We test the fundamental performance of FE2009bal with test pulse. Analog output linearity is secured for input charge from -1pC to 1pC. Time walk is less than 6ns. Threshold level variety between each channels is about 300 mV . This variety is calibrated with DAC and reduced to about 20mV. The measured ENC is about 6000 electrons @Cd = 100pF. Requirement ASD FE2009bal Ch 4 16 Input dynamic range ±1pC -1.22.0pC ±1pC Noise (Cd=100pF) < 5000e 3000e ~6000e Timewalk (10fC-1pC) <10ns 6.5ns (20fC-1pC) <6ns Power Consumption <20mW 59mW/ch 18mW/ch SPICE Simulation 0 100 200 300 400 time(ns) 3.0 2.0 1.0 0 V(V) Digital output ( 0.1~ 0.4pC impulse inputs ) We have developed new ASIC chips : FE2009bal in collaboration with KEK. The chips work regularly and we successfully observe tracks of Cosmic-ray muon. Now the mass-production have been started and we are developing 128ch readout board. With this board, we will measure the whole track of Cosmic- ray muon to estimate detector efficiency for MIP, and obtain MeV γ –ray image. And we will develop readout system for fright model. 0 100 200 300 400 time(ns) 800 600 400 200 0 V(mV) Analog sum output ( 0.1~ 0.4pC impulse inputs ) Comparison of ASD and FE2009bal (10cm) 3 ETCC @Sanriku, JPN (2006) SMILE-[4] (30cm) 3 ETCC (~2012) SMILE-Operation test of ETCC @ 35km Measurement of Diffuse cosmic and atmospheric gamma rays 512ch readout ATLAS TGC ASD[5] (Bipolar Process) : 59mW/ch Analog sum-out linearity. Photograph of the detector setup. Max : 6ns (10fC-300fC) Input Charge(fC) Time walk (ns) 10fC Requirement for new ASIC Low power Consumption : < 20mW Small size and high integration; Time walk : < 10ns Input dynamic range : -1pC ~ 1pC ENC<5000e @ Cd = 100pF We took signals from the 10cm 3 µ-TPC using FE2009bal chips. We measured signals from only 2.5 × 2.5cm 2 region of the µ-PIC (64ch, both of anode and cathode) . We have successfully measured spectrum of X-ray from 109 Cd source, and got tracks of cosmic-ray muon. In a balloon-borne experiment the power consumption and space is seriously restricted. In addition, to detect track of recoil electron accurately, the time-walk of digital output have to be less than 10ns for input dynamic-range from 10fC to 1pC. Thus we need low-power and fast readout system. Photograph of SMILE Input Analog Sum (50Ω load) Qin=0.5pC 400mV 200ns Measured analog signal Measured time walk. Detector Status Gas : Ar 90% + C 2 H 6 10% 1atm (sealed) Effectine Volume : 2.5cm×2.5cm×10cm Stable gas gain : ~40000 Effective area FE2009bal chips 10cm Roadmap of the SMILE (40cm) 3 ETCC Super pressure balloon10days 4000ch readout Observation of Crab or Cyg X-1 1500ch readout Input dynamic range -1pC ~ pC Input Charge(C) [1] H.Kubo et al., Nucl.Instr.Meth.,A513(2003)94. [2] A.Ochi et al., Nucl.Instr.Meth.,A471(2001)264. [3] A.Ochi et al., Nucl.Instr.Meth.,A478(2002)196. [4] A. Takada et al., J. Phys. Soc. Jpn.,78(2009) Supplement A 161. [5] O.Sasaki and M.Yoshida,IEEE Trans.Nucl.Sci.,46(1999). After calibration : 20mV Before calibration : 300mV Threshold level variety

Upload: others

Post on 28-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of a Read-Out System Using CMOS ASICs for the ...€¦ · Timewalk (10fC-1pC)

Development of a Read-Out System Using CMOS ASICsfor the μ-PIC, Micro-Pixel Gaseous Chamber

S. Iwakia, H. Kuboa, M. Tanakab, Y. Fujitab, H. Ohwadac, N. Higashia, S. Kabukia,S. Kurosawaa, K. Miuchia, K. Nakamuraa, J. D. Parkera, T. Sawanoa,

A. Takadad, M. Takahashia, T. Tanimoria, K. Taniuea, and K. Uenoe

a Department of Physics, Kyoto University, b IPNS,KEK, c BeeBeans Technologies Co. Ltdd Scientific Balloon Laboratory, JAXA/ISAS, eCosmic Radiation Laboratory, RIKEN

* e-mail address: [email protected]

1, Our Detector ~ Electron Tracking Compton Camera ~

References

4, TPC Performance

400μm

AbstractFor astronomical observations, we have been developing an Electron Tracking Compton Camera (ETCC) based on a gaseous time projection chamber (µ-TPC) with a micro pixel gaseous chamber (µ-PIC), and scintillation cameras. With this camera, we have the balloon born experiment named SMILE. In a balloon-borne experiment, the power consumption is seriously restricted by capacity of batteries.

To detect celestial gamma-rays, our Compton camera needs large detection volume and the number of readout channels of µ-PIC reaches more than 3000. Thus the power consumption has become a serious problem. We have developed low power CMOS Front-End ASICs and read-out systems with these chips for a gaseous detector in collaboration with KEK. The chip integrated 16 channels of a charge sensitive front-end amplifier, a shaper, and a discriminator has been designed and fabricated with the TSMC 0.5 µm process. The power consumption is about 18mW/ch. The performance of this chip coupled with a 10×10×15cm3 μ-TPC has been tested. In this poster, we introduce this new ASIC chip and report the fundamental performance of the TPC with this chips .

Schematic view of the µ-PIC (left) photographs of 10cm2µ-PIC, (bottom right)and its microphotograph (upper right).

For the sub MeV to MeV gamma-ray observation in astronomy, we have been developing an Electron Tracking Compton Camera (ETCC). The Camera consists of a gaseous time projection chamber (µ-TPC) [1] with µ-PIC [2][3], which detects the 3D-track and the energy of the recoil electron, and a scintillation camera, which detects the absorption point and the energy of the scattered gamma ray. By using these four pieces of information, we can completely reconstruct the Compton scattering event by event, and obtain a fully ray-traced gamma-ray image[1].With this camera, we have planned the balloon experiments, SMILE.

Schematic view of the ETCC.

Observed tracks of cosmic-ray muon.

ATLAS TGC ASD 64ch board4ch/chip

2, Circuit Design & SPICE Simulation of New ASIC Chip : FE2009bal

Specification • TSMC 0.5μm CMOS process• Preamplifier with a gain of 0.6V/pC•16ch sum amp with a gain of 0.8V/pC• Comparator with CMOS 2.5V output• 6bit DAC for calibrate channel variety.• ENC ~ 6000 electrons at CD = 100pF• Required voltage +/-2.5V,GND •16 channels in 1 chip•Power Consumption : 18mW/ch

4mm

Microphotograph of theFE2009bal.

PZCPreamp ShaperABuf

Comparatoranalog ch0

analog ch15

Inputch0

digital ch0

ch7

Amp

× 30

6bitDAC

digital ch15

CAL

Sumamp

VthCAL/DAC

Control

AOUT

DOUT1

(CMOS2.5V)

DOUT16

Cf=1.1pF

Block diagram of the FE2009bal

We have developed CMOS ASIC for μ-TPC with KEK and improved the power consumption. The new ASIC chips, named “FE2009bal” , use “TSMC 0.5μm CMOS process” .The IC contains 16 channels in a LQFP100-pin package.In this chip, the power consumption is reduced to 18mW/ch, and other parameters are all satisfy our requirement.

Energy Spectrum of X-ray from 109Cd .

3, Chip Performance

6, Summary & Future Works

22keVRes.=20%(FWHM)

We test the fundamental performance of FE2009bal with test pulse. Analog output linearity is secured for input charge from -1pC to 1pC. Time walk is less than 6ns.Threshold level variety between each channels is about 300 mV . This variety is calibrated with DAC and reduced to about 20mV. The measured ENC is about 6000 electrons@Cd = 100pF.

Requirement ASD FE2009balCh 4 16Input dynamic range

±1pC -1.2~2.0pC ±1pC

Noise (Cd=100pF)

< 5000e ~3000e ~6000e

Timewalk(10fC-1pC)

<10ns ~6.5ns(20fC-1pC)

<~6ns

Power Consumption

<20mW 59mW/ch 18mW/ch

SPICE Simulation

0 100 200 300 400time(ns)

3.0

2.0

1.0

0

V(V)

Digital output( 0.1~ 0.4pC impulse inputs )

We have developed new ASIC chips : FE2009bal in collaboration with KEK. The chips work regularly and we successfully observe tracks of Cosmic-ray muon.

Now the mass-production have been started and we are developing 128ch readout board. With this board, we will measure the whole track of Cosmic-ray muon to estimate detector efficiency for MIP, and obtain MeV γ –ray image. And we will develop readout system for fright model.

0 100 200 300 400time(ns)

800

600

400

200

0

V(mV)

Analog sum output( 0.1~ 0.4pC impulse inputs )

Comparison of ASD and FE2009bal

(10cm)3 ETCC @Sanriku, JPN (2006)SMILE-Ⅰ[4]

(30cm)3 ETCC (~2012) SMILE-Ⅱ

Operation test of ETCC @ 35kmMeasurement of Diffuse cosmic

and atmospheric gamma rays512ch readoutATLAS TGC ASD[5](Bipolar Process) : 59mW/ch

Analog sum-out linearity.

Photograph of the detector setup.

Max : 6ns(10fC-300fC)

Input Charge(fC)

Tim

e w

alk

(ns)

10fC

Requirement for new ASIC•Low power Consumption : < 20mW•Small size and high integration;•Time walk : < 10ns•Input dynamic range : -1pC ~ 1pC•ENC<5000e @ Cd = 100pF

We took signals from the 10cm3 µ-TPC using FE2009bal chips. We measured signals fromonly 2.5 × 2.5cm2 region of the µ-PIC (64ch, both of anode and cathode) . We have successfully measured spectrum of X-ray from 109Cd source, and got tracks of cosmic-ray muon.

In a balloon-borne experiment the power consumption and space is seriously restricted. In addition, to detect track of recoil electron accurately, the time-walk of digital output have to be less than 10ns for input dynamic-range from 10fC to 1pC. Thus we need low-power and fast readout system.

Photograph of SMILE

Input

Analog Sum (50Ω load)

Qin=0.5pC

400mV

200ns

Measured analog signal

Measured time walk.

Detector Status•Gas : Ar 90% + C2H6 10% 1atm (sealed)•Effectine Volume : 2.5cm×2.5cm×10cm•Stable gas gain : ~40000

Effective area

FE2009balchips

10cm

Roadmap of the SMILE

(40cm)3 ETCCSuper pressure balloon~10days4000ch readout

Observation of Crab or Cyg X-11500ch readout

Input dynamic range-1pC ~ 1pC

Input Charge(fC)

[1] H.Kubo et al., Nucl.Instr.Meth.,A513(2003)94.[2] A.Ochi et al., Nucl.Instr.Meth.,A471(2001)264.[3] A.Ochi et al., Nucl.Instr.Meth.,A478(2002)196.[4] A. Takada et al., J. Phys. Soc. Jpn.,78(2009) Supplement A 161. [5] O.Sasaki and M.Yoshida,IEEE Trans.Nucl.Sci.,46(1999).

After calibration : 20mV

Before calibration : 300mV

Threshold level variety