MRI Image Artifacts

and their Remedies

Hsiao-Wen Chung (鍾孝文), Ph.D., Professor

Dept. Electrical Engineering, National Taiwan Univ.

Dept. Radiology, Tri-Service General Hospital

Page 2: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

2 of 160

Example of MRI Artifacts

Expected image Motion ghosts

Page 3: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

3 of 160

MRI Artifacts

• Images find something that actually

does not exist in the patient

• Images do not find things that

actually exist in the patient

• MRI is know to contain quite a lot of

different artifacts

Page 4: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

4 of 160

Why Artifacts ?

• Retina receive photic stimulations

• Films receive photic exposure

• Anything that can get films exposed

is shown on the photograph and

seen by humans

Page 5: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

5 of 160

Why MRI Artifacts ?

• Complicated MRI image formation

• Factors unrelated to physiological

conditions but affecting image

formation can become visible on the

resulting MRI

Page 6: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

6 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 7: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

7 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 8: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

8 of 160

Artifacts from B0 or B1

• Nonuniform magnetization or RF

excitation = nonuniform signal

• Rarely seen in modern clinical MRI

– Surface coil receiving profile is

widely known to be nonuniform

and hence not treated as artifacts

Page 9: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

9 of 160

Comparison of Different RF Coils

Body coil Head coil 3-in surface

Page 10: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

10 of 160

One RF-related Artifact

• Flow void (in spin-echo)

• 900 excitation and 1800 refocusing

pulses are at different times

• Out-of-slice flow causes dislocation

and signal loss

Page 11: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

11 of 160

2D Spin Echo Sequence

Different timing for 900 and 1800 pulses

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

900 1800

Page 12: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

12 of 160

Flow-Void in Spin-echo

• 900 – 1800 – signal receiving

• Timing difference

• Incomplete refocusing due to

flowing blood out of the slice

• Black-blood images (flow void)

Page 13: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

13 of 160

Bright-blood & Black-blood Images

Gradient-echo Spin-echo

Page 14: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

14 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 15: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

15 of 160

Spatial Encoding Artifact

• Gradient for spatial encoding

• Local magnetic field strength

= frequency = location

• Frequency variations = location

misregistration

Page 16: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

16 of 160

Chemical-shift Artifacts

• Fat and water protons have

inherently different resonance

frequencies

• Location misregistration

• Will be addressed later today

Page 17: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

17 of 160

Chemical-shift Artifacts

R-L freq encoding A-P freq encoding

Page 18: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

18 of 160

RF Leakage

• EM wave interference from outside

– Larmor freq in FM band for radio

• Usually isolated by RF shielding

• Opened scan room door could result

in RF leakage

Page 19: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

19 of 160

RF Leakage

Zipper artifacts along phase direction

A-P : freq encoding

S-I : phase encoding

Page 20: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

20 of 160

Zipper Artifacts

• Zipper artifacts not necessarily

caused by RF leakage !

• Un-encoded stimulated echo ? It

gets complicated

Page 21: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

21 of 160

Blood Flow Artifacts

• Flow-related enhancement

– Detailed in a future lecture

• Flow void (spin-echo)

• Displacement artifacts

Page 22: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

22 of 160

Two Encoding Directions

• Phase and frequency encoding occur

at different times

• Blood flows to another location

before frequency encoding ?

• The vessels seem to be “displaced”

• Displacement artifacts

Page 23: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

23 of 160

2D Gradient Echo Sequence

Different timing for two spatial encodings

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

< 900

Page 24: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

24 of 160

Displacement Artifacts

Displaced vessels especially for oblique ones

Note:

MRI displacement

artifacts occur for

in-plane flow

Page 25: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

25 of 160

Geometric Distortions

• Large range B0 inhomogeneity

• Resonance frequency changed

location mis-mapped

– Also sampling frequency related

– Particularly severe for EPI along

phase direction

Page 26: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

26 of 160

Large Range B0 Inhomogeneity

Shifting of resonance frequency locally

Image voxel :

Page 27: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

27 of 160

B0 Geometric Distortions

Arrow-head pattern along freq direction

Page 28: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

28 of 160

Distortions from Susceptibility

Distortion related to readout bandwidth

Round glass tubes

(diamagnetic)

CuSO4 solution

(paramagnetic)

1.5 Tesla

Gradient echo

Arrow-head pattern

Page 29: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

29 of 160

Severe Distortions from Iron Hair Pin

Spin echo Gradient echo

Page 30: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

30 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 31: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

31 of 160

Signal Receiver Gain

• Receiver magnifies the weak

(~uV) MR signals

• Receiver gain often adjusted

automatically for clinical MRI

• Too high or too low artifacts

Page 32: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

32 of 160

Receiver Gain Artifacts

Desired normal image Receiver gain too high

Page 33: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

33 of 160

Aliasing Artifacts

• Sampling frequency too low

– High frequency mistaken as low

– “Front” mistaken as “back”

• Increasing sampling frequency

solves the problem

Page 34: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

34 of 160

Frequency Encoding

Freq = location; Amplitude = proton density

Bo

Page 35: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

35 of 160

The MRI Signal Received (Echo)

Signals = sum of all frequencies

+ + + + =

Page 36: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

36 of 160

The Sampling Process

One data point every fixed interval

. . .

Page 37: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

37 of 160

Sampling Too Slow : Freq Mistaken

+ + + +

=

. . .

Page 38: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

38 of 160

Mistaken of Location

Bo

Page 39: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

39 of 160

Aliasing Artifacts

FOV too small BW increased

Page 40: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

40 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 41: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

41 of 160

Motion Ghosts

• Nonuniform signal intensity for each

repetition due to (periodic) motion

• k-space signal “modulated” by motion

• Modulation “frequency” reflected in image

“location”

• Multiple objects appear

Page 42: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

42 of 160

Pulse Sequence and k-space

RF

Gz

Gy

Gx

t

t

t

t

kx

ky

TR

TR

Page 43: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

43 of 160

Breathing Motion & Signal Modulation

Signal intensity changed by motion

Image slice location

Page 44: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

44 of 160

Signal intensity in k-space

kx

ky

Signal stronger ...

weaker ...

Page 45: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

45 of 160

Ghosts from Periodic Motion

Motion ghosts Respiratory gating

Page 46: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

46 of 160

The Modulation Pattern

• Periodic motion : clear ghosts

– Single frequencies in k-space

• Aperiodic motion : multiple

overlapped (blurred) ghosts

– Too many frequencies in k-space

Page 47: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

47 of 160

Ghosts from Respiratory Motion

Motion ghosts Respiratory gating

Page 48: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

48 of 160

Ghosts from Cardiac Motion

Motion ghosts ECG gating

Page 49: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

49 of 160

Artifacts from Aperiodic Motion

Desired image Motion ghosts

Page 50: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

50 of 160

Artifacts from Periodic Motion

Pulsation from abdominal aorta

Note :

1. Ghosts only for the

moving tissues

2. Always along phase

direction

Page 51: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

51 of 160

Motion Ghosts from CSF Flow

No Flow Comp Flow Comp

Page 52: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

52 of 160

Remedies

• Synchronized detection

– Respiratory or ECG gating

• Faster scan + breath-hold

• Ultrafast scanning

Page 53: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

53 of 160

Blood Flow Artifacts

• Flow-related enhancement

– Detailed in a future lecture

• Flow void (spin-echo)

• Displacement artifacts

Page 54: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

54 of 160

Flow-Related Enhancement

Neck Abdomen

Page 55: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

55 of 160

Cross Talk

• Frequently encountered in multi-angle

multi-slice imaging

• Some tissues excited multiple times within

one single TR

– TR effectively shortened low signal

• Also seen in contiguous-slice acquisition

Page 56: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

56 of 160

Single-slice Pulse Sequence Expanded

TR >> TE : scanner mostly idle

Gp

B1 t

t

...

...

Gs t ...

Gr t ...

Page 57: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

57 of 160

Adding Other Slices …

Making use of the idle time

Gp

B1 t

t

...

...

Gs t ...

Gr t ...

Page 58: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

58 of 160

… More Slices Added

Multi-slice imaging (scan time not lengthened)

Gp

B1 t

t

...

...

Gs t ...

Gr t ...

Page 59: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

59 of 160

Cross Talk for Multi-Angle Imaging

Excited by both slices

TR ~ halved

Page 60: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

60 of 160

Cross Talk

Lumbar image Sacrum image

Page 61: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

61 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 62: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

62 of 160

k-space Discontinuity

• Image reconstruction = Fourier

transform = Combination using

sine & cosine waveforms

• Abrupt discontinuity = ringing

Page 63: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

63 of 160

Artifacts from Minor Data Error

k-space MR image

Value too large

(kx = - 29, ky = - 41)

Page 64: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

64 of 160

Remedies

• Minor error anyway

• Replaced using neighboring data

• Reconstructed images often very

satisfactory

Page 65: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

65 of 160

Another Type of k-space Discontinuity

Full k-space No data in outer k-space

discontinuity

Page 66: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

66 of 160

Gibb’s Ringing (Truncation Artifacts)

256x256 256x128

Page 67: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

67 of 160

Gibb’s Ringing

• Also called truncation artifacts

• k-space discontinuity due to data

omission

• Filling the k-space with more

data removes the artifacts

Page 68: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

68 of 160

Gibb’s Ringing (Truncation Artifacts)

256x128 256x192 256x256

Page 69: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

69 of 160

Syringomyelia (hole in spinal cord) ?

256x128 256x256

Page 70: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

70 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 71: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

71 of 160

Not Only The Above !

• Image composed of “pixels”

• Huge number of protons in one pixel

• Signal = vector sum from all proton

magnetization

• What if inconsistent behavior ?

Page 72: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

72 of 160

Dephasing

• Protons = magnetization vectors

• Magnetization vectors oriented in all

different directions = 0 vector sum

– Reduced signals (dark)

• Intra-voxel phase dispersion

Page 73: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

73 of 160

Effects of TE on Dephasing

TE = 9 msec TE = 18 msec

Page 74: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

74 of 160

B0 Inhomogeneity

• Short T2* : Signal loss in gradient-echo

– Spin-echo unaffected due to 1800

refocusing pulse

– Signal loss more severe with long TE

– Also related to image resolution

Page 75: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

75 of 160

Air-Tissue Interface

Air distorts surrounding magnetic flux lines

Tissue :

diamagnetic

Air :

paramagnetic

Tissue :

diamagnetic

Page 76: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

76 of 160

Small Range Susceptibility Effect

B0 inhomogeneous in one voxel short T2*

Image voxel :

Page 77: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

77 of 160

Effects of Slice Thickness

B0 inhomogeneity is related to resolution

paramagnetic

Thin slice

Thick slice

Page 78: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

78 of 160

Effects of Slice Thickness

3 mm 5 mm 10 mm

Page 79: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

79 of 160

T2* Signal Loss in Hematoma

PDWI T2WI

Page 80: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

80 of 160

T2* Signal Loss in Hemorrhage

T1 PD T2 GrE

Page 81: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

81 of 160

Chemical-shift Dephasing

• Fat and water protons have inherent

different resonance frequencies

• 1.5 Tesla 220 Hz

– In-phase to out-phase every 2.27 msec

• Voxels containing both water fat have

variable signal intensities

Page 82: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

82 of 160

Chemical-shift Signal Loss

In-phase Out-phase

Page 83: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

83 of 160

Chemical-shift Artifacts

and Fat Suppression

Hsiao-Wen Chung (鍾孝文), Ph.D., Professor

Dept. Electrical Engineering, National Taiwan Univ.

Dept. Radiology, Tri-Service General Hospital

Page 84: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

84 of 160

Chemical Shift Effects

• Water and fat protons inherently

have slightly different resonance

frequencies

– Shielding effects from electron

cloud

– Difference in electron negativity

Page 85: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

85 of 160

Chemical Shift Phenomena

Shielding effects from electron cloud

O H C H

Oxygen nucleus attracts electron cloud,

reducing shielding effects on hydrogen nucleus

Carbon nucleus (methyl protons):

less electron negativity

Page 86: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

86 of 160

Proton MR Spectrum from Leg

Water & fat proton frequencies differ by 3.5 ppm

H2O

protons -CH2-

protons

from fat

ppm

Page 87: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

87 of 160

Chemical Shift Artifacts

• Spatial encoding in MRI :

– Field = frequency = location

• Water and fat protons have different

resonance frequencies

• Inherent location mis-registration

Page 88: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

88 of 160

Location = Magnetic Field = Frequency

Different frequencies imply different locations ??

water

fat

Bo

Page 89: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

89 of 160

Chemical Shift Artifacts

R-L freq encoding A-P freq encoding

Page 90: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

90 of 160

The Annoying Fat

• Inconsistent location with water tissues

• Inconsistent intensity in gradient-echo

• Fat is usually bright, obscuring lesion

– Short T1, moderate T2

Page 91: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

91 of 160

How Far Is It Displaced ?

• Qualitative description :

• Sampling needed before Fourier transform

• Sampling takes time

• Frequency difference more prominent as

time gets longer

• Displacement related to sampling speed

Page 92: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

92 of 160

How Far Is Artifact Displaced ?

Fast sampling,

frequency difference

less obvious

Slow sampling,

frequency difference

more obvious

Page 93: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

93 of 160

Quantitative Calculation

• Concept of “bandwidth per pixel” :

• Echo Fourier-transformed to form image

• Image = spectrum of echo

• A pixel occupies a spectral “interval”

– Bandwidth per pixel

Page 94: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

94 of 160

Image = Spectrum of Echo Signal

Every pixel occupies a “bandwidth”

Fourier

transform

time-varying

echo

pixel

Page 95: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

95 of 160

Chemical Shift Artifacts

• Displacement in number of pixels =

220 Hz / BW per pixel

– 3.5 ppm @ 1.5T = 220 Hz

• 32 KHz sampling freq ~ 125 Hz/pixel

• 220 / 125 ~ 2 pixels

Page 96: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

96 of 160

Fat Artifacts & Sampling Frequency

32 Hz/pixel 64Hz/pixel 128 Hz/pixel

Also note SNR difference !

Page 97: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

97 of 160

Fat Displacement & SNR

• High sampling freq low SNR

• Low sampling freq large

chemical shift displacement

Page 98: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

98 of 160

Freq & Phase Encoding

• No chemical shift displacement along the

phase encoding direction

• Identical time interval between RF

excitation to sampling

• Chemical shift artifacts occur only along

the freq encoding direction

Page 99: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

99 of 160

No Chemical Shift Artifact along Phase

t

t

t ...

RF

Gp

Gr

t

t ...

Gr

Gp

t RF

Identical time interval

between RF excitation

and readout

Page 100: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

100 of 160

EPI Is an Exception !

• EPI : continual sampling after one RF

• Long interval between phase encodings

• Severe chemical shift displacement along

phase direction for EPI

• Freq direction: less than one pixel

displacement due to very fast sampling

Page 101: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

101 of 160

EPI : Continual Sampling after One RF

RF

Gz

Gy

Gx

t

t

t

t

kx

ky

Large time interval between phase encodings

Page 102: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

102 of 160

EPI Fat Suppression

No Fat Sat With Fat Sat

Page 103: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

103 of 160

What Artifact Is This?

Soul out of skull ?

Answer :

Chemical shift

(fat) dislocation

artifact in EPI

Phase encoding

direction : A/P

Too severe that

EPI must use fat

suppression

Page 104: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

104 of 160

Likewise …

• Extremely severe fat artifact in EPI

• All other off-resonance artifacts get

“magnified” in EPI as well

– Susceptibility-induced geometric

distortion

Page 105: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

105 of 160

Distortions from Susceptibility

Distortion made visible using low bandwidth

Round glass tubes

(diamagnetic)

CuSO4 solution

(paramagnetic)

1.5 Tesla

Gradient echo

Arrow-head pattern

Page 106: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

106 of 160

Severe Distortions Using Iron Hair Pin

Spin echo Gradient echo

Page 107: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

107 of 160

But in EPI, no need for magnification

TSE T2WI EPI T2WI

Page 108: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

108 of 160

EPI Geometric Distortions

• Especially severe near skull base

– Ear, nose (air) …

• Stringent shimming requirement

• EPI less used outside brain

(although still some …)

Page 109: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

109 of 160

Not Only The Above !

• Image composed of “pixels”

• Huge number of protons in one pixel

• Signal = vector sum from all proton

magnetization

• What if inconsistent behavior ?

Page 110: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

110 of 160

Dephasing

• Protons = magnetization vectors

• Magnetization vectors oriented in all

different directions = 0 vector sum

– Reduced signals (dark)

• Intra-voxel phase dispersion

Page 111: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

111 of 160

Chemical Shift Dephasing

• Inherently different resonance freq

• 3.5 ppm 220 Hz at 1.5 Tesla

– In-phase changed to out-of-phase

every 2.27 msec

• Pixels having both fat and water show

varying signal intensity

Page 112: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

112 of 160

Chemical Shift Dephasing

In-phase to out-of-phase every 2.27 msec

z'

y' x'

z' z'

fat

water

RF t

Page 113: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

113 of 160

Chemical Shift Dephasing

In-phase Out-of-phase

Page 114: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

114 of 160

The Annoying Fat

• Inconsistent location with water tissues

• Inconsistent intensity in gradient-echo

• Fat is usually bright, obscuring lesion

– Short T1, moderate T2

Page 115: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

115 of 160

Signal Intensity vs. TE (Gradient Echo)

TE = 5 TE = 7 TE = 9 TE = 11

Page 116: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

116 of 160

Similar Behavior in Vertebral Bodies

Red marrow ~ water Yellow marrow ~ fat

TE = 9 msec TE = 10 msec TE = 11 msec TE = 12 msec TE = 13 msec

TE = 14 msec TE = 15 msec TE = 16 msec TE = 17 msec TE = 18 msec

Page 117: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

117 of 160

Note

• Chemical shift dephasing occurs in

gradient-echo images only

• Signal varies as a function of TE

• Spin-echo images unaffected due to

the refocusing 1800 pulse

Page 118: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

118 of 160

2D Gradient Echo Sequence

Chemical shift dephasing is present

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

< 900

Page 119: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

119 of 160

2D Spin Echo Sequence

No chemical shift dephasing

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

900 1800

Page 120: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

120 of 160

The Annoying Fat

• Inconsistent location with water tissues

• Inconsistent intensity in gradient-echo

• Fat is usually bright, obscuring lesion

– Short T1, moderate T2

Page 121: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

121 of 160

The Solution

• Why not just suppressing fat signal ?

– Fat suppression (Fat SAT)

• Comparison with before-suppression

provides diagnostic information

– Fatty or water-based mass?

Page 122: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

122 of 160

Comparison of Fat Suppression

Bone lesion clearly depicted after fat-SAT

Page 123: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

123 of 160

How to Suppress Fat ?

• Making use of the artifact origin

– Difference in frequencies

– Short T1 of fat

Page 124: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

124 of 160

CHESS Fat Suppression

• Chemical shift selective

• RF pulse to excite fat only without

touching water protons

• Use strong gradient to “spoil” the signal

• Perform imaging afterwards

Page 125: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

125 of 160

CHESS Principles

water

fat

900 fat only

RF

Gz

Gy

Gx

t

t

t

t

spin-echo fat-SAT

ppm

ppm

Page 126: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

126 of 160

Spin-Echo Fat Suppression

No fat sat With fat sat

Page 127: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

127 of 160

1-3-3-1 CHESS Pulse

• Simple means for frequency-

selective excitation

• Used in spatial modulation MRI

(SPAMM) for cardiac imaging

• You’ll see it in your homework set …

Page 128: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

128 of 160

Fat-SAT & Water-SAT

• Exactly the same principle

• Excite water without touching fat

• Strong gradient to “spoil” the excited

water signal

Page 129: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

129 of 160

Comparison of Fat-SAT & Water-SAT

Spin-echo Fat-SAT Water-SAT

Page 130: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

130 of 160

B0 & B1 Requirements

• Uniform B1 for complete saturation

– Good volume excitation RF coil

• Uniform B0 for consistent frequency

– Shimming before imaging

Page 131: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

131 of 160

Effects of RF Homogeneity

z'

y' x'

z'

y' x'

RF = 900

RF < 900

z'

z'

after spoiling

with strong gradient

residual

fat

Page 132: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

132 of 160

Spectra Before & After Shimming

water

H2O fat

-CH2-

ppm

Page 133: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

133 of 160

B0 Homogeneity Needed

• Fat-SAT according to frequency

– Frequency directly related to B0

• Air & tissue have different susceptibilities

– Incomplete fat-SAT very frequently

encountered near sinus

Page 134: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

134 of 160

Dixon (In- & Out-phase)

• Again using frequency difference

• Adjust TE to obtain in-phase (water+fat) &

out-phase (water-fat) gradient-echo

– Water image = in-phase + out-phase

– Fat image = in-phase – out-phase

Page 135: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

135 of 160

Dixon Method

z'

y' x'

z' z'

fat

water

RF t

Image #1 = water - fat

Image #2 = water + fat

Page 136: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

136 of 160

Remember Chemical Shift Dephasing ?

In-phase Out-of-phase

Page 137: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

137 of 160

The Dixon Method

• B0 homogeneity required as well

• Needs in- and out-phase images

twice scan time

– unless sampling within 2.27 msec

• Called in- (out-) phase clinically

Page 138: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

138 of 160

Water & Fat Images After Calculation

In-phase water image fat image

Page 139: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

139 of 160

Another Example in Abdomen

Out-phase water image fat image

Page 140: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

140 of 160

Multi-point Dixon Method

• If other factors interfere B0

– More variables come into play

• Multiple images at different TEs

– E.g. water + fat + B0 map

Page 141: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

141 of 160

How to Suppress Fat ?

• Making use of the artifact origin

– Difference in frequencies

– Short T1 of fat

Page 142: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

142 of 160

Short TI Inversion Recovery

• STIR imaging

• 1800 inversion (sounds familiar ?)

• Wait for T1 relaxation (TI : inversion time)

• Imaging started when fat passes null point

• TI ~ about 160 msec at 1.5T

Page 143: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

143 of 160

STIR Principles

TI

RF

Gz

Gy

Gx

t

t

t

t

spin-echo IR

fat

gray

matter

CSF

z'

y' x'

z'

y' x'

Page 144: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

144 of 160

STIR Fat-Sat Image (Lipoma)

T1-weighted image STIR image

Page 145: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

145 of 160

STIR & FLAIR Principles

Adjust TI to suppress different tissues

fat

gray matter

CSF

TI (STIR)

TI (FLAIR)

fluid attenuated inversion recovery

Page 146: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

146 of 160

FLAIR Depicting Lesion (Infarction)

T1WI T2WI FLAIR

Page 147: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

147 of 160

STIR Properties

• Homogeneous B0 not required

– T1 not a strong function of B0

• Image shows some T1 influence

• Inverse T1 weighting

– Opposite to common short-TR T1-

weighted images

Page 148: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

148 of 160

T1-weighting in STIR

• Long T1 tissues : bright

Short T1 tissues : dark

• CSF > gray matter > white matter

• Contrary to common spin-echo T1-

weighted images

Page 149: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

149 of 160

Why Inverse T1 weighting ?

CSF signal larger than gray matter at short TI

fat

gray matter

CSF

TI

Long T1 tissues not decayed much yet

Page 150: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

150 of 160

STIR Fat-sat Spin-Echo

Inversion recovery for fat-SAT

1.5 Tesla

IR Spin-echo

TI = 150 msec

TR = 2000

TE = 20

Periorbital fat is

suppressed.

Page 151: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

151 of 160

Fat-SAT Comparison

• CHESS : homogeneous B0 & B1

• Dixon : twice scan time (?)

• STIR : contrast altered

Page 152: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

152 of 160

Conventional Fat-SAT

• Suppression “preparation”

• Imaging follows suppression

Page 153: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

153 of 160

Special RF Pulses

• Shinnar-LaRoux RF pulse design

• k-space RF pulse design

• One RF to achieve > two purposes

• Example: simultaneous spatial & spectral

selection

Page 154: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

154 of 160

Spatial & Spectral Selection

Cost : 20 ~ 40 msec duration just for the RF

Page 155: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

155 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

Page 156: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

156 of 160

Other MRI Artifacts ?

• Come on ! Plenty of them !

• Many artifacts show no obvious origin or

remedies

– Just like human diseases

• What should I do if encountering strange

artifacts ?

Page 157: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

157 of 160

What Artifacts ?

Why do I get this ? Why do I get this ?

Page 158: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

158 of 160

What Artifacts ?

Thick water waves ? Thin water waves ?

Page 159: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

159 of 160

Artifacts Rule-of-Thumb

• Make sure reproducibility first

• Reasoning + adjusting parameters +

experiments, then back again

• Cost : Time + efforts

• Blame the manufacturer ?? Perhaps

never solved …

Page 160: MRI Image Artifacts and their Remedies1 of 160 MRI Image Artifacts and their Remedies Hsiao-Wen Chung (鍾孝文), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ

160 of 160

MRI Image Artifacts

and their Remedies

Hsiao-Wen Chung (鍾孝文), Ph.D., Professor