mri image artifacts and their remedies1 of 160 mri image artifacts and their remedies hsiao-wen...

Post on 25-Feb-2020

3 Views

Category:

Documents

1 Downloads

Preview:

Click to see full reader

TRANSCRIPT

1 of 160

MRI Image Artifacts

and their Remedies

Hsiao-Wen Chung (鍾孝文), Ph.D., Professor

Dept. Electrical Engineering, National Taiwan Univ.

Dept. Radiology, Tri-Service General Hospital

2 of 160

Example of MRI Artifacts

Expected image Motion ghosts

3 of 160

MRI Artifacts

• Images find something that actually

does not exist in the patient

• Images do not find things that

actually exist in the patient

• MRI is know to contain quite a lot of

different artifacts

4 of 160

Why Artifacts ?

• Retina receive photic stimulations

• Films receive photic exposure

• Anything that can get films exposed

is shown on the photograph and

seen by humans

5 of 160

Why MRI Artifacts ?

• Complicated MRI image formation

• Factors unrelated to physiological

conditions but affecting image

formation can become visible on the

resulting MRI

6 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

7 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

8 of 160

Artifacts from B0 or B1

• Nonuniform magnetization or RF

excitation = nonuniform signal

• Rarely seen in modern clinical MRI

– Surface coil receiving profile is

widely known to be nonuniform

and hence not treated as artifacts

9 of 160

Comparison of Different RF Coils

Body coil Head coil 3-in surface

10 of 160

One RF-related Artifact

• Flow void (in spin-echo)

• 900 excitation and 1800 refocusing

pulses are at different times

• Out-of-slice flow causes dislocation

and signal loss

11 of 160

2D Spin Echo Sequence

Different timing for 900 and 1800 pulses

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

900 1800

12 of 160

Flow-Void in Spin-echo

• 900 – 1800 – signal receiving

• Timing difference

• Incomplete refocusing due to

flowing blood out of the slice

• Black-blood images (flow void)

13 of 160

Bright-blood & Black-blood Images

Gradient-echo Spin-echo

14 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

15 of 160

Spatial Encoding Artifact

• Gradient for spatial encoding

• Local magnetic field strength

= frequency = location

• Frequency variations = location

misregistration

16 of 160

Chemical-shift Artifacts

• Fat and water protons have

inherently different resonance

frequencies

• Location misregistration

• Will be addressed later today

17 of 160

Chemical-shift Artifacts

R-L freq encoding A-P freq encoding

18 of 160

RF Leakage

• EM wave interference from outside

– Larmor freq in FM band for radio

• Usually isolated by RF shielding

• Opened scan room door could result

in RF leakage

19 of 160

RF Leakage

Zipper artifacts along phase direction

A-P : freq encoding

S-I : phase encoding

20 of 160

Zipper Artifacts

• Zipper artifacts not necessarily

caused by RF leakage !

• Un-encoded stimulated echo ? It

gets complicated

21 of 160

Blood Flow Artifacts

• Flow-related enhancement

– Detailed in a future lecture

• Flow void (spin-echo)

• Displacement artifacts

22 of 160

Two Encoding Directions

• Phase and frequency encoding occur

at different times

• Blood flows to another location

before frequency encoding ?

• The vessels seem to be “displaced”

• Displacement artifacts

23 of 160

2D Gradient Echo Sequence

Different timing for two spatial encodings

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

< 900

24 of 160

Displacement Artifacts

Displaced vessels especially for oblique ones

Note:

MRI displacement

artifacts occur for

in-plane flow

25 of 160

Geometric Distortions

• Large range B0 inhomogeneity

• Resonance frequency changed

location mis-mapped

– Also sampling frequency related

– Particularly severe for EPI along

phase direction

26 of 160

Large Range B0 Inhomogeneity

Shifting of resonance frequency locally

Image voxel :

27 of 160

B0 Geometric Distortions

Arrow-head pattern along freq direction

28 of 160

Distortions from Susceptibility

Distortion related to readout bandwidth

Round glass tubes

(diamagnetic)

CuSO4 solution

(paramagnetic)

1.5 Tesla

Gradient echo

Arrow-head pattern

29 of 160

Severe Distortions from Iron Hair Pin

Spin echo Gradient echo

30 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

31 of 160

Signal Receiver Gain

• Receiver magnifies the weak

(~uV) MR signals

• Receiver gain often adjusted

automatically for clinical MRI

• Too high or too low artifacts

32 of 160

Receiver Gain Artifacts

Desired normal image Receiver gain too high

33 of 160

Aliasing Artifacts

• Sampling frequency too low

– High frequency mistaken as low

– “Front” mistaken as “back”

• Increasing sampling frequency

solves the problem

34 of 160

Frequency Encoding

Freq = location; Amplitude = proton density

Bo

35 of 160

The MRI Signal Received (Echo)

Signals = sum of all frequencies

+ + + + =

36 of 160

The Sampling Process

One data point every fixed interval

. . .

37 of 160

Sampling Too Slow : Freq Mistaken

+ + + +

=

. . .

38 of 160

Mistaken of Location

Bo

39 of 160

Aliasing Artifacts

FOV too small BW increased

40 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

41 of 160

Motion Ghosts

• Nonuniform signal intensity for each

repetition due to (periodic) motion

• k-space signal “modulated” by motion

• Modulation “frequency” reflected in image

“location”

• Multiple objects appear

42 of 160

Pulse Sequence and k-space

RF

Gz

Gy

Gx

t

t

t

t

kx

ky

TR

TR

43 of 160

Breathing Motion & Signal Modulation

Signal intensity changed by motion

Image slice location

44 of 160

Signal intensity in k-space

kx

ky

Signal stronger ...

weaker ...

45 of 160

Ghosts from Periodic Motion

Motion ghosts Respiratory gating

46 of 160

The Modulation Pattern

• Periodic motion : clear ghosts

– Single frequencies in k-space

• Aperiodic motion : multiple

overlapped (blurred) ghosts

– Too many frequencies in k-space

47 of 160

Ghosts from Respiratory Motion

Motion ghosts Respiratory gating

48 of 160

Ghosts from Cardiac Motion

Motion ghosts ECG gating

49 of 160

Artifacts from Aperiodic Motion

Desired image Motion ghosts

50 of 160

Artifacts from Periodic Motion

Pulsation from abdominal aorta

Note :

1. Ghosts only for the

moving tissues

2. Always along phase

direction

51 of 160

Motion Ghosts from CSF Flow

No Flow Comp Flow Comp

52 of 160

Remedies

• Synchronized detection

– Respiratory or ECG gating

• Faster scan + breath-hold

• Ultrafast scanning

53 of 160

Blood Flow Artifacts

• Flow-related enhancement

– Detailed in a future lecture

• Flow void (spin-echo)

• Displacement artifacts

54 of 160

Flow-Related Enhancement

Neck Abdomen

55 of 160

Cross Talk

• Frequently encountered in multi-angle

multi-slice imaging

• Some tissues excited multiple times within

one single TR

– TR effectively shortened low signal

• Also seen in contiguous-slice acquisition

56 of 160

Single-slice Pulse Sequence Expanded

TR >> TE : scanner mostly idle

Gp

B1 t

t

...

...

Gs t ...

Gr t ...

57 of 160

Adding Other Slices …

Making use of the idle time

Gp

B1 t

t

...

...

Gs t ...

Gr t ...

58 of 160

… More Slices Added

Multi-slice imaging (scan time not lengthened)

Gp

B1 t

t

...

...

Gs t ...

Gr t ...

59 of 160

Cross Talk for Multi-Angle Imaging

Excited by both slices

TR ~ halved

60 of 160

Cross Talk

Lumbar image Sacrum image

61 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

62 of 160

k-space Discontinuity

• Image reconstruction = Fourier

transform = Combination using

sine & cosine waveforms

• Abrupt discontinuity = ringing

63 of 160

Artifacts from Minor Data Error

k-space MR image

Value too large

(kx = - 29, ky = - 41)

64 of 160

Remedies

• Minor error anyway

• Replaced using neighboring data

• Reconstructed images often very

satisfactory

65 of 160

Another Type of k-space Discontinuity

Full k-space No data in outer k-space

discontinuity

66 of 160

Gibb’s Ringing (Truncation Artifacts)

256x256 256x128

67 of 160

Gibb’s Ringing

• Also called truncation artifacts

• k-space discontinuity due to data

omission

• Filling the k-space with more

data removes the artifacts

68 of 160

Gibb’s Ringing (Truncation Artifacts)

256x128 256x192 256x256

69 of 160

Syringomyelia (hole in spinal cord) ?

256x128 256x256

70 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

71 of 160

Not Only The Above !

• Image composed of “pixels”

• Huge number of protons in one pixel

• Signal = vector sum from all proton

magnetization

• What if inconsistent behavior ?

72 of 160

Dephasing

• Protons = magnetization vectors

• Magnetization vectors oriented in all

different directions = 0 vector sum

– Reduced signals (dark)

• Intra-voxel phase dispersion

73 of 160

Effects of TE on Dephasing

TE = 9 msec TE = 18 msec

74 of 160

B0 Inhomogeneity

• Short T2* : Signal loss in gradient-echo

– Spin-echo unaffected due to 1800

refocusing pulse

– Signal loss more severe with long TE

– Also related to image resolution

75 of 160

Air-Tissue Interface

Air distorts surrounding magnetic flux lines

Tissue :

diamagnetic

Air :

paramagnetic

Tissue :

diamagnetic

76 of 160

Small Range Susceptibility Effect

B0 inhomogeneous in one voxel short T2*

Image voxel :

77 of 160

Effects of Slice Thickness

B0 inhomogeneity is related to resolution

paramagnetic

Thin slice

Thick slice

78 of 160

Effects of Slice Thickness

3 mm 5 mm 10 mm

79 of 160

T2* Signal Loss in Hematoma

PDWI T2WI

80 of 160

T2* Signal Loss in Hemorrhage

T1 PD T2 GrE

81 of 160

Chemical-shift Dephasing

• Fat and water protons have inherent

different resonance frequencies

• 1.5 Tesla 220 Hz

– In-phase to out-phase every 2.27 msec

• Voxels containing both water fat have

variable signal intensities

82 of 160

Chemical-shift Signal Loss

In-phase Out-phase

83 of 160

Chemical-shift Artifacts

and Fat Suppression

Hsiao-Wen Chung (鍾孝文), Ph.D., Professor

Dept. Electrical Engineering, National Taiwan Univ.

Dept. Radiology, Tri-Service General Hospital

84 of 160

Chemical Shift Effects

• Water and fat protons inherently

have slightly different resonance

frequencies

– Shielding effects from electron

cloud

– Difference in electron negativity

85 of 160

Chemical Shift Phenomena

Shielding effects from electron cloud

O H C H

Oxygen nucleus attracts electron cloud,

reducing shielding effects on hydrogen nucleus

Carbon nucleus (methyl protons):

less electron negativity

86 of 160

Proton MR Spectrum from Leg

Water & fat proton frequencies differ by 3.5 ppm

H2O

protons -CH2-

protons

from fat

ppm

87 of 160

Chemical Shift Artifacts

• Spatial encoding in MRI :

– Field = frequency = location

• Water and fat protons have different

resonance frequencies

• Inherent location mis-registration

88 of 160

Location = Magnetic Field = Frequency

Different frequencies imply different locations ??

water

fat

Bo

89 of 160

Chemical Shift Artifacts

R-L freq encoding A-P freq encoding

90 of 160

The Annoying Fat

• Inconsistent location with water tissues

• Inconsistent intensity in gradient-echo

• Fat is usually bright, obscuring lesion

– Short T1, moderate T2

91 of 160

How Far Is It Displaced ?

• Qualitative description :

• Sampling needed before Fourier transform

• Sampling takes time

• Frequency difference more prominent as

time gets longer

• Displacement related to sampling speed

92 of 160

How Far Is Artifact Displaced ?

Fast sampling,

frequency difference

less obvious

Slow sampling,

frequency difference

more obvious

93 of 160

Quantitative Calculation

• Concept of “bandwidth per pixel” :

• Echo Fourier-transformed to form image

• Image = spectrum of echo

• A pixel occupies a spectral “interval”

– Bandwidth per pixel

94 of 160

Image = Spectrum of Echo Signal

Every pixel occupies a “bandwidth”

Fourier

transform

time-varying

echo

pixel

95 of 160

Chemical Shift Artifacts

• Displacement in number of pixels =

220 Hz / BW per pixel

– 3.5 ppm @ 1.5T = 220 Hz

• 32 KHz sampling freq ~ 125 Hz/pixel

• 220 / 125 ~ 2 pixels

96 of 160

Fat Artifacts & Sampling Frequency

32 Hz/pixel 64Hz/pixel 128 Hz/pixel

Also note SNR difference !

97 of 160

Fat Displacement & SNR

• High sampling freq low SNR

• Low sampling freq large

chemical shift displacement

98 of 160

Freq & Phase Encoding

• No chemical shift displacement along the

phase encoding direction

• Identical time interval between RF

excitation to sampling

• Chemical shift artifacts occur only along

the freq encoding direction

99 of 160

No Chemical Shift Artifact along Phase

t

t

t ...

RF

Gp

Gr

t

t ...

Gr

Gp

t RF

Identical time interval

between RF excitation

and readout

100 of 160

EPI Is an Exception !

• EPI : continual sampling after one RF

• Long interval between phase encodings

• Severe chemical shift displacement along

phase direction for EPI

• Freq direction: less than one pixel

displacement due to very fast sampling

101 of 160

EPI : Continual Sampling after One RF

RF

Gz

Gy

Gx

t

t

t

t

kx

ky

Large time interval between phase encodings

102 of 160

EPI Fat Suppression

No Fat Sat With Fat Sat

103 of 160

What Artifact Is This?

Soul out of skull ?

Answer :

Chemical shift

(fat) dislocation

artifact in EPI

Phase encoding

direction : A/P

Too severe that

EPI must use fat

suppression

104 of 160

Likewise …

• Extremely severe fat artifact in EPI

• All other off-resonance artifacts get

“magnified” in EPI as well

– Susceptibility-induced geometric

distortion

105 of 160

Distortions from Susceptibility

Distortion made visible using low bandwidth

Round glass tubes

(diamagnetic)

CuSO4 solution

(paramagnetic)

1.5 Tesla

Gradient echo

Arrow-head pattern

106 of 160

Severe Distortions Using Iron Hair Pin

Spin echo Gradient echo

107 of 160

But in EPI, no need for magnification

TSE T2WI EPI T2WI

108 of 160

EPI Geometric Distortions

• Especially severe near skull base

– Ear, nose (air) …

• Stringent shimming requirement

• EPI less used outside brain

(although still some …)

109 of 160

Not Only The Above !

• Image composed of “pixels”

• Huge number of protons in one pixel

• Signal = vector sum from all proton

magnetization

• What if inconsistent behavior ?

110 of 160

Dephasing

• Protons = magnetization vectors

• Magnetization vectors oriented in all

different directions = 0 vector sum

– Reduced signals (dark)

• Intra-voxel phase dispersion

111 of 160

Chemical Shift Dephasing

• Inherently different resonance freq

• 3.5 ppm 220 Hz at 1.5 Tesla

– In-phase changed to out-of-phase

every 2.27 msec

• Pixels having both fat and water show

varying signal intensity

112 of 160

Chemical Shift Dephasing

In-phase to out-of-phase every 2.27 msec

z'

y' x'

z' z'

fat

water

RF t

113 of 160

Chemical Shift Dephasing

In-phase Out-of-phase

114 of 160

The Annoying Fat

• Inconsistent location with water tissues

• Inconsistent intensity in gradient-echo

• Fat is usually bright, obscuring lesion

– Short T1, moderate T2

115 of 160

Signal Intensity vs. TE (Gradient Echo)

TE = 5 TE = 7 TE = 9 TE = 11

116 of 160

Similar Behavior in Vertebral Bodies

Red marrow ~ water Yellow marrow ~ fat

TE = 9 msec TE = 10 msec TE = 11 msec TE = 12 msec TE = 13 msec

TE = 14 msec TE = 15 msec TE = 16 msec TE = 17 msec TE = 18 msec

117 of 160

Note

• Chemical shift dephasing occurs in

gradient-echo images only

• Signal varies as a function of TE

• Spin-echo images unaffected due to

the refocusing 1800 pulse

118 of 160

2D Gradient Echo Sequence

Chemical shift dephasing is present

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

< 900

119 of 160

2D Spin Echo Sequence

No chemical shift dephasing

z gradient

RF (B1) t

t

y gradient t

x gradient t ...

900 1800

120 of 160

The Annoying Fat

• Inconsistent location with water tissues

• Inconsistent intensity in gradient-echo

• Fat is usually bright, obscuring lesion

– Short T1, moderate T2

121 of 160

The Solution

• Why not just suppressing fat signal ?

– Fat suppression (Fat SAT)

• Comparison with before-suppression

provides diagnostic information

– Fatty or water-based mass?

122 of 160

Comparison of Fat Suppression

Bone lesion clearly depicted after fat-SAT

123 of 160

How to Suppress Fat ?

• Making use of the artifact origin

– Difference in frequencies

– Short T1 of fat

124 of 160

CHESS Fat Suppression

• Chemical shift selective

• RF pulse to excite fat only without

touching water protons

• Use strong gradient to “spoil” the signal

• Perform imaging afterwards

125 of 160

CHESS Principles

water

fat

900 fat only

RF

Gz

Gy

Gx

t

t

t

t

spin-echo fat-SAT

ppm

ppm

126 of 160

Spin-Echo Fat Suppression

No fat sat With fat sat

127 of 160

1-3-3-1 CHESS Pulse

• Simple means for frequency-

selective excitation

• Used in spatial modulation MRI

(SPAMM) for cardiac imaging

• You’ll see it in your homework set …

128 of 160

Fat-SAT & Water-SAT

• Exactly the same principle

• Excite water without touching fat

• Strong gradient to “spoil” the excited

water signal

129 of 160

Comparison of Fat-SAT & Water-SAT

Spin-echo Fat-SAT Water-SAT

130 of 160

B0 & B1 Requirements

• Uniform B1 for complete saturation

– Good volume excitation RF coil

• Uniform B0 for consistent frequency

– Shimming before imaging

131 of 160

Effects of RF Homogeneity

z'

y' x'

z'

y' x'

RF = 900

RF < 900

z'

z'

after spoiling

with strong gradient

residual

fat

132 of 160

Spectra Before & After Shimming

water

H2O fat

-CH2-

ppm

133 of 160

B0 Homogeneity Needed

• Fat-SAT according to frequency

– Frequency directly related to B0

• Air & tissue have different susceptibilities

– Incomplete fat-SAT very frequently

encountered near sinus

134 of 160

Dixon (In- & Out-phase)

• Again using frequency difference

• Adjust TE to obtain in-phase (water+fat) &

out-phase (water-fat) gradient-echo

– Water image = in-phase + out-phase

– Fat image = in-phase – out-phase

135 of 160

Dixon Method

z'

y' x'

z' z'

fat

water

RF t

Image #1 = water - fat

Image #2 = water + fat

136 of 160

Remember Chemical Shift Dephasing ?

In-phase Out-of-phase

137 of 160

The Dixon Method

• B0 homogeneity required as well

• Needs in- and out-phase images

twice scan time

– unless sampling within 2.27 msec

• Called in- (out-) phase clinically

138 of 160

Water & Fat Images After Calculation

In-phase water image fat image

139 of 160

Another Example in Abdomen

Out-phase water image fat image

140 of 160

Multi-point Dixon Method

• If other factors interfere B0

– More variables come into play

• Multiple images at different TEs

– E.g. water + fat + B0 map

141 of 160

How to Suppress Fat ?

• Making use of the artifact origin

– Difference in frequencies

– Short T1 of fat

142 of 160

Short TI Inversion Recovery

• STIR imaging

• 1800 inversion (sounds familiar ?)

• Wait for T1 relaxation (TI : inversion time)

• Imaging started when fat passes null point

• TI ~ about 160 msec at 1.5T

143 of 160

STIR Principles

TI

RF

Gz

Gy

Gx

t

t

t

t

spin-echo IR

fat

gray

matter

CSF

z'

y' x'

z'

y' x'

144 of 160

STIR Fat-Sat Image (Lipoma)

T1-weighted image STIR image

145 of 160

STIR & FLAIR Principles

Adjust TI to suppress different tissues

fat

gray matter

CSF

TI (STIR)

TI (FLAIR)

fluid attenuated inversion recovery

146 of 160

FLAIR Depicting Lesion (Infarction)

T1WI T2WI FLAIR

147 of 160

STIR Properties

• Homogeneous B0 not required

– T1 not a strong function of B0

• Image shows some T1 influence

• Inverse T1 weighting

– Opposite to common short-TR T1-

weighted images

148 of 160

T1-weighting in STIR

• Long T1 tissues : bright

Short T1 tissues : dark

• CSF > gray matter > white matter

• Contrary to common spin-echo T1-

weighted images

149 of 160

Why Inverse T1 weighting ?

CSF signal larger than gray matter at short TI

fat

gray matter

CSF

TI

Long T1 tissues not decayed much yet

150 of 160

STIR Fat-sat Spin-Echo

Inversion recovery for fat-SAT

1.5 Tesla

IR Spin-echo

TI = 150 msec

TR = 2000

TE = 20

Periorbital fat is

suppressed.

151 of 160

Fat-SAT Comparison

• CHESS : homogeneous B0 & B1

• Dixon : twice scan time (?)

• STIR : contrast altered

152 of 160

Conventional Fat-SAT

• Suppression “preparation”

• Imaging follows suppression

153 of 160

Special RF Pulses

• Shinnar-LaRoux RF pulse design

• k-space RF pulse design

• One RF to achieve > two purposes

• Example: simultaneous spatial & spectral

selection

154 of 160

Spatial & Spectral Selection

Cost : 20 ~ 40 msec duration just for the RF

155 of 160

MRI Image Formation

• Magnetization

• RF excitation

• Spatial encoding repeat N times

• Signal receiving

• Image calculation

156 of 160

Other MRI Artifacts ?

• Come on ! Plenty of them !

• Many artifacts show no obvious origin or

remedies

– Just like human diseases

• What should I do if encountering strange

artifacts ?

157 of 160

What Artifacts ?

Why do I get this ? Why do I get this ?

158 of 160

What Artifacts ?

Thick water waves ? Thin water waves ?

159 of 160

Artifacts Rule-of-Thumb

• Make sure reproducibility first

• Reasoning + adjusting parameters +

experiments, then back again

• Cost : Time + efforts

• Blame the manufacturer ?? Perhaps

never solved …

160 of 160

MRI Image Artifacts

and their Remedies

Hsiao-Wen Chung (鍾孝文), Ph.D., Professor

Dept. Electrical Engineering, National Taiwan Univ.

Dept. Radiology, Tri-Service General Hospital

top related