    Error lists/Troubleshooting 3-1

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Error list 58XX

    Display Cause Remedy

    Error 1 Error in measurement system for

    rotational speed.

    No rotor recognized after 10

    measurements.

    No starting magnet (SM+L1)

    recognized when calculating the

    rotational speed.

    Rational speed for the measurement is

    not reached, abortion after 12 sec.The motor or the frequency converter

    could be defective.

    The time for one rotation is < 4 ms

    (errors).

    Device brakes.

    After repair:

    False connection of cable U2/V2/W2

    from motor to converter.

    Check the rotational system on the

    rotor.

    Service program 8.

    Check/replace motor cap incl. speed

    sensor, check connections.

    Replace "Control" PCB,

    check connections at plug X10.

    Service program 9

    converter

    motor run.

    Customer service program 9

    Check connections.

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    2/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-2 Error lists/Troubleshooting

    Error 2 Imbalance sensor is defective.The idle value is outside of the limits or

    the measuring value is absolutely incor-

    rect. Error occurs only if an incorrect idle

    value is applied for at least 3 sec.

    Device does not start or brake.

    The old Imbalance-recognition PCB is

    still in use (cable colors at plug X 10:

    white/black/red/blue).

    Is the customers rotor damaged orloaded asymmetrically?

    Is the centrifuge placed on a

    horizontal, stable and resonance-free

    work surface?

    Was the centrifuge moved and just put

    into operation? If so, let the centrifuge

    run at maximum speed.

    Is the sensor idle value stable and

    within the tolerance (670 100)?

    Adjust imbalance PCB and carry outimbalance adjustment. Replace

    imbalance PCB in case of no success.

    Was the imbalance PCB adjusted

    correctly after being replaced or after

    the motor being replaced?

    Following replacement of the motor,

    the PCB Imbalance or the PCB

    Control, was a fresh imbalance

    calibration performed?

    Was the correct rotor with the correctweighs used for an imbalance

    adjustment?

    Was the plug reconnected correctly to

    the PCB after a repair?

    Check plug X10.

    Use new Imbalance-recognition PCB

    from -02 onwards (cable colors at plug

    X 10: White/brown/green/yellow).

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    3/36

    Error lists/Troubleshooting 3-3

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Error 3no rotor No rotor pulses during run oracceleration.

    Missing rotor, measuring system for

    rotational speed defective, drive

    defective.

    Rotor not fastened.

    No pulses after start. Device brakes

    for 10 sec.

    Pulse failure during run. Free deceler-

    ation requires 8 min!

    Error remains active after a powersupply interruption.

    Software version 3.3 to 3.7

    when using brake level "0" or "1".

    Fasten rotor.Keep device switched on and start

    again after > 12 min.

    Service program 8.

    Check the state of the rotational

    speed sensor and its connections,

    replace if necessary.

    Service program 9

    Check converter, replace if necessary.

    Check motor, replace if necessary.

    Replace "Control" PCB,

    check connector X10 on PCB "Con-

    trol".

    Install software version as of 3.8.

    Error 4 Sensor for lid locking mechanism (Hall

    sensor, position recognition) defective.

    During run device brakes.

    At standstill with the lid closed, start is

    not possible.

    Service program 4.

    The voltage on the sensor should be

    approx. 2.5 V 0.2 V without the

    influence of a magnetic field("Control 58XX" PCB).

    Check the sensor position.

    Error 5 No report from lid during run.

    Changed measuring value of

    Hall sensor.

    Device brakes, even when the lid is open.

    Service program 4.

    Check the fastening and the

    adjustment of the microswitches.

    Carry out a continuity check of the

    microswitches while they are closed.

    Check the state of the magnets on the

    junction plate.

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    4/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-4 Error lists/Troubleshooting

    Error 6Overload Undervoltage or overvoltage.Error in frequency converter (FC), thermal

    overload.

    Not ready after switching on the

    frequency converter.

    FC has carried out a RESET, caused

    by mains interruption.

    Device stops without braking

    Check mains supply!(Does starting of rotor or cooling

    aggregate cause deviations?)

    Customer service program 9

    view the error table of the

    converter.

    Check ribbon cable from converter to

    X9 of the PCB Control.

    Check the overtemperature sensor of

    the motor and its connections to the

    converter.

    Check the brake resistors, replace

    them if necessary (150 each).

    Check motor fan, replace if necessary.

    Service program 6.

    (Only present in refrigerated

    centrifuges.)

    Check the converter, replace it if

    necessary.

    Customer service program 3.

    See "Problems with defective

    frequency converter, motor, brake

    resistors, overload, Error 6",

    page 3-17.

    Error 7 Overspeed monitoring.

    Maximum rotational speed of the rotor

    is exceeded.

    There are regulator deviations of

    800 1/min from the nominal value.

    Perhaps resulting error from Error 25.

    Device stops without braking.

    Service program 9.

    Check the motor, replace it if

    necessary.

    Replace the converter if necessary.

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    5/36

    Error lists/Troubleshooting 3-5

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Error 8 Acceleration ramps are not met. Slippage monitoring (comparison

    rotational field-actual rotational speed)

    during acceleration.

    Device brakes.

    Service program 11.Determination of the acceleration

    time.

    Replace the converter if necessary.

    Check the motor, replace it if

    necessary.

    Error 9 Writing error EEPROM. Re-initialize the EEPROM with service

    program 5.

    Replace the "Control" PCB.

    Error 10 Reading error EEPROM.

    Error display and setting of standard

    settings. If there is an error when reading

    the setting for imbalance the device

    cannot start (error 2!). Writing and reading

    operations are only carried out when the

    device is at a standstill (after switching

    on, after a run, manual rotor recognition)

    Re-initialize the EEPROM with service

    program 5.

    Replace the "Control" PCB.

    Error 11 Watchdog reset, software or hardware

    error.

    No actions possible, after a reset the port

    lines are constantly reset to 1.

    Replace the EPROM or the

    PROCESSOR (IC1).

    Replace the "Control" PCB.

    Error 12 EDP error

    Serial PC interface: error during trans-

    mission.

    Device can be operated by hand.

    Check the interface PCB and the plug

    connections for correct positioning.

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    6/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-6 Error lists/Troubleshooting

    Error 13 EDP errorSerial PC interface: reception error, parity

    error, wrong code number, check sum

    error.

    Check the interface PCB and the plugconnections for correct positioning.

    Error 14 EPROM check sum error after mains on.

    Error program memory, EPROM

    defective.

    No further actions possible.

    Replace the EPROM or the

    PROCESSOR (IC1).

    Replace the "Control" PCB.

    Error 15

    Rotor

    Imbalance error at n > 2000 1/min

    Device brakes with normal ramp; device

    was pushed or rotor not fixed.

    refer to ERROR 2.

    Rotor not fixed tightly.

    Device was pushed.

    Table is not stable.

    Error 16 Error during request to FC.

    FC not ready, drive relay has dropped.

    Error after telegram repetition.

    Device switches off stops without

    braking.

    Customer service program 9

    view the error table of the

    converter.

    Check cable and connections at plug

    X9 and X2.

    Check the overtemperature sensor of

    the motor and its connections to the

    converter.

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    7/36

    Error lists/Troubleshooting 3-7

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Error 17 Error during transmission of selecttelegram to FC.

    Queue overflow FC telegrams.

    Error after 3 repetitions of a select

    telegram.

    Device switches off stops without

    braking.

    Customer service program 9 view the error table of the

    converter.

    Check cable and connections at plug

    X9 and X2.

    Check the overtemperature sensor of

    the motor and its connections to the

    converter .

    Error 18

    overtemp.

    Overtemperature of the rotor chamber.

    After the regulation (temp.) the interval

    of 5 C is exceeded.

    Coolant can not be aspirated, ambient

    temperature is too low.

    After starting a run, the chamber tem-

    perature exceeds the nominal value or

    during a run, the chamber tempera-

    ture exceeds the value (nominal

    value + 10 C).

    Refrigeration aggregate does not

    work.

    An FC with extended temperature

    range was replaced by an old FC.

    Rotor chamber temperature > 50C.

    Device switches off braked stop.

    Room temperature too high?

    Direct sunlight?

    Wait 4 hours following transport.

    Bring the centrifuge into a room with a

    room temperature of 22 C.

    Is ventilation ensured by sufficient

    clearence at the sides of the device?

    Which rotor is used? Can the desired

    temperature be achieved with this

    rotor and this adapters?

    (See Operating manual/Accessories

    guide)

    Is the lid sealing damaged?

    Is the lid closed correctly? (Air entry

    may lead to heavy icing in the rotor

    chamber. The regulation is interfered,

    full performance is not reached).

    Is the condenser fan contaminated,

    defective? Check with service

    program 6.

    Service program 6,

    check temperature sensors,

    condenser and bottom of rotor

    chamber.

    PCB Control defective?

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    8/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-8 Error lists/Troubleshooting

    Error 19 Overtemperature at the condenser tem-perature > 75 C.

    Fan is blocked / does not start, air

    intake is covered.

    Temperature sensor defective.

    Device switches off braked stop, the

    refrigeration aggregate is switched off,

    the fan keeps running.

    Let the refrigeration aggregate cool

    down, check the air intake.

    Check whether the condenser fan rotateswhen the aggregate is switched on.

    Service program 6. Check temperature sensors, condenser

    and bottom of rotor chamber.

    Check the condenser fan, replace it if

    necessary.

    Check all plug connections X3, X5,

    X10.

    Check whether the refrigeration

    aggregate is filled.

    Error 20 Temperature sensor in the rotor chamber

    is defective.

    Device switches off braked stop.

    Refrigeration aggregate switches off.

    Service program 6.

    Check sensor in rotor chamber.

    Replace the sensor or the "Control"

    PCB.

    Error 21 Temperature sensor on condenser

    defective (NTC).

    Measuring value is not within < 8

    C > 91 C.

    Device switches off braked

    deceleration, refrigeration keeps running.

    Service program 6.

    Check temperature sensor at

    condenser.

    Replace the sensor or the "Control"

    PCB.

    Error 22 No 50 Hz signals for fan control.

    During run and stand-by. The fan is

    controlled with full power.

    After mains on: The device is not rec-

    ognized as a refrigerated device!!

    In case of sporadic occurrence:

    Interference in the environment,

    caused, for example, by other devices.

    Interference frequencies on the power

    line.

    Replace the "Control" PCB.

    Remove interference in the environ-

    ment. Test the centrifuge in another

    room with a different device circuit.

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    9/36

    Error lists/Troubleshooting 3-9

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Error 24Clear

    Memory

    Program memory is full. Erase program if required,or start new run with Start or Prog.

    Error 25

    Interrupt

    Interrupt.

    Mains interrupted during a run.

    Is mains voltage generally too low or too

    high?

    Do deviations exist?

    Cable in the device is damaged, e.g. after

    replacement of the motor cap/rotational

    speed sensor or PCB imbalance

    detection.

    Restart.

    Check mains supply (during start of

    cooling aggregate as well).

    Check cables in the device and look

    for cable damage, particularly near the

    motor.

    Error 31 x 0: Timeout while closing the lid.

    1: Timeout while opening the lid.

    2: Undefined position when opening.

    3: Monitoring of microswitches for lid.

    4: Released lid does not open.

    5: Error during lid test after power-on.

    Check the connections on the sensor

    board and plug X10.

    Display Cause Remedy

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    10/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-10 Error lists/Troubleshooting

    Errors from the service programs

    Display Cause

    Error 32 Hardware error: imbalance measuring

    system.

    Check the connections on the sensor

    board and plug X10.

    Error 33 Measuring values outside of the

    measuring range, too little space between

    sensor and magnet.

    Check the connections on the sensor

    board and plug X10.

    Error 34 Faulty measuring series.

    Mean value not within limits or

    distance too short.

    Scattering run 13 too great.

    Scattering run 46 too great.

    Distance sensor magnet too great.

    Rotor is unbalanced during imbalance

    adjustment.

    Motor continues to vibrate.

    Check the connections on the sensorboard and STV X10.

    See Error 2.

    Adjust distance Sensor Magnet,

    see Sec. 4.

    Clean pegs and nuts of the rotor/

    bucket and grease slightly with

    Eppendorf pivot grease.

    Test another rotor.

    When using Rotor A-4-81:use magnet carrier imbalance

    detection with 9x13 mm magnet.

    Speed sensor cable jammed.

    Replace shock mounts for motor.

    Perform imbalance calibration on a

    solid bench.

    Error 35 Imbalance sensor idle value exceeds thelimit (670 100).

    Check the connections on the sensorboard and plug X10.

    see Error 34

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    11/36

    Error lists/Troubleshooting 3-11

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Error 36 EEPROM not yet initialized. Reinitialize EEPROM with serviceprogram 5.

    Change PCB "Control".

    Error 37 Voltage version of frequency converter

    and Centrifuge do not correspond.

    undefined

    display, nobackground

    illumination

    Display module or 26-pole cable

    defective, 5 V power supply faulty

    Check all cables.

    Change PCB Control pad or PCB

    Control.

    Display Cause

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    12/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-12 Error lists/Troubleshooting

    Warnings / Error displays

    (are not stored in the EEPROM)

    Display Cause

    Voltage

    false

    W-Error 27

    Mains voltage of FC incorrect.

    Check voltage version.

    SPEED W-Error 29

    SPEED the set rotational speed is faster

    than the nominal rotational speed of the rotor

    in use.

    The device stops the run after rotor

    recognition.

    Enter new nominal speed.

    200 rpm New rotor was used in devices with older soft-

    ware

    Install new software and carry out

    imbalance adjustment.

    IMBAL W-Error 30IMBAL Imbalance at < 2000 1/min.

    The device brakes.

    No imbalance calibration was carried out after

    installing rotor A-4-81.

    Check rotor and loading. Test bucketsand adapters with a scale.

    Clean pegs and nuts of the rotor/

    bucket and grease slightly with

    Eppendorf pivot grease.

    Test another rotor.

    Carry out imbalance calibration.

    Rotor See Error 15.

    Caution with open lid locking mechanism is actuated! Risk of squeezing!

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    13/36

    Error lists/Troubleshooting 3-13

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Pressopen

    LIFT LID

    W-Error 31The locking mechanism for the lid is in an

    undefined position.

    Lid is released and does not lift up

    automatically.

    Actuate the OPEN button. Open lid manually via emergency

    release using the rotor key.

    Bring lid mechanism into different

    position via emergency release using

    the rotor key.

    Lid mat/reinforcement may be

    warped. Align manually, or replace if

    required.

    Smooth locking eye of lid mat/

    reinforcement using fine sandpaper.

    Sealing of rotor chamber may have

    become rigid. Replace if required.

    The gas spring is too sluggish.

    Grease with "Esso S420", or replace

    if required.

    Grease lid mechanism (i.e. with

    "Molykote") in the area of the Y-slider

    and the two Pertinax plates (there

    might be a roughness).

    Replace motor of lid latch.

    Replace Control PCB (too little

    power).

    Magnets on Y-slider fell off.

    Replace Y-slider.

    CLOSE

    LID

    W32

    Command to close the lid.

    Close lid again firmly.

    Display Cause

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    14/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-14 Error lists/Troubleshooting

    Overload Drive overloaded. See Error 6. Mains voltage too low.

    CLEAn

    ro

    W-Error 32

    200 runs

    Error appears after run 200/201/202

    (from software 2.0 onwards).

    Clean rotor, rotor cone, motor

    cone and chamber,

    grease rotor axle slightly!

    With the swing-bucket rotor,

    lightly grease the pivots and

    grooves of the bucket with

    pivot grease.

    Open lid.

    Standby

    off

    Centrifuge not used for 6 hours

    (from software 2.0 onwards).

    Open lid and close again.

    Display Cause

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    15/36

    Error lists/Troubleshooting 3-15

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Changerotor W-Error 28Load-cycle count of 100.000 reached

    (from software 3.0 onwards with PCB in SMD

    technology).

    The warning is displayed 3 times following the

    run after 98,000, 99,000 and 99,600 runs.

    It appears after every run after 100,000 runs.

    Notes:

    The "Change rotor" message is conceived of

    as a warning and provides the customer and

    the service technician the opportunity to

    evaluate the condition of the rotor.

    Use of a rotor with load change limitation that

    has exceeded the 100,000th load change is

    not permitted!

    It must be noted that the load change numbers

    (CLyy) can only be reliably evaluated when:

    1. In case of repairs the data of the

    defective PCB Control has been taken

    over by the new PCB Control. (Saving

    does not take place in the EPROM

    program, but instead only on the PCB.)

    2. The load change counter has been resetwhile exchanging a rotor.

    3. The customer uses only a rotor of the

    same type in the centrifuge.

    4. The customer uses several centrifuges

    and each rotor is allocated to one

    centrifuge.

    5. The fixed-angle rotors CL13 and 17 have

    no load change limitation. It is not

    necessary to discontinue operation.

    Only the fixed-angle rotors CL14 and 18

    and all swing-bucket rotors have a load

    change limitation.

    Only centrifuges equipped with the PCB

    Control in SMD technology as of software

    version 3.0 enable a load change number

    (CLxx).).

    Evaluate the load change table withservice program SE 14.

    The load change numbers (CLyy) are

    measured separately for each rotor

    type.

    Immediately discontinue operation of

    the rotor if necessary, see notes.

    Prior to using a new rotor, delete the

    load change table for the rotor type with

    service program SE 14.

    Display Cause

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    16/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-16 Error lists/Troubleshooting

    no rotor See Error 3.

    Over-

    temp.

    See Error 18.

    CCI

    false

    An old FC with software 2.1 was used. Use new FC with extended temperature

    range.

    Clear

    Memory

    See Error 24.

    Display Cause

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    17/36

    Error lists/Troubleshooting 3-17

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Problems with defective frequency converter, motor, brake resistors, overload, Error 6

    1. Mains voltage stable at customers site? (Error 25 possible.)

    Mains voltage generally too low or too high?

    2. Room temperature too high, max. 35C ? Direct sunlight?

    3. Sufficient space for air ventilation at sides of device?

    Minimum 15 cm at both sides and 10 cm at rear side.

    4. Do the brake resistors still have approx. 150 Ohm each?

    5. Is the motor fan working (only present in refrigerated centrifuges)?

    6. Service Manual, "Service function, Program 13, Viewing the error table".

    7. Service Manual, "Customer service program 7, Last error report".

    8. Turn motor by hand and listen to ball bearing noise.

    A sluggish or loud motor may result in higher current, thus damaging the frequency con-

    verter and/or the brake resistors!

    9. Service Manual, "Customer service program 2, FC current during a run".

    xx Phase current in phase line to motor

    230 V:Typical value after start is xx = 42 (4.2 A)

    and after reaching maximum speed xx = 24 (2.4 A).

    120 V:Typical value after start is xx = 84 (8.4 A)

    and after reaching maximum speed xx = 48 (4.8 A).

    10. With Error 6: Service Manual, Customer service program 9, current error indication

    (selectable via cursor key only directly after error)

    xx error number e.g. 06

    zz 0 =no error message7=overcurrent FC

    1 =undervoltage8=interference EPROM FC

    2 =overvoltage10=no release

    4 =overtemperature FCFA=reset by internal software or hardware

    5 =overtemperature motorFF=serial interface time out

    Check the non-refrigerated centrifuges 5804/5810 to see whether the foam parts in the device are still

    suitable to maintain the required ventilation flow for cooling.

    2 : x x y y y z z z z z

    9 : E r x x y y z z

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    18/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-18 Error lists/Troubleshooting

    Customer service programs

    By actuating simultaneously the PLUS and MINUS buttons (cursor keys) the device is switched over to

    the service indication mode. This does not depend on the operational mode the device is currently in.

    The run of the centrifuge is not influenced by this.

    The indications are continued by pressing the PLUS button.

    During a run only the indications no. 1...6 are accessible (from START to opening the lid or resetting to

    the nominal value indication with START).

    During a fault the indication automatically jumps to 9. After that you can start again switching to different

    indications (error indication with display of the error location and for error 6 of the FC error numbers).

    The FC parameters can be checked during a standstill and while the lid is closed (FC on!) with the

    indication 2 and by actuating the MINUS button.

    1: Temperatures

    Devices without refrigeration system

    Devices with refrigeration system

    x.x Condenser temperature, refrigeration aggregate switches off at 75 C

    * means the refrigeration aggregate is switched on

    yy.y nominal temperature rotor chamber sensor

    zz.z corrected nominal value = regulator value. Rotor-dependant!

    2: FC parameters current during a run

    xx Motor current 33 3.3 A (phase current in the phase line to the motor)

    y F = FC on, display for fixed-angle rotor (only for factory setting)

    yyy parameter 9, display for swing-bucket rotor (only for factory setting)

    zzzzz parameter 3 rotational field - rotational speed/control value (only for factory setting)

    2: FC parameter during a standstill (for factory settings only)

    1 :

    1 : x x * y y y z z z

    2 : x x y y y z z z z z

    2 : P x x V = y y y y y

    =

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    19/36

    Error lists/Troubleshooting 3-19

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3: FC parameters temperature during a run

    xxx temperature frequency converter

    FC oldxxx =digital value displayed

    (switching off at 128, value for 20 C = approx. 83)

    FC newxxx =display in C x factor 10

    (switching off at approx. 75 C, display for 20 C = 200)

    * FC temperature warning

    yyy parameter 4: acceleration ramp (for factory checks only)

    zzz parameter 5:braking ramp (for factory checks only)

    4: Rotor recognition

    xxx Rotor relation (rotor table, rotor relations, see last page of Section 3)

    y number of recognition measurements

    zz Rotor number (rotor table, rotor relation, see last page of Section 3)

    5: Imbalance monitoring

    xxx Imbalance value (current sensor idle value during standstill)

    Recommended service setting = 670 30 Please note, important!Max. tolerance during standstill = 670 100 Please note, important!

    zzz saved idle value of sensor (from last imbalance calibration)

    with lid opened

    5: Imbalance monitoring

    with lid closed

    xxx Imbalance value (current sensor idle value during standstill)

    Recommended service setting = 670 30 Please note, important!Max. tolerance during standstill = 670 100 Please note, important!

    xxx imbalance value during run = current sensor value (cannot be evaluated during a run)

    yyy maximum appeared imbalance (only active during run)

    zzz imbalance cut-off value for the rotor used

    3 : x x x * y y y z z z

    4 : x x x y z z

    5 : x x x r u = z z z

    5 : x x x y y y z z z

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    20/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-20 Error lists/Troubleshooting

    6: Lid locking

    uuu mean value of sensor lid locking position

    x1,x2 lid report left and right

    yy O = locking mechanism in upper dead centre=closed

    U = locking mechanism in lower dead centre=open

    = undefined position

    zzz actual sensor value (Hall sensor junction plate)

    7: Last error report

    ww last error

    xx error report before last etc.

    8: Version

    xxx voltage version set

    yy software version

    zzzzz checksum EPROM (for factory settings only)

    9: Current error indication (selectable via cursor key only directly after error)

    xx error numberyy no. for locating the cause (for factory settings only)

    zz FC error number, if the frequency converter switches off with error 6

    0 =no error message

    1 =undervoltage

    2 =overvoltage

    4 =overtemperature FC

    5 =overtemperature motor

    7 =overcurrent FC

    8 =interference EEROM FC

    10 =no release

    FA =reset by internal software and hardware

    FF =serial interface timeout

    6 : u u u x1 x2 y z z z

    7 : w w x x y y z z

    8 : x x x y y z z z z

    9 : E r x x y y z z

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    21/36

    Error lists/Troubleshooting 3-21

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Service programs 58XX

    General description

    The service programs run independently from the normal operating modes.

    You enter the program selection by simultaneously pressing the PLUS-/MINUS key and switching the

    centrifuge on.

    Display:

    xx Service program number

    The required program number can be set by actuating the PLUS or MINUS key.

    You can open the lid with the OPEN button.

    All programs are started with the START button.

    Indicated errors can be erased with the START button.

    If "error 36" is displayed immediately after calling up the service programs the EEPROM must be

    initialized (see section Initializing).

    The service mode can only be quit by switching off the device.

    S E x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    22/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-22 Error lists/Troubleshooting

    Program overview

    Program no. Program contents

    1

    2

    3

    4

    5

    6

    7

    8

    9

    10

    11

    12

    13

    14

    15

    16

    17

    18

    Setting the serial address

    Setting the voltage version

    Checking the display and the keypad

    Testing the lid function / lid switches

    Initializing

    Test refrigeration aggregate / fan / temperature sensor

    Cycle test

    Rotor recognition

    Test frequency converter

    Calibration imbalance sensor

    Determination of the acceleration and braking times

    Overspeed cut-off test

    Viewing the error table

    Display of the number of starts / the number of runs / load-cycle display

    Display of the switching off values for imbalance / sensor idle value / powersupply

    Deleting of the operating hours / error table

    (factory only)

    (factory only)

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    23/36

    Error lists/Troubleshooting 3-23

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Description of the individual service programs

    Setting of the serial interface

    START Program 1 (optional)

    xxxxx = RS232

    BBUS Selection by means of the PLUS button

    disable

    Factory setting: disable

    Start next step

    xx device address 20...29 (Hex)

    Factory setting: 20 H

    By actuating the PLUS button the address to be set is continuously counted upwards.

    Acknowledgement of the selected device address with START.

    END end of program actuate the START button.

    Voltage version (set after replacing PCB Control)

    START Program 2

    Start SE 2 Voltage version is displayed

    Button PLUS Selection of the voltage version 230 115 100

    Button PROG Value is stored in the EEPROM

    (display 'ok')

    Button START/STOP End of program

    Check of the display and keypad

    START Program 3

    1. Check of the display

    1.5-s sound on and all digits displaying the pattern 0 0 0 0

    2. Check of the keypad

    Input request '>' the button designation is displayed as text.

    End of program by means of the STOP button (as a last button).

    1 x x x x x

    2. n x x

    S P x x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    24/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-24 Error lists/Troubleshooting

    Lid function test

    START Program 4

    l, r left and right limit switch (1 = pressed / 0 = open)

    ss position OT in upper dead centre

    UT in lower dead centre

    un undefined intermediate position

    xxx current sensor value

    yyy sensor idle value (mean value)

    Keys:

    OPEN lid open/closed normal function

    PLUS pulse by pulse forwards

    MINUS pulse by pulse backwards

    SPEED erase counter for continuous lid latch test.

    STOP end of program

    Initializing

    START Program 5

    STOP button no initializing, return to program selection

    PROG button initializing is carried out

    The nominal values set and the operating conditions are stored in the EEPROM after each run of the

    centrifuge. The limits for switching off (imbalance) and statistical data and fixed programs are also

    stored in the EEPROM.

    When calling up the program via the serial interface code no. 64, additionally the interface is set to

    20 H/disable.

    y y y l r s s x x x

    I n i t 2

    The following is deleted: The following is not deleted:

    Error table Imbalance cut-off values (SE 10)

    Number of connections (IO) Number of load changes per rotor (Clyy)

    Operating hours (h) Voltage version

    Total number of centrifuge runs (CL) Interface setting

    Programmed customer fixed programs

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    25/36

    Error lists/Troubleshooting 3-25

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Switching the aggregates / temperature sensor test (only for refrigerated devices)

    START Program 6

    Switching the aggregates

    Key:

    TEMP refrigerating aggregate on/off

    PROG motor fan on/off

    SPEED frequency converter on/off (no control of FC, only relay on PCB audible)

    TIME Condenser fan on/off

    xx fan step (0...16)

    the step can be changed by means of the PLUS and MINUS buttons

    Fast coolswitching over to temperature display

    xx.x temperature of sensor in rotor chamber

    yy condenser temperature

    room temperature when centrifuge in standby

    Procedure for cooling problems:

    Does the refrigerating aggregate switch? Does the condenser fan start? Fan level 016 Does the sensor in rotor chamber work? (Check with thumb, replace if necessary) Is the condenser temperature sensor working?

    Check refrigerating system: 5804 R, Rotor A4-44, 5000 rpm

    5810 R, Rotor A4-62, 4000 rpm

    Fill 50-ml tubes with "ethylene glycol".

    Start program, select temperature +6 C, "Fast Temp"

    Start run

    measure after 45 min with 5804 R approx. +4 C to +8

    C5810 R approx. +4 C to +8 C

    Observe customer service program 1

    xx = condenser temperature approx. +40 C to +50 C (refrigerating aggregate switches off

    at +75 C)

    * = refrigerating aggregate is switched on

    yyy = actual temperature of sensor in rotor chamber

    zzz = correction nominal value (not important for service)

    K A L M F U L K S = x x

    T e m p = x x x T K = y y

    1 : x x * y y y z z z

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    26/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-26 Error lists/Troubleshooting

    Is the refrigerating system tight? Does the valve switch? Repair if necessary and refill cooling agent

    Quantity cooling agent R134A

    5804 R 400 g

    5810 R 470 g

    Repairing the refrigerating system

    The spare parts are mounted at your own risk!

    Repair must be carried out by a specialist for refrigerating devices!

    Observe safety regulations of your country!

    Cycle program

    START Program 7

    This program is used for a continuous test (e.g. factory test) of the centrifuge. Always 3 runs with a

    subsequent interval of 1 min are carried out. After the 4th run the interval is 10 min. The FC is switchedoff after each run.

    The nominal values of the run can be changed after the start of the service program (pre-setting:

    nominal rotational speed of the rotor, 10 min, 4 C).

    Display of the pre-set nominal values, it is possible to change them

    Start cycle run

    Display after each run

    xx interval time in min/s

    An individual run can be aborted by means of the STOP button. The lid can be opened during the

    interval.

    You can only quit the cycle program by switching off the device.

    4 0 4 0 0 0 1 0

    P A U S E x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    27/36

    Error lists/Troubleshooting 3-27

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Rotor recognition test / rotor pulses

    START Program 8

    Slowly rotate the rotor < 100 1/min

    SPEED pulse display

    _ gap passes the sensor

    magnet passes the sensor

    STOP end of pulse display

    Turn rotor slowly < 100 rpm

    TIME display of rotor relation

    xxx rotor relation

    y number of measurements (2 y < 10)

    zz rotor number 1...20

    STOP ends subprogram and SE 8

    Frequency converter test (FC)

    When the upper part of the housing has been taken off, the check must be done without the

    rotor!

    START Program 9

    Check voltage version before entering FC parameters (differing data records)!

    Actuation SPEED buttonswitching FC on/off

    PLUS button Run at approx. 120 rpm, lid opened, FC must be switched on.

    MINUS buttonbraking

    b converter ready, from switching on the FC to READY indication approx. 1 s will pass

    xxx FC on/off

    yyy rotational field speed, if FC is switched on

    End = STOP button; the FC is switched off, return to program selection

    r o t a t e

    _ _ _ _ _ _ _ _ _ _ _

    r v = x x x y z z

    F U b x x x y y y

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    28/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-28 Error lists/Troubleshooting

    Imbalance calibration carried out by a service technician

    START Program 10

    The centrifuge determines the imbalance cut-off value of the individual rotors from all rotors in a rotorgroup.

    Oneangled rotor (W)and oneswing-bucket rotor (A)must be used from each rotor group. As a

    rule, this should be carried out with the standard rotors.

    Exception: Use (A) Rotor A-4-81 if avaliable.

    Standard rotors: Centrifuge 5804 / 5804R (W) F34-6-38 and (A) Rotor A-4-44

    Standard rotors: Centrifuge 5810 / 5810R (W) F34-6-38 and (A) Rotor A-4-62

    Prepare the weights: G1/G2/G3

    It is possible to weigh the weights yourself using Eppendorf tubes. Filled 2.0 ml tubes (approx. 3.2 g) are

    recommended. The weight of the tube must also be taken into consideration.

    With swing-bucket rotors, these should be placed flat, loose and without an adapter on the bottom of

    the swing-bucket. With the angled rotor F34-6-38, water should be pipetted directly into the empty

    bores. MTP rotors can be loaded and then filled with one MTP each.

    Procedure for imbalance calibration

    1. Insert fixed-angle rotor(W). (Use the present rotor lid).

    2. Call up the service program: Hold down both cursor keys and switching on the device.

    Display : SE 1

    3. Using the cursor key, call up SE 10

    4. Insert the tube with G1.

    5. Close lid. Start run: Start

    6. Three calibration runs are performed automatically. The lid opens automatically.

    7. Insert weight G2.

    8. Close the lid and wait until three further calibration runs have been performed.

    Lid opens automatically.

    9. Insert weight G3. Close lid. Start run: Start (Centrifuge must show "Imbal" and switch off.)

    10. Open lid: Open.

    11. Remove fixed-angle rotor (W) and insert swing-bucket rotor (A).

    12. Using the corresponding imbalance amounts G1, G2, G3, carry out the procedure described in

    points 2. to 9

    13. Switch off the device. Remove the imbalance weights and then switch on the centrifuge again as

    normal.

    Note:

    An imbalance adjustment must be carried out every time the motor, PCB Control or PCB Imbalance is

    replaced or when the rotor is inserted!

    After the motor has been replacedl or when the centrifuge has not been used for a long period, it should

    run for approximately five minutes at maximum speed in order to ensure

    that the vibration mounts have the normal degree of hardness.

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    29/36

    Error lists/Troubleshooting 3-29

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Overview of rotors required and weights: G1/G2/G3

    Fixed-angle rotors (W)

    Centrifuges: 5804 / 5804 R / 5810 / 5810 R

    Swing-bucket rotors (A)

    Centrifuges: 5804

    Centrifuges: 5804 R

    Centrifuges: 5810 / 5810 R

    Rotor designation G1 G2 G3

    W F-34-6-38 6g 10g 12g

    W F-45-30-11 6g 10g 15g

    W FA-45-30-11 6g 10g 15g

    W T-60-11 7.5g 15g 22.5g

    W F-45-48-PCR Without PCR stripsa PCR strip filled

    with waterNot required

    Rotor designation G1 G2 G3

    A A-4-44 16g 22g 25g

    A A-2-DWP 16g 22g 25g

    A A-2-MTP 16g 22g 25g

    Rotor designation G1 G2 G3

    A A-4-44 12g 18g 25g

    A A-2-DWP 12g 18g 25g

    A A-2-MTP 12g 18g 25g

    Rotor designation G1 G2 G3

    A A-4-62 16g 22g 30g

    A A-4-44 16g 22g 30g

    A A-2-DWP 16g 22g 30g

    A A-4-62 MTP 16g 22g 30g

    A A-2-MTP 16g 22g 30g

    A A-4-81 16g 22g 30g

    A A-4-81 MTP 16g 22g 30g

    A A-4-81 Flex 16g 22g 30g

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    30/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-30 Error lists/Troubleshooting

    Determination of the acceleration and braking times (for factory checks)

    START Program 11

    completely equip the rotor(the measured rotational speed is calculated by the program after the rotor recognition = 97 % of

    the max. rotational speed of the rotor in use)

    close the lid

    START run

    Display of the acceleration time in seconds (at 97 % of the rotational speed nominal value)

    STOP start of the measurement of the braking time

    Display braking time

    STOP return to the program selection

    Overspeed cut-off test (Test with Rotor)

    START Program 12

    Close the lid

    The software sets a rotational speed above the switching-off limit. If the device operates correctly, itswitches itself off when reaching the switching-off rotational speed and stops without braking.

    If the device is not switched off, the run must be stopped immediately by hand!

    The device must be switched off!

    On switching on again the centrifuge brakes and is then again in the normal operational mode.

    Overspeed cut-off:

    The overspeed cut-off is about 300 1/min above the nominal speed of the rotor in use.

    3 0 0 2 0 0 0 1 1

    H L x x x

    B L x x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    31/36

    Error lists/Troubleshooting 3-31

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Viewing the error table

    START Program 13

    The error that occurred first during a run is stored.display of the errors during a run

    PROG transition to the step mode

    START continuing step by step

    STOP return to the program selection

    xx error no.

    yyy no. of errors

    Display for the number of starts / number of runs

    valid for centrifuges incl. software version 2.1

    START Program 14

    xxxx no. of starts

    START next display

    xxxx no. of runs

    STOP return to program selection

    E r r o r x x y y y

    E A x x x x

    C L x x x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    32/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-32 Error lists/Troubleshooting

    valid for centrifuges with software from version 3.0 onwards

    The load cycles are recorded separately for each rotor type.

    STARTProgram 14

    xxx no. of starts

    START next display

    xxxx no. of runs completely

    Key PROG erases the number

    START next display

    yy rotor number CL

    xxxx no. of load cycles per rotor type

    START next rotor number CL

    Key PROG erases the number n for this rotor only

    Cursor keyschange the number of load cycles in steps of 500

    START next display

    xxxx no. of operating hours

    START end of program

    E A x x x x

    C L x x x x x x

    C L y y x x x x x x

    h x x x x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    33/36

    Error lists/Troubleshooting 3-33

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Viewing the values stored in the EEPROM (for factory checks)

    START program 15

    yy rotor number

    xxx switching-off values for all rotors (no. 2, 3, 4, 5, 8, 13, 14, 17, 18)

    always continue switching with START

    START

    xxx idle value imbalance sensor (measurement during the imbalance calibration)

    Start

    xxx max. imbalance value for 30g-run G3 with rotor A-4-81 (from software 3.4 onwards)

    Start

    xx rotor number calibration rotor fixed-angle rotor

    Start

    xx rotor number calibration rotor swing-bucket rotor

    Start

    xx idle value position sensor locking mechanism for lid

    Start

    xxx voltage version

    U y y x x x

    U r x x x

    U 3 x x x

    R W x x

    R A x x

    D V x x

    S P x x x

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    34/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-34 Error lists/Troubleshooting

    Start

    XX P0, P1, PL

    test words (check of the test process)

    STOP return to the program selection

    Deleting the error table

    START program 16

    PROG error table and the operating hours are being erased

    acknowledgement display ok

    STOP abortion without deleting

    X X y y

    E r a s e E r r o r

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    35/36

    Error lists/Troubleshooting 3-35

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    Rotor table, Rotor relations

    *1

    Usable from software version 3.0 onwards

    *2

    From software version 3.3 onwards allowed up to 3700 rpm as well for older rotors labeled 3600 rpm

    *3

    Can be used from software version 3.5 onwards

    *4

    Usable from software version 3.7 onwards

    *5

    Centrifuge 5804 / 5804 R / 5810 only 11000 rpm

    Rotor num-

    ber

    CL

    max. speed/rcf

    max. radiusRotor designation Rotor relation

    1

    23700 / 2250

    147 mmA-2-DWP *

    1*

    2 153.9143...163

    34000 / 2253

    126 mmA-2-MTP

    192.3182...202

    44000 / 3220

    180 mmA-4-62

    (rectangular bucket)230.7

    220...240

    44000 / 2750

    154 mmA-4-62

    (MTP carrier)230.7

    220...240

    54000 / 3220

    180 mmA-4-81 *

    3

    (rectangular buckets)269.2

    255...283

    54000 / 2915

    163 mmA-4-81 *

    3

    (MTP carrier)269.2

    255...283

    54000 / 2900

    153 mmA-4-81 *

    3

    (Flex bucket)269.2

    255...283

    6

    7

    85000 / 4500

    161 mmA-4-44

    384.6375...395

    9

    1011

    12

    1312000 / 15294

    95 mmF-45-48-PCR *

    4 577563...591

    1412000 *

    5/ 18514

    115 mmF-34-6-38 *

    5 615.4605...625

    15

    16

    1714000 / 20817

    95 mmF-45-30-11

    FA-45-30-11730.7

    721...741

    1814000 / 16435

    75 mmT-60-11 *

    1 769.2759...779

    19

    20

    MSTM/0408

  • 5/21/2018 E5810R_kap03

    36/36

    Centrifuge 5804/5804 R and 5810/5810 R

    3 Error lists / Troubleshooting

    3-36 Error lists/Troubleshooting MSTM/0408