mios_jisl
TRANSCRIPT
-
8/7/2019 MIOS_JISL
1/45
Minolta Innovative Optical SystemMinolta Innovative Optical System
TechnologyTechnology
-
8/7/2019 MIOS_JISL
2/45
Contents
Numerical Gloss Control
Numerical UV Control
Minolta Innovative Optical SystemMinolta Innovative Optical System
-
8/7/2019 MIOS_JISL
3/45
Objects that composed of the same materials, may have
variances in colors due to differences in the gloss of the
surfaces.
For example, when the surface condition of a shiny or
high gloss sample is being changed using a sandpaper,
it will cause the sample to look duller in color.
Color & GlossColor & Gloss
Before After
-
8/7/2019 MIOS_JISL
4/45
SpecularComponent IncludedSpecularComponent Included
SC
I The total amount of reflected light including the totalamount of specular component is taken into accountfor the measurement. The total amount of light is thesame and therefore the reflectance data of matte and
glossy surface is the same. It measures the total colour appearance independent
of the surface conditions.
When to useWhen to use:
Recommended for most CCM measurements Can be selected for certain QC measurements
-
8/7/2019 MIOS_JISL
5/45
SpecularComponent ExcludedSpecularComponent Excluded
SCE The specular component is excluded. Only the diffused
reflected light is taken into account for the measurement
and the reflectance data of the matte and glossy surface is
different. A matte surface is brighterbut less saturated
and a glossy surface will show a darkerbut highersaturation.
These results are close to visual perception
When to useWhen to use:
Recommended for most QC measurements Can be selected for certain CCM measurements too,
provided the surface is smooth and flat
If the surface is not smooth and flat, some light rays mayreflect in the directional reflection path which is not
wanted.
-
8/7/2019 MIOS_JISL
6/45
Traditional SCI/SCE Switching
Numerical Gloss Control Method
Numerical Gloss Control Advantage
Traditional Method vs Numerical Gloss Control
Numerical Gloss ControlNumerical Gloss Control
Contents
-
8/7/2019 MIOS_JISL
7/45
Traditional MethodTraditional Method
Specular Component Included and Excluded (SCI/SCE)
Physical gloss trap is opened and closed
mechanically
Switching time is necessary (much longer thanmeasuring time)
Moving parts are a potential danger ofbreak-down
Simultaneous measurements (completely same
conditions of sample) are impossible.
-
8/7/2019 MIOS_JISL
8/45
SCE SCI
Gloss Trap Closed
Sample
TraditionalTraditional SCISCI/SCE Measurement/SCE Measurement
-
8/7/2019 MIOS_JISL
9/45
SCE SCI
Traditional SCI/Traditional SCI/SCESCE MeasurementMeasurement
Gloss Trap Opened
Sample
-
8/7/2019 MIOS_JISL
10/45
SCE SCI
Gloss Trap Closed
Sample
TraditionalTraditional SCISCI/SCE Measurement/SCE Measurement
-
8/7/2019 MIOS_JISL
11/45
SCE SCI
Gloss Trap Opened
Sample
Traditional SCI/Traditional SCI/SCESCE MeasurementMeasurement
-
8/7/2019 MIOS_JISL
12/45
Illumination
Two flashes are activated
Diffuse Directional
Numerical Gloss ControlNumerical Gloss Control
-
8/7/2019 MIOS_JISL
13/45
First illumination - diffused
Sample
Numerical Gloss ControlNumerical Gloss Control
-
8/7/2019 MIOS_JISL
14/45
Sample
Numerical Gloss ControlNumerical Gloss Control
Second illumination - directional
-
8/7/2019 MIOS_JISL
15/45
Illumination
Two flashes are activated
Diffuse Directional
Calculation
With the data obtained by diffused illumination reflection and
directional illumination reflection, the CPU calculate the SCI
and SCE data simultaneously
Numerical Gloss ControlNumerical Gloss Control
-
8/7/2019 MIOS_JISL
16/45
Diffused light
deflection
Sample
Numerical Gloss ControlNumerical Gloss Control
-
8/7/2019 MIOS_JISL
17/45
Directional light
reflection
Sample
Numerical Gloss ControlNumerical Gloss Control
-
8/7/2019 MIOS_JISL
18/45
Numerical Gloss ControlNumerical Gloss Control
NGC calculation
SCE
Reflection due to diffuseReflection due to diffuse
illuminationillumination
SCI
Reflection by directionalReflection by directional
illuminationillumination
SCI
SCE
Simultaneously
-
8/7/2019 MIOS_JISL
19/45
Exclusive and patented
Practical and time saving
fast sampling - SCI & SCE in 4 seconds
Fully reliable no mechanism
Numerical Gloss ControlNumerical Gloss Control
Contents
-
8/7/2019 MIOS_JISL
20/45
Traditional Method Numerical Gloss Control
Switching between SCI and SCEtakes longer time
Quick measurement (SCI and SCEtakes 4 seconds)
Completely same conditions ofsample for SCI/SCE measurementis impossible
Completely same conditions ofsample for SCI/SCE measurement
Moving parts are subjected tobreakdown
No moving part
Numerical Gloss ControlNumerical Gloss Control
Contents
-
8/7/2019 MIOS_JISL
21/45
BLACKLIGHT
A blacklight is an illumination using wavelengths mostly
outside the visible regions of the spectrum. Example of
such light is the UV lamp.
It emits energy in the ultraviolet region
This light is use to illuminate fluorescent samples like
fluorescent jigsaw puzzle, while shirts or socks, to
cause them to appear striking or glow even when theyare in a dark room.
Fluorescent Material & WhitenessFluorescent Material & Whiteness
-
8/7/2019 MIOS_JISL
22/45
FLUORESCENT MATERIAL
Fluorescent material are chemicals that absorbs the
energy emitted by the backlight and re-emits it as light
in the visible region.
The sample appears white when it reflects all
wavelengths in the visible regions at nearly 100%.
When the reflectance is less in the blue wavelength,
the sample will appear to be yellowish. Fluorescent materials, sometimes referred to as
Optical Brighteners, are added to the samples
Fluorescent Material & WhitenessFluorescent Material & Whiteness
-
8/7/2019 MIOS_JISL
23/45
As fluorescent materials are used in many products like
paper and white shirts, whiteness measurement
becomes a must in these industries to monitor the
quality of the product as well as to ensure unnecessary
lost.
Correct measurement of whiteness demands an
instrument equipped with an adjustment device
capable of regulating the UV intensity.
The instrument is regarded as calibrated, if the amountof UV falling onto the sample is similar to that
encountered in daylight.
Whiteness MeasurementWhiteness Measurement
-
8/7/2019 MIOS_JISL
24/45
Traditional UV Adjustment Method
Numerical UV Adjustment Method
Numerical UV Control Merit
Traditional Method vs. Numerical UV Control
Anti-Triplet Effect
Numerical UV ControlNumerical UV Control
Contents
-
8/7/2019 MIOS_JISL
25/45
UV filter moves up and down to change UV content
of illumination to meet the value of fluorescent
calibration standard
Total energy is adjusted for one colorimetric index,
e.g., CIE WI with D65.
The procedure is based on trial and error to find the
exact filter position
Tedious
Traditional MethodTraditional Method
-
8/7/2019 MIOS_JISL
26/45
UV Cut Filter
100% UV
Fluorescent standard
Reflection
Calibration value
Illumination
Traditional UV CalibrationTraditional UV Calibration
-
8/7/2019 MIOS_JISL
27/45
Reflection
Calibration value
Illumination
Fluorescent Standard
UV Cut
Filter
Partial UV
UV Cut Filter
Traditional UV CalibrationTraditional UV Calibration
-
8/7/2019 MIOS_JISL
28/45
Traditional UV CalibrationTraditional UV Calibration
UV Cut Filter
0% UV
Fluorescent Standard
Reflection
Calibration value
Illumination
-
8/7/2019 MIOS_JISL
29/45
Maybe calibrated
UV Cut Filter
Adjusted UV
Fluorescent Standard
Traditional UV CalibrationTraditional UV Calibration
Reflection
Calibration value
Illumination
-
8/7/2019 MIOS_JISL
30/45
Two flashes
First: With 100% UV radiation
Second: 0% UV radiation
Numerical UV ControlNumerical UV Control
-
8/7/2019 MIOS_JISL
31/45
400 nm UV Cut Filter
Fluorescent standard
Numerical UV CalibrationNumerical UV Calibration
First illumination with UV
-
8/7/2019 MIOS_JISL
32/45
400 nm UV Cut Filter
Fluorescent standard
Numerical UV CalibrationNumerical UV Calibration
First illumination without UV
-
8/7/2019 MIOS_JISL
33/45
Two illumination
First: With 100% UV radiation
Second: 0% UV radiation)
Calculation
With the data obtained by 100% UV and 0% UV, the CPU
calculates the ratio to meet the calibration data in each
wavelength.
Numerical UV ControlNumerical UV Control
-
8/7/2019 MIOS_JISL
34/45
ReflectionIllumination
Numerical UV CalibrationNumerical UV Calibration
First illumination @ 100% UV
-
8/7/2019 MIOS_JISL
35/45
Numerical UV CalibrationNumerical UV Calibration
IlluminationSecond illumination @ 0% UV
ReflectionIllumination
-
8/7/2019 MIOS_JISL
36/45
Reflection
100% UV
Calibration Value
0% UV
a
b
Numerical UV CalibrationNumerical UV Calibration
-
8/7/2019 MIOS_JISL
37/45
For measurement, there are 2 illumination
The ratios obtained by 100% UV and 0% UV in
calibration, are adapted to the data obtained by
measurement to calculate the UV adjustedmeasurement value.
UV adjustment can be done in various ways:-
Whiteness Index
Whiteness Index & Tint
Profile
Granz & Griesser
Numerical UV MeasurementNumerical UV Measurement
-
8/7/2019 MIOS_JISL
38/45
Reflection100% UV
0% UV
Adjusted Measurement
valuea
b
Numerical UV MeasurementNumerical UV Measurement
-
8/7/2019 MIOS_JISL
39/45
Exclusive and patented
Practical and time saving
Total UV measurement process drastically simplified
Multiple information for UV analysis by as singlemeasurement
Parameter setting without hardware manipulation
No human error
Reliable No mechanical parts
Numerical UV Calibration MeritNumerical UV Calibration Merit
Return
-
8/7/2019 MIOS_JISL
40/45
Tr iti l t ri l C tr l
Calibrati tak s l r
ti
Instant calibrati n
Fl r sc nt sampl as tb pos b manyill mination ringcalibration
Only t o ill mination forcalibration
Mec anical parts s bject towearand tear omec anical parts
imitedparameters witfrequent manipulation
Manyparameters/42nm cutoff, ardand
soft illumination
ri l C tr lri l C tr l
R t r
-
8/7/2019 MIOS_JISL
41/45
AntiAnti--Triplet EffectTriplet Effect
Triplet Effect
Certain fluorescent materials has triplet effect, this
will cause a dip of the spectral graph just after the
fluorescent region as shown in the graph. Due to a high intensity illumination by the xenon flash
on the sample for a very short time.
Will not happen in normal illumination condition
-
8/7/2019 MIOS_JISL
42/45
Intensive Flash
AntiAnti--Triplet EffectTriplet Effect
-
8/7/2019 MIOS_JISL
43/45
Soft Flash
AntiAnti--Triplet EffectTriplet Effect
-
8/7/2019 MIOS_JISL
44/45
AntiAnti--Triplet EffectTriplet Effect
Soft Flash
Triplet EffectEffect
Return
Intensive Flash
-
8/7/2019 MIOS_JISL
45/45
Return
Manish Kapadia
Chief manager
JISL